Tem<br />

O 2 <strong>Transmitter</strong> <strong>4220X</strong><br />

<strong>The</strong> <strong>smart</strong> 2-<strong>wire</strong> <strong>transmitter</strong><br />

for industrial use<br />

ORP<br />

O 2<br />

pH<br />

Turb<br />

• Dissolved oxygen<br />

• Temperature<br />

• Ambient air pressure<br />

• 4...20 mA current<br />

output<br />

• Switching contact or<br />

second current output<br />

• Sensor diagnostics<br />

• PI-controller<br />

• Simple calibration<br />

• Calibration record<br />

• Logbook<br />

• HART ® -communication<br />

• Measurement recorder<br />

• II 2 (1) G EEx ib [ia]<br />

IIC T6<br />

• First choice 2-<strong>wire</strong> process<br />

<strong>transmitter</strong> for applications in<br />

hazardous areas<br />

• Safe and reliable measurement<br />

due to continuous sensor and<br />

<strong>transmitter</strong> diagnostics<br />

• Easy to use with plain text<br />

dialogue operation<br />

• Ruggedized <strong>transmitter</strong> suitable<br />

for outdoor installation


O 2 <strong>Transmitter</strong> <strong>4220X</strong>,<br />

advanced 2-<strong>wire</strong> process <strong>transmitter</strong>.<br />

O 2 <strong>4220X</strong><br />

O 2 Measurement input<br />

input for amperometric sensor,<br />

e.g. METTLER TOLEDO oxygen<br />

sensors<br />

O 2<br />

P air<br />

mA /<br />

HART ®<br />

Output 1 (Current Loop)<br />

galvanically isolated current loop (power<br />

supply required), 4 to 20 mA, user defined<br />

as current output for the corresponding<br />

process variables, digital communication<br />

via FSK modulation to HART ®<br />

protocol<br />

Pressure sensor<br />

built-in pressure sensor for<br />

measurement of the ambient<br />

air pressure<br />

Temperature input<br />

input for NTC temperature<br />

sensor<br />

Temp<br />

mA /<br />

Output 2 (Passive)<br />

galvanically isolated output user defined<br />

as 0(4) to 20 mA current output for the<br />

corresponding process variables or as<br />

switching or controller output (current<br />

sink, power supply required)<br />

Exceptionally wide range of<br />

functions<br />

<strong>The</strong> METTLER TOLEDO O 2 <strong>Transmitter</strong><br />

features the following:<br />

– Simultaneous measurement of O 2<br />

partial pressure, temperature and<br />

ambient air pressure.<br />

– Via HART ® communication the<br />

O 2 <strong>4220X</strong> can be configured and<br />

remote controlled from the control<br />

room. Measured values and messages<br />

are retrievable at any time<br />

allowing integration in fully automatic<br />

production systems.<br />

– In addition to the standard loop<br />

current, the unit provides a passive<br />

current output that can also be<br />

defined as a switching output.<br />

It allows to transmit a further measured<br />

variable to the control<br />

room, monitor limit values, or<br />

create a process control system<br />

using the integrated controller.<br />

A series of safety functions<br />

monitors the correct measurement<br />

– A self-test checks the operation of<br />

the instrument and provides data for<br />

QM documentation in accordance<br />

with DIN ISO 9000.<br />

– Sensocheck ® continuously monitors<br />

the state of the sensor.<br />

– <strong>The</strong> adjustable calibration timer<br />

tracks the calibration schedule.<br />

<strong>The</strong> following functions provide<br />

documentary support<br />

– Two user-defined variables are<br />

recorded and simultaneously<br />

represented by two graphs on the<br />

display. This allows to follow the<br />

process sequence and gives important<br />

information on process<br />

control and monitoring.<br />

– <strong>The</strong> logbook records the last 200<br />

events, with date and time, e.g.<br />

calibrations, warning and failure<br />

messages, auxiliary power failure,<br />

etc. QM documentation in compliance<br />

with DIN ISO 9000 ff. is<br />

therefore possible.<br />

–<strong>The</strong> calibration records show all<br />

relevant data of the last calibration<br />

for documentation to GMP.<br />

–<strong>The</strong> sensor statistics provide an<br />

overview of the condition of the<br />

sensor. Changes in sensor data<br />

between calibrations can be<br />

checked and the values even<br />

compared with those of the first<br />

calibration.<br />

Four different calibration modes<br />

are available<br />

– Automatic calibration in airsaturated<br />

water<br />

– Automatic calibration in air<br />

