07.03.2014 Views

(SOPs): GC/MS ANALYSES AND DATA PROCESSING METHODS

(SOPs): GC/MS ANALYSES AND DATA PROCESSING METHODS

(SOPs): GC/MS ANALYSES AND DATA PROCESSING METHODS

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

ST<strong>AND</strong>ARD OPERATING PROCEDURES (<strong>SOPs</strong>):<br />

<strong>GC</strong>/<strong>MS</strong> <strong>ANALYSES</strong> <strong>AND</strong> <strong>DATA</strong> <strong>PROCESSING</strong> <strong>METHODS</strong> FOR<br />

ORGANOCHLORINE CONTAMINANTS IN TRADITIONAL FOODS<br />

Analyte list<br />

1. Select the contaminants to measure in traditional foods. Suggested compounds:<br />

1.1. Chlordanes: alpha-chlordane, gamma-chlordane, cis-nonachlor, trans-nonachlor and nonachlor<br />

III<br />

1.2. DDT compounds: p,p'-DDT, p,p'-DDE and p,p'-DDD<br />

1.3. PCB congeners: 18, 28, 44, 52, 66, 101, 105, 118, 128, 138, 153, 170, 180, 187, 195, 206 and<br />

209<br />

1.4. Hexachlorobenzene<br />

1.5. Toxaphenes: 2 congeners, an octachlorotoxaphene and a nonachlorotoxaphene<br />

<strong>GC</strong>/<strong>MS</strong> Instrumental Analyses<br />

2. Select the instrument settings and conditions for gas chromatography/mass spectrometry in order to<br />

measure the analytes selected above.<br />

<strong>GC</strong>/<strong>MS</strong> settings:<br />

injection volume<br />

injection type<br />

<strong>GC</strong> inlet temperature<br />

inlet purged time<br />

carrier gas<br />

constant head pressure<br />

septum purge flow<br />

split vent flow<br />

column dimensions<br />

column phase<br />

detector temperature<br />

3-µL<br />

splitless<br />

270° C<br />

1 min.<br />

Helium<br />

40 psi<br />

1 mL/min.<br />

50 mL/min.<br />

60 mm long x 0.25 mm internal diameter<br />

0.25 µm film-thickness "DB-5"<br />

300° C<br />

<strong>GC</strong> oven temperature program:<br />

80° C for 1 min.<br />

80° C to 150° C at 10° C/min.<br />

150° C to 190° C at 0.5° C/min.<br />

190° C to 315° C at 3° C/min.<br />

315° C for 15 min.


<strong>MS</strong> selected-ion monitoring:<br />

Approximate<br />

Time range (min.)<br />

Dwell<br />

Window<br />

Mass Time Analyte or Internal Standard<br />

1 24-31 207 1000 Tetrachloro-meta-xylene (<strong>GC</strong> Internal<br />

Standard<br />

2 31-37 284 1000 Hexachlorobenzene<br />

3 37-55 256 1000 PCB 18 and 28<br />

4 55-65 292 1000 PCB 44 and 52<br />

5 65-68 326 1000 PCB 103 (Internal Standard)<br />

6 68-71 115 1000 Oxychlordane<br />

7 71-75 292 1000 PCB 66<br />

8 75-78 373 1000 gamma-chlordane<br />

9 78-87 326 333 PCB 101<br />

409 333 nonachlor III and trans-nonachlor<br />

373 333 alpha-chlordane<br />

10 87-92 246 1000 p,p'-DDE<br />

11 92-97 329 1000 octachlorotoxaphene<br />

12 97-99.6 326 1000 PCB 118<br />

13 99.6-103 409 500 cis-nonachlor<br />

235 500 p,p'-DDD<br />

14 103-107 360 500 PCB 153<br />

326 500 PCB 105<br />

15 107-110 235 500 p,p'-DDT<br />

360 500 PCB 138<br />

16 110-116 394 333 PCB 187<br />

327 333 nonachlorotoxaphene<br />

360 333 PCB 128<br />

17 116-122 394 500 PCB 180<br />

430 500 PCB 170<br />

18 122-126 430 1000 PCB 195<br />

19 126-129 464 1000 PCB 206<br />

20 129-132 498 1000 PCB 209<br />

1. Adjust the Time ranges as necessary (e.g., after <strong>GC</strong> maintenance) based on the retention times of<br />

the analytes.<br />

2. Analyze sets of samples along with five or more calibration standards (referred to as multi-level<br />

standards)<br />

2.1. Each multilevel calibration standard contains the analytes at a different concentration so that the<br />

concentrations of the calibration standards span the range of analyte concentrations expected to<br />

be found in the tissue samples.<br />

2.2. Monitor the <strong>GC</strong>/<strong>MS</strong> reproducibility using the mid-level calibration standard. Analyze a mid-level<br />

calibration standard at the beginning, middle and end of the series of samples and standards.<br />

Data Processing<br />

3. Calculate the concentrations of the analytes relative to the Internal Standard in the samples using the<br />

quadratic regression of signal versus concentration in the multilevel standards. [Note that this<br />

calibration/calculation process will have its own SOP once the analysis has been validated and the<br />

August 17, 2000 2


calculation databases have been developed.]<br />

4. Calculate the lower quantitation limits for analytes in samples using the <strong>GC</strong>/<strong>MS</strong> minimum area<br />

equivalent to the area of the analytes in the lowest level calibration standard analyzed with the set.<br />

5. Calculate the % recovery of the Internal Standard PCB 103 using the <strong>GC</strong> Internal Standard<br />

Tetrachloro-meta-xylene.<br />

6. Calculate the estimated concentration of total PCBs by multiplying 2 times the sum of the<br />

concentrations of the 17 PCB congeners.<br />

- DRAFT -<br />

August 17, 2000 3

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!