Service Manual Analyzer for Combi Screen® 11 SYS urine ... - Alere
Service Manual Analyzer for Combi Screen® 11 SYS urine ... - Alere
Service Manual Analyzer for Combi Screen® 11 SYS urine ... - Alere
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
<strong>Service</strong> <strong>Manual</strong> <strong>Combi</strong>Scan 100<br />
6<br />
Press button (mostright) under<br />
„Print”. The stored worklist is<br />
printed. The display will change back<br />
to the main menu automatically.<br />
Yes<br />
Press button under the<br />
„ChkMeas” (middle). “Check mode<br />
measurement” is diplayd and check<br />
measurement is per<strong>for</strong>med.<br />
The device prints out the result of<br />
the check measurement. „Valid<br />
check” or „Invalid check”.<br />
Then returns to the main menu<br />
automatically.<br />
Yes<br />
Press the ”START” button. The<br />
display goes back to the main menu.<br />
Yes<br />
Press button under „Menu”<br />
(rightmost). Further MENU items<br />
are displayed (Setup, MeasMode,<br />
Memory).<br />
Yes<br />
Main units of the device are<br />
operating proper. Different<br />
parameters should be checked<br />
or set (i.e. date, time, language)<br />
according to the user‘s manual.<br />
Yes<br />
Visual test: visible damages of the<br />
upper housing, printer cover, plastic<br />
push buttons, plexi plate.<br />
No<br />
No<br />
No<br />
No<br />
No<br />
No<br />
No<br />
No<br />
Push button defect<br />
SW error<br />
Error messages<br />
(E7)<br />
Push button defect<br />
SW error<br />
SW error<br />
Error messages<br />
SW error<br />
No Problem unsolvable by the<br />
service<br />
No Replacement of the<br />
assembled upper housing.<br />
See section 5.<br />
Repair of the push button<br />
defect, section 3.6.<br />
SW error handling,<br />
section 3.7.<br />
Error message description,<br />
section 5.<br />
Repair of the push button<br />
defect, section 3.6.<br />
SW error SW error handling,<br />
section 3.7.<br />
SW error handling,<br />
section 3.7.<br />
Error message description,<br />
section 5.<br />
SW error handling,<br />
section 3.7.<br />
The equipment must be<br />
returned <strong>for</strong> repair.