(g) Reverse Osmosis (RO) Module<br />
<br />
<br />
<br />
(Refer to Water Quality).<br />
Type of RO membrane: Spiral Wound Polyamide, TEC of<br />
Polysulfone or equivalent.<br />
The recovery rate of RO system shall be at least 50%.<br />
<br />
Standard water treatment system shall have the following<br />
parameters displayed:<br />
•<br />
•<br />
•<br />
•<br />
•<br />
•<br />
Conductivity of permeate<br />
<br />
<br />
Raw water pressure<br />
Guard-in & guard-out pressure<br />
RO (membrane) system-in & system-out pressure<br />
<br />
Water sample ports shall be available for sampling at the following<br />
points:<br />
•<br />
•<br />
•<br />
•<br />
•<br />
•<br />
•<br />
<br />
Post second carbon column<br />
Post softener column/Pre-RO module<br />
Immediate post RO module<br />
First point in the distribution loop<br />
Last point in the distribution loop<br />
Last point of the dialyzer-reprocessing loop.<br />
<br />
In the event of RO pump failure, the softened water shall be<br />
diverted into the 0.2 microns Bacterial Filter as temporary measure.<br />
However, this shall not exceed 24 hours.<br />