<strong>DICOM</strong> <strong>Conformance</strong> <strong>Statement</strong><br />

Table A.4.2-19<br />

Valid Image Attributes to display image<br />

Tag Name VR Description<br />

0008,0018 SOP Instance UID UI displayed in case of error<br />

0008,0023 Image Date DA Displayed<br />

0008,0033 Image Time TM Displayed<br />

0008,0080 Institution Name LO Displayed<br />

0008,1050 Performing Physicians Name PN Displayed<br />

0008,1090 Manufacturer’s Model Name LO ||<br />

0018,0060 KVP DS Displayed<br />

0018,1151 X-Ray Tube Current IS Displayed<br />

0018,1000 Device Serial Number LO ||<br />

0018,1020 Software Versions LO ||<br />

0020,0013 Image Number IS C-arm creates own number<br />

0020,4000 Image Comments LT max 64 character displayed<br />

0028,0002 Samples per pixel US 1,3<br />

0028,0004 Photometric Interpretation CS MONOCROME1,MONOCHROME2,RGB<br />

0028,0008 Number of Frames IS<br />

0028,0010 Rows US max 3800<br />

0028,0011 Columns US max 3800<br />

0028,0100 Bits Allocated US 8 or 16<br />

0028,0101 Bits Stored US all smaller Bits Allocated<br />

0028,0102 High Bit US Bits Stored – 1<br />

0028,0103 Pixel Representation US 0 (unsigned integer), 1(2‘s complement)<br />

0028,0006 Planar Configuration US 0, if Samples per pixel =3.. R1G1B1, ....<br />

0028,0106 Smallest Image Pixel Value US ||<br />

0028,0107 Largest Image Pixel Value US ||<br />

0028,1050 Window Center DS only first value<br />

0028,1051 Window Width DS only first value<br />

0028,1052 Rescale Intercept DS ||<br />

0028,1053 Rescale Slope DS ||<br />

0028,1055 Window Center&Width LO ||<br />

Explanation<br />

7FE0,0010 Pixel Data OB/OW only IMPLICIT LITTLE ENDIAN<br />

A.4.2.3 Workflow Application Entity Specification<br />

A.4.2.3.1 SOP Classes<br />

The Ziehm NetPort AE provides Standard <strong>Conformance</strong> to the following SOP<br />

Classes:<br />

Table A.4.2-19<br />

SOP CLASSES FOR AE WORKLIST and MPPS<br />

SOP Class Name SOP Class UID SCU SCP<br />

Modality Worklist Information Model – FIND 1.2.840.10008.5.1.4.31 Yes No<br />

Modality Performed Procedure Step 1.2.840.10008.3.1.2.3.3 Yes No<br />

Form: EW_<strong>DICOM</strong>_CONF.doc<br />

A.4.2.3.2 Association Policies (see A.4.2.1.2)<br />

A.4.2.3.3 Association Initiation Policy<br />

A.4.2.3.3.1 Activity – Worklist<br />

File: conforma215.doc<br />

Page 20 of 54