<strong>Service</strong> <strong>Manual</strong> <strong>Combi</strong>Scan 100<br />

6<br />

Press button (mostright) under<br />

„Print”. The stored worklist is<br />

printed. The display will change back<br />

to the main menu automatically.<br />

Yes<br />

Press button under the<br />

„ChkMeas” (middle). “Check mode<br />

measurement” is diplayd and check<br />

measurement is per<strong>for</strong>med.<br />

The device prints out the result of<br />

the check measurement. „Valid<br />

check” or „Invalid check”.<br />

Then returns to the main menu<br />

automatically.<br />

Yes<br />

Press the ”START” button. The<br />

display goes back to the main menu.<br />

Yes<br />

Press button under „Menu”<br />

(rightmost). Further MENU items<br />

are displayed (Setup, MeasMode,<br />

Memory).<br />

Yes<br />

Main units of the device are<br />

operating proper. Different<br />

parameters should be checked<br />

or set (i.e. date, time, language)<br />

according to the user‘s manual.<br />

Yes<br />

Visual test: visible damages of the<br />

upper housing, printer cover, plastic<br />

push buttons, plexi plate.<br />

No<br />

No<br />

No<br />

No<br />

No<br />

No<br />

No<br />

No<br />

Push button defect<br />

SW error<br />

Error messages<br />

(E7)<br />

Push button defect<br />

SW error<br />

SW error<br />

Error messages<br />

SW error<br />

No Problem unsolvable by the<br />

service<br />

No Replacement of the<br />

assembled upper housing.<br />

See section 5.<br />

Repair of the push button<br />

defect, section 3.6.<br />

SW error handling,<br />

section 3.7.<br />

Error message description,<br />

section 5.<br />

Repair of the push button<br />

defect, section 3.6.<br />

SW error SW error handling,<br />

section 3.7.<br />

SW error handling,<br />

section 3.7.<br />

Error message description,<br />

section 5.<br />

SW error handling,<br />

section 3.7.<br />

The equipment must be<br />

returned <strong>for</strong> repair.