<strong>LI</strong>-<strong>820</strong> CO 2 Analyzer<br />

For <strong>OEM</strong> Applications<br />

Introducing the <strong>LI</strong>-<strong>820</strong><br />

Long recognized as a market leader in high<br />

quality, portable infrared gas analyzers for<br />

environmental research, <strong>LI</strong>-COR, Inc. now<br />

introduces the <strong>LI</strong>-<strong>820</strong> CO 2 Analyzer, a low<br />

cost, low maintenance CO 2 analyzer for <strong>OEM</strong><br />

applications.<br />

Choose Your Measurement<br />

Range<br />

The <strong>LI</strong>-<strong>820</strong> is a single path, dual wavelength,<br />

non-dispersive infrared (NDIR) gas analyzer<br />

with ± 1 ppm resolution. This compact,<br />

lightweight analyzer features interchangeable<br />

optical benches; your choice of a 5.5" (14 cm)<br />

bench with CO 2 measurement ranges of 0-<br />

1000 or 0-2000 ppm, and a 2" (5 cm) bench<br />

with measurement ranges of 0-5000 or 0-<br />

20000 ppm.<br />

Sophisticated Measurement<br />

Technology<br />

A built-in pressure sensor and a thermostatically<br />

controlled optical path provide excellent<br />

stability over time, temperature and pressure<br />

changes. This reduces the need for frequent<br />

zero and span adjustments.<br />

Choose Your Application<br />

The <strong>LI</strong>-<strong>820</strong> is well suited for industrial and<br />

other continuous CO 2 monitoring applications.<br />

High/low alarms can be used to control CO 2<br />

levels or to notify operators of abnormally<br />

high or low CO 2 conditions. Activated alarms<br />

are also indicated by LED lights on the<br />

instrument top panel. The <strong>LI</strong>-<strong>820</strong> is perfect for<br />

incorporating into larger complex or multi-gas<br />

instruments requiring a reliable <strong>OEM</strong> CO 2<br />

analyzer.<br />

Multiple Data Outputs<br />

For complete and easy integration of the<br />

<strong>LI</strong>-<strong>820</strong> into other equipment controls for <strong>OEM</strong><br />

applications, the <strong>LI</strong>-<strong>820</strong> features a communication<br />

protocol based on a subset of the<br />

eXtensible Markup Language (XML). XML<br />

is a simple text-based language allowing<br />

communication with the <strong>LI</strong>-<strong>820</strong> using simple<br />

text strings sent bi-directionally between a data<br />

acquisition system and the <strong>LI</strong>-<strong>820</strong> over the RS-<br />

232 port. As an example, sending this<br />

command:<br />

?<br />

instructs the <strong>LI</strong>-<strong>820</strong> to send the most recent set<br />

of data values. In this manner, the <strong>LI</strong>-<strong>820</strong> can<br />

be polled for data at user-defined intervals,<br />

globally reconfigured, or configured to<br />

perform automatic calibration routines.<br />

A choice of parameters (CO 2 , cell pressure, or<br />

cell temperature) can be output as linearized<br />

voltage or current signals. In addition, 0-5V<br />

alarm relay outputs can be connected to<br />

audible alarms, relay switches, automatic<br />

dialers, pumps, or valves in industrial settings.<br />

The <strong>LI</strong>-<strong>820</strong> also comes with Windows®compatible<br />

interface software for instrument<br />

configuration, control, data collection, and<br />

display.<br />

Simple Maintenance<br />

The optical bench can be easily cleaned with<br />

the cleaning kit provided; simply remove the<br />

source and detector housings and swab out the<br />

optical path. This reduces down-time due to<br />

contamination, and eliminates the need for<br />

factory recalibration.<br />

●<br />

●<br />

●<br />

●<br />

●<br />

●<br />

●<br />

●<br />

Safety monitoring<br />

Leak detectors<br />

Indoor air quality<br />

TOC analyzers<br />

Greenhouse control systems<br />

Growth chambers<br />

Fruit storage<br />

Meteorological studies


<strong>LI</strong>-<strong>820</strong> Operating Principle<br />

The <strong>LI</strong>-<strong>820</strong> analyzer is a single beam, dual<br />

wavelength NDIR sensor. The CO 2 measurement<br />

is a function of the absorption of IR<br />

energy as it travels through the optical path.<br />

Source<br />

Gold Plated<br />

Parabolic Reflector<br />

Gold Plated<br />

Optical Path<br />

Pressure Transducer<br />

Filters<br />

3.95 & 4.26 µm<br />

Detector<br />

The IR source emits radiation into the optical<br />

path, where CO 2 absorbs photons of a certain<br />

wavelength. The CO 2 sampling channel uses<br />

an optical filter centered at 4.26 micrometers,<br />

corresponding to the absorption band for CO 2 ,<br />

while the reference channel uses a nonabsorbing<br />

band at 3.95 micrometers. The<br />

detector measures the absorption of the<br />

infrared beam passing through the optical<br />

path; the ratio of the sample and reference<br />

signals indicates the amount of light absorption<br />

by CO 2 , and thus, the gas concentration.<br />

Thermistor<br />

Ribbon Cable<br />

Connector<br />

<strong>Gas</strong> Inlet<br />

Heating Element<br />

Broad Band IR Source<br />

<strong>Gas</strong> Outlet<br />

Heating Element<br />

14 cm and 5 cm<br />

optical benches<br />

Thermistor<br />

Interchangeable optical benches swap out<br />

easily by means of two source connectors on<br />

the main circuit board; swapping the benches<br />

takes only a few minutes.<br />

Ordering Information<br />

<strong>LI</strong>-<strong>820</strong> CO 2 Analyzer:<br />

Includes <strong>LI</strong>-<strong>820</strong> CO 2 Analyzer with choice of 5.5” (14 cm) or 2” (5 cm) optical bench, spare parts kit, 2 disposable air filters, cleaning kit, Windows®<br />

communications software, and a 9-pin RS-232 communications cable. A power source is required.<br />

800-401 AC Power Supply (110 VAC):<br />

18 VDC output, 800 mA, for indoor use only. The AC power supply is required if the <strong>LI</strong>-<strong>820</strong> will not be wired directly to an external source. See<br />

Power Requirements in the specifications.<br />

Specifications<br />

Measurement Range: 0 – 1000 ppm, 0 – 2000 ppm with 14 cm optical bench<br />

0 – 5000 ppm, 0 – 20000 ppm with 5 cm optical bench<br />

Measurement Principle: Non-Dispersive Infrared<br />

Accuracy: