(j) Disinfection of Distribution Loops<br />
<br />
<br />
<br />
A minimum ofsix (6) monthly<br />
recommendation) chemical disinfection of distribution loop<br />
including the connections to dialysis machine shall be done using<br />
peracetic acid 2-3% or chlorine dioxide especially when materials<br />
of distribution loop are not heat resistant.<br />
Weekly heat disinfection of the tank and distribution loop is<br />
recommended for a system which incorporates a heater and uses<br />
heat resistant piping.<br />
<br />
velocity of 1.5 feet per second (FPS) for direct feed system and 3.0<br />
FPS for indirect feed system.<br />
Additional disinfection may be needed in the following<br />
circumstances:<br />
(i)<br />
(ii)<br />
(iii)<br />
(iv)<br />
(v)<br />
Installation of new system<br />
Upgrading of existing system<br />
Out-break of pyrogenic reaction<br />
Breach of the closed loop system.<br />
When microbial testing of treated water reach action level<br />
(refer section 5.3.5).<br />
3.4 Reprocessing<br />
3.4.1 Dialyser Reprocessing Machine<br />
<br />
The reprocessing machine shall be approved by regulatory<br />
authorities in the USA or equivalent.<br />
<br />
The reprocessing machine shall be a fully automated integrated<br />
<br />
<br />
For reprocessing of dialyser, this shall include testing for total cell<br />
volume (TCV), membrane integrity and perform disinfection as per<br />
AAMI standard.<br />
<br />
<br />
<br />
<br />
Haemodialysis Quality and Standards 21