Collimators for Radiography and Fluoroscopy - Siemens
Collimators for Radiography and Fluoroscopy - Siemens
Collimators for Radiography and Fluoroscopy - Siemens
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
<strong>Collimators</strong> <strong>for</strong> <strong>Radiography</strong> <strong>and</strong> <strong>Fluoroscopy</strong><br />
Product Description<br />
The purpose of the multileaf collimator is to collimate<br />
exactly the radiation beam emerging from the X-ray tube<br />
assembly onto the size of the object to be displayed.<br />
Collimation is manual with the aid of the field light<br />
localizer (except A 01 II <strong>and</strong> AF 01 II) or motor-driven by<br />
automatic <strong>for</strong>mat collimation system.<br />
The current source-image distance (SID) <strong>and</strong> the size<br />
of the collimated radiation field are displayed on a LCD<br />
display except A 01 II, AF 01 II.<br />
A laser line localizer is available <strong>for</strong> centering the bucky<br />
except <strong>for</strong> type A 01 II, AF 01 II.<br />
For optimizing the image <strong>and</strong> radiation quality three<br />
different Cu-filter can be selected.<br />
Features<br />
◾ Lamellae close to the source to shield off extrafocal<br />
radiation<br />
◾ Additional prefilters integrated in the collimator <strong>for</strong><br />
manual (AL 01 II eL) or motorized operation (AL 02 II eL,<br />
A 01 II, AF 01 II, AFL 01 II eL)<br />
◾ Bucky light generated by laser results in a sharp<br />
light line <strong>for</strong> all source-image distances (not A 01 II <strong>and</strong><br />
AF 01 II)<br />
2<br />
◾ Display of field sizes optionally in cm (default) or inch<br />
where available<br />
◾ Rotation by ± 45° with a stop in 0° position by swiveling<br />
flange (not A 01 II <strong>and</strong> AF 01 II)<br />
◾ Tape measure integrated in a user friendly way in the<br />
collimator (not A 01 II <strong>and</strong> AF 01 II)<br />
◾ Two foam protected accessory rails facilitate attachment<br />
of additional devices <strong>and</strong> further filters (not A 01<br />
II <strong>and</strong> AF 01 II)<br />
Options <strong>and</strong> accessories<br />
Set of three-field templates <strong>for</strong><br />
SID 90 – 110 cm / 36 – 42‘’<br />
SID 110 – 130 cm / 42 – 50‘’<br />
SID 130 – 175 cm / 50 – 70‘’<br />
SID 175 – 220 cm / 70 – 90‘’<br />
The collimator is prepared <strong>for</strong> installation of a DAP<br />
chamber as well as a CANopen DAP chamber.<br />
Customizing<br />
In order to fulfill your particular requirements, we also<br />
provide you with customized versions of this product.