<strong>Membrane</strong> <strong>Separation</strong> <strong>Technology</strong><br />

<strong>for</strong> <strong>Research</strong> <strong>and</strong> <strong>Quality</strong> <strong>Control</strong><br />

C E R T I F I E D<br />

DIN EN ISO 9001<br />

Q U A L I T Y<br />

S Y S T E M<br />

Reg. No. 1476-02<br />

C E R T I F I E D<br />

Q U A L I T Y<br />

S Y S T E M<br />

Sartorius AG, <strong>Separation</strong> <strong>Technology</strong><br />

Laboratory Filtration


Major Applications<br />

<strong>and</strong> Product Descriptions<br />

The right filter <strong>for</strong> every application<br />

Sartorius offers a wide selection of filter types, filter holders <strong>and</strong> ready-to-use<br />

filtration units. There are certainly reasons <strong>for</strong> you to use these products if you<br />

are engaged in analysis, research, quality control or production. But which are<br />

the right ones <strong>for</strong> you?<br />

The choice naturally depends on your particular application.<br />

Are you interested in the particles, colloids, cells, microoorganisms or proteins<br />

which are retained on the filter surface?<br />

Do you need a visibly clear, particle-free, sterile or pyrogen-free filtrate?<br />

Must cells be harvested, or proteins concentrated or purified?<br />

Or do you need an adsorptive, porous support <strong>for</strong> blotting or diagnostic kits?<br />

We have listed recommended products <strong>for</strong> a number of typical applications on<br />

the next two pages, <strong>and</strong> have given brief introductions to five major<br />

application areas (listed top right) on pages 4 to 8. Should you have questions<br />

on these, or require advice on other filtration or separation needs, please<br />

contact your nearest Sartorius representative. Addresses, telephone <strong>and</strong> telefax<br />

numbers are given on the back page.<br />

We wish you every success,<br />

Sartorius <strong>Separation</strong> <strong>Technology</strong><br />

New!<br />

8<br />

55<br />

60<br />

61<br />

62<br />

63<br />

79<br />

85<br />

Sterile, single use Funnels <strong>and</strong> Monitors <strong>for</strong> quicker <strong>and</strong> simpler<br />

vacuum filtration of samples <strong>for</strong> colony counts<br />

Centrisart-C30 Micro-Concentrators <strong>for</strong> centrifugal filtration of<br />

volumes of up to 3 ml<br />

Sartocon Slice crossflow filtration system <strong>for</strong> scaleable results from<br />

1 to 100 liters<br />

Sartocon 3 hydraulic crossflow system <strong>for</strong> larger volume concentration<br />

<strong>and</strong> purification<br />

Ultrasart X. The spiral flow cell which avoids the drawbacks of<br />

stirred cells<br />

Perisys 70 Benchtop System <strong>for</strong> optimum yields with Sartocon<br />

Micro crossflow units<br />

Sartopore Capsules <strong>for</strong> faster pressure filtration <strong>and</strong> broader<br />

chemical compatibility<br />

<strong>Membrane</strong> Adsorbers <strong>for</strong> the rapid purification <strong>and</strong> concentration<br />

of biomolecules<br />

Contents<br />

Major Applications<br />

Concentration/Purification<br />

Air/Gas filtration<br />

Microbiological QC<br />

Sample cleaning<br />

Sterilizing filtration<br />

(pages 4 to 8)<br />

Product descriptions<br />

Syringe units<br />

<strong>and</strong> filter holders<br />

(pages 9 to 17)<br />

Vacuum units<br />

<strong>and</strong> filter holders<br />

(pages 18 to 29)<br />

Pressure units<br />

<strong>and</strong> filter holders<br />

(pages 30 to 48)<br />

Air filtration units<br />

<strong>and</strong> filter holders<br />

(pages 49 to 53)<br />

Centrifugal <strong>and</strong><br />

crossflow units<br />

(pages 54 to 64)<br />

<strong>Membrane</strong> filters<br />

<strong>and</strong> ultrafilters<br />

(pages 65 to 81)<br />

<strong>Membrane</strong> Adsorbers<br />

Air sampling <strong>and</strong><br />

integrity testing instruments<br />

(pages 85 to 99)<br />

Chemical Compatibility Guides<br />

(pages 82, 83, 84)<br />

1


Sterile venting units.<br />

Capsules, Holders <strong>and</strong> Housings<br />

<strong>for</strong> air filtration<br />

Air filter holders / units<br />

The Midisart unit (bottom right) <strong>and</strong> the Sartofluor Capsule (middle left) are mainly used <strong>for</strong> sterile venting. Both are<br />

ready <strong>for</strong> use, <strong>and</strong> repeatedly autoclavable <strong>for</strong> re-use. The stainless steel housing (top centre) is used <strong>for</strong> the same<br />

purpose, in combination with Sartofluor Mini Cartridges (see page 81).<br />

The stainless steel holder <strong>for</strong> 47 mm membranes (middle right) is mainly used <strong>for</strong> in-line air/gas filtration, as are the<br />

smaller 25 mm holder <strong>and</strong> the Mini Cartridge housings which are not shown here (see pages 52 <strong>and</strong> 53).<br />

The Gelatine Filter Unit (bottom left) quantitatively collects airborne microorganisms. It is described on page 88.<br />

49


Sterile venting unit<br />

Midisart ®<br />

2000 Sterile Venting Units,<br />

light in weight <strong>and</strong> easy to connect<br />

Re-usable complete filtration units with<br />

naturally hydrophobic PTFE membrane<br />

<strong>for</strong> reliable sterile venting of small<br />

fermenters <strong>and</strong> of containers <strong>for</strong> culture<br />

media.<br />

Midisart 2000 units have been<br />

designed <strong>for</strong> maximum h<strong>and</strong>ling ease<br />

<strong>and</strong> reliability. Tapered hose barbs<br />

ensure a simple, secure hold <strong>for</strong> 6 to<br />

12 mm i.d. tubing. The light weight,<br />

just 20 g, will not weigh down <strong>and</strong><br />

kink tubing. The membrane is rein<strong>for</strong>ced<br />

with polypropylene gauze <strong>for</strong> stability<br />

to pressures of up to 3 bar. The 20 cm 2<br />

large filter area allows high air flow<br />

rates at low differential pressures.<br />

Each unit is printed with lot number<br />

<strong>and</strong> individual piece number on the<br />

housing <strong>for</strong> total security <strong>and</strong> traceability.<br />

50<br />

SPECIFICATIONS <strong>for</strong> Midisart 2000 units<br />

Biosafety All materials pass USP Plastics Test Class VI<br />

Bubble point Min. values with isopropanol are 1.4 bar (140 kPa) <strong>for</strong> the<br />

0.2 µm unit (1.1 bar after autoclaving) <strong>and</strong> 0.9 bar (90 kPa) <strong>for</strong><br />

the 0.45 µm unit<br />

Connectors Choice of hose nipples <strong>for</strong> 6 to 12 mm i.d.tubing (with slip-fit <strong>for</strong><br />

luer syringes) or 1/8" male NPT<br />

Hold-up volume Approx. 0.5 ml after (1 ml be<strong>for</strong>e) bubble point<br />

Filtration area 20 cm2 Air flow rates Typical values <strong>for</strong> 0.2 µm pore sizes are:<br />

1.1 l/min at 0.02 bar (1.8 l/min <strong>for</strong> 0.45 µm),<br />

2.9 l/min at 0.05 bar (4.6 l/min <strong>for</strong> 0.45 µm),<br />

5.0 l/min at 0.1 bar (8.5 l/min <strong>for</strong> 0.45 µm)<br />

Housing 62 mm diameter<br />

Materials PTFE membrane filter, strengthened with polypropylene gauze.<br />

Polypropylene housing<br />

Limits Max. recommended operating pressure, 3 bar (300 kPa).<br />

Max. temperature, 134°C<br />

Sterilization By autoclaving at 121°C (at least 20 times) or 134°C.<br />

The E <strong>and</strong> G packs are presterilized with ethylene oxide<br />

Priming volume Approx. 3 ml<br />

Water wetting Water penetration point, 4.0 bar (0.2 µm), 3.0 bar (0.45 µm)<br />

ORDER NUMBERS <strong>for</strong> Midisart 2000 units<br />

a) with hose nipple connectors:<br />

17804 E 0.45 µm, sterile, individually packed, (pack of 12)<br />

17804 G 0.45 µm, sterile, individually packed, (pack of 25)<br />

17805 E 0.2 µm, sterile, individually packed, (pack of 12)<br />

17805 G 0.2 µm, sterile, individually packed, (pack of 25)<br />

17805 UPN 0.2 µm, bulk packed, non-sterile (pack of 100)<br />

b) with 1/8" male NPT connectors:<br />

17804 NPE 0.45 µm, sterile, individually packed, (pack of 12)<br />

17804 NPG 0.45 µm, sterile, individually packed, (pack of 25)<br />

17805 NPE 0.2 µm, sterile, individually packed, (pack of 12)<br />

17805 NPG 0.2 µm, sterile, individually packed, (pack of 25)<br />

Minisart-HY Ready-to-Connect Units <strong>for</strong> sterile air-venting bottles<br />

