This quartz degradation is accelerated by alkaline impurities present in the<br />

catalyst. Even very low amounts of these impurities reduce the live-time of the tube<br />

drastically. This adverse reaction was not detected during the analytical runs of the<br />

new product.<br />

The good quality of the catalyst was evaluated with a large number of<br />

analyses. The chrome oxide region of the quartz tube was carefully observed and<br />

compared to the degradation and the internal thickness of the silicate deposits with<br />

different catalysts present on the consumable market.<br />

The tests were performed on the New Chrome Oxide after analyses of<br />

acetanilide samples. After 40, 150 and 250 runs the quartz tube was removed from<br />

the furnace. Visual aspects and glass porosity were observed and documented. In<br />

the following figures photos of the same quartz tube, after 40, 150 and 250 tests, are<br />

reported.<br />

Fig. 2: Quartz tube with new Cr2O3 after 40 runs<br />

Fig. 3: Quartz tube with new Cr2O3 after 150 runs