21.01.2014 Views

Varain FRG-720 Bayard Alpert Pirani Gauge - Ideal Vacuum Products

Varain FRG-720 Bayard Alpert Pirani Gauge - Ideal Vacuum Products

Varain FRG-720 Bayard Alpert Pirani Gauge - Ideal Vacuum Products

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

4 Operation<br />

When the supply voltage is applied, the measurement signal is<br />

available between pins 2 (+) and 12 (–) (relationship between<br />

measurement signal and pressure → 14 and [1]).<br />

The Profibus gauge can also be operated via the corresponding<br />

fieldbus interface (Profibus) (→ [1] for details and further<br />

functions).<br />

Allow for a stabilizing time of ≈10 minutes. Once the gauge has<br />

been switched on, permanently leave it on irrespective of the<br />

pressure.<br />

4.1 Gas Type Dependence<br />

The measurement value is gas type dependent. The pressure<br />

reading applies to dry air, O 2 , CO and N 2 . For other gases, it has<br />

to be converted (→ 15 and [1]).<br />

tqma03e1 (2010-05) 23