25.01.2014 Views

P. Reiter

P. Reiter

P. Reiter

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

Summary:<br />

Detector module and Preamplifiers<br />

• A. Wiens, IKP Cologne<br />

• A. Goergen, CEA Saclay<br />

• D. Quirion, Canberra<br />

• G. Pascovici, IKP Cologne<br />

P. <strong>Reiter</strong> IKP University Cologne AGATA week 8-11 July 2008 Uppsala


Detector Delivery Owner Status<br />

A001* 11/2005 GANIL rejected – hole trapping<br />

A001 06/2006 GANIL accepted<br />

A002* 04/2006 TU Munich rejected – leakage currents<br />

A002 06/2007 TU Munich accepted<br />

A003* 11/2006 Liverpool rejected – leakage currents<br />

A003 10/2007 Liverpool accepted<br />

A004 09/2007 Ankara rejected – segments out of specification<br />

A004* 07/2008 Ankara CAT at IKP, Cologne<br />

A005 12/2007 Stockholm CAT at CEA, Saclay & IKP<br />

B001* 11/2005 Padova rejected – segments out of specification<br />

B001 06/2007 Padova rejected – leakage currents<br />

B002 08/2006 GANIL accepted<br />

B003 06/2007 Liverpool accepted<br />

B004 12/2007 Ankara rejected – missing segment in capsule<br />

B004* 06/2008 Ankara CAT at IKP, Cologne<br />

B005 12/2007 Stockholm CAT at CEA, Saclay<br />

C001 12/2005 Padova accepted<br />

C002 07/2006 GANIL accepted<br />

C003 03/2007 Liverpool accepted, broken, leakage currents<br />

C004 09/2007 Ankara accepted<br />

C005 12/2007 Stockholm CAT at CEA, Saclay


Summary customer acceptance tests<br />

B001 06/07 Padova Rejected - damage of contacts of segments A3 & C3<br />

A002 06/07 München Accepted<br />

C004 09/07 Ankara Accepted<br />

A004 09/07 Ankara Rejected – six segments out of E specs<br />

A003 10/07 Liverpool Accepted<br />

B004 12/07 Ankara Rejected - missing segment<br />

B004 06/08 Ankara Tested - not in specification<br />

A005 12/07 Stockholm Test ongoing (Saclay & Cologne)<br />

C005 12/07 Stockholm Test at Saclay<br />

B005 12/07 Stockholm Test at Saclay<br />

A004 07/08 Ankara Test at Cologne<br />

• Detector tests since last AGATA week<br />

•Tests are done at CEA, Saclay and IKP, Cologne


Summary available detector<br />

Available detectors for AGATA triple clusters<br />

A1, A2, A3<br />

B2, B3<br />

C2<br />

Detectors for period fall and winter 2008/09<br />

• 13 detectors may be available<br />

A1, A2, A3, A4, A5, A6<br />

B1, B2, B3, B4, B5, B6<br />

C1, C2, C4, C5<br />

• A6, B1, B6 delivery July 2008<br />

• redelivery of turkish detectors A4, B4, C4<br />

Future detectors<br />

• January - June 2009 italian detectors C6, A7<br />

• detector repair of C3<br />

• summer 2008 order of german detectors A8, B7, C7


First AGATA triple cluster detector<br />

IKP Cologne, H.-G. Thomas CTT Montabaur


A001 – FWHM [keV]<br />

1.6<br />

1.4<br />

1.2<br />

1<br />

0.8<br />

0.6<br />

1.6<br />

0.4<br />

1.4<br />

0.2<br />

1.2<br />

0<br />

1<br />

Tripel Cluster 1 Energies: Single vs Triple cryostat<br />

59,6 keV Single 59.6 keV Triple Specification low energy<br />

3<br />

1332,5 keV Single 1332,5 keV Triple Specification high energy<br />

2.5<br />

2<br />

1.5<br />

Excellent energy resolution in new triple cluster<br />

0.5<br />

2.5<br />

1<br />

3<br />

0<br />

2<br />

A1 A2A3A4A5A6<br />

B1 B2B3B4B5B6<br />

C1 C2C3C4C5C6<br />

D1 D2D3D4D5D6<br />

E1 E2E3E4E5E6<br />

F1 F2F3F4F5F6<br />

Core<br />

A1 A2A3A4A5A6<br />

B1 B2B3B4B5B6<br />

C1 C2C3C4C5C6<br />

D1 D2D3D4D5D6<br />

E1 E2E3E4E5E6<br />

F1 F2F3F4F5F6<br />

1.5<br />

1332,5 keV Single 1332,5 keV Triple Specification high energy<br />

0.5<br />

2.5<br />

0<br />

2<br />

A1 A2A3A4A5A6<br />

B1 B2B3B4B5B6<br />

C1 C2C3C4C5C6<br />

D1 D2D3D4D5D6<br />

E1 E2E3E4E5E6<br />

F1 F2F3F4F5F6<br />

Core<br />

