Software architecture<br />

Figure 20. State Sending results flow chart<br />

4.6 Calculating parameters<br />

This section provides a summary of the formulas used to calculate the spirometry parameters. These formulas are embedded<br />

in the MCU software.<br />

With each new sample taken by the ADC, the next calculations are performed:<br />

If ADC Value > Signal offset:<br />

If ADC Value < Signal offset:<br />

If (FlowX >4.7), then flowfactor = 0<br />

If (FlowX ≤4.7), then flowfactor = (4.7―FlowX) / 2.7<br />

Instant flow = FlowX +(FlowX*flowfactor)<br />

Instant volume = Instant flow * sampling period<br />

Volume = Volume + instant volume<br />

• ADC Value—The value read from the ADC.<br />

• Conversion factor—This is a factor that converts the pressure measured <strong>with</strong> the sensor to flow units. It depends on the<br />

mouthpiece used. Using the sensor mentioned in Spirometry, this factor is 0.0136.<br />

• Signal offset—The baseline of the measurement. This value corresponds to an air flow of 0 (no differential pressure on<br />

the sensor).<br />

• Sampling period—The time between consecutive samples (seconds).<br />

• Instant flow—The corresponding instantaneous flow of the current sample (liters/second).<br />

<strong>Spirometer</strong> <strong>Demo</strong> <strong>with</strong> <strong>Freescale</strong> <strong>Microcontrollers</strong>, Rev. 1, 12/2012<br />

20 <strong>Freescale</strong> Semiconductor, Inc.