<strong>and</strong> small containers<br />

These 26 mm units contain a polyester-strengthened 0.2 µm PTFE membrane<br />

in a Cyrolite housing with luer lock connectors (female top, male bottom).<br />

SPECIFICATIONS <strong>for</strong> Minisart-HY<br />

Air flow rate Approx. 1.4 l/min at ∆p = 0.1 bar<br />

Bubble point Min. value with isopropanol, 1.2 bar<br />

Burst pressure Housing holds up to at least 6.0 bar<br />

Filtration area 5.3 cm 2<br />

Water wetting Penetration pressure ≥ 4.0 bar<br />

ORDER NUMBERS <strong>for</strong> Minisart-HY<br />

16596 HYK Sterile, individually packed (pack of 50)<br />

16596 HYQ Bulk packed, non-sterile (pack of 500)


Capsule filters<br />

Sartofluor ®<br />

Capsules with PTFE membrane<br />

<strong>for</strong> maximum security in sterile venting<br />

Sartofluor Capsules are ready-toconnect,<br />

pre-tested, complete filter<br />

units. The PTFE membrane is pleated<br />

to obtain the largest possible usable<br />

filtration area in the small, translucent<br />

polypropylene housing. The two vents<br />

in the upstream side of the housing<br />

ensure good steam passage <strong>and</strong><br />

corresponding certainty of sterilization<br />

of the Capsules by autoclaving.<br />

The extreme hydrophobicity of the<br />

PTFE membrane ensures maximum<br />

reliability, even when filtering moist<br />

air. The high air flow rates of the<br />

membranes <strong>and</strong> the large filtration<br />

area allow effective sterile venting at<br />

low differential pressures.<br />

Differential pressure (mbar)<br />

300<br />

250<br />

200<br />

150<br />

100<br />

50<br />

0.05 m 2<br />

0.1 m 2<br />

0.2 m 2<br />

0<br />

0 10 20 30 40 50 60<br />

Air flow rate (Nm3 /h)<br />

The excellent chemical compatibility of<br />

the PTFE <strong>and</strong> polypropylene materials<br />

makes Sartofluor Capsules additionally<br />

useful <strong>for</strong> the filtration of those<br />

acids, bases <strong>and</strong> non-aqueous<br />

solvents <strong>for</strong> which other Capsule types<br />

cannot be used.<br />

SPECIFICATIONS <strong>for</strong> Sartofluor Capsules<br />

Biological safety All materials pass USP Plastics Test Class VI<br />

Bubble point Minimum values, wetted with 60% isopropanol,<br />

1.5 bar (150 kPa) <strong>for</strong> 0.1 µm pore size,<br />

1.0 bar (100 kPa) <strong>for</strong> 0.2 µm pore size,<br />

0.6 bar (60 kPa) <strong>for</strong> 0.45 µm pore size<br />

Chemical compatibility See page 84<br />

Connectors 10 mm hose nipples or 1"-11/2" Sanitary flanges<br />

Differential pressure Max. 4 bar at 20°C, 2 bar at 80°C<br />

Operating pressure Max. 4 bar at 20°C<br />

Filtration area 0.05 m 2 , 0.1 m 2 , or 0.2 m 2<br />

Flow rates For 0.2 µm Capsules, see the diagram bottom left<br />

Materials PTFE membrane filter, polypropylene housing <strong>and</strong> supporting/drainage<br />

layers<br />

Sterilization By autoclaving at 121°C or 134°C<br />

Water penetration Occurs at approx. 4.5 bar (450 kPa) with 0.2 µm pore<br />

size Capsules<br />

ORDER NUMBERS <strong>for</strong> Sartofluor Capsules<br />

a) Sartofluor Capsules with inlet <strong>and</strong> outlet hose nipples:<br />

5181358T7 OO B 0.1 µm, 0.05 m2 5181358T8 OO B<br />

(pack of 5)<br />

0.1 µm, 0.1 m2 (pack of 5)<br />

5181358T9 OO A 0.1 µm, 0.2 m2 (pack of 4)<br />

5181307T7 OO B 0.2 µm, 0.05 m2 (pack of 5)<br />

5181307T8 OO B 0.2 µm, 0.1 m2 (pack of 5)<br />

5181307T9 OO A 0.2 µm, 0.2 m2 (pack of 4)<br />

5181306T9 OO A 0.45 µm, 0.2 m2 (pack of 4)<br />

b) Sartofluor Capsules with inlet <strong>and</strong> outlet Sanitary flanges:<br />

5181307T7 SS B 0.2 µm, 0.05 m2 (pack of 5)<br />

5181307T8 SS B 0.2 µm, 0.1 m2 (pack of 5)<br />

5181307T9 SS A 0.2 µm, 0.2 m2 (pack of 4)<br />

Capsules with other connector combinations may be<br />

available according to requirements. Details on request.<br />

51


52<br />

SPECIFICATIONS<br />

Connectors, 10 mm hose nipples<br />

Filter required, 25 mm PTFE membrane<br />

(type 118)<br />

Filtration area, 3 cm2 Flow rates <strong>for</strong> air at ∆p = 1 bar<br />

(100 kPa), 0.5 l/min with 0.2 µm,<br />

1.0 l/min with 0.45 µm pore size filters<br />

Materials, stainless steel except <strong>for</strong> silicone<br />

O-ring (21x2 mm) <strong>and</strong> aluminium<br />

closing ring<br />

Operating pressure, max. 5 bar<br />

(500 kPa)<br />

Sterilization, by autoclaving<br />

(121°C or 134°C) or dry heat (180°C)<br />

Weight, 170 g<br />

REPLACEMENT PARTS<br />

are shown in the diagram<br />

SPECIFICATIONS<br />

Connectors, 10 mm hose nipples<br />

Filter required, 47 mm PTFE membrane<br />

(type 118)<br />

Filtration area, 13 cm2 Flow rates <strong>for</strong> air at p = 0.3 bar<br />

(300 kPa), 0.5 l/min with 0.2 µm,<br />

1.0 l/min with 0.45 µm pore size filters<br />

Materials, stainless steel except <strong>for</strong><br />

silicone O-ring (42x3 mm), PTFE <strong>and</strong><br />

Viton valve seals<br />

Operating pressure, max. 20 bar<br />

(2,000 kPa)<br />

Sterilization, by autoclaving<br />

(121°C or 134°C) or dry heat (180°C)<br />

Weight, approx. 490 g<br />

REPLACEMENT PARTS<br />

are shown in the diagram<br />

Air/Gas filter holders<br />

25 mm <strong>and</strong> 47 mm Stainless Steel Filter<br />

Holders <strong>for</strong> in-line air/gas filtration<br />

25 mm In-line Filter Holder<br />

St<strong>and</strong>ardly supplied with stainless<br />

steel hose nipples which are<br />

precisely machined to seal into the<br />

holder body without sealing rings.<br />

Other connectors, available as<br />

accessories, fit the holder onto<br />

reducing valves or pumps with G1/4<br />

threaded nut (order no. 01030) or<br />

G3/8 threaded nut (01029) or onto<br />

pressure tanks with G3/8 male<br />

thread (00177).<br />

47 mm In-line Filter Holder<br />

The inlet side valve is convenient <strong>for</strong><br />

the intermittent run-off of retained water<br />

when filtering gases, or <strong>for</strong> air-venting<br />

when the holder is used as an in-line<br />

sampler <strong>for</strong> liquids. The hose nipples<br />

can be replaced by other connectors,<br />

available as accessories, to fit the<br />

holder onto reducing valves or pumps<br />

with G3/8 threaded nut (order<br />

no. 17089), pressure tanks with<br />

G3/8 male thread (17069) or taps<br />

with G3/4 male thread (17068).<br />

ORDER NUMBER<br />

16251 In-line Stainless Steel<br />

Holder <strong>for</strong> 25 mm Filters<br />

ORDER NUMBER<br />

16254 In-line Stainless Steel<br />

Holder <strong>for</strong> 47 mm Filters


Housings <strong>for</strong> sterile air venting<br />

<strong>and</strong> <strong>for</strong> air/gas filtration<br />

a) Air venting housing <strong>for</strong><br />

Mini Cartridges<br />

The base plate has a bayonet-lock<br />

on top <strong>for</strong> the Mini Cartridge, two<br />

large cut-outs in the sides <strong>for</strong> free air<br />

flow into <strong>and</strong> out of the housing bell,<br />

<strong>and</strong> a threaded connector nut below.<br />

The bell has slits in the top <strong>for</strong> rapid,<br />

complete drying-out of the Mini<br />

Cartridge after autoclaving.<br />

Mini Cartridge housings <strong>for</strong><br />

air/gas filtration<br />

The Mini Cartridge is held in a<br />

bayonet lock on a plate situated<br />

above the level of the inlet opening<br />

to prevent incoming steam from<br />

impinging directly on it at the start of<br />

the sterilization cycle, so ensuring the<br />

longest possible service life of the<br />

Mini cartridge. The base has space<br />

to collect condense water <strong>and</strong> a<br />

drain plug.<br />

Mini Cartridge Housings<br />

SPECIFICATIONS <strong>for</strong> the venting housing<br />

Connector DN 25 beverage type female threaded nut<br />

Filter 0.2 µm Sartofluor Mini Cartridge<br />

Materials AISI 316Ti stainless steel base plate <strong>and</strong> housing bell. Steel clamp<br />

Operating limits Max. operating pressure, 10 bar (1000 kPa),<br />

max. temperature, 180°C<br />

Height Approx. 186 mm (from base to top)<br />

Weight Approx. 700 g<br />

ORDER NUMBER <strong>for</strong> the venting housing<br />

350019V15-K025A Mini Cartridge Housing <strong>for</strong> sterile venting<br />

SPECIFICATIONS <strong>for</strong> air/gas filtration housings<br />

Connectors DN 15 beverage type or R1/2" female threads (as in photo)<br />

Filter Sartofluor Mini Cartridge (or a Sartopure-PP Mini Cartridge <strong>for</strong><br />

particle removal, or as a prefilter)<br />

Materials AISI 316Ti stainless steel base <strong>and</strong> top part. Steel clamp.<br />

Silicone O-ring (75.8 x 3.53 mm) in housing base <strong>and</strong> silicone<br />

gaskets (16 x 3 mm) in the DN 15 connecting pieces<br />

Operating limits Max. operating pressure, 10 bar (1000 kPa),<br />

max. temperature, 180°C<br />

Weight Approx. 1.3 kg<br />

Width Approx. 133 mm<br />

ORDER NUMBERS <strong>for</strong> air/gas filtration housings:<br />

360019P15KG015A Mini Cartridge housing <strong>for</strong> air/gas filtration, with<br />

DN 15 beverage type connectors<br />

360019P15II012A Mini Cartridge housing <strong>for</strong> air/gas filtration, with<br />

R1/2" female thread connectors<br />

53


Centrifugal <strong>and</strong> Crossflow Units<br />

<strong>for</strong> the concentration, purification<br />

<strong>and</strong> removal of proteins<br />

54<br />

Ultrafiltration<br />

The photograph above shows four ready-to-use units. From top to bottom, an EasyFlow unit with 1,000 cm 2 of pleated<br />

filter area <strong>for</strong> up to 10 liter volumes, a Sartocon-Micro Crossflow unit with 50 cm 2 of filtration area <strong>for</strong> up to 1 liter<br />

volumes, <strong>and</strong> a 2.5 ml capacity Centrisart-I centrifugal filter above the smaller 0.4 ml Centrisart-C4 microconcentrator.


Centrifugal units<br />

Centrisart ®<br />

-C4 <strong>and</strong> Centrisart ®<br />

-C30<br />

Micro-Concentrators <strong>for</strong> volumes of<br />

up to 0.4 <strong>and</strong> 3 ml<br />

Centrisart-C4 microconcentrators are<br />

designed <strong>for</strong> quick <strong>and</strong> efficient small<br />

volume ultrafiltration or microfiltration.<br />

Five types are available. Major<br />

applications are the concentration<br />

<strong>and</strong> purification of proteins <strong>and</strong><br />

DNA, buffer exchange, desalting, the<br />

removal of low molecular weight<br />

substances such as primers, non-incorporated<br />

nucleotides <strong>and</strong> excess<br />

linkers (5, 10, 20 or 100 kD units),<br />

as well as the removal of agarose<br />

particles from eluted DNA <strong>and</strong> the<br />

retention of cells <strong>and</strong> bacteria<br />

(0.2 µm unit)<br />

Identification of the units is quick <strong>and</strong><br />

easy. The MWCO (or pore size) <strong>and</strong><br />

an abbreviation <strong>for</strong> the filter material<br />

are printed on the outer (filtrate) tube,<br />

<strong>and</strong> an etched area can be used <strong>for</strong><br />

individual labelling.<br />

The larger Centrisart-C30 units are<br />

used <strong>for</strong> the same applications when<br />

larger sample volumes, up to 3 ml,<br />

are to be processed.<br />

SPECIFICATIONS <strong>for</strong> Centrisart-C4 units<br />

Capacity Max. 0.4 ml<br />

Centrifugal <strong>for</strong>ce Max. 10,000 x g<br />

Filtration area 0.28 cm2 Final volume 5 to 10 µl of concentrate<br />

Hold-up volume Less than 4 µl<br />

Materials Polypropylene centrifuge tube, cap <strong>and</strong> inner tube. Cellulose<br />

triacetate ultrafilter (5,000, 10,000 <strong>and</strong> 20,000 D MWCO),<br />

polysulfone ultrafilter (100,000 D MWCO) or cellulose acetate<br />

membrane filter (0.2 µm pore size). Silicone O-ring<br />

Per<strong>for</strong>mance Depends upon the Centrisart-C4 unit used <strong>and</strong> the initial volume<br />

<strong>and</strong> protein concentration of the sample. In general, 0.4 ml sample<br />

volumes can be concentrated 100 fold within 15 to 45 minutes.<br />

Particle removal with 0.2 µm units takes 30 minutes at most<br />

SPECIFICATIONS <strong>for</strong> Centrisart C30 units<br />

Capacity Max. 3 ml<br />

Centrifugal <strong>for</strong>ce Max. 5,000 x g<br />

Dimensions 87 mm total length, 16.4 mm outer diameter of filtrate tube<br />

Filtration area 0.78 cm2 Final volume 30 - 50 µl or more of concentrate<br />

Materials Polypropylene filtrate tube, filter cup <strong>and</strong> cap. Cellulose triacetate<br />

ultrafilter (5,000, 10,000 <strong>and</strong> 20,000 MWCO units), polysulfone<br />

ultrafilter (100,000 MWCO unit) or cellulose acetate<br />

membrane filter (0.2 µm pore size unit). Silicone O-Ring.<br />

Operating limits +4°C to +40°C<br />

Per<strong>for</strong>mance The time required <strong>for</strong> concentration depends upon the MWCO<br />

used as well as the composition <strong>and</strong> initial volume of the sample.<br />

In general, 3 ml sample volumes can be concentrated up to<br />

100-fold within 30 to 60 minutes. Particle removal with 0.2 µm<br />

units takes about 15 minutes.<br />

ORDER NUMBERS <strong>for</strong> Centrisart-C4 units<br />

13229 C4-G 5,000 D MWCO (pack of 25)<br />

13239 C4-G 10,000 D MWCO (pack of 25)<br />

13249 C4-G 20,000 D MWCO (pack of 25)<br />

13269 C4-G 100,000 D MWCO (pack of 25)<br />

13207 C4-G 0.2 µm (pack of 25)<br />

ORDER NUMBERS <strong>for</strong> Centrisart C30 units<br />

13239-C30 G 10,000 D MWCO, pack of 25<br />

13249-C30 G 20,000 D MWCO, pack of 25<br />

13269-C30 G 100,000 D MWCO, pack of 25<br />

13207-C30 G 0.2 µm pore size, pack of 25<br />

55


Centrifugal units<br />

Centrisart ®<br />

I Units <strong>for</strong> the rapid preparation<br />

of protein-free ultrafiltrates from up to 2.5 ml<br />

sample volumes<br />

Ready-to-use units <strong>for</strong> quick <strong>and</strong><br />

convenient small volume ultrafiltration<br />

in fixed angle or swinging bucket<br />

centrifuges. They are particularly<br />

effective <strong>for</strong> the removal of interfering<br />

proteins from samples in which low<br />

molecular weight substances are to<br />

be determined, e.g. <strong>for</strong> the determination<br />

of creatinine in serum, <strong>and</strong> <strong>for</strong><br />

drug binding studies. The ultrafiltrate<br />

is collected in the inner tube, where<br />

it is readily accessible.<br />

Centrisart I consists of an outer centrifuge<br />

tube, in which the sample is<br />

placed, <strong>and</strong> a free-sliding inner tube,<br />

closed on the underside with an<br />

ultrafilter. During centrifugation, the<br />

inner tube is pulled down against the<br />

sample, so that liquid is <strong>for</strong>ced<br />

through the ultrafilter, while particles<br />

<strong>and</strong> retained proteins are simultaneously<br />

pulled away from it. This is so<br />

effective in preventing blockage that<br />

even whole blood samples can be<br />

deproteinated.<br />

56<br />

Ultrafiltrate<br />

Ultrafilter<br />

Sample<br />

Centrifugal<br />

<strong>for</strong>ce<br />

Direction of<br />

ultrafiltration<br />

Particles etc.<br />

SPECIFICATIONS <strong>for</strong> Centrisart I units<br />

Capacity For 2.5 ml sample volume<br />

Centrifugal <strong>for</strong>ce Max. 2,500 G (swinging bucket rotor) or 2,000 G<br />

(fixed angle rotor)<br />

Dimensions Max. outer diameter 14 mm, overall length 93 mm<br />

Filtration area 0.79 cm2 Final volume 100 µl or more of concentrate<br />

Materials Polystyrene centrifuge tube, cellulose proprionate inner floater<br />

tube, polyethylene cap, cellulose triacetate (5, 10 <strong>and</strong> 20 kD<br />

units) or polysulfone ultrafilters (100 kD <strong>and</strong> 300 kD units)<br />

Per<strong>for</strong>mance Depends upon sample composition, starting volume <strong>and</strong> unit<br />

used. Generally, 2.5 ml samples can be completely filtered<br />

within 30 to 90 minutes.<br />

1.0<br />

0.5<br />

ml<br />

15 30 45<br />

60<br />

min<br />

ORDER NUMBERS <strong>for</strong> Centrisart I units<br />

The curves show ultrafiltration runs with<br />

1 ml starting volumes of:<br />

● Water<br />

■ 2.5% BSA<br />

◆ Bovine serum<br />

▲ Bovine blood<br />

13209 E Trial pack with 3 each of the 5,000 D, 10,000 D,<br />

20,000 D <strong>and</strong> 100,000 D units<br />

13229 E 5,000 D MWCO (pack of 12)<br />

13239 E 10,000 D MWCO (pack of 12)<br />

13249 E 20,000 D MWCO (pack of 12)<br />

13269 E 100,000 D MWCO (pack of 12)<br />

13279 E 300,000 D MWCO (pack of 12)


Crossflow units<br />

Sartocon ®<br />

-Micro.<br />

For faster <strong>and</strong> safer concentration of<br />

volumes of up to one liter<br />

These h<strong>and</strong>-size, ready-to-connect<br />

crossflow ultrafiltration units concentrate<br />

or purify 50 ml to 1 litre volumes.<br />

They bridge the gap between centrifugal<br />

units <strong>and</strong> crossflow module<br />

systems, <strong>and</strong> offer the ease of h<strong>and</strong>ling<br />

<strong>and</strong> user-safety of closed,<br />

complete units. They are supplied<br />

presterilized <strong>and</strong> pyrogen-free, <strong>and</strong><br />

require no assembly work. The<br />

tangential flow across the filter is<br />

gentle on the sample, avoiding the<br />

foaming which occurs in stirred cells.<br />

Every housing is tested to ensure that<br />

it is leakproof, to prevent product<br />

leaks during filtration, <strong>and</strong> every<br />

manufactured unit is integrity tested<br />

prior to packaging. Sartocon-Micro<br />

units contain the same membranes as<br />

the modules <strong>for</strong> Sartocon Crossflow<br />

Systems, so that they are ideal to test<br />

the per<strong>for</strong>mance of membranes <strong>for</strong><br />

large scale filtration processes.<br />

0,25<br />

0,20<br />

0,15<br />

0,10<br />

0,05<br />

Flux (ml/min x cm 2 ) with<br />

0.1% Albumin in 0.9% NaCl<br />

Sartocon-Mini<br />

Sartocon-Micro<br />

5 10 15 20 time (min)<br />

SPECIFICATIONS <strong>for</strong> Sartocon-Micro units<br />

Connectors Hose barbs of 4 mm outer diameter<br />

Cytotoxicity Non-toxic as determined with L929 mouse fibroblast cells <strong>and</strong><br />

with MRC-5 human embryonic lung cells<br />

Extractables Pass USP Plastics Class VI Test<br />

Filter area 50 cm2 Flow rates <strong>for</strong><br />

a) water Approx. 10 ml/min (5,000 D), 25 ml/min (10,000 D),<br />

35 ml/min (30,000 D), 110 ml/min (50,000 D), 125 ml/min<br />

(100,000 D), 700 ml/min (300,000 D), 1050 ml/min<br />

(0.2 µm), all at 2 bar (200 kPa)<br />

b) 1% albumin Approx. 6 ml/min (10,000 D unit at 2 bar), unchanged after<br />

cleaning with 0.1 N NaOH<br />

Hold-up volume 0.5 ml on concentrate side, 1.7 ml on filtrate side<br />

Materials Polyethersulfone membrane filter or ultrafilter, polypropylene housing,<br />

