series GPC - Merkantile
series GPC - Merkantile
series GPC - Merkantile
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
<strong>series</strong> <strong>GPC</strong><br />
S2C143E3-GPG3<br />
• PRESSURE CONTROL<br />
Ranges <strong>GPC</strong>3-A002 <strong>GPC</strong>3-A005 <strong>GPC</strong>3-A010 <strong>GPC</strong>3-A025 <strong>GPC</strong>3-A050 <strong>GPC</strong>3-A100<br />
Absolute 1.25 hPa/350 kPa 2.5 hPa/600 kPa 0.5 kPa/1.1 MPa 1.25 kPa/2.6 MPa 2. 5 kPa/5.1 MPa 5 kPa/10.1 MPa<br />
Gauge 0 hPa/250 kPa 0 hPa/500 kPa 0 kPa/1 MPa 0 kPa/2.5 MPa 0 kPa/5 MPa 0 kPa/10 MPa<br />
Absolute Sub<br />
Ranges<br />
Gauge Sub<br />
Ranges<br />
0.125 -180 kPa<br />
0.125 - 260 kPa<br />
0.125 - 350 kPa<br />
0 - 80 kPa<br />
0 - 160 kPa<br />
0 - 250 kPa<br />
0.25 - 260 kPa<br />
0.25 - 350 kPa<br />
0.25 - 600 kPa<br />
0 - 160 kPa<br />
0 - 250 kPa<br />
0 - 500 kPa<br />
0.5 - 450 kPa<br />
0.5 - 800 kPa<br />
0.5 – 1,100 kPa<br />
0 - 350 kPa<br />
0 - 700 kPa<br />
0 – 1,000 kPa<br />
0 - 800 kPa<br />
0 - 1,600 kPa<br />
0 - 2,500 kPa<br />
0 - 800 kPa<br />
0 - 1,600 kPa<br />
0 - 2,500 kPa<br />
2.5 – 1,700 kPa<br />
2.5 – 2,600 kPa<br />
2.5 – 5,100 kPa<br />
0 – 1,600 kPa<br />
0 – 2,500 kPa<br />
0 – 5,000 kPa<br />
5 – 3,600 kPa<br />
5 – 7,100 kPa<br />
5 – 10,100 kPa<br />
0 – 3,500 kPa<br />
0 – 7,000 kPa<br />
0 –10,000 kPa<br />
• METROLOGICAL SPECIFICATION<br />
Accuracy*: ± 0.005% SR f.s.**<br />
Stability (90 days): ± 0.002% SR f.s.**<br />
Response time: < 0.2 seconds<br />
Temperature effect: fully compensated<br />
Resolution: 1 ppm f.s.<br />
* Accuracy statements include the combined errors of<br />
repeatability, linearity, hysteresis and temperature after<br />
zeroing at the operating temperature<br />
** SRFS stands for sub range full scale<br />
• TECHNICAL SPECIFICATIONS<br />
Control accuracy<br />
± 0.002% FS<br />
Dwell time<br />
Fast mode: 10 sec max between two<br />
points within 0.5% FS and FS<br />
Fine mode: user programmable<br />
Stability control<br />
Fast mode: 15 sec after a pressure point is<br />
set for stabilisation within ± 0.01% FS<br />
Fine mode: stabilisation within the dwell<br />
time<br />
Circuit volume<br />
optimally up to 500 cc (Circuits with larger<br />
volumes increase the generation time)<br />
• TECHNICAL DATA<br />
Operating temperature : 10 to 45°C<br />
Pressure medium : clean, dry noncorrosive<br />
gas<br />
Computer interfaces: RS232 and IEEE488<br />
Power supply: 115-230VAC, 50-60Hz<br />
Dimensions (LxWxH) : 48.8 x 50 x 17.7 cm<br />
Weight : 22.5 kg<br />
• STANDARD SUPPLY<br />
Standard Supply includes:<br />
• Basic Unit<br />
• Power Cable<br />
• Certificate of Compliance<br />
• Certificate “COFRAC”<br />
• Operating Manual<br />
• OPTIONS<br />
Pascal TM Calibration Software Package<br />
• Computer Interface to <strong>GPC</strong> & DPG <strong>series</strong><br />
• Calibration Data Acquisition<br />
• Calibration Test Report print-out<br />
• Calibration report storage<br />
• Reference: Pascal.18d.demo<br />
• RECOMMENDED COMPLEMENTS<br />
Digital Primary Pressure Standard Mod.<br />
DPG5 or DPG8<br />
• ACCESSORIES<br />
Gas Bottle (as per operating requirements<br />
(by sub-suppliers)<br />
Pressure Regulator, Manifold, Pneumatic<br />
Connections defined by application and<br />
requirement on a job by job basis<br />
• SCANDURA: Synonymous of Calibration<br />
Scandura is a customer’s oriented enterprise that<br />
stays aside all Instrument Maintenance Engineers<br />
to better serve their requirements and making<br />
their job more accurate and valuable.<br />
Although nearly forty years of challenging and<br />
competitive activity all over the world, have<br />
definitively given Scandura an international role,<br />
the roots and the fruits remain “Made in Italy”. A<br />
symbol of sophisticated solutions ensured by an<br />
experienced, creative, flexible and open minded<br />
team of people that is pleased to respond to your<br />
needs whatever your needs are.<br />
Dott. Ing.SCANDURA & FEM S.r.l.<br />
Via Ambrosoli 8<br />
20090 RODANO MILLEPINI – ITALY<br />
++39 02 95320021<br />
Fax ++39 02 95328231<br />
E-Mail: scandura@scandura.it<br />
www.scandura.it<br />
Printed in Italy<br />
Agent:<br />
We reserve the rights to alter at any time the technical<br />
specifications.<br />
SXY917E0-CAT