10.01.2015 Views

MRI Artifacts: Mechanism and Control

MRI Artifacts: Mechanism and Control

MRI Artifacts: Mechanism and Control

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

<strong>MRI</strong> <strong>Artifacts</strong>: <strong>Mechanism</strong> <strong>and</strong> <strong>Control</strong><br />

Chun Ruan<br />

Abstract<br />

A wide variety of artifacts is routinely encountered on MR images. This article<br />

presents the cause, appearance, diagnostic effect, <strong>and</strong> available remedies for the<br />

artifacts that are most frequently observed on MR images <strong>and</strong> are of greatest clinical<br />

significance. Combined with routine preventive maintenance of imaging equipment,<br />

consistent quality control, <strong>and</strong> appropriate selection of imaging parameters, awareness<br />

of the manifestations of these artifacts will allow image quality <strong>and</strong> diagnostic<br />

interpretation to be optimized.<br />

I. Introduction<br />

Magnetic resonance imaging (<strong>MRI</strong>) is widely used in medical diagnosis for its various<br />

advantageous features, such as high-resolution capability, the ability to produce an<br />

arbitrary anatomic cross-sectional image, <strong>and</strong> high tissue contrast. Unfortunately,<br />

there are many potential sources of image artifacts associated with the technology of<br />

<strong>MRI</strong>. They can potentially degrade images sufficiently to cause inaccurate diagnosis.<br />

Many MR artifacts are neither obvious nor underst<strong>and</strong>able from previous experience<br />

with conventional types of imaging. While some MR artifacts are machine specific,<br />

the majority are inherent in the imaging method itself.<br />

MR imaging artifacts can be grouped into two general categories. First, there are<br />

artifacts that are hardware related. These artifacts are relatively<br />

uncommon—fortunately, because they are often difficult to diagnose <strong>and</strong> usually<br />

require service personnel to correct. The second category consists of artifacts related<br />

to the patient or under operator control. This category is encountered much more<br />

commonly <strong>and</strong> may often be easily prevented or corrected once they are recognized.<br />

II. Technique <strong>and</strong> Methods<br />

A. Motion <strong>Artifacts</strong><br />

Motion is the most prevalent source of MR imaging artifacts. As the name implies,<br />

motion artifacts are caused by motion of the imaged object or a part of the imaged<br />

object during the imaging sequence. Motion results in two effects on MR images.<br />

View-to-view effects are caused by motion that occurs between the acquisitions of<br />

successive phase-encoding steps. The inconsistent location <strong>and</strong> signal intensity of<br />

spins that move as phase-encoding data are acquired result in phase errors. When<br />

motion is periodic—that is, occurs in a regular pattern—the result is complete or<br />

incomplete replication of the moving tissue, commonly referred to as ghosting<br />

artifacts. These artifacts are observed along the phase-encoding direction of the<br />

image, regardless of the direction in which the motion actually occurred. Periodic<br />

physiologic motions that commonly result in ghosting artifacts include cardiac<br />

motion, respiratory motion, vascular pulsation, <strong>and</strong> cerebrospinal fluid (CSF)<br />

pulsation. 1<br />

Motion occurring between the time of radiofrequency (RF) excitation <strong>and</strong> echo<br />

collection (i.e., within-view) results in a lack of coherent phase among the population<br />

1


of moving spins at the time of echo formation. 2 This incoherence manifests as<br />

blurring <strong>and</strong> increased image noise. Unlike phase errors encountered in view-to-view<br />

motion effects, this within-view effect is expressed throughout the image. It is most<br />

frequently associated with r<strong>and</strong>om motion, as can occur with gastrointestinal<br />

peristalsis, swallowing, coughing, eye motion, <strong>and</strong> gross patient movement. 3<br />

1. Respiratory motion<br />

Respiratory motion results in ghosting artifacts <strong>and</strong> blurring that can obscure or<br />

simulate lesions. A variety of methods have been used to reduce the effect of<br />

respiratory motion artifacts. Mechanical methods, such as use of an abdominal or<br />

thoracic binder or taking images with the patient in a prone position, are intended to<br />

restrict the amplitude of respiratory motion. However, these maneuvers often produce<br />

patient discomfort <strong>and</strong> may therefore have counterproductive effects; Signal<br />

averaging is the use of multiple data acquisitions to improve the signal-to-noise ratio<br />

(SNR) of the image ( Fig.1). In this process, the prominence of ghosting artifacts is<br />

reduced by approximately the square root of the number of signal averages obtained.<br />

