Operating instructions <strong>and</strong> service manual<br />

h/p/cosmos <strong>running</strong>-<strong>machine</strong> series:<br />

N 150/50 (mercury)<br />

N 170-190/65 (quasar / pulsar)<br />

N 200-450/75-300 (venus / saturn)<br />

<strong>running</strong> <strong>machine</strong> model name<br />

h/p/cosmos mercury<br />

h/p/cosmos mercury med<br />

h/p/cosmos quasar<br />

h/p/cosmos quasar med<br />

h/p/cosmos pulsar<br />

h/p/cosmos pulsar 3p<br />

h/p/cosmos stratos<br />

h/p/cosmos stratos med<br />

h/p/cosmos stellar<br />

h/p/cosmos stellar med<br />

incl. lt-models<br />

<strong>running</strong> <strong>machine</strong><br />

manual<br />

<strong>running</strong> <strong>machine</strong> model name<br />

h/p/cosmos locomotion 150/50 E med<br />

h/p/cosmos locomotion 150/50 DE med<br />

h/p/cosmos Kistler Gaitway II F<br />

h/p/cosmos Kistler Gaitway II S<br />

h/p/cosmos venus - all sizes<br />

h/p/cosmos saturn - all sizes<br />

manual-version<br />

MCU5-v1.06<br />

Revision 20120210<br />

manual order-no.:<br />

[cos14310m5-v1.06hpc-en]<br />

h/p/cosmos sports & medical gmbh Am Sportplatz 8 DE 83365 Nussdorf-Traunstein / Germany<br />

phone +49 86 69 86 42 0 fax +49 86 69 86 42 49 email@h-p-cosmos.com www.h-p-cosmos.com<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh author: fh created 15.02.2012 printed 15.02.2012 page: 1 of 213<br />

development, production, sales & service<br />

h/p/cosmos sports & medical gmbh<br />

Am Sportplatz 8<br />

DE 83365 Nussdorf-Traunstein<br />

Germany<br />

phone +49 86 69 86 42 0<br />

fax +49 86 69 86 42 49<br />

service@h-p-cosmos.com<br />

www.h-p-cosmos.com


file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh author: fh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 2 of 213


Dear customer,<br />

We would like to express our gratitude for putting your trust in us, in deciding for this top of the range <strong>running</strong> <strong>machine</strong>.<br />

Since 1988 we have been developing <strong>and</strong> manufacturing <strong>running</strong> <strong>machine</strong>s for sports <strong>and</strong> medical application. When it<br />

comes down to technology, design <strong>and</strong> safety, we have set extremely high st<strong>and</strong>ards for ourselves.<br />

Because the <strong>running</strong> <strong>machine</strong> is a motor-driven device, you must pay special attention to the mentioned safety regulations. If<br />

proper notice is taken of the safety regulations, the operation of our <strong>running</strong> <strong>machine</strong>s is almost without any risk. The neglect<br />

of the safety regulations could result in dangerous situations <strong>and</strong> serious accidents. Therefore please read the installation<br />

<strong>and</strong> operation manual <strong>and</strong> the danger precautions before taking the device into operation.<br />

Some simple maintenance (no repair work!), as described, can easily be done by yourself. We recommend to call our<br />

competent service team or entering into a maintenance contract for a routine service in an interval of 6 or 12 months. A form<br />

for registration of your institution <strong>and</strong> device is included in the delivery. In order to be able to supply you with the latest<br />

technical information <strong>and</strong> service, it is important for you to fill out the form. Therefore please fill out the form for registration<br />

immediately <strong>and</strong> send it back via fax.<br />

The instruction manual as a firm part of the delivery has to be accessible for the user at any time. This instruction manual<br />

has been written with great care. Should you, however, still find any details, which do not correspond with your device,<br />

please give us notice, so that we can correct any mistakes as soon as possible. Subject to alterations without prior notice,<br />

errors <strong>and</strong> omissions excepted. E & OE.<br />

We wish you a lot of fun <strong>and</strong> success while exercising <strong>and</strong> working with your h/p/cosmos <strong>running</strong> <strong>machine</strong>.<br />

h<br />

Franz Harrer<br />

President<br />

h/p/cosmos sports & medical gmbh<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh author: fh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 3 of 213


[1.]Introduction<br />

model view of some of the h/p/cosmos <strong>running</strong> <strong>machine</strong> series<br />

h/p/cosmos locomotion 150/50 DE med<br />

h/p/cosmos mercury<br />

h/p/cosmos quasar<br />

h/p/cosmos pulsar<br />

h/p/cosmos venus & h/p/cosmos saturn<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh author: fh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 4 of 213<br />

introduction


h/p/cosmos saturn 300/125r<br />

h/p/cosmos saturn 450/300rs<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh author: fh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 5 of 213<br />

introduction<br />

This manual is only valid for the firmware-/software version noted on the first page of this manual <strong>and</strong> only for the original<br />

configuration of the first delivery of the <strong>machine</strong>.<br />

Firmwareupdates, softwareupdates, changes of the system configuration or retrofittings of additional equipment or<br />

accessories can result in invalidity of this manual. In case of alterations of the device or the additional equipment always the<br />

actual version of the manual or the according additional information has to be considered.


[2.]Content<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh author: fh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 6 of 213<br />

content<br />

[1.] Introduction ........................................................................................................................................................................ 4<br />

[2.] Content ............................................................................................................................................................................... 6<br />

[3.] Introduction ...................................................................................................................................................................... 12<br />

[3.A] Description of the device <strong>and</strong> its intended application ................................................................................................... 12<br />

[3.B] Position on the treadmill ................................................................................................................................................. 12<br />

[3.C] Treadmill-Patient-Interface ............................................................................................................................................. 12<br />

[3.D] Materials used ................................................................................................................................................................ 13<br />

[3.E] Device components, including software <strong>and</strong> accessories .............................................................................................. 13<br />

[3.F] Mechanical characteristics ............................................................................................................................................. 14<br />

[3.G] Performance claims & performance data ....................................................................................................................... 14<br />

[3.H] Describe how the device achieves its intended purpose ................................................................................................ 14<br />

[3.I] Specific safety or performance claims made for the device: .......................................................................................... 15<br />

[3.J] Emergency buttons: ....................................................................................................................................................... 15<br />

[3.K] Pictures of safety devices ............................................................................................................................................... 16<br />

[4.] Safety, warnings, precautions, prohibitions ................................................................................................................. 17<br />

[4.A] Safety precautions, safety regulations, prohibitions <strong>and</strong> warnings ................................................................................. 17<br />

[4.B] Exit the device in emergency case; emergency dismount.............................................................................................. 18<br />

[4.C] Emergency Off, Emergency Stop <strong>and</strong> Safety Stop Devices <strong>and</strong> functions .................................................................... 18<br />

[4.D] General Instructions / Safety .......................................................................................................................................... 21<br />

[4.E] Instructions for safety <strong>and</strong> operation .............................................................................................................................. 22<br />

[4.E1] Electric safety / safety classes / degree of protection ........................................................................................ 22<br />

[4.E2] Patient security according to VDE 0750 (IEC 60601-1) type B R applied part ............................................... 22<br />

[4.E3] Potential equalization ......................................................................................................................................... 22<br />

[4.E4] Connection of units installed outside medically utilized locations ...................................................................... 23<br />

[4.E5] <strong>Medical</strong> electrical systems ................................................................................................................................. 23<br />

[4.E6] Subject's surroundings ....................................................................................................................................... 23<br />

[4.E7] Protection against the danger of electrical shocks ............................................................................................. 24<br />

[4.E8] Environmental requirements .............................................................................................................................. 24<br />

[4.F] <strong>Medical</strong> device software ................................................................................................................................................. 25<br />

[4.G] Connection of external host <strong>and</strong>/or auxiliary devices via interface ................................................................................. 26<br />

[4.H] Programmable, electrical medical systems (PEMS)....................................................................................................... 27<br />

[4.H1] In-house production according to § 12 MPG ..................................................................................................... 27<br />

[4.I] Economic life-time .......................................................................................................................................................... 33<br />

[4.J] Symbols used ................................................................................................................................................................. 33<br />

[4.K] Safety st<strong>and</strong>ards, norms ................................................................................................................................................ 33<br />

[4.K1] VDE / IEC EN norm ........................................................................................................................................... 34<br />

[4.K2] The c mark for sports devices ........................................................................................................................ 34<br />

[4.K3] The n mark for medical devices ........................................................................................................... 34<br />

[4.K4] EMC Electromagnetic Compatibility ................................................................................................................... 34<br />

[4.K5] General advice ................................................................................................................................................... 34<br />

[4.L] h/p/cosmos medical <strong>running</strong> <strong>machine</strong>s - MDD medical device directive. Safety st<strong>and</strong>ards ........................................... 35


file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh author: fh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 7 of 213<br />

content<br />

Forbidden use .......................................................................................................................................................................... 38<br />

[5.] Application, Indication, Contra-Indication ..................................................................................................................... 39<br />

[5.A] Fields of application / intended use ................................................................................................................................ 39<br />

[5.A1] Professional application in sports <strong>and</strong> fitness (with c mark on nameplate) ................................................... 39<br />

[5.A2] Professional application in the medical field (with n mark on nameplate) ............................................ 39<br />

[5.B] Risks of any application .................................................................................................................................................. 39<br />

[5.C] Risk <strong>and</strong> benefit of treadmill application ......................................................................................................................... 40<br />

[5.C1] Summary / potential risk remains ....................................................................................................................... 41<br />

[5.D] Contraindications for any application: ............................................................................................................................ 41<br />

[5.E] Application treadmill ergometry: Cardiology ECG Stress Test on the treadmill ............................................................. 42<br />

[5.E1] Indications .......................................................................................................................................................... 42<br />

[5.E2] Benefit ................................................................................................................................................................ 42<br />

[5.F] Application gait analysis <strong>and</strong> rehabilitation on the treadmill with force measurement plates ......................................... 43<br />

[5.F1] Indications .......................................................................................................................................................... 43<br />

[5.F2] Benefit ................................................................................................................................................................ 43<br />

[5.G] Application gait analysis <strong>and</strong> rehabilitation on the treadmill with neurologic patients ..................................................... 44<br />

[5.G1] Indications .......................................................................................................................................................... 44<br />

[5.G2] Benefit ................................................................................................................................................................ 44<br />

[6.] Installation & commissioning ......................................................................................................................................... 45<br />

[6.A] Transport, unpacking <strong>and</strong> packaging ............................................................................................................................. 45<br />

[6.A1] Transport to upper or lower floors <strong>and</strong> through narrow doors ............................................................................ 45<br />

[6.A2] Transport with trolleys ........................................................................................................................................ 47<br />

[6.A3] Transport of oversize <strong>running</strong> <strong>machine</strong>s ............................................................................................................ 47<br />

[6.B] Mechanical installation ................................................................................................................................................... 47<br />

[6.C] Electrical installation ....................................................................................................................................................... 49<br />

[6.C1] Name plate ......................................................................................................................................................... 50<br />

[6.C2] Electric safety measurements <strong>and</strong> „First Measured Values“ ............................................................................. 51<br />

[6.C3] Potential equalization ......................................................................................................................................... 51<br />

[6.C4] Plug assignment at the UserTerminal ................................................................................................................ 52<br />

[6.D] Starting the <strong>running</strong> <strong>machine</strong>s ........................................................................................................................................ 56<br />

[6.D1] Unlocking of the emergency-stop switch ........................................................................................................... 56<br />

[6.D2] Switching the device on ..................................................................................................................................... 56<br />

[6.D3] Switching the device off ..................................................................................................................................... 57<br />

[6.E] Emergency-Stop / Emergency-Off ................................................................................................................................. 58<br />

[6.F] Installation, Overview, Checklist ..................................................................................................................................... 59<br />

[7.] Operation .......................................................................................................................................................................... 61<br />

[7.A] UserTerminal, keyboard <strong>and</strong> display .............................................................................................................................. 61<br />

[7.A1] The keyboard ..................................................................................................................................................... 62<br />

[7.A2] The display ......................................................................................................................................................... 62<br />

[7.A3] The display (st<strong>and</strong>ard configuration) .................................................................................................................. 63<br />

[7.B] Mode of operation .......................................................................................................................................................... 64<br />

[7.B1] General notes .................................................................................................................................................... 64<br />

[7.B2] Acceleration levels / acceleration intensity ........................................................................................................ 64


file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh author: fh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 8 of 213<br />

content<br />

[7.B3] Manual mode ..................................................................................................................................................... 65<br />

[7.B4] Automatic mode ................................................................................................................................................. 66<br />

[7.B5] Profile Mode / Pre-programmed training profiles ............................................................................................... 66<br />

[7.B6] Possibilities of interfering into the profile mode manually .................................................................................. 68<br />

[7.B7] Profile outline ..................................................................................................................................................... 69<br />

[7.B8] Cardio mode – general notes ............................................................................................................................. 72<br />

[7.B9] Cardio mode - Field of application / control ....................................................................................................... 72<br />

[7.B10] Cardio mode – Operation ................................................................................................................................... 74<br />

[7.B11] Possibilities of interfering with the Cardio Mode manually ................................................................................. 76<br />

[7.B12] Test mode <strong>and</strong> free program profiles – General notes ...................................................................................... 77<br />

[7.B13] Test profiles / Field of application ...................................................................................................................... 77<br />

[7.B14] Test mode - Operation ....................................................................................................................................... 90<br />

[7.B15] The UKK 2 km walk test ..................................................................................................................................... 90<br />

[7.B16] Programing example: UKK walk test ................................................................................................................. 92<br />

[7.B17] Programming-example: free definable profile .................................................................................................... 93<br />

[7.B18] Possibilities of interfering in test mode manually ............................................................................................... 97<br />

[8.] Optional settings .............................................................................................................................................................. 98<br />

[8.A] Settings of control <strong>and</strong> display options ........................................................................................................................... 98<br />

[8.A1] Select user options ............................................................................................................................................ 98<br />

[8.A2] List of user options / optional functions .............................................................................................................. 99<br />

[8.A3] St<strong>and</strong>ard settings for the user options ............................................................................................................. 105<br />

[8.B] Administrator options .................................................................................................................................................... 106<br />

[8.B1] List of the administrator options ....................................................................................................................... 106<br />

[8.B2] St<strong>and</strong>ard settings of administrator options ....................................................................................................... 119<br />

[8.B3] List of <strong>running</strong> <strong>machine</strong>s types ........................................................................................................................ 121<br />

[9.] Maintenance <strong>and</strong> safety inspections ........................................................................................................................... 122<br />

[9.A] Preventive maintenance ............................................................................................................................................... 122<br />

[9.B] Immediate maintenance ............................................................................................................................................... 123<br />

[9.C] Regular inspections/examinations ................................................................................................................................ 123<br />

[9.C1] Visual inspections ............................................................................................................................................ 123<br />

[9.C2] Protective Earth Resistance RPE measurement ............................................................................................... 124<br />

[9.C3] Isolation Resistance RISO measurement .......................................................................................................... 124<br />

[9.C4] Equivalent (alternative) leakage current IEDL measurement ............................................................................. 125<br />

[9.C5] Leakage current measurement ........................................................................................................................ 125<br />

[9.C6] Electric Safety Tester <strong>and</strong> measurements ....................................................................................................... 126<br />

[9.C7] Building installation: Electric Checks, Protective Earth Function, RCD, Leakage Current .............................. 126<br />

[9.D] Lubrication of the <strong>running</strong> belt/<strong>running</strong> surface ............................................................................................................ 127<br />

[9.D1] Devices with UserTerminal <strong>and</strong> without automatic oil-pump ........................................................................... 127<br />

[9.D2] Lubricant test ................................................................................................................................................... 129<br />

[9.D3] Devices without UserTerminal <strong>and</strong> without automatic oil-pump ...................................................................... 130<br />

[9.D4] Devices with UserTerminal & automatic oil-pump (models "r" for cycle & wheelchair) .................................... 131<br />

[9.D5] Set back of oil message MCU 4 without UserTerminal .................................................................................... 133<br />

[9.D6] Set back of oil message MCU with UserTerminal ............................................................................................ 133<br />

[9.E] Control <strong>and</strong> tightening of the <strong>running</strong> belt .................................................................................................................... 134


file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh author: fh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 9 of 213<br />

content<br />

[9.F] Adjustment (centering) of the <strong>running</strong> belt ................................................................................................................... 135<br />

[9.G] Control <strong>and</strong> tightening of the driving belt ...................................................................................................................... 136<br />

[9.G1] Driving belt with poly-V-belt system ................................................................................................................. 136<br />

[9.G2] Driving belt with timing-belt system .................................................................................................................. 138<br />

[9.H] Side step platforms: Check for non-slipping ................................................................................................................. 138<br />

[9.I] Hygiene, cleansing, cleansing of the interior ................................................................................................................ 138<br />

[9.J] Cleansing <strong>and</strong> adjustment of the speed-sensor/Light-barrier ....................................................................................... 139<br />

[9.K] Cleansing <strong>and</strong> adjustment of the light-barrier at belt-re-entry-zones ........................................................................... 140<br />

[9.L] Spare parts & consumables ......................................................................................................................................... 141<br />

[10.] Trouble shooting .......................................................................................................................................................... 143<br />

[10.A] Mechanical / Noise problems ....................................................................................................................................... 143<br />

[10.B] Running belt adjustment problems ............................................................................................................................... 143<br />

[10.C] Fuses 143<br />

[10.D] Interference factor ........................................................................................................................................................ 144<br />

[10.D1] Electrostatic Discharge .................................................................................................................................... 144<br />

[10.D2] Source of Interference ..................................................................................................................................... 144<br />

[10.E] Voltage at the device housing / electric shock ............................................................................................................. 145<br />

[10.E1] Open (interrupted) earth wire / ground ............................................................................................................. 145<br />

[10.F] Malfunctions of the heart rate measurement system POLAR ...................................................................................... 145<br />

[10.F1] Troubleshooting heart rate measurement system POLAR .............................................................................. 146<br />

[10.G] Malfunctions of the RS232 interface ............................................................................................................................ 146<br />

[10.G1] Troubleshooting <strong>and</strong> testing of the RS232 interface ........................................................................................ 146<br />

[10.H] Malfunctions of the control unit ..................................................................................................................................... 146<br />

[10.I] Error messages / Error codes ...................................................................................................................................... 147<br />

[10.I1] Devices with UserTerminal .............................................................................................................................. 147<br />

[10.I2] Devices without UserTerminal ......................................................................................................................... 153<br />

[10.J] Set back of error message ........................................................................................................................................... 153<br />

[10.J1] Devices with UserTerminal .............................................................................................................................. 153<br />

[10.J2] Devices without UserTerminal ......................................................................................................................... 153<br />

[11.] Technical data .............................................................................................................................................................. 154<br />

[11.A] Running <strong>machine</strong>s, 150/50 ........................................................................................................................................... 154<br />

[11.B] Running <strong>machine</strong>s, 170-190/65 ................................................................................................................................... 156<br />

[11.C] Running <strong>machine</strong>s, oversize ........................................................................................................................................ 158<br />

[11.D] EMC electromagnetic compatibility. Guidance <strong>and</strong> manufacturer's declaration ........................................................... 160<br />

[11.E] Compatibility interface .................................................................................................................................................. 162<br />

[11.E1] Interface additional information: ....................................................................................................................... 163<br />

[12.] Accessories <strong>and</strong> options ............................................................................................................................................ 165<br />

[12.A] Arm support for body weight support <strong>and</strong> safety .......................................................................................................... 165<br />

[12.A1] Danger precaution <strong>and</strong> operation of the adjustable arm supports ................................................................... 165<br />

[12.A2] Reproducibility of adjustments ......................................................................................................................... 166<br />

[12.B] Additional keyboard <strong>and</strong> add. stop-button for arm support <strong>and</strong> remote control............................................................ 167<br />

[12.C] H<strong>and</strong>rails adjustable with gear tooth locking elements ................................................................................................ 168<br />

[12.C1] Danger precaution <strong>and</strong> operation .................................................................................................................... 169


file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh author: fh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 10 of 213<br />

content<br />

[12.C2] Reproducibility of adjustments ......................................................................................................................... 169<br />

[12.D] H<strong>and</strong>rails adjustable. Parallel bars with gas struts <strong>and</strong> locking lever ........................................................................... 170<br />

[12.D1] Safety precautions <strong>and</strong> operation .................................................................................................................... 170<br />

[12.D2] Adjustment of the bar h<strong>and</strong>rails ....................................................................................................................... 170<br />

[12.D3] Reproducibility of adjustments ......................................................................................................................... 172<br />

[12.E] Therapist’s seat <strong>and</strong> foot support adjustable ................................................................................................................ 172<br />

[12.E1] Safety precautions <strong>and</strong> operation .................................................................................................................... 172<br />

[12.E2] Operation <strong>and</strong> adjustment ................................................................................................................................ 172<br />

[12.F] Special h<strong>and</strong>rails .......................................................................................................................................................... 174<br />

[12.G] h/p/cosmos airwalk se 135 (available as of 10/2008) ................................................................................................... 175<br />

[12.H] Unweighting system h/p/cosmos airwalk ...................................................................................................................... 176<br />

[12.H1] Vests for subjects for h/p/cosmos airwalk ........................................................................................................ 176<br />

[12.I] h/p/cosmos robowalk ® exp<strong>and</strong>er .................................................................................................................................. 177<br />

[12.J] Wheelchair ramp .......................................................................................................................................................... 177<br />

[12.K] h/p/cosmos wheel chair stabilizer for oversize <strong>running</strong> <strong>machine</strong>s ................................................................................ 177<br />

[12.L] Safety arch with chest belt <strong>and</strong> harness ....................................................................................................................... 178<br />

[12.M] Running belt rotation reversible for downhill walk ............................................................................................ 179<br />

[12.M1] Pull Cord Actuator (lanyard + magnet)............................................................................................................. 180<br />

[12.M2] Change from forward to reverse rotation ......................................................................................................... 181<br />

[12.M3] Change from reverse to forward rotation ......................................................................................................... 181<br />

[12.N] Bottle holder ................................................................................................................................................................. 181<br />

[12.O] Retrofitting UserTerminal ............................................................................................................................................. 181<br />

[12.P] Retrofitting / additional safety stop / emergency stop ................................................................................................... 183<br />

[12.Q] Monitor / laptop arm ..................................................................................................................................................... 184<br />

[12.R] Safety multiple socket .................................................................................................................................................. 184<br />

[12.S] Digital (serial) RS232 interface / USB converter .......................................................................................................... 185<br />

[12.S1] Interface cable RS232 for connection to PC .................................................................................................... 185<br />

[12.S2] Interface cable RS232 for connection to ECG Schiller .................................................................................... 185<br />

[12.S3] Interface cable for connection Ergospirometry JAEGER/VIASYS/CardinalHealth/CareFusion OXYCON ...... 186<br />

[12.T] Printer protocol ............................................................................................................................................................. 187<br />

[12.U] POLAR Heart rate measurement system ..................................................................................................................... 188<br />

[12.U1] Note from POLAR for patients with pacemaker ............................................................................................... 188<br />

[12.U2] Chest belt <strong>and</strong> transmitter ................................................................................................................................ 188<br />

[12.U3] Radius of transmission ..................................................................................................................................... 188<br />

[12.U4] Battery of the transmitter .................................................................................................................................. 189<br />

[12.U5] Correct placement of transmitter belt ............................................................................................................... 189<br />

[12.U6] Transmission problems .................................................................................................................................... 189<br />

[12.V] PC Software h/p/cosmos para control ® 4.1 .................................................................................................................. 189<br />

[12.W] PC Software h/p/cosmos para graphics ® ......................................................................................................... 191<br />

[12.X] PC Software h/p/cosmos para analysis ® ...................................................................................................................... 192<br />

[12.Y] PC Software h/p/cosmos para motion .......................................................................................................................... 193<br />

[13.] Certificates ................................................................................................................................................................... 194<br />

[13.A] Certificate of the TSA (TÜV) according to ISO 9001 .................................................................................................... 194<br />

[13.B] Certificate of the TSA (TÜV) according to EN ISO 13485 ............................................................................................ 195<br />

[13.C] EC Certificate of the TSA (TÜV) ................................................................................................................................... 196<br />

[13.D] Certificate of compliance for sports <strong>and</strong> fitness devices .............................................................................................. 197


file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh author: fh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 11 of 213<br />

content<br />

[13.E] Certificate for medical devices ..................................................................................................................................... 198<br />

[14.] Disposal ........................................................................................................................................................................ 201<br />

[14.A] Disassemble <strong>and</strong> cut up ............................................................................................................................................... 201<br />

[14.B] h/p/cosmos <strong>running</strong> <strong>machine</strong> ....................................................................................................................................... 201<br />

[14.C] h/p/cosmos service oil / grease .................................................................................................................................... 201<br />

[14.D] h/p/cosmos <strong>running</strong> belt ............................................................................................................................................... 201<br />

[15.] Contact .......................................................................................................................................................................... 202<br />

[15.A] Service department ...................................................................................................................................................... 202<br />

[15.B] Sales department ......................................................................................................................................................... 202<br />

[15.C] Headquaters ................................................................................................................................................................. 202<br />

[16.] Appendix – Instruction <strong>and</strong> commissioning ............................................................................................................. 203<br />

[16.A] Instruction into general operation ................................................................................................................................. 203<br />

[16.B] Instruction into functions ............................................................................................................................................... 204<br />

[16.C] Instruction into maintenance works .............................................................................................................................. 206<br />

[16.D] Advice <strong>and</strong> support documents .................................................................................................................................... 207<br />

[16.E] Confirmation of commissioning <strong>and</strong> introduction / instruction ...................................................................................... 208<br />

[16.F] Service report – cos15531 ........................................................................................................................................... 209<br />

[16.G] Control protocol – cos11690en-02 ............................................................................................................................... 210<br />

[16.H] Wiring diagram load circuit quasar med 5.0 / pulsar 5.0 .............................................................................................. 211<br />

[16.I] Wiring diagram control circuit quasar med 5.0 / pulsar 5.0 – page 1 ........................................................................... 212<br />

[16.J] Wiring diagram control circuit quasar med 5.0 / pulsar 5.0 – page 2 ........................................................................... 213


[3.]Introduction<br />

[3.A] Description of the device <strong>and</strong> its intended application<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 12 of 213<br />

introduction<br />

A treadmill is an exercise <strong>machine</strong> for <strong>running</strong> or walking while staying in one place.<br />

The h/p/cosmos treadmill device is a stationary treadmill which is used for training <strong>and</strong> testing The medical versions can<br />

also be used for diagnostics <strong>and</strong> therapy applications.<br />

The <strong>machine</strong> provides a moving platform with a wide conveyor belt <strong>and</strong> an electric motor. The belt moves to the rear<br />

allowing a person to walk or run an equal, <strong>and</strong> necessarily opposite, velocity. The rate at which the belt moves is the<br />

rate of walking or <strong>running</strong>. Thus, the speed of <strong>running</strong> may be controlled <strong>and</strong> measured.<br />

[3.B] Position on the treadmill<br />

The normal position of the walking or <strong>running</strong> subject should be at the front 40% of the <strong>running</strong> deck length. Up to 70%<br />

of the front of the <strong>running</strong> deck length can be utilized. The rear 30% of the <strong>running</strong> belt length shall not be used <strong>and</strong><br />

shall remain as safety space. Safety harnesses such as safety stop cord type (safety lanyard with pull-cord actuator)<br />

<strong>and</strong>/or safety arch with harness, chest belt <strong>and</strong> stop function <strong>and</strong>/or unweighting systems with patient vest <strong>and</strong> harness<br />

shall be adjusted in a way, that the subject can not use the rear 30% of the <strong>running</strong> deck length.<br />

[3.C] Treadmill-Patient-Interface<br />

Example:<br />

Model h/p/cosmos quasar med<br />

A: 170 cm = 100%: total length of <strong>running</strong> surface<br />

B: 68 cm = 40%: recommended position for heel strike<br />

C: 51 cm = 30%: tolerated position for exercise<br />

D: 51 cm = 30%: not tolerated position for exercise<br />

shall remain unused for safety space<br />

Duration of use of the medical versions is approx. 5 to 25 minutes per patient per day, depending on the prescription of<br />

the doctor, the condition of the patient <strong>and</strong> depending on the st<strong>and</strong>ardized or individual load protocol (for example<br />

BRUCE Protocol).<br />

Contact with the body is occurring through walking shoes on the <strong>running</strong> belt, through h<strong>and</strong> palms on h<strong>and</strong>rail grips in<br />

some cases when the patient holds the h<strong>and</strong>rail, through chest belt in case the patient wears the chest belt <strong>and</strong> harness<br />

for heart rate detection <strong>and</strong>/or for fall stop prevention, on forearms in case the patient utilizes the optional arm support<br />

<strong>and</strong> through fingers in case the patient is allowed to press keyboard <strong>and</strong>/or emergency button on the UserTerminal of<br />

the device. There is no physical contact to organs, tissues or body fluids.


