6.2.2 Operating <strong>and</strong> calibration problems<br />

Problem Possible cause Correction<br />

No display appears when the unit is<br />

switched on<br />

Unit switched off for too short a period<br />

of time<br />

After switching the unit off, wait approx. 10 seconds<br />

before restarting<br />

Pressure display unstable Filter time constant too low Increase the filtering (→ section 4.7.7)<br />

Switching functions (relays) flutter Hysteresis too small Modify the threshold values (→ section 4.7.4)<br />

Increase the filtering (→ section 4.7.7)<br />

Pirani <strong>pressure</strong> display too high Pirani <strong>gauge</strong> contaminated Calibrate the measuring circuit<br />

Pirani measurement circuit cannot be calibrated<br />

Combination measurement plug-in<br />

board / <strong>gauge</strong> cable / <strong>gauge</strong> is not<br />

<strong>com</strong>patible<br />

Cold cathode <strong>pressure</strong> display too high Contaminated or moist connector insulation<br />

Clean the <strong>gauge</strong> (→ [4] ... [6])<br />

Replace the <strong>gauge</strong><br />

Select correct <strong>com</strong>bination (→ [3])<br />

Pirani <strong>gauge</strong> severely contaminated Clean the <strong>gauge</strong> (→ [4] ... [6])<br />

Replace the <strong>gauge</strong><br />

Clean or replace connector<br />

Humidity ( leak current) Keep humidity low, keep the unit switched on<br />

Cold cathode <strong>pressure</strong> display too low Cold cathode <strong>gauge</strong> contaminated Clean the <strong>gauge</strong> (→ [7] ... [10])<br />

«no P» is displayed No plug-in board has been installed Install the appropriate plug-in board 2)<br />

In<strong>com</strong>prehensible display Plug-in board not screwed down Tighten the screws<br />

Contacts contaminated / bent Clean / carefully straighten contacts 2)<br />

BG 800 <strong>300</strong> BE / A (0105) <strong>TPG</strong> <strong>300</strong> 39