Multisignal LC/MS Analysis for Compound ... - Agilent Technologies
Multisignal LC/MS Analysis for Compound ... - Agilent Technologies
Multisignal LC/MS Analysis for Compound ... - Agilent Technologies
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
application<br />
<strong>Multisignal</strong> <strong>LC</strong>/<strong>MS</strong> <strong>Analysis</strong> <strong>for</strong> <strong>Compound</strong> Screening<br />
Christine Miller<br />
Introduction<br />
Acquiring <strong>MS</strong> in<strong>for</strong>mation using different acquisition<br />
modes within a single sample analysis is a<br />
powerful way to improve productivity in <strong>LC</strong>/<strong>MS</strong><br />
compound screening and method development.<br />
It speeds analyses and makes it easier to use<br />
a single, generic method <strong>for</strong> <strong>LC</strong>/<strong>MS</strong> screening.<br />
The enhanced <strong>Agilent</strong> 1100 <strong>LC</strong>/<strong>MS</strong>D provides<br />
the ability to cycle through as many as four<br />
different acquisition modes on a scan-by-scan<br />
basis within a single sample analysis. Each<br />
acquisition mode is user-definable and can be<br />
customized <strong>for</strong> specific needs. This flexibility<br />
allows many combinations of acquisition modes<br />
including high/low energy in-source collisioninduced<br />
dissociation (CID), positive/negative<br />
polarity switching, and selected ion monitoring<br />
(SIM)/scan modes.<br />
This note shows examples of using positive/<br />
negative polarity switching to confirm<br />
compound identification and to screen<br />
mixtures.<br />
Experimental<br />
All experiments were done using an <strong>Agilent</strong><br />
1100 Series <strong>LC</strong>/<strong>MS</strong>D system that was comprised<br />
of a binary pump, vacuum degasser, autosampler,<br />
thermostatted column compartment<br />
with column-switching valve, diode-array<br />
detector, and an enhanced <strong>LC</strong>/<strong>MS</strong>D. The<br />
<strong>LC</strong>/<strong>MS</strong>D was used with either the electrospray<br />
ionization (ESI) or atmospheric pressure<br />
chemical ionization (APCI) source. Complete<br />
system control and data evaluation were done<br />
on the <strong>Agilent</strong> ChemStation <strong>for</strong> <strong>LC</strong>/<strong>MS</strong>.<br />
Reagent grade chemicals and HP<strong>LC</strong> grade<br />
solvents were used in preparing mobile phases<br />
and standards.<br />
Tuning the <strong>LC</strong>/<strong>MS</strong>D is the process of adjusting<br />
parameters <strong>for</strong> sensitivity, mass resolution,<br />
and mass accuracy. A commercially available<br />
stable mixture of compounds (p/n G2421A<br />
and G2422A) is used <strong>for</strong> tuning. The autotune<br />
process on the <strong>LC</strong>/<strong>MS</strong>D automatically delivers<br />
the tune mix, optimizes parameters in both<br />
positive and negative ionization modes, and<br />
then creates a tune file that contains optimized<br />
parameters <strong>for</strong> both ionization modes. There<strong>for</strong>e,<br />
data can be acquired in positive/negative<br />
polarity switching mode with a single tune file.<br />
Results and Discussion<br />
<strong>Analysis</strong> parameters were developed with conditions<br />
that would allow the <strong>for</strong>mation of both<br />
negative and positive ions. In the electrospray<br />
ionization process, the mobile phase emerging<br />
in the electric field is charged and then the<br />
charged liquid is sprayed into droplets. Analyte<br />
ions in solution migrate to the droplet surfaces<br />
and, as the droplets are evaporated, gas phase<br />
ions are released. To maximize ion <strong>for</strong>mation,<br />
the mobile phase needs to be conductive so that<br />
the liquid can be highly charged. In addition,<br />
the mobile phase should be at a pH that will<br />
promote analyte ion <strong>for</strong>mation. Because the<br />
evaporation process will cause changes in the<br />
droplet pH that affect ionization, 1 volatility of<br />
mobile phase components is also an important<br />
consideration. For example, sulfamethizole,<br />
which has a pKa of 5.45, gives a better negative<br />
mode response with 0.1% acetic acid (pH 4.5)<br />
than with 10 mM ammonium acetate (pH 5.5).<br />
Both acetic acid and ammonium acetate will<br />
make the droplets conductive and promote<br />
ionization. However, because acetic acid is more<br />
volatile, the pH in the droplets will increase<br />
as the acetic acid concentration is reduced,<br />
producing a condition that is more favorable <strong>for</strong><br />
the <strong>for</strong>mation of negative ions of sulfamethizole.<br />
In APCI, ions are generated in the gas phase.<br />
Solvent and analytes are vaporized; the solvent<br />
is ionized by corona discharge; and the charge<br />
is transferred from the solvent to the analyte<br />
molecules. Using a protic solvent such as<br />
methanol will generally aid in the ionization<br />
process.
