C VE Achieva GB.qxd
C VE Achieva GB.qxd
C VE Achieva GB.qxd
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
ACHIEVA<br />
Portable Volume Ventilator with Pressure Support
ACHIEVA Portable Ventilator<br />
ACHIEVA Portable Ventilators provide positive pressure support for a variety<br />
of clinical conditions in pediatric and adult patients. A pressure support<br />
capability aids the weaning process and improves patient comfort by reducing<br />
work of breathing. Extended battery operation and a lightweight, compact<br />
design enhance patient quality of life. An internal modem facilitates remote<br />
data transfer to clinicians, and an internal dial-in PEEP with flow/pressure<br />
trigger sensitivity promotes patient synchrony. ACHIEVA ventilators are<br />
highly efficient in terms of oxygen and power consumption compared with<br />
other portable ventilators. They are easy to operate, convenient to use and<br />
cost-effective to own.<br />
Product Features<br />
• Sturdy components and compact<br />
design for easy portability and<br />
uncompromised reliability.<br />
• Offers both volume and<br />
pressure control ventilation.<br />
• Flow and/or Pressure triggering<br />
for enhanced patient synchrony.<br />
• Internal dial-in PEEP takes the<br />
guesswork out of setting pressure<br />
trigger sensitivity.<br />
• Pressure Support to enhance<br />
weaning and improve patient<br />
comfort and patient ventilator/<br />
synchrony.<br />
• Expiratory Sensitivity improves<br />
patient synchrony by allowing the<br />
clinician to adjust the percent of<br />
peak inspiratory flow that will<br />
cycle the ventilator from inspiration<br />
to expiration in Pressure Support.<br />
• Internal and optional external<br />
battery for extended mobility.<br />
• Internal O2 blending for convenient<br />
and accurate delivery of 21% to<br />
100% FiO2 (PSO2 Model)<br />
• Flow Acceleration Limit.<br />
• Internal memory allow access to<br />
stored data, providing a tool to<br />
troubleshoot alarms and evaluate<br />
effectiveness of patient settings.<br />
• Can be used with pediatric<br />
patients as small as 5 kg.<br />
Specifications<br />
Dimensions / Weight<br />
27.3 cm x 33.8 cm x 39.6 cm / Less than<br />
14.6 kg<br />
Power Line / Power Consumption<br />
100-240 VAC, 50/60 Hz, universal AC input<br />
Maximum: 75 W Nominal: 10 W<br />
Optional External Battery: Approx. 20 hours<br />
operation under normal load.<br />
Internal Battery: At least 4 hours operation<br />
under normal load ; Backup use only<br />
Type — Positive Pressure Volume Ventilator<br />
Motor: Brushless induction<br />
Pump: Piston, 50 to 2200 ml tidal volume<br />
Front Panel Controls/Connections<br />
Alarm Silence/Reset, Battery Test, Standby,<br />
Ventilate, Menu/Esc, Start/Enter, Up and<br />
Down Arrow Keys, Patient Pressure Port,<br />
Patient Air Port: 22 mm O.D./15 mm I.D.,<br />
Exhalation Valve Port.<br />
Rear Panel Components<br />
O2 Port on backpanel: 9/16-18, DISS 1240<br />
THD, inlet filter, external battery, communication<br />
port (RS232), remote alarm, modem<br />
jack (US & Canada only), power entry module.<br />
Setting Parameters<br />
Ventilation Mode: Assist/Control, SIMV, SPON<br />
Tidal Volume: 50 to 2200 ml (1750 ml SIMV)<br />
Inspiratory Time: 0.2 to 5.0 seconds<br />
Inspiratory Trigger: Flow from 3 to 25 LPM.<br />
Pressure from 1 to 15 cm H2O/hPa<br />
Expiratory Trigger: 15 to 55% of peak<br />
inspiratory flow<br />
Breath Rate: 1 to 80 BPM<br />
Pressure: 0 to 50 cm H2O/hPa<br />
PEEP: 0 and 3 to 20 cm H2O/hPa<br />
Low Pressure Alarm: 1 to 59 cm H2O/hPa<br />
High Pressure Alarm: 2 to 80 cm H2O/hPa<br />
O2 Level: 21 to 100%<br />
Flow Acceleration: ON/OFF<br />
Backup rate (10 BPM) in SPON: ON/OFF<br />
Alarms<br />
Low Pressure/Apnea, High Pressure,<br />
Setting Error, Power Switchover,<br />
Low Power, O2 Failure<br />
Alerts<br />
External battery, Internal battery, Battery<br />
charge, Alarm control, Assist/Spontaneous<br />
Environment<br />
Operating Altitude: 0 to 4500 m<br />
Storage: -20°C to 50°C, 10% to 90% RH<br />
non condensing ; allow at least 1 hour<br />
temperature stabilization before use when<br />
moving from nonoperating to operating<br />
environment ; the battery must be<br />
recharged every 30 days.<br />
Ordering information<br />
REF<br />
Y-ACH+E*<br />
Y-ACHPS+E*<br />
Y-ACHPSO2+E*<br />
Description<br />
ACHIEVA<br />
ACHIEVA PS<br />
ACHIEVA PSO2<br />
Accessories<br />
Y-6267 Reusable adult circuit - metal<br />
Y-6268 Reusable adult circuit - plastic<br />
Y-6463 Disposable adult circuit<br />
Y-6464 Disposable pediatric circuit<br />
Y-6109 Flat pak air inlet filter<br />
L-006197-000 Bacteria filter (disposable)<br />
Optional Accessories<br />
Y-CGVP24-E External battery pack<br />
Y-6353 Exhalation valve<br />
L-6202 O2 enrichment kit<br />
Y-6109-02 O2 adapter for Flatpak filter<br />
L-6861 Trach support arm w/<br />
Fisher & Paykel bracket<br />
Y-102064-00 Ventilator wheelchair tray<br />
Y-6441-ACH Ventilator Cart<br />
* ACHIEVA Ventilators are available with English,<br />
French, German, Italian and Spanish .<br />
Please contact your local Tyco Healthcare<br />
representative for more informations.<br />
Use only PURITAN BENNETT accessories with<br />
the ventilators.<br />
Tyco Healthcare UK Limited - 154 Fareham Road - GOSPORT Hampshire - PO 13 0AS - UK<br />
Tel: +44 (0) 1329 224226 - Fax: +44 (0) 1329 224334 - www.tycohealthcare.com<br />
Trademarks of Tyco Healthcare Group LP or its affiliate © 2006 Tyco Healthcare Group LP. All rights reserved.<br />
C-<strong>VE</strong>-<strong>Achieva</strong>/<strong>GB</strong> - 05/2006