A1 A2A3A4A5A6<br />

B1 B2B3B4B5B6<br />

C1 C2C3C4C5C6<br />

D1 D2D3D4D5D6<br />

E1 E2E3E4E5E6<br />

F1 F2F3F4F5F6<br />

Core<br />

0.8<br />

Core<br />

0.6<br />

1.6<br />

0.4<br />

1.4<br />

0.2<br />

1.2<br />

0<br />

1<br />

59.6 keV Single 59.6 keV Triple Specification low energy<br />

1.5<br />

1<br />

1332,5 keV Single 1332,5 keV Triple Specification high energy<br />

0.5<br />

0<br />

A1 A2A3A4A5A6<br />

B1 B2B3B4B5B6<br />

C1 C2C3C4C5C6<br />

D1 D2D3D4D5D6<br />

E1 E2E3E4E5E6<br />

F1 F2F3F4F5F6<br />

Core<br />

A1 A2A3A4A5A6<br />

B1 B2B3B4B5B6<br />

C1 C2C3C4C5C6<br />

D1 D2D3D4D5D6<br />

E1 E2E3E4E5E6<br />

F1 F2F3F4F5F6<br />

Core<br />

0.8<br />

0.6<br />

0.4<br />

0.2<br />

0<br />

Resolution 60keV line Resolution 1.33MeV line<br />

Mean values segments @ 60 keV :<br />

A001 : 1011 +/- 53 eV<br />

B002 : 1039 +/- 70 eV<br />

C002 : 965 +/- 1<br />

3<br />

63 eV<br />

especially at low energies (electronic noise):<br />

59,6 keV Single 59,6 keV Triple Specification low energy<br />

- glued detector → less microphonics<br />

- HEKO → AGATA preamp<br />

- electronic assembly (grounding, cabling)


A001 in Triple Cryostat 5000 V<br />

Measured Cross talk<br />

Core to Seg Xtalk (theory)<br />

0<br />

0 200 400 600 800 1000 1200 1400<br />

Smallest capacities in ring 2<br />

Xtalk amplitudes from seg "i" to seg "j"<br />

[%]<br />

-0,0005<br />

-0,001<br />

-0,0015<br />

-0,002<br />

A2<br />

B2<br />

C2<br />

D2<br />

E2<br />

F2<br />

-0,0025<br />

segment "i" + 36x segment "j"


No cross talk between detectors in Triple Cryostat<br />

0.05%<br />

A001 B002 C002<br />

A001<br />

0.00%<br />

-0.05%<br />

-0.10%<br />

-0.15%<br />

-0.20%<br />

-0.25%<br />

0 200 400 600 800 1000 1200 1400 1600<br />

Sector A<br />

Sector F<br />

Sector A Sector F Sector A Sector F<br />

Core B<br />

Core C<br />

0.05%<br />

B002<br />

0.00%<br />

-0.05%<br />

-0.10%<br />

-0.15%<br />

0 200 400 600 800 1000 1200 1400 1600<br />

Core A<br />

Core C<br />

-0.20%<br />

-0.25%<br />

0.05%<br />

C002<br />

0.00%<br />

-0.05%<br />

-0.10%<br />

-0.15%<br />

0 200 400 600 800 1000 1200 1400 1600<br />

Core A<br />

Core B<br />

-0.20%<br />

-0.25%


AGATA Triple Clusters: Summary and outlook<br />

AGATA Triple Cluster Detectors operational:<br />

-first triple:<br />

• cryostat, mechanics, vacuum o.k.<br />

• seg-preamps, core preamp & pulser o.k.<br />

• energy resolution o.k.<br />

• no cross talk between detectors<br />

•second triple: commissioning phase<br />

Outlook:<br />

-- transport to Legnaro of ATC1<br />

- in July 2008<br />

-- assembly of third triple<br />

- constrained by available detectors<br />

ATC1<br />

ATC2<br />

ATC3


AGATA Core - upgraded front end electronics<br />

‣ Upgraded Charge Sensitive Preamplifier<br />

- new frequency compensation for Single & Dual Gain Core<br />

‣ Reworked Dual Gain Core, reconfigurable:<br />

- either Single or Dual; either LV-CMOS or LVDS<br />

‣ Programmable Spectroscopic Pulser<br />

- as a tool for self-calibrating energy, time, dead time, time<br />

alignment, detector characterization etc.)<br />

G. Pascovici for the AGATA Preamplifier- and Detector Group<br />

F. Zocca, A. Puglia, D. Bazzacco


Conclusions<br />

• very low noise, very wide dynamic<br />

range charge-sensitive pre-amplifier<br />

• spectroscopic range successfully extended<br />

by more than one order of magnitude with<br />

new TOT technique up to 180 MeV<br />

• two modes of operations four sub-ranges:<br />

0-5 (20) MeV and 5(20)-180 MeV<br />

• very clean transfer function<br />

improvement: ‘Dominant pole compensation’<br />

• Programmable Spectroscopic Pulser