High Resolution Anorectal<br />

Manometry <strong>Balloon</strong><br />

<strong>Attachment</strong> Instructions<br />

Using Polyurethane B<strong>and</strong>s<br />

(Part # A86-4100)<br />

High Resolution Anorectal<br />

Manometry <strong>Balloon</strong><br />

REMOVAL Instructions<br />

Using Polyurethane B<strong>and</strong>s<br />

(Part # A86-4100)<br />

7 1<br />

7. Squeeze the h<strong>and</strong>le of the ligator to release the<br />

b<strong>and</strong> onto the probe. Check to make sure the<br />

b<strong>and</strong> is aligned properly in the groove.<br />

1. Gather equipment: Suture <strong>Removal</strong> Scissors <strong>and</strong><br />

Forceps<br />

2<br />

8<br />

Tie-on B<br />

2. Gently stabilize b<strong>and</strong> with forceps taking care not<br />

to apply pressure to the sensors.<br />

8. Slide the tip of the balloon down so that tie-on B<br />

is visible in the nipple at the end of the stimulation<br />

balloon<br />

3<br />

9<br />

3. Slide suture removal scissors under each b<strong>and</strong><br />

<strong>and</strong> cut. Remove balloon from probe <strong>and</strong> discard<br />

balloon.<br />

9. In the same manner as for tie-on A, attach a b<strong>and</strong><br />

onto the probe at tie-on B.<br />

S<strong>and</strong>hill <strong>Scientific</strong>, Inc<br />

9150 Commerce Center Circle, #500<br />

Highl<strong>and</strong>s Ranch, CO 80129 USA<br />

800.468.4556 303.470.7020<br />

Fax: 303.470.2975<br />

PM-11-002, Rev 1.1