<strong>Multisignal</strong> <strong>LC</strong>/<strong>MS</strong> <strong>Analysis</strong> <strong>for</strong> <strong>Compound</strong> Screening<br />
Under the appropriate <strong>LC</strong>/<strong>MS</strong> conditions, some<br />
molecules will produce both positive and negative ions.<br />
The ESI-<strong>LC</strong>/<strong>MS</strong> analysis of a mixture of four sulfonamide<br />
antibiotics shows a response in both ionization<br />
modes (Figure 1). The mass spectra <strong>for</strong> sulfamethizole<br />
show the protonated molecular ion [M+H] + at m/z 271<br />
200000<br />
175000<br />
150000<br />
125000<br />
100000<br />
75000<br />
50000<br />
25000<br />
0<br />
1000000<br />
800000<br />
600000<br />
400000<br />
200000<br />
0<br />
Negative mode<br />
Positive mode<br />
16<br />
14<br />
12<br />
10<br />
8<br />
6<br />
4<br />
2<br />
0<br />
80<br />
60<br />
40<br />
20<br />
0<br />
Negative mode<br />
269.0<br />
100 271.0<br />
Positive mode<br />
1<br />
2<br />
1<br />
2<br />
1 2 3 4 5 6 7 8 9<br />
[M-H]<br />
200 250 300<br />
–<br />
[M+H]<br />
+<br />
H 2N<br />
1. Sulfamethizole<br />
2. Sulfamethazine<br />
3. Sulfachlorpyridazine<br />
4. Sulfamethoxine<br />
Figure 1. ESI-<strong>LC</strong>/<strong>MS</strong> analysis of sulfonamide antibiotics using<br />
positive/negative switching mode.<br />
O<br />
S<br />
S<br />
NH N<br />
O<br />
Figure 2. Mass spectra of sulfamethizole from the chromatogram<br />
in Figure 1.<br />
3<br />
3<br />
4<br />
4<br />
2<br />
and the deprotonated molecular ion [M-H] – at m/z 269<br />
(Figure 2). Having results from both positive and negative<br />
ionization in a single analysis provides confirmation<br />
of peak molecular weight without increasing analysis<br />
time or sample use.<br />
N<br />
m/z<br />
min<br />
ANALYSIS METHOD:<br />
Chromatographic Conditions<br />
Column: 15 × 3 mm<br />
Zorbax ® SB-C18, 3.5 µm<br />
(p/n 863954-302)<br />
Mobile phase: A = 0.1% acetic acid<br />
in water<br />
B = 0.1% acetic acid<br />
in acetonitrile<br />
Gradient: start with 20% B<br />
at 3 min, 20% B<br />
at 5 min, 50% B<br />
Flow rate: 0.6 ml/min<br />
Column temperature: 40°C<br />
Injection volume: 5 µl<br />
Diode-array detector: signal 270, 10 nm<br />
<strong>MS</strong> Conditions<br />
Source: ESI<br />
Drying gas flow: 11 l/min<br />
Nebulizer: 45 psig<br />
Drying gas temperature: 350°C<br />
Vcap: 3000 V (positive);<br />
2250 V (negative)<br />
Stepsize: 0.1<br />
Peakwidth: 0.09 min<br />
Time filter: On<br />
<strong>MS</strong> Signal 1: Ion mode: Negative<br />
Scan: 150–400 amu<br />
Fragmentor: 50 V<br />
<strong>MS</strong> Signal 2: Ion mode: Positive<br />
Scan: 150–400 amu<br />
Fragmentor: 50 V
<strong>Multisignal</strong> <strong>LC</strong>/<strong>MS</strong> <strong>Analysis</strong> <strong>for</strong> <strong>Compound</strong> Screening<br />
Because some molecules may respond in only the positive<br />
or negative ionization mode, screening methods<br />
that incorporate both modes are very useful. Figure 3<br />
shows the analysis of a mixture of polymer additives by<br />
Norm.<br />
3000000<br />
2500000<br />
2000000<br />
1500000<br />
1000000<br />
500000<br />
0<br />
positive<br />
negative<br />
1<br />
2<br />
2 4 6 8 10 12 14 16 18<br />
3<br />
4<br />
5<br />
6<br />
min<br />
3<br />
APCI-<strong>LC</strong>/<strong>MS</strong>. Some of the polymer additives ionize in<br />
both modes as with the sulfa drugs, but butylated<br />
hydroxytoluene (BHT) only responds well in negative<br />
mode. Moreover, some of the additives that respond in<br />
