SARA – AUTONOMOUS SPECTROSCOPIC STATION AGS71xS/M<br />
The SARA Autonomous Spectroscopic Station <strong>AGS710</strong>S/M / AGS711S/M is a radiation monitoring station which combines both<br />
spectroscopic gamma detection with data logging capabilities and autonomous operation due to solar power supply and wireless data<br />
transmission. <strong>Data</strong> communication between the <strong>AGS710</strong>S/M / AGS711S/M and Network Monitoring Center (NMC) uses the GPRS<br />
service of cellular networks through a secure VPN. Two models are available: the AGS71xS for stationary use is designed for fixed<br />
installation and unlimited operation that is achieved by the solar panel and a back-up battery. The mobile station AGS71xM can be<br />
quickly deployed in case of emergency (battery-based temporary use). The station consists of a detector unit and a base unit. This<br />
independency from any wired connections eases the installation of the station at any place of interest and permits simple relocation<br />
if necessary. SARA is able to detect even minor changes of the composition of the nuclear spectra in the environment. This improves<br />
significantly the recognition of artificial isotopes, which is very important for the radiation early warning system, for example. Not<br />
only does it support the fast detection of artificial radiation but it can also identify the nuclear isotopes. It is designed for outdoor use<br />
even in harsh environments and for continuous operation without any maintenance. The hermetically sealed detector unit<br />
guarantees optimal protection for the detector and the electronics against environmental conditions. The NaI(Tl) -based scintillation<br />
detector together with the MCA (multi-channel analyzer) provide high energy resolution under ambient temperature operation<br />
conditions. The integrated embedded Linux-PC enables online isotope identification and versatile data exchange through several<br />
interfaces. The standardized ANSI N42.42 protocol enables the use of many spectra evaluation software programs. The integrated<br />
web server facilitates easy data access using a web browser. SARA also measures the gamma dose rate. For an extended gamma do se<br />
rate range, the model AGS711S/M comes with an additional Geiger-Müller-detector (GM). Total gamma dose rate measurement of<br />
SARA is calibrated according to Ambient Dose Equivalent Rate H*(10).<br />
GENERAL TECHNICAL DATA<br />
Detector type<br />
NaI(Tl) Scintillator<br />
(for LaBr see IGS91x)<br />
Detector size 1.5” x 1.5”<br />
Dose rate range<br />
1 nSv/h to 100 µSv/h<br />
(0.1 µrem/h to 10 mrem/h)<br />
Accuracy ±10 %<br />
Extended (AGS711S/M) 1 nSv/h to 100 mSv/h<br />
(0.1 µrem/h to 10 rem/h)<br />
Detection limit *<br />
(typ. background,10 min)
FUNCTIONS<br />
Acquisition of the gamma spectra every 5 or 10 min<br />
(storage of 144 values = 0.5 resp. 1 day)<br />
In-situ isotope identification every 5 or 10 min<br />
Calculation of 1 h average spectra<br />
(storage of 96 values = 4 days)<br />
Calculation of 24 h average spectra<br />
(storage of 32 values = 32 days)<br />
Detection of gamma dose rate every 5 or 10 min<br />
(storage of 1024 values = 3.5 resp. 7 days)<br />
Calculation of 1h average of gamma dose rate<br />
(storage of 1024 values = 42 days)<br />
Extended dose rate range with additional GM detector<br />
(model AGS711S/M)<br />
Free configurable isotope library<br />
Isotope based alarm management<br />
Integrated detector accuracy test<br />
Temperature stabilization of energy spectra based on K40<br />
Supervision of detectors and electronic devices<br />
Overload protection of detector<br />
Precipitation detection (with optional rain sensor)<br />
<strong>Data</strong> access and parameter setting with web browser<br />
FEATURES<br />
Fast detection of very low artificial radiation<br />
Online spectrum analysis<br />
In-situ isotope identification<br />
Standardized data protocol ANSI N42.42 (XML-based)<br />
No infrastructure required<br />
Absolute unattended operation<br />
Easy to maintain - neither consumables nor wear parts<br />
Operation under harsh environmental conditions<br />
Rugged design (IP68 / IP65)<br />
Wireless data communication (GPRS)<br />
Easy to install (model S) or to deploy (model M) & set-up<br />
Mobile type and application supported by GPS (option)<br />
Detector verification supported automatically (TS710)<br />
SPECTRA OF AM-241, CS-137 AND CO-60<br />
RELATED PRODUCTS<br />
IGS71x Spectroscopic Gamma Detector<br />
AGS71xF Spectroscopic Monitoring Station<br />
<strong>AGS710</strong>S/M/07EN/08/2013<br />
FUNCTION DIAGRAM OF MONITORING STATION<br />
TECHNICAL DATA GM DETECTOR (AGS711S/M)<br />
Detector<br />
Range<br />
GM detector 70 017 A<br />
50 µSv/h to 100 mSv/h<br />
(5 mrem/h to 10 rem/h)<br />
Accuracy ±15 %<br />
Sensitivity<br />
36 counts min -1 / µSv/h<br />
(360 cpm per mrem/h)<br />
Intrinsic background < 140 nSv/h (< 14 µrem/h)<br />
Energy range<br />
50 keV to 1.3 MeV<br />
INTERFACES<br />
GPRS<br />
LAN Ethernet 10 Mbit/s<br />
RS232 (for service)<br />
OPTIONS<br />
GPS / Mobile application supported automatically<br />
RD: Rain detector<br />
GM: Geiger Müller detector / model AGS711S/M<br />
Solar power supply<br />
Mounting platform (removable) for model S<br />
Transportation box for model M<br />
TS710: Test Set for detectors verification and testing<br />
ORDERING INFORMATION<br />
The Spectroscopic Monitoring Station AGS71xS/M consists of a<br />
Spectroscopic Gamma Detector IGS71x and the Base Unit. The<br />
station can be elected as follows:<br />
SOL<br />
<strong>AGS710</strong>S-SOL<br />
x<br />
<strong>AGS710</strong>S<br />
<strong>AGS710</strong>M-SOL<br />
x<br />
<strong>AGS710</strong>M<br />
Additionally the IGS71x can be selected as follows:<br />
LAN GPRS GPS RD<br />
IGS710-GPRS-GPS-RD x x x x<br />
IGS710-LAN-GPRS x x<br />
IGS710-LAN-GPRS-GPS x x x<br />
IGS710-LAN-GPRS-RD x x x<br />
IGS711 comes with additional GM detector. In this case use also<br />
AGS711S/M as station type.<br />