[3.D] Materials used<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 13 of 213<br />

introduction<br />

Mainly steel tubes for frame with powder coating, conventional rubber belt for <strong>running</strong> belt, conventional frequency<br />

inverter <strong>and</strong> drive motor for belt speed control, conventional LCD Display <strong>and</strong> keyboard for speed <strong>and</strong> elevation control,<br />

conventional electric <strong>and</strong> electronic components <strong>and</strong> conventional plastic parts.<br />

[3.E] Device components, including software <strong>and</strong> accessories<br />

Heavy duty drive system with high capacity <strong>and</strong> inverter controlled drive motor, optional elevation system for simulating<br />

gradient / elevation (uphill <strong>and</strong> some <strong>machine</strong>s optional downhill), UserTerminal (where applicable) with clear LCD<br />

display <strong>and</strong> keyboard, emergency button, firmware included in the UserTerminal, interface for possible connection to<br />

ECG or other medical devices, coscom interface protocol, optional safety arch with fall stop prevention, chest belt <strong>and</strong><br />

harness, optional arm support with forearm rest for additional stabilization of the patient, optional wheelchair ramp <strong>and</strong><br />

adjustable h<strong>and</strong>rails or other shape h<strong>and</strong>rails if required, optional POLAR heart rate monitor.<br />

There is no external PC software for medical diagnosis <strong>and</strong> medical evaluation purpose supplied by h/p/cosmos.<br />

Software products such as h/p/cosmos para graphics, h/p/cosmos para analysis <strong>and</strong> h/p/cosmos para motion are only<br />

for fitness applications.<br />

The h/p/cosmos para control software is also not for medical evaluation <strong>and</strong> diagnosis but for remote control of the<br />

treadmills load parameter <strong>and</strong> for maintenance issues.<br />

The h/p/cosmos treadmill system consists of several single components, listed in the following. Furthermore h/p/cosmos<br />

offers options, accessories <strong>and</strong> products that can be used together with the treadmill.<br />

[1] UserTerminal (cockpit) with display <strong>and</strong> keyboard<br />

[2] safety stop cord type (lanyard with magnetic switch)<br />

[3] emergency-stop (mushroom style)<br />

[4] heartrate-measurement-system (receiver board)<br />

[5] h<strong>and</strong>rail crossbar (front)<br />

[6] h<strong>and</strong>rails (side)<br />

[7] motor hood/cover<br />

[8] tread plate with slip resistant surface<br />

[9] <strong>running</strong> surface (gliding plate/deck)<br />

[10] <strong>running</strong> belt<br />

[11] rollers<br />

[12] roller cover<br />

[13] base frame<br />

[14] elevation element<br />

[15] option/accessory: safety arch with fall-stop prevention<br />

[16] option/accessory: chest-belt/harness for safety arch<br />

[17] option/accessory: arm support<br />

[18] option/accessory: keyboard for arm support<br />

[19] option/accessory: stop button for arm support<br />

example: h/p/cosmos quasar med treadmill with optional safety arch <strong>and</strong> optional arm support.


[3.F] Mechanical characteristics<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 14 of 213<br />

introduction<br />

Running surface of the model “h/p/cosmos quasar med” for example is 170 x 65 cm / 66.0 x 25.6 inches,<br />

frame size 230 x 105 x 140 cm (90.6 x 41.3 x 55.1 inches).<br />

Running surface of the model “h/p/cosmos pulsar” for example is 190 x 65 cm / 74.8 x 25.6 inches, frame size 250 x 105<br />

x 140 cm (98.4 x 41.3 x 55.1 inches), <strong>machine</strong>’s weight approx. 400 kg (884 lbs), max. subject weight on <strong>running</strong> deck:<br />

200 kg (440lbs). The treadmill has both side h<strong>and</strong>rails <strong>and</strong> is in compliance with the norm EN 957-1 / EN 957-6. The<br />

overall impression is very sturdy, reliable <strong>and</strong> safe.<br />

[3.G] Performance claims & performance data<br />

The treadmill is simulating various adjustable speeds <strong>and</strong> various adjustable gradients to the subject. Performance data<br />

see operation manual technical specifications.<br />

The recommended <strong>and</strong> right endurance <strong>and</strong> dose of the load parameter (speed <strong>and</strong> elevation) <strong>and</strong> specific application<br />

in medical diagnostics <strong>and</strong> therapy is decided <strong>and</strong> supervised by the doctor.<br />

The treadmill is not calculating any medical parameter, is not performing any medical diagnosis <strong>and</strong> is not giving any<br />

recommendations for treatment.<br />

The medical treadmill can support the doctor <strong>and</strong> patient in their decision to perform stress tests <strong>and</strong> in physical<br />

movement of walking <strong>and</strong> <strong>running</strong>.<br />

Load simulation such as speed <strong>and</strong> elevation are non critical, as long as the device is used by adults <strong>and</strong> healthy<br />

subjects in similar way to load <strong>and</strong> stress in daily life (walking <strong>and</strong> <strong>running</strong>) <strong>and</strong> on a deck size of approx. L: 150cm x W:<br />

50cm until L: 190cm x W: 65cm which is an “unwritten st<strong>and</strong>ard” for medical <strong>and</strong> professional treadmills. The load <strong>and</strong><br />

condition to the subject will be very much similar independent from such size <strong>and</strong> configuration details, since the used<br />

load <strong>and</strong> condition to the patient will be independent from the detailed possible treadmill specs, because in diagnostics<br />

<strong>and</strong> therapy load protocols such as speed <strong>and</strong> elevation <strong>and</strong> duration are mainly st<strong>and</strong>ardized. See protocols such as<br />

Bruce, Naughton, Cooper, Ellestad A, Gardner, etc.<br />

[3.H] Describe how the device achieves its intended purpose<br />

The patient is asked to follow the speed of the belt <strong>and</strong> keep his position on the deck. By walking <strong>and</strong>/or <strong>running</strong> the<br />

physical condition of the patient can be analysed through treadmill stress test protocol <strong>and</strong> additional ECG or VO2max.<br />

(metabolic cart) diagnostic device. In physiotherapy <strong>and</strong> neurologic application the patient can be additionally secured<br />

<strong>and</strong> un-weighted with body weight support system <strong>and</strong> the gait can be analysed <strong>and</strong> corrected under safe <strong>and</strong> reproducible<br />

conditions.


[3.I] Specific safety or performance claims made for the device:<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 15 of 213<br />

introduction<br />

Manufacturer’s statement:<br />

For all risk applications or any use with patients, where falling would lead to very dangerous situations (for example after<br />

hip replacement operations, patients with neurologic problems, patients connected to invasive probes, maximum load<br />

tests, etc.) a fall prevention system like a safety arch with chest harness <strong>and</strong> stop-function or an unweighting system (for<br />

example the h/p/cosmos airwalk) is obligatory for such applications. Automatic stop of the <strong>running</strong> belt in the moment a<br />

fall occurs. The chest belt secures the subject <strong>and</strong> prevents falling over to the front. The chest belt straps do not pinch or<br />

restrict the athlete. The chest belt is connected via harness with the emergency stop at the crossbar of the safety arch. If<br />

there is more than approx. 8 kg / 17.6 lbs traction pulling on the harness, the switch will stop the <strong>running</strong> belt<br />

immediately. The length of the harness is individually adjustable.<br />

Unweighting systems as an accessory (for example the h/p/cosmos airwalk) can be used for body weight support of the<br />

patient, in case the patient should not bear his full body weight on his legs, for example after hip operation.<br />

A safety stop cord type (e.g. safety lanyard with magnetic switch) can be attached to the subject. The safety stop cord<br />

type shall limit the range of motion of the subject to the 70% of the front part of the <strong>running</strong> deck length.<br />

The actuator cord shall have a suitable length so that the actuator releases/triggers at a position not more than 70 % of<br />

the length of the <strong>running</strong> surface.<br />

In case the subject cannot catch up with the speed <strong>and</strong> is falling behind to the rear 30% of the <strong>running</strong> deck length, or in<br />

case the subject is falling, the safety stop cord type will trigger a stop function of the treadmill <strong>and</strong> will stop the <strong>running</strong><br />

belt.<br />

Caution! The safety stop cord type can not prevent from falling <strong>and</strong> is therefore not as safe as the safety arch with chest<br />

belt <strong>and</strong> fall prevention or the unweighting system!<br />

[3.J] Emergency buttons:<br />

The big <strong>and</strong> well visible emergency-stop-button is st<strong>and</strong>ard at all models, placed at the h<strong>and</strong>rail or UserTerminal for<br />

easy access. An additional emergency-stop-button can be retro-fitted at all h/p/cosmos <strong>running</strong> <strong>machine</strong>s, for example<br />

to be duplicated on the opposite site h<strong>and</strong>rail.<br />

Extract from the norm EN 957-6:<br />

The push-button type or pull cord safety stop actuator shall be in a reachable position in front of the user <strong>and</strong> in an area<br />

of ± 180 mm parallel to the centreline of the <strong>running</strong> surface.<br />

If the push-button type safety stop actuator is outside this area it shall be duplicated on the opposite side except when a<br />

pull-cord actuator is provided.<br />

h/p/cosmos can supply a second <strong>and</strong> additional push button emergency stop button <strong>and</strong>/or a pull-cord actuator (safety<br />

lanyard with magnetic stop) <strong>and</strong>/or a safety arch with chest belt, harness <strong>and</strong> fall stop function.<br />

At all models of the category medical the main push button emergency-stop is simultaneously an emergency-off, that<br />

means by pushing this button the power supply of the device gets cut off.


[3.K] Pictures of safety devices<br />

fall prevention system: safety arch<br />

with harness <strong>and</strong> chest belt <strong>and</strong> stop<br />

function<br />

emergency stop button on the<br />

UserTerminal of the treadmill.<br />

Triggers bi-polar disconnection from<br />

the mains (power supply).<br />

Additional emergency stop push<br />

button can be mounted on the<br />

opposite side h<strong>and</strong>rail of the<br />

treadmill.<br />

h/p/cosmos arm support with<br />

integrated additional keyboard <strong>and</strong><br />

integrated additional stop button<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 16 of 213<br />

introduction<br />

h/p/cosmos airwalk unweighting<br />

systems for body weight support, can<br />

also be used as fall prevention<br />

system.<br />

A safety stop cord type (safety lanyard with magnetic switch)<br />

The treadmill also has st<strong>and</strong>ard safety features such as emergency button, potential isolation transformer <strong>and</strong> potential<br />

equalisation pin.<br />

potential isolation transformer potential equalisation pin<br />

Additional safety related information, articles <strong>and</strong> announcements see manufacturer’s website:<br />

http://www.h-p-cosmos.com/en/safety/index.htm


[4.]Safety, warnings, precautions, prohibitions<br />

[4.A] Safety precautions, safety regulations, prohibitions <strong>and</strong> warnings<br />

safety, warnings, precautions, prohibitions<br />

N Use medical <strong>running</strong> <strong>machine</strong>s <strong>and</strong> accessories only under permanent supervision of your doctor or/<strong>and</strong> medical<br />

staff. Use sports <strong>running</strong> <strong>machine</strong>s <strong>and</strong> accessories only under permanent supervision of your trainer.<br />

N Using the <strong>running</strong> <strong>machine</strong> <strong>and</strong> the accessories without supervisory staff <strong>and</strong> introduction is forbidden.<br />

N Before using the <strong>running</strong> <strong>machine</strong>, read the operation manual, safety precautions <strong>and</strong> safety regulations carefully.<br />

N Start with slow walking. After some minutes increase the speed slowly according to your fitness level.<br />

N It is strongly recommended to visit a doctor before using an exercise device.<br />

N Always train with sports clothing <strong>and</strong> sports or <strong>running</strong> shoes (no spikes) - never run barefoot.<br />

N At the time of use the emergency-stop has to be in reach for the subject <strong>and</strong> the supervisory staff at any time.<br />

N Use emergency-stop only if in danger or in danger of falling.<br />

N At set elevation (e.g. by pressing the stop-button or the emergency-off-button, power failure, etc.), the bodyweight<br />

of the subject <strong>and</strong> the force of gravity may cause an acceleration of the <strong>running</strong> belt. Do not step on the slant of<br />

roller when getting on or off the <strong>running</strong> surface; the gravity force might cause a movement of the <strong>running</strong> belt.<br />

N Dangerous capture areas/gaps at the elevation system as well as at the rear end <strong>and</strong> the sides of the <strong>running</strong> belt.<br />

(At reverse belt rotation: danger as well at the motor hood <strong>and</strong> the front area).<br />

N Please make sure that, in case of a fall, subjects with long hair, wide clothes, jewellery, etc. do not get caught in<br />

the capture area at the rear end of the <strong>running</strong> belt. Possibly use a hairnet <strong>and</strong> make the subject aware of the<br />

danger or secure the person with a safety harness against fall. Do not deposit any clothes, towels, jewellery, etc.<br />

on the <strong>running</strong> <strong>machine</strong> – because of danger of tripping over <strong>and</strong> danger of capturing.<br />

N Children <strong>and</strong> animals are not allowed to use the <strong>running</strong> <strong>machine</strong> <strong>and</strong> must not get too near (4 m distance) to it.<br />

N Using the <strong>running</strong> <strong>machine</strong> under the influence of alcohol, drugs <strong>and</strong>/or anesthetic is prohibited.<br />

N WARNING: Heart rate monitoring systems can be inaccurate. Incorrect <strong>and</strong>/or forbidden use <strong>and</strong>/or overtraining<br />

can lead to serious injury or even death. If you feel close to fainting, the training has to be stopped immediately.<br />

N Interrupt the training immediately if you start feeling sick, dizzy or pain <strong>and</strong> see your doctor.<br />

N Subjects with a cardiac pacemaker or those who suffer from any kind of physical restriction must see their doctor<br />

<strong>and</strong> get a permission before using the <strong>running</strong> <strong>machine</strong>.<br />

N The subject has to be secured by a safety arch with fall stop <strong>and</strong> chest belt while performing sprints, max.<br />

endurance test <strong>and</strong> while training on wide <strong>running</strong> surfaces <strong>and</strong>/or other higher dangers.<br />

N Please keep a safety area of at least 2 m in length <strong>and</strong> 1 m width directly behind the <strong>running</strong> <strong>machine</strong>. For <strong>running</strong><br />

<strong>machine</strong>s equipped with reverse-belt-rotation keep a safety area of 2 m in length <strong>and</strong> 1 m width in the front <strong>and</strong> the<br />

back of the <strong>running</strong> <strong>machine</strong> <strong>and</strong> always use the pull-rope emergency-stop or better the chest belt with fall-stop.<br />

N Explicit warning of improper <strong>and</strong> forbidden use.<br />

N Ensure that all cables (power connection, interface, potential equalization, etc.) <strong>and</strong> accessories are installed<br />

properly <strong>and</strong> safely <strong>and</strong> that nobody can stumble or fall over the cables <strong>and</strong>/or accessories.<br />

N Only connect accessories <strong>and</strong> software <strong>and</strong> host equipment if it is approved as compatible by all manufacturers.<br />

N Pay attention to further safety regulations <strong>and</strong> operation instructions in the appendix of this instruction manual. Pay<br />

special attention to all instructions for operation, maintenance <strong>and</strong> service also in manuals of all accessories.<br />

N In case of any detected <strong>and</strong>/or assumed malfunctions <strong>and</strong>/or defects or unreadable safety warning labels the<br />

device has to be taken out of operation immediately, the device has to be marked <strong>and</strong> secured against operation<br />

<strong>and</strong> the supplier <strong>and</strong> authorized service personnel has to be informed in writing immediately.<br />

N Disregard of warnings, disregard of intended <strong>and</strong> forbidden use, safety precautions <strong>and</strong> also unauthorized or lack<br />

of maintenance <strong>and</strong>/or regular safety checks may lead to injuries or death <strong>and</strong>/or can damage the device <strong>and</strong> will<br />

result in loss of any liability <strong>and</strong> warranty.<br />

N Do not drink <strong>and</strong> eat any lubrication material <strong>and</strong> no other material or parts of this device, <strong>and</strong> keep all lubrication<br />

material <strong>and</strong> all other parts <strong>and</strong> packing away from children <strong>and</strong> animals.<br />

N These safety precautions <strong>and</strong> regulations are to be displayed within sight of the <strong>running</strong> <strong>machine</strong>. The warnings,<br />

safety regulations <strong>and</strong> dangers have to be pointed out to every user <strong>and</strong> operator. The manufacturer does not<br />

undertake any liability for any injury to persons or damage to property.<br />

N Valuable articles (like watches, emblazonments, etc.) must be abated before using the device <strong>and</strong> must also not<br />

be in near environment of the device.<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 17 of 213


safety, warnings, precautions, prohibitions<br />

N When using the treadmill for load tests, a medical doctor <strong>and</strong> a defibrillator has to be within reach at any time.<br />

Therefore pay special attention to your country-specific regulations.<br />

[4.B] Exit the device in emergency case; emergency dismount<br />

Emergency dismount (e.g. in danger of stumbling <strong>and</strong>/or falling, etc.) can be done e.g. by grabbing the front h<strong>and</strong>lebar<br />

or both side h<strong>and</strong>rails <strong>and</strong> jumping with both feet on the foot rails (step-platforms). In any emergency case the<br />

emergency button has to be pressed instantly!<br />

[4.C] Emergency Off, Emergency Stop <strong>and</strong> Safety Stop Devices <strong>and</strong> functions<br />

Based on the norm EN 957-6, all power driven treadmills shall be equipped with a safety stop switch, which shall include<br />

a push-button operated switch <strong>and</strong>/or a pull-cord operated switch.<br />

The push-button type or pull cord safety stop actuator shall be in a reachable position in front of the user <strong>and</strong> in an area<br />

of ± 180 mm parallel to the centreline of the <strong>running</strong> surface.<br />

If the push-button type safety stop actuator is outside this area it shall be duplicated on the opposite side except when a<br />

pull-cord actuator is provided. Operating the safety stop shall result in the <strong>machine</strong> cutting off the main power of the<br />

motor <strong>and</strong> the inclination system without relying on any software functions.<br />

The contacts of manually operated safety stop devices shall ensure positive opening. The achievement of contacts<br />

separation shall be the direct result of a specified movement of the switch actuator through non resilient members.<br />

Emergency stop relays or manually reset fail-safe relays may be used.<br />

It shall not be possible to restore the circuit until the actuator of the safety stop device has been manually reset. After<br />

restoring the circuit the equipment shall restart with st<strong>and</strong>still <strong>and</strong> the incline shall stay the same or reset.<br />

Where several safety stop devices are provided, the circuit shall not be restored until all actuators previously operated<br />

have been reset. Pull-cord safety stop actuators shall release with a maximum force of 50 % of the force required to<br />

release the attachment device when attached to the user according to the manufacturers' recommendation.<br />

The actuator cord shall have a suitable length so that the actuator releases at a position not more than 70 % of the<br />

length of the <strong>running</strong> surface.<br />

All h/p/cosmos <strong>running</strong> <strong>machine</strong>s have at least ONE electric emergency stop device, which result in the <strong>machine</strong><br />

cutting off the main power of the motor <strong>and</strong> the inclination system without relying on any software functions.<br />

However, in a number of cases <strong>and</strong> situations it is reasonable <strong>and</strong> important to have additional safety stop devices<br />

<strong>and</strong> quick stop devices, which can even use the electric power to stop the drive motor faster <strong>and</strong> which can maintain<br />

interface communication with connected host devices, such as PC, ECG, etc. when used in medical field.<br />

Therefore h/p/cosmos devices can be equipped with a number of additional emergency stop devices <strong>and</strong> functions<br />

which are not cutting off the main power.<br />

This means that h/p/cosmos is in compliance with the norm EN 957-6 <strong>and</strong> offers additional safety features.<br />

picture type of stop function interface<br />

communication<br />

Safety Stop =<br />

Emergency Stop<br />

on UserTerminal<br />

(st<strong>and</strong>ard on all<br />

<strong>machine</strong>s with<br />

UserTerminal)<br />

N palm switch is hit<br />

N <strong>running</strong> belt & inclination<br />

system stop fast<br />

N <strong>machine</strong> is cut off from main<br />

power of drive motor <strong>and</strong><br />

inclination system<br />

N display, keyboard <strong>and</strong><br />

interface inactive<br />

communication<br />

interrupted<br />

during cut off<br />

from main<br />

power<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 18 of 213<br />

how to restart<br />

N wait 1 minute<br />

N pull palm switch<br />

N push “power on”<br />

button / main<br />

power<br />

N press START key,<br />

N or new comm<strong>and</strong><br />

through interface


Safety Stop =<br />

Emergency Stop<br />

External <strong>and</strong>/or<br />

additional<br />

emergency stop<br />

mushroom head<br />

type.<br />

Quick Stop<br />

External <strong>and</strong>/or<br />

additional<br />

emergency stop<br />

mushroom head<br />

type, connected<br />

to the “quick stop<br />

port” of the<br />

UserTerminal of<br />

oversize model<br />

families<br />

200-450/75-300<br />

(venus/saturn).<br />

Quick Stop<br />

on Safety Arch<br />

with chest belt,<br />

harness/rope <strong>and</strong><br />

fall stop<br />

prevention.<br />

= pull-cord<br />

operated switch<br />

N mushroom switch is hit<br />

N <strong>running</strong> belt & inclination<br />

system stop fast<br />

N <strong>machine</strong> is cut off from main<br />

power of drive motor <strong>and</strong><br />

inclination system<br />

N display, keyboard <strong>and</strong><br />

interface inactive<br />

N mushroom switch is hit<br />

N <strong>running</strong> belt & inclination<br />

system stop extremely fast<br />

N drive motor is used to stop<br />

the <strong>running</strong> belt very fast<br />

N <strong>machine</strong> is NOT cut off from<br />

main power of drive motor<br />

<strong>and</strong> inclination system<br />

N display active <strong>and</strong> shows<br />

“PULL STOP”<br />

N interface active, but no start,<br />

speed or elevation signal<br />

executed<br />

N keyboard inactive<br />

N subject wears chest belt.<br />

safety rope is pulled with<br />

more than approx. 8 kg<br />

N <strong>running</strong> belt & inclination<br />

system stop very fast<br />

N drive motor is used to stop<br />

the <strong>running</strong> belt very fast on<br />

the oversize <strong>machine</strong>s<br />

venus/saturn<br />

N <strong>machine</strong> is NOT cut off from<br />

main power of drive motor<br />

<strong>and</strong> inclination system<br />

N display active <strong>and</strong> shows<br />

“PULL STOP”<br />

N interface active, but no start,<br />

speed or elevation signal<br />

executed<br />

N keyboard inactive<br />

safety, warnings, precautions, prohibitions<br />

communication<br />

interrupted<br />

during cut off<br />

from main<br />

power<br />

communication<br />

NOT interrupted<br />

during Quick-<br />

Stop stage<br />

communication<br />

NOT interrupted<br />

during Quick-<br />

Stop <strong>and</strong>/or<br />

“pull stop” stage<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 19 of 213<br />

N wait 1 minute<br />

N pull mushroom<br />

switch<br />

N push “power on”<br />

button / main<br />

power<br />

N press START key,<br />

N or new comm<strong>and</strong><br />

through interface<br />

N pull mushroom<br />

switch<br />

N press START key,<br />

N or new comm<strong>and</strong><br />

through interface<br />

N release weight<br />

from safety rope<br />

N press START key,<br />

N or new comm<strong>and</strong><br />

through interface


Pull-cord<br />

actuator (lanyard<br />

with<br />

magnetic switch<br />

on the<br />

UserTerminal <strong>and</strong><br />

waist belt on the<br />

subject)<br />

Additional stop<br />

button for arm<br />

support,<br />

incorporated on<br />

top of the<br />

h<strong>and</strong>grip<br />

quick stop<br />

through light<br />

barriers<br />

at belt re-entry<br />

zones, only at<br />

oversize model<br />

families<br />

200-450/75-300<br />

(venus/saturn).<br />

N magnet is pulled off through<br />

the cord<br />

N <strong>running</strong> belt & inclination<br />

system stop fast<br />

N <strong>machine</strong> is NOT cut off from<br />

main power of drive motor<br />

<strong>and</strong> inclination system<br />

N display active <strong>and</strong> shows<br />

“PULL STOP”<br />

N interface active, but no start,<br />

speed or elevation signal<br />

executed<br />

N keyboard inactive<br />

N Stop button is pushed<br />

N <strong>running</strong> belt & inclination<br />

system stop fast<br />

N <strong>machine</strong> is NOT cut off from<br />

main power of drive motor<br />

<strong>and</strong> inclination system<br />

N display active <strong>and</strong> shows<br />

“PULL STOP”<br />

N interface active, but no start,<br />

speed or elevation signal<br />

executed<br />

N keyboard inactive<br />

N light beam is interrupted<br />

N <strong>running</strong> belt & inclination<br />

system stop extremely fast<br />

N drive motor is used to stop<br />

the <strong>running</strong> belt very fast<br />

N <strong>machine</strong> is NOT cut off from<br />

main power of drive motor<br />

<strong>and</strong> inclination system<br />

N display active <strong>and</strong> shows<br />

“PULL STOP”<br />

N interface active, but no start,<br />

speed or elevation signal<br />

executed<br />

N keyboard inactive<br />

safety, warnings, precautions, prohibitions<br />

communication<br />

NOT interrupted<br />

during<br />

“pull stop” stage<br />

communication<br />

NOT interrupted<br />

during<br />

“pull stop” stage<br />

communication<br />

NOT interrupted<br />

during Quick-<br />

Stop stage<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 20 of 213<br />

N reconnect magnet<br />

to its position on<br />

UserTerminal<br />

N press START key,<br />

N or new comm<strong>and</strong><br />

through interface<br />

N release push force<br />

from the button<br />

N press START key,<br />

N or new comm<strong>and</strong><br />

through interface<br />

N remove obstacle<br />

from light beam<br />

N press START key,<br />

N or new comm<strong>and</strong><br />

through interface<br />

In all above mentioned cases all displays will reset to the value “0”, independent from the type of stop function<br />

<strong>and</strong> independent from the way, if the <strong>machine</strong> was cut off from main power or if the main power remained.


[4.D] General Instructions / Safety<br />

This instruction manual is part of the device <strong>and</strong> must always be accessible for every user.<br />

safety, warnings, precautions, prohibitions<br />

Exact observance of the instruction manual is a requirement for the appropriate operation of an h/p/cosmos device.<br />

Holding the h<strong>and</strong>rails during exercise may effect the exercise results (e.g. heart rate, ECG, EMG, oxygen uptake, etc.).<br />

Under normal conditions <strong>and</strong> for healthy subjects we recommend to hold the h<strong>and</strong>rails only for emergency or when<br />

needed for safety or unweighting.<br />

Important notes, warnings <strong>and</strong> safety precautions are marked with the opposite sign. It also<br />

reminds you of concerns which have to be considered for measurements <strong>and</strong> connection with<br />

other devices.<br />

The safety of your subjects <strong>and</strong> the compliance to the mentioned accuracy of measurement can only be achieved, if the<br />

consumption goods, sensors <strong>and</strong> detectors, described in the instruction manual, are being used <strong>and</strong> the recommended<br />

preventive maintenance <strong>and</strong> safety checks are performed by personnel authorised by h/p/cosmos.<br />

Absorption-, cleansing- <strong>and</strong> disinfecting chemicals are only to be kept in the appropriate reservoirs. At the connection of<br />

gas pressure container it has to be examined about its MOT-admissibility (TÜV) first. Pull the power plug before<br />

cleansing or disinfecting electric devices.<br />

Safety, reliability <strong>and</strong> function can only be achieved if ...<br />

N installation, commissioning, instruction, extension, alteration <strong>and</strong> repair is performed by authorized people;<br />

N the room for installation corresponds to the DIN <strong>and</strong> VDE installation directives;<br />

N the device is plugged in at a socket with a protection system;<br />

N the room for installation corresponds to the requirements of the surroundings for the device;<br />

N the device after the instruction is being used in accordance with the instruction manual.<br />

The entry of fluid into the device has to be removed immediately by the authorized customer service <strong>and</strong> a safety<br />

examination has to be performed. Damaged socket connections, wires <strong>and</strong> pressure control switches <strong>and</strong>/or any other<br />

damaged parts <strong>and</strong>/or labels have to be replaced instantly by trained personnel or by authorised persons.<br />

In case of any detected <strong>and</strong>/or assumed malfunctions <strong>and</strong>/or defects or unreadable safety warning<br />

labels the device has to be taken out of operation immediately, the device has to be marked <strong>and</strong><br />

secured against operation (for example pull power plug <strong>and</strong> affix warning/defect label on the power<br />

plug) <strong>and</strong> the supplier <strong>and</strong> authorised service personnel has to be informed in writing immediately.<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 21 of 213


[4.E] Instructions for safety <strong>and</strong> operation<br />

safety, warnings, precautions, prohibitions<br />

The regulations mentioned in this chapter refer to the Federal Republic of Germany. For all other states take into<br />

account the national legislation. If the <strong>running</strong> <strong>machine</strong> is a sports device (only c mark without 4-digit number), it must<br />

not be connected with a medical product <strong>and</strong> must not be used for medical purposes. Due to higher safety requirements<br />

medical instruments <strong>and</strong> devices are not allowed to be connected via extension cable <strong>and</strong>/or multiple plug socket with<br />

the power supply, <strong>and</strong> a medical product must not be connected with a sports device either.<br />

Disregard of warnings, disregard of intended <strong>and</strong> forbidden use, safety precautions <strong>and</strong> also<br />

unauthorized or lack of maintenance <strong>and</strong>/or regular safety checks may lead to injuries or even<br />

death <strong>and</strong>/or can damage the device <strong>and</strong> will result in loss of any liability <strong>and</strong> warranty<br />

[4.E1] Electric safety / safety classes / degree of protection<br />

In order to protect the subject (athlete <strong>and</strong>/or patient) <strong>and</strong> the personnel, the association of German electrotechnicians<br />

Inc, (VDE) has published special directives for medical used rooms <strong>and</strong> electro-medical devices.<br />

Devices with a power supply therefore have to, in order to prevent the passing on of the mains voltage over to<br />

touchable metal pieces, be equipped with not only a reliable isolation of the parts being under voltage but also<br />

additional safety precautions. The VDE-association divides it into so called safety classes.<br />

Of the, for electro-medical devices, licensed safety classes are mainly used the safety class I M (i.e. safety<br />

precautions with protective wiring), <strong>and</strong> the safety class II, (i.e. safety measures without protective wiring but with<br />

double isolation): Devices of the safety class I M are devices, where the metallic casing-parts are connected with<br />

the protective wire of the line net via the safety contact. In the case of an isolation error the inserted fuse element<br />

switches off. h/p/cosmos <strong>running</strong> <strong>machine</strong>s have:<br />

degree of protection against electrical shock: class I M<br />

degrees of protection provided by enclosures: IP 20<br />

[4.E2] Patient security according to VDE 0750 (IEC 60601-1) type B R applied part<br />

If your h/p/cosmos device has a c mark with a 4-digit number (e.g. n) at the identification plate, it is a<br />

medical device. At your h/p/cosmos medical device the patient is protected/isolated from the mains in accordance<br />

with the safety st<strong>and</strong>ard referring to the requirements for leakage current values according to IEC 60601-1 applied<br />

part type B R. The connection of further mains-operated units to your h/p/cosmos-unit may cause that all the<br />

leakage currents add <strong>and</strong> the safety of the patient is reduced. Due to this, the connection of further units may only<br />

be carried out on consultation with h/p/cosmos <strong>and</strong> if the connected equipment is explicitly declared as compatible<br />

by all relevant manufacturers.<br />

[4.E3] Potential equalization<br />

The connection of e.g. EDP units to the V24 / RS 232 / USB / LAN / FireWire interface of the computer or of other<br />

independent measuring stations can only be carried out in medically utilized locations according to VDE 0107 / EN<br />

ISO 60601-1-1. Parallel connection of the potential equalization ensures, that in case of insulation failure none of<br />

the connected units exceeds a voltage of 10mV <strong>and</strong> thus endangers the patient.<br />

For this purpose all of the medical h/p/cosmos <strong>running</strong> <strong>machine</strong>s are equipped with a connection fixture with which<br />

the equipotent conductor, being part of the st<strong>and</strong>ard equipment, is connected to the equipotent bus bar of the<br />

room.<br />

St<strong>and</strong>ard potential equalization cable (length 5m) with 4mm 2 cross-section [cos10223]<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 22 of 213


[4.E4] Connection of units installed outside medically utilized locations<br />

safety, warnings, precautions, prohibitions<br />

If units installed outside medically utilized locations (e. g. external printer, host computer etc.) are connected to an<br />

h/p/cosmos unit installed in a medically utilized location the VDE 0750 regulations must be observed. Connection<br />

only via …<br />

1. Optical fibre or optocoupler (4 kV checked <strong>and</strong> approved) or …<br />

2. Protective isolation via insulation transformer according to IEC 60601-1, appendix k. These units are also to be<br />

connected to the equipotent conductor.<br />

Due to the increased safety requirements medical devices may not be connected via extension leads or multi-way<br />

connectors.<br />

[4.E5] <strong>Medical</strong> electrical systems<br />

The medical electrical system (later in this text referred as “system”) is a combination of several devices, of which<br />

at least one is a medical electrical equipment. A combination occurs when devices are connected via so called<br />

“functional connections” or via a mobile multiple socket. Not only electrical connections are defined as “functional<br />

connections” but also those for the transmission of signals, electrical energy <strong>and</strong>/or substances. Thus already the<br />

operation of two or more devices on the same multiple outlet is defined as a system. Even a mechanical<br />

connection or a wireless connection produces a medical electrical system.<br />

Basically it is imperative that a system is as safe as a single medical electrical equipment. The requirements in the<br />

medical devices act <strong>and</strong> the applicable regulations (for example MPBetreibV) must be met. The system as a whole<br />

must ensure that within the patients environment or at contact with patients the same level of safety prevails, as<br />

determined by IEC 60601-1 (VDE 0750 Part 1), IEC 60601-1-1, IEC 60601-1-2, IEC 60601-1-4 <strong>and</strong> IEC 60601-1-<br />

6.<br />

[4.E6] Subject's surroundings<br />

The surroundings of the subject must have a gap of 1.5 m as it has been proofed by experience. This has been<br />

laid down here as the surrounding.<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 23 of 213


[4.E7] Protection against the danger of electrical shocks<br />

Casing <strong>and</strong> cover: Parts of non-medical<br />

electric devices within the surrounding of<br />

the subject, which, after having taken off<br />

the coverings etc. without the use of tools<br />

for a routine maintenance etc., are in<br />

touchable reach, have to operate with a<br />

voltage that does not exceed 25 volt<br />

alternate voltage <strong>and</strong> 60 volt direct<br />

voltage, which is produced by a separate<br />

source as described in IEC 60601-1.<br />

According to this example the leakage<br />

current would flow from the electric device<br />

to the earthened subject via the trainer/doctor.<br />

safety, warnings, precautions, prohibitions<br />

Electrical equipment [3] <strong>and</strong> a subject [1] must never be touched by the doctor or trainer [2]<br />

at the same time.<br />

[4.E8] Environmental requirements<br />

Running <strong>machine</strong>s are not to be used in medically utilized rooms with a danger of explosions or in easily<br />

inflammable atmospheres. The devices must not be installed near to e.g. an x-ray device, motors or transformer<br />

with high connection power, as the electric <strong>and</strong> magnetic interference can falsify measurements or even make<br />

them impossible. High voltage lines must be avoided nearby the device. h/p/cosmos electrical devices with mains<br />

connections must neither be used in wet <strong>and</strong> humid areas (e.g. swimming pool, sauna, etc.) nor in environmental<br />

chambers.<br />

If not stated otherwise in the delivery information h/p/cosmos devices are designed for operation in normal climatic<br />

surroundings (IEC 60601-1):<br />

N Temperature: + 10 ° ... + 40 ° C<br />

N Relative humidity: 30 ... 70 % (non condensing!)<br />

N Air pressure: 700 ... 1060 mbar<br />

N Maximum operating altitude: approx. 10,000 feet (3000m), without pressurization<br />

The <strong>running</strong> <strong>machine</strong> has to be protected from high humidity. Venting slots are not to be covered; otherwise it<br />

would hinder the air circulation. Store the devices at a temperature of – 20 ° ... + 50 ° C.<br />

All devices can be stored without power-connection <strong>and</strong> without operation for a period of 6 – 9 months. After this<br />

period batteries in the device may be discharged <strong>and</strong> maintenance <strong>and</strong>/or a new setup/programming of the device<br />

could be required by authorized h/p/cosmos technician.<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 24 of 213


[4.F] <strong>Medical</strong> device software<br />

safety, warnings, precautions, prohibitions<br />

Software safety classification according to EN 62304 medical device software; software lifecycle processes:<br />

The manufacturer has to assign a software safety class to any software, according to consequences caused by the software<br />

in form of a danger on the patient, the applier or third parties.<br />

The software safety classes, based on the severity, are assigned as follows:<br />

Class A: No injury or damage to health possible<br />

Class B: No SERIOS INJURY possible<br />

Class C: DEATH or SERIOS INJURY possible<br />

Without risk reducing measurements <strong>and</strong> risk reducing design a software for treadmills <strong>and</strong> interface protocols would have to<br />

be classified as Class C most dangerous with death result, because for example undesired <strong>and</strong> uncontrolled acceleration of<br />

the <strong>running</strong> belt can always cause falling off with the possible result of a broken neck. Due to a possible technical<br />

malfunction of a measurement (e.g. heart rate measurement), which has to be theoretically be taken into account, the patient<br />

may be overloaded with the possible result of death.<br />

Also on the part of the dem<strong>and</strong> on SOUP (software of unknown provenance) it has to be stated, that medical treadmill<br />

ergometers <strong>and</strong> their software use SOUP components. So the frequency inverter, motor regulation, firmware <strong>and</strong> parts of the<br />

pc software are components of medical treadmill ergometers <strong>and</strong> the treadmill manufacturer has no access to design,<br />

validation <strong>and</strong> maintenance of these components of suppliers. For this reason always the worst case of a malfunction has to<br />

be taken into account.<br />

h/p/cosmos classifies the internal firmware <strong>and</strong> software components as Class B due to a number of risk reducing design<br />

features <strong>and</strong> measurments.<br />

Healthy people in sports are usually able to activate the emergency stop button of the treadmill <strong>and</strong> this way to switch off the<br />

treadmill quickly in case of emergency caused by a malfunction. Thus, the emergency switch off has to work completely<br />

without software <strong>and</strong> has to brake the power supply of the treadmill. For patients <strong>and</strong> medical users fast reactions of the<br />

patient <strong>and</strong> therewith a fast switch off of the treadmill are not guaranteed because for a patient also very light acceleration<br />

may cause in falling off. To control the risk for patients <strong>and</strong> persons with any disabilities a fall prevention system (e.g. safety<br />

arch with chest belt, harness <strong>and</strong> fall stop rope) which prevents falling off (catches the patient) has to be used for<br />

applications with a higher risk of falling (e.g. patients after hip replacement, neurologic patients, cardiac patients, maximum<br />

load tests, etc.) or in cases falling could result in serious injuries, for example if the patient is connected to invasive probes.<br />

For the acceleration of the drive motor <strong>and</strong> <strong>running</strong> belt safety delay design measures have been incorporated, which<br />

prevent from extreme fast acceleration in case of any malfunction.<br />

Warning labels based on EN 957-6 about possible malfunction <strong>and</strong> inaccuracy of heart rate sensors are placed on the<br />

treadmill <strong>and</strong> in the operation manual.<br />

Furthermore the patient on the medical treadmill has to be observed permanently <strong>and</strong> the medical staff has to be within 1.5m<br />

around the patient.<br />

The risk analysis <strong>and</strong> risk control are included in the risk management according to EN 14971.<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 25 of 213


[4.G] Connection of external host <strong>and</strong>/or auxiliary devices via interface<br />

safety, warnings, precautions, prohibitions<br />

Many optional <strong>and</strong> host devices (e.g. ECG, spirometry devices, EMG, etc.) which can be connected to the treadmill via<br />

interface, are not made by <strong>and</strong> not supplied by h/p/cosmos.<br />

Only devices <strong>and</strong> software, which are explicitly declared as compatible by the treadmill manufacturer <strong>and</strong> the<br />

manufacturer of the host device (e.g. ECG, spirometry devices, EMG, etc.) are allowed to be connected to the treadmill.<br />

The summary of clinical data <strong>and</strong> the risk management of h/p/cosmos is covering only the medical treadmill ergometer<br />

<strong>and</strong> the “intersection” of the risks, which arise out of the stress test in ergometry in connection with the host devices (e.g.<br />

ECG, spirometry devices, EMG, etc.).<br />

The summary of clinical data <strong>and</strong> the risk management of h/p/cosmos is NOT covering any measured parameter <strong>and</strong>/or<br />

vital functions which are measured <strong>and</strong>/or detected by host devices (e.g. ECG, spirometry devices, EMG, etc.) which<br />

are connected to the treadmill.<br />

For this purpose there are clinical evaluations <strong>and</strong> risk management processes to be established <strong>and</strong> performed by the<br />

respective manufacturers of these host devices on individual base.<br />

Accessory equipment connected to the analog <strong>and</strong> digital interfaces must be certified according to the respective IEC<br />

st<strong>and</strong>ards, e.g. IEC 950 for data processing equipment <strong>and</strong> IEC 60601-1 for medical equipment.<br />

Furthermore all configurations shall comply with the valid version of the system st<strong>and</strong>ard IEC 60601-1-1, IEC 60601-1-2,<br />

IEC 60601-1-4 <strong>and</strong> IEC 60601-1-6. Always use IEC 60601-1 approved potential isolation components when linking<br />

medical devices via interface.<br />

It is not allowed to connect medical devices (for instance ECG-systems) to sports <strong>running</strong> <strong>machine</strong>s.<br />

Everybody who connects additional equipment to the signal input part or signal output part of a medical device<br />

configures a medical system, <strong>and</strong> is therefore responsible that the system complies with the requirements of the valid<br />

version of the system st<strong>and</strong>ard IEC 60601-1-1, IEC 60601-1-2, IEC 60601-1-4 <strong>and</strong> IEC 60601-1-6.<br />

h/p/cosmos only approves the interface protocol „coscom v3“ for medical applications <strong>and</strong> herewith confirms the -since<br />

March 21, 2010 (MDD / Directive 2007/47/EC) m<strong>and</strong>atory- conformity of coscom v3 according to the norms EN 62304<br />

(Life Cycle Requirements for <strong>Medical</strong> Device Software; <strong>Medical</strong> Device Software Life Cycle Processes) <strong>and</strong> EN 14971<br />

(risk management for medical devices) as well as the compliance to a very high safety st<strong>and</strong>ard.<br />

All other interface protocols (older versions coscom v2, coscom v1, coscomekg.dll, Trackmaster, Pacer, etc.) are not<br />

validated by h/p/cosmos according to the norms EN 62304 <strong>and</strong> EN 14971 <strong>and</strong> must not be used for medical applications<br />

(e.g. ergometry, stress-tests, cardiorespiratory diagnostic, performance diagnostic, etc.) since March 21, 2010 in<br />

connection with h/p/cosmos treadmills or h/p/cosmos OEM treadmills, except where explicitly approved <strong>and</strong> confirmed in<br />

writing by h/p/cosmos <strong>and</strong> the respective manufacturer!!<br />

We strongly recommend to update to the latest h/p/cosmos coscom v3 interface protocol. The latest coscom.dll <strong>and</strong><br />

implementation notes can be found on www.coscom.org. In case you have any questions please contact<br />

service@h-p-cosmos.com<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 26 of 213


[4.H] Programmable, electrical medical systems (PEMS)<br />

[4.H1] In-house production according to § 12 MPG<br />

safety, warnings, precautions, prohibitions<br />

In case a user creates a connection of st<strong>and</strong>ard components, to support or even perform diagnose <strong>and</strong> appraisal, in<br />

terms of in-house production according to § 12 MPG (German <strong>Medical</strong> Devices Act) he creates a system <strong>and</strong> therefore<br />

has to perform <strong>and</strong> verify a simplified conformity validation process. An in-house production is existent as the<br />

manufacturer creates a product or system, which is used for diagnostics <strong>and</strong>/or therapy, himself.<br />

The in-house production of medical products is regulated in the 3rd MPG revision law. According to this the § 12 is only<br />

valid for devices <strong>and</strong> systems, which are created <strong>and</strong> only applied at the operating place. That means, as well the<br />

manufacturer as the user are not allowed to put these devices <strong>and</strong> systems into circulation or give them to third parties<br />

on the single European market.<br />

A simplified conformity evaluation process means the user does not need to create a declaration of conformity or to<br />

inform the notified body, but has to demonstrate the compliance to the essential requirements incl. clinical evaluation,<br />

risk management <strong>and</strong> the dem<strong>and</strong> of documentation.<br />

Dem<strong>and</strong>s out of EN 60601-1 3rd edition, chapter 14.13 <strong>and</strong> annex H as well as chapter 16:<br />

Connection between programmable, electrical, medical systems (PEMS) <strong>and</strong> other devices via network or data sharing<br />

respectively.<br />

The norm claims the manufacturer to inform the responsible organization, which means the applier (hospital, medical<br />

practice, etc.), about following risks in the technical description:<br />

• The connection of a PEMS to a network or data sharing respectively, that includes other devices may cause in<br />

unknown risks for patients, users or third parties.<br />

• The applier (hospital, medical practice, etc.) should determine, analyze, evaluate <strong>and</strong> control those risks (the<br />

adequate tool is the risk management according to DIN EN 14971).<br />

• The applier must be informed that the following changes in the network or data sharing respectively, may lead<br />

to new risks <strong>and</strong> therefore dem<strong>and</strong> new analyses.<br />

Changes in network or data sharing respectively may include following procedures:<br />

• Adding new devices to the network or data sharing respectively<br />

• Removing devices from the network or data sharing respectively<br />

• Devices, connected to the network or data sharing respectively have to meet the state of the art<br />

• Improvement of devices connected to the network or data sharing respectively<br />

Important information:<br />

• In case the manufacturer declares multiple sockets as valid, these have not to be located on the floor, to<br />

prevent soaking of liquids <strong>and</strong> mechanical damage.<br />

h/p/cosmos herewith informs about all above <strong>and</strong> below mentioned risks, warnings <strong>and</strong><br />

obligations.<br />

Responsibility for system integration:<br />

The operator of ME equipment <strong>and</strong> ME systems (ME=medical-electrical), has to appoint a so-called system integrator,<br />

responsible for taking care of the tasks that result from the norm. This justifies this claim so that even ME devices are<br />

used, which were not primarily designed to work with other devices or ME-devices or ME-systems. Therefore, the norm<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 27 of 213


safety, warnings, precautions, prohibitions<br />

dem<strong>and</strong>s a systems integrator, in practice, also known as a system administrator who is responsible for ensuring that all<br />

individual ME device to work satisfactorily in an integrated system.<br />

The system integrator has to perform the following tasks <strong>and</strong> must have knowledge of:<br />

• How to use the integrated system?<br />

• What are the requirements for the performance of the integrated system?<br />

• How the proposed system configuration looks like?<br />

• What restrictions exist regarding the exp<strong>and</strong>ability of the system?<br />

• Documents relating to the specifications of all ME devices <strong>and</strong> other devices to integrate.<br />

• What is the performance of each ME unit <strong>and</strong> other devices?<br />

• How does the information flow within <strong>and</strong> around the system?<br />

Manufacturers can usually not take over the task of a system integrator in the hospital because they do not have the<br />

complete information <strong>and</strong> data that were previously listed. The norm limits the liability of a manufacturer to supply the<br />

required information on their device, they can not be split between different manufacturers. An operator such as a<br />

hospital or medical practice can of course hire a manufacturer or maybe even a service provider to integrate their<br />

system. In this case, the entire system gets an ME system, derive from the responsibility of the manufacturer or service<br />

provider, is to create a properly integrated system.<br />

The system integrator should have the expertise <strong>and</strong> experience to identify dangers <strong>and</strong> assess the most likely result<br />

from the integration of a system <strong>and</strong> to ensure that remaining potential (rest-) risks are identified during the operation of<br />

the system.<br />

For the system integrator this means<br />

• He has to plan the integration of all ME-devices or ME-systems <strong>and</strong> all non-medical devices according to the<br />

manufacturer’s instructions.<br />

• He has to perform the risk management for the integrated system.<br />

• He has to forward any manufacturers information that is needed for the safe use of the integrated system to<br />

the applier (hospital or medical practice).<br />

The norm requires that manufacturer information include information <strong>and</strong> warnings about hazards that can arise from<br />

configuration changes (upgrades, updates). This requires that manufacturers report all information regarding software<br />

updates <strong>and</strong> upgrades, as well as hardware modifications to the designated system integrator.<br />

The operator should ideally create a complete documentation of both network-based systems, including telemedicine<br />

systems, <strong>and</strong> changes (technological changes, software changes, upgrades, etc.) or to update the document<br />

accordingly.<br />

For this job the system intergrator has to know <strong>and</strong> know how to apply the risk management norm EN 14971.<br />

The following possible sources of risk in a network or data network may occur:<br />

• Data loss<br />

• inappropriate data exchange<br />

• Data Corruption<br />

• Inappropriate time allocation data<br />

• unexpected data received<br />

• non-authorized access to data<br />

• destructive data.<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 28 of 213


safety, warnings, precautions, prohibitions<br />

Examples of ME equipment <strong>and</strong> ME systems on the network <strong>and</strong> data network:<br />

• modality connectivity via network to PACS <strong>and</strong> WEB-based electronic image distribution<br />

• server-based database for long-term ECG devices or stress-test systems<br />

• telemedicine applications<br />

• Wi-Fi connectivity of patient monitors<br />

• et.al.<br />

Appendix A of DIN EN 14971 includes questions to identify characteristics of a medical device that may have security<br />

implications. The application of this Annex on the causes <strong>and</strong> risks of networks <strong>and</strong> data network should at least,<br />

according to DIN EN 60601-1, include the following considerations or risk <strong>and</strong> potential causes respectively:<br />

• Telemedicine <strong>and</strong> tele-service with external access to internal network or data network of an operator (hospital<br />

or other)<br />

• Remote service of modalities manufacturers<br />

• compatibility of operating systems<br />

• changes <strong>and</strong> upgrades of the software (operating system, applications, etc.)<br />

• The impact <strong>and</strong> consequences of patch management<br />

• Interface management (e.g. intolerance of 10 MB network cards with a 100 MB Ethernet network or other)<br />

• Connections (modification of hardware, network connector)<br />

• protocols such as DICOM, HL7 in the network or data network<br />

• Packet address structure <strong>and</strong> b<strong>and</strong>width<br />

• Heterogeneous network topology<br />

• Normal network traffic <strong>and</strong> b<strong>and</strong>width requirements<br />

• peak network load<br />

• security <strong>and</strong> long-term readability of data carriers<br />

• Security against destructive software, unauthorized software updates or upgrades<br />

• Maximum acceptable response time<br />

• Acceptable error rate of the network or data network<br />

• availability during planned <strong>and</strong> unplanned maintenance<br />

• inconsistency of interfaces <strong>and</strong> formats, which can lead to loss of accuracy during the data transfer<br />

• etc.<br />

Annex D of the DIN EN 14971 describes examples of risks <strong>and</strong> other factors in connection between the ME device <strong>and</strong><br />

network or data network:<br />

• What predictable abuse can occur?<br />

• Is the connection to the network or data network carried out in accordance to the intended use or purpose<br />

according to § 3 para 10 of the <strong>Medical</strong> Devices Act, or is it in contradiction?<br />

• Could a false data flow occur from or to any connected PEMS?<br />

• What shall the medical data transferred through the network or data network achieve or what shall happen to<br />

them? What happens if the network or data network collapses during data transmission?<br />

• Can there be any deviations from specified operating characteristics of each participating PEMS?<br />

• What qualities <strong>and</strong> operating characteristics does a PEMS have <strong>and</strong> how <strong>and</strong> in what form may they be<br />

influenced by the network or data network?<br />

• Does a complete description of the parameters of the network or data network exist, such as network topology,<br />

configuration, parameters, b<strong>and</strong>width (100 MB Ethernet, 1GB Ethernet, etc.), etc.?<br />

• Can an overloading of the network or data connections within the network node occur?<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 29 of 213


safety, warnings, precautions, prohibitions<br />

• Is the network designed to ensure load? Does the number of planned network nodes suffice or is there<br />

redundancy? Is there a structured network cabling?<br />

• Can use errors occur <strong>and</strong> if so, which? What training, <strong>and</strong> capabilities, does the operator need to have to<br />

manage <strong>and</strong> administer the network properly?<br />

• How do the configuration <strong>and</strong> patch management of the network <strong>and</strong> connected PEMS occur? Does regular<br />

servicing change the characteristics <strong>and</strong> properties of the network or data network such as remote service?<br />

What influence do remote service, patch management, etc. have to the connected PEMS such arrangements,<br />

etc.? Does the system administrator takes care of the authorization or approval of patches on the operating<br />

system level, virus protection, etc., <strong>and</strong> does he examine the effects on PEMS <strong>and</strong> network?<br />

• Do the medical data arrive at the right place at the right destination in their complete form? May there be<br />

unforeseen changes, the user acknowledges in time?<br />

• Is there a readily available documentation for all hard- <strong>and</strong> software components with all updates?<br />

IEC 60601-1 classifies networks <strong>and</strong> data networks as shown in Figure 11 according to criteria A, B <strong>and</strong> C to<br />

identify a statement about the consequences <strong>and</strong> required response times. Reaction time means for connection<br />

of a PEMS to a network or data network, the time lag between the occurrence of a fault in the network or data<br />

network <strong>and</strong> the occurrence of patient harm. Table 1 contains potential risks based on severity <strong>and</strong> response time<br />

in case of data loss or alteration of data in a network or data network.<br />

Severity Reaction time Network class Example<br />

Death or serious injury Second(s) A Infusion (closed loop), lack control of operation<br />

robot<br />

Minute(s) A Missing alarm transfer of an intensive care network<br />

Hour(s) A/B Wrong therapy data on dialysis <strong>machine</strong> or<br />

ventilator<br />

Medium injury Second(s) A Wrong alarm transfer, lack control of operation<br />

robot<br />

Minute(s) A/B Wrong alarm transfer, lack control of operation<br />

robot<br />

Hour(s) B Image mutilation, loss of therapy protocol<br />

Light injury Second(s) B<br />

Minute(s) B Loss of a x-ray image<br />

Hour(s) B/C Malfunction of a tele-medic connection<br />

minor Second(s) C<br />

Minute(s) C<br />

Hour(s) C Breakdown of a tele-medic connection<br />

Table 1: Possible risks by severity <strong>and</strong> reaction time in case of data loss or disturbances of the network or data<br />

network respectively.<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 30 of 213


Picture 11: classification of a network or data network in categories A, B <strong>and</strong> C<br />

Class „C“-network or data network:<br />

safety, warnings, precautions, prohibitions<br />

All time-critic processes <strong>and</strong> applications for those a malfunction or interruption would lead to a critical situation<br />

for the patient as in an intensive care network are categorized as class C. Such a network should not be<br />

connected to the general hospital network, because such a connection might create uncontrollable risks. The<br />

accessibility of such an isolated (floating) network has to be very high, interruptions must occur very seldom. For<br />

such a network the manufacturer / deliverer is the only responsible, he has to define the specifications of the<br />

network nodes to be used.<br />

Note: In practice such an isolation of networks in hospitals can be realized only by very high technical effort or not<br />

at all. Depending on the manufacturer the patient monitors in intensive care can also show x-ray images,<br />

laboratory data <strong>and</strong> other patient related information; this requires a connection of the intensive care network <strong>and</strong><br />

the general hospital network to transfer data like x-ray images form the PACS. In this case the responsibility of<br />

such networks is clearly transferred to the operator.<br />

Class „B“ network or data network:<br />

This category of networks or data networks includes all not time-critical applications or processes, which h<strong>and</strong>le<br />

therapeutic or diagnostic patient data. Such a network may be connected to another network like a hospital<br />

network via a defined <strong>and</strong> controllable or save interface. The st<strong>and</strong>ards for such a network are high, so<br />

interruptions shall last only for short times.<br />

The specifications of such a network are either defined by the manufacturer or the system administrator defined<br />

by the operator. As class B networks are usually radiologic networks, it is a problem to define the responsibility for<br />

such a system when several manufacturers allocate different modules.<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 31 of 213


Class „A“ network or data network:<br />

safety, warnings, precautions, prohibitions<br />

The general hospital network is an example for class A, such a network or data network carries general<br />

applications including administrative or demographic patient data. The norm accepts a longer time falling out<br />

because in hospital usually are alternatives.<br />

For such a system the system administrator defined by the operator is responsible.<br />

In practice such clearly defined, closed categorization / classification will not exist. The following example will<br />

show the reason why in practice usually exists a mixture of the three classifications:<br />

A radiology network of class B sends images <strong>and</strong> diagnostic findings from the radiology network via the general<br />

hospital network of class A to a PACS Server (Picture <strong>and</strong> Communication System). The radiologic images are<br />

accessible to all users can be opened in the operating room, intensive care <strong>and</strong> so on. Some manufacturers of<br />

intensive care observation systems that, according to the norm, belong to class C offer the possibility to show xray<br />

images on the bed side monitor on intensive care. In the hospitals daily routine a mixture of all three network<br />

classes occurs. It is important that the 3 rd edition of DIN EN 60601-1 h<strong>and</strong>les the increasing networking of<br />

medical technical devices with networks <strong>and</strong> server based databases, <strong>and</strong> shows a possible solution to cover the<br />

complexity <strong>and</strong> potential risks of tele-medical techniques in form of the dem<strong>and</strong>ed system administrator or the risk<br />

management according to DIN EN 14971.<br />

Telemedicine is not yet described explicitly in DIN EN 60601-1 3 rd edition, but the chapter about network <strong>and</strong> data<br />

network includes an approach how to h<strong>and</strong>le this development.<br />

The st<strong>and</strong>ard requires that when using a network or data network with the aim of exchanging data between<br />

PEMS <strong>and</strong> PEMS or other IT devices (such as servers <strong>and</strong> databases) as well the manufacturer as the operator<br />

have the required knowledge to build such networks, to serve <strong>and</strong> control, with all related processes <strong>and</strong><br />

functions.<br />

The St<strong>and</strong>ard requires, for example, manufacturers or suppliers of PEMS <strong>and</strong> / or networks <strong>and</strong> data networks to<br />

choose the configuration of their products so that they meet internationally known network st<strong>and</strong>ards such as<br />

Ethernet, Fast Ethernet, GigaBitEthernet, FDDI, <strong>and</strong> others, <strong>and</strong> the available b<strong>and</strong>width according to the<br />

intended use or the purpose in accordance with § 3 para 10 MPG use is appropriate <strong>and</strong> that they achieve the<br />

optimum performance for their application.<br />

The third Edition of DIN EN 60601-1 requires that a hospital as an operator, represented by a system integrator,<br />

<strong>and</strong> the PEMS manufacturer discuss any major technical parameters to ensure a reliable installation of a PEMS<br />

operating in a network or data network. This procedure is required to avoid unacceptable risks as possible.<br />

