O2 Transmitter 4220X The smart 2-wire transmitter ... - Mettler-Toledo
O2 Transmitter 4220X The smart 2-wire transmitter ... - Mettler-Toledo
O2 Transmitter 4220X The smart 2-wire transmitter ... - Mettler-Toledo
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
Tem<br />
O 2 <strong>Transmitter</strong> <strong>4220X</strong><br />
<strong>The</strong> <strong>smart</strong> 2-<strong>wire</strong> <strong>transmitter</strong><br />
for industrial use<br />
ORP<br />
O 2<br />
pH<br />
Turb<br />
• Dissolved oxygen<br />
• Temperature<br />
• Ambient air pressure<br />
• 4...20 mA current<br />
output<br />
• Switching contact or<br />
second current output<br />
• Sensor diagnostics<br />
• PI-controller<br />
• Simple calibration<br />
• Calibration record<br />
• Logbook<br />
• HART ® -communication<br />
• Measurement recorder<br />
• II 2 (1) G EEx ib [ia]<br />
IIC T6<br />
• First choice 2-<strong>wire</strong> process<br />
<strong>transmitter</strong> for applications in<br />
hazardous areas<br />
• Safe and reliable measurement<br />
due to continuous sensor and<br />
<strong>transmitter</strong> diagnostics<br />
• Easy to use with plain text<br />
dialogue operation<br />
• Ruggedized <strong>transmitter</strong> suitable<br />
for outdoor installation
O 2 <strong>Transmitter</strong> <strong>4220X</strong>,<br />
advanced 2-<strong>wire</strong> process <strong>transmitter</strong>.<br />
O 2 <strong>4220X</strong><br />
O 2 Measurement input<br />
input for amperometric sensor,<br />
e.g. METTLER TOLEDO oxygen<br />
sensors<br />
O 2<br />
P air<br />
mA /<br />
HART ®<br />
Output 1 (Current Loop)<br />
galvanically isolated current loop (power<br />
supply required), 4 to 20 mA, user defined<br />
as current output for the corresponding<br />
process variables, digital communication<br />
via FSK modulation to HART ®<br />
protocol<br />
Pressure sensor<br />
built-in pressure sensor for<br />
measurement of the ambient<br />
air pressure<br />
Temperature input<br />
input for NTC temperature<br />
sensor<br />
Temp<br />
mA /<br />
Output 2 (Passive)<br />
galvanically isolated output user defined<br />
as 0(4) to 20 mA current output for the<br />
corresponding process variables or as<br />
switching or controller output (current<br />
sink, power supply required)<br />
Exceptionally wide range of<br />
functions<br />
<strong>The</strong> METTLER TOLEDO O 2 <strong>Transmitter</strong><br />
features the following:<br />
– Simultaneous measurement of O 2<br />
partial pressure, temperature and<br />
ambient air pressure.<br />
– Via HART ® communication the<br />
O 2 <strong>4220X</strong> can be configured and<br />
remote controlled from the control<br />
room. Measured values and messages<br />
are retrievable at any time<br />
allowing integration in fully automatic<br />
production systems.<br />
– In addition to the standard loop<br />
current, the unit provides a passive<br />
current output that can also be<br />
defined as a switching output.<br />
It allows to transmit a further measured<br />
variable to the control<br />
room, monitor limit values, or<br />
create a process control system<br />
using the integrated controller.<br />
A series of safety functions<br />
monitors the correct measurement<br />
– A self-test checks the operation of<br />
the instrument and provides data for<br />
QM documentation in accordance<br />
with DIN ISO 9000.<br />
– Sensocheck ® continuously monitors<br />
the state of the sensor.<br />
– <strong>The</strong> adjustable calibration timer<br />
tracks the calibration schedule.<br />
<strong>The</strong> following functions provide<br />
documentary support<br />
– Two user-defined variables are<br />
recorded and simultaneously<br />
represented by two graphs on the<br />
display. This allows to follow the<br />
process sequence and gives important<br />
information on process<br />
control and monitoring.<br />
– <strong>The</strong> logbook records the last 200<br />
events, with date and time, e.g.<br />
calibrations, warning and failure<br />
messages, auxiliary power failure,<br />
etc. QM documentation in compliance<br />
with DIN ISO 9000 ff. is<br />
therefore possible.<br />
–<strong>The</strong> calibration records show all<br />
relevant data of the last calibration<br />
for documentation to GMP.<br />
–<strong>The</strong> sensor statistics provide an<br />
overview of the condition of the<br />
sensor. Changes in sensor data<br />
between calibrations can be<br />
checked and the values even<br />
compared with those of the first<br />
calibration.<br />
Four different calibration modes<br />
are available<br />
– Automatic calibration in airsaturated<br />
water<br />
– Automatic calibration in air<br />
– Manual entry of saturation value<br />
– Manual entry of sensor data<br />
Additional functions<br />
– Instrument design meeting EMC<br />
demands<br />
– EMC-protected to NAMUR requirements<br />
– High-quality graphics display<br />
– Self-explanatory graphical userinterface<br />
to NAMUR specification<br />
– Permanent data storage during<br />
power failure<br />
– Manual entry of pressure<br />
– Automatic O 2 salinity correction<br />
– Digital input filter for transient<br />
suppression<br />
– Self-contained clock<br />
Warranty<br />
–3 years warranty for <strong>transmitter</strong><br />
– 1 year warranty for accessories<br />
Together with sensors and<br />
housings from METTLER TOLEDO,<br />
complete measurement systems<br />
can be configured to provide a<br />
high level of process safety.
