PLS LB PM FEA, PM Photometry


Modern requirements for state<br />

of <strong>the</strong> art <strong>photometry</strong><br />

Drinking water & waste water analysis<br />

with excellence<br />

Gunter Decker<br />

Product Manager Food & Environmental Analysis<br />

Merck KGaA, Darmstadt<br />

Germany<br />

June 03, 2008 Complete solutions Page 2


ISO 14001: 2005_02<br />

4.5.1. Monitoring and Measurement<br />

Citation chapter 4.5.1:<br />

• “The organization shall ensure that calibrated or<br />

verified monitoring and measurement equipment<br />

is used and maintained i and shall retain associated<br />

records.”<br />

June 03, 2008 Complete solutions Page 3


What kind of Analytical Quality Assurance is<br />

required for getting g good results and meeting<br />

international standards?<br />

• Installation (e.g. IQ, OQ, PQ) must be done after receiving <strong>the</strong><br />

<strong>system</strong>.<br />

• Instruments must be regularly checked.<br />

• Accessories like e.g. certified glass ware must be re-checked.<br />

• Pipettes, balances and o<strong>the</strong>r equipment must be regularly checked<br />

and calibrated.<br />

• The staff performance incl. <strong>the</strong> chemicals and/or test kits must be<br />

checked by running some standards and testing <strong>the</strong> recovery rate.<br />

• Interfering substances for different samples must be checked.<br />

– Dilution, Spiking and e.g. multi Spiking <strong>the</strong>refore is required.<br />

June 03, 2008 Complete solutions Page 4


Analytical Quality Assurance<br />

SelfCheck of NOVA 30/60/400 & <strong>Pharo</strong> 100/300<br />

• When <strong>the</strong> instruments are switched on, automatically <strong>the</strong> photometers<br />

start with a SelfCheck procedure. The instruments are checking <strong>the</strong><br />

following steps:<br />

• Optical <strong>system</strong><br />

- lamp<br />

- shaft for cell tests incl. <strong>the</strong> barcode reading <strong>system</strong><br />

- reagent test kit shaft incl. <strong>the</strong> diodes for identifying <strong>the</strong> cells sizes<br />

- optical <strong>system</strong> incl. photomultiplier<br />

• Electronic <strong>system</strong> - Check of all electronic compounds<br />

• Check of <strong>the</strong> complete <strong>system</strong><br />

- <strong>the</strong> <strong>system</strong> is measuring against air and comparing <strong>the</strong> result against<br />

<strong>the</strong> stored (in-build internal) tolerance.<br />

• Only when all steps are successfully passed <strong>the</strong> instrument is going into<br />

<strong>the</strong> measuring mode or <strong>the</strong> measuring can be started.<br />

June 03, 2008 Complete solutions Page 5


Analytical Quality Assurance<br />

Hardware control – AQA level 1<br />

• How to test <strong>the</strong> instruments?<br />

- Therefore we created a test kits using round cells – PhotoCheck.<br />

• What are <strong>the</strong> advantages for customers using PhotoCheck?<br />

- The PhotoCheck solutions are checked by an reference<br />

spectrophotometer (primary standard) which is controlled by<br />

NIST standards.<br />

- PhotoCheck is traceable to NIST and has a shelf life of 3 years.<br />

• All lot certificates are in <strong>the</strong> Internet.<br />

• With PhotoCheck customers are able to check <strong>the</strong> hardware by<br />

<strong>the</strong>mselves. PhotoCheck meets <strong>the</strong> requirement of ISO 9001, ISO<br />

14001 & ISO 17025.<br />

June 03, 2008 Complete solutions Page 6


Printout of <strong>the</strong> PhotoCheck results from a<br />

<strong>Pharo</strong> / NOVA<br />

AQA-Check of meter AQA1<br />

29.08.2007 10:23<br />

Operator:<br />

AQA-Interval<br />

12 w<br />

AQA-Check AQA1 ok<br />

L-Check<br />

ok<br />

Test solution Meas. value Dimension Theor. val. Tolerance Result<br />

445-1 0.211 A 0.200 0.020 ok<br />

445-2 0.641 A 0.650 0.030 ok<br />

445-3 1.120120 A 1.100100 0.040040 ok<br />

445-4 1.520 A 1.500 0.050 ok<br />

525-1 0.270 A 0.280 0.020 ok<br />

525-2 0.698 A 0.710 0.030 ok<br />

525-3 1.130130 A 1.142142 0.040040 ok<br />

525-4 1.573 A 1.580 0.050 ok<br />

690-1 0.215 A 0.230 0.020 ok<br />

690-2 0.442 A 0.450 0.030 ok<br />

690-3 0.931 A 0.950 0.040040 ok<br />

690-4 1.480 A 1.460 0.050 ok<br />

June 03, 2008 Complete solutions Page 7


Documentation of <strong>the</strong> results using<br />

PhotoCheck<br />

• Control Chart<br />

Page 2 of <strong>the</strong> package insert<br />

from PhotoCheck<br />

• Add <strong>the</strong> measured absorbance<br />

b<br />

values and sign <strong>the</strong> sheet.<br />

• This meets <strong>the</strong> requirements<br />

according ISO 9001, 14001 &<br />

ISO 17025<br />

June 03, 2008 Complete solutions Page 8


Analytical Quality Assurance –<br />

<strong>System</strong> control – AQA level 2<br />

• What is necessary to monitor <strong>the</strong> complete <strong>system</strong>?<br />

• Customers should run standards.<br />

- To know that <strong>the</strong> handling is correct and that <strong>the</strong>y are able<br />

to get correct results<br />

- Frequency depends on <strong>the</strong> customer and should be<br />

reflecting <strong>the</strong> importance of <strong>the</strong> results<br />

• Customers should test <strong>the</strong> recovery rate to identify matrix<br />

influences<br />

• That’s <strong>the</strong> reason why we create CombiCheck<br />

June 03, 2008 Complete solutions Page 9


Analytical Quality Assurance –<br />

<strong>System</strong> control – AQA level 2<br />

• CombiCheck – <strong>the</strong> multi usable AQA tool to check <strong>the</strong><br />

complete <strong>system</strong>.<br />

• Two solutions are coming with <strong>the</strong> box<br />

- R-1 (ready-to-use standard)<br />

- R-2 (addition solution for spiking to test <strong>the</strong> recovery)<br />