– Manual entry of saturation value<br />

– Manual entry of sensor data<br />

Additional functions<br />

– Instrument design meeting EMC<br />

demands<br />

– EMC-protected to NAMUR requirements<br />

– High-quality graphics display<br />

– Self-explanatory graphical userinterface<br />

to NAMUR specification<br />

– Permanent data storage during<br />

power failure<br />

– Manual entry of pressure<br />

– Automatic O 2 salinity correction<br />

– Digital input filter for transient<br />

suppression<br />

– Self-contained clock<br />

Warranty<br />

–3 years warranty for <strong>transmitter</strong><br />

– 1 year warranty for accessories<br />

Together with sensors and<br />

housings from METTLER TOLEDO,<br />

complete measurement systems<br />

can be configured to provide a<br />

high level of process safety.


Clear advantages.<br />

Easy to use<br />

Operation is menu-driven by operator prompting in plain<br />

text dialog. Information text provides you with an accompanying<br />

explanation during operation.<br />

Measurement Recorder<br />

currently measured value<br />

1st channel<br />

measured variable<br />

data and time<br />

functional check active<br />

currently measured value<br />

2nd channel<br />

measured variable<br />

more data<br />

available<br />

<strong>The</strong> integrated measurement recorder simultaneously<br />

records two selectable process variables and represents<br />

them graphically on the display. Measurement range,<br />

recording method and scanning interval are user-defined<br />

within wide limits. <strong>The</strong> last 500 values measured are<br />

available with time and date, either graphically or as<br />

numerical values.<br />

Combined Current, Switching and Controller Output<br />

measured values<br />

alarms<br />

limit values<br />

controller<br />

output 2<br />

current output<br />

switching output<br />

Apart from the current loop, the O 2 <strong>4220X</strong> optionally<br />

provides an additional passive output. This output can be<br />

used as current, switching or controller output.<br />

As current output (0 to 20 mA or 4 to 20 mA), it is userdefinable<br />

for the different process variables. In addition,<br />

alarm and warning messages can be outputed via a<br />

22 mA signal.<br />

As switching output, it releases warnings and alarms or<br />

limit values.<br />

You can also use the output as an analog controller or as<br />

a switching controller output.


Ordering Information<br />

<strong>Transmitter</strong><br />

O 2 <strong>Transmitter</strong> <strong>4220X</strong><br />

Designation<br />

O 2 <strong>4220X</strong><br />

Options<br />

Second current output (passive) 487<br />

Controller function (only with opt. 487) 483<br />

Logbook 354<br />

Key-lockable cover 432<br />

Measurement recorder 448<br />

HART ® communication 467<br />

Mounting accessories<br />

Mounting plate (not required for direct wall mounting) ZU 0136<br />

Bracket kit (ZU 0136 mounting plate required) ZU 0125<br />

Protective hood (ZU 0136 mounting plate required) ZU 0157<br />

Protective case ZU 0158<br />

Bracket kit for protective case ZU 0220<br />

Further accessories<br />

Power supply/isolator for 24 V AC/DC supply<br />

WG 20 A2<br />

Repeater power supply for 90 to 253 V AC (optionally 24 V AC/DC)<br />

WG 21 A7<br />

Repeater power supply for HART ® transmission WG 21 A7/470<br />

Please order the O 2 <strong>Transmitter</strong> <strong>4220X</strong> by specifying the instrument designation and the option numbers,<br />

e.g. O 2 <strong>4220X</strong>/354/448/487<br />

METTLER TOLEDO. <strong>The</strong> competent answer to your QA questions.<br />

For over 40 years, products from METTLER TOLEDO have measured in processes in biotechnology,<br />

chemistry, in food processing and breweries. Dependable monitoring of pH, redox, or O 2 values<br />

through state-of-the-art technology and superior production quality – in the interests of users<br />

and the environment.<br />

ISO<br />

9001<br />

certified<br />

ISO<br />

14001<br />

certified<br />

Management System<br />

certified according to<br />

ISO 9001 / ISO 14001<br />

INTERNET<br />

http://www.mtpro.com<br />

Sales and service:<br />

<strong>Mettler</strong>-<strong>Toledo</strong> GmbH, Process Analytics, Industrie Nord, CH-8902 Urdorf, Tel +41-44-736 22 11, Fax +41-44-736 26 36<br />

Subject to technical changes. 11/04 © <strong>Mettler</strong>-<strong>Toledo</strong> GmbH. Printed in Switzerland. 52 120 793

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!