In the Event of Failure…<br />

Safety Device and Error Code<br />

This unit has an automatic diagnosis function built in the controller and safety devices independent of the<br />

controller. The table below shows the cause and the solution method when the safety device operates.<br />

Error Code:<br />

When an abnormal condition occurs, an error code appears and the alarm lamp lights in the controller,<br />

the buzzer sounds simultaneously. Record the error code and turn off the power of device<br />

immediately.<br />

Safety Device Notify Cause/Solution<br />

Sensor trouble detection<br />

SSR short-circuit detection<br />

Heater disconnecting detection<br />

Memory error<br />

Internal communication error<br />

Overheating<br />

Measurement temperature<br />

error<br />

“ALARM” lamp lights on,<br />

“Er.01” appears<br />

“ALARM” lamp lights on,<br />

“Er.02” appears<br />

“ALARM” lamp lights on,<br />

“Er.03” appears<br />

“ALARM” lamp lights on,<br />

“Er.15” appears<br />

“ALARM” lamp lights on,<br />

“Er.17” appears<br />

“ALARM” lamp lights on,<br />

“Er.19” appears<br />

“ALARM” lamp lights on,<br />

“--------” appears<br />

• Temperature sensor is broken or<br />

disconnected.<br />

• Make a call for service.<br />

• Triac is in short-circuit<br />

• Make a call for service.<br />

• Heater is disconnected.<br />

• Make a call for service.<br />

• Failure in internal memory.<br />

• Make a call for service.<br />

• Failure in internal communication or<br />

temperature inputting circuit.<br />

• Make a call for service.<br />

• Overheating prevention device is in<br />

operation.<br />

• Reset the power supply, and then<br />

adjust the setting temperature of the<br />

overheating protection device.<br />

• If the state does not recover, make a<br />

call for service.<br />

• Measurement value is out of display<br />

range.<br />

• Make a call for service.<br />