and the cannula barb.<br />

Battery Error<br />

Service Needed<br />

External Power<br />

Low<br />

The system is not properly<br />

communicating with the battery<br />

The compressor has reached<br />

maximum speed<br />

The input voltage is less than 17V<br />

If the problem persists, return the<br />

concentrator for service.<br />

Remove and reinsert the battery to<br />

clear the error. If the error persists,<br />

replace the battery<br />

The compressor is at its end of life or<br />

there is a leak in the concentrator<br />

Remove power to the concentrator and<br />

remove and reinsert the battery.<br />

Try another external power source<br />

such as DC power<br />

6. CONTACTS FOR MORE INFORMATION<br />

<strong>Inogen</strong> Website: Patient Issues: http://www.inogenone.com<br />

Corporate:<br />

http://www.inogen.net<br />

<strong>Inogen</strong> :<br />

326 Bollay Dr<br />

Goleta, CA 93117<br />

1-866-765-2800<br />

clientservices@inogen.net<br />

CONFIDENTIAL and<br />

PROPRIETARY<br />

96-02735-00-01 revC<br />

INOGEN ONE <strong>G2</strong> OXYGEN CONCENTRATOR<br />

TECHNICAL MANUAL<br />