Of course, imaging time increases linearly with the number of signal averages; With<br />

respiratory triggering, data are collected only during a limited portion of the<br />

respiratory cycle, usually near end-expiration, when respiratory movement is minimal.<br />

The major drawback of this technique is its marked prolongation of imaging time,<br />

because so much of the time is not being used productively for data acquisition;<br />

Respiratory ordered phase encoding involves monitoring the patient’s respiratory<br />

cycle during imaging using a bellows device. Unlike respiratory triggering, however,<br />

with this method data are collected in a continuous fashion. This method does not<br />

restrict the operator’s selection of TR, <strong>and</strong> it does not significantly extend imaging<br />

time; Gradient moment nulling involves the application of additional gradient pulses<br />

to correct for phase shifts among a population of moving protons at the time of echo<br />

collection. This method corrects for constant-velocity motion <strong>and</strong> helps reduce the<br />

signal loss <strong>and</strong> ghosting associated with such movement. Use of gradient moment<br />

nulling requires a nominal TE, which may preclude its use on T1-weighted pulse<br />

sequences. This method does not prolong image acquisition time; Another method to<br />

reducing the signal intensity of moving tissue is the use of fat suppression methods.<br />

With this method, not only subcutaneous fat but also mediastinal, mesenteric,<br />

retroperitoneal, <strong>and</strong> other stores of internal fat are suppressed <strong>and</strong> are thus less<br />

capable of generating ghosting artifacts.<br />

2


2. Cardiac motion<br />

Figure 1. The image on the left exhibits respiratory motion as blurring of the structures<br />

as well as motion-induced ghosting. By using multiple averages motion can be reduced in<br />

the same way that multiple averages increase the signal to noise ratio. The image on the<br />

right was obtained with 16 averages.<br />

Cardiac motion produces a series of ghost artifacts along the phase-encoding<br />

direction of the image, in addition to blurring <strong>and</strong> signal loss of cardiac <strong>and</strong><br />

juxtacardiac structures. 4 The major approach for reducing cardiac motion artifacts is<br />

electrocardiographic triggering, in which data collection is synchronized with cardiac<br />

phase ( Fig. 2). This synchronization enables cardiac tissue to be located in a<br />

consistent position as each successive phase-encoding step is acquired, resulting in<br />

increased tissue signal intensity <strong>and</strong> decreased phase errors. Other approaches include<br />

the use of fast imaging sequences that reduce the opportunity for motion during data<br />

acquisition, gradient moment nulling, <strong>and</strong> spatial RF presaturation pulses.<br />

3. Vascular Pulsation<br />

Figure 2. The image on the left was acquired without any form of motion compensation<br />

technique for cardiac motion. The image on the right was obtained using cardiac gating.<br />

Vascular pulsation artifacts are recognized by their alignment with the responsible<br />

vessel along the phase-encoding direction of the image. These artifacts reproduce the<br />

cross-sectional size <strong>and</strong> shape of the responsible vessel, but not necessarily its signal<br />

intensity. Spatial RF presaturation pulses applied outside the field of view help reduce<br />

the signal intensity of inflowing blood <strong>and</strong>, hence reduce the resultant pulsation<br />

artifact. Other practical methods for reducing the prominence of vascular pulsation<br />

artifacts include positioning the section of interest in the middle of a multisection<br />

acquisition, thus reducing any potential entry phenomenon, <strong>and</strong> maximizing the<br />

saturation of flowing spins.<br />

B. Susceptibility <strong>Artifacts</strong><br />

Susceptibility artifacts occur as the result of microscopic gradients or variations in<br />

the magnetic field strength that occurs near the interfaces of substance of different<br />

magnetic susceptibility ( Fig. 3). Large susceptibility artifacts are commonly seen<br />

surrounding ferromagnetic objects inside of diamagnetic materials (such as the human<br />

body). These gradients cause dephasing of spins <strong>and</strong> frequency shifts of the<br />

surrounding tissues. The net results are bright <strong>and</strong> dark areas with spatial distortion of<br />

surrounding anatomy. These artifacts are worst with long echo times <strong>and</strong> with<br />

3


gradient echo sequences.<br />

Susceptibility artifacts can be made less prominent by performing imaging at low<br />

magnetic field strength, using smaller voxels, decreasing echo time, <strong>and</strong> increasing<br />

receiver b<strong>and</strong>width. Gradient-echo <strong>and</strong> echo-planar sequences should be avoided,<br />

because they accentuate susceptibility artifacts. The use of spin-echo <strong>and</strong> particularly<br />

fast spin-echo sequences should be considered.<br />

C. Chemical Shift <strong>Artifacts</strong><br />

Figure 3. An axial <strong>MRI</strong> of the head in a patient with mascara on her eyelids.<br />

Susceptibility artifacts from the mascara obscure the front half of the<br />

globes.<br />

A chemical shift artifact is caused by the difference in chemical shift (Larmor<br />

frequency) of fat <strong>and</strong> water. The artifact manifests itself as a misregistration between<br />

the fat <strong>and</strong> water pixels in an image ( Fig. 4). The effect being that fat <strong>and</strong> water spins<br />

in the same voxel are encoded as being located in different voxels. The magnitude of<br />

the effect is proportional on the magnitude of the Bo field <strong>and</strong> inversely proportional<br />

to the sampling rate in the frequency encoding direction. For a constant sampling rate,<br />

the larger Bo, the greater the effect. 5 Chemical shift artifacts are typically observed<br />

along the frequency-encoding direction but can also occur along the slice-selection<br />

direction of the image.<br />

Chemical shift artifact can be reduced by performing imaging at low magnetic field<br />

strength, by increasing receiver b<strong>and</strong>width, or by decreasing voxel size. The artifacts<br />

tend to be more prominent on T2-weighted than on T1-weighted images. Fat<br />

suppression methods often eliminate visible artifacts, <strong>and</strong> gradient reorientation can<br />

redirect chemical shift artifacts to another portion of the image.<br />

4


D. Wrap Around <strong>Artifacts</strong><br />

Figure 4. This artifact is shown in an axial image of a kidney where the<br />

bright border along the top of the kidney <strong>and</strong> the dark border along the<br />

bottom of the kidney represent the artifact.<br />

A wrap around artifact is the occurrence of a part of the imaged anatomy, which is<br />

located outside of the field of view, inside of the field of view ( Fig. 5). This artifact is<br />

caused by the selected field of view being smaller than the size of the imaged object.<br />

Or more specifically the digitization rate is less than the range of frequencies in the<br />

FID or echo. 6 The solution to a wrap around artifact is to choose a larger field of<br />

view, adjust the position of the image center, or select an imaging coil which will not<br />

excite or detect spins from tissues outside of the desired field of view.<br />

Figure 5. The first image shows wrap-around of the back of the head on to the front of the<br />

head, where the phase-encoded direction is anterior-posterior. The second image has the phase<br />

<strong>and</strong> frequency directions reversed resulting in absence of the aliasing artifact. Oversampling<br />

was used in the frequency direction to eliminate the aliasing.<br />

E. Partial Volume <strong>Artifacts</strong><br />

A partial volume artifact is any artifact which is caused by the size of the image<br />

voxel. It occurs when multiple tissue types are encompassed within a single voxel.<br />

For example, if a small voxel contains only fat or water signal, <strong>and</strong> a larger voxel<br />

might contain a combination of the two, the large voxel possess a signal intensity<br />

equal to the weighted average of the quantity of water <strong>and</strong> fat present in the voxel.<br />

Volume averaging is most likely to occur in the slice-selection direction of the image,<br />

which has the largest voxel dimension. It also occurs when structures are oriented<br />

obliquely to the imaging plane <strong>and</strong> when structures move in <strong>and</strong> out of a given section<br />

during image acquisition. Volume averaging can simulate abnormalities, decrease the<br />

visualization of low-contrast abnormalities, <strong>and</strong> blur or distort affected structures.<br />

Partial volumeaveraging is usually recognized by careful analysis of adjacent<br />

images. Decreasing voxel size, particularly reducing section thickness, can be useful<br />

if further confirmation is required ( Fig.6). Three-dimensional Fourier transform<br />

imaging is particularly useful, because it provides thin sections with no intervening<br />

gaps <strong>and</strong> is conductive to reformatting in alternate imaging planes; Simple acquisition<br />

of additional two-dimensional images in alternate imaging planes is very helpful for<br />

resolving issues relating to partial volume averaging. Retrospective reformatting of<br />

two-dimensional data can also be performed using an interpolation algorithm <strong>and</strong> can<br />

be of further assistance.<br />

5


Figur e 6. These two axial T1-weighted images of the head were obtained at exactly the same location,<br />

yet the second image shows the VII <strong>and</strong> VIII cranial nerves while the first does not. The reason<br />

for the vanishing nerve is explained by partial volume averaging. The first slice was obtained<br />

with a thickness of 10 mm while the second was at a thickness of 3 mm.<br />

F. Gibbs Ringing <strong>Artifacts</strong><br />

Gibbs ringing artifacts are bright or dark lines that are seen parallel <strong>and</strong> adjacent to<br />

borders of abrupt intensity change. The ringing is caused by incomplete digitization of<br />

the echo. This means the signal has not decayed to zero by the end of the acquisition<br />

window, <strong>and</strong> the echo is not fully digitized. This artifact is seen in images when a<br />

small acquisition matrix is used. Solutions include use of a higher resolution imaging<br />

matrix <strong>and</strong> filtration methods ( Fig.7). Gradient reorientation will displace the artifacts<br />

to another portion of the image.<br />

Figure 7. The fine lines visible in the image on the left are due to undersampling of the high spatial<br />

frequencies. This results in a "ringing" type of artifact following these borders in the phase direction (R<br />

to L in this image). This problem can be easily fixed by taking more samples such as the image on the<br />

right with 256 phase encodes.<br />

G. Zebra Stripes<br />

Zebra stripes can be observed along the periphery of gradient-echo images where<br />

there is an abrupt transition in magnetization at the air-tissue interface. They are<br />

accentuated by aliasing that results from the use of a relatively small field of view.<br />

Solutions include exp<strong>and</strong>ing the field of view, using spin-echo pulse sequences, or<br />

using oversampling techniques to reduce aliasing.<br />

H. Slice-overlap <strong>Artifacts</strong><br />

6


The slice-overlap artifact is the loss of signal seen in an image from a multi-angle,<br />

multi-slice acquisition, as is obtained commonly in the lumbar spine. If the slices<br />

obtained at different disk spaces are not parallel, then the slices may overlap. If two<br />

levels are done at the same time, e.g., L4-5 <strong>and</strong> L5-S1, then the level acquired second<br />

will include spins that have already been saturated. This causes a b<strong>and</strong> of signal loss<br />

crossing horizontally in image, usually worst posteriorly. Therefore, overlap of<br />

sections within areas of diagnostic interest should be carefully avoided.<br />

I. RF Overflow <strong>Artifacts</strong><br />

RF overflow artifacts cause a non-uniform, washed-out appearance to an image. This<br />

artifact occurs when the signal received by the scanner from the patient is too intense<br />

to be accurately digitized by the analog-to-digital converter. Autoprescanning usually<br />

adjusts the receiver gain to prevent this from occurring but if the artifact still occurs,<br />

the receiver gain can be decreased manually.<br />

J. Entry Slice Phenomenon<br />

Entry slice phenomenon occurs when unsaturated spins in blood first enter into a<br />

slice or slices. It is characterized by bright signal in a blood vessel (artery or vein) at<br />

the first slice that the vessel enters. Usually the signal is seen on more than one slice,<br />

fading with distance. This artifact has been confused with thrombosis with disastrous<br />

results. The characteristic location <strong>and</strong> if necessary, the use of gradient echo flow<br />

techniques can be used to differentiate entry slice artifacts from occlusions.<br />

K. Zipper <strong>Artifacts</strong><br />

There are various causes for zipper artifacts in images. Most of them are related to<br />

hardware or software problems beyond the physicist immediate control. The zipper<br />

artifacts that can be controlled easily are those due to RF entering the scanning room<br />

when the door is open during acquisition of images. 7 RF from some radio transmitters<br />

will cause zipper artifacts that are oriented perpendicular to the frequency axis of<br />

image. Broad-b<strong>and</strong> noise degrades the entire image, whereas narrow frequency noise<br />

produces linear b<strong>and</strong>s that transverse the phase-encoding direction of the image.<br />

