• ECMO initiated<br />
• Flow of 2.8 LPM achieved<br />
• Fi02 1.0, sweep gas of 4.0 liters,<br />
• Arterial oxygen saturation 98% on oxysat monitor, CDI 100<br />
• Negative pressure -56 mmhg (measured pre Rotoflow cone), with<br />
chatter noted in lines<br />
• To bring down the inlet pressure, Voluven 250 mls given , pressure<br />
dropped to – 27 mmhg<br />
• ACT drawn with results of 360 seconds<br />
• Heparin infusion started at 1600 units per hour<br />
• ACT’s remained in the 160-210 sec range with minor adjustments of<br />
heparin drip. 1500-1800 1800 units per hour<br />
• ACT’s drawn every hour