ethylene vinyl acetate tubing<br />

Operating limits Max. 2 bar pressure, 50°C temperature<br />

Priming volume Sartocon-Micro 4 ml. System with tubing 20-25 ml<br />

Processing volume 50 ml to 1 liter<br />

Test results with 5,000 D, 10,000 D <strong>and</strong><br />

50,000 D Sartocon-Micro units:<br />

MWCO Concentration of<br />

= 50,000 89 kD MAb (0.32%)<br />

● = 10,000 BSA (1%)<br />

= 5,000 Lysozyme (1%)<br />

Flux<br />

(ml/min)<br />

21<br />

0<br />

0 5 10 15 20 25 30<br />

Time (min)<br />

ORDER NUMBERS <strong>for</strong> Sartocon-Micro units, sterile packed<br />

15629-00-1 5,000 D MWCO (pack of 1)<br />

15639-00-1 10,000 D MWCO (pack of 1)<br />

15659-00-1 30,000 D MWCO (pack of 1)<br />

15650-00-1 50,000 D MWCO (pack of 1)<br />

15669-00-1 100,000 D MWCO (pack of 1)<br />

15679-00-1 300,000 D MWCO (pack of 1)<br />

15607-00-1 0.2 µm pore size (pack of 1)<br />

ORDER NUMBERS <strong>for</strong> accessories:<br />

17802 Holding clamp <strong>for</strong> Sartocon-Micro Unit<br />

RECOMMENDED ACCESSORIES,<br />

see 1a, page 63<br />

18<br />

15<br />

12<br />

9<br />

6<br />

3<br />

57


Large area units<br />

EasyFlow ®<br />

single use units with pleated<br />

filter <strong>for</strong> liter volumes<br />

A ready-to-connect EasyFlow unit<br />

<strong>and</strong> a peristaltic pump represent the<br />

simplest way to ultrafilter laboratory<br />

scale batches. Applications include<br />

protein concentration, diafiltration,<br />

virus retention <strong>and</strong> pyrogen removal.<br />

In addition to the ultrafilter units, an<br />

0.2 µm membrane filter unit is also<br />

available <strong>for</strong> cell harvesting.<br />

EasyFlow units are supplied sterile<br />

<strong>and</strong> pyrogen free. They are intended<br />

<strong>for</strong> single use, but can be cleaned<br />

<strong>and</strong> sanitized to some extent. The<br />

pleated ultrafilter enables a large<br />

filter area to be incorporated in a<br />

compact, practical housing, <strong>and</strong><br />

ensures a favourable cost per liter<br />

relationship, even when used <strong>for</strong><br />

direct, low-pressure filtration. The<br />

housing has two connectors on the<br />

inlet side, to allow use of the unit in<br />

a tangential flow mode when the<br />

highest possible throughput is<br />

required.<br />

58<br />

B<br />

A<br />

SPECIFICATIONS <strong>for</strong> EasyFlow Units<br />

Connectors Inlet <strong>and</strong> filtrate outlet, tubing conectors of 6 mm outer<br />

diameter. Air vent (or concentrate outlet), tubing conector<br />

of 5 mm outer diameter<br />

Dimensions Length 240 mm, width 36 mm, height 36 mm<br />

Filtration area 0.1 m 2<br />

Flow rate For water, typically 450 ml/min with the 20,000 cut-off unit<br />

Hold-up volume 10 ml<br />

Materials Cellulose triacetate or polysulfone ultrafilter, or cellulose<br />

acetate membrane filter, <strong>and</strong> polypropylene supporting<br />

material in a Cyrolite (CY/RO Industries trademarked<br />

MBS-copolymer) housing<br />

Operating limits Max. pressure 1 bar (100 kPa), max. temperature 50°C<br />

Per<strong>for</strong>mance Depends on many factors, including the initial volume <strong>and</strong><br />

protein concentration/composition of the sample <strong>and</strong> the cutoff<br />

used. One liter of a 1% albumin solution can be concentrated<br />

ten-fold within 10 minutes using a 20,000 cut-off unit.<br />

Processable volume Min. 100 ml, max. 10 liters<br />

ORDER NUMBERS <strong>for</strong> EasyFlow Units, sterile packed<br />

1. with cellulose triacetate ultrafilters:<br />

14529-OS-1-V 5,000 D MWCO (pack of 2)<br />

14549-OS-1-V 20,000 D MWCO (pack of 2)<br />

2. with polysulfone ultrafilters:<br />

14639-OS-1-V 10,000 D MWCO (pack of 2)<br />

14669-OS-1-V 100,000 D MWCO (pack of 2)<br />

14679-OS-1-V 300,000 D MWCO (pack of 2)<br />

3. with cellulose acetate membrane filter:<br />

12307-OS-1-V 0.2 µm pore size (pack of 2)<br />

RECOMMENDED ACCESSORIES,<br />

see 1b, page 63


Ultrafiltration bags/units<br />

Collodion Bags <strong>for</strong> protein concentration<br />

<strong>and</strong> DNA dialysis Ultrasart ®<br />

-D20 Units<br />

<strong>for</strong> the LAL-Test<br />

Collodion Bags<br />

Collodion Bags are used <strong>for</strong> gentle,<br />

small volume protein concentration,<br />

immersed in water in a glass holder<br />

connected to vacuum. The time taken<br />

depends upon the degree of concentration<br />

required <strong>and</strong> the starting<br />

volume. 5 ml of spinal fluid are<br />

concentrated down to about 20 µl in<br />

1.5 to 2 hours. The smaller Micro<br />

Collodion Bags of 1 ml capacity,<br />

used in a pressure system, allow far<br />

quicker protein concentration. 1 ml<br />

of spinal fluid can be concentrated to<br />

4 µl within about 4 minutes.<br />

The use of Sartorius Collodion Bags<br />

in the isolation of high molecular<br />

weight DNA from mammalian cells is<br />

described in ”Molecular Cloning. A<br />

Laboratory Manual“, Sambrook,<br />

Fritsch <strong>and</strong> Maniatis, Cold Spring<br />

Harbor Laboratory Press, 1989. The<br />

procedure is more gentle, <strong>and</strong> gives<br />

higher DNA yields, than<br />

phenol/chloro<strong>for</strong>m extraction.<br />

Collodion bags are also often<br />

preferred to dialysis tubing because<br />

they are firm bags which can be<br />

simply suspended in the dialysis<br />

beaker <strong>and</strong> give high recoveries<br />

even with sample volumes of less<br />

than 0.5 ml.<br />

SPECIFICATIONS <strong>for</strong> Collodion Bags<br />

Capacity, 8 ml<br />

Material, cellulose nitrate ultrafilter of<br />

12,000 D MWCO<br />

Max. vacuum, 50 mm Hg<br />

Packaging, supplied immersed in<br />

20% ethanol in a screw-cap plastic<br />

container<br />

Size, length 75 mm, diameter 15 mm<br />

SPECIFICATIONS <strong>for</strong><br />

Micro Collodion Bags<br />

Capacity, 1 ml<br />

Material, cellulose nitrate ultrafilter of<br />

12,000 D MWCO<br />

Max. pressure, 10 bar (1000 kPa)<br />

Packaging, supplied immersed in 20%<br />

ethanol in a screw-cap plastic container<br />

Size, length 27 mm, diameter 8.4 mm<br />

ORDER NUMBER<br />

<strong>for</strong> Collodion Bags<br />

13200E<br />

(Pack of 12)<br />

ORDER NUMBER<br />

<strong>for</strong> Micro<br />

Collodion Bags<br />

13202K<br />

(Pack of 50)<br />

RECOMMENDED ACCESSORIES<br />

are described on page 64<br />

Ultrasart-D20<br />

The limulus test is frequently used in<br />

pharmaceutical quality control.<br />

Ultrasart-D20 pressure ultrafiltration<br />

units allow samples containing<br />

interfering substances to be freed<br />

from them within 15 to 30 minutes,<br />

so allowing them to be LAL-tested<br />

reliably <strong>and</strong> without reducing the<br />

sensitivity of the test.<br />

SPECIFICATIONS<br />

Chemical compatibility, aqueous<br />

solutions of pH 3-9, 2 hours contact<br />

with 1 molar amino acids<br />

Filtration area, 5.3 cm2 Flow rate <strong>for</strong> water at 2 bar pressure,<br />

2 ml/min<br />

Materials, cellulose triacetate ultrafilter<br />

of 20,000 D MWCO <strong>and</strong> 100%<br />

endotoxin retention, Luran 368 R<br />

(polycarbonate) housing<br />

Max. sample volume, 15 ml<br />

ORDER NUMBER<br />

16520C Ultrasart-D20 Units,<br />

presterilized <strong>and</strong><br />

pyrogen-free<br />

(pack of 6)<br />

ACCESSORIES REQUIRED<br />

are described on page 64<br />

59


Sartocon Slice cassettes are pictured<br />

next to the larger Sartocon cassettes<br />

(see page opposite) in the photo on<br />

the right. The same construction <strong>and</strong><br />

materials are used in each of these<br />

cassette types, <strong>and</strong> the flow path<br />

lengths are identical, so that scale-up<br />

or scale-down is perfectly linear<br />

throughout the range of applications,<br />

from protein purification <strong>and</strong> concentration<br />

to cell harvesting <strong>and</strong> blood<br />

fractionation. Validation requirements,<br />

<strong>and</strong> their high cost, are<br />

greatly reduced or entirely eliminated.<br />

All cassettes have an excellent<br />

chemical compatibility, covering a<br />

wide pH range, <strong>and</strong> are autoclavable<br />

or in-line steamable, so that they can<br />

be easily <strong>and</strong> efficiently cleaned <strong>and</strong><br />

sterilized prior to re-use.<br />

The Sartocon Slice holding system<br />

can accomodate up to three Sartocon<br />

Slice cassettes. It is, like the Sartocon<br />

3 system (page opposite), designed<br />

<strong>for</strong> maximum per<strong>for</strong>mance <strong>and</strong> ease<br />

of cleaning. The new easy-to-connect<br />

design with in-line pressure transmitters<br />

eliminates the need <strong>for</strong> cumbersome<br />

T-connectors.<br />

60<br />

Crossflow Filtration<br />

Sartocon ®<br />

Slice<br />

The new pilot-scale crossflow system<br />

<strong>for</strong> batches of 1 to 100 liters<br />

SPECIFICATIONS <strong>for</strong> Sartocon Slice cassettes<br />

Biosafety All materials pass USP Plastics Test Class VI<br />

Compatibility pH 2-14 (Hydrosart), pH 1-14 (polyethersulfone)<br />

Filtration area 0.1 m2 Limits Max. 4 bar inlet pressure. Max. 50°C operating temp.<br />

Sterilization In-line steaming (121°C, 30 min) or autoclaving (121°C, 110 min)<br />

ORDER NUMBERS <strong>for</strong> the Sartocon Slice holding system<br />

17521---001 Sartocon Slice holding device (without accessories)<br />

17521---101 Sartocon Slice with accessory kit <strong>for</strong> microfiltration<br />

17521---102 Sartocon Slice with accessory kit <strong>for</strong> ultrafiltration<br />

ORDER NUMBERS <strong>for</strong> Sartocon Slice cassettes<br />

3051860601W--SG Hydrosart, 0.45 µm pore size<br />

3051860701W--SG Hydrosart, 0.2 µm pore size<br />

3051467901E--SG Polyethersulfone, 300,000 D MWCO<br />

3051466901E--PG Polyethersulfone, 100,000 D MWCO<br />

3051465001E--SG Polyethersulfone, 50,000 D MWCO<br />

3051465901E--SG Polyethersulfone, 30,000 D MWCO<br />

3051463901E--SG Polyethersulfone, 10,000 D MWCO<br />

3051462901E--SG Polyethersulfone, 5,000 D MWCO


Crossflow Filtration<br />

Sartocon ®<br />

3<br />

The new hydraulic crossflow system<br />

<strong>for</strong> batches of 10 to 10,000 liters<br />

The Sartocon 3 holding device<br />

(shown on the right with the maximum<br />

load of 10 cassettes) is designed <strong>for</strong><br />

easy h<strong>and</strong>ling <strong>and</strong> maximum convenience.<br />

An easy-to-use, water-based<br />

hydraulic compression system replaces<br />

clamps. All connections are to a<br />

single 316L stationary plate, uniquely<br />

suited <strong>for</strong> hard piping <strong>and</strong> allowing<br />

easy access, which, together with the<br />

highly polished surfaces (Ra ² 0.6 µm),<br />

simplifies cleaning.<br />

The hydraulic compression system<br />

avoids over-compression <strong>and</strong> possible<br />

cassette damage. It seals the system<br />

in seconds, <strong>and</strong> maintains consistent<br />

compression <strong>for</strong>ce.<br />

The three different cassette types<br />

allow the optimum cassette to be<br />

selected <strong>for</strong> a specific application.<br />

Major features of Sartorius Crossflow<br />

cassettes:<br />

Hydrosart cassettes<br />

O low binding capacity<br />

O pH range of 2 to 14<br />

O high flux rate<br />

O long service life<br />

O easy to clean <strong>and</strong> recycle<br />

O steam sterilizable<br />

Polyethersulfone cassettes<br />

O pH range of 1 to 14<br />

O long service life<br />

O easy to clean<br />

O steam sterilizable<br />

Cellulose triacetate cassettes<br />

O low binding capacity<br />

O easy to clean <strong>and</strong> recycle<br />

O high flux rate<br />

O high process reliability<br />

SPECIFICATIONS <strong>for</strong> Sartocon cassettes<br />

Biosafety All materials pass USP Plastics Test Class VI<br />

Compatibility pH 2-14 (Hydrosart), pH 1-14 (polyethersulfone),<br />

pH 4-8 (cellulose triacetate)<br />

Filtration area 0.6 m2 (MF cassette), 0.7 m2 (UF cassette)<br />

Flow rates Typical values <strong>for</strong> water (2 bar inlet <strong>and</strong> 0.5 bar retentate outlet pressure,<br />

free filtrate outlet). Permeate values in l/h/m2 (retentate in l/h):<br />

Hydrosart: 0.45 µm = 1,400 (680), 0.2 µm = 1,300 (680).<br />

Polyethersulfone: 300 kD = 942 (490), 100 kD = 598 (467),<br />

30 kD = 567 (820), 10 kD = 139 (731). Cellulose triacetate:<br />

20 kD = 200 (850), 10 kD = 57 (630), 5 kD = 12 (1,000)<br />

Limits Max. 4 bar inlet pressure. Max. 50°C operating temp.<br />

Sterilization Hydrosart <strong>and</strong> polyethersulfone cassettes, by in-line steaming<br />

(121°C, 30 min) or autoclaving (121°C, 110 min)<br />

ORDER NUMBERS <strong>for</strong> the Sartocon 3 holding system<br />

17546---001 Sartocon 3 holding device (without accessories)<br />

17546---101 Sartocon 3 with accessory kit <strong>for</strong> microfiltration<br />

17546---102 Sartocon 3 with accessory kit <strong>for</strong> ultrafiltration<br />

ORDER NUMBERS <strong>for</strong> Sartocon cassettes<br />

3021860606W--SG Hydrosart, 0.45 µm pore size<br />

3021860706W--SG Hydrosart, 0.2 µm pore size<br />

3021467907E--SG Polyethersulfone, 300,000 D MWCO<br />

3021466907E--PG Polyethersulfone, 100,000 D MWCO<br />

3021465007E--SG Polyethersulfone, 50,000 D MWCO<br />

3021465907E--SG Polyethersulfone, 30,000 D MWCO<br />

3021463907E--SG Polyethersulfone, 10,000 D MWCO<br />

3021462907E--SG Polyethersulfone, 5,000 D MWCO<br />

3021454907E--SG Cellulose triacetate, 20,000 D MWCO<br />

3021453907E--SG Cellulose triacetate, 10,000 D MWCO<br />

3021452907E--SG Cellulose triacetate, 5,000 D MWCO<br />

61


Ultrasart ® X<br />

The spiral flow cell which avoids<br />

the drawbacks of stirred cells<br />

In ultrafiltration, liquid movement<br />

across the filter surface is necessary<br />

to prevent a filter-blocking layer of retained<br />

material from building-up on it.<br />

This is achieved in stirred cells by the<br />

rapid rotation of a stirrer bar. Ultrasart<br />

X uses the more gentle <strong>and</strong> effective<br />

approach, recirculation across the filter<br />

in spiral flow mode.<br />

The extra top part next to an assembled<br />

Ultrasart X (above) is inverted to<br />

show the spiral recirculation channel.<br />

As shown below, the recirculation inlet<br />

<strong>and</strong> outlet are both on the top part, allowing<br />

simple connection to a peristaltic<br />

pump.<br />

Whereas adding more sample to a<br />

stirred cell entails stopping the run<br />

<strong>and</strong> opening the cell, Ultrasart X offers<br />

far more flexibility. The retentate<br />

is always readily accessible, <strong>and</strong> can<br />

be added to as required without interrupting<br />

filtration.<br />

62<br />

Cell <strong>for</strong> Disc Ultrafilters<br />

APPLICATIONS <strong>for</strong> Ultrasart X include:<br />

O Concentration of proteins <strong>and</strong> viruses. Purification of proteins<br />

O Desalting. Diafiltration<br />

O Virus removal. Particle removal<br />

SPECIFICATIONS <strong>for</strong> Ultrasart X<br />

Compatibility compatible with aqueous solutions of pH 1-14, including 1N<br />

NaOH <strong>and</strong> 30% <strong>for</strong>maldehyde, as well as to solvents such as<br />

ethylene glycol <strong>and</strong> isopropanol<br />

Connectors hose barbs <strong>for</strong> approx. 2.5 mm i. d. hose<br />

Depyrogenation circulate 1N NaOH through the system <strong>for</strong> 30 minutes at room<br />

temperature<br />

Dimensions 69 mm outer diameter<br />

Filter diameter 47 mm<br />

Hold-up volume approx. 0.5 ml (5-10 ml with tubing)<br />

Materials Delrin* top part, base part <strong>and</strong> closing ring. Silicone O-ring<br />

(*Delrin is a trademark of E. I. DuPont Nemours & Co)<br />

Operating conditions <strong>for</strong> use with a peristaltic pump or similar with a pumping per<strong>for</strong>mance<br />

of max. 100 ml/min.<br />

Per<strong>for</strong>mance depends on the initial volume <strong>and</strong> protein concentration/composition<br />

of the sample, the circulation rate <strong>and</strong> the MWCO<br />

used. Fifty ml of a 1% BSA solution can be concentrated ten-fold<br />

within 60 minutes using a 10,000 MWCO ultrafilter at a circulation<br />

rate of 28 ml/min.<br />

On the left, the improved maintenance of<br />

initial flow rate with Ultrasart X compared<br />

to a stirred cell when concentrating 50 ml<br />

of 1% BSA with 10,000 MWCO ultrafilters.<br />

Sterilization alone, or with a mounted Polysulfone ultrafilter, by autoclaving,<br />

at 121°C, 30 min<br />

Working volume 10 to 100 ml<br />

ORDER NUMBER <strong>for</strong> Ultrasart X<br />

16668 Ultrasart X ultrafiltration cell, complete with holder<br />

Select the 47 mm diameter ultrafilters required from the order number list on p.78.<br />

Ultrasart X can also be used <strong>for</strong> particle removal with 0.2 mm pore size cellulose<br />

acetate membrane filters (order number 11107-047N, pack of 100)


Accessories <strong>for</strong> EasyFlow <strong>and</strong><br />

Sartocon ®<br />

Micro<br />

1. Recommended<br />

Accessories <strong>for</strong> complete<br />

filtration systems<br />

1a. For Sartocon Micro Units<br />

Each unit is supplied with a tubing set<br />

containing ethylene vinyl acetate<br />

tubing of 5 mm inner diameter,<br />

closing caps <strong>and</strong> a hose clip <strong>for</strong><br />

regulation of the pressure. Additional<br />

requirements are a peristaltic pump<br />

<strong>and</strong> pump tubing (e.g. pump 16697<br />

with tubing 16817), <strong>and</strong> an anodized<br />

aluminium, reusable clamp (17802),<br />

in which the Sartocon-Micro unit is<br />

slipped <strong>for</strong> a firm hold during use, as<br />

shown in the photo below.<br />

The Perisys 70 Benchtop System is a<br />

complete, compact system <strong>for</strong> Sartocon<br />

Micro with optimized tubing lengths<br />

<strong>for</strong> increased yields.<br />

1b. For EasyFlow Units<br />

Liquid can be fed into these from a<br />

pressure tank <strong>for</strong> low pressure filtration,<br />

or they can be used with a peristaltic<br />

pump 16697 <strong>and</strong> tubing 16817 in<br />

a limited crossflow mode.<br />

2. Descriptions of<br />

Accessories<br />

Peristaltic Pump, 50 l/h<br />

Order number: 16697<br />

Watson-Marlow pump type 302/SA<br />

with pumphead 501 RI. Recommended<br />

<strong>for</strong> use with Sartocon Micro units<br />

<strong>and</strong> with EasyFlow units, together<br />

with Verderprene tubing of 4.8 mm<br />

inner diameter <strong>and</strong> 1.6 mm wall<br />

thickness (order number 16817 <strong>for</strong> a<br />

length of 1.5 m).<br />

Specifications:<br />

Length x height x width:<br />

300 x 180 x 125 mm<br />

Line voltage, 100-240 V, 50-60 Hz<br />

(no adjustment necessary)<br />

Pump head, with 2 springloaded<br />

rollers<br />

Pumping rate, up to 900 ml/min with<br />

4.8 x 1.6 mm tubing, type 16817<br />

Speed, max. 220 rpm (control<br />

range, 10 to 100%)<br />

Tubing requirements, 0.5 to 8.0 inner<br />

diameter, 1.6 mm wall thickness.<br />

Weight, 5.2 kg<br />

Accessories/Ultrafiltration<br />

Perisys 70 Benchtop System<br />

<strong>for</strong> Sartocon Micro<br />

Order number: 17205<br />

To obtain maximum yields with an<br />

ultrafiltration system, the pump,<br />

container, ultrafiltration unit <strong>and</strong><br />

tubing system must be designed to<br />

match each other exactly <strong>and</strong> to<br />

keep the fluid pathways as short as<br />

possible. This has been excellently<br />

achieved in Perisys 70, a complete<br />

system which requires little bench<br />

space (only 185 x 235 mm), is easy<br />

to assemble <strong>and</strong> use <strong>and</strong> ensures<br />

highest recoveries.<br />

The dead volume of the tubing is less<br />

than 0.5 ml. The 1 liter glass container<br />

has 100 ml graduations <strong>and</strong> can<br />

be autoclaved together with the<br />

tubing <strong>for</strong> certain sterilization <strong>and</strong><br />

avoidance of cross-contamination.<br />

The peristaltic pump housing is made<br />

of stainless steel, the pumphead of<br />

POM. The pump supplies flow rates<br />

of 0.4 to 620 ml/min <strong>and</strong> accepts<br />

tubing of 3.2 to 6.4 mm internal<br />

diameter.<br />

Further details on request.<br />

63


Accessories <strong>for</strong> Collodion Bags<br />

<strong>and</strong> Ultrasart ®<br />

-D20<br />

Holders <strong>for</strong> Collodion Bags<br />

Order numbers:<br />

16304 Round-bottom glass holder<br />

16305 Flat-bottom glass holder<br />

26314 6-Place manifold<br />

The round- <strong>and</strong> flat-bottom holders<br />

shown above each accept one single<br />

Collodion Bag. The 6-place manifold<br />

is supplied with 6 round-bottom<br />

holders in a stainless steel st<strong>and</strong>,<br />

connecting tubing, <strong>and</strong> a glass<br />

manifold having taps on each of the<br />

6 hose nipple connectors to the<br />

holders <strong>for</strong> individual control of<br />

vacuum.<br />

Holders <strong>and</strong> manifold require a<br />

source of vacuum <strong>and</strong> connecting<br />

hose, e.g. type 16623.<br />

Replacement part numbers are shown<br />

in the diagram below.<br />

64<br />

3-Place Pressure Manifold <strong>for</strong> Micro<br />

Collodion Bags<br />

Order number: 16538<br />

Supplied with three polycarbonate<br />

supports of 1 ml capacity, each <strong>for</strong> a<br />

single Micro Collodion Bag. After<br />

insertion of a bag, a support is<br />

simply snapped into one of the three<br />

quick-connectors on the manifold. The<br />

connector is only open when a<br />

support is in position. Connection of<br />

the quick-connect nipple pressure inlet<br />

of the manifold to a pressure source<br />

requires a pressure hose with G1/4<br />

Accessories / Ultrafiltration<br />

threaded nut (order number 16929),<br />

or with G3/8 threaded nut (order<br />

number 16931), according to the<br />

male thread of the pressure source<br />

outlet. The maximum operating<br />

pressure is 10 bar (1000 kPa).<br />

Additional polycarbonate supports<br />

of 1 ml capacity are available as a<br />

pack of 12 under the order number<br />

16536E. When larger volumes are<br />

to be filtered, a polycarbonate<br />

support of 5 ml capacity can be<br />

used, order number 16597.<br />

Replacement parts:<br />

6980632 Pressure gauge<br />

6980633 Gasket <strong>for</strong> gauge<br />

3-Place Pressure Manifold <strong>for</strong><br />

Ultrasart-D20 Units<br />

Order number: 16506<br />

Complete with 3 liter pressure tank<br />

<strong>and</strong> connecting hose with G3/8<br />

threaded nut <strong>for</strong> connection to the<br />

pressure source. Ultrasart-D20 units<br />

can be screwed directly into the ports<br />

of the manifold, each of which has a<br />

tap <strong>for</strong> individual control of rinse<br />

liquid flow or pressure <strong>and</strong> of air<br />

venting. Two manifolds can be<br />

connected together using an adapter,<br />

order number 17152, plus a nipple,<br />

order number 17153.<br />

Maximum operating pressure is<br />

5 bar (500 kPa). The manifold with<br />

connecting hose to the pressure tank<br />

<strong>and</strong> the pressure tank (with pressure<br />

gauge removed) can be depyrogenated<br />

by up to 200°C heat.


Sartorius – Investing in the future<br />

In order to do justice to the requirements<br />

of the future <strong>and</strong> to set strict<br />

st<strong>and</strong>ards <strong>for</strong> the environmentally<br />

protective production of membrane<br />

filters, cartridges <strong>and</strong> cassettes,<br />

Sartorius began in 1989 constructing<br />

new production facilities on an industrial<br />

complex covering 104,000 m2 .<br />

The first building phase was completed<br />

in 1991, <strong>and</strong> the factory went into<br />

operation, equipped with new,<br />

innovative control technology <strong>and</strong> a<br />

manufacturing process which was<br />

developed by Sartorius <strong>and</strong> sponsored<br />

by the Federal Department of the<br />

Environment with the sum of 3 million<br />

marks. This process is environmentally<br />

friendly <strong>and</strong> low in emissions,<br />

incorporating a cost-efficient solvent<br />

recycling system which recovers<br />

99.9% of the solvents used <strong>and</strong><br />

recycles 99% of them in production.<br />

We supply <strong>Quality</strong><br />

As was the case in the past, Sartorius<br />

sees its future in gearing <strong>Quality</strong><br />

Assurance to meet specific customer<br />

requirements. For years our production<br />

methods <strong>and</strong> quality assurance have<br />

100<br />

con<strong>for</strong>med to GMP st<strong>and</strong>ards <strong>and</strong><br />

have been audited by experts from<br />

multi-national customers. As manufacturer<br />

of ”sterile medical devices<br />

<strong>and</strong> surgical products“ Sartorius has<br />

been awarded the D. H. Certificate<br />

of Registration. The Sartorius<br />

<strong>Separation</strong> Technologies Division<br />

was certified according to DIN ISO<br />

9001 by the DQS (German Society<br />

<strong>for</strong> the Certification of <strong>Quality</strong><br />

We offer worldwide service<br />

Management Systems) as of May<br />

1994, thus completing coverage of<br />

DIN ISO 9001 certification of<br />

Sartorius AG in Göttingen.<br />

We work with customers<br />

Sartorius is not just a filter manufacturer<br />

but is able to provide its customers<br />

with complete system solutions,<br />

including support with the validation<br />

of their production processes.


Order Number Index<br />

Order Number Index<br />

Order number: Page: Order number: Page: Order number: Page:<br />

A100X 93<br />

A15X 93<br />

A5F 93<br />

C100X 93<br />

C120 94<br />

C15X 93<br />

C550 94<br />

C5500X10 94<br />

C5F 93<br />

C600X5 94<br />

D100X 93<br />

D120 94<br />

D15X 93<br />

D550 94<br />

D5500X10 94<br />

D5F 93<br />

D600X5 94<br />

Q100X 93<br />

Q120 94<br />

Q15X 93<br />

Q550 94<br />

Q5500X10 94<br />

Q5F 93<br />

Q600X5 94<br />

S100X 93<br />

S120 94<br />

S15X 93<br />

S550 94<br />

S5500X10 94<br />

S5F 93<br />

S600X5 94<br />

00118 21<br />

00124 22, 23<br />

00177 52<br />

00179 38<br />

01029 52<br />

01030 52<br />

01324 17<br />

01325 17<br />

11104 66<br />

11105 66<br />

11106 66<br />

11107 66<br />

11301 68-69<br />

11302 68-69<br />

11303 68-69<br />

11304 68-69<br />

11305 68-69<br />

11306 68-69<br />

11306BL 72<br />

11307 68-69<br />

11307BL 72<br />

11342 68-69<br />

11347 68-69<br />

11358 68-69<br />

11403 74, 75<br />

11404 74, 75<br />

11405 74, 75<br />

11406 74, 75<br />

11407 74, 75<br />

11803 73<br />

11806 73<br />

11807 73<br />

11842 73<br />

12301 66<br />

12301-050 ACN 28<br />

12302 66<br />

12303 66<br />

12307 0S1V 58<br />

12342 66<br />

12602-080 ALK 97<br />

12806 BL 72<br />

12807 BL 72<br />

13001 75<br />

13004 74, 75<br />

13005 74, 75<br />

13006 74, 75<br />

13101 75<br />

13106 74, 75<br />

13107 74, 75<br />

13200 E 59<br />

13202 K 59<br />

13207 C30G 55<br />

13207 C4G 55<br />

13209 E 56<br />

13229 C4G 55<br />

13229 E 56<br />

13239 C30G 55<br />

13239 C4G 55<br />

13239 E 56<br />

13249 C30G 55<br />

13249 C4G 55<br />

13249 E 56<br />

13269 C30G 55<br />

13269 C4G 55<br />

13269 E 56<br />

13279 E 56<br />

13400 71<br />

13420 71<br />

13430 71<br />

13440 71<br />

13806 74, 75<br />

13906 74, 75<br />

14003-.... 76<br />

14005-.... 76<br />

14006-.... 76<br />

14008-.... 76<br />

14051-.... 76<br />

14053-.... 76<br />

14054-.... 76<br />

14055-.... 76<br />

14056-.... 76<br />

14057-.... 76<br />

14058-.... 76<br />

14059-.... 76<br />

14061-.... 76<br />

14062-.... 76<br />

14063-.... 76<br />

14064-.... 76<br />

14065-.... 76<br />

14066-.... 76<br />

14067-.... 76<br />

14068-.... 76<br />

14069-.... 76<br />

14070-.... 76<br />

14072-.... 76<br />

14074-.... 76<br />

14075-.... 76<br />

14078-.... 76<br />

14079-.... 76<br />

14080-.... 76<br />

14082-.... 76<br />

14083-.... 76<br />

14096-.... 76<br />

14097-.... 76<br />

14132 K 21<br />

14529 78<br />

14529 0S1V 58<br />

14539 78<br />

14549 78<br />

14549 0S1V 58<br />

14639 78<br />

14639 0S1V 58<br />

14659 78<br />

14669 78<br />

14669 0S1V 58<br />

14679 78<br />

14679 0S1V 58<br />

15629 001 57<br />

15639 001 57<br />

15650 001 57<br />

15659 001 57<br />

15669 001 57<br />

15679 001 57<br />

16201 24<br />

16214 15<br />

16219 24<br />

16220 24<br />

16249 38<br />

16251 52<br />

16254 52<br />

16274 40<br />

16275 39<br />

16276 41<br />

16276-3 94<br />

16277 42<br />

16277-7 94<br />

16286 99<br />

16296---05 98<br />

16304 64<br />

16305 64<br />

16306 21<br />

16307 22<br />

16309 23<br />

101


Order Number Index<br />

16315 21<br />

16316 22<br />

16506 64<br />

16508 B 36<br />

16510 20<br />

16511 20<br />

16514 E 14<br />

16517 E 15<br />

16520 C 59<br />

16522 29<br />

16523 29<br />

16534 K 12<br />

16534 Q 12<br />

16536 E 64<br />

16538 64<br />

16540 37<br />

16555 K 13<br />

16555 Q 13<br />

16574 14<br />

16579 37<br />

16592 K 13<br />

16592 Q 13<br />

16596 HYK 50<br />

16596 HYQ 50<br />

16597 64<br />

16606 27<br />

16610 27<br />

16611 27<br />

16612 28<br />

16615 28<br />

16617 46<br />

16618 17<br />

16619 17<br />

16620 17<br />

16622 17<br />

16623 29<br />

16625 17, 28, 48<br />

16633 47<br />

16634 48<br />

16639 17<br />

16644 E 17<br />

16645 E 17<br />

16646 E 17<br />

16647 E 17<br />

16656 48<br />

16660 44<br />

16662 46<br />

16668 62<br />

16671 29<br />

16672 27<br />

16673 27<br />

16685 16<br />

16692 28<br />

16695 28<br />

16697 63<br />

16740 97<br />

16743 96<br />

16744 96<br />

102<br />

16803 44<br />

16807 28<br />

16817 63<br />

16823 48<br />

16824 25<br />

16826 27<br />

16828 25<br />

16829 27<br />

16831 25<br />

16832 25<br />

16836 26<br />

16837 26<br />

16863 45, 47<br />

16880 44<br />

16881 44<br />

16929 64<br />

16931 48, 64<br />

16970 97<br />

16976 97<br />

16999 48<br />

17004 27<br />

17005 27<br />

17006 27<br />

17016 44<br />

17017 44, 45<br />

17019 44<br />

17033 44,48<br />

17036 44<br />

17037 97<br />

17038 38<br />

17039 37<br />

17051 37<br />

17068 52<br />

17069 52<br />

17070 47<br />

17085 97<br />

17088 97<br />

17089 44, 45, 52<br />

17090 45<br />

17091 48<br />

17098 44, 46<br />

17099 44, 46<br />

17105 37<br />

17108 D 21<br />

17109 21<br />

17145 38<br />

17146 22<br />

17147 22<br />

17148 21<br />

17149 21<br />

17150 44, 48<br />

17152 64<br />

17153 64<br />

17168 97<br />

17170 45, 47<br />

17173 26<br />

17174 26<br />

17175 26<br />

Order Number Index<br />

Order number: Page: Order number: Page: Order number: Page:<br />

17205 63<br />

17521---001 60<br />

17521---101 60<br />

17521---102 60<br />

17528-080 ACD 96<br />

17528-080 BZD 96<br />

17530 46<br />

17531 46<br />

17532 46<br />

17533 46<br />

17534 46<br />

17535 46<br />

17536 46<br />

17546---001 61<br />

17546---101 61<br />

17546---102 61<br />

17558 K 11<br />

17558 Q 11<br />

17559 K 11<br />

17559 Q 11<br />

17573 K 11<br />

17573 Q 11<br />

17573 ZYQ 11<br />

17574 K 11<br />

17574 Q 11<br />

17574 ZYQ 11<br />

17575 K 11<br />

17575 Q 11<br />

17576 K 11<br />

17576 Q 11<br />

17593 K 13<br />

17593 Q 13<br />

17594 K 13<br />

17594 Q 13<br />

17597 K 12<br />

17597 Q 12<br />

17598 K 13<br />

17598 Q 13<br />

17635 47<br />

17636 47<br />

17649 29<br />

17655 97<br />

17656 97<br />

17657 97<br />

17660 97<br />

17661 97<br />

17662 97<br />

17712 97<br />

17747 35<br />

17748 35<br />

17749 35<br />

17750 35<br />

17761 K 10<br />

17761 Q 10<br />

17761 ZYQ 10<br />

17762 K 10<br />

17762 Q 10<br />

17762 ZYQ 10


Order Number Index<br />

17764 K 10<br />

17764 Q 10<br />

17765 K 10<br />

17765 Q 10<br />

17766 35<br />

17771 35<br />

17773 35<br />

17799 17<br />

17801 96<br />

17802 57, 63<br />

17804 E 50<br />

17804 G 50<br />

17804 M 28<br />

17804 NPE 50<br />

17804 NPG 50<br />

17805 E 50<br />

17805 G 50<br />

17805 UPN 50<br />

17805 NPE 50<br />

17805 NPG 50<br />

17820 K 11<br />

17820 Q 11<br />

17821 K 10<br />

17821 Q 10<br />

17822 K 10<br />

17822 Q 10<br />

17823 K 12<br />

17823 Q 12<br />

17824 K 13<br />

17824 Q 13<br />

17825 Q 13<br />

17829 K 13<br />

17829 Q 13<br />

18041 D 19<br />

18043 E 19<br />

18044 E 19<br />

18045 E 19<br />

18046 E 19<br />

18047 D 19<br />

18052 D 31<br />

18053 D 31<br />

18056 D 31<br />

18058 D 31<br />

18059 46<br />

18099 17, 31<br />

18113 29<br />

18406 67<br />

18407 67<br />

24002 21<br />

25006 70<br />

25006 BL 72<br />

25007 70<br />

25007 BL 72<br />

26314 64<br />

302145...... 61<br />

302146...... 61<br />

302186...... 61<br />

305146...... 60<br />

305186...... 60<br />

331019P15TT112A 43<br />

332019P15TT112A 43<br />

350019V15-K025A 53<br />

360019P15II015A 53<br />

360019P15KG015A 53<br />

5181306.... 51<br />

5181307... 51<br />

5181358.... 51<br />

5181506.... 81<br />

5181507.... 81<br />

5181558.... 81<br />

5231306...... 33<br />

5231307...... 33<br />

except <strong>for</strong>:<br />

5231307H5NOV 32<br />

5231307H5OOB 32<br />

5231307H5OOV 32<br />

5231307H8VNB 35<br />

5231307H8VOB 35<br />

5231307H8VUB 35<br />

5231307H8VZB 35<br />

5231307HNOB 32<br />

5401306... 79<br />

5401307... 79<br />

5521302... 34<br />

5521303... 34<br />

5521307... 34<br />

5521320... 34<br />

5521502... 80<br />

5521503... 80<br />

5521507... 80<br />

5521520... 80<br />

5531303... 34<br />

5531307... 34<br />

5531503... 80<br />

5531507... 80<br />

5601304... 34<br />

5601305... 34<br />

5601504... 80<br />

5601505... 80<br />

5621304... 34<br />

5621305... 34<br />

5621306... 34<br />

5621307... 34<br />

5621504... 80<br />

5621505... 80<br />

5621506... 80<br />

5621507... 80<br />

6980101 24<br />

6980102 24, 25<br />

6980103 24, 25<br />

6980104 24, 25<br />

6980110 20, 36<br />

6980115 21<br />

6980116 21<br />

6980117 21<br />

6980119 22, 23<br />

Order Number Index<br />

Order number: Page: Order number: Page: Order number: Page:<br />

6980120 22<br />

6980121 22<br />

6980123 22<br />

6980151 38, 40<br />

6980176 52<br />

6980178 38, 52<br />

6980180 38, 52<br />

6980225 20<br />

6980226 20<br />

6980227 20<br />

6980228 20<br />

6980229 20<br />

6980230 20<br />

6980232 20, 29, 36<br />

6980233 20<br />

6980234 20<br />

6980235 20, 27<br />

6980236 20<br />

6980264 25<br />

6980267 27<br />

6980268 27<br />

6980270 25, 27<br />

6980271 25, 27<br />

6980272 25, 27<br />

6980383 36<br />

6980389 47<br />

6980390 47<br />

6980395 47<br />

6980396 47<br />

6980407 48<br />

6980415 47<br />

6980420 47<br />

6980569 14<br />

6980570 15<br />

6980573 29<br />

6980574 29<br />

6980586 29<br />

6980588 29<br />

6980589 29<br />

6980590 29<br />

6980591 29<br />

6980595 15<br />

6980632 64<br />

6980633 64<br />

6980656 39, 40, 41,<br />

42, 43, 52<br />

6980657 43<br />

6980700 37<br />

6980701 37<br />

6980702 39, 40<br />

6980703 37<br />

6980704 37<br />

6980705 37<br />

6980706 37<br />

6980707 39, 40, 41<br />

6980708 39, 40, 41<br />

6980711 39, 40, 41<br />

6980712 37, 39, 41<br />

103


Order Number Index<br />

Order number: Page:<br />

6980713 37<br />

6980714 42<br />

6980715 42<br />

6980716 42<br />

6980717 39, 40, 41,<br />

42, 52<br />

6980718 42<br />

6980719 42<br />

6980721 52<br />

6980722 39, 40, 41,<br />

42, 43, 52<br />

6980737 38, 52<br />

6980740 41<br />

6980767 43<br />

6980801 38, 39, 40,<br />

52<br />

6980810 64<br />

6980825 64<br />

6981001 24, 25<br />

6981002 24, 25<br />

6981003 24, 25<br />

6981004 24<br />

6981031 52<br />

6981033 52<br />

6981032 52<br />

6981034 52<br />

6981063 24, 25<br />

6981064 24, 25<br />

6981065 24, 25<br />

6981075 64<br />

6981076 64<br />

6981077 64<br />

6981078 64<br />

6981079 64<br />

6981080 64<br />

6981083 64<br />

6981090 20<br />

6981139 28<br />

6981288 38<br />

6981314 48<br />

6981320 17<br />

6981321 17<br />

6981322 17<br />

6981323 17<br />

6981540 29<br />

6982001 39, 40, 41<br />

6982002 40<br />

6982003 52<br />

6982005 38, 52<br />

6982006 38, 52<br />

6982012 39<br />

6982020 38<br />

6982022 42<br />

6982029 48<br />

6982036 40<br />

6982043 39, 40<br />

6982044 42<br />

6982060 48<br />

104<br />

6982061 48<br />

6982070 39, 40, 41<br />

6982071 39, 40, 41<br />

6982072 39, 40, 41<br />

6982077 42<br />

6982078 42<br />

6982079 42<br />

6982119 24, 25<br />

6983001 23<br />

6983002 23<br />

6983003 23<br />

6983004 23<br />

6983005 23<br />

6983006 22<br />

6983008 21<br />

6983009 22, 23<br />

6983010 22<br />

6985000 37<br />

6985001 37<br />

6985002 37<br />

6985004 36<br />

6985010 37<br />

6985011 37<br />

6985060 29<br />

6985093 47<br />

6985128 45, 48<br />

6985131 47<br />

6985149 81<br />

6985150 81<br />

6985151 81<br />

6985157 43<br />

6985183 39, 40, 41,<br />

42, 43<br />

6985184 39, 40, 41,<br />

42<br />

6985216 48<br />

6986006 46<br />

6986017 28<br />

6986055 38, 40<br />

6986070 17<br />

6986071 17<br />

6986073 17<br />

6986078 17<br />

6986079 17<br />

6986080 17<br />

6986081 17<br />

6986083 38<br />

6986084 38<br />

6986090 48<br />

6986091 48<br />

6986092 48<br />

6986105 28<br />

6986110 47<br />

6986111 47<br />

6986112 47<br />

6986113 47<br />

6986114 47<br />

6986115 47<br />

Order Number Index<br />

6986116 47<br />

6986117 47<br />

6986118 47<br />

6986119 47<br />

6986120 16<br />

6986121 16<br />

6986124 16<br />

6986125 16<br />

6986126 16<br />

6986127 16<br />

6986129 47<br />

6986130 47<br />

6986131 47<br />

6986132 47<br />

6986133 47<br />

6986137 47<br />

6986138 47<br />

6986139 16<br />

6988093 48<br />

6988094 46


Product Name Index<br />

AC membranes 74-75<br />

Air sampler 96<br />

All-Glass holder 23<br />

Anaerobic container 29<br />

Biosart 250 Funnels 8<br />

Biosart 100 Monitors 8<br />

Black membranes 74-75<br />

Blotting membranes 72<br />

Calibration unit 97<br />

Capsules 32-35,<br />

51, 79<br />

Cellulose acetate membranes 66<br />

Cellulose nitrate membranes 68-69,<br />

72,<br />

74-75<br />

Cellulose triacetate cassettes 61<br />

Cellulose triacetate ultrafilters 78<br />

Centrifugal units 55, 56<br />

Centrisart-C4 55<br />

Centrisart-C30 55<br />

Centrisart-I 56<br />

Collodion bags 59<br />

Colony counter 29<br />

Crossflow cassettes 60, 61<br />

Crossflow systems 60, 61<br />

Crossflow units 57<br />

Dialysis bags 59<br />

Diaphragm pump 46<br />

Disc filters 66-70,<br />

72-75,<br />

94<br />

Dosing syringe 16<br />

EasyFlow 58<br />

Forceps 17, 28,<br />

48<br />

Gelatin filter units 96<br />

Glass fibre filters 71<br />

Glass holders 21, 22,<br />

23<br />

Gray membranes 74-75<br />

Green membranes 75<br />

Gridded membranes 74-75<br />

H<strong>and</strong> vacuum pump 27<br />

Hydrophobic edge<br />

membranes 74-75<br />

Hydrophobic membranes 73<br />

Hydrosart cassettes 60, 61<br />

Impeller pumps 46, 47<br />

Incubator 29<br />

Integrity testers 98, 99<br />

Manifolds 25, 27<br />

MD 8 Air Sampler 96<br />

<strong>Membrane</strong> Adsorbers 85-94<br />

<strong>Membrane</strong> filters 66-70,<br />

72-75<br />

Micro collodion bags 59<br />

Midisart 2000 50<br />

Mini Cartridges 80, 81<br />

Mini Cartridge housings 43, 53<br />

Minicheck-BP 17<br />

Minisart syringe units 10-13<br />

Minisart-GF 13<br />

Minisart-HY 50<br />

Minisart-plus 12, 13<br />

Minisart-RC 10<br />

Minisart, st<strong>and</strong>ard 0.2 µm 12<br />

Minisart, st<strong>and</strong>ard, 0.45-5 µm 13<br />

Minisart-SRP 11<br />

NC-Extra 72<br />

Needles 17<br />

Nitrocellulose membranes 72<br />

Nutrient Pad Sets 76-77<br />

Peristaltic pump 63<br />

Perisys 70 Benchtop 63<br />

Polyamide membranes 70<br />

Polycarbonate holders 14, 15,<br />

20, 36<br />

Polyester separators 71<br />

Polyethersulfone cassettes 60, 61<br />

Polysulfone ultrafilters 78<br />

Prefilters 71<br />

Prefilter attachment 28<br />

Prefilter Capsules 34<br />

Prefilter Mini Cartridges 80<br />

Pressure filter holders 36-42,<br />

52, 94<br />

Pressure filtration units 31-35,<br />

50, 51,<br />

94<br />

Pressure hose 48<br />

Pressure tanks 46<br />

PTFE holders 14, 37<br />

PTFE membranes 73<br />

Pumps 27, 28,<br />

46, 63<br />

Quick connectors 48<br />

RC membrane filters 67<br />

Sartobind discs, modules 94<br />

Sartobran 300 Capsules 32<br />

Sartobran-P Capsules 33<br />

Sartocheck 3 99<br />

Sartocheck Junior BPplus 98<br />

Product Name Index<br />

Product name: Page: Product name: Page: Product name: Page:<br />

Sartoclean-CA Capsules 34<br />

Sartoclean-CA Mini Cartridges 80<br />

Sartoclean-GF Capsules 34<br />

Sartoclean-GF Mini Cartridges 80<br />

Sartocon Micro 57<br />

Sartocon Slice 60<br />

Sartocon 3 61<br />

Sartofluor Capsules 51<br />

Sartofluor Mini Cartridges 81<br />

Sartolab-P20 31<br />

Sartolab-P20plus 31<br />

Sartolab-V units 19<br />

Sartopore Capsules 79<br />

Sartopore Mini Cartridges 79<br />

Sartopure-GF Capsules 34<br />

Sartopure-GF Mini Cartridges 80<br />

Sartopure-PP Capsules 34<br />

Sartopure-PP Mini Cartridges 80<br />

School Kit 21<br />

Spiral flow cell 62<br />

Stainless steel holders 15, 24,<br />

25,<br />

38-42,<br />

52, 94<br />

Sterile venting units 50, 51<br />

Sterility test system 29<br />

Syringe filter holders 14, 15<br />

Syringe filtration units 10-13<br />

Syringes 17<br />

Suction flasks 27<br />

Three-way valve 17<br />

Ultrafilters 78<br />

Ultrafiltration units 55-58<br />

Ultrasart D20 59<br />

Ultrasart X 62<br />

Vacusart 28<br />

Vacuum filter holders 20-25<br />

Vacuum filtration units 19<br />

Vacuum hose 29<br />

Vacuum pumps 27, 28<br />

Vent filters 50, 51<br />

Wash water Capsules 35<br />

Water jet pump 27<br />

Woulffe's bottle 27<br />

For the purpose of product<br />

development we reserve the right to<br />

make changes without notice.


How to contact<br />

Sartorius <strong>Separation</strong> <strong>Technology</strong><br />

1<br />

Sartorius Australia Pty. Ltd.<br />

Unit 17, 104 Ferntree Gully Road,<br />

Oakleigh, Vic. 3166/Australia<br />

† (03) 5 44 0811,<br />

ƒ (03) 5 44 48 28<br />

Postal address: P.O. Box 84,<br />

Chadstone Centre, Chadstone,<br />

Vic. 3148/Australia<br />

Sartorius GmbH<br />

Franzosengraben 12,<br />

1030 Wien/Austria<br />

† (2 22) 7 96 57 63 14,<br />

ƒ (2 22) 7 96 57 63 44<br />

Sartorius Canada Inc.<br />

2179 Dunwin Drive, Units 4 + 5,<br />

Mississauga, Ontario L5L 1X2/Canada<br />

† (4 16) 5 69 79 77<br />

ƒ (416) 5 69 70 21<br />

Sartorius S. A.<br />

11, Avenue du 1er Mai,<br />

91127 Palaiseau Cedex/France<br />

† (1) 69 19 21 00,<br />

ƒ (1) 69 20 09 22<br />

Sartorius Limited<br />

Longmead Business Centre,<br />

Blenheim Road, Epsom, Surrey<br />

KT19 9QN/Great Britain<br />

† (13 72) 74 58 11<br />

ƒ (13 72) 72 07 99<br />

2<br />

3<br />

Write, call or fax the Sartorius Subsiduary in your Country.<br />

In other countries, contact the Agency<br />

or Representative <strong>for</strong> your country at<br />

the address on the adhesive label.<br />

Sartorius Biotech (I) Pvt. Ltd.<br />

3-4, 33-34, Patel Chinnigappa<br />

Industrial Estate, Sunkadakatte<br />

Magadi Main Road. Bangalore<br />

560 091/India<br />

† (91) 8 03 30 26 55<br />

ƒ (91) 8 03 30 57 81<br />

Sartorius<br />

5 Sussex Street, Dun Laoghaire,<br />

Co. Dublin/Irel<strong>and</strong><br />

† (1) 2 84 44 99<br />

ƒ (1) 2 84 45 99<br />

Sartorius S.p.A.<br />

Via Botticelli, 11/Rosso,<br />

50132 Firenze/Italy<br />

† (55) 57 75 41<br />

ƒ (55) 57 24 03<br />

Sartorius K.K., Tokyo, Japan<br />

No. 3 Hoya Building,<br />

8–17 Kamitakaido 1-chome,<br />

Suginami-ku, Tokyo 168/Japan<br />

† (3) 33 29 33 66<br />

ƒ (3) 33 29 28 82<br />

Sartorius-Instrumenten B.V.<br />

Edisonbaan 22–24,<br />

3439 MN Nieuwegein/Netherl<strong>and</strong>s<br />

† (34 02) 5 30 01<br />

ƒ (34 02) 5 29 17<br />

Postbus 1265, 3430 BG<br />

Nieuwegein/Netherl<strong>and</strong>s<br />

Or write, call or fax Company Headquarters in Germany:<br />

Sartorius AG<br />

37070 Goettingen, Germany<br />

† (551) 3 08-0,<br />

ƒ (551) 3 08-2 89<br />

Laboratory Filtration<br />

Sartorius Inc. Puerto Rico<br />

P.O. Box 6, Yauco, USA, Puerto Rico<br />

00698, Road 128, Intersection 376,<br />

Bda. Arturo Lluberas, Yauco, PR 00698<br />

† (8 09) 8 56 50 20<br />

ƒ (8 09) 8 56 12 92<br />

Sartorius S. A.<br />

Isabel Colbr<strong>and</strong> 10-12<br />

Planta 4, Nave 121<br />

28050 Madrid/Spain<br />

† (1) 3 58 85 66<br />

ƒ (1) 3 58 88 04<br />

Sartorius Seperasyon Teknolojileri Ltd.<br />

Sti. (Turkey)<br />

Ko‚suyolu, Mehmet Afkan Sk. No. 39,<br />

81020 Kadıkoy – Istanbul/Turkey<br />

† (2 16) 3 26 08 00,<br />

ƒ (2 16) 3 26 52 51<br />

Sartorius Corporation, USA<br />

<strong>Separation</strong>s Division,<br />

131 Heartl<strong>and</strong> Boulevard, Edgewood,<br />

New York 11717, USA<br />

† (8 00) 3 68 71 78<br />

ƒ (516) 2 54 42 49<br />

Status: March 1997<br />

Specifications subject to change without notice. Printed in Germany on paper bleached without any use of chlorine · M<br />

Publication No.: SL-0001-e97043 · Order No.: 85030-500-72

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!