hexane and then, 6 g of activated florisil was poured into the column (Florisil 60-100 mesh from<br />

FISCHER, activated overnight at 130C). A gentle shaking of the column was necessary to settle<br />

the florisil inside the column. Two cm of sodium sulphate was then added <strong>on</strong> top. Hexane was<br />

poured into the column and when the hexane reaches the top of the filling, the stopcock was<br />

closed, a 250 mL round bottom flask was placed under the column and the evaporated sample<br />

extract was transferred to the top of the column using a Pasteur pipette. As the sample passes<br />

through the column, the sample c<strong>on</strong>tainer was rinsed with 3 x 2 mL porti<strong>on</strong>s of hexane (from a<br />

premeasured 50 mL of hexane) and rinses were also transferred to the column. The remaining<br />

hexane was poured into the reservoir of the column and the eluti<strong>on</strong> of fracti<strong>on</strong> 1 was performed<br />

at a steady drip of 3 mL/min. When the hexane reached the top of the filling, the stopcock was<br />

turned off, the flask with fracti<strong>on</strong> 1 was replaced with another 250 mL flask, the column was<br />

eluted with 50 mL of 15% DCM/Hexane (v/v) and fracti<strong>on</strong> 2 was collected. The procedure was<br />

then repeated for fracti<strong>on</strong> 3 using 150 mL of 60% DCM/Hexane (v/v). Each of the 3 fracti<strong>on</strong>s<br />

was separately rotary-evaporated (after adding 5 mL of iso-octane – Optima grade) to 1.0 mL. 1<br />

uL of the extract was analysed by a Agilent 6890 GC-MS (scan mode) fitted with a HP 5MS<br />

column (0.25mm x 0.25 um x 30m), the appropriate reference standards and the Chem-Stati<strong>on</strong><br />

Software.<br />

Analysis of Mirror Samples<br />

Fish and Cow‟s Milk samples were collected and each <strong>on</strong>e was splitted to two –A and B. Fish<br />

sample A and Cow‟s Milk sample A were sent to Dr. Estaben‟s <strong>Laboratory</strong> in Spain. B-samples<br />

were analyzed at he <strong>Pesticide</strong> <strong>Research</strong> <strong>Laboratory</strong>.<br />

Mirror Samples:<br />

Results for UNEP Mirror Samples<br />

PCB<br />

C<strong>on</strong>geners<br />

Fish/(ng/kg) Cow’s<br />

Milk/(ng/kg)<br />

PCB18 38.5 nd<br />

PCB17 17 nd<br />

PCB31 110.8 nd