An open system for accurate measurement<br />

of soil <strong>CO</strong> 2 flux<br />

<strong>CFX</strong>-2<br />

<strong>Flux</strong> <strong>System</strong><br />

<strong>Soil</strong> <strong>CO</strong> 2<br />

For over 20 years, PP <strong>System</strong>s has<br />

been manufacturing high quality<br />

instrumentation for researchers worldwide.<br />

The <strong>CFX</strong>-2 system is suitable for<br />

unattended operation in the field and<br />

designed to give accurate measurement<br />

of soil net <strong>CO</strong> 2 flux. It is an "Open<br />

<strong>System</strong>" where the measurements are<br />

based on concentration differences<br />

between air entering and leaving the<br />

chamber and the flow rate. The<br />

chamber is designed to minimize the<br />

internal over-pressure to such a degree<br />

that normal soil-atmosphere exchanges<br />

are maintained. It also has an integral<br />

stainless steel ring to provide a good<br />

seal with the soil surface.<br />

<strong>System</strong> Features<br />

• Stand-alone operation<br />

• Lightweight and field portable<br />

• Accurate, integral <strong>CO</strong> 2 analyzer and<br />

H 2 O sensor<br />

• Open system measurement<br />

• User friendly operation<br />

• Simple setup<br />

• Full data logging, storage and output<br />

of data<br />

The Control Interface Module (CIM)<br />

includes an integrated <strong>CO</strong> 2 /H 2 O analyzer,<br />

a mass flow controlled air supply providing<br />

100 l/min of ambient air to an opaque<br />

chamber, datalogging/storage capability<br />

and keyboard for setting up the system<br />

and retrieval of stored data.<br />

Close-up view of the Control Interface<br />

Module (CIM)<br />

Data Sheet


Technical Specifications<br />

For Further Information, Please<br />

Contact Us At:<br />

North America<br />

PP <strong>System</strong>s<br />

110 Haverhill Rd., Suite 301<br />

Amesbury, MA 01913 U.S.A.<br />

Tel: +1 978-834-0505<br />

Fax: +1 978-834-0545<br />

Europe<br />

PP <strong>System</strong>s<br />

Unit 2, Glovers Court<br />

Bury Mead Rd.<br />

Hitchin, Herts SG5 1RT UK<br />

Tel: +44 1462-453411<br />

Fax: +44 1462-431090<br />

Email: sales@ppsystems.com<br />

URL: www.ppsystems.com<br />

08/14/06<br />

Analysis Method<br />

Non-dispersive infrared, configured as an<br />

absolute absorptiometer with microprocessor<br />

control of linearization.<br />

Measurement Range<br />

<strong>CO</strong> 2 : 0-2,000 ppm (µmol mol -1 )<br />

H 2 O: 0-100%<br />

Measurements are automatically corrected for<br />

temperature and pressure.<br />

Accuracy<br />

<strong>CO</strong> 2 : Better than 1% of span concentration<br />

over the calibrated range, but limited by the<br />

accuracy of the calibration gas mixture.<br />

H 2 O: 2%<br />

Linearity<br />

Better than 1.0% throughout the range.<br />

Stability<br />

Automatic Zero at regular intervals, corrects<br />

for sample cell contamination, source and<br />

detector ageing and pre-amplifier gain<br />

changes.<br />

Sampling Pump<br />

Integral DC pump.<br />

Air Filter<br />

Filtered sample line (hydrophobic).<br />

Calibration<br />

Default value preset in factory (built-in initialization).<br />

Automatic calibration by keypad if<br />

required.<br />

Real Time Clock<br />

Accuracy > 1 minute per month at 25 o C,<br />

operating temperature 0-70 o C. Automatic<br />

correction for month end and leap years.<br />

Recording<br />

Manual (by keypress) or automatic at user<br />

selected intervals between 1 and 250 minutes.<br />

Data Storage<br />

512K Battery backed RAM (1,000 records).<br />

Environmental Sensor Interface<br />

Input available for use with the <strong>CFX</strong>-2 <strong>Soil</strong><br />

Respiration Chamber.<br />

Digital (RS232) Output<br />

9600 baud/8 data bits, 1 start bit/2 stop bits/<br />

no parity. ASCII format.<br />

Display<br />

High contrast 2 x 16 character LCD.<br />

Power Supply<br />

External 12V DC (User supplied)<br />

Gas Connections<br />

Two gas ports (Reference/Chamber) for use<br />

with 1/8” (.125) ID tubing.<br />

Housing<br />

High impact, IP65 enclosure with handle.<br />

Dimensions<br />

Control Interface Module:<br />

25.5 cm (H) x 9 cm (D) x 18 cm (W)<br />

Air Supply Chimney:<br />

63 cm (H) x 5.5 cm (D)<br />

<strong>CFX</strong>-2 <strong>Soil</strong> Respiration Chamber:<br />

20 cm (Diameter) x 12 cm (Height)<br />

Surface Area: Approx. 350 cm 2<br />

Weight<br />

Control Interface Module:<br />

2.8 kg. (including chimney)<br />

<strong>CFX</strong>-2 <strong>Soil</strong> Respiration Chamber:<br />

1.7 kg.<br />

Optional Accessory<br />

<strong>CFX</strong> <strong>Soil</strong> Temperature Probe<br />

PP <strong>System</strong>s is continuously updating its<br />

products and reserves the right to amend<br />

product specifications without notice.