Operative sequence<br />

A protocol in 5 main steps.<br />

1 Rm hand file<br />

Place selected solvent in the pulp chamber.<br />

Operating procedure<br />

■ 1/4 turn with pressure<br />

directed towards apex.<br />

■ Removal.<br />

Function<br />

■ Used to dig or break the hard layer<br />

of filling material.<br />

■ Allows the centring and alignment<br />

of the next instrument.<br />

■ As soon as possible check the working length.<br />

■ If possible check the apical patency.<br />

■ At the end of the preparation, ensure the complete removal<br />

of the debris from the previous filling using an X-ray.<br />

■ In curved canals:<br />

- Same precaution as for any kind of NiTi instrument must be observed<br />

to avoid false canal creation.<br />

- Introduction of a pre-curved hand file MMC n°10 is necessary prior<br />

to penetration with R-Endo ® files.<br />