haemodialysis_quality_standards - Rotary International District 3310
haemodialysis_quality_standards - Rotary International District 3310
haemodialysis_quality_standards - Rotary International District 3310
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
5.4.5 Limits and Action Level<br />
Maximum Allowed<br />
<br />
<br />
CFU level < 100 CFU/ml<br />
Endotoxin level < 0.25 EU/ml<br />
Action Level<br />
<br />
<br />
CFU level > 50 CFU/ml<br />
Endotoxin Level> 0.125EU/ml<br />
(Ref: AAMI/ISO 23500: 2011)<br />
If Action Levels are observed, disinfection and retesting shall be done<br />
immediately to restore the <strong>quality</strong> into acceptable level.<br />
5.4.6 Laboratory<br />
All samples shall be sent to an accredited laboratory recognized by the<br />
Director General of Health.<br />
5.4.7 Record<br />
All the results shall be properly documented and made available for<br />
inspection.<br />
Haemodialysis Quality and Standards 27