Table H.4 of the st<strong>and</strong>ard gives an example of a collection of parameters that are needed to describe a network<br />

or a data network to document <strong>and</strong> define. This list represents however only a draft, which is regarded as the<br />

start of a comprehensive documentation.<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 32 of 213


Summary:<br />

safety, warnings, precautions, prohibitions<br />

The operation of programmable electrical medical systems (PEMS) requires the observance of <strong>and</strong> compliance<br />

with the <strong>Medical</strong> Devices Act <strong>and</strong> safety st<strong>and</strong>ards, as described in the technical rules (st<strong>and</strong>ards) to ensure<br />

quality <strong>and</strong> safe patient care. In particular, the third Edition of DIN EN 60601-1 is aimed much more to the<br />

operator of telemedicine connections than the second Edition. To operate a safe <strong>and</strong> high quality tele-medicine<br />

as tele-neurology network, the safety st<strong>and</strong>ards should be met.<br />

[4.I] Economic life-time<br />

The economic life time of the product is determined at common usage <strong>and</strong> application to 20 years, provided that after 10<br />

years all electrical parts <strong>and</strong> component are renewed (at damage or unusual heavy duty possibly earlier) <strong>and</strong> the<br />

recommended maintenance intervals are kept. Every maintenance <strong>and</strong> repair work needs to be carried out by<br />

authorized h/p/cosmos technicians. For expendable parts a shorter life-time is due.<br />

[4.J] Symbols used<br />

All symbols used comply with the according norms IEC417, IEC878, EN957-1 <strong>and</strong> Council Directive 2002/96/EC.<br />

Read manufacturer’s guide, advises,<br />

instructions <strong>and</strong> manual<br />

Warning / safety precautions<br />

Pay attention to accompanying instructions<br />

M<br />

R Devices of the type B P<br />

J<br />

T<br />

[4.K] Safety st<strong>and</strong>ards, norms<br />

Dangerous electric voltage<br />

Alternating current (AC)<br />

Warning of hot surfaces<br />

Risk/danger of entanglement<br />

N Earth<br />

Protection ground<br />

Potential compensation<br />

Symbol for collection, treatment, recycling<br />

<strong>and</strong> disposal of waste electrical <strong>and</strong><br />

electronic equipment (WEEE) as set out in<br />

Directive 2002/96/EC of 27 January 2003<br />

of the European Parliament <strong>and</strong> of the<br />

Council on waste electrical <strong>and</strong> electronic<br />

equipment are necessary to reduce the<br />

waste management problems linked to the<br />

heavy metals concerned <strong>and</strong> the flame<br />

retardants concerned.<br />

All h/p/cosmos <strong>running</strong>-<strong>machine</strong>s are produced according to strict safety- <strong>and</strong> quality controls. The <strong>running</strong> <strong>machine</strong><br />

must not be used, if the certificates/permissions <strong>and</strong> the (according to the list of technical specifications) defined safety<br />

st<strong>and</strong>ards do not correspond with the local <strong>and</strong> country-specific requirements. The country-specific requirements must<br />

be compared before initial operation. Operate <strong>running</strong> <strong>machine</strong> only, if the requirements correspond. Full list of<br />

applicable norms see also in the CE-certificate of compliance.<br />

The st<strong>and</strong>ards with date in this manual with reference to design <strong>and</strong> construction characteristics, refer to the up to date<br />

valid versions at the time of manufacture.<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 33 of 213


safety, warnings, precautions, prohibitions<br />

The st<strong>and</strong>ards with date in this manual applying to routine measurements <strong>and</strong> continual safety relevant measures (e.g.<br />

electrical measurements to determine leakage current, insulation resistances etc.) can <strong>and</strong> will change with time.<br />

Therefore the up to date valid versions of the respective st<strong>and</strong>ards at the time of measurement <strong>and</strong> their limit values,<br />

which can deviate then from the procedures <strong>and</strong> limit values in this manual, apply.<br />

[4.K1] VDE / IEC EN norm<br />

The <strong>running</strong>-<strong>machine</strong>s for the professional application in sports <strong>and</strong> fitness have been built corresponding to the<br />

DIN EN 60335-1 (VDE 0700) guidelines. The <strong>running</strong>-<strong>machine</strong>s for the professional application in the medical area<br />

have been built corresponding to the DIN EN 60601-1 (VDE 0750) guidelines.<br />

[4.K2] The c mark for sports devices<br />

The c mark on the nameplate of the sports <strong>and</strong> fitness <strong>running</strong> <strong>machine</strong>s confirms the compliance to the council<br />

directive 89/336/EEC appendix I (EMC electromagnetic compatibility). The audit has been conducted after the<br />

criterion of jamming <strong>and</strong> interference immunity.<br />

[4.K3] The n mark for medical devices<br />

The n mark on the nameplate of the <strong>running</strong> <strong>machine</strong> for medical application confirms the compliance to<br />

the council directive 93/42/EEC <strong>and</strong> 2007/47/EC (MDD <strong>Medical</strong>-Device-Directive).<br />

[4.K4] EMC Electromagnetic Compatibility<br />

The <strong>running</strong>-<strong>machine</strong>s for the professional application in sports <strong>and</strong> fitness (with c mark) have been built<br />

corresponding to the European Union Council Directive 89/336/EEC (EMC) <strong>and</strong> the low voltage directive<br />

73/23/EEC. Norms: EMC: EN 55011; EN 61000-6-1; EN 61000-3-2; EN 61000-3-3, Safety: EN 60950<br />

The <strong>running</strong>-<strong>machine</strong>s for medical applications (with n mark) have been built corresponding to the<br />

European Union Council Directive 93/42/EEC <strong>and</strong> 2007/47/EC<br />

Norms: EMC: EN 60601-1-2; EN 61000-3-2; EN 61000-3-3, Safety: EN 60601-1<br />

Certification bodies: Mitsubishi Electric Europe / EMC-Laboratroy, Mündelheimer Weg 35, DE 40472 Düsseldorf<br />

EMV Testhaus GmbH, Gustav - Hertz - Strasse 35, DE 94315 Straubing / Germany,<br />

J. Schmitz GmbH, DE 83022 Rosenheim / Germany<br />

Strong electromagnetic fields, transmitters <strong>and</strong> interferences, which are above the normal tolerance, can interfere<br />

with measuring fuctions <strong>and</strong> displays of the <strong>running</strong> <strong>machine</strong>s <strong>and</strong> can lead to other malfunctions.<br />

Follow details to EMC values, guidance <strong>and</strong> manufacturers declaration in this manual in chapter „technical data“<br />

[4.K5] General advice<br />

The device is assigned to overvoltage-category II <strong>and</strong> pollution level I.<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 34 of 213


safety, warnings, precautions, prohibitions<br />

[4.L] h/p/cosmos medical <strong>running</strong> <strong>machine</strong>s - MDD medical device directive. Safety st<strong>and</strong>ards<br />

Over 24 years of experience in the production of treadmill-ergometers <strong>and</strong> complex systems with accessories <strong>and</strong><br />

special options have made h/p/cosmos the absolute treadmill specialist. Safety has a very high priority for us –<br />

above all because we know from experience what kind of accidents can occur <strong>and</strong> which consequences might<br />

result from these accidents.<br />

On the one h<strong>and</strong> a treadmill is a very effective instrument in rehabilitation, therapy <strong>and</strong> diagnostics. However, like<br />

in many areas of daily life (i.e. all kinds of traffic, may it be by car, train or bus or the h<strong>and</strong>ling of electrical <strong>machine</strong><br />

tools etc.) the use of a treadmill involves dangers which can be minimized by abiding strictly to safety procedures,<br />

but can never be fully ruled out. As a manufacturer it is h/p/cosmos’ duty, <strong>and</strong> as the distributor or enduser/operator<br />

of the <strong>running</strong> <strong>machine</strong> it is your duty, to minimize those dangers as far as possible.<br />

In medical applications many additional risks (patients with pacemakers, ECG lead <strong>and</strong> electrode connections to<br />

the skin, health <strong>and</strong> physical limitations, high workloads during stress tests, medication, etc.) have to be<br />

considered. Therefore medical treadmills with interface <strong>and</strong> interactive heart rate based load controls are classified<br />

as risk class IIb medical devices. They have specific additional design characteristics compared to fitness<br />

treadmills, <strong>and</strong> can only be sold <strong>and</strong> serviced by authorized <strong>and</strong> certified distributors. h/p/cosmos has the<br />

responsibility to ensure that the systems are correctly installed (with usage instruction) <strong>and</strong> maintained. Therefore<br />

we have implemented a medical device observation <strong>and</strong> reporting system based on the statutory obligations<br />

contained in, Directive 93/42/EEC <strong>and</strong> 2007/47/EC MDD (medical device directive), ISO 9001 <strong>and</strong><br />

EN 13485 st<strong>and</strong>ard <strong>and</strong> our internal safety policy <strong>and</strong> risk management system. This is of utmost importance <strong>and</strong><br />

all our distributors must sign a contract that they will comply with these important international rules for medical<br />

devices, particularly for class IIb devices <strong>and</strong> their respective statutory requirements.<br />

h/p/cosmos fulfils its duty in different ways:<br />

N All medical treadmill-ergometer device-families have to undergo an internal product-type examination as well<br />

as external tests performed by the respective examination institutes such as for example the TUV Product<br />

Service in Munich / Germany or the EMC laboratory. Our medical treadmill-ergometers meet the norms <strong>and</strong><br />

guidelines stipulated for medical applications such as the, MDD, Directive 93/42/EEC <strong>and</strong> 2007/47/EC for<br />

medical devices, IEC EN 60601-1 (VDE 0750), VDE 0751, IEC EN 60601-1-2 (EMC), IEC EN 60601-1-4, EN<br />

957-1, EN 957-6.<br />

N All h/p/cosmos medical treadmill-ergometers must undergo a several hour non-stop test after production with<br />

strict safety tests, high-voltage-tests <strong>and</strong> electrical as well as mechanical safety checks. All tests are recorded,<br />

checked by a second independent employee <strong>and</strong> filed. h/p/cosmos treadmills have to pass the checks 100%<br />

without exception.<br />

N All medical h/p/cosmos treadmill-ergometers have an Emergency Off Switch (current off with mains<br />

disconnection, not only the “st<strong>and</strong>ard software controlled emergency stop button”), power supply with<br />

increased voltage resistance, electrical installation <strong>and</strong> wiring according to IEC 60601-1, a potential<br />

equalization pin <strong>and</strong> a potential equalization transformer.<br />

N h/p/cosmos has a quality management <strong>and</strong> risk management system according to ISO 9001 & ISO 13485 as<br />

well as medical device observation <strong>and</strong> incident reporting system.<br />

N In the course of maintenance work h/p/cosmos technicians or authorised service personnel check the<br />

electrical safety of the treadmill together with all safety relevant details <strong>and</strong> measure the accuracy of speed<br />

<strong>and</strong> elevation of the treadmills at the customer’s premises.<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 35 of 213


safety, warnings, precautions, prohibitions<br />

N h/p/cosmos trains all relevant service technicians <strong>and</strong> sales personnel as medical device consultants with<br />

respective certification.<br />

N The user manuals of h/p/cosmos treadmills contain all important information <strong>and</strong> safety advice as well as<br />

maintenance instructions <strong>and</strong> user regulations. The treadmills have a label with the date of the next safety<br />

check / maintenance.<br />

N h/p/cosmos offers numerous options <strong>and</strong> accessories such a safety arch with harness for fall prevention,<br />

chest belt with automatic stop, arm supports or the unweighting system h/p/cosmos airwalk, which increase<br />

safety for the patient during the use of a treadmill-ergometer. http://www.h-p-cosmos.com/en/safety/index.htm<br />

All <strong>running</strong> <strong>machine</strong>s <strong>and</strong> other <strong>machine</strong>s with moving belts <strong>and</strong> parts, hold a great potential of danger at the belt<br />

re-entry zones on the sides <strong>and</strong> at the end of the <strong>running</strong> surface, as the gap between <strong>running</strong> belt <strong>and</strong> frame is<br />

frequently too wide. To reduce the risk for users, h/p/cosmos <strong>running</strong> <strong>machine</strong>s have a gap of less than 4 mm <strong>and</strong><br />

it is marked clearly with additional danger precaution stickers.<br />

Therefore we take it as our responsibility to inform our partners <strong>and</strong> customers about the background <strong>and</strong><br />

necessity only to use medical products in the medical application field. Statutory requirements only permit us to<br />

supply medically classified equipment for medical applications.<br />

<strong>Medical</strong> Device Directive Article 1 Definitions, scope 2 a.)<br />

“medical device” means any instrument, apparatus, appliance, material or other article, whether used alone or in<br />

combination, including the software necessary for its proper application intended by the manufacturer to be used<br />

for human beings for the purpose of: Diagnosis, prevention, monitoring, treatment or alleviation of disease,<br />

diagnosis, monitoring, treatment, alleviation of or compensation for any injury or h<strong>and</strong>icap, investigation,<br />

replacement or modification of the anatomy or of a physiological process, control of conception, <strong>and</strong> which does<br />

not achieve its principal intended action in or on the human body by pharmacological, immunological or metabolic<br />

means, but which may be assisted in its function by such means;<br />

The use of a product not released for medical use by the manufacturer in a medical application can result in<br />

increased risks for the patients <strong>and</strong> for the operator – especially in the case of an accident on the treadmill<br />

ergometer. In this case all parties involved (your insurance company, your professional association, possibly<br />

patients lawyers <strong>and</strong> courts) will question negligence <strong>and</strong> liability. Therefore the use of a sports device for medical<br />

purposes bears increased risks for the distributor, customer, operator, end user <strong>and</strong> last but not least the patient.<br />

Despite that some of the above mentioned laws <strong>and</strong> guidelines are only binding within the European Union, the<br />

patient’s needs, the risks <strong>and</strong> the required safety procedures are the same all over the world. We ask all parties<br />

involved to help in improving safety.<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 36 of 213


Some examples for safety features at h/p/cosmos devices:<br />

Potential isolation<br />

transformer<br />

Opto coupler board Potential equalization<br />

connector<br />

<strong>Medical</strong> devices are equipped with an Emergency-Off-<br />

Button instead of a conventional Safety-Stop-Button. The<br />

emergency off button disconnects the device from the<br />

power supply when pushing the button in emergency<br />

cases. Sport devices mostly are only equipped with an<br />

emercency-stop-button, which only stops the <strong>machine</strong> but<br />

does not disconnet it completely from the mains.<br />

safety, warnings, precautions, prohibitions<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 37 of 213<br />

Shielded motor<br />

cable<br />

Frequency inverter metal housing (for shielding)<br />

Danger zone sticker


Forbidden use<br />

safety, warnings, precautions, prohibitions<br />

N All prohibitions in the chapter "safety precautions, safety regulations, prohibition <strong>and</strong> warnings"<br />

N Do not modify the <strong>running</strong> <strong>machine</strong>s <strong>and</strong> do not connect to other equipment which is not explicitly declared as<br />

compatible by all involved manufactuers.<br />

N The <strong>running</strong> <strong>machine</strong> must not be used without being carefully instructed by specialist staff <strong>and</strong> without having<br />

received notes on the safety regulations.<br />

N Interrupt the training immediately if you start feeling sick or dizzy <strong>and</strong> see your doctor.<br />

N Subject with a cardiac pacemaker or who suffers from any kind of physical restriction must see a doctor before<br />

using the <strong>running</strong> <strong>machine</strong> <strong>and</strong> ask for allowance.<br />

N Children <strong>and</strong> animals are not allowed to use the <strong>running</strong> <strong>machine</strong> <strong>and</strong> must not get near (4m safety area) to it.<br />

N Other use, then explicitly listed directed use at the point field of application.<br />

N In case of any detected <strong>and</strong>/or assumed malfunctions <strong>and</strong>/or defects or unreadable safety warning labels the<br />

device has to be taken out of operation, the device has to be marked <strong>and</strong> secured against operation <strong>and</strong> the<br />

supplier <strong>and</strong> authorized service personnel has to be informed in writing.<br />

N Every kind of overload of test person / subject / patient is forbidden.<br />

N Every kind of use with increased risk, e.g. sprint or medical application with patients <strong>and</strong> increased risk of falling, is<br />

forbidden without using additional safety precautions <strong>and</strong> safety equipment, such as e.g. safety-arch with fall-stop.<br />

N In case of existence of one or more of the following contraindications (see also chapter “contra indications”) we<br />

strongly suggests to see a medical doctor or a specialist before using the <strong>running</strong> <strong>machine</strong>: pregnancy, acute<br />

thrombosis, heart- <strong>and</strong>/or arteries disease, fresh wounds e.g. after surgery, artificial joints <strong>and</strong> limbs, acute<br />

fracture, damaged disc or traumatic caused disease of the spine, diabetes, epilepsy, inflammations, acute<br />

migraine <strong>and</strong> any kind of tumors. Please contact the manufacturer h/p/cosmos, if your doctor or specialist needs<br />

further information about the <strong>running</strong> <strong>machine</strong>.<br />

N The automatic operation (modes profile, cardio, test, remote control via PC <strong>and</strong> peripheral devices) is forbidden, if<br />

the health <strong>and</strong> condition of the test person / subject / patient does not allow it <strong>and</strong> if a doctor has not authorized it.<br />

Disregard may cause injuries <strong>and</strong> dangerous health problems <strong>and</strong> even death.<br />

N All use with wheels (cycling, wheelchair, inline-skating or roller-ski) is forbidden at st<strong>and</strong>ard <strong>running</strong> surfaces <strong>and</strong> -<br />

belts. Such applications are only allowed at special <strong>running</strong> <strong>machine</strong>s marked with „r“(e.g. h/p/cosmos saturn<br />

300/100r). Additionally a safety harness with fall-stop is prescribed. The brakes at the bike respectively wheelchair<br />

must be deactivated (e.g. demount brake suspension) <strong>and</strong> the stop at the inline-skates must be demounted during<br />

the exercise on the <strong>running</strong> <strong>machine</strong>. Do not forget to reactivate brakes before using this equipment outdoor again.<br />

N Never use <strong>running</strong> shoes or other shoes with spikes or studs at the st<strong>and</strong>ard <strong>running</strong> surface resp. <strong>running</strong> belt.<br />

These applications are only allowed on special <strong>running</strong> <strong>machine</strong>s with suffix „rs“ which have as an option<br />

additionally an extra thick <strong>running</strong> belt. A safety arch with chest belt <strong>and</strong> fall-stop is also obligatory in this case.<br />

N In environmental conditions, others than specified in the chapter „technical specifications“ <strong>and</strong> “environmental<br />

conditions“, h/p/cosmos devices must not be used (e.g. in wet <strong>and</strong> humid areas, swimming pools, sauna,<br />

environmental chambers, high-pressure-, low-pressure, altitude- <strong>and</strong> oxygen-chambers, etc.).<br />

N Do not jump onto the rotating <strong>running</strong> belt. Do not jump off the rotating <strong>running</strong> belt (not even towards the front).<br />

Do not stop moving or turn around on the rotating <strong>running</strong> belt. Never run sideways or backwards. Do not do<br />

anything which could interfere with your balance. Do not set too high loads (speed, elevation).<br />

At the suspicion of unauthorized access or other reasons to lock the <strong>running</strong> <strong>machine</strong>, it has to be<br />

locked for starting: See „option 40“ in chapter „option settings“. With option 41 ... 44 you can lock<br />

also separate modes (manual, profile, cardio, test).<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 38 of 213


[5.]Application, Indication, Contra-Indication<br />

[5.A] Fields of application / intended use<br />

application, indication, contra-indication<br />

All <strong>running</strong>-<strong>machine</strong>s are built corresponding to the classes S,I,A based on EN 957-1 / EN 957-6.<br />

The indications, discussed in the following, can be treated by any h/p/cosmos treadmill version, although there are more<br />

or less recommended solutions.<br />

[5.A1] Professional application in sports <strong>and</strong> fitness (with c mark on nameplate)<br />

The <strong>running</strong> <strong>machine</strong> models designed for sports <strong>and</strong> fitness have not been tested for application in medical areas<br />

<strong>and</strong> therefore are not qualified for that field of application. They must not be connected to other medical devices.<br />

You can apply these models in endurance training walking <strong>and</strong> <strong>running</strong>.<br />

[5.A2] Professional application in the medical field (with n mark on nameplate)<br />

The <strong>running</strong> <strong>machine</strong> models designed for professional application in the medical field are also qualified for sports<br />

<strong>and</strong> fitness use as well as for application in the medical field.<br />

N Endurance training walking <strong>and</strong> <strong>running</strong><br />

N Endurance tests <strong>and</strong> EMG-measuring with the subject in the laboratory (only with safety arch / fall stop)<br />

N Ergometry on the Running-Machine, Exercise- ECG <strong>and</strong> Ergospirometry (only with safety arch / fall stop)<br />

N Gait training<br />

N Recreational fitness training<br />

N Diagnosis <strong>and</strong> prognosis of cardiovascular disease, specifically coronary artery disease (CAD) through<br />

treadmill stress test.<br />

N Therapy of orthopaedic problems, for example Arthritis in foot ankle, Femoropatellares Syndrom, knee joint<br />

problems, anterior cruciate ligament plastic through treadmill therapy.<br />

N Gait analysis <strong>and</strong> gait correction for stroke patients through treadmill therapy.<br />

Children are only allowed to use the <strong>running</strong> <strong>machine</strong> under constant supervision of a<br />

medical doctor <strong>and</strong> only when safeguarded by safety harness with fall-stop. Additionally<br />

h/p/cosmos suggests the use of adjustable h<strong>and</strong>rails.<br />

[5.B] Risks of any application<br />

The dosing of loading the patient in form of velocity, stride length, kind of course of motion on the treadmill,<br />

elevation, duration, total distance as well as loading / unloading of the patient by an unloading system fall to the<br />

medical doctor in connection with the therapist. The treadmill’s manufacturer cannot make any declaration or<br />

recommendation, because it is individually depending on the patient, the clinical picture, the degree of disability<br />

<strong>and</strong> the progress of rehabilitation.<br />

In case of overload of the patient by wrong application <strong>and</strong>/or by wrong estimation <strong>and</strong>/or by unexpected problems<br />

also in correct application it may come to following risks, problems <strong>and</strong> injuries:<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 39 of 213


application, indication, contra-indication<br />

[5.C] Risk <strong>and</strong> benefit of treadmill application<br />

The use of treadmills is supported by medical doctors all around the world for more than 50 years now. Since all<br />

available specialists literature points out the benefit of treadmill application explicitly, <strong>and</strong> the risks are well controllable,<br />

the benefit exeeds the risks. The manufacturer releases the medical product for the described applications <strong>and</strong><br />

indications.<br />

Clinical studies are obsolete <strong>and</strong> have not been ordered by h/p/cosmos by now, since there is a huge amount of<br />

literature, over 50 years of experience in application worldwide <strong>and</strong> the performance data according to EN 957-6.<br />

The medical treadmills may be operated only in medical facilities by medical personnel. The safety requirements for<br />

medical electrical equipment - in particular requirements for medical electrical equipment <strong>and</strong> medical electrical systems<br />

for the medical care in a home environment (IEC 60601-1-11) - are not considered in the medical device, h/p/cosmos<br />

treadmill ergometer.<br />

An application of the h/p/cosmos medical products treadmills in a home environment <strong>and</strong> / or in the<br />

open <strong>and</strong> / or by not medically trained personnel is not allowed.<br />

Overload of the patient by wrong application <strong>and</strong>/or by misinterpretation <strong>and</strong>/or by unexpected problems during correct<br />

application may result in the following risks, problems <strong>and</strong>/or injuries:<br />

Problem & injury risk Risik control<br />

Injuries of the muscoskeletal system (joints, tendons, 1. Note in instructions <strong>and</strong> warning labels on the user<br />

ligaments, muscles, bones) due to overloading <strong>and</strong>/or terminal, that the use of the device may only occur<br />

fall over <strong>and</strong> skin abrasions, bruises, sprains <strong>and</strong> torn under supervision by medical or therapeutic staff.<br />

ligaments, fractures <strong>and</strong> broken neck in worst case 2. Proper loading dose must be prescribed by doctors<br />

resulting in death.<br />

or therapists as qualified personnel only.<br />

3. Notes in instructions <strong>and</strong> warning labels on the user<br />

terminal that operating personnel must stay in<br />

permanent patient environment of 1.5 m.<br />

4. Recommendation <strong>and</strong> application of weight support<br />

system <strong>and</strong> / or fall stop device, such as the<br />

h/p/cosmos safety arch with harness <strong>and</strong> <strong>and</strong><br />

automatic shut-off of the <strong>running</strong> belt.<br />

5. Comprehensive instruction of the patient before the<br />

start of treadmill therapy with an explanation that the<br />

drive of the <strong>running</strong> belt is very strong <strong>and</strong> does not<br />

stop automatically, if the patient remains st<strong>and</strong>ing.<br />

Overload of the cardiovascular system in extreme 1. Note in user’s manual for correct load dosing by a<br />

situations with cardiac arrest <strong>and</strong> death.<br />

medical doctor or therapist.<br />

2. Note to keep a defibrillator ready for reanimation.<br />

Country-specific this is already st<strong>and</strong>ard.<br />

3. Integrated cardio mode, which reduces the load after<br />

reaching the recommended maximum heart rate <strong>and</strong><br />

stops the treadmill in case overload or breakdown of<br />

the heart rate measurement.<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 40 of 213


application, indication, contra-indication<br />

In case of correct load dosing (according to the clinical picture <strong>and</strong> condition of the patient) <strong>and</strong> intended use, the<br />

risks are comparable to the risks at walking / <strong>running</strong> on the ground: Falling off the treadmill with skin abrasions,<br />

bruises, torn ligaments, fractures <strong>and</strong> in worst case e.g. broken neck resulting in death.<br />

[5.C1] Summary / potential risk remains<br />

Especially for ergometry (stress-test) application an extra risk for the patient stays since it is part of the test to<br />

stress the patient until cardiovascular problems occur. The risk stays even at presence of trained, medical staff<br />

<strong>and</strong> even at availability of medical equipment for reanimation. Thereby the higher risk cannot be reduced even by<br />

the knowledge of contra indications.<br />

Restrictions of cognition or loading capacity cannot be excluded by contraindications also.<br />

[5.D] Contraindications for any application:<br />

Before treadmill application (e.g. stress test, therapy, etc.) contraindications have tob e excluded.<br />

Contraindications include pregnancy, acute thrombosis, heart- <strong>and</strong>/or arteries disease, fresh wounds e.g. after<br />

surgery, artificial joints <strong>and</strong> limbs, acute fracture, damaged disc or traumatic caused disease of the spine, diabetes,<br />

epilepsy, inflammations, acute migraine <strong>and</strong> any kind of tumors.<br />

The division of absolute <strong>and</strong> relative contraindications has proved itself of value. In case of relative<br />

contrainidcations a try of application may be started if the possible benefit exceeds the risk of application. The<br />

decision has to be made by the medical doctor.<br />

Absolute contraindications:<br />

N Acute myocarinal infarction<br />

N Instable Angina pektoris<br />

N Cardiac arrhythmia pathology <strong>and</strong>/or limited hemodynamics<br />

N Symptomatic massive aorta stenosis<br />

N Uncompensated heart insufficiency<br />

N Acute pulmonary embolilsm<br />

N Acute myocarditis<br />

N Acute pericarditis<br />

N Acute aorta dissection<br />

Relative Contraindications:<br />

N Main artery disease<br />

N Cardiac valve disease of moderate severity<br />

N Known electrolyte disturbance<br />

N Arterial hypertonia (RR > 200 mm Hg syst. > 110 mm Hg diast.)<br />

N Tachyarrhythmia or Bradyarrhythmia<br />

N Hypertrophic cardiomyopathy <strong>and</strong> other kind of defluxionwayobstruction<br />

N Higher degree AV-blocking<br />

N Physical <strong>and</strong>/or mental disabilities<br />

Also see following guidelines: Deutsche Gesellschaft für Kardiologie – Herz- und Kreislaufforschung e. V:<br />

http://leitlinien.dgk.org/<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 41 of 213


application, indication, contra-indication<br />

It cannot be excluded that further contraindications may occur. This has to be evaluated by the responsible<br />

medical doctor.<br />

[5.E] Application treadmill ergometry: Cardiology ECG Stress Test on the treadmill<br />

(not for sports <strong>machine</strong>s; only for medical <strong>running</strong> <strong>machine</strong>s with n mark on nameplate)<br />

In order to check reaction of the patient’s cardiovascular system to physical load, the patient is confronted with steadily<br />

increasing load on the treadmill.<br />

There are certain given protocols that give an increasing load to the patient either by<br />

N increasing velocity at constant elevation,<br />

N increasing elevation at constant velocity or<br />

N a combination of both.<br />

The h/p/cosmos treadmill is equipped with several load protocols that can be chosen in the “test-mode” (see chapter<br />

”operation”). You also have the possibility to create a sef-defined profile, either using the “graded test” or the “free<br />

deinable user profiles”.<br />

Also see guidelines of German or American Heart Associations:<br />

Deutsche Gesellschaft für Kardiologie – Herz- und Kreislaufforschung e. V: http://leitlinien.dgk.org/<br />

American Heart Association. http://www.heart.org <strong>and</strong> http://circ.ahajournals.org/content/96/1/345.full<br />

[5.E1] Indications<br />

If controlling the treadmill via ECG, pay special attention to the ECG manufacturer’s manual!<br />

The summary of clinical data <strong>and</strong> the risk management of h/p/cosmos is NOT covering any<br />

measured parameter <strong>and</strong>/or vital functions which are measured <strong>and</strong>/or detected by host devices<br />

(e.g. ECG, spirometry devices, EMG, etc.) which are connected to the treadmill.<br />

Suspicion of cardiovascular diseases (diagnostics <strong>and</strong> therapy)<br />

[5.E2] Benefit<br />

Diseases of the cardiovascular system can be detected <strong>and</strong> treated early <strong>and</strong> in a controlled surrounding. For the<br />

patient it is better to sustain a cardiac arrest caused by high load, in a clinic under supervision of a medical doctor<br />

with medical equipment <strong>and</strong> a defibrillator, than at work, at home or during sports.<br />

The benefit of this application can only be achieved in combination with an external ECG that is not delivered by<br />

h/p/cosmos.<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 42 of 213


application, indication, contra-indication<br />

[5.F] Application gait analysis <strong>and</strong> rehabilitation on the treadmill with force measurement plates<br />

(not for sports <strong>machine</strong>s; only for medical <strong>running</strong> <strong>machine</strong>s with n mark on nameplate)<br />

The h/p/cosmos Kistler gaitway is an h/p/cosmos treadmill of mercury size, equipped with two Kistler force<br />

measurement plates; it is the only h/p/cosmos treadmill with these features. The system is measuring the ground<br />

reaction force <strong>and</strong> thereby collecting a huge amount of gait parameter. With the Kistler gaitway software the total force<br />

path can be displayed. Experts know to interpret the force path <strong>and</strong> use it as basis for therapy.<br />

[5.F1] Indications<br />

N Pathologic gait pattern<br />

N Arthritis in ankles<br />

N Femoropatellar syndrome<br />

N Knee joint problems<br />

N anterior cruciate ligament reconstruction<br />

[5.F2] Benefit<br />

Some examples for the clinical benefit (h/p/cosmos gaitway with KISTLER force measurement plates:<br />

After four weeks of gait therapy you can see a considerable improvement of gait symmetry, socumented by a<br />

comparison of the ground reaction force.<br />

Source: 20061114_h-p-cosmos_gaitway_gait_analysis_workshop_Radovanovic_en.ppt<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 43 of 213


Further literature about gait analysis, approving the benefit of using treadmills:<br />

http://www.ortholine.de/09_nagel.pdf<br />

Instrumentale Ganganalyse – Praktische Anwendung und Versorgungsbeispiele<br />

Versorgungsbeispiel 1: Bewegungsanalyse und Einlagen<br />

Dr. rer. medic. A. Nagel / Dr. med. A. Spitz<br />

[5.G] Application gait analysis <strong>and</strong> rehabilitation on the treadmill with neurologic patients<br />

application, indication, contra-indication<br />

(not for sports <strong>machine</strong>s; only for medical <strong>running</strong> <strong>machine</strong>s with n mark on nameplate)<br />

For neurologic application the h/p/cosmos locomotion with unweighting system airwalk se is the recommended solution.<br />

Neurologic gait therapy could be performed using any h/p/cosmos treadmill, but with less comfort.<br />

Manual locomotion therapy is the basis of physiologic treatment of neuologic patients. The patient walks on the treadmill<br />

while the therapist is sitting next to it, supporting <strong>and</strong> correcting the patients motion.<br />

As for neurologic patients only very low speed is needed, it makes most sense to control the treadmill in the “manual<br />

mode” at a speed of 0.1 to 6.0 km/h.<br />

[5.G1] Indications<br />

Pathologic gate pattern, e.g. after a stroke<br />

(gait analysis <strong>and</strong> gait correction as well as rehabilitation of stroke patients)<br />

[5.G2] Benefit<br />

With the aid of additional safety equipment such as an unweighting system or arm support gait training can be<br />

carried out in a controlled surrounding on the treadmill.<br />

The additional safety equipment prevents from falling <strong>and</strong> the duration of therapy can be increased.<br />

By motion support in manual locomotion therapy the still adaptive neuronal system is stimulated <strong>and</strong> for many<br />

patients an improvement of the motion pattern can be reached under safe circumstances.<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 44 of 213


[6.]Installation & commissioning<br />

[6.A] Transport, unpacking <strong>and</strong> packaging<br />

installation & commissioning<br />

When receiving the <strong>machine</strong> in a crate or unpacked, make sure the <strong>machine</strong>, the accessories <strong>and</strong>/or the packaging is<br />

not damaged. If you discover any damage <strong>and</strong>/or missing parts make a note on the packing-list / delivery note of the<br />

carrier. Inform h/p/cosmos <strong>and</strong> your dealer immediately in writing about any damages <strong>and</strong>/or missing parts.<br />

The manufacturer does not undertake any liability for any damage, complaints <strong>and</strong> missing parts<br />

which is not reported immediately by delivery on the packing-list / delivery note.<br />

Before you unpack the <strong>machine</strong> <strong>and</strong> accessories read instructions on the crate. Make sure that the <strong>machine</strong>, power<br />

connection cable or any optional equipment will not be damaged during unpacking. Pay special attention on small parts,<br />

so that you do not dispose them or any instructions with the packaging. Within Germany mostly all devices are delivered<br />

<strong>and</strong> assembled by h/p/cosmos directly or by an authorised forwarder. If delivered by h/p/cosmos the packaging will be<br />

taken back <strong>and</strong> recycled.<br />

If the <strong>running</strong> <strong>machine</strong> is being delivered by a carrier, you can recycle the packaging yourself or send it back to the<br />

manufacturer (transportation is to be paid by the customer). Often a recyclable transport tool or packaging or a<br />

transport-fuse (metal angle with screws) is included in the extent of delivery.<br />

Ask your dealer <strong>and</strong> the carrier to take the packaging <strong>and</strong> the recyclable transport tool back to your dealer or to<br />

h/p/cosmos at your own costs. In some cases a credit note can be granted. Special packaging <strong>and</strong>/or carrier<br />

constructions must not be disposed unauthorized.<br />

[6.A1] Transport to upper or lower floors <strong>and</strong> through narrow doors<br />

With the st<strong>and</strong>ard models (sizes <strong>running</strong> surface 150/50cm, 170/65cm, 190/65cm / except adjustable h<strong>and</strong>rails)<br />

the h<strong>and</strong>rail at the left is easy to remove, as there are no cables inside. The h<strong>and</strong>rail at the right can be<br />

unscrewed, can be put 90° to the left ("lay it on the <strong>running</strong> belt") <strong>and</strong> then the h<strong>and</strong>rail has to be temporarily fixed<br />

with special transport fuse construction (available at h/p/cosmos) in order to avoid damages during the transport.<br />

Under the condition that the right h<strong>and</strong>rail lies flat <strong>and</strong><br />

temporarily fixed on the <strong>running</strong> surface, the whole<br />

<strong>machine</strong> can be put 90° to the right to an upright<br />

position. So the width of the <strong>machine</strong> is fairly reduced<br />

<strong>and</strong> can be easily moved through narrow doors or<br />

windows.<br />

A special transport tool is available to move the<br />

<strong>machine</strong> through aisles - see below (available as loan<br />

object at h/p/cosmos).<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 45 of 213


installation & commissioning<br />

Transport angle construction <strong>running</strong> surface 150/50cm [cos10971]<br />

Transport angle construction <strong>running</strong> surfaces 170 <strong>and</strong> 190/65cm [cos10314]<br />

Transport angle construction <strong>running</strong> surface 200 ... 300/75 ... 125cm [cos14090]<br />

Transport tool for construction size <strong>running</strong> surface 150/50cm [cos10972]<br />

Transport tool for construction size <strong>running</strong> surfaces 170 <strong>and</strong> 190/65cm [cos13794]<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 46 of 213


[6.A2] Transport with trolleys<br />

For an easy transport put 2 usual trolleys under the device (platform<br />

approximately 30cm x 30cm with 4 maneuverable rolls each) to make the<br />

device “mobile”.<br />

Trolley with 4 rolls for max. 300kg [cos13016]<br />

Trolley with 4 rolls for max. 500kg [cos13672]<br />

[6.A3] Transport of oversize <strong>running</strong> <strong>machine</strong>s<br />

installation & commissioning<br />

For big devices it might be the best <strong>and</strong> only possibility to move them through the window with the help of a special<br />

basket <strong>and</strong> a heavy mobile crane.<br />

[6.B] Mechanical installation<br />

Transports of heavy devices must only be carried out by authorised staff in compliance with<br />

the safety st<strong>and</strong>ards. Otherwise there is serious danger for people <strong>and</strong> devices.<br />

N In order to ensure proper installation <strong>and</strong> safety, either the manufacturer, an authorised service crew or an<br />

authorised dealer must always perform transport <strong>and</strong> installation of the devices<br />

N For reasons of safety please make sure that there is a safety zone of at least L: 2.5m x W: 1.0m behind <strong>and</strong><br />

L: 1.0m x B: 1.0m to the front the <strong>running</strong> <strong>machine</strong>. Please note, that with the optional equipment reversible<br />

<strong>running</strong> belt rotation you also have to ensure a safety zone of L: 2.5m x W: 1.0m in front of the <strong>running</strong> <strong>machine</strong>.<br />

N Put an aerobic mat (or something similar) in front of <strong>and</strong> behind the <strong>running</strong> <strong>machine</strong> to protect the subject when<br />

falling. Put the device on a rubber mat as large as the device to protect the floor from dust <strong>and</strong> scratches, to<br />

ensure a firm st<strong>and</strong> <strong>and</strong> to reduce noise.<br />

N The provided space for the <strong>running</strong> <strong>machine</strong> must be even <strong>and</strong> horizontal.<br />

N Models with leveling sockets (adjustable "feet") at the rear of the <strong>running</strong> <strong>machine</strong> have to be adjusted so that they<br />

have a firm st<strong>and</strong>, otherwise it could lead to noises like knocking or rattling during the training. Check the weight<br />

load pressure on the leveling sockets (weight of the <strong>running</strong> <strong>machine</strong> at the rear), by trying to lift the frame off the<br />

ground at the rear of the <strong>running</strong> <strong>machine</strong> one side after the other. That way you can find out, whether you have<br />

got the same load on the two sockets. The levelling sockets <strong>and</strong> the front wheels of the elevation system may<br />

cause a print in the flooring. In case of sensitive flooring please use h/p/cosmos floor mats for protection:<br />

Floor protection mat 150/50 for <strong>running</strong> <strong>machine</strong>s with deck size 150x50 cm [cos14005]<br />

Floor protection mat 170/65 for <strong>running</strong> <strong>machine</strong>s with deck size 170x65 cm [cos14042]<br />

Floor protection mat 190/65 for <strong>running</strong> <strong>machine</strong>s with deck size 190x65 cm [cos14043]<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 47 of 213


installation & commissioning<br />

Other sizes available on request.<br />

N For models without leveling sockets please make sure that the floor condition is even <strong>and</strong> horizontal. The <strong>machine</strong><br />

st<strong>and</strong>s with the entire ground frame on the floor.<br />

N Models with a <strong>running</strong> surface of 200 x 75 cm or bigger: Make sure the floor condition is even <strong>and</strong> horizontal. The<br />

<strong>machine</strong> st<strong>and</strong>s with the entire ground frame on the floor. There are leveling sockets incorporated, but they should<br />

only be used in case of problems with <strong>and</strong> uneven floor.<br />

N Before you switch on the device you need to de-install the mechanical transportation fuse, fixing the upper <strong>and</strong><br />

lower frame.<br />

N The bearing capacity of floor <strong>and</strong> ceiling in the building must be higher than the weight of the <strong>machine</strong>. It has to be<br />

approved for the h/p/cosmos device by an authorized body of the operator. Example h/p/cosmos saturn 4.0<br />

300/100r: Dimensions ground frame <strong>running</strong> <strong>machine</strong>: L: 3400 mm x B: 1490 mm = 5.066 m² platform. Net weight<br />

of <strong>running</strong> <strong>machine</strong>: 1300 kg, static bodyweight of subject: 200 kg, dynamical weight of subject: 1200 kg (up to 6<br />

times the bodyweight), over all weight load of system at platform: 2500 kg = bearing capacity of platform: 493.48<br />

kg (~ 500 kg) / m².<br />

N At h/p/cosmos gaitway with integrated force plate the transportation locks have to be removed after erection. Drive<br />

elevation up to max of 25 %. Loosen the four M8 screws, indicated with the red arrows at the underside of the aluforce<br />

plates <strong>and</strong> remove them. Drive elevation down to 0 %. Loosen the locknuts at the levelling sockets. Now<br />

level out the <strong>running</strong> <strong>machine</strong> with the help of the Kistler Software <strong>and</strong> then tighten the locknuts again.<br />

N The <strong>running</strong> belt has to be checked <strong>and</strong> if necessary adjusted (see maintenance instructions), so that it is placed<br />

exactly in the middle of the two rollers, after installation or change of location.<br />

N If the <strong>running</strong> <strong>machine</strong> is installed safely <strong>and</strong> horizontal, it can be plugged in (see installation instructions) <strong>and</strong><br />

taken into operation.<br />

N It is recommended to lubricate the <strong>running</strong> surface with 40ml Silicone-Oil before using it for the first time. See<br />

separate chapter "maintenance".<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 48 of 213


[6.C] Electrical installation<br />

installation & commissioning<br />

An overload or an voltage drop (even short-time) of more than 20 % of the mains voltage might<br />

cause malfunctions <strong>and</strong>/or defects <strong>and</strong> might totally switch-off the <strong>running</strong> <strong>machine</strong>. At such high<br />

voltage drops or at total power failure the <strong>running</strong> <strong>machine</strong> switches off <strong>and</strong> the <strong>running</strong> belt stops.<br />

To start again you have to switch-on the <strong>running</strong> <strong>machine</strong> at the main „ON“ switch <strong>and</strong> than press<br />

START at the keyboard.<br />

N Installation of any electrical device <strong>and</strong> h/p/cosmos <strong>running</strong>-<strong>machine</strong>s must only be carried out at a voltage power<br />

connection with ground wire (earthing) including leakage-current protection-switch (interrupter) <strong>and</strong> according to<br />

VDE 0100 or/<strong>and</strong> the currently valid regulations <strong>and</strong> directives. Requirements for special locations, areas <strong>and</strong><br />

establishments (e.g. medically used areas) must be strictly adhered. PE-connection (protection-earth-contact) is<br />

stipulated at all <strong>running</strong>-<strong>machine</strong>s. The voltage drop between the beginning of the consumer’s installation <strong>and</strong> the<br />

wall socket must not be more than 4% (DIN VDE 0100-520). It is the consignee’s <strong>and</strong> user’s personal responsibility<br />

to check the correct functions of the mains connection including the outlets <strong>and</strong> an authorized electrical engineer<br />

must check these points for perfect functions regularly (1 ... 4 years). Inspections of the electrical installations<br />

within the building are not incumbent on the supplier h/p/cosmos.<br />

N Read the nameplate of the <strong>machine</strong> before connecting the <strong>machine</strong> to the power outlet, just in case the <strong>machine</strong><br />

was designed for special voltage supply (e.g. 110 Volts <strong>and</strong> 25 Ampere). As st<strong>and</strong>ard, a usual power supply of<br />

230 Volt / AC, 50/60 Hz is sufficient for most of the <strong>running</strong> <strong>machine</strong>s.<br />

N The bigger <strong>running</strong> <strong>machine</strong>s (size <strong>running</strong> surface 190/65cm 3p up to 300/125cm)<br />

require a 3 phase AC power supply: 3 x 400 Volt / 32 Amp (<strong>running</strong> <strong>machine</strong><br />

190/65cm 3p 16 Amp) fused (read name of the <strong>machine</strong> just to make sure) <strong>and</strong> righth<strong>and</strong>ed<br />

polarized (clockwise rotation field).<br />

Pictures power-plug (on device) <strong>and</strong> power-outlet:<br />

CEE power outlet (socket), 5-pin, 16 Ampere<br />

[cos11092]<br />

CEE power outlet (socket), 5-pin, 32 Ampere<br />

[cos11090]<br />

Power supply for h/p/cosmos pulsar 3p<br />

Power supply for h/p/cosmos venus <strong>and</strong> saturn<br />

400 Volt AC 3~/N/PE f: 50/60 Hz Imax: 16.0 A 400 Volt AC 3~/N/PE f: 50/60 Hz Imax: 32.0 A<br />

3 phase current with clockwise rotation field.<br />

3 phase current with clockwise rotation field.<br />

16 Ampere fuse (C-characteristic)<br />

32 Ampere fuse (C-characteristic)<br />

The plug-in has 5 plugs:<br />

The plug-in has 5 plugs:<br />

L1 = Phase 1 (black wire)<br />

L1 = Phase 1 (black wire)<br />

L2 = Phase 2 (brown wire)<br />

L2 = Phase 2 (brown wire)<br />

L3 = Phase 3 (black or grey wire)<br />

L3 = Phase 3 (black or grey wire)<br />

N = Neutral (blue wire)<br />

N = Neutral (blue wire)<br />

PE = Ground / Earth (yellow-green wire)<br />

PE = Ground / Earth (yellow-green wire)<br />

The mentioned wire colours are applicable for Germany <strong>and</strong> different colours may be applicable in other countries.<br />

Voltage values for 3-phase voltage power supply:<br />

from L1 to L2: 400 Volt, from L1 to L3: 400 Volt, from L2 to L3: 400 Volt<br />

from L1 to N: 230 Volt, from L2 to N: 230 Volt, from L3 to N: 230 Volt,<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 49 of 213


installation & commissioning<br />

N After having switched on a 3 phase <strong>machine</strong>, an elevation of 0 % has to be adjusted automatically. If this is not the<br />

case, the unit is to be switched off immediately <strong>and</strong> the two phases of the wall socket have to be changed so that<br />

the device is supplied with right-h<strong>and</strong>ed polarized power. Otherwise the elevation cannot be set correctly <strong>and</strong> the<br />

power supply of the motor for controlling the elevation is switched off via the limit stop switch at the ground frame.<br />

N Use a usual 16 Ampere expulsion fuse with C-tripping-circuit for your house distribution. For <strong>running</strong> <strong>machine</strong>models<br />

with a construction size of the <strong>running</strong> surface of 200/75cm up to 300/125cm you will need a 32 Ampere<br />

expulsion fuse with C-tripping-circuit. If however, the expulsion fuses switch off when starting the device, the circuit<br />

has to be secured with a blowout fuse or an expulsion fuse with another tripping-circuit (e.g.: K–fuse). Read name<br />

plate <strong>and</strong> technical data of the <strong>machine</strong> just to make sure.<br />

N For further questions please ask your electrical engineer or h/p/cosmos.<br />

N Before installing the <strong>running</strong> <strong>machine</strong> please compare the specifications on the nameplate concerning the mains<br />

voltage <strong>and</strong> the mains frequency with your local characteristics. Connection only if identical.<br />

N Check the main lead, the voltage power supply outlet <strong>and</strong> ground wire protection-contacts before plugging it in.<br />

Damaged leads <strong>and</strong> couplers <strong>and</strong> defective or dirty contacts have to be exchanged immediately. Rubber-leads can<br />

get porous <strong>and</strong> friable after some years.<br />

[6.C1] Name plate<br />

N Plug the <strong>running</strong> <strong>machine</strong>s directly into the wall socket. Each <strong>running</strong> <strong>machine</strong> has to be<br />

connected to a separate circuit. The socket has to be marked with name <strong>and</strong> serial number of<br />

the <strong>running</strong> <strong>machine</strong>. The use of an extension cables or a multiple plug sockets is not<br />

allowed.<br />

N Electrical devices with mains connections must neither be used in wet <strong>and</strong> humid areas (e.g.<br />

swimming pool, sauna, etc.) nor in environmental chambers.<br />

picture left: name plate example sports device, model: h/p/cosmos quasar<br />

picture right: name plate example medical device, model: h/p/cosmos saturn 300/125r<br />

For full details please refer to the chapter „technical data“ in this manual.<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 50 of 213


[6.C2] Electric safety measurements <strong>and</strong> „First Measured Values“<br />

Immediately at first installation at the customer’s site an electric safety test<br />

<strong>and</strong> measurement has to be performed for „Protective Earth Resistance“,<br />

„Isolation Resistance“ <strong>and</strong> „Leakage Current“ <strong>and</strong> the values have to be<br />

recorded on a special protocol [cos11690xx] <strong>and</strong> marked with „first<br />

measured values“. For details please refer to the chapter „maintenance /<br />

inspections“ in this manual. One copy of this protocol [cos11690xx]<br />

remains at the owner’s manual <strong>and</strong> the original of the protocol with the<br />

„first measured values“ shall be sent to the manufacturer h/p/cosmos.<br />

Picture: Example for electric safety testing device based on IEC 601-1<br />

[6.C3] Potential equalization<br />

The potential equalization cable [cos10223] must be connected with the<br />

connector on the device plug <strong>and</strong> with the potential-compensationbearing<br />

within the medically used room. During installation, connecting or<br />

disconnecting the potential equalization, the device (<strong>running</strong> <strong>machine</strong>)<br />

must not be connected to the power supply. Protection against electric<br />

shock has to be provided in the finished product/device.<br />

installation & commissioning<br />

N When used in the medical field, all devices within the system have to be connected via a<br />

potential-compensation-cable in „star-connection“ with the potential-compensation-<br />

bearing within the medically used room.<br />

N First connect the potential compensation (potential equalization) with the corresponding<br />

plug pin (next to the main switch at the front) <strong>and</strong> then connect the mains plug.<br />

N During electric safety measurements <strong>and</strong> test (leakage currents, etc.) the potential<br />

equalization cable has to be disconnected temporarily.<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 51 of 213


[6.C4] Plug assignment at the UserTerminal<br />

installation & commissioning<br />

If the UserTerminal was disassembled during transport please connect the cables according to following<br />

illustration:<br />

N Devices with MCU5 control unit (as of manufacturing date 12/2007)<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 52 of 213


N Devices with MCU4 control unit (until manufacturing date 12/2007)<br />

installation & commissioning<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 53 of 213


installation & commissioning<br />

N For devices with MCU4 control unit (until manufacturing date 12/2007) <strong>and</strong> reverse belt rotation<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 54 of 213


Picture top:<br />

MCU5-control-board <strong>and</strong> display board inside the UserTerminal<br />

Picture right:<br />

Sticker for plug references at the back plate of the UserTerminal<br />

Above: Drawing cable connection for ON/OFF button <strong>and</strong> emergency stop in<br />

the UserTerminal<br />

Right h<strong>and</strong> side: Ground connection at the UserTerminal<br />

Note: GND / GND = ground<br />

installation & commissioning<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 55 of 213


[6.D] Starting the <strong>running</strong> <strong>machine</strong>s<br />

[6.D1] Unlocking of the emergency-stop switch<br />

installation & commissioning<br />

In order to be able to switch on the device, you may have to unlock the Emergency-stop switch first by pulling it<br />

(depending on the type) or turning it around (to the left h<strong>and</strong> side, see symbol).<br />

[6.D2] Switching the device on<br />

If the <strong>running</strong> <strong>machine</strong> is at an angle of elevation > 0 when switching it on, it will<br />

automatically drive into the position zero (display: ELEVATION „INIT“). So please take care<br />

before switching on, that while driving down the elevation no harm can be done to persons<br />

<strong>and</strong> there are no objects under the <strong>running</strong> <strong>machine</strong>.<br />

No. Illustration Description<br />

[01]<br />

[02]<br />

[03]<br />

Release all emergency stop buttons of the treadmill-ergometer.<br />

picture left unlatching by turning<br />

picture below unlatching by pulling<br />

The expulsion fuse at the front of the device (front terminal below<br />

the hood) has to be switched on.<br />

Switch on the <strong>running</strong> <strong>machine</strong> by using the white key (until 2003<br />

green "ON" or "I" key). at the FrontTerminal. At some models built<br />

before the year 12/2007 the location of the ON-/OFF-key could<br />

also be the UserTerminal.<br />

The indicator light within the key is glowing (if the indicator does<br />

not flash up, please check the power supply, the expulsion fuse<br />

<strong>and</strong> the Emergency-stop).<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 56 of 213


installation & commissioning<br />

Devices for <strong>Sports</strong> <strong>and</strong> Fitness manufacturing before manufacturing date 12/2007<br />

The <strong>running</strong> <strong>machine</strong> can be switched on directly at the expulsion fuse at the FrontTerminal (frontal section below<br />

the hood) to the condition "st<strong>and</strong>-by mode" (if the emergency-stop has been activated a flashing notice „PULL<br />

STOP“ appears on the display.).<br />

Expulsion fuse <strong>and</strong><br />

main switch 2- or 3-pole<br />

[6.D3] Switching the device off<br />

ON / OFF button<br />

with control lamp<br />

ON / OFF button<br />

until 2003<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 57 of 213<br />

Main switch from construction<br />

size <strong>running</strong> surface<br />

200...300/75...125cm onwards<br />

N The ON-/OFF-intervals must not be shorter than 1-2 minutes (at models with 3-phase<br />

connection: 2-3 minutes). Otherwise it could lead to interference in the adjustment of<br />

the motor or to the failure of the backup.<br />

N The models for medical application with the isolation transformer have an inrush-<br />

current limiter (surge guard). Too short switch on / switch off intervals will lead to a<br />

deactivation of the inrush current limiter <strong>and</strong> results in an overload of the circuit fuse.<br />

N For the professional use, where the devices are often being used daily, we recommend<br />

to switch on the devices in the morning, <strong>and</strong> leave it in the st<strong>and</strong>-by mode during the<br />

day.<br />

Switch off the <strong>running</strong> <strong>machine</strong> by pressing the black key (until 2003 red “O” key) at the UserTerminal (some<br />

models at the front of the device). The indicator within the key is extinguished. At some models built before<br />

12/2007 the location of the ON-/OFF-key could also be the UserTerminal.<br />

Devices for <strong>Sports</strong> <strong>and</strong> Fitness manufacturing before 12/2007<br />

The <strong>machine</strong> can be switched off directly with the explusion fuse at the FrontTerminal (frontal section below the<br />

hood).


