(g) Reverse Osmosis (RO) Module<br />

<br />

<br />

<br />

(Refer to Water Quality).<br />

Type of RO membrane: Spiral Wound Polyamide, TEC of<br />

Polysulfone or equivalent.<br />

The recovery rate of RO system shall be at least 50%.<br />

<br />

Standard water treatment system shall have the following<br />

parameters displayed:<br />

•<br />

•<br />

•<br />

•<br />

•<br />

•<br />

Conductivity of permeate<br />

<br />

<br />

Raw water pressure<br />

Guard-in & guard-out pressure<br />

RO (membrane) system-in & system-out pressure<br />

<br />

Water sample ports shall be available for sampling at the following<br />

points:<br />

•<br />

•<br />

•<br />

•<br />

•<br />

•<br />

•<br />

<br />

Post second carbon column<br />

Post softener column/Pre-RO module<br />

Immediate post RO module<br />

First point in the distribution loop<br />

Last point in the distribution loop<br />

Last point of the dialyzer-reprocessing loop.<br />

<br />

In the event of RO pump failure, the softened water shall be<br />

diverted into the 0.2 microns Bacterial Filter as temporary measure.<br />

However, this shall not exceed 24 hours.<br />