• CombiCheck can be used for Spectroquant ® Picco, Multy,<br />

NOVA, <strong>Pharo</strong> and e.g. also spectrophotometers.<br />

• Advantage:<br />

The NOVA and <strong>Pharo</strong> photometers are pre-programmed with<br />

<strong>the</strong> target values of <strong>the</strong> R-1 solution from CombiCheck.<br />

June 03, 2008 Complete solutions Page 10


How is <strong>the</strong> CombiCheck reagent R-1 working?<br />

CombiCheck R-1 = multi parameter solution in one bottle<br />

• Instead of using <strong>the</strong> sample, add<br />

<strong>the</strong> same quantity of <strong>the</strong> R-1<br />

(standard solution) to <strong>the</strong> test and<br />

measure <strong>the</strong> result.<br />

• result = same as expected ☺<br />

protocol result in control chart<br />

• result ≠ different as expected <br />

- error in <strong>the</strong> <strong>system</strong>, check <strong>system</strong>,<br />

if error not found - service<br />

CombiCheck 10, R-1<br />

25.0 ± 6.0mg/l Cl<br />

80 ± 12mg/l COD<br />

4.00 ± 0.30mg/l NH4-N<br />

N<br />

2.50 ± 0.40mg/l NO3-N<br />

0.80 ± 0.0808 mg/l PO4-PP<br />

100 ± 15mg/l SO4<br />

The measuring errors from CombiCheck are less that 0.5 %<br />

June 03, 2008 Complete solutions Page 11


Control Chart of <strong>the</strong> CombiCheck<br />

to document <strong>the</strong> measured results of R-1<br />

m<br />

KONTROLLKARTE / CONTROL CHART<br />

COD 1.1454014540 10 - 150<br />

NOVA 60, serial no. 99450026<br />

CombiCheck<br />

Test: Meßbereich / Measuring range: mg/l<br />

Photometer:<br />

<br />

<br />

<br />

92<br />

80<br />

68<br />

mg/l<br />

mg/l<br />

mg/l<br />

x<br />

x<br />

x<br />

x<br />

x<br />

x<br />

x<br />

x<br />

x<br />

x<br />

x<br />

x<br />

x<br />

x<br />

x x x<br />

x<br />

x<br />

x<br />

Vertrauensb sbereich /<br />

Confidence<br />

interval<br />

Meßwert / Measurement<br />

value<br />

82 83 81 79 82 77 77 81 80 86 85 74 91 73 70 71 70 66 73 70<br />

Datum / Date<br />

Feb. 2005, 01.<br />

04. 05. 06. 07. 08. 11.<br />

12.<br />

13. 14.<br />

15. 18. 19. 20.<br />

21. 22. 25. 26. 27.<br />

28.<br />

Prüfer / Tester<br />

De De De De De De De De De De De XY XY XY XY XY XY XY XY XY<br />

Merck KGaA, D-64271 Darmstadt<br />

7.91146.7591/02-61874047 msp. 05/98<br />

June 03, 2008 Complete solutions Page 12


Analytical Quality Assurance<br />

Print out of a AQA2-Check of a <strong>Pharo</strong> / NOVA using<br />

CombiCheck or standard solutions<br />

AQA-Check of meter AQA2<br />

25.05.05 13:57<br />

Operator:<br />

AQA-Interval<br />

2 w<br />

Method Meas.value Dimension unit Theor. val. Tolerance Result<br />

14558 1.95 mg/l NH4-N 2.00 0.20 ok<br />

14555 5200 mg/l COD 5000 400 ok<br />

14540 78 mg/l COD 80 12 ok<br />

14540 65 mg/l COD 80 12 Error<br />

14563 2.4 mg/l NO3-N 2.5 0.4 ok<br />

14541 780 mg/l COD 750 75 ok<br />

14828 0.49 mg/l Cl2 0.50 0.05 ok<br />

14557 0.77 mg/l F- 0.80 0.07 ok<br />

June 03, 2008 Complete solutions Page 13


How is <strong>the</strong> CombiCheck reagent R-2 working?<br />

CombiCheck R-2 = multi parameter solution in one bottle<br />

• How to use CombiCheck: CombiCheck 10, R-2<br />

- First step: Measure <strong>the</strong> sample<br />

- Second step: take a <strong>new</strong> tube and add<br />

<strong>the</strong> sample and 0.1 ml of <strong>the</strong> R-2.<br />

Measure and compare <strong>the</strong> results.<br />

• result = in <strong>the</strong> tolerance range ☺<br />

- <strong>the</strong> recovery is nearly 100 %<br />

• result ≠ out of <strong>the</strong> tolerance range <br />

- <strong>the</strong> recovery is lower or higher 100 %<br />

– matrix effect, e.g. sample<br />

preparation necessary<br />

concentration and<br />

confidence interval<br />

Cat. no.<br />

25 ± 6 mg/l Cl - 1.1473014730<br />

30 ± 8 mg/l COD 1.14540<br />

3.00 ± 0.25 mg/l NH4-N N 1.1455814558<br />

0.60 ± 0.07 mg/l PO4-P<br />

0.30 ± 0.05 mg/l PO4-P<br />

1.14543<br />

1.1484814848<br />

sample = 72 mg/l COD<br />

sample + 0.1 ml = 105 mg/l COD<br />

For pipetting of 0.1 ml very precise pipettes are required.<br />

June 03, 2008 Complete solutions Page 14


Control Chart of <strong>the</strong> CombiCheck<br />

to document <strong>the</strong> measured results of R-2<br />

m<br />

KONTROLLKARTE / CONTROL CHART<br />

CombiCheck<br />

Test: Meßbereich / Measuring range: mg/l<br />

Photometer:<br />

COD 1.14540 10 - 150<br />

NOVA 60, serial no. 99450026<br />

<br />

38<br />

mg/l<br />

x<br />

x<br />

<br />

<br />

30<br />

22<br />

mg/l<br />

mg/l<br />

x<br />

x x<br />

x<br />

x<br />

x<br />

x<br />

x<br />

x<br />

x<br />

x<br />

x<br />

x<br />

x<br />

x<br />

x<br />

x<br />

x<br />

Vertrauensbereich<br />

/<br />

Confidence interva val<br />

Meßwert / Measurement 28 29 28 29 32 27 27 31 30 28 30 23 37 33 23 34 31 39 24 23<br />

value<br />

Feb. 2002, 01 04 05 06 07 08 11 12 13 14 15 18 19 20 21 22 25 26 27 28<br />

Datum / Date<br />

Prüfer / Tester<br />

. De De . De . De . De . De . De . De . De . De . De . XY . X.<br />

X.<br />

X.<br />

X.<br />

XY . . XY XY . XY .<br />

Y Y Y Y<br />

Merck KGaA, D-64271 Darmstadt<br />

7.91146.7591/02-61874047 msp. 05/98<br />

June 03, 2008 Complete solutions Page 15


Analytical Quality Assurance –<br />

<strong>System</strong> control – AQA level l 2<br />

What are <strong>the</strong> advantages for customers using CombiChecks?<br />

• CombiChecks are ready-to-use test kits for <strong>system</strong> control<br />