[6.E] Emergency-Stop / Emergency-Off<br />

When in danger of falling or in any<br />

emergency case please press the<br />

red button of the emergency-stop<br />

at the UserTerminal. At devices<br />

without UserTerminal the red<br />

emergency-stop-button is located<br />

directly at the h<strong>and</strong>rail.<br />

In order to switch on the device<br />

again, unlock the Emergency-stop<br />

by pulling or by turning it slightly<br />

to the left (depending on the<br />

type).<br />

installation & commissioning<br />

picture left unlatching by turning<br />

picture below unlatching by pulling<br />

Running <strong>machine</strong>s for sports <strong>and</strong> fitness with manufacturing date before 12/2007 will be ready for operation immediately<br />

(pay attention to the display). Running <strong>machine</strong>s with manufacturing date after 12/2007 resp. all treadmills for medical<br />

application (independet of manufacturing date) have to be switched on with the white main power on „I“ key (until 2003<br />

green “ON” or “I” key) again. Before switching it on again, however, wait for at least 1 to 2 minutes (3 phase devices: 2<br />

to 3 minutes). As mentioned above „Switching the device off“.<br />

N Use the emergency-stop only if in danger.<br />

N The emergency-stop is not to be used as a normal stop-key.<br />

picture type of stop function interface<br />

communication<br />

Safety Stop =<br />

Emergency Stop<br />

on UserTerminal<br />

(st<strong>and</strong>ard on all<br />

<strong>machine</strong>s with<br />

UserTerminal)<br />

Safety Stop =<br />

Emergency Stop<br />

External <strong>and</strong>/or<br />

additional<br />

emergency stop<br />

mushroom head<br />

type.<br />

N palm switch is hit<br />

N <strong>running</strong> belt & inclination<br />

system stop fast<br />

N <strong>machine</strong> is cut off from main<br />

power of drive motor <strong>and</strong><br />

inclination system<br />

N display, keyboard <strong>and</strong><br />

interface inactive<br />

N mushroom switch is hit<br />

N <strong>running</strong> belt & inclination<br />

system stop fast<br />

N <strong>machine</strong> is cut off from main<br />

power of drive motor <strong>and</strong><br />

inclination system<br />

N display, keyboard <strong>and</strong><br />

interface inactive<br />

communication<br />

interrupted<br />

during cut off<br />

from main<br />

power<br />

communication<br />

interrupted<br />

during cut off<br />

from main<br />

power<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 58 of 213<br />

how to restart<br />

N wait 1 minute<br />

N pull palm switch<br />

N push “power on”<br />

button / main<br />

power<br />

N press START key,<br />

N or new comm<strong>and</strong><br />

through interface<br />

N wait 1 minute<br />

N pull mushroom<br />

switch<br />

N push “power on”<br />

button / main<br />

power<br />

N press START key,<br />

N or new comm<strong>and</strong><br />

through interface


[6.F] Installation, Overview, Checklist<br />

installation & commissioning<br />

The authorized technician (distributor or service partner) has to check whether the installation of the device has been<br />

performed properly, before the customer is introduced into the operation of the h/p/cosmos device. For detailed<br />

information <strong>and</strong> instructions for installation <strong>and</strong> commissioning please refer to all respective individual chapters in this<br />

manual. Among others, following points save to be performed or controlled:<br />

No. Illustration Description<br />

[01]<br />

[02]<br />

[03]<br />

[04]<br />

Compare the delivery note with the installed <strong>running</strong> <strong>machine</strong>.<br />

Are all parts like <strong>running</strong> <strong>machine</strong>, accessories, service box <strong>and</strong><br />

<strong>running</strong> <strong>machine</strong> folder complete <strong>and</strong> without damage.<br />

Adjust the levelling sockets.<br />

Tighten all fixation screws of the h<strong>and</strong>rails.<br />

Check the wall socket: Optical check for damages, burned<br />

contacts, proper fixation <strong>and</strong> good ground contact (no paint, dirt,<br />

deformation)<br />

Check out the correct pin assignment of the wall socket with a<br />

socket-tester with direct indication (see picture left).<br />

As testing-device we recommend the Testavit Schuki 3 –<br />

[cos15900] - Separate instruction available at h/p/cosmos.<br />

Check the electrical connection of the device: Direct connection to<br />

the wall socket with separate fuse, no danger of stumbling, no<br />

extension cable, no multiple plug socket, separate circuit for the<br />

treadmill. If a correct connection to the wall socket is not possible,<br />

note this on the delivery note, inform the customer <strong>and</strong> put the<br />

device out of operation (if necessary) <strong>and</strong> safe<br />

accomplished �<br />

against start-up.<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 59 of 213<br />

accomplished �<br />

accomplished �<br />

accomplished �


[05]<br />

[06]<br />

[07]<br />

[08]<br />

[09]<br />

installation & commissioning<br />

Immediately at first installation an electric safety test / measurment<br />

has to be performed for „Protective Earth Resistance“, „Isolation<br />

Resistance“ <strong>and</strong> „Leakage Current“ <strong>and</strong> the values have to be<br />

recorded on a special protocol [cos11690xx] <strong>and</strong> marked with „first<br />

measured values“. For details please refer to chapter<br />

„maintenance / inspections“. One copy of this protocol<br />

[cos11690xx] remains at the owner’s manual <strong>and</strong> the original with<br />

the „first measured values“ shall be sent to h/p/cosmos. The<br />

potential equalization cable has to be<br />

accomplished �<br />

connected AFTER the measurements.<br />

Lubricate the <strong>running</strong> <strong>machine</strong> before introduction according to<br />

instruction with 30 ml special silicon oil which is part of the “service<br />

box”.<br />

Check the tension of the <strong>running</strong> belt (no slip).<br />

Adjust the <strong>running</strong> belt properly before instruction with the<br />

customer.<br />

The allen-key must be pulled out of the screw<br />

immediately after usage due to risk of injury!<br />

Test the function of the heart rate transmission with the help of a<br />

simulator or a POLAR sender. Search for interferences <strong>and</strong><br />

possible reasons.<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 60 of 213<br />

accomplished �<br />

accomplished �<br />

accomplished �<br />

accomplished �


[7.]Operation<br />

[7.A] UserTerminal, keyboard <strong>and</strong> display<br />

If your <strong>running</strong> <strong>machine</strong> has no UserTerminal (no display, no keyboard), it can only<br />

be controlled via interface RS 232, e.g. via ECG, Ergospirometry, PC with software<br />

h/p/cosmos para graphics ® or h/p/cosmos para control ® . Important security advices,<br />

warnings, system requirements <strong>and</strong> a list of the compatible peripheric devices you<br />

can find in this manual under RS232 interface <strong>and</strong> option functions RS232 protocol<br />

or at our homepage www.coscom.org<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 61 of 213<br />

operation<br />

More recent computers will have only an USB interface instead of the RS232 interface. For that case an „USB to RS232<br />

interface-adapter-cable“ is available at h/p/cosmos under order number cos12769. To control via USB interface the<br />

processor must be a Pentium 1.8 GHz or higher.<br />

To use all the functions described below <strong>and</strong> for maintenance <strong>and</strong> service we recommend the PC software h/p/cosmos<br />

para control ® (Freeware). It is possible to purchase a keyboard or an external UserTerminal as optional equipment for<br />

these models, which can be connected to the <strong>running</strong> <strong>machine</strong> via interface RS232. Option 1: UserTerminal MCU4<br />

remote control with 5 meter cable: h/p/cosmos order number cos10002. Option 2: additional keyboard 6 keys with 2<br />

meter cable: h/p/cosmos order number [cos10106]. When using an additional keyboard it is necessary (depending on<br />

the model) to install a additional wiring <strong>and</strong> a connector, or the UserTerminal backplate has to be re-fitted/removed.<br />

At the suspicion of unauthorized access or other reasons to lock the <strong>running</strong> <strong>machine</strong>, it has to be<br />

locked for starting: See „option 40“ in chapter „option settings“. With option 41 ... 44 you can lock<br />

also separate modes (manual, profile, cardio, test).<br />

Oversize <strong>running</strong> <strong>machine</strong>s <strong>and</strong> devices without UserTerminal<br />

The oversize <strong>running</strong> <strong>machine</strong>s series<br />

h/p/cosmos venus <strong>and</strong> h/p/cosmos<br />

saturn have an external control unit<br />

with integrated UserTerminal.<br />

It is possible to re-fit lt-models <strong>and</strong> to<br />

equpip oversize <strong>running</strong> <strong>machine</strong>s with<br />

an additional UserTerminal at the<br />

h<strong>and</strong>rail.<br />

Attachment for retrofitting UserTerminal at h<strong>and</strong>rail [cos13514]<br />

Monitor arm (movable-) for PC monitors or laptops [cos13321]<br />

Laptop computer [cos13476]<br />

Roll set (4 rolls) for external control unit [cos14184]<br />

pic. top: Retrofitting UserTerminal at h<strong>and</strong>rail<br />

pic. left: External UserTerminal control unit<br />

with options: rolls set <strong>and</strong> monitor-/laptop arm


[7.A1] The keyboard<br />

Basic functions<br />

Alteration of elevation<br />

Program steps<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 62 of 213<br />

operation<br />

Do not lean on the UserTerminal. Do not put any pressure on the display. Press the keys softly. As<br />

confirmation you will hear a beeping sound.<br />

Modus-Übernahme Basic functions (Manual, Profile, Cardio, Test)<br />

Selection of mode (manual, profile, cardio, test)<br />

Alteration of parameters (speed, acceleration / deceleration)<br />

Selection of options, program parameter<br />

The keys may also have special functions within different modes. See the instructions manual mode, profile mode, cardio<br />

mode <strong>and</strong> test mode<br />

[7.A2] The display<br />

The display consists of 6 four digit LCD-displays, which show speed, exercise period, (energy consumption,<br />

wattage <strong>and</strong> MET by turns), exercise distance, elevation (inclination) <strong>and</strong> heart rate. Next to the LCD-displays you<br />

will find light-emitting diodes (LED) in addition, which supply information about mode, measuring unit <strong>and</strong> so on.<br />

Flashing displays indicate:<br />

N select option (LED �)<br />

N alterable parameter (LCD)<br />

N interference / error<br />

Basic functions<br />

Stop<br />

Discontinuance of operation<br />

Basic functions<br />

Confirmation of mode (manual, profile, cardio, test)<br />

Confirmation of parameter (speed, time, pulse, a.s.)<br />

Confirmation of options (measuring unit)<br />

Start (of selected mode)<br />

Modification of acceleration / deceleration


The indicated values remain, after stopping the <strong>running</strong> <strong>machine</strong>, until ...<br />

... the <strong>running</strong> <strong>machine</strong> has been started again with the c -key<br />

... the mode (manual, profile, cardio, test) has been changed<br />

... the display has been initiated by pressing the b -key once more.<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 63 of 213<br />

operation<br />

If the <strong>running</strong> <strong>machine</strong> has been stopped by reducing the speed with the a -key („PAUSE“-position), then the<br />

display continues with the previous values after starting again.<br />

[7.A3] The display (st<strong>and</strong>ard configuration)<br />

Speed<br />

LED Status Unit Definition<br />

max �<br />

m/min � m/min 0.1<br />

km/h � km/h 0.01<br />

m/s � m/s 0.01<br />

mph � mph 0.01<br />

Running distance<br />

LED Status Unit Definition<br />

m � m 0.1<br />

km � km 0.1<br />

miles � miles 0.1<br />

Elevation<br />

Program sequence (Step) / -number (No)<br />

LED Status Unit Definition<br />

% � % 0.1<br />

∠ � ° 0.1<br />

Step �<br />

1<br />

No �<br />

1<br />

� LED on / � LED off<br />

Display LED Status Unit Definition<br />

MET *<br />

MET � MET 0.1<br />

Energy consumption Energy � KJ 1<br />

Wattage<br />

Power � Watt 1<br />

Fitness-Index (UKK)<br />

Index 1<br />

* One metabolic equivalent (MET) is defined as the consumption of oxygen<br />

while sedentary life. Two MET correspond to the consumption while walking<br />

with ca. 3.5 km/h <strong>and</strong> 8 MET correspond to the consumption while <strong>running</strong><br />

with ca. 13 km/h. Displayed not for all models.<br />

Runing time<br />

LED Status Unit Definition<br />

mm:ss<br />

hh:mm<br />

Display LED Status Unit Definition<br />

Heart rate (max.) � � 1/min 1<br />

Heart rate (min.) � � 1/min 1<br />

Sex<br />

Sex<br />

M = male<br />

F = female<br />

� M/F<br />

Age<br />

Age � year 1<br />

Weight<br />

Weight � kg 1<br />

Height of subject H ....<br />

cm 1<br />

1<br />

1


[7.B] Mode of operation<br />

[7.B1] General notes<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 64 of 213<br />

operation<br />

The <strong>running</strong> <strong>machine</strong> is equipped with 4 modes of operation: manual, profile, cardio <strong>and</strong> test. Some of the modes<br />

of operation can also be changed online into another while the <strong>running</strong> <strong>machine</strong> is in operation.<br />

Select the Mode first (see 4 LED) with a or d <strong>and</strong> confirm the Mode with c<br />

The incorporated series interface is always active. This means, that you can send <strong>and</strong> receive data/comm<strong>and</strong>s at<br />

any time (parallel) <strong>and</strong> during any mode. Always the latest comm<strong>and</strong> will be executed. No matter if the comm<strong>and</strong><br />

came via interface or from the UserTerminal during one of the 4 modes.<br />

[7.B2] Acceleration levels / acceleration intensity<br />

7 different acceleration levels <strong>and</strong> deceleration levels are available for all modes <strong>and</strong> for remote control via<br />

interface <strong>and</strong> h/p/cosmos coscom protocol. Max. acceleration level <strong>and</strong> minimum acceleration level see chapter:<br />

User options / optional functions. The following diagram shows the acceleration levels at a <strong>running</strong> <strong>machine</strong> with<br />

maximum speed of 40km/h. The acceleration levels are programmable for every single program step of the<br />

individual profile <strong>and</strong> tests. The acceleration levels are accessible via interface <strong>and</strong> optional PC software<br />

h/p/cosmos para control <strong>and</strong> h/p/cosmos para graphics.<br />

Explanation:<br />

[a] Speed in km/h,<br />

[b] Time in seconds 0 to maximum<br />

speed: Intensity<br />

[1] 131 sec., [2] 66 sec., [3] 33 sec.,<br />

[4] 16 sec., [5] 8 sec., [6] 5 sec. <strong>and</strong><br />

[7] 3 sec.<br />

Slow acceleration (level 1 or 2) is for<br />

rehabilitation <strong>and</strong> for fitness for<br />

beginners. Medium acceleration (level 3<br />

or 4) is for well trained subjects for<br />

speeding up or slowing down pretty fast.<br />

High acceleration (level 5 to 7) is only for<br />

top athletes for sprint training <strong>and</strong> other<br />

competitive sports applications.<br />

Therefore the high accelerations are only<br />

accessible via special release in the<br />

optional functions, protected by access<br />

key.<br />

Never set too high loads (speed, acceleration, elevation). High loads are forbidden, if the<br />

health <strong>and</strong> the condition of the test person / patient do not allow them <strong>and</strong> a doctor has not<br />

authorized these loads. Disregard may cause injuries <strong>and</strong> dangerous health problems. At<br />

high loads <strong>and</strong> identifiable risks meet the appropriate higher <strong>and</strong> reasonable safety<br />

precautions.


[7.B3] Manual mode<br />

Initiation: Running belt is not moving. One of the modes LED � is flashing: (manual, profile, cardio, test)<br />

[7.B3] Step no. Activity Keys Response / Display<br />

[01] Select operational mode manual<br />

[02] Start operational mode manual<br />

[03] Alteration of speed<br />

[04] Alteration of acceleration <strong>and</strong><br />

deceleration level for speed<br />

[05] Pause mode<br />

[06] Alteration of elevation<br />

a or d<br />

c<br />

a or d<br />

press <strong>and</strong> hold<br />

a or d<br />

press several times<br />

<strong>and</strong> then hold<br />

a<br />

f or e<br />

press <strong>and</strong> hold<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 65 of 213<br />

operation<br />

Alteration of the mode (manual, profile, cardio, test) until<br />

� manual is flashing<br />

A) In case of deactivation (st<strong>and</strong>ard) in the optional<br />

functions:<br />

� manual, km/h, m <strong>and</strong> % are glowing<br />

� met, energy <strong>and</strong> power are glowing by turns<br />

Running belt accelerates up to the pre-selected startingspeed<br />

(st<strong>and</strong>ard: 0.5 km/h, can be changed in the<br />

optional mode), start of measuring<br />

% indicates present heart rate: P. 40 ... P. 220. The<br />

display indicates every single heart-beat with a flashing<br />

dot behind the P<br />

B) In case of activation in the optional functions:<br />

% indicates: 65 flashing (for 65 kg body weight)<br />

� weight is glowing<br />

in order to enter with a or d <strong>and</strong> then c the body<br />

weight for an accurate calculation of the POWER <strong>and</strong><br />

ENERGY consumption.<br />

Further steps see A)<br />

Speed is being increased / reduced (0-max.) as long as<br />

the key get pressed. If Speed is being reduced down to<br />

„0“= Pause-position:<br />

= indicates: PAUS. Current values will be<br />

sustained.<br />

Example: For acceleration or deceleration level “3”,<br />

press 3 times the corresponding key <strong>and</strong> then hold.<br />

Note: Maximum 7 levels, limited by the maximum<br />

acceleration level selected in the Optional Mode /<br />

st<strong>and</strong>ard value = 4.<br />

If Speed is being reduced down to „0“ = Pause position<br />

= indicates: "PAUS". Running belt stops. All<br />

displays stop. Current values will be sustained.<br />

When restart with d or c all display values will be<br />

"added".<br />

Elevation is being increased / reduced (0 ... max.) as<br />

long as the key get pressed.


[7.B3] [07] Switch over to the next mode or<br />

one mode back<br />

[08] Stop <strong>running</strong> belt<br />

[7.B4] Automatic mode<br />

c<br />

<strong>and</strong> simultaneous<br />

a or d<br />

b<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 66 of 213<br />

operation<br />

Switch (Jump) over to the next Mode, or jump to the<br />

previous Mode. This is a perfect way to change to other<br />

Modes without stop of the belt.<br />

Running-Belt stops.<br />

(Deceleration time can be adjusted in the optional<br />

functions)<br />

Display values <strong>and</strong> pre-selected mode remain for 2<br />

minutes <strong>and</strong> will then be deleted (reset) automatically<br />

The automatic mode is for starting a training profile, a heart rate dependant speed control or a test profile.<br />

Caution: The treadmill belt will start moving automatically <strong>and</strong> the speed <strong>and</strong> elevation will be changed<br />

automatically after one of the 3 modes (profile, cardio, test) has been started. You <strong>and</strong> all other users need to get<br />

familiar with the details <strong>and</strong> risks of these modes (e.g. the max. speed <strong>and</strong> elevation of these modes) for not facing<br />

the danger of too high speeds <strong>and</strong> injuries.<br />

Additionally there is the possibility to disable each mode (manual, profile, cardio, test) in the user-option-mode (see<br />

chapter “optional settings”. This can be required in special cases for additional safety or to reduce functions of<br />

more simple operation.<br />

N Automatic operation (mode profiles, cardio, test, remote control via PC <strong>and</strong> peripheric<br />

devices) is forbidden, if the health <strong>and</strong> the condition of the test person / patient do not<br />

allow them <strong>and</strong> a doctor has not authorized these loads. Disregard may cause injuries <strong>and</strong><br />

dangerous health problems.<br />

N At automatic operation the test person / patient <strong>and</strong> supervisory staff have to impropriate<br />

exact knowledge about the expected loads before starting <strong>and</strong> anticipate an automatic<br />

load alternation (speed, acceleration, decelaration, elevation <strong>and</strong> STOP) at any time. It is<br />

necessary to meet the appropriate higher <strong>and</strong> reasonable safety precautions.<br />

[7.B5] Profile Mode / Pre-programmed training profiles<br />

Profile-mode offers 6 different programs, which, with the help of various speed <strong>and</strong> elevation combinations,<br />

simulates easy <strong>running</strong> training as well as a cross-country run.<br />

In case of activation (st<strong>and</strong>ard deactivated) in the user-options (OP11): The maximum speed, maximum elevation<br />

<strong>and</strong> the duration of the 6 profiles can be selected ("scaled"), in order to provide 6 x 6 x 6 (216 total) variations of<br />

profiles. St<strong>and</strong>ard set-up is without scaling, for quicker <strong>and</strong> easier access.<br />

All of the 6 profiles are predefined <strong>and</strong> cannot be changed in the memory. For creating your personal profile:<br />

Test No 21 ... 28 are user defined (free programmable Profiles).<br />

For a better <strong>and</strong> easier documentation we recommend to use the serial printer interface or to connect an external<br />

PC in order to use the h/p/cosmos para graphics ® for documentation.


Initiation: Running-Belt is not moving. One of the modes LED � is flashing: (manual, profile, cardio, test)<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 67 of 213<br />

operation<br />

[7.B5] Step no. Activity Keys Response / Display<br />

[01] Select operational mode profile<br />

Alteration of the mode (manual, profile, cardio, test) until<br />

[02] Start operational mode profile<br />

[03] Select profile no. 1 ... 6<br />

[04] Start operational mode profile:<br />

The selected profile no. 1 ... 6<br />

[05] Stop <strong>running</strong>-belt<br />

a or d<br />

c<br />

a or d<br />

c<br />

b<br />

� profile is flashing<br />

� profile is glowing<br />

� speed max. is glowing<br />

� program no. is glowing<br />

= indicates: max. speed in this profile<br />

! indicates: time in this profile<br />

$ indicates: profil no. 1<br />

� profile is glowing<br />

� speed max. is glowing<br />

� program no. is glowing<br />

= indicates: max. speed in this profile<br />

! indicates: time in this profile.<br />

$ indicates: profil no. 1 ... 6<br />

A) In case of deactivation (st<strong>and</strong>ard) in the optional<br />

functions:<br />

� profile is glowing<br />

� % is glowing, by turns with LED step<br />

� KJ <strong>and</strong> Watt are glowing by turns<br />

Running belt accelerates up to the starting-speed of the<br />

first program step, according to the "table of programs".<br />

5 seconds before the next program step there is an<br />

acoustic countdown. Start of measuring. All displays<br />

indicate the actual values. The belt stops automatically<br />

according to the profile duration. The belt can be<br />

stopped at any time with the help of b.<br />

B) In case of activation in the optional functions:<br />

" indicates: SC 3 (scaling 1 ... 6) is flashing<br />

= indicates max. speed according to the scaling (for<br />

max. speed in this profile)<br />

� speed max. is glowing<br />

! indicates time according to the scaling (for<br />

duration of this Profile)<br />

In order to enter with a or d <strong>and</strong> then c das the<br />

scaling (max. Speed, Elevation, Time <strong>and</strong> Distance).<br />

Herewith you can create more variations out of the 6<br />

st<strong>and</strong>ard Profiles. Continue with A)<br />

Running belt stops.<br />

(Deceleration time can be adjusted in the optional<br />

functions)


[7.B6] Possibilities of interfering into the profile mode manually<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 68 of 213<br />

operation<br />

A Profile can be changed on-line, but not in the memory location. For personal profiles use test-profiles no. 21 – 24<br />

Initiation: Operational mode: Profile, <strong>running</strong>-belt is moving.<br />

[7.B6] Step no. Activity Keys Response / Display<br />

[01] Alteration of speed<br />

a or d<br />

Speed is being increased / reduced (0-max.) as long as<br />

the key get pressed. If SPEED is being reduced down to<br />

press <strong>and</strong> hold „0“ = PAUSE-position. Current values will be sustained.<br />

[02] Alteration of acceleration <strong>and</strong><br />

deceleration of speed<br />

[03] Pause Mode = Interfere a<br />

program step<br />

[04] Alteration of elevation<br />

[05] Switch over to the next program<br />

step or one program step back.<br />

[06] Switch over to the next mode or<br />

one mode<br />

c<br />

press several times<br />

<strong>and</strong> then hold<br />

a<br />

f or e<br />

press <strong>and</strong> hold<br />

c<br />

<strong>and</strong> simultaneous<br />

e or f<br />

c<br />

<strong>and</strong> simultaneous<br />

a or d<br />

Alteration of the speed is only online for this single step.<br />

The speed can not be changed in the memory. Next<br />

program step as in the profile definied.<br />

Example: For acceleration or deceleration level “3”,<br />

press 3 times the corresponding key <strong>and</strong> then hold.<br />

Note: Maximum 7 levels, limited by the maximum<br />

acceleration level selected in the Optional Mode /<br />

st<strong>and</strong>ard value = 4.<br />

= indicates: "PAUS". Running belt stops. All<br />

displays stop. Current values will be sustained.<br />

When restart with d or c all display values will be<br />

added.<br />

Elevation is being increased / reduced (0-max.) as long<br />

as the key get pressed.<br />

Alteration of the elevation is only online for this single<br />

step. The elevation can not be changed in the memory.<br />

Next program step as in the profile definied.<br />

Jump over to the next program step, or jump to the<br />

previous program step.<br />

Jump over to the next mode, or jump to the previous<br />

Mode. This is a perfect way to change to other modes<br />

without stop of the belt. For example to start manual<br />

mode for a "cool-down" period without stop of the belt<br />

For a better documentation we recommend to use a printer linked to the serial interface or/<strong>and</strong> to use an external PC with the<br />

Software h/p/cosmos para graphics ® .


[7.B7] Profile outline<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 69 of 213<br />

operation<br />

Profile 1-3: time dependent, without elevation<br />

Profile 4-6: time dependent, with elevation<br />

As an optional feature with some models on memory location 30...99 additional program profiles may be included.<br />

The following chart contains the st<strong>and</strong>ard values, if the profiles have not been scaled. Activation of the scaling<br />

function see chapter “optional functions”.<br />

[7.B7] profile 1<br />

Endurance interval for beginners = m/sec = km/h ! min $ %<br />

Warming up 1.8 6.5 4.0 0.0<br />

High 1 2.5 9.0 0.5 0.0<br />

Low 1 1.8 6.5 3.0 0.0<br />

High 2 2.5 9.0 0.5 0.0<br />

Low 2 1.8 6.5 3.0 0.0<br />

High 3 2.5 9.0 0.5 0.0<br />

Low 3 1.8 6.5 3.0 0.0<br />

High 4 2.5 9.0 0.5 0.0<br />

Low 4 1.8 6.5 3.0 0.0<br />

High 5 2.5 9.0 0.5 0.0<br />

Low 5 1.8 6.5 3.0 0.0<br />

High 6 2.5 9.0 0.5 0.0<br />

Low 6 1.8 6.5 3.0 0.0<br />

profile 2<br />

25 min<br />

Endurance interval st<strong>and</strong>ard = m/sec = km/h ! min $ %<br />

Warming up 2.0 7.2 5.0 0.0<br />

High 1 2.5 9.0 3.0 0.0<br />

Low 1 2.0 7.2 2.0 0.0<br />

High 2 2.5 9.0 3.0 0.0<br />

Low 2 2.0 7.2 2.0 0.0<br />

High 3 2.5 9.0 3.0 0.0<br />

Low 3 2.0 7.2 2.0 0.0<br />

High 4 2.5 9.0 3.0 0.0<br />

Low 4 2.0 7.2 2.0<br />

25 min<br />

0.0


[7.B7] profile 3<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 70 of 213<br />

operation<br />

Progressive intervall = m/sec = km/h ! min $ %<br />

Warming up 2.8 10.1 4.0 0.0<br />

High 1 3.2 11.5 2.0 0.0<br />

Low 1 2.8 10.1 2.0 0.0<br />

High 2 3.6 13.0 1.0 0.0<br />

Low 2 2.8 10.1 2.0 0.0<br />

High 3 4.0 14.4 1.0 0.0<br />

Low 3 2.8 10.1 2.0 0.0<br />

High 4 3.6 13.0 1.0 0.0<br />

Low 4 2.8 10.1 2.0 0.0<br />

High 5 3.2 11.5 1.0 0.0<br />

Low 5 2.8 10.1 2.0 0.0<br />

High 6 3.2 11.5 1.0 0.0<br />

Low 6 2.8 10.1 4.0 0.0<br />

profile 4<br />

25 min<br />

Endurance intervall for beginners = m/sec = km/h ! min $ %<br />

Warming up 1.8 6.5 4.0 0.0<br />

High 1 2.5 9.0 0.5 5.0<br />

Low 1 1.8 6.5 3.0 0.0<br />

High 2 2.5 9.0 0.5 10.0<br />

Low 2 1.8 6.5 3.0 0.0<br />

High 3 2.5 9.0 0.5 10.0<br />

Low 3 1.8 6.5 3.0 0.0<br />

High 4 2.5 9.0 0.5 10.0<br />

Low 4 1.8 6.5 3.0 0.0<br />

High 5 2.5 9.0 0.5 10.0<br />

Low 5 1.8 6.5 3.0 0.0<br />

High 6 2.5 9.0 0.5 10.0<br />

Low 6 1.8 6.5 3.0<br />

25 min<br />

0.0


[7.B7] profile 5<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 71 of 213<br />

operation<br />

Endurance interval st<strong>and</strong>ard = m/sec = km/h ! min $ %<br />

Warming up 2.0 7.2 5.0 5.0<br />

High 1 2.5 9.0 3.0 0.0<br />

Low 1 2.0 7.2 2.0 10.0<br />

High 2 2.5 9.0 3.0 0.0<br />

Low 2 2.0 7.2 2.0 10.0<br />

High 3 2.5 9.0 3.0 0.0<br />

Low 3 2.0 7.2 2.0 10.0<br />

High 4 2.5 9.0 3.0 0.0<br />

Low 4 2.0 7.2 2.0<br />

25 min<br />

10.0<br />

profile 6<br />

Progressive intervall / elevation = m/sec = km/h ! min $ %<br />

Warming up 2.8 10.1 4.0 0.0<br />

High 1 3.2 11.5 2.0 10.0<br />

Low 1 2.8 10.1 2.0 0.0<br />

High 2 3.6 13.0 1.0 7.5<br />

Low 2 2.8 10.1 2.0 0.0<br />

High 3 4.0 14.4 1.0 5.0<br />

Low 3 2.8 10.1 2.0 0.0<br />

High 4 3.6 13.0 1.0 7.5<br />

Low 4 2.8 10.1 2.0 0.0<br />

High 5 3.2 11.5 1.0 5.0<br />

Low 5 2.8 10.1 2.0 0.0<br />

High 6 3.2 11.5 1.0 5.0<br />

Low 6 2.8 10.1 4.0<br />

25 min<br />

0.0


[7.B8] Cardio mode – general notes<br />

Heart rate controlled work load („pulse-control“ cardio-training)<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 72 of 213<br />

operation<br />

N Pay attention to all safety instructions <strong>and</strong> warnings in the instruction manual of this<br />

<strong>machine</strong> <strong>and</strong> of all accessories <strong>and</strong> options involved.<br />

N Heart rate monitoring systems can be inaccurate.<br />

N Overtraining can lead to serious injury or death.<br />

N If you feel close to fainting, the training is to be stopped immediately.<br />

N Stop training immediately if feeling sick or dizzy, pain or in case of any other health<br />

problem <strong>and</strong> see your doctor.<br />

N In the case of detected or suspected interference of the wireless heart rate transmission<br />

do not use the cardio mode.<br />

N The operation of the cardio mode is forbidden, if the health <strong>and</strong> the condition of the test<br />

person / patient do not allow it <strong>and</strong> a doctor has not authorized these loads. Disregard may<br />

cause injuries <strong>and</strong> dangerous health problems.<br />

N At automatic <strong>and</strong> cardio operation the test person / patient <strong>and</strong> supervisory staff have to<br />

impropriate exact knowledge about the expected loads before starting <strong>and</strong> anticipate an<br />

automatic load alternation (speed, acceleration, decelaration, elevation <strong>and</strong> STOP) at any<br />

time. It is necessary to meet the appropriate higher <strong>and</strong> reasonable safety precautions.<br />

[7.B9] Cardio mode - Field of application / control<br />

The Cardio Mode controls the <strong>running</strong> <strong>machine</strong> so that the heart rate of the subject stays within a predefined zone.<br />

Therefore the subject has to wear a corresponding chest belt.<br />

Adjust the belt length so that the belt fits<br />

tightly but does not confine you. The belt<br />

should not loosen while exercising. Close the<br />

belt with transmitter placed outwards<br />

(POLAR-Logo in right position).<br />

In order to allow an optimal skin contact the<br />

skin should be moist. Contact gel, as used for<br />

ECG, is an excellent solution. Moisten the<br />

two electrodes <strong>and</strong> the skin with water or<br />

contact gel, which is available at chemistries.<br />

Place the transmitter so that it is right below the pectoral muscle (chest), as is shown in the illustration above. For<br />

ca. 95 ... 98 % of the subjects the "normal" placement of the transmitter belt is recommended.


file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 73 of 213<br />

operation<br />

The following parameters can be programmed:<br />

N The desired max. heart rate (pulse) during the training<br />

N The desired min. heart rate (pulse) during the training<br />

N The max. allowed speed for the cardio mode. (Elevation can be set manually online during the exercise)<br />

