LI-820 OEM CO2 Gas Analyser Brochure (PDF) - John Morris Scientific
LI-820 OEM CO2 Gas Analyser Brochure (PDF) - John Morris Scientific
LI-820 OEM CO2 Gas Analyser Brochure (PDF) - John Morris Scientific
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
<strong>LI</strong>-<strong>820</strong> CO 2 Analyzer<br />
For <strong>OEM</strong> Applications<br />
Introducing the <strong>LI</strong>-<strong>820</strong><br />
Long recognized as a market leader in high<br />
quality, portable infrared gas analyzers for<br />
environmental research, <strong>LI</strong>-COR, Inc. now<br />
introduces the <strong>LI</strong>-<strong>820</strong> CO 2 Analyzer, a low<br />
cost, low maintenance CO 2 analyzer for <strong>OEM</strong><br />
applications.<br />
Choose Your Measurement<br />
Range<br />
The <strong>LI</strong>-<strong>820</strong> is a single path, dual wavelength,<br />
non-dispersive infrared (NDIR) gas analyzer<br />
with ± 1 ppm resolution. This compact,<br />
lightweight analyzer features interchangeable<br />
optical benches; your choice of a 5.5" (14 cm)<br />
bench with CO 2 measurement ranges of 0-<br />
1000 or 0-2000 ppm, and a 2" (5 cm) bench<br />
with measurement ranges of 0-5000 or 0-<br />
20000 ppm.<br />
Sophisticated Measurement<br />
Technology<br />
A built-in pressure sensor and a thermostatically<br />
controlled optical path provide excellent<br />
stability over time, temperature and pressure<br />
changes. This reduces the need for frequent<br />
zero and span adjustments.<br />
Choose Your Application<br />
The <strong>LI</strong>-<strong>820</strong> is well suited for industrial and<br />
other continuous CO 2 monitoring applications.<br />
High/low alarms can be used to control CO 2<br />
levels or to notify operators of abnormally<br />
high or low CO 2 conditions. Activated alarms<br />
are also indicated by LED lights on the<br />
instrument top panel. The <strong>LI</strong>-<strong>820</strong> is perfect for<br />
incorporating into larger complex or multi-gas<br />
instruments requiring a reliable <strong>OEM</strong> CO 2<br />
analyzer.<br />
Multiple Data Outputs<br />
For complete and easy integration of the<br />
<strong>LI</strong>-<strong>820</strong> into other equipment controls for <strong>OEM</strong><br />
applications, the <strong>LI</strong>-<strong>820</strong> features a communication<br />
protocol based on a subset of the<br />
eXtensible Markup Language (XML). XML<br />
is a simple text-based language allowing<br />
communication with the <strong>LI</strong>-<strong>820</strong> using simple<br />
text strings sent bi-directionally between a data<br />
acquisition system and the <strong>LI</strong>-<strong>820</strong> over the RS-<br />
232 port. As an example, sending this<br />
command:<br />
?<br />
instructs the <strong>LI</strong>-<strong>820</strong> to send the most recent set<br />
of data values. In this manner, the <strong>LI</strong>-<strong>820</strong> can<br />
be polled for data at user-defined intervals,<br />
globally reconfigured, or configured to<br />
perform automatic calibration routines.<br />
A choice of parameters (CO 2 , cell pressure, or<br />
cell temperature) can be output as linearized<br />
voltage or current signals. In addition, 0-5V<br />
alarm relay outputs can be connected to<br />
audible alarms, relay switches, automatic<br />
dialers, pumps, or valves in industrial settings.<br />
The <strong>LI</strong>-<strong>820</strong> also comes with Windows®compatible<br />
interface software for instrument<br />
configuration, control, data collection, and<br />
display.<br />
Simple Maintenance<br />
The optical bench can be easily cleaned with<br />
the cleaning kit provided; simply remove the<br />
source and detector housings and swab out the<br />
optical path. This reduces down-time due to<br />
contamination, and eliminates the need for<br />
factory recalibration.<br />
●<br />
●<br />
●<br />
●<br />
●<br />
●<br />
●<br />
●<br />
Safety monitoring<br />
Leak detectors<br />
Indoor air quality<br />
TOC analyzers<br />
Greenhouse control systems<br />
Growth chambers<br />
Fruit storage<br />
Meteorological studies
<strong>LI</strong>-<strong>820</strong> Operating Principle<br />
The <strong>LI</strong>-<strong>820</strong> analyzer is a single beam, dual<br />
wavelength NDIR sensor. The CO 2 measurement<br />
is a function of the absorption of IR<br />
energy as it travels through the optical path.<br />
Source<br />
Gold Plated<br />
Parabolic Reflector<br />
Gold Plated<br />
Optical Path<br />
Pressure Transducer<br />
Filters<br />
3.95 & 4.26 µm<br />
Detector<br />
The IR source emits radiation into the optical<br />
path, where CO 2 absorbs photons of a certain<br />
wavelength. The CO 2 sampling channel uses<br />
an optical filter centered at 4.26 micrometers,<br />
corresponding to the absorption band for CO 2 ,<br />
while the reference channel uses a nonabsorbing<br />
band at 3.95 micrometers. The<br />
detector measures the absorption of the<br />
infrared beam passing through the optical<br />
path; the ratio of the sample and reference<br />
signals indicates the amount of light absorption<br />
by CO 2 , and thus, the gas concentration.<br />
Thermistor<br />
Ribbon Cable<br />
Connector<br />
<strong>Gas</strong> Inlet<br />
Heating Element<br />
Broad Band IR Source<br />
<strong>Gas</strong> Outlet<br />
Heating Element<br />
14 cm and 5 cm<br />
optical benches<br />
Thermistor<br />
Interchangeable optical benches swap out<br />
easily by means of two source connectors on<br />
the main circuit board; swapping the benches<br />
takes only a few minutes.<br />
Ordering Information<br />
<strong>LI</strong>-<strong>820</strong> CO 2 Analyzer:<br />
Includes <strong>LI</strong>-<strong>820</strong> CO 2 Analyzer with choice of 5.5” (14 cm) or 2” (5 cm) optical bench, spare parts kit, 2 disposable air filters, cleaning kit, Windows®<br />
communications software, and a 9-pin RS-232 communications cable. A power source is required.<br />
800-401 AC Power Supply (110 VAC):<br />
18 VDC output, 800 mA, for indoor use only. The AC power supply is required if the <strong>LI</strong>-<strong>820</strong> will not be wired directly to an external source. See<br />
Power Requirements in the specifications.<br />
Specifications<br />
Measurement Range: 0 – 1000 ppm, 0 – 2000 ppm with 14 cm optical bench<br />
0 – 5000 ppm, 0 – 20000 ppm with 5 cm optical bench<br />
Measurement Principle: Non-Dispersive Infrared<br />
Accuracy: