O 2 <strong>Transmitter</strong> <strong>4220X</strong>,<br />
advanced 2-<strong>wire</strong> process <strong>transmitter</strong>.<br />
O 2 <strong>4220X</strong><br />
O 2 Measurement input<br />
input for amperometric sensor,<br />
e.g. METTLER TOLEDO oxygen<br />
sensors<br />
O 2<br />
P air<br />
mA /<br />
HART ®<br />
Output 1 (Current Loop)<br />
galvanically isolated current loop (power<br />
supply required), 4 to 20 mA, user defined<br />
as current output for the corresponding<br />
process variables, digital communication<br />
via FSK modulation to HART ®<br />
protocol<br />
Pressure sensor<br />
built-in pressure sensor for<br />
measurement of the ambient<br />
air pressure<br />
Temperature input<br />
input for NTC temperature<br />
sensor<br />
Temp<br />
mA /<br />
Output 2 (Passive)<br />
galvanically isolated output user defined<br />
as 0(4) to 20 mA current output for the<br />
corresponding process variables or as<br />
switching or controller output (current<br />
sink, power supply required)<br />
Exceptionally wide range of<br />
functions<br />
<strong>The</strong> METTLER TOLEDO O 2 <strong>Transmitter</strong><br />
features the following:<br />
– Simultaneous measurement of O 2<br />
partial pressure, temperature and<br />
ambient air pressure.<br />
– Via HART ® communication the<br />
O 2 <strong>4220X</strong> can be configured and<br />
remote controlled from the control<br />
room. Measured values and messages<br />
are retrievable at any time<br />
allowing integration in fully automatic<br />
production systems.<br />
– In addition to the standard loop<br />
current, the unit provides a passive<br />
current output that can also be<br />
defined as a switching output.<br />
It allows to transmit a further measured<br />
variable to the control<br />
room, monitor limit values, or<br />
create a process control system<br />
using the integrated controller.<br />
A series of safety functions<br />
monitors the correct measurement<br />
– A self-test checks the operation of<br />
the instrument and provides data for<br />
QM documentation in accordance<br />
with DIN ISO 9000.<br />
– Sensocheck ® continuously monitors<br />
the state of the sensor.<br />
– <strong>The</strong> adjustable calibration timer<br />
tracks the calibration schedule.<br />
<strong>The</strong> following functions provide<br />
documentary support<br />
– Two user-defined variables are<br />
recorded and simultaneously<br />
represented by two graphs on the<br />
display. This allows to follow the<br />
process sequence and gives important<br />
information on process<br />
control and monitoring.<br />
– <strong>The</strong> logbook records the last 200<br />
events, with date and time, e.g.<br />
calibrations, warning and failure<br />
messages, auxiliary power failure,<br />
etc. QM documentation in compliance<br />
with DIN ISO 9000 ff. is<br />
therefore possible.<br />
–<strong>The</strong> calibration records show all<br />
relevant data of the last calibration<br />
for documentation to GMP.<br />
–<strong>The</strong> sensor statistics provide an<br />
overview of the condition of the<br />
sensor. Changes in sensor data<br />
between calibrations can be<br />
checked and the values even<br />
compared with those of the first<br />
calibration.<br />
Four different calibration modes<br />
are available<br />
– Automatic calibration in airsaturated<br />
water<br />
– Automatic calibration in air<br />
– Manual entry of saturation value<br />
– Manual entry of sensor data<br />
Additional functions<br />
– Instrument design meeting EMC<br />
demands<br />
– EMC-protected to NAMUR requirements<br />
– High-quality graphics display<br />
– Self-explanatory graphical userinterface<br />
to NAMUR specification<br />
– Permanent data storage during<br />
power failure<br />
– Manual entry of pressure<br />
– Automatic O 2 salinity correction<br />
– Digital input filter for transient<br />
suppression<br />
– Self-contained clock<br />
Warranty<br />
–3 years warranty for <strong>transmitter</strong><br />
– 1 year warranty for accessories<br />
Together with sensors and<br />
housings from METTLER TOLEDO,<br />
complete measurement systems<br />
can be configured to provide a<br />
high level of process safety.