(shelf life 2 years).<br />

• Each pack CombiCheck contains R-1 (standard solution) and R-2<br />

(addition solution – spiking). They are prepared by <strong>the</strong> high-precision<br />

weighing of standard substances.<br />

• All analytical scales used are regularly calibrated and monitored against<br />

standards of <strong>the</strong> PTB (Physikalisch Technische Bundesanstalt, Germany)<br />

and NIST (National Institute t of Standards d and Technology, USA).<br />

• The standard substances can be traced back directly to e.g. <strong>the</strong><br />

respective NIST standard reference material - CombiChecks are<br />

traceable to NIST<br />

- Certificates of quality and <strong>the</strong> lot certificate are in <strong>the</strong> Internet.<br />

• With CombiCheck our customers are able to do a control of <strong>the</strong> whole<br />

<strong>system</strong> which is required from ISO 9001, ISO 14001 & ISO 17025.<br />

June 03, 2008 Complete solutions Page 16


Certificate of Quality for CombiCheck<br />

1.14676.000114676 Spectroquant ® CombiCheck 10<br />

zur Qualitätsprüfung photometrischer Methoden / for quality check of photometric methods /<br />

para comprobar la calidad de métodos fotométricos<br />

Spectroquant ® CombiCheck 10 ist ein gebrauchsfertiger Mehrelement-Standard zur Überprüfung der<br />

Spectroquant ® Testsätze. Jede Packung enthält eine Standardlösung und eine Additionslösung. Diese<br />

werden durch genaue Einwaage von Standardsubstanzen hergestellt. Alle Waagen werden regelmäßig gegen<br />

PTB (Physikalisch Technische Bundesanstalt, Deutschland) und NIST Standards (National Institute of Standards<br />

and Technology, USA) kalibriert und überwacht. Die Standardsubstanzen sind z.B. auf Standard Referenz<br />

Material von NIST rückführbar.<br />

Spectroquant ® CombiCheck 10 is a ready-to-use multiparameter standard for checking <strong>the</strong> Spectroquant ® test<br />

kits. Each pack contains one standard solution and one addition solution, which are prepared by <strong>the</strong> highprecision<br />

weighing in of standard substances. All analytical scales used are regularly calibrated and monitored<br />

against standards of <strong>the</strong> PTB (Physikalisch Technische Bundesanstalt, Germany) and NIST (National Institute of<br />

Standards and Technology, USA). The standard substances can be traced back directly to e.g. <strong>the</strong> respective<br />

NIST standard reference material.<br />

Spectroquant ® CombiCheck 10 es un patrón multielemento listo para el uso para comprobar los kits de tests<br />

Spectroquant ® . Cada envase contiene una solución patrón y una solución de adición. Éstas son preparadas<br />

pesando exactamente las sustancias patrón. Todas las balanzas se calibran y controlan regularmente frente a<br />

patrones PTB (Physikalisch-Technische Bundesanstalt, Alemania) y patrones NIST (National Institute of<br />

Standards and Technology, EUA). Las sustancias patrón son referibles p.ej., al material de referencia estándar<br />

de NIST.<br />

Parameter /<br />

Parameter /<br />

Parámetro<br />

Sollwert und Vertrauensbereich /<br />

concentration and confidence interval /<br />

valor nominal e intervalo de confianza<br />

Standardlösung Reagenz R-1<br />

Standard Solution Reagent R-1<br />

Solución patrón del reactivo R-1<br />

einsetzbar für Art. /<br />

can be used for test kits /<br />

aplicable para art.<br />

Ammonium / Ammonium / 4.00 ± 0.30 mg/l (de) NH4-N 1.14558.0001<br />

Amonio<br />

Chlorid / Chloride / Cloruros 25 ±6mg/l(de) Cl 1.14730.000114730 0001<br />

CSB / COD / DQO 80<br />

80<br />

80<br />

Nitrat / Nitrate / Nitratos 2.50<br />

2.50<br />

Phosphor / Phosphorous /<br />

Fósforo<br />

Sulfat / Sulfate / Sulfatos 100<br />

100<br />

100<br />

± 12 mg/l CSB / COD / de DQO<br />

± 12 mg/l CSB / COD / de DQO<br />

± 12 mg/l CSB / COD / de DQO<br />

± 0.25 mg/l (de) NO3-N<br />

± 0.25 mg/l (de) NO3-N<br />

1.09772.0001<br />

1.14540.0001<br />

1.18751.0001<br />

1.14556.0001<br />

1.14773.0001<br />

• Certificates of Quality<br />

are available for all<br />

CombiChecks<br />

Additionslösung Reagenz R-2 2( (zur Aufstockung der Probe)<br />

0.80 ± 0.08 mg/l (de) PO4-P<br />

1.14543.0001<br />

0.80 ± 0.08 mg/l (de) PO4-P<br />

1.14848.0001<br />

± 15 mg/l (de) SO4<br />

1.00617.0001<br />

± 15 mg/l (de) SO4<br />

1.14548.0001<br />

± 15 mg/l (de) SO4<br />

1.14791.0001<br />

• This certificate shows<br />

<strong>the</strong> quality of <strong>the</strong><br />

3.00 ± 0.25 mg/l (de) NH4-N 1.14558.0001 CombiCheck and is a<br />

± 8 mg/l CSB / COD / de DQO<br />

1.09772.0001<br />

basic information to<br />

± 8 mg/l CSB / COD / de DQO<br />

1.14540.0001<br />

± 8 mg/l CSB / COD / de DQO<br />

1.18751.0001<br />

± 0.20 mg/l (de) NO3-N<br />

1.14556.0001<br />

± 0.40 mg/l (de) NO3-N<br />

1.14773.0001 meet <strong>the</strong> quality<br />

0.60 ± 0.07 mg/l (de) PO4-P<br />

1.14543.0001<br />

0.30 ± 0.05 mg/l (de) PO4-P<br />

1.14848.0001<br />

± 15 mg/l (de) SO4<br />

1.00617.0001<br />

± 5 mg/l (de) SO4<br />

1.14548.0001<br />

± 10 mg/l (de) SO4<br />

1.14791.0001<br />

criteria of <strong>the</strong> ISO 9001,<br />

ISO 14001 and o<strong>the</strong>r<br />

regulations.<br />

Addition Solution Reagent R-2 (for spiking <strong>the</strong> sample)<br />

Solución de adición del reactivo R-2 (para aumentar la concentración de la muestra)<br />