Clear advantages.<br />
Easy to use<br />
Operation is menu-driven by operator prompting in plain<br />
text dialog. Information text provides you with an accompanying<br />
explanation during operation.<br />
Measurement Recorder<br />
currently measured value<br />
1st channel<br />
measured variable<br />
data and time<br />
functional check active<br />
currently measured value<br />
2nd channel<br />
measured variable<br />
more data<br />
available<br />
<strong>The</strong> integrated measurement recorder simultaneously<br />
records two selectable process variables and represents<br />
them graphically on the display. Measurement range,<br />
recording method and scanning interval are user-defined<br />
within wide limits. <strong>The</strong> last 500 values measured are<br />
available with time and date, either graphically or as<br />
numerical values.<br />
Combined Current, Switching and Controller Output<br />
measured values<br />
alarms<br />
limit values<br />
controller<br />
output 2<br />
current output<br />
switching output<br />
Apart from the current loop, the O 2 <strong>4220X</strong> optionally<br />
provides an additional passive output. This output can be<br />
used as current, switching or controller output.<br />
As current output (0 to 20 mA or 4 to 20 mA), it is userdefinable<br />
for the different process variables. In addition,<br />
alarm and warning messages can be outputed via a<br />
22 mA signal.<br />
As switching output, it releases warnings and alarms or<br />
limit values.<br />
You can also use the output as an analog controller or as<br />
a switching controller output.
Ordering Information<br />
<strong>Transmitter</strong><br />
O 2 <strong>Transmitter</strong> <strong>4220X</strong><br />
Designation<br />
O 2 <strong>4220X</strong><br />
Options<br />
Second current output (passive) 487<br />
Controller function (only with opt. 487) 483<br />
Logbook 354<br />
Key-lockable cover 432<br />
Measurement recorder 448<br />
HART ® communication 467<br />
Mounting accessories<br />
Mounting plate (not required for direct wall mounting) ZU 0136<br />
Bracket kit (ZU 0136 mounting plate required) ZU 0125<br />
Protective hood (ZU 0136 mounting plate required) ZU 0157<br />
Protective case ZU 0158<br />
Bracket kit for protective case ZU 0220<br />
Further accessories<br />
Power supply/isolator for 24 V AC/DC supply<br />
WG 20 A2<br />
Repeater power supply for 90 to 253 V AC (optionally 24 V AC/DC)<br />
WG 21 A7<br />
Repeater power supply for HART ® transmission WG 21 A7/470<br />
Please order the O 2 <strong>Transmitter</strong> <strong>4220X</strong> by specifying the instrument designation and the option numbers,<br />
e.g. O 2 <strong>4220X</strong>/354/448/487<br />
METTLER TOLEDO. <strong>The</strong> competent answer to your QA questions.<br />
For over 40 years, products from METTLER TOLEDO have measured in processes in biotechnology,<br />
chemistry, in food processing and breweries. Dependable monitoring of pH, redox, or O 2 values<br />
through state-of-the-art technology and superior production quality – in the interests of users<br />
and the environment.<br />
ISO<br />
9001<br />
certified<br />
ISO<br />
14001<br />
certified<br />
Management System<br />
certified according to<br />
ISO 9001 / ISO 14001<br />
INTERNET<br />
http://www.mtpro.com<br />
Sales and service:<br />
<strong>Mettler</strong>-<strong>Toledo</strong> GmbH, Process Analytics, Industrie Nord, CH-8902 Urdorf, Tel +41-44-736 22 11, Fax +41-44-736 26 36<br />
Subject to technical changes. 11/04 © <strong>Mettler</strong>-<strong>Toledo</strong> GmbH. Printed in Switzerland. 52 120 793