New Guide <strong>Orthodontic</strong> for <strong>Orthodontic</strong> <strong>Screw</strong> <strong>Screw</strong> Placement<br />
Please read and observe the following guide carefully for successful placement.<br />
<strong>Osstem</strong> <strong>Orthodontic</strong> <strong>Screw</strong> was developed by the R&D center of <strong>Osstem</strong> Implant to help your successful placement with its superior quality.<br />
1. Drilling<br />
New Design Concept<br />
of the orthodontic screw, In this case, initial drilling is recommended.<br />
● Good Initial Stability<br />
● The placement methods using drilling process of the products are shown below;<br />
● Good Self Drilling & Self<br />
● Make Tapping<br />
sure to inject a saline solution during drilling to prevent overheating.<br />
● Good Feeling of <strong>Screw</strong> ● The Implantation<br />
recommended drilling speed is 800rpm.<br />
● Good Strength<br />
● Small Head Size<br />
● Simple Gingiva Shape<br />
● The design of this product dose not require drilling before placement.<br />
However, if the bone is very hard, the excessive placement torque may lead to the failure<br />
∅1.3<br />
∅1.3<br />
Drilling Depth<br />
Only Cortical<br />
Drilling Depth<br />
Only Cortical<br />
1.4<br />
1.6<br />
Using ∅1.3 Drill<br />
Ø0.8(022” Wire)<br />
(Soft Bone) (Normal Bone) (Hard Bone)<br />
Low Profile and<br />
Smooth Curved Head<br />
Simple Gingiva Shape<br />
∅1.5<br />
∅1.5<br />
1.8<br />
Drilling Tapered Depth Body to Increase<br />
Only Cortical<br />
Bone Support<br />
Drilling Depth<br />
Only Cortical<br />
Using ∅1.5 Drill<br />
Strengthened Body Design<br />
(Soft Bone) Powerful Self (Normal Threading Bone)<br />
(0.7 Pitch)<br />
(Hard Bone)<br />
2. <strong>Orthodontic</strong> <strong>Screw</strong> ● For manual placement, mount the Driver Tip on the Universal Handle, and connect the screw.<br />
Placement<br />
● Turn in the clockwise direction Sharp tip with for force, at the initial placement speed of one revolution<br />
per five (5) seconds. When Self Drill thread Function begins to be formed in the bone, apply a rotating force only<br />
for further placement.<br />
● When using an Engine, mount the Machine Driver on the Hand Piece, and connect<br />
Simple Head the screw. Place the orthodontic screw at the speed lower than Through 20rpm. Hole<br />
● Insert the screw until the driver tip touches the gingival tissue, and then remove the driver.<br />
02<br />
12 MS Inplant <strong>System</strong>