The <strong>running</strong> <strong>machine</strong> starts with the starting speed. In order to keep the heart rate within the pre-selected heart<br />

rate zone, the <strong>machine</strong> increases or decreases the speed automatically if necessary.<br />

The speed will not exceed the max. defined speed. If the max. speed has been obtained <strong>and</strong> the heart rate of the<br />

subject is still below the target zone, the elevation will be increased to achieve the right target hart rate of the<br />

subject. On the other h<strong>and</strong>, if the max. Heart-Rate has been reached in order to give the subject relief, first the<br />

elevation <strong>and</strong> then the speed is going to be reduced.<br />

If the <strong>running</strong> <strong>machine</strong> does not receive a heart rate signal when starting the cardio mode it will start with the set<br />

starting speed <strong>and</strong> does not alter this speed. Every 30 seconds a warning signal is audible, after 1 minute the<br />

<strong>running</strong> <strong>machine</strong> cancels the mode (at MCU5 firmware-versions < V1.03.1 or MCU4 firmware-versions < V4.04.2<br />

the <strong>running</strong> <strong>machine</strong> cancels after 2 minutes). If the heart rate signal fails during the operation in the cardio mode<br />

the <strong>running</strong> <strong>machine</strong> will run in the actual set speed <strong>and</strong> angle of inclination. In this case beeper allert signal will be<br />

activated every 30 seconds. The <strong>running</strong> <strong>machine</strong> will run down speed to 2 km/h <strong>and</strong> the mode will be terminated<br />

after 1 minute (at MCU5 firmware-versions < V1.03.1 or MCU4 firmware-versions < V4.04.2 the <strong>running</strong> <strong>machine</strong><br />

chancels after 2 minutes).<br />

max. speed (km/h) Control<br />

2 ... max. Speed control first. After having reached the speed, which is maximally<br />

allowed, the elevation will be controlled.<br />

Examples:<br />

a) Set max. speed 4.0 km/h if you want the subject only to walk <strong>and</strong> to control the load by means of elevation.<br />

b) Set max. speed 20.0 km/h if you want the subject to control the load by means of speed only without elevation.<br />

Chart for heart rate controlled work load at the heart rate lower level<br />

Heart rate difference actual > < set value Speed Elevation Time<br />

0 ... 5 0.2 km/h 0.1 % 25 Seconds<br />

5.1 ... 15 0.4 km/h 0.2 % 25 Seconds<br />

15.1 ... 30 0.6 km/h 0.4 % 25 Seconds<br />

30.1 ... 50 0.8 km/h 0.8 % 20 Seconds<br />

> 50 1.0 km/h 1.0 % 20 Seconds<br />

Chart for heart rate controlled work load at the heart rate upper level<br />

Heart rate difference actual > < set value Speed Elevation Time<br />

0 ... 5 0.3 km/h 0.3 % 12 Seconds<br />

5.1 ... 15 0.8 km/h 0.8 % 12 Seconds<br />

15.1 ... 30 1.0 km/h 1.0 % 10 Seconds<br />

30.1 ... 50 1.5 km/h 1.2 % 8 Seconds<br />

> 50 2.0 km/h 1.6 % 7 Seconds


[7.B10] Cardio mode – Operation<br />

[7.B10]<br />

The Cardio-Training cannot be performed without a POLAR chest belt <strong>and</strong> transmitter.<br />

Initiation: Running-Belt is not moving. One of the modes LED � is flashing: (manual, profile, cardio, test)<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 74 of 213<br />

operation<br />

Step no. Activity Keys Response / Display<br />

[01] Select operational mode cardio<br />

Alteration of the mode (manual, profile, cardio, test) until<br />

[02] Confirm operational cardio<br />

mode<br />

a or d<br />

c<br />

[03] Change max. speed for Cardio<br />

mode a or d<br />

[04] Confirm max. speed for cardio<br />

mode<br />

[05] Select personal age<br />

c<br />

a or d<br />

� cardio is flashing<br />

� cardio is glowing<br />

� speed max. is glowing<br />

= indicates 6.0 is (for max. speed during cardio<br />

mode). This default value (here 6.0) can be changed in<br />

the optional functions<br />

� % is glowing<br />

� energy is glowing<br />

= indicates flashing: 2.0 ... max. speed of the<br />

<strong>running</strong> <strong>machine</strong> (for max. allowed speed during cardio<br />

mode)<br />

Examples:<br />

a) Set max. Speed 4.0 if you want the subject only to<br />

walk <strong>and</strong> to control the load by means of elevation.<br />

b) Set max. Speed 20.0 if you want the subject to control<br />

the load by means of speed only without elevation.<br />

� cardio is glowing<br />

� speed max. is glowing<br />

= indicates 2.0 ... max. speed of the <strong>machine</strong> (for<br />

max. allowed speed during cardio mode)<br />

� years is glowing<br />

% indicates flashing: 35 (for the subject's age)<br />

� years is glowing<br />

% indicates flashing: 0 ... 100(for the age), from<br />

Firmware V3.02.4: 18 ... 100


[7.B10] [06] Confirm personal age<br />

[07] Change upper level of heart rate<br />

[08] Confirm upper level of heart rate<br />

[09] Change lower level of heart rate<br />

[10] Confirm lower level of heart rate<br />

<strong>and</strong> start of the <strong>running</strong>-belt<br />

[11] Stop <strong>running</strong> belt<br />

c<br />

a or d<br />

c<br />

a or d<br />

c<br />

b<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 75 of 213<br />

operation<br />

� � is glowing<br />

% indicates flashing the suggested max. level of the<br />

heart rate. Calculated with the formula: 180 minus age<br />

Important: According to your health condition <strong>and</strong><br />

according to the advises of your medical doctor you<br />

can/must change this level. If no changes required<br />

confirm the value with c<br />

� � is glowing<br />

% indicates flashing max. level of the heart rate<br />

� � is glowing<br />

% indicates flashing the suggested min. level of the<br />

heart rate. Calculated with the formula: max. level of<br />

heart rate minus 10<br />

� � is glowing<br />

% flashing min. level of the heart rate<br />

Running belt starts from 0 up to a speed of 2 km/h<br />

� cardio is glowing<br />

� or � is glowing, indicating that the present heart rate<br />

is too low or too high<br />

% indicates present Heart Rate: P. 40 ... P. 220. The<br />

display indicates every single heartbeat with a flashing<br />

dot behind the “P”. Speed <strong>and</strong> elevation of the <strong>running</strong><br />

<strong>machine</strong> is controlled automatically (see separate table).<br />

Running belt stops.<br />

(Deceleration time can be adjusted in the optional<br />

functions)<br />

For a better documentation we recommend to use a printer linked to the serial interface RS232 or/<strong>and</strong> to use an external PC<br />

with the Software h/p/cosmos para graphics ® .


[7.B11] Possibilities of interfering with the Cardio Mode manually<br />

Initiation: Operational mode: Cardio. Running belt is moving.<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 76 of 213<br />

operation<br />

[7.B11] Step no. Activity Keys Response / Display<br />

[01] Alteration of speed<br />

a or d<br />

= indicates 0 ... max. speed of the <strong>machine</strong><br />

(for max. allowed speed during cardio mode) as long as<br />

press <strong>and</strong> hold the key get pressed.<br />

Alteration of the Speed<br />

automatically changes the<br />

allowed max. speed during<br />

Cardio Mode<br />

[02] Pause mode -<br />

Interfere the cardio mode<br />

[03] Alteration of elevation<br />

[04] Changing the max. level of heart<br />

rate<br />

[05] Switch over to the next mode or<br />

one mode back<br />

a<br />

f or e<br />

press <strong>and</strong> hold<br />

c<br />

<strong>and</strong> simultaneous<br />

e or f<br />

c<br />

<strong>and</strong> simultaneous<br />

a or d<br />

Examples:<br />

a) Set max. Speed 4.0 if you want the subject only to<br />

walk <strong>and</strong> to control the load by means of elevation.<br />

b) Set max. Speed 20.0 if you want the subject to control<br />

the load by means of speed only without elevation.<br />

N Speed is being increased / reduced (0-max.).<br />

N If SPEED is being reduced down to „0“ = PAUSEposition:<br />

= indicates: PAUS. Current values will be<br />

sustained.<br />

= indicates: "PAUS". Running belt stops. All<br />

displays stop. Current values will be sustained.<br />

When restart with d or c all display values will be<br />

"added".<br />

Elevation is being increased / reduced (0-max.) as long<br />

as the key get pressed.<br />

Important: Alteration of the elevation can have impact on<br />

the heart rate of the subject <strong>and</strong> therefore can result in<br />

compensation via automatic speed control.<br />

� � is glowing<br />

% indicates the max. � level of the heart rate /<br />

the min. � level of the heart rate is automatically<br />

changed within the selected range.<br />

Jump over to the next mode to the previous mode. This<br />

is a perfect way to change to other modes without stop<br />

of the <strong>running</strong> belt. For example to start manual mode<br />

for a "cool-down" period without stop of the belt.


[7.B12] Test mode <strong>and</strong> free program profiles – General notes<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 77 of 213<br />

operation<br />

N Some test profiles (e.g. CONCONI-test, STEP-test, COOPER-test, etc.) are endurance tests (max. load <strong>and</strong><br />

max. heart rate tests) <strong>and</strong> should only be performed after consultation of a medical doctor <strong>and</strong> under<br />

supervision of trained staff.<br />

N Proper warm-up <strong>and</strong> cool-down periods have to be considered.<br />

N Use a safety harness with a safe chest belt system (optional equipment) for the subject when making max.<br />

load <strong>and</strong> max. heart rate tests.<br />

[7.B13] Test profiles / Field of application<br />

Test mode can perform several different (pre-defined <strong>and</strong> self defined) tests. Information about the tests can be<br />

found in the enclosure. The UKK walk test (No 1) is the only test where the <strong>running</strong> <strong>machine</strong> makes a calculation<br />

<strong>and</strong> an evaluation (Fitness Index). All other test profiles only make the load control. There is no automatic<br />

evaluation done by the <strong>running</strong> <strong>machine</strong>. The evaluation has to be done by host equipment (e.g. ECG,<br />

Ergospirometry, etc.) or external PC Software (e.g. POLAR Analysis Software).<br />

[7.B13] Test no. Description test / profil Comment / Programing<br />

1 UKK 2-km walk test<br />

(UKK-Institute Tampere/Finl<strong>and</strong>)<br />

2 Graded test<br />

Fitness test with evaluation <strong>and</strong> display of fitness index. Walk<br />

for 2km as fast as you can. After 2km the fitness index is<br />

indicated.<br />

Endurance test (max. Load Test) with the parameters<br />

N Start speed<br />

N Step length in min:sec.<br />

N Increment / Step height<br />

N Accelereration level<br />

N Break time in min:sec<br />

(e.g. for determination of anaerobic threshold by means of<br />

taking blood samples during the load-brakes <strong>and</strong> lactate<br />

analysis with external lactate analysing equipment.<br />

St<strong>and</strong>ard load profile:<br />

N Start speed: 8.0 km/h, must be changed according the<br />

conditions of the subject<br />

N Step length: 3 min. (can be changed)<br />

N Increment: 2.0 km/h (can be changed)<br />

N Acceleration level: 4 (adjustable from 1 to 5)<br />

N Break time: 30 sec. (can be changed)<br />

STOP must be activated manually by the medical doctor.<br />

Continuation next page ...


[7.B13] 2 Graded test Continuation ...<br />

3 Conconi test<br />

e.g. for determination of anaerobic threshold via<br />

heart rate curve of the subject<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 78 of 213<br />

operation<br />

In an already pre-defined step-test it is possible, to set the preprogrammed<br />

pause-times individually during the test, i.e. the<br />

pause can be prolonged or ended immediately.<br />

End pause manually:<br />

Once the test profile has reached the pause mode <strong>and</strong> the<br />

<strong>running</strong> belt has come to halt, you can select the next program<br />

step immediately with START – either with warning sound<br />

signals at the beginning by pressing START once or without by<br />

pressing START twice. This is done independently of the predefined<br />

pause-times <strong>and</strong> has no influence on the following run<br />

of the program, i.e. the following program steps remain<br />

unaffected. By pressing the START button the next speed<br />

step scheduled is started. The pause can also be ended with<br />

the „+“ button. In this case the speed set depends on the point<br />

of release of the „+“ button.<br />

Prolong pause:<br />

Once the test profile has reached the pause mode <strong>and</strong> the<br />

<strong>running</strong> belt has come to halt, it is possible to freeze pause<br />

with the „-“ button.<br />

= indicates: „PAUS“<br />

To end pause you can either press START or „+“ - see end<br />

pause manually.<br />

Endurance test (max. heart rate test)<br />

St<strong>and</strong>ard load profile:<br />

N Starting speed: 8.0 km/h, must be changed according the<br />

conditions of the subject<br />

N Circuit (lap length): 200m (can be changed)<br />

N Increment: 0.5 km/h (can be changed)<br />

STOP must be activated manually by the medical doctor when<br />

the subject is fully exhausted.


[7.B13] 4 Bruce protocol<br />

e.g. for ECG stress test<br />

5 Naughton protocol<br />

e.g. for ECG stress test<br />

6 Balke protocol<br />

e.g. for ECG stress test<br />

7 Cooper protocol<br />

e.g. for ECG stress test<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 79 of 213<br />

operation<br />

Step Duration Speed Elevation<br />

1 3:00 min 2.7 km/h 10 %<br />

2 4.0 km/h 12 %<br />

3 5.4 km/h 14 %<br />

4 6.7 km/h 16 %<br />

5 8.0 km/h 18 %<br />

6 8.8 km/h 20 %<br />

7 9.6 km/h 22 %<br />

Step Duration Speed Elevation<br />

1 3:00 min 3.0 km/h 0.0 %<br />

2 3.5 %<br />

3 7.0 %<br />

4 10.5 %<br />

5 14.0 %<br />

6 17.5 %<br />

Step Duration Speed Elevation<br />

1 2:00 min 5.0 km/h 2.5 %<br />

2 5.0 %<br />

3 7.5 %<br />

4 10.0 %<br />

5 12.5 %<br />

6 15.0 %<br />

7 17.5 %<br />

8 20.0 %<br />

9 22.5 %<br />

10 * 25.0 %<br />

* <strong>running</strong> <strong>machine</strong>s with sizes 150/50 cm max. 24%<br />

N Start at 5.3 km/h <strong>and</strong> 0% elevation<br />

N after 1 minutes: elevation increase to 2 %<br />

N after another minute the elevation will be increased by<br />

1% every minute<br />

N when elevation is 25%: elevation stays constant <strong>and</strong> the<br />

speed is going to be increased by 0.32 km/h every minute<br />

STOP must be activated manually by the medical doctor when<br />

the subject is fully exhausted


[7.B13] 8 Ellestad A Protokoll<br />

e.g. for ECG stress test<br />

9 Ellestad B Protokoll<br />

e.g. for ECG stress test<br />

10 Ramp profile<br />

(not for all models available)<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 80 of 213<br />

operation<br />

Step Duration Speed Elevation<br />

1 3:00 min 2.7 km/h 10.0 %<br />

2 4.8 km/h<br />

3 6.4 km/h<br />

4 8.0 km/h<br />

Step Duration Speed Elevation<br />

1 3:00 min 2.7 km/h 10.0 %<br />

2 4.8 km/h 10.0 %<br />

3 6.4 km/h 10.0 %<br />

4 8.0 km/h 10.0 %<br />

5 8.0 km/h 15.0 %<br />

6 9.6 km/h 15.0 %<br />

Ramp profile with 2 parameters:<br />

N Target speed: St<strong>and</strong>ard: 10.0 km/h; adjustable from 0 to<br />

maximum speed of the treadmill.<br />

N Time for reaching target speed in seconds: St<strong>and</strong>ard:<br />

10 seconds; adjustable from 0 to 99 seconds


[7.B13] 11<br />

The „Gardner Testprotocol“<br />

for application Angiology.<br />

The Gardner Testprotocol serves for evaluation of<br />

the maximum walking distance of Peripheral<br />

Arterial Disease Patients with Intermittent<br />

Claudication.<br />

The test has to be performed under permanent<br />

supervision of a doctor <strong>and</strong> a fall prevention<br />

system (e.g. safety arch with chest belt) has to be<br />

used.<br />

The patient first st<strong>and</strong>s on the side-footboards of<br />

the <strong>running</strong> <strong>machine</strong> <strong>and</strong> not on the belt. Start test<br />

profile 11 <strong>and</strong> the belt speeds up to 3.2 km/h. As<br />

the patient steps on the <strong>running</strong> belt the doctor<br />

press the START key again <strong>and</strong> by pressing the<br />

START key the second time the displays will be<br />

reset to zero.<br />

After completing the test the results can be printed<br />

on a host printer, if connected to the interface of<br />

the <strong>running</strong> <strong>machine</strong> directly.<br />

Step Duration<br />

min:sec<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 81 of 213<br />

operation<br />

Speed Elevation Total Time<br />

min:sec<br />

Pre-Testphase. Patient st<strong>and</strong>ing on the footboards, not on the<br />

belt<br />

0 until 3.2 km/h 0 until START<br />

START<br />

is pressed<br />

is pressed<br />

Testphase. Patient steps onto the <strong>running</strong> belt.<br />

1 02:00 3.2 km/h 0 2:00<br />

2 02:00 3.2 km/h 2 4:00<br />

3 02:00 3.2 km/h 4 6:00<br />

4 02:00 3.2 km/h 6 8:00<br />

5 02:00 3.2 km/h 8 10:00<br />

6 02:00 3.2 km/h 10 12:00<br />

7 02:00 3.2 km/h 12 14:00<br />

8 02:00 3.2 km/h 14 16:00<br />

9 02:00 3.2 km/h 16 18:00<br />

10 02:00 3.2 km/h 18 20:00<br />

11 30:00 3.2 km/h 18 50:00<br />

12 - 20 not engaged / reserved for following extensions <strong>and</strong> up-dates<br />

21 - 28 free definable user profiles max. 40 Program steps / not scaleable<br />

Additional profiles in the test mode from firmware 3.02.3 – only optionally selectable, locked by default<br />

[7.B13] 70 Training<br />

Interval 600 m / 14.5 ... 17.5 km/h<br />

Step Distance/time Speed Elevation Acceleration<br />

1 600 m 14.5 km/h 0 % 3<br />

2 1:30 min 0.0 km/h<br />

3 600 m 15.5 km/h<br />

4 1:30 min 0.0 km/h<br />

5 600 m 16.5 km/h<br />

6 1:30 min 0.0 km/h<br />

7 600 m 17.5 km/h<br />

8 1:30 min 0.0 km/h


[7.B13] 71 Training<br />

Interval 500 m / 15.0 ... 19.0 km/h<br />

72 Training<br />

Interval 400 m / 16.0 ... 21.0 km/h<br />

73 Training<br />

Interval 300 ... 600 m<br />

15.0 ... 21.0 km/h – Pyramid<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 82 of 213<br />

operation<br />

Step Distance/time Speed Elevation Acceleration<br />

1 500 m 15.0 km/h 0 % 3<br />

2 1:15 min 0.0 km/h<br />

3 500 m 16.0 km/h<br />

4 1:15 min 0.0 km/h<br />

5 500 m 17.0km/h<br />

6 1:15 min 0.0 km/h<br />

7 500 m 18.0 km/h<br />

8 1:15 min 0.0 km/h<br />

9 500 m 19.0 km/h<br />

10 1:15 min 0 km/h<br />

Step Distance/Time Speed Elevation Acceleration<br />

1 400 m 16.0 km/h 0 % 3<br />

2 1:00 min 0.0 km/h<br />

3 400 m 17.0 km/h<br />

4 1:00 min 0.0 km/h<br />

5 400 m 18.0 km/h<br />

6 1:00 min 0.0 km/h<br />

7 400 m 19.0 km/h<br />

8 1:00 min 0.0 km/h<br />

9 400 m 20.0 km/h<br />

10 1:00 min 0.0 km/h<br />

11 400 m 21.0 km/h<br />

12 1:00 min 0.0 km/h<br />

Step Distance/Time Speed Elevation Acceleration<br />

1 600 m 15.0 km/h 0 % 3<br />

2 1:30 min 0.0 km/h<br />

3 500 m 17.0 km/h<br />

4 1:15 min 0.0 km/h<br />

5 400 m 19.0 km/h<br />

6 1:00 min 0.0 km/h<br />

7 300 m 21.0 km/h<br />

8 0:45 min 0.0 km/h<br />

9 300 m 21.0 km/h<br />

10 0:45 min 0.0 km/h<br />

11 400 m 19.0 km/h<br />

12 1:00 min 0.0 km/h<br />

13 500 m 17.0 km/h<br />

14 1:15 min 0.0 km/h<br />

15 600 m 15.0 km/h<br />

16 1:30 min 0.0 km/h


[7.B13] 74 Training<br />

Interval 300 ... 600 m<br />

15.0 ... 21.0 km/h<br />

75 Training<br />

Interval 300 m / 20.0 km/h<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 83 of 213<br />

operation<br />

Step Distance/Time Speed Elevation Acceleration<br />

1 600 m 15.0 km/h 0 % 3<br />

2 1:30 min 0.0 km/h<br />

3 300 m 21.0 km/h<br />

4 0:45 min 0.0 km/h<br />

5 600 m 15.0 km/h<br />

6 1:30 min 0.0 km/h<br />

7 300 m 21.0 km/h<br />

8 0:45 min 0.0 km/h<br />

9 600 m 15.0 km/h<br />

10 1:30 min 0.0 km/h<br />

11 300 m 21.0 km/h<br />

12 0:45 min 0.0 km/h<br />

13 600 m 15.0 km/h<br />

14 1:30 min 0.0 km/h<br />

15 300 m 21.0 km/h<br />

16 0:45 min 0.0 km/h<br />

Step Distance/Time Speed Elevation Acceleration<br />

1 300 m 20.0 km/h 0 % 3<br />

2 0:45 min 0.0 km/h<br />

3 300 m 20,0 km/h<br />

4 0:45 min 0.0 km/h<br />

5 300 m 20.0 km/h<br />

6 0:45 min 0.0 km/h<br />

7 300 m 20.0 km/h<br />

8 0:45 min 0.0 km/h<br />

9 300 m 20.0 km/h<br />

10 0:45 min 0.0 km/h<br />

11 300 m 20.0 km/h<br />

12 0:45 min 0.0 km/h<br />

13 300 m 20,0 km/h<br />

14 0:45 min 0,0 km/h<br />

15 300 m 20,0 km/h<br />

16 0:45 min 0,0 km/h


[7.B13] 76 Walk-Protocol 13 Steps<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 84 of 213<br />

operation<br />

Step Distance/Time Speed Elevation Acceleration<br />

1 01:00 min 1.8 km/h 0 % 1<br />

2 01:00 min 2.4 km/h<br />

3 01:00 min 3.0 km/h<br />

4 01:00 min 3.6 km/h<br />

5 01:00 min 4.3 km/h<br />

6 01:00 min 4.9 km/h<br />

7 01:00 min 5.5 km/h<br />

8 01:00 min 6.1 km/h<br />

9 01:00 min 6.7 km/h<br />

10 01:00 min 7.3 km/h<br />

11 01:00 min 7.9 km/h<br />

12 01:00 min 8.5 km/h<br />

13 03:00 min 1.8 km/h<br />

77 - 79 not engaged / reserved for following extensions <strong>and</strong> up-dates<br />

80 Profile VO2 / 10 k<br />

81 Profile VO2 / 11 k<br />

Step Duration Speed Elevation Acceleration<br />

1 2:00 min 8.0 km/h 0 % 1<br />

2 2:00 min 10.0 km/h 0 %<br />

3 1:00 min 10.0 km/h 2 %<br />

4 1:00 min 10.0 km/h 4 %<br />

5 1:00 min 10.0 km/h 6 %<br />

6 1:00 min 10.0 km/h 8 %<br />

7 1:00 min 10.0 km/h 10 %<br />

8 1:00 min 11.0 km/h 10 %<br />

9 2:00 min 11.0 km/h 11 %<br />

10 2:00 min 11.0 km/h 12 %<br />

Step Duration Speed Elevation Acceleration<br />

1 2:00 min 9.0 km/h 0 % 1<br />

2 2:00 min 11.0 km/h 0 %<br />

3 1:00 min 11.0 km/h 2 %<br />

4 1:00 min 11.0 km/h 4 %<br />

5 1:00 min 11.0 km/h 6 %<br />

6 1:00 min 11.0 km/h 8 %<br />

7 1:00 min 11.0 km/h 10 %<br />

8 1:00 min 12.0 km/h 10 %<br />

9 2:00 min 12.0 km/h 11 %<br />

10 2:00 min 12.0 km/h 12 %


file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 85 of 213<br />

operation<br />

[7.B13] 82 Profil VO2 / 12 k<br />

Step Duration Speed Elevation Acceleration<br />

1 2:00 min 10,0 km/h 0 % 1<br />

2 2:00 min 12,0 km/h 0 %<br />

3 1:00 min 12,0 km/h 2 %<br />

4 1:00 min 12,0 km/h 4 %<br />

5 1:00 min 12,0 km/h 6 %<br />

6 1:00 min 12,0 km/h 8 %<br />

7 1:00 min 12,0 km/h 10 %<br />

8 1:00 min 13,0 km/h 10 %<br />

9 2:00 min 13,0 km/h 11 %<br />

10 2:00 min 13,0 km/h 12 %<br />

83 not engaged / reserved for following extensions <strong>and</strong> up-dates<br />

84 Profile VO2 / 14 k<br />

Step Duration Speed Elevation Acceleration<br />

1 2:00 min 12.0 km/h 0 % 1<br />

2 2:00 min 14.0 km/h 0 %<br />

3 1:00 min 14.0 km/h 2 %<br />

4 1:00 min 14.0 km/h 4 %<br />

5 1:00 min 14.0 km/h 6 %<br />

6 1:00 min 14.0 km/h 8 %<br />

7 1:00 min 14.0 km/h 10 %<br />

8 1:00 min 15.0 km/h 10 %<br />

9 2:00 min 15.0 km/h 11 %<br />

10 2:00 min 15.0 km/h 12 %<br />

85 Profile Super Balke<br />

Step Duration Speed Elevation Acceleration<br />

1 1:00 min 6.0 km/h 4 % 1<br />

2 1:00 min 6 %<br />

3 1:00 min 8 %<br />

4 1:00 min 10 %<br />

5 1:00 min 12 %<br />

6 1:00 min 14 %<br />

7 1:00 min 16 %<br />

8 1:00 min 18 %<br />

9 2:00 min 20 %<br />

10 2:00 min 22 %<br />

11 2:00 min 24 %<br />

86 ... 89 not engaged / reserved for following extensions <strong>and</strong> up-dates<br />