Ammonium / Ammonium /<br />

Amonio<br />

Chlorid / Chloride / Cloruros 25 ± 6 mg/l (de) Cl 1.14730.0001<br />

CSB / COD / DQO 45<br />

30<br />

45<br />

Nitrat / Nitrate / Nitratos 1.50<br />

2.00<br />

Phosphor / Phosphorous /<br />

Fósforo<br />

Sulfat / Sulfate / Sulfatos 100<br />

40<br />

80<br />

Darmstadt, 21.09.2004<br />

g<br />

Merck KGaA, Qualitätskontrolle / Quality control / Laborleiter / Head of Lab. /<br />

Control de calidad Jefe de laboratorio<br />

http://<strong>photometry</strong>.merck.de<br />

• All CombiChecks are<br />

traced back to NIST<br />

June 03, 2008 Complete solutions Page 17


Lot Certificate for CombiCheck<br />

Chargenzert if ikat<br />

Lot Certificate • Certificado de lote<br />

Spectroquant ® CombiCheck 20<br />

zur Qualitätsprüfung photometrischer Methoden / for quality check of photometric methods /<br />

para comprobar la calidad de métodos fotométricos<br />

1.14675.0001 Charge / Lot / Lote: OC499922<br />

Spectroquant ® CombiCheck 20 ist ein gebrauchsfertiger Mehrelement-Standard zur Überprüfung der<br />

Spectroquant ® Testsätze. Jede Packung enthält eine Standardlösung und eine Additionslösung. Diese<br />

werden durch genaue Einwaage von Standardsubstanzen hergestellt. Alle Waagen werden regelmäßig gegen<br />

PTB (Physikalisch Technische Bundesanstalt, Deutschland) und NIST Standards (National Institute of<br />

Standards and Technology, USA) kalibriert und überwacht. Die Standardsubstanzen sind z.B. auf Standard<br />

Referenz Material von NIST rückführbar.<br />

Spectroquant ® CombiCheck 20 is a ready-to-use multiparameter standard for checking <strong>the</strong> Spectroquant ® test<br />

kits. Each pack contains one standard solution and one addition solution, which are prepared by <strong>the</strong> highprecision<br />

weighing in of standard substances. All analytical scales used are regularly calibrated and monitored<br />

against standards of <strong>the</strong> PTB (Physikalisch Technische Bundesanstalt, Germany) and NIST (National Institute<br />

of Standards and Technology, USA). The standard substances can be traced back directly to e.g. <strong>the</strong> respective<br />

NIST standard reference material.<br />

Spectroquant ® CombiCheck 20 es un patrón multielemento listo para el uso para comprobar los kits de tests<br />

Spectroquant ® . Cada envase contiene una solución patrón y una solución de adición. Éstas son preparadas<br />

pesando exactamente las sustancias patrón. Todas las balanzas se calibran y controlan regularmente frente a<br />

patrones PTB (Physikalisch-Technische h h i h Bundesanstalt, Alemania) y patrones NIST (National Institute t of<br />

Standards and Technology, EUA). Las sustancias patrón son referibles p.ej., al material de referencia estándar<br />

de NIST.<br />

Parameter / Sollwert und Vertrauensbereich / Messergebnis /<br />

eingesetzte Testsätze<br />

Parameter / concentration and confidence interval /<br />

Result /<br />

zur Überprüfung /<br />

Parámetro valor nominal e intervalo de confianza Resultado<br />

used test kits for checking /<br />

kits de tests usados para la<br />

comprobación<br />

Standardlösung Reagenz R-1<br />

Standard Solution Reagent R-1<br />

Solución patrón del reactivo R-1<br />

Ammonium / 12.0 ± 1.0 mg/l NH4-N /<br />

mg/l NH4-N /<br />

1.14544.0001<br />

Ammonium /<br />

de NH4-N /<br />

de NH4-N /<br />

Amonio<br />

Chlorid / Chloride / 60 ± 10 mg/l Cl /<br />

mg/l Cl /<br />

1.14730.0001<br />

Cloruros<br />

de Cl<br />

de Cl<br />

CSB / COD / DQO 750 ± 75 mg/l CSB / COD / de mg/l CSB / COD /<br />

1.14541.0001<br />

DQO<br />

de DQO<br />

Nitrat / Nitrate / 9.0 ± 0.9 mg/l NO3-N /<br />

mg/l NO3-N /<br />

1.14563.0001<br />

Nitratos<br />

de NO3-N<br />

de NO3-N<br />

Phosphor /<br />

8.0 ± 0.7 mg/l PO4-P /<br />

mg/l PO4-P /<br />

1.14729.0001<br />

Phosphorous /<br />

de PO4-P<br />

de PO4-P<br />

Fósforo<br />

Sulfat / Sulfate / 500 ± 75 mg/l SO4 /<br />

mg/l SO4 /<br />

1.14564.0001<br />

Sulfatos<br />

de SO4<br />

de SO4<br />

Additionslösung Reagenz R-2 (zur Aufstockung der Probe)<br />

Addition Solution Reagent R-2 2(Spike sample)<br />

Solución de adición del reactivo R-2 (para aumentar la concentración de la muestra)<br />

Ammonium /<br />

8.0 ± 0.8 mg/l NH4-N /<br />

mg/l NH4-N /<br />

1.14544.0001<br />

Ammonium /<br />

de NH4-N /<br />

de NH4-N /<br />

Amonio<br />

Chlorid / Chloride / 40 ± 7 mg/l Cl /<br />

mg/l Cl /<br />

1.14730.0001<br />

Cloruros<br />

de Cl<br />

de Cl<br />

CSB / COD / DQO 200 ± 40 mg/l CSB / COD / de mg/l CSB / COD /<br />

1.14541.0001<br />

DQO<br />

de DQO<br />

Nitrat / Nitrate / 7.5 ± 0.8 mg/l NO3-N /<br />

mg/l NO3-N /<br />

1.14563.0001<br />

Nitratos<br />

de NO3-N<br />

de NO3-N<br />

Phosphor /<br />

5.0 ± 0.5 mg/l PO4-P /<br />

mg/l PO4-P /<br />

1.14729.0001<br />

Phosphorous /<br />

de PO4-P<br />

de PO4-P<br />

Fósforo<br />

Sulfat / Sulfate / 150 ± 30 mg/l SO4 /<br />

mg/l SO4 /<br />

1.14564.0001<br />

Sulfatos<br />

de SO4<br />

de SO4<br />

• Lot certificates are<br />

available for all<br />

CombiCheck test kits<br />

• CombiChecks are<br />

traceable to NIST.<br />

Mindestens verwendbar bis: / Minimum shelf life: /<br />

Utilizable como mínimo hasta el: 25.08.2007<br />

Datum der Freigabe: / Date of release: /<br />

Fecha de liberación: 26.08.2004<br />

Merck KGaA, Qualitätskontrolle / Quality control / Laborleiter / Head of Lab. /<br />

