Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
<strong>octet</strong> System Choices<br />
Analyte MW<br />
> 150 D<br />
Analyte MW<br />
> 5,000 D<br />
SENSITIVITY<br />
Label-Free Kinetic Characterization<br />
all <strong>octet</strong> instruments provide sample temperature control from 4 °c above ambient up to 40 °c in 1 °c increments. the kinetic constant measurement<br />
ranges are system- and assay-dependent. a typical range for each system is shown below.<br />
<strong>octet</strong> SYSteMS Kinetic conStant rangeS<br />
Application Specification Octet RED384 Octet QK384 Octet RED96 Octet QKe<br />
Protein-Protein Kinetics ka, M -1 s -1 10 2 – 10 7 10 3 – 10 7 10 2 – 10 7 10 3 – 10 7<br />
kd, s -1 10 -6 – 10 -1 10 -6 – 10 -1 10 -6 – 10 -1 10 -6 – 10 -1<br />
KD, M 1 mM – 10 pM 0.1 mM – 10 pM 1 mM – 10 pM 0.1 mM – 10 pM<br />
Protein-Small Molecule Kinetics ka, M -1 s -1 10 2 – 10 7 n/a 10 2 – 10 7 n/a<br />
Data acquisition rate options Data points per<br />
second (Hz)<br />
Octet QK e<br />
Octet RED96<br />
• 96-well plates<br />
• 1 sample plate position<br />
• 8-channel readout<br />
• Biosensor re-racking<br />
THROUGHPUT<br />
Octet QK384<br />
• 96- and 384-well plates<br />
• 2 sample plate positions<br />
• 16-channel readout<br />
• Automated plate handling<br />
• Biosensor re-racking<br />
• ≥40 µL/well (384TW microplate)*<br />
kd, s -1 10 -6 – 10 -1 n/a 10 -6 – 10 -1 n/a<br />
Octet RED384<br />
*The 384TW microplate is a black, polypropylene, 384-well tilted<br />
bottom plate designed for use on the Octet RED384 and Octet<br />
QK384 instruments. The 384TW microplate enables use of sample<br />
volumes as low as 40 µL, reduces background signal variability and<br />
improves assay sensitivity.<br />
KD, M 1 mM – 10 pM n/a 1 mM – 10 pM n/a<br />
2, 5 and 10 Hz 0.3 and 0.6 Hz 2, 5 and 10 Hz 0.3 and 0.6 Hz