Solutions include identifying <strong>and</strong> removing external RF sources, ensuring that the<br />

door to the imaging room remains closed, <strong>and</strong> verifying the integrity of the magnet<br />

room enclosure <strong>and</strong> associated seals.<br />

L. Cross-Excitation<br />

Cross-excitation is caused by the imperfect shape of RF slice profiles, which leads to<br />

the unintended excitation of adjacent tissue. This excitation results in the saturation of<br />

such tissue, manifest as decreased signal intensity <strong>and</strong> decreased contrast that can<br />

hinder lesion detection. One way to avoid this artifact is to introduce an intersection<br />

gap that is 10% to 50% of the prescribed section thickness. Another method is<br />

interleaved image acquisition, in which odd-numbered sections are initially acquired,<br />

followed by acquisition of even-numbered sections. 8 Also, optimized RF pulses that<br />

have a more rectangular slice profile can be implemented.<br />

7


Figure 8.The dark b<strong>and</strong>s visible on the T1-weighted Spin Echo axial image are due to the<br />

intersection of other axial images through the image. These are acquired in a multi-planar<br />

fashion <strong>and</strong> thus cause pre-excitation (saturation) of the protons in the area where the slices<br />

intersect. The sagittal image illustrates where the slices were obtained <strong>and</strong> how they intersect<br />

posteriorly.<br />

M. Shading<br />

Shading artifacts manifest as foci of relatively reduced signal intensity involving a<br />

portion of the image. Abnormalities contained in the shaded portion of the MR image<br />

may be obscured. 9 There are many potential causes for this artifact, including partial<br />

volume averaging malfunction of the RF transmitter, amplifier, or receiver, excessive<br />

RF absorption, etc. 10, 11 To minimize shading artifacts, the anatomy of interest should<br />

be centered within the magnet, within the coil, <strong>and</strong> within the group of sections to be<br />

acquired.<br />

III. Conclusions<br />

<strong>Artifacts</strong> are common in magnetic resonance imaging. Most occur as a result of<br />

interactions of multiple factors, especially motion. MR artifacts primarily cause image<br />

degradation, although they can occasionally mimic pathological lesions. Some MR<br />

artifacts, such as those caused by periodic respiratory or vascular motion, may<br />

obscure important diagnostic findings. These <strong>and</strong> other MR artifacts may be<br />

minimized in some instances by proper selection of the directions of the phase <strong>and</strong><br />

frequency encoding gradients. The ability to select <strong>and</strong> vary the direction of these<br />

gradients is a useful option in MR imaging. Investigations in the origins of <strong>MRI</strong><br />

artifacts can not only lead to further underst<strong>and</strong>ing of the imaging process itself, but<br />

can also improve the quality <strong>and</strong> diagnostic accuracy of MR examination.<br />

1 Saloner D, “Flow <strong>and</strong> motion,” Magn Reson Imaging Clin N Am 1999 Nov;7(4):699-715.<br />

2 Barish MA <strong>and</strong> Jara H, “Motion artifact control in body MR imaging,” Magn Reson Imaging Clin N<br />

Am 1999 May;7(2):289-301.<br />

3 Hedley M, Yan H, “ Motion artifact suppression: a review of post-processing techniques,” Magn<br />

Reson Imaging 1992;10(4):627-35.<br />

4 Huber ME, Hengesbach D <strong>and</strong> Botnar RM, “ Motion artifact reduction <strong>and</strong> vessel enhancement for<br />

free-breathing navigator-gated coronary MRA using 3D k-space reordering,” Magn Reson Med 2001<br />

Apr;45(4):645-52.<br />

5 Altbach MI, Trouard TP <strong>and</strong> Van<br />

de Walle R, “Chemical-shift imaging utilizing the positional shifts along the readout gradient<br />

direction,” IEEE Trans Med Imaging 2001 Nov;20(11):1156-66.<br />

6 Tsai CM <strong>and</strong> Nishimura DG, “<br />

Reduced aliasing artifacts using variable-density k-space sampling trajectories,” Magn Reson Med<br />

2000 Mar;43(3):452-8.<br />

7 Mugler JP 3rd, “Overview of MR imaging pulse<br />

sequences,” Magn Reson Imaging Clin N Am 1999 Nov;7(4):661-97.<br />

8


8 Clark JA 2nd <strong>and</strong> Kelly WM, “Common artifacts encountered in magnetic resonance imaging,” Radiol<br />

Clin North Am 1988 Sep;26(5):893-920.<br />

9 Jones RW <strong>and</strong> Witte RJ, “Signal intensity artifacts in clinical MR imaging,” Radiographics 2000 May-<br />

Jun;20(3):893-901.<br />

10 Mirowitz SA, “MR imaging artifacts.<br />

Challenges <strong>and</strong> solutions,” Magn Reson Imaging Clin N Am 1999 Nov;7(4):717-32.<br />

11 Mirowitz SA, “MR imaging artifacts. Challenges <strong>and</strong> solutions,” Magn Reson Imaging Clin N Am<br />

1999 Nov;7(4):717-32.<br />

9

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!