Control de calidad Jefe de laboratorio<br />

http://<strong>photometry</strong>.merck.de<br />

June 03, 2008 Complete solutions Page 18


Analytical Quality Assurance<br />

Printout of measured results of <strong>Pharo</strong> / NOVA<br />

after using PhotoCheck and CombiCheck<br />

no. of meas. I.D. Date time meas. Dimension unit AQA<br />

value number value<br />

001 14739 10.08.0708 07 09:03:50 1.2 mg/l NH4-NN AQA2<br />

002 14558 10.08.07 09:50:05 7.7 mg/l NH4-N AQA2<br />

003 14560 10.08.07 10:31:35 21 mg/l COD AQA2<br />

004 14541 10.08.07 10:36:13 280 mg/l COD AQA2<br />

005 14541 10.08.07 10:38:39 670 mg/l COD AQA2<br />

006 14555 10.08.07 10:40:21 5280 mg/l COD AQA2<br />

007 14543 10.08.07 11:30:53 0.35 mg/l PO4-P AQA2<br />

008 14729 10.08.07 11:46:47 12.5 mg/l PO4-P AQA2<br />

009 00598 10.08.07 14:56:33 0.355 mg/l Cl2 AQA1<br />

010 00602 10.08.0708 07 14:58:36 0.709 mg/l Cl2 AQA1<br />

011 14761 11.08.07 10:05:44 0.029 mg/l Fe AQA2<br />

012 14758 11.08.07 11:23:53 0.043 mg/l Cr AQA2<br />

013 00856 11.08.07 11:56:28 0.052 mg/l PHE AQA1<br />

014 14563 11.08.07 12:02:33 3.40 mg/l NO3-N N AQA2<br />

015 14558 11.08.07 12:09:42 6.9 mg/l NH4-N AQA2<br />

June 03, 2008 Complete solutions Page 19


Spectroquant ® Multy Colorimeter<br />

SQ Multy Colorimeter offers an easy start t into water analysis:<br />

• Covers all important parameters for waste water and drinking water analysis -<br />

more than 120 methods are pre-calibrated<br />

• Complete, portable and flexible solution – comes with a case and can be used<br />

with mains or power pack (rechargeable batteries), all accessories are<br />

included<br />

• 16 and 24 mm round cells can be used for measuring<br />

• Use of ready-to-use Standards (CombiCheck) possible<br />

• Documentation of <strong>the</strong> results via PC or printer<br />

• Method update via internet t is free of charge<br />

• Different citation forms are integrated<br />

• Use <strong>the</strong> Spectroquant ® tests<br />

• New: Programming of own methods<br />

is possible (update)<br />

NEW<br />

Multy<br />

June 03, 2008 Complete solutions Page 20


SQ Multy Colorimeter - dual beam technology:<br />

Detector<br />

Measuring beam<br />

Sample cell<br />

Interference<br />

filter<br />

Reference<br />

beam<br />

Detector<br />

LED<br />

Permeable<br />

mirror<br />

June 03, 2008 Complete solutions Page 21


Spectroquant ® NOVA photometers<br />

• Easy to handle because of bar-code reading and stored blank values<br />

• Automatically identification <strong>system</strong> for cell size (10, 20 and 50 mm)<br />

• Analytical Quality Assurance <strong>system</strong> available<br />

• Flexible because of <strong>the</strong> use of cell and reagent tests and <strong>the</strong>refore low,<br />

medium and higher levels can be measured<br />

• Documentation of <strong>the</strong> results via PC or printer<br />

• Open and flexible <strong>system</strong> for <strong>the</strong> future<br />

(170 methods already pre-calibrated)<br />

NOVA 60 and NOVA 400<br />

June 03, 2008 Complete solutions Page 22


How to measure tests on NOVA<br />

labelled?<br />

inserted?<br />

recognized?<br />

Unique: barcode reading<br />

and automatic cell size<br />

recognition for reagent<br />

tests !!!<br />

Measuring + result shown<br />

on display<br />

June 03, 2008 Complete solutions Page 23


How to measure cell tests<br />

(all NOVA photometers)<br />

labelled?<br />

inserted?<br />

recognized?<br />

Exception<br />

no<br />

manual<br />

input<br />

of<br />

method<br />

yes<br />

Measuring + results shown in display<br />

June 03, 2008 Complete solutions Page 24


Optical path NOVA 30 and 60<br />

Diaphragm<br />

Round Cell<br />

Rectangular cells<br />

Mirrors<br />

(partly light permeable)<br />

Lamp<br />

Lens<br />

Lens<br />

Full light is<br />

passing <strong>the</strong> cells<br />

Filtermoduls with Measuring<br />

Elements (each with 4 Filters)<br />

“Unused” light, which would<br />

o<strong>the</strong>rwise be wasted is used as<br />

reference beam<br />

The monochromator<br />

(filters) is at <strong>the</strong> end, just<br />

before <strong>the</strong> detector<br />

June 03, 2008 Complete solutions Page 25


Optical path NOVA 30 and 60<br />

Lamp<br />

Diaphragm<br />

Lens<br />

Round Cell<br />

Lens<br />

Rectangu<br />

lar cells<br />

Mirrors<br />

(partly light permeable)<br />

Filtermoduls with Measuring<br />

Elements (each with 4 Filters)<br />

Advantage of NOVA:<br />

• Very robust, because no<br />

moving gparts<br />

• Fast measurement - at all<br />

wavelengths at <strong>the</strong> same<br />

time<br />

• Full light intensity passes cell,<br />

light is split at <strong>the</strong> end<br />

-No problems with straylight,<br />

even in full sun light no lid is<br />

necessary<br />

• Measures easily high<br />

absorbance<br />

• Only specific wavelengths<br />

available<br />

June 03, 2008 Complete solutions Page 26


The <strong>new</strong> Spectroquant ® <strong>Pharo</strong> line:<br />

NEW<br />

Cat. No.:<br />

Name:<br />

1.00706.0001 Spectroquant ® Spectrophotometer <strong>Pharo</strong> 100<br />

320 – 1100 nm, 4 nm<br />

1.00707.0001 Spectroquant ® UV-VIS Spectrophotometer <strong>Pharo</strong> 300<br />

190 – 1100 nm, 4 nm<br />

June 03, 2008 Complete solutions Page 27


Single beam spectrometer<br />

Single beam<br />

In single beam instruments you have to measure<br />

<strong>the</strong> reference before you measure your sample!<br />

3<br />

1<br />

2<br />

This device separates <strong>the</strong> light in<br />

The emitted light is focused via beams<br />

<strong>the</strong> entrance<br />

of different<br />

wavelengths The non and absorbed focuses light <strong>the</strong> passes single beams <strong>the</strong> sample<br />

slit to <strong>the</strong> dispersion device.<br />

successive and reached through <strong>the</strong> <strong>the</strong> detector. exit slit on <strong>the</strong> sample.<br />

June 03, 2008 Complete solutions Page 28


Optical path of SQ <strong>Pharo</strong> 100:<br />

Entrance slit<br />

Halogen lamp<br />

Lens<br />

Lid<br />

Round cell<br />

Optical Filter<br />

Monochromator<br />

Shutter<br />

Rectangular cell<br />

• Light of one wavelength is passing <strong>the</strong> cell<br />

Lens<br />

- less energy, not likely to destroy biological samples<br />

- stray light is a more important, lid necessary<br />

Photodiode<br />

• Shutter: cutting of light beam, stray light (ambient light) can be measured<br />

and compensated.<br />

• Optical filter: reduction of stray light through suppression of light of <strong>the</strong><br />

„wrong“ wavelength.<br />

June 03, 2008 Complete solutions Page 29


Optical path of SQ <strong>Pharo</strong> 300:<br />

Entrance slit<br />

Xenon flash lamp<br />

Lens<br />

Lid<br />

Round cell<br />

Optical Filter<br />

Monochromator<br />

Rectangular cell<br />

Lens<br />

Photodiode<br />

• Similar construction, but without shutter<br />

June 03, 2008 Complete solutions Page 30


Tungsten-Halogen lamp<br />

Intensity Spectrum of <strong>the</strong> Tungsten-Halogen Lamp<br />

• The beside picture shows a typical<br />

lamp. They are easy to handle,<br />

relatively cheap and have a good<br />

intensity in <strong>the</strong> visible range.<br />

The intensity in UV range is very low.<br />

Therefore <strong>the</strong> tungsten-halogen<br />

lamps are only used in <strong>the</strong> VIS range.<br />

• The lifespan is ~1000h.<br />

• Customers can changed <strong>the</strong> lamp<br />

easily. Prices are ~ 60-100 EUR.<br />

• Tungsten-Halogen lamp is used in<br />

<strong>Pharo</strong> 100.<br />

June 03, 2008 Complete solutions Page 31


Pulsed xenon lamp<br />

Intensity Spectrum of <strong>the</strong> Xenon Lamp<br />

• The pictures show some typical<br />

types of <strong>the</strong>se lamps.<br />

• They have good intensity in UV and<br />

VIS range.<br />

• They need no warm up time and<br />

produce only few heat, but <strong>the</strong>y<br />

have a medium noise.<br />

• They are quite expensive<br />

(~800 EUR) but have a very<br />

long lifespan (5 x 10 8 flashes),<br />

correspond to min. 13.000h<br />

continuous operation.<br />

• Service required for lamp exchange.<br />

• Pulsed xenon lamp is used in<br />

<strong>Pharo</strong> 300.<br />

June 03, 2008 Complete solutions Page 32


Deuterium lamp<br />

• The beside picture shows a typical<br />

deuterium lamp.<br />

• Good intensity in UV range.<br />

• Average lifespan of ~2000h.<br />

• Price is about ~400-600 EUR,<br />

means ~half of xenon flash lamp.<br />

• A combination of Tungsten-Halogen<br />

and Deuterium lamp is used in<br />

most of <strong>the</strong> commercial available<br />

UV-VIS spectrophotometers.<br />

Intensity Spectrum of <strong>the</strong> Deuterium Arc Lamp<br />

June 03, 2008 Complete solutions Page 33


The features of <strong>the</strong> Spectroquant ® <strong>Pharo</strong> line:<br />

Actual date and time<br />

Access to all<br />

Easy access to<br />

Photometer<br />

general settings AQA functions information<br />

(software version)<br />

Easy to handle<br />

through intuitive<br />

menu navigation<br />

• User are guided<br />

through <strong>the</strong> menu<br />

• In many cases<br />

<strong>the</strong>y only have to<br />

select between<br />

given possibilities<br />

June 03, 2008 Complete solutions Page 34


Measuring concentration:<br />

Result in<br />

big, easy to<br />

read letters<br />

5 digit cat.no.<br />

Type of cell used<br />

Easy access to<br />

change citation<br />

form<br />

Parameter<br />

name<br />

Measuring<br />

range<br />

Easy access to<br />

change <strong>the</strong> unit<br />

(e.g. ppm, mmol/l)<br />

June 03, 2008 Complete solutions Page 35


Measuring concentration:<br />

Easy control functions for users:<br />

Dilution factor<br />

(settings done<br />

by user)<br />

Turbidity<br />

correction<br />

If an own blank is used, it<br />

is shown in <strong>the</strong> display<br />

Batch of blank can be<br />

stored as well<br />

Display of absorbance<br />

possible (no separate<br />

measurement<br />

necessary)<br />

June 03, 2008 Complete solutions Page 36


The features of <strong>the</strong> Spectroquant ® <strong>Pharo</strong> line:<br />

Easy handling through pop up menus<br />

June 03, 2008 Complete solutions Page 37


How to measure tests on <strong>Pharo</strong> (and NOVA)<br />

Automatic<br />

selection of<br />

measuring range<br />

June 03, 2008 Complete solutions Page 38


How to measure tests on <strong>Pharo</strong><br />

Adapterless and automatic cell size recognition<br />

√<br />

X X<br />

√ √ X<br />

√ √ √<br />

1cm 2cm 5cm<br />

June 03, 2008 Complete solutions Page 39


The features of <strong>the</strong> Spectroquant ® <strong>Pharo</strong> line:<br />

User management:<br />

User rights<br />

in detail<br />

NEW<br />

June 03, 2008 Complete solutions Page 40


The features of <strong>the</strong> Spectroquant ® <strong>Pharo</strong> line:<br />

• Extended instrument supported AQA<br />

- AQA1: checking <strong>the</strong> photometer<br />

- minimum with PhotoCheck<br />

- can be enlarged with e.g. CertiPUR<br />

® UV/VIS standards<br />

- AQA2: checking <strong>the</strong> complete <strong>system</strong><br />

- with Standards, e.g. CombiCheck<br />

- MatrixCheck: checking <strong>the</strong> influence of <strong>the</strong> sample<br />

(interferences)<br />

- by using CombiCheck R2 solution or<br />

- multiple dilution or stocking up<br />

- guided multiple stocking and dilution procedure<br />

June 03, 2008 Complete solutions Page 41


The need for MatrixCheck<br />

• What is <strong>the</strong> matrix effect?<br />

– Besides <strong>the</strong> parameter, which needs to be analysed,<br />

<strong>the</strong> sample may include o<strong>the</strong>r substances, which can<br />

interfere with <strong>the</strong> current form of analysis.<br />

– Details are described in <strong>the</strong> package inserts in <strong>the</strong><br />

chapter “influence of foreign substances”.<br />

– The effect of several interferences can only be checked<br />

by diluting or spiking <strong>the</strong> sample.<br />

June 03, 2008 Complete solutions Page 42


MatrixCheck by spiking<br />

on SQ <strong>Pharo</strong><br />

NEW<br />

• To detect a matrix effect, you<br />

can spike your sample with a<br />

selected standard solution.<br />

• Therefore you have to prepare<br />

up to three solutions with<br />

different mixing ratio of sample<br />

and standard solution.<br />

• The left round cell includes just<br />

<strong>the</strong> sample (after colour<br />

reaction), <strong>the</strong> o<strong>the</strong>rs includes <strong>the</strong><br />

sample and standard solution in<br />

a predefined d mixing i ratio.<br />

• First <strong>the</strong> sample is measured,<br />

before <strong>the</strong> MatrixCheck is<br />

performed.<br />

June 03, 2008 Complete solutions Page 43


MatrixCheck by spiking<br />

on SQ <strong>Pharo</strong><br />

NEW<br />

• After supplying <strong>the</strong> mixing ration<br />

data, <strong>the</strong> meter is calculating <strong>the</strong><br />

target value for each round cell.<br />

• Each cell has to be measured<br />

one after <strong>the</strong> o<strong>the</strong>r.<br />

• The actual - measured - value is<br />

displayed.<br />

• The meter is calculating <strong>the</strong><br />

recovery.<br />

• If <strong>the</strong> recovery is in a range of<br />

e.g. ± 10 %, <strong>the</strong>re are no<br />

interferences coming from <strong>the</strong><br />

matrix.<br />

June 03, 2008 Complete solutions Page 44


MatrixCheck by dilution<br />

• An o<strong>the</strong>r method to detect matrix<br />

effects is to dilute <strong>the</strong> sample<br />

with a predefined volume of<br />

distilled water.<br />

• Up to three predefined dilutions<br />

are required.<br />

• The left round cell includes just<br />

<strong>the</strong> sample (after colour<br />

reaction), <strong>the</strong> o<strong>the</strong>rs includes <strong>the</strong><br />

sample and distilled water in a<br />

predefined mixing ratio<br />

• First <strong>the</strong> sample is measured,<br />

before <strong>the</strong> MatrixCheck by<br />

dilution is performed.<br />

June 03, 2008 Complete solutions Page 45


MatrixCheck by dilution<br />

• After supplying <strong>the</strong> mixing ratio<br />

data for dilution, <strong>the</strong> meter is<br />

calculating <strong>the</strong> target value for<br />

each cell.<br />

• Each cell has to be measured<br />

one after <strong>the</strong> o<strong>the</strong>r.<br />

• The actual – measured - value is<br />

displayed.<br />

• The meter is calculating <strong>the</strong><br />

recovery.<br />

• If <strong>the</strong> recovery is in a range of<br />

e.g. ± 10 %, <strong>the</strong>re are no<br />

interferences coming from <strong>the</strong><br />

matrix. (tolerance can be set by<br />

<strong>the</strong> user)<br />

June 03, 2008 Complete solutions Page 46


Cells<br />

• Our standard 1,2 and 5cm cells are made from optical glass and are tested<br />

for all our Spectroquant ® test kits (for use on SQ NOVA, SQ <strong>Pharo</strong> etc.).<br />

• The <strong>new</strong>er ones are marked, because it is very important to use <strong>the</strong> same<br />

orientation for zero adjustment and measurements.<br />

• For <strong>the</strong> UV range we introduced a 1cm Quartz cell (item 1.00784.0001)<br />

NEW<br />

June 03, 2008 Complete solutions Page 47


Cells<br />

NEW<br />

50-mm halfmicro cell (item 1.73502.0001)<br />

Advantage:<br />

• Minimum quantity for <strong>the</strong> “normal” 50-mm<br />

rectangular cell is approx. 10 ml. Therefore<br />

<strong>the</strong> volume of sample and reagents have to<br />

be doubled in <strong>the</strong> past for some test kits<br />

(approx. 10 items).<br />

• The <strong>new</strong> 50-mm halfmicro cell works with 5<br />

ml volume – doubling not anymore required.<br />

• User can use <strong>the</strong> full number of<br />

determinations and saves money<br />

June 03, 2008 Complete solutions Page 48


Spectroquant ® Test Kits - Environmental Testing<br />

• Comprehensive product<br />

range of<br />

» Spectroquant ® cell tests<br />

» Spectroquant ® reagent<br />

tests<br />

• around 150 tests<br />

• for water, waste water,<br />

environmental analysis,<br />

in-process control<br />

June 03, 2008 Complete solutions Page 49


Spectroquant ® Cell Tests<br />

• Highly convenient and user-friendly. Virtually<br />

all of <strong>the</strong> reagents necessary for <strong>the</strong> analysis<br />

are contained in <strong>the</strong> cells.<br />

• The packaging concept offers <strong>the</strong> cells ideal<br />

protection from fluctuations in temperature<br />

and moisture during transport.<br />

• The external label shows information about<br />

<strong>the</strong> content and gives valuable information<br />

regarding safety and risk issues, <strong>the</strong> batch<br />

number, and <strong>the</strong> shelf-life of <strong>the</strong> test kit (up to<br />

three years, without refrigerator).<br />

• The cell label shows <strong>the</strong> parameters, item<br />

number and contains <strong>the</strong> bar code, which<br />

ensures that no mistakes can occur.<br />

• Reagents that must be added for <strong>the</strong><br />

determination or for <strong>the</strong> pre-treatment of <strong>the</strong><br />

samples are contained in easy-to handle<br />

dose-metering bottles.<br />

June 03, 2008 Complete solutions Page 50


Spectroquant ® Reagent Tests<br />

• Inexpensive, ready-to-use reagent mixtures in<br />

a user-friendly design.<br />

• Selection of <strong>the</strong> appropriate p cell format makes<br />

it easy to vary <strong>the</strong> measuring range as you<br />

require.<br />

• The AutoSelector contains <strong>the</strong> bar code for<br />

<strong>the</strong> automatic selection of <strong>the</strong> correct method<br />

• Ideal for e.g. drinking water analysis: <strong>the</strong><br />

lower limit of detection (LLD) can be extended<br />

into <strong>the</strong> ppb range by <strong>the</strong> use of 50 mm cells.<br />

• The reagents are ei<strong>the</strong>r in liquid or in powder<br />

form and are added to <strong>the</strong> sample using a<br />

dropper, pipette, or metering spoon.<br />

• Extremely long shelf-life (up to three years)<br />

and <strong>the</strong> avoidance of <strong>the</strong> necessity for<br />

storage in a refrigerator.<br />

June 03, 2008 Complete solutions Page 51


Lot Certificate for Spectroquant ® Tests<br />

►<br />

• Lot certificates are available for<br />

all photometric test kits<br />

• As Merck is a certified supplier,<br />

<strong>the</strong> lot certificates could be used<br />

from Merck. Means:<br />

– We are doing <strong>the</strong> quality control<br />

– Customers have to do just a<br />

quality check of <strong>the</strong> incoming test<br />

kit when such certificates t are<br />

available<br />

• Means less time and money<br />

– Meets ISO 9001 - 4.10 + 4.11<br />

– Meets ISO 14001 - 4.5.1 + 4.5.3<br />

• http://<strong>photometry</strong>.merck.de<br />

June 03, 2008 Complete solutions Page 52


Certificate of Quality for Spectroquant ® Tests<br />

Eignung der Spectroquant ® Testsätze zur Selbstüberwachung<br />

Applicability of Spectroquant ® Test Kits for Self-Monitoring.<br />

Aptitud de los equipos de ensayo Spectroquant ® para autovigilancia<br />

• Certificates of Quality are available for<br />

all photometric test kits<br />

• This certificates give <strong>the</strong> information<br />

summarising several lots. Customers<br />

Spectroquant can see all quality details of our test<br />

® COD Cell Test, Cat. No. 1.14541<br />

25 – 1500 mg/l CSB / COD / kits and <strong>the</strong> precise how we are able<br />

DQO<br />

17 mg/l CSB / COD / DQO<br />

to produce. This information is<br />

6.9 mg/l CSB / COD / DQO<br />

necessary for <strong>the</strong> official use of <strong>the</strong><br />

18 mg/l CSB / COD / DQO<br />

method.<br />

Die Verfahrenskenndaten für den unten genannten Testsatz wurden gemäß ISO 8466-1 und DIN 38402 A51<br />

”Kalibrierung von Analysenverfahren” bei der Produktionsendkontrolle ermittelt.<br />

The test characteristic data of <strong>the</strong> procedure of <strong>the</strong> following test kit were determined in accordance with ISO<br />

8466-1 and DIN 38402 A51 ”Calibration of analysis methods” during <strong>the</strong> production control process.<br />

Los datos característicos del procedimiento para el equipo de ensayo abajo citado se determinaron según ISO<br />

8466-1 y DIN 38402 A51 ”Calibración de procedimientos analíticos” durante el control final de producción.<br />

Spectroquant ® CSB-Küvettentest, Art.-Nr. 1.14541<br />

Spectroquant ® Test en cubetas DQO, Art. Núm. 1.14541<br />

Messbereich / Measuring Range / Intervalo de medida<br />

Empfindlichkeit: 0.010 E (Extinktion) =<br />

Sensitivity: 0.010 A (absorbance) =<br />

Sensibilidad: 0.010 A (absorbancia) =<br />

Nachweisgrenze<br />

Lower Limit of Detection (LLD)<br />

Límite de detección<br />

<strong>Best</strong>immungsgrenze<br />

Method Detection Limit (MDL)<br />

Límite de determinación<br />

Vertrauensbereich (95 % Wahrscheinlichkeit) (Mittelwert aller Chargen)<br />

Confidence Interval (P = 95 %) (average value of lots)<br />

Intervalo de confianza (95 % de probabilidad) (valor medio de todos los<br />

lotes)<br />

Verfahrensstandardabweichung (Mittelwert aller Chargen)<br />

Standard Deviation of Procedure (average value of lots)<br />

Desviación estándar del procedimiento (valor medio de todos los lotes)<br />

Verfahrensvariationskoeffizient (Mittelwert aller Chargen)<br />

Variation Coefficient of Procedure (average value of lots)<br />

Coeficiente de variación del procedimiento (valor medio de todos los lotes)<br />

Anzahl Produktionschargen zur Berechnung<br />

Number of Lots for calculation<br />

Número de lotes de producción para el cálculo<br />

Genauigkeit / Accuracy / Exactitud<br />

± 14 mg/l CSB / COD / DQO<br />

± 5.6 mg/l CSB / COD / DQO<br />

± 0.73 %<br />

30<br />

± 30 mg/l CSB / COD / DQO<br />

– Certificates are regularly updated<br />

• http://<strong>photometry</strong>.merck.de<br />

Merck KGaA, Darmstadt, 10.10.2004<br />

Susen Püschel<br />

Merck KGaA, 64271 Darmstadt, Germany<br />

►<br />

June 03, 2008 Complete solutions Page 53


End of <strong>the</strong> Presentation<br />

Modern requirements for state<br />

of <strong>the</strong> art <strong>photometry</strong><br />

Drinking water analysis with excellence<br />

Gunter Decker<br />

Product Manager Food & Environmental Analysis<br />

Merck KGaA, Darmstadt<br />

Germany<br />

June 03, 2008 Complete solutions Page 54

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!