SARA – AUTONOMOUS SPECTROSCOPIC STATION AGS71xS/M<br />

The SARA Autonomous Spectroscopic Station <strong>AGS710</strong>S/M / AGS711S/M is a radiation monitoring station which combines both<br />

spectroscopic gamma detection with data logging capabilities and autonomous operation due to solar power supply and wireless data<br />

transmission. <strong>Data</strong> communication between the <strong>AGS710</strong>S/M / AGS711S/M and Network Monitoring Center (NMC) uses the GPRS<br />

service of cellular networks through a secure VPN. Two models are available: the AGS71xS for stationary use is designed for fixed<br />

installation and unlimited operation that is achieved by the solar panel and a back-up battery. The mobile station AGS71xM can be<br />

quickly deployed in case of emergency (battery-based temporary use). The station consists of a detector unit and a base unit. This<br />

independency from any wired connections eases the installation of the station at any place of interest and permits simple relocation<br />

if necessary. SARA is able to detect even minor changes of the composition of the nuclear spectra in the environment. This improves<br />

significantly the recognition of artificial isotopes, which is very important for the radiation early warning system, for example. Not<br />

only does it support the fast detection of artificial radiation but it can also identify the nuclear isotopes. It is designed for outdoor use<br />

even in harsh environments and for continuous operation without any maintenance. The hermetically sealed detector unit<br />

guarantees optimal protection for the detector and the electronics against environmental conditions. The NaI(Tl) -based scintillation<br />

detector together with the MCA (multi-channel analyzer) provide high energy resolution under ambient temperature operation<br />

conditions. The integrated embedded Linux-PC enables online isotope identification and versatile data exchange through several<br />

interfaces. The standardized ANSI N42.42 protocol enables the use of many spectra evaluation software programs. The integrated<br />

web server facilitates easy data access using a web browser. SARA also measures the gamma dose rate. For an extended gamma do se<br />

rate range, the model AGS711S/M comes with an additional Geiger-Müller-detector (GM). Total gamma dose rate measurement of<br />

SARA is calibrated according to Ambient Dose Equivalent Rate H*(10).<br />

GENERAL TECHNICAL DATA<br />

Detector type<br />

NaI(Tl) Scintillator<br />

(for LaBr see IGS91x)<br />

Detector size 1.5” x 1.5”<br />

Dose rate range<br />

1 nSv/h to 100 µSv/h<br />

(0.1 µrem/h to 10 mrem/h)<br />

Accuracy ±10 %<br />

Extended (AGS711S/M) 1 nSv/h to 100 mSv/h<br />

(0.1 µrem/h to 10 rem/h)<br />

Detection limit *<br />

(typ. background,10 min)


FUNCTIONS<br />

Acquisition of the gamma spectra every 5 or 10 min<br />

(storage of 144 values = 0.5 resp. 1 day)<br />

In-situ isotope identification every 5 or 10 min<br />

Calculation of 1 h average spectra<br />

(storage of 96 values = 4 days)<br />

Calculation of 24 h average spectra<br />

(storage of 32 values = 32 days)<br />

Detection of gamma dose rate every 5 or 10 min<br />

(storage of 1024 values = 3.5 resp. 7 days)<br />

Calculation of 1h average of gamma dose rate<br />

(storage of 1024 values = 42 days)<br />

Extended dose rate range with additional GM detector<br />

(model AGS711S/M)<br />

Free configurable isotope library<br />

Isotope based alarm management<br />

Integrated detector accuracy test<br />

Temperature stabilization of energy spectra based on K40<br />

Supervision of detectors and electronic devices<br />

Overload protection of detector<br />

Precipitation detection (with optional rain sensor)<br />

<strong>Data</strong> access and parameter setting with web browser<br />

FEATURES<br />

Fast detection of very low artificial radiation<br />

Online spectrum analysis<br />

In-situ isotope identification<br />

Standardized data protocol ANSI N42.42 (XML-based)<br />

No infrastructure required<br />

Absolute unattended operation<br />

Easy to maintain - neither consumables nor wear parts<br />

Operation under harsh environmental conditions<br />

Rugged design (IP68 / IP65)<br />

Wireless data communication (GPRS)<br />

Easy to install (model S) or to deploy (model M) & set-up<br />

Mobile type and application supported by GPS (option)<br />

Detector verification supported automatically (TS710)<br />

SPECTRA OF AM-241, CS-137 AND CO-60<br />

RELATED PRODUCTS<br />

IGS71x Spectroscopic Gamma Detector<br />

AGS71xF Spectroscopic Monitoring Station<br />

<strong>AGS710</strong>S/M/07EN/08/2013<br />

FUNCTION DIAGRAM OF MONITORING STATION<br />

TECHNICAL DATA GM DETECTOR (AGS711S/M)<br />

Detector<br />

Range<br />

GM detector 70 017 A<br />

50 µSv/h to 100 mSv/h<br />

(5 mrem/h to 10 rem/h)<br />

Accuracy ±15 %<br />

Sensitivity<br />

36 counts min -1 / µSv/h<br />

(360 cpm per mrem/h)<br />

Intrinsic background < 140 nSv/h (< 14 µrem/h)<br />

Energy range<br />

50 keV to 1.3 MeV<br />

INTERFACES<br />

GPRS<br />

LAN Ethernet 10 Mbit/s<br />

RS232 (for service)<br />

OPTIONS<br />

GPS / Mobile application supported automatically<br />

RD: Rain detector<br />

GM: Geiger Müller detector / model AGS711S/M<br />

Solar power supply<br />

Mounting platform (removable) for model S<br />

Transportation box for model M<br />

TS710: Test Set for detectors verification and testing<br />

ORDERING INFORMATION<br />

The Spectroscopic Monitoring Station AGS71xS/M consists of a<br />

Spectroscopic Gamma Detector IGS71x and the Base Unit. The<br />

station can be elected as follows:<br />

SOL<br />

<strong>AGS710</strong>S-SOL<br />

x<br />

<strong>AGS710</strong>S<br />

<strong>AGS710</strong>M-SOL<br />

x<br />

<strong>AGS710</strong>M<br />

Additionally the IGS71x can be selected as follows:<br />

LAN GPRS GPS RD<br />

IGS710-GPRS-GPS-RD x x x x<br />

IGS710-LAN-GPRS x x<br />

IGS710-LAN-GPRS-GPS x x x<br />

IGS710-LAN-GPRS-RD x x x<br />

IGS711 comes with additional GM detector. In this case use also<br />

AGS711S/M as station type.<br />

Technical contents are subject to change without notice!<br />

ENVINET GmbH<br />

Hans-Pinsel-Str. 4, 85540 Haar (Munich), Germany<br />

Tel +49 89 456657-0<br />

info@<strong>envinet</strong>.com, www.<strong>envinet</strong>.com www.<strong>envinet</strong>.com