1. Irganox 245<br />
2. BHT<br />
3. Tinuvin 328<br />
4. Irganox 1010<br />
5. Irganox 1078<br />
6. Irgafos 168<br />
Figure 3. APCI-<strong>LC</strong>/<strong>MS</strong> analysis of a mixture of polymer additives.<br />
100<br />
80<br />
60<br />
40<br />
20<br />
0<br />
5<br />
4<br />
3<br />
2<br />
1<br />
0<br />
205.2<br />
A<br />
473.2<br />
200 400 600 800 1000<br />
647.3<br />
[M–173] –<br />
[M+H] +<br />
Positive mode<br />
Negative mode<br />
m/z<br />
100<br />
80<br />
60<br />
40<br />
20<br />
0<br />
60<br />
50<br />
40<br />
30<br />
20<br />
10<br />
0<br />
233.1<br />
B<br />
267.0<br />
352.1<br />
415.1<br />
471.1<br />
563.1<br />
527.2<br />
581.1<br />
731.2<br />
675.2<br />
619.1<br />
200 400 600 800 1000<br />
787.3<br />
Negative mode<br />
Positive mode<br />
[M–H] –<br />
Figure 4. Mass spectra <strong>for</strong> (A) Irgafos 168 and (B) Irganox 1010<br />
from chromatogram in Figure 3.<br />
1175.7<br />
m/z<br />
ANALYSIS METHOD:<br />
Chromatographic Conditions<br />
Column: 15 × 3 mm<br />
Zorbax ® SB-C18,<br />
3.5 µm (p/n 863954-302)<br />
Mobile phase: A = water<br />
B= methanol<br />
Gradient: start with 75% B<br />
at 5 min, 90% B<br />
at 14 min, 100%B<br />
Flow rate: 0.8 ml/min<br />
Column temperature: 50°C<br />
Injection volume: 5 µl of 400 ppm<br />
per component<br />
Diode-array detector: Signal 220, 10 nm<br />
<strong>MS</strong> Conditions<br />
Source: APCI<br />
Drying gas flow: 5 l/min<br />
Nebulizer: 60 psig<br />
Drying gas temperature: 350°C<br />
Vaporizer: 400°C<br />
Vcap: 3000 V (positive);<br />
3000 V (negative)<br />
Corona: 4 µA (positive);<br />
20 µA (negative)<br />
Stepsize: 0.1<br />
Peakwidth: 0.15 min<br />
Time filter: On<br />
<strong>MS</strong> Signal 1: Ion mode: Negative<br />
Scan: 200–1200 amu<br />
Fragmentor: 150 V<br />
<strong>MS</strong> Signal 2: Ion mode: Positive<br />
Scan: 200–1200 amu<br />
Fragmentor: 80 V
<strong>Multisignal</strong> <strong>LC</strong>/<strong>MS</strong> <strong>Analysis</strong> <strong>for</strong> <strong>Compound</strong> Screening<br />
both modes show fragmentation in one of the modes.<br />
Irgafos 168 shows just the [M+H] + ion in positive<br />
mode, but in negative mode shows a fragment ion<br />
at m/z 205 and an ion from rearrangement of the<br />
remaining molecule at m/z 473 (Figure 4A). Similarly,<br />
Irganox 1010 shows just the [M-H] – ion in negative<br />
mode but shows extensive fragmentation in positive<br />
mode (Figure 4B).<br />
This ability to acquire data using alternating positive<br />
and negative ionization modes maximizes the in<strong>for</strong>mation<br />
from each analysis.<br />
Reference<br />
1. Zhou, S., Edwards, A. G., Cooke, K. D., and<br />
Van Berkel, G. J., Analytical Chemistry 1999,<br />
71, 769–776.<br />
Author<br />
Christine Miller is an applications chemist at <strong>Agilent</strong><br />
<strong>Technologies</strong> in Palo Alto, Cali<strong>for</strong>nia.<br />
<strong>Agilent</strong> <strong>Technologies</strong> shall not be liable <strong>for</strong> errors contained<br />
herein or <strong>for</strong> incidental or consequential damages in connection<br />
with the furnishing, per<strong>for</strong>mance or use of this material.<br />
In<strong>for</strong>mation, descriptions and specifications in this publication<br />
are subject to change without notice.<br />
Copyright © 2000<br />
<strong>Agilent</strong> <strong>Technologies</strong><br />
All rights reserved.<br />
Reproduction and adaptation is prohibited.<br />
Printed in the U.S.A. January 2000<br />
(23) 5968-8658E