90 Test profile<br />

Step Duration Speed Elevation Acceleration<br />

1 3:00 min 7.5 km/h 0 % 1<br />

2 7.5 km/h 3 %<br />

3 7.5 km/h 6 %<br />

4 8.0 km/h 6 %<br />

5 8.0 km/h 9 %<br />

6 8.0 km/h 12 %<br />

7 8.0 km/h 15 %<br />

8 8.0 km/h 18 %<br />

9 8.0 km/h 21 %<br />

10 8.0 km/h 24 %


[7.B13] 91 Test profile<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 86 of 213<br />

operation<br />

Step Duration Speed Elevation Acceleration<br />

1 3:00 min 9.0 km/h 0 % 1<br />

2 0:30 min 0.0 km/h<br />

3 3:00 min 10.8 km/h<br />

4 0:30 min 0.0 km/h<br />

5 3:00 min 12.6 km/h<br />

6 0:30 min 0.0 km/h<br />

7 3:00 min 14.4 km/h<br />

8 0:30 min 0.0 km/h<br />

9 3:00 min 16.2 km/h<br />

10 0:30 min 0.0 km/h<br />

11 3:00 min 18.0 km/h<br />

12 0:30 min 0.0 km/h<br />

13 3:00 min 19.8 km/h<br />

14 0:30 min 0.0 km/h<br />

15 3:00 min 21.6 km/h<br />

16 0:30 min 0.0 km/h<br />

17 3:00 min 23.4 km/h<br />

18 0:30 min 0.0 km/h<br />

19 3:00 min 25.2 km/h<br />

20 0:30 min 0.0 km/h<br />

21 3:00 min 27.0 km/h<br />

22 0:30 min 0.0 km/h<br />

23 3:00 min 28.8 km/h<br />

24 0:30 min 0.0 km/h<br />

25 3:00 min 30.6 km/h


[7.B13] 92 Test profile<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 87 of 213<br />

operation<br />

Step Duration Speed Elevation Acceleration<br />

1 3:00 min 9.0 km/h 2 % 1<br />

2 0:30 min 0.0 km/h 2 %<br />

3 3:00 min 10.8 km/h 2 %<br />

4 0:30 min 0.0 km/h 2 %<br />

5 3:00 min 12.6 km/h 2 %<br />

6 0:30 min 0.0 km/h 2 %<br />

7 3:00 min 14.4 km/h 2 %<br />

8 0:30 min 0.0 km/h 4 %<br />

9 3:00 min 14.4 km/h 4 %<br />

10 0:30 min 0.0 km/h 6 %<br />

11 3:00 min 14.4 km/h 6 %<br />

12 0:30 min 0.0 km/h 8 %<br />

13 3:00 min 14.4 km/h 8 %<br />

14 0:30 min 0.0 km/h 10 %<br />

15 3:00 min 14.4 km/h 10 %<br />

16 0:30 min 0.0 km/h 12 %<br />

17 3:00 min 14.4 km/h 12 %<br />

18 0:30 min 0.0 km/h 14 %<br />

19 3:00 min 14.4 km/h 14 %<br />

20 0:30 min 0.0 km/h 16 %<br />

21 3:00 min 14.4 km/h 16 %<br />

22 0:30 min 0.0 km/h 18 %<br />

23 3:00 min 14.4 km/h 18 %<br />

24 0:30 min 0.0 km/h 20 %<br />

25 3:00 min 14.4 km/h 20 %


[7.B13] 93 Test profile<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 88 of 213<br />

operation<br />

Step Duration Speed Elevation Acceleration<br />

1 2:30 min 8.7 km/h 3 % 3<br />

2 2:20 min 5.6 km/h 9 % 1<br />

3 0:10 min 7.8 km/h 5 % 1<br />

4 1:00 min 13.7 km/h 0 % 1<br />

5 1:00 min 9.4 km/h 0 % 3<br />

6 2:30 min 9.4 km/h 3 % 3<br />

7 2:20 min 6.3 km/h 9 % 1<br />

8 0:10 min 8.5 km/h 5 % 1<br />

9 1:00 min 14.4 km/h 0 % 1<br />

10 1:00 min 10.1 km/h 0 % 3<br />

11 2:30 min 10.1 km/h 3 % 3<br />

12 2:20 min 7.0 km/h 9 % 1<br />

13 0:10 min 9.2 km/h 5 % 1<br />

14 1:00 min 15.1 km/h 0 % 1<br />

15 1:00 min 10.8 km/h 0 % 3<br />

16 2:30 min 10.8 km/h 3 % 3<br />

17 2:20 min 7.7 km/h 9 % 1<br />

18 0:10 min 9.9 km/h 5 % 1<br />

19 1:00 min 15.8 km/h 0 % 1<br />

20 1:00 min 11.5 km/h 0 % 3<br />

21 2:30 min 11.5 km/h 3 % 3<br />

22 2:20 min 8.4 km/h 9 % 1<br />

23 0:10 min 10.6 km/h 5 % 1<br />

24 1:00 min 16.5 km/h 0 % 1<br />

25 1:00 min 12.2 km/h 0 % 3<br />

26 2:30 min 12.2 km/h 3 % 3<br />

27 2:20 min 9.1 km/h 9 % 1<br />

28 0:10 min 11.3 km/h 5 % 1<br />

29 1:00 min 17.2 km/h 0 % 1<br />

30 1:00 min 12.9 km/h 0 % 3<br />

31 2:30 min 12.9 km/h 3 % 3<br />

32 2:20 min 9.8 km/h 9 % 1<br />

33 0:10 min 12.0 km/h 5 % 1<br />

34 1:00 min 17.9 km/h 0 % 1<br />

35 1:00 min 13.6 km/h 0 % 3<br />

36 2:30 min 13.6 km/h 3 % 3<br />

37 2:20 min 10.5 km/h 9 % 1<br />

38 0:10 min 12.7 km/h 5 % 1<br />

39 1:00 min 18.6 km/h 0 % 1<br />

40 1:00 min 14.3 km/h 0 % 3<br />

41 2:30 min 14.3 km/h 3 % 3<br />

42 2:20 min 11.2 km/h 9 % 1<br />

43 0:10 min 13.4 km/h 5 % 1<br />

44 1:00 min 19.3 km/h 0 % 1<br />

45 1:00 min 15.0 km/h 0 % 3


[7.B13] 94 Test profile<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 89 of 213<br />

operation<br />

Step Duration Speed Elevation Acceleration<br />

1 2:30 min 7.2 km/h 3.5 % 1<br />

2 2:00 min 6.9 km/h 10.5 %<br />

3 1:00 min 0.0 km/h 0 %<br />

4 1:00 min 7.8 km/h 10.5 %<br />

5 1:00 min 8.1 km/h<br />

6 1:00 min 8.4 km/h<br />

7 1:00 min 8.6 km/h<br />

8 0:45 min 8.9 km/h<br />

9 0:45 min 9.2 km/h<br />

10 0:45 min 9.4 km/h<br />

11 0:30 min 9.5 km/h<br />

12 0:30 min 9.6 km/h<br />

13 0:30 min 9.8 km/h<br />

14 0:30 min 9.9 km/h<br />

15 0:10 min 10.1 km/h<br />

16 0:10 min 10.2 km/h<br />

17 0:10 min 10.4 km/h<br />

18 0:10 min 10.5 km/h<br />

19 0:10 min 10.7 km/h<br />

20 0:10 min 10.8 km/h<br />

21 0:10 min 10.9 km/h<br />

For a better documentation we recommend to use a printer linked to the serial interface RS232 or/<strong>and</strong> to use an external PC<br />

with the Software h/p/cosmos para graphics ® .


[7.B14] Test mode - Operation<br />

Initiation: <strong>running</strong> <strong>machine</strong> is in st<strong>and</strong>by mode. � LED mode is flashing.<br />

[7.B14] Step no. Activity Keys Response / Display<br />

[01] Select mode test<br />

a or d<br />

Selection of mode (manual, profile, cardio, test)<br />

until � test is flashing<br />

[02] Confirm mode test<br />

[03] Select test number<br />

[04] Confirm test program<br />

[05] Stop test program <strong>and</strong> <strong>running</strong>belt<br />

[7.B15] The UKK 2 km walk test<br />

c<br />

a or d<br />

c<br />

b<br />

$ indicates program no. "Pr. 1 " is flashing<br />

$ indicates program no. „Pr. 1" ... "Pr. 24" is<br />

flashing<br />

� N o is glowing<br />

N o 01 ... 10 predefined<br />

N o 11 ... 20 reserved<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 90 of 213<br />

operation<br />

N o 21 ... 28 user defined (free programmable profiles)<br />

Display <strong>and</strong> reaction depend on the selected type of<br />

program / test profile<br />

Running-Belt stops.<br />

(Deceleration time can be adjusted in the optional<br />

functions)<br />

The UKK 2km walk test (UKK are the initials for Urho Kaleka Kekkonen, founder of the UKK Institute in<br />

Tampere/Finl<strong>and</strong>) makes it possible under certain assumptions to determine the approximately max. cardio<br />

respiratory efficiency (VO2 max) <strong>and</strong> therefore the physical condition of the subject. On the basis of this result a<br />

fitness-index is calculated with the help of a complex formula taking age, height, weight, test duration <strong>and</strong> average<br />

heart-rate into consideration, which tells us whether the physical condition of the subject is above or below<br />

average. The value of 100 as the average serves as the base. If somebody's index-value is 90 then his physical<br />

condition is slightly below average. An index-value of 110 on the other h<strong>and</strong> would be above average.<br />

The test explains itself by its name. In order to perform the test after having done some warming up <strong>and</strong> some<br />

stretching walk for 2 km as fast as possible (do not run). It is important to keep the walking speed as high as is<br />

needed for the subject to achieve 80% of their max. heart rate ( 220 – age). The test duration <strong>and</strong> heart rate are<br />

measured immediately after finishing the 2 km distance. At h/p/cosmos devices the heart rate is measured every<br />

500m due to higher accuracy. For calculating the index value the average value of the 4 measured values is used.<br />

As soon as the 2 km distance has been absolved the speed gets reduced to half the speed at the end of the test<br />

(2000 m). The <strong>running</strong> <strong>machine</strong> operates as described under point "Manual-Mode" (the UP/DOWN-keys are in<br />

operation again) with one exception: The display INDEX instead of indicating the wattage <strong>and</strong> energy<br />

consumption, indicates the test result (calculation following). The LED INDEX is flashing (top right h<strong>and</strong> side).


The test-result is the index-value which is calculated as follows (according to sex):<br />

Men INDEX VALUE = 420 + A x 0.2 – (T x 0.19338 + HR x 0.56 + [W : (H2 ) x 2.6])<br />

Women INDEX VALUE = 304 + A x 0.4 – (T x 0.1417 + HR x 0.32 + [W : (H2 ) x 1.1])<br />

Explanation<br />

A (Age) = Age in years<br />

HR (Heart Rate) = average heart rate during the test in bpm<br />

Example: Age: 50 years / Weight: 105 kg / Height: 188 cm<br />

Duration of test: 17:34 = 1054 seconds<br />

Heart rate 500m = 158. 1000m = 156. 1500m = 160 <strong>and</strong> 2000m = 155<br />

Average (158+156+160+155) : 4 = 157 bpm<br />

Index value = 420+50x0.2–(1054x0.19338+157x0.56+(105:(1.882 )x2.6)) =<br />

420+10-(203.8+87.9+29.7x2.6)=430–368.9 = _61_<br />

T (Time) = needed for 2 km in seconds<br />

W (Weight) = Weight in kg<br />

H (Height) = Height in cm<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 91 of 213<br />

operation<br />

An index-value of less than 70 is a result quite below average (average=100). Conclusion: Our subject is in need<br />

of some endurance training. Basically the test is suitable for all physically healthy persons between the age of 20<br />

<strong>and</strong> 65. The results will be less accurate with people who are overweight. Athletes usually do not reach the<br />

required heart rate when performing the test. The results will also be less accurate with elderly people.<br />

Original the UKK walk test has not been designed for individual tests but for a larger amount of subjects. i.e.<br />

testing of a crowd of people at the same time by sending them off in 1/2 minute intervals on the 2 km-distance<br />

track. Therefore originally the UKK walk test has not been designed as a typical test for the <strong>running</strong> <strong>machine</strong>.<br />

Advantages of performing the UKK walk test on the <strong>running</strong> <strong>machine</strong>:<br />

N Being able to measure the precise time <strong>and</strong> distance of the test (it is exactly 2 km)<br />

N Heart-Rate control during the test<br />

N Constant supervision of the subject by a doctor or trainer possible<br />

N Automatic display of the fitness-index <strong>and</strong> documentation with the help of either a printer or PC-connection<br />

A serial printer or a PCL-printer with parallel/centronics plug und converter [cos10056] can be connected directly to<br />

the h/p/cosmos <strong>running</strong> <strong>machine</strong>. Following data can be printed after the UKK walk test even without using any<br />

PC: UKK fitness index, date, time, test duriation, distance, elevation, heart rate, age, sex, body-mass-index (BMI).<br />

See also chapter optional equipment / accessories (interface / printer) in this manual.


[7.B16] Programing example: UKK walk test<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 92 of 213<br />

operation<br />

Attention: Before start of the UKK walk test it is important to do some warm up exercise <strong>and</strong> you must have<br />

already found your personal maximum walk-speed.<br />

Initiation: <strong>running</strong> <strong>machine</strong> is in st<strong>and</strong>by mode. � LED mode is flashing.<br />

[7.B16] Step no. Activity Keys Response / Display<br />

[01] Select Mode Test<br />

a or d<br />

Selection of mode (manual, profile, cardio, test)<br />

until � test is flashing<br />

[02] Confirm Mode Test<br />

[03] Confirm test program no 1<br />

UKK walk test<br />

[04] Select female<br />

[05] Confirm female<br />

[06] Select age 30 years<br />

[07] Confirm 30 years<br />

[08] Select weight 55 Kg<br />

[09] Confirm 55 Kg<br />

[10] Select height 170 cm<br />

[11] Confirm 170 cm<br />

c<br />

c<br />

a or d<br />

c<br />

a or d<br />

c<br />

a or d<br />

c<br />

a or d<br />

c<br />

[12] Select personal max. walkingspeed<br />

a or d<br />

$ indicates program no. "Pr. 1 " is flashing<br />

� Sex is glowing<br />

% indicates sex: "M" (male) flashing<br />

� Sex is glowing<br />

% indicates sex: "F" (female) flashing<br />

� Age is glowing<br />

% indicates age: "35" flashing<br />

� Age is glowing<br />

% indicates age: "30" flashing<br />

� Weight is glowing<br />

% indicates weight: "65" flashing<br />

� Weight is glowing<br />

% indicates weight: "55" flashing<br />

% indicates height: "H175" flashing<br />

% indicates height: "H170" flashing<br />

The <strong>running</strong>-<strong>machine</strong> starts<br />

= indicates speed: "0.0" flashing<br />

% indicates indicates present heart rate<br />

All other displays also indicate the present parameters.<br />

In case % indicates: "P 0" flashing <strong>and</strong> at the same<br />

time an acoustic signal appears, the <strong>running</strong> <strong>machine</strong><br />

cannot receive the heart rate of the subject. Check chest<br />

belt system.


[7.B16] [13] 2000 meter walking. Walk<br />

fast as you can, but not <strong>running</strong>.<br />

[14] Stop manually -<br />

after cool-down of approx.<br />

5 min. <strong>and</strong> after notification of<br />

test result (Fitness index <strong>and</strong><br />

time)<br />

b<br />

[7.B17] Programming-example: free definable profile<br />

profile<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 93 of 213<br />

operation<br />

Automatic reduction of speed to 50 % of the manually<br />

selected maximum walking-speed after 2000 meters<br />

� INDEX is glowing<br />

" indicates test result: Fitness index according the<br />

UKK Research Institute<br />

! indicates test result: total time for the 2000 meter<br />

End of the UKK 2 km Walk Test<br />

For a better documentation we recommend to use a<br />

printer linked to the serial interface RS232 or/<strong>and</strong> to use<br />

an external PC with the Software h/p/cosmos para<br />

graphics ® .<br />

Programming <strong>and</strong> correction of a free programmed test is possible in test mode „ Pr. 21" up to „Pr. 28". Starting a<br />

free programmed test is also possible in test mode „Pr. 21" up to „Pr. 28". The below listed program profile is<br />

supposed to be available on storage facility test mode "Pr. 21" <strong>and</strong> therefore it has to be created on test mode "Pr.<br />

21" according to the below listed sequence of input.<br />

Up to 40 program steps can be stored per program. In case there are more program steps required, ask for the<br />

optional PC-Software h/p/cosmos para graphics ® which is able to control all functions of the <strong>running</strong> <strong>machine</strong> via<br />

host PC <strong>and</strong> also to serve for on-line monitoring of all data. During programming the displays indicate the values of<br />

the present program step <strong>and</strong> not the value of the total distance or total time programmed so far for this profile.<br />

In case of entering a false value, or in case you want to alternate a profile, you can "scroll" the profile steps with<br />

the help of the keys e <strong>and</strong> f. With the help of the keys d <strong>and</strong> a you can alternate the values. During<br />

programming procedure the display " indicates the acceleration level " Acc X " for the single program step.<br />

During the short sprint (see below listed program step no. 4) the <strong>machine</strong> speeds up at acceleration level 4 <strong>and</strong><br />

reduces the speed at level 2 after the sprint distance of 200 meters. During programming procedure the display<br />

% indicates the present program step " St X ".<br />

Sequence Step no. Level<br />

Warming up 1<br />

Acceleration<br />

Speed<br />

=<br />

Distance<br />

§<br />

Time<br />

!<br />

1 5.0 km/h 5:00 min. 0 %<br />

Speed increase 2 1 8.0 km/h 2:30 min. 0 %<br />

Easy up-hill run 3 1 8.0 km/h 4:00 min. 5 %<br />

Sprint with fast acceleration 4 4 16.0 km/h 200 m 0 %<br />

Cool-down 5 2 6.0 km/h 5:00 min. 0 %<br />

Stop 6 1 0.0 km/h 0:00 min. 0 %<br />

Elevation<br />

$


Initiation: Running-Machine is in st<strong>and</strong>by mode. � LED mode is flashing.<br />

[7.B17] Step no. Activity Keys Response / Display<br />

[01] Select Mode Test<br />

a or d<br />

Selection of mode (manual, profile, cardio, test)<br />

until � test is flashing<br />

[02] Confirm Mode Test<br />

[03] Select program no.<br />

[04] Confirm program no.<br />

[05] Select speed 5.0 km/h<br />

[06] Confirm speed<br />

[07] Conform distance 0 -<br />

(this step will be programmed<br />

after time)<br />

[08] Select time 5 minutes<br />

[09] Confirm time<br />

[10] Confirm time - 00 seconds<br />

[11] Confirm elevation 0 %<br />

[12] Confirm acceleration level 1<br />

[13] Select speed 8.0 km/h<br />

[14] Confirm speed<br />

[15] Confirm distance 0 (this step will<br />

be programmed after time)<br />

[16] Select time 2 minutes<br />

[17] Confirm time<br />

[18] Select time 30 seconds<br />

c<br />

a or d<br />

until display "Pr. 21<br />

" is flashing<br />

c<br />

for at least 5 sec.<br />

a or d<br />

c<br />

c<br />

a or d<br />

c<br />

c<br />

c<br />

c<br />

a or d<br />

c<br />

c<br />

a or d<br />

c<br />

a or d<br />

$ indicates program no. "Pr. 1 " is flashing<br />

$ indicates program no.: " Pr. 21" is flashing<br />

= indicates speed: Value "0.0" is flashing<br />

% indicates program step:" St. 1 "<br />

§ indicates distance: Value "0" is flashing<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 94 of 213<br />

operation<br />

! indicates time (minutes): Value "0:00" is flashing<br />

! indicates time (seconds): Value "5:00" is flashing<br />

$ indicates elevation: Value "0.0" is flashing<br />

" indicates acceleration level: "Acc. 1" is flashing<br />

= indicates speed: Value "0.0" is flashing<br />

% indicates program step: " St. 2<br />

§ indicates distance: Value "0" is flashing<br />

! indicates time (minutes): Value "0:00" is flashing<br />

! indicates time (seconds): Value "2:00" is flashing


[7.B17] [19] Confirm time<br />

[20] Confirm elevation 0 %<br />

[21] Confirm acceleration level 1<br />

[22] Select speed 8.0 km/h<br />

[23] Confirm speed<br />

[24] Confirm distance 0 (this step will<br />

be programmed after time)<br />

[25] Select time 4 minutes<br />

[26] Confirm time<br />

[27] Confirm time 00 seconds<br />

[28] Select elevation 5 %<br />

[29] Confirm elevation 5 %<br />

[30] Confirm acceleration level 1<br />

[31] Select speed 16.0 km/h<br />

[32] Confirm speed<br />

c<br />

c<br />

c<br />

a or d<br />

c<br />

c<br />

a or d<br />

c<br />

c<br />

a or d<br />

c<br />

c<br />

a or d<br />

c<br />

[33] Select distance 200 m (this step<br />

will be programmed after time) a or d<br />

[34] Confirm distance<br />

[35] Confirm elevation 0 %<br />

[36] Select acceleration level 4<br />

[37] Confirm acceleration level<br />

[38] Select der Speed 6.0 km/h<br />

c<br />

c<br />

a or d<br />

c<br />

a or d<br />

$ indicates elevation: Value "0.0" is flashing<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 95 of 213<br />

operation<br />

" indicates acceleration level: "Acc. 1" is flashing<br />

= indicates speed: Value "0.0" is flashing<br />

% indicates program step:" St. 3 "<br />

§ indicates distance: Value "0" is flashing<br />

! indicates time (minutes): Value "0:00" is flashing<br />

! indicates time (seconds): Valuet "4:00" is flashing<br />

$ indicates elevation: Value "0.0" is flashing<br />

" indicates acceleration level: "Acc. 1" is flashing<br />

= indicates speed: Value "0.0" is flashing<br />

% indicates program step:" St. 4 "<br />

§ indicates distance: Value "0" is flashing<br />

$ indicates den elevation: Value "0" is flashing<br />

" indicates acceleration level: "Acc. 1" is flashing<br />

= indicates speed: Value "0.0" is flashing<br />

% indicates program step:" St. 5 "


[7.B17] [39] Confirm speed<br />

[40] Confirm distance 0 (this step will<br />

be programmed after time)<br />

[41] Select time 5 minutes<br />

[42] Confirm time<br />

[43] Confirm time 00 seconds<br />

[44] Confirm elevation 0 %<br />

c<br />

c<br />

a or d<br />

c<br />

c<br />

c<br />

[45] Select acceleration level<br />

("deceleration") 2. a or d<br />

[46] Confirm acceleration level<br />

[47] Confirm stop-speed 0 km/h <strong>and</strong><br />

save program file<br />

c<br />

c for more<br />

than 5 seconds<br />

§ indicates distance: Value "0" is flashing<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 96 of 213<br />

operation<br />

! indicates time (minutes): Value "0:00" is flashing<br />

! indicates time (seconds): Value "5:00" is flashing<br />

$ indicates elevation: Value "0.0" is flashing<br />

" indicates acceleration level: "Acc. 1" is flashing<br />

In this program step the speed will be reduced from 16<br />

km/h to 6 km/h at an acceleration level "2”<br />

= indicates speed: Value "0.0" is flashing<br />

% indicates program step:" St. 6 "<br />

End of Programming.<br />

The profile can be started later on in test mode Pr. 21<br />

For a better documentation we recommend to use a printer linked to the serial interface RS 232 or/<strong>and</strong> to use an external PC<br />

with the Software h/p/cosmos para graphics ® .


[7.B18] Possibilities of interfering in test mode manually<br />

Initiation: Operational mode: test, <strong>running</strong>-belt is moving.<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 97 of 213<br />

operation<br />

[7.B18] Step no. Activity Keys Response / Display<br />

[01] Alteration of speed<br />

a or d<br />

Speed is being increased / reduced (0-max.) as long as<br />

key get pressed. If SPEED is being reduced down to „0“<br />

press <strong>and</strong> hold = PAUSE-position:<br />

= indicates: PAUS. Current values will be<br />

sustained. Only valid for the actual program step, no<br />

alteration in the memory. Next program step as in the<br />

profile defined.<br />

[02] Alteration of elevation<br />

Not activated in every test<br />

f or e<br />

press <strong>and</strong> hold<br />

Elevation is being increased / reduced (0-max.) as long<br />

as key get pressed. Only valid for the actual program<br />

step, no alteration in the memory. Next program step as<br />

in the profile definied.<br />

[03] Alteration of acceleration level Press several times Example: For acceleration or deceleration level “3”,<br />

(only during increase or<br />

decrease of speed)<br />

c<br />

press 3 times the corresponding key <strong>and</strong> then hold.<br />

Note: Maximum 7 levels, limited by the maximum<br />

acceleration level selected in the Optional Mode /<br />

st<strong>and</strong>ard value = 4. During deceleration this function<br />

alternates the degree of deceleration.<br />

[04] Switch over to the next program<br />

step or one program step back<br />

(not activated in every test).<br />

c<br />

Jump over to the next program step or jump to the<br />

previous program step<br />

[05] Switch over to the next mode or<br />

one mode back<br />

(not activated in every test)<br />

<strong>and</strong> simultaneous<br />

e or f<br />

c<br />

<strong>and</strong> simultaneous<br />

a or d<br />

Jump over to the next mode or to the previous mode.<br />

This is a perfect way to change from test mode to<br />

manual mode (cool down) without stop of the belt.<br />

For a better documentation we recommend to use a printer linked to the serial interface RS 232 or/<strong>and</strong> to use an external PC<br />

with the Software h/p/cosmos para graphics ® .


[8.]Optional settings<br />

[8.A] Settings of control <strong>and</strong> display options<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 98 of 213<br />

optional settings<br />

Optional settings serve for example as an acknowledgement of error notes on the display, or for the st<strong>and</strong>ard setting of<br />

the device: selection of RS232 interface protocol, etc.<br />

Remark for devices without UserTerminal: For optional settings on devices without UserTerminal an external<br />

UserTerminal or a connected PC with the software h/p/cosmos para control ® is required.<br />

[8.A1] Select user options<br />

Initiation: <strong>running</strong> belt is not moving. One of the modes LED � is flashing: (manual, profile, cardio, test)<br />

[8.A1] Step no. Activity Keys Response / Display<br />

[01] Select option mode (OPxx) for<br />

users – the option mode for a<br />

= indicates: OP 01 flashing (for option no. 01)<br />

! <strong>and</strong> " indicating: E.rE SEt (for error reset)<br />

administrators is only accessible <strong>and</strong> simultaneuosly<br />

for authorised service<br />

technicians, not for users d<br />

<strong>and</strong> simultaneuosly<br />

b<br />

[02] Select option no.<br />

[03] Confirm option no.<br />

Example: OP 01 (OIL & other<br />

error codes): Confirmation &<br />

deleting by pressing:<br />

Example: OP 02 (Distance /<br />

km): Confirmation by pressing<br />

Example: OP 08<br />

(STOP-time/ "slow-down-time"):<br />

a) change value by pressing<br />

b) confirm value by pressing<br />

for at least 3<br />

seconds.<br />

a or d<br />

c<br />

c<br />

c<br />

a) a or d<br />

b) c<br />

= indicates: OP 01 ... OP 53 flashing (for option no.<br />

01 ... 53)<br />

! <strong>and</strong> " indicate: a short explanation of the<br />

option, e.g.: E.rE SEt (for Error reset)<br />

Read, confirm <strong>and</strong> set options according to the following<br />

option list. All adjustable values are going to be shown<br />

flashing. All non-adjustable values are not going to be<br />

shown flashing.<br />

Display indicates: donE (confirmation)


[8.A2] List of user options / optional functions<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 99 of 213<br />

optional settings<br />

For devices without UserTerminal: An external UserTerminal or PC software h/p/cosmos para control ® is required<br />

to control the <strong>running</strong> <strong>machine</strong> via RS232 interface.<br />

[8.A2] Option Description Comment / Display<br />

OP01 Setting back (deleting) of error<br />

The required maintenance work has to be performed before deleting<br />

messages<br />

error messages. The h/p/cosmos service department has to be<br />

contacted before one of the service displays or error codes will be<br />

deleted!<br />

Confirmation in display § indicating "donE"<br />

Info: This option only sets back the error message! If the error still<br />

exists, you are not able to set back the error message. In this case<br />

consult an authorised service engineer<br />

The following intervals are set back by this operation:<br />

E.01: Oil-interval A-OP 35<br />

E.02: Service-time-interval A-OP 37<br />

E.02: Service-distance-intervall A-OP 38<br />

OP02 Total distance covered (km)<br />

Note: This option resets exclusively the causing variable. If for<br />

example the E.02 service-time-interval is being set back with this<br />

operation, the distance-interval is not set back <strong>and</strong> vice versa! Only<br />

Administrator-option OP 47 resets all three values implicitly!<br />

§ <strong>and</strong> $ indicates: total distance covered in km<br />

OP03 Indication of total hours of operation (h)<br />

= st<strong>and</strong>-by-time including runtime of<br />

motor/ <strong>running</strong> belt<br />

OP04 Indication of total hours of operation (h)<br />

= runtime of motor/<strong>running</strong> belt<br />

% indicates km<br />

§ <strong>and</strong> $ show: operation hours<br />

% reports: h<br />

§ <strong>and</strong> $ report: operation hours<br />

% shows: h<br />

OP05 Indication of Firmware-Version <strong>and</strong> date = reports “OP05” ! reports “typE”<br />

OP06 Adjustment of actual date <strong>and</strong> the real<br />

time clock<br />

" indicates device type, e.g. “1.4”<br />

§ indicates: „MCU 5“ $ reports version, e.g. “1.01.1”<br />

% reports default type, e.g. 1.3<br />

! shows: rtc for Real Time Clock<br />

$ reports flashing: date / time, year, month, date, hours, minutes,<br />

seconds


file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 100 of 213<br />

optional settings<br />

[8.A2] OP07 Acoustic heart rate signal This function is normally used to control the regularity of the heart rate<br />

or to find reasons for transmitting problems (like mobile phones or<br />

computer monitors).<br />

§ indicates: OFF or ON<br />

OFF: No acoustic heart-rate-signal<br />

ON: Acoustic heart-rate-signal for every beat<br />

OP08 Stop Time / Deceleration Time after § indicates: Stop time in sec.<br />

STOP key in relation to the max. speed<br />

$ indicates: sec for seconds<br />

OP09 Start Speed (Modus Manual <strong>and</strong> Cardio)<br />

for feedback after START key has been<br />

pressed.<br />

This value can be reduced to 0.0 km/h<br />

for advanced users.<br />

OP11 Scaling of the profiles in profile mode<br />

(not for mode test)<br />

Adjustable from 2.... 30 seconds<br />

= indicates: Start speed in km/h<br />

� max blinking<br />

� set unit is blinking<br />

Adjustable from 0.0 km/h ... 5.0 km/h<br />

OP12 Unit for display of speed = indicates: Unit for speed<br />

... without decimal place:<br />

§ shows: scaling possibilities<br />

0: no scaling (st<strong>and</strong>ard)<br />

1: Scaling 1...6, which is shown in the mode profile at the display<br />

INDEX, refers to all parameters (speed, elevation, time)<br />

2: Scaling 1...6 occurs to each parameter (speed, elevation, time)<br />

individually<br />

3 = m/min<br />

... with one decimal place:<br />

0 = km/h 1 = m/s 2 = mph 23 = m/min<br />

... with two decimal places:<br />

20 = km/h 21 = m/s 22 = mph<br />

� km/h, m/s, mph or m/min is blinking<br />

OP13 Unit for display of distance § indicates: Unit for distance<br />

0 = km 1 = miles 2 = m<br />

� m, km, or miles is blinking<br />

OP14 Unit for angle of elevation $ indicates: Unit for elevation<br />

0 = % (percent) 1 = ° (degree)<br />

� % or ° blinking<br />

OP15 Subject's Body Weight (default value) § indicates: 10 ... 250 (estimated weight)<br />

� weight blinking<br />

The personal body weight is necessary for a more correct calculