COMBITUBE<br />
Esophageal/Tracheal<br />
Double Lumen Airway
COMBITUBE<br />
Esophageal/Tracheal Double-Lumen Airway<br />
Rapid intubation of difficult airways<br />
Insert the COMBITUBE airway blindly<br />
into the mouth. Advance gently until<br />
the teeth or alveolar ridges are aligned<br />
between the two black rings.<br />
IInflate line 1 through the blue pilot<br />
balloon leading to the large pharyngeal<br />
cuff with 100ml of air for the 41 French<br />
or 85ml of air for the 37 French using<br />
the large syringe provided. Inflate<br />
line 2 through the white pilot balloon<br />
leading to the smaller distal cuff with<br />
approximately 15ml of air for the 41<br />
French or 12ml of air for the 37 French<br />
using the small syringe provided.<br />
Since an esophageal placement of<br />
the COMBITUBE airway is most likely,<br />
begin ventilation by attaching the<br />
ventilation device to the longer blue<br />
connecting tube marked No. 1.<br />
If auscultation of breath sounds<br />
is positive, indicating esophageal<br />
placement, continue ventilation.<br />
If esophageal placement is confirmed,<br />
the second clear connecting tube marked<br />
No. 2 con be used for gastric suction<br />
with the suction catheter provided.<br />
If auscultation of breath sounds is<br />
negative, a tracheal placement has<br />
been accomplished. The ventilation<br />
device should then be connected to<br />
the shorter clear connecting tube<br />
marked No. 2. The COMBITUBE<br />
airway is then functioning as a normal<br />
tracheal tube.<br />
Standard Tray<br />
Single<br />
Roll up Kit<br />
Ordering Information<br />
REF Description Size Quantity<br />
5-18537 COMBITUBE Standard Trays 37 Fr 4/box<br />
5-18541 COMBITUBE Standard Trays 41 Fr 4/box<br />
5-18237 COMBITUBE Singles 37 Fr 4/box<br />
5-18241 COMBITUBE Singles 41 Fr 4/box<br />
5-18437 COMBITUBE Roll-Ups 37 Fr 4/box<br />
5-18441 COMBITUBE Roll-Ups 41 Fr 4/box<br />
5-18141 COMBITUBE Demonstration Airway 41 Fr 1/box<br />
When a difficult airway must be established, time and<br />
accuracy are critical. Now you can be assured of rapid<br />
intubation and effective ventilation with the COMBITUBE<br />
airway from KENDALL. Unlike traditional endotracheal<br />
tubes, the COMBITUBE airway is designed to establish<br />
a nonsurgical patient airway when placed into either the<br />
trachea or the esophagus.<br />
●<br />
●<br />
●<br />
●<br />
●<br />
●<br />
Unique double-lumen design allows for rapid<br />
establishment of an effective airway through either<br />
esophageal or tracheal placement.<br />
Blind placement eliminates the need for laryngoscope.<br />
Pharyngeal balloon inflates to hold device firmly in place<br />
and helps prevent the escape of gas through the nose or<br />
mouth.<br />
Full-Iength lumen allows for suctioning of gastric contents<br />
with no interruption of patient ventilation should the<br />
COMBITUBE airway be placed in the esophagus.<br />
Esophageal cuff inflates to seal the esophagus so that<br />
gas does not enter the stomach and gastric contents are<br />
not aspirated.<br />
Appropriate for pre-hospital, surgical and emergency use.<br />
Trademark of Tyco Healthcare Group LP or its affiliate © 2005 Tyco Healthcare Group LP. All rights reserved.<br />
C-<strong>AN</strong>-<strong>Combtub</strong>/<strong>GB</strong><br />
Tyco Healthcare UK Limited<br />
154 Fareham Road<br />
GOSPORT Hampshire<br />
PO 13 0AS - UK<br />
Tel: +44 (0) 1329 224226<br />
Fax: +44 (0) 1329 224334<br />
www.tycohealthcare.co.uk<br />
uksales@emea.tycohealthcare.com<br />
Tyco Healthcare<br />
Regional Export<br />
154 Fareham Road<br />
GOSPORT Hampshire<br />
PO 13 0AS UK<br />
Tel: +44 (0) 1329 224258<br />
Fax: +44 (0) 1329 224086<br />
Tyco Healthcare Nederland BV<br />
Hogeweg 105<br />
NL-5301 LL ZALTBOMMEL<br />
The NETHERL<strong>AN</strong>DS<br />
Tel: +31 418 57 66 00<br />
Fax: +31 418 57 67 91<br />
TycoHealthNed@TycoHealth.com<br />
www.tycohealthcare.nl<br />
Tyco Healthcare Pty Ltd<br />
Corporate Park North<br />
379 Roan Cresent<br />
Randjespark, Midrand<br />
P.O. Box 8108 - SA<br />
Halfway House, 1685<br />
Tel: +27 (0) 11 542 9500<br />
Fax: +27 (0) 11 542 9501<br />
Tyco Healthcare Belgium N.V.-S.A.<br />
Generaal de Wittelaan 9/5<br />
B-2800 MECHELEN<br />
BELGIUM<br />
Tel: +32 (0) 15 29 44 50<br />
Fax: +32 (0) 15 29 44 55<br />
tycohealthcare.Belgium@emea.<br />
tycohealthcare.com<br />
Tyco Healthcare Norden AB<br />
Hemvärnsgatan 9, Box 54<br />
SE-171 74 Solna<br />
SWEDEN<br />
Tel: +46 (0) 8585 60 505<br />
Fax: +46 (0) 8585 60 506<br />
Tyco Healthcare Danmark A/S<br />
Langebrogade 6E, 4.th.<br />
DK-1411 Køpenhavn K<br />
DENMARK<br />
Tel: +45-70 27 52 50<br />
Kundeservice +45-70 27 52 50<br />
Fax: +45-70 27 51 50<br />
Tyco Healthcare Norge AS<br />
Lensmannslia 4<br />
N-1386 Asker<br />
NORWAY<br />
Tel: +47-66 85 22 00<br />
Fax: +47-66 85 22 11<br />
Tyco Healthcare Finland Oy<br />
Pursimiehenkatu 26-30 C,<br />
PB 407<br />
FIN-00150 Helsinki<br />
FINL<strong>AN</strong>D<br />
Puh +358-(0)9-6226 840<br />
Asiakaspalvelu<br />
+358-(0)9-6226 8418<br />
Fax: +358-(0)9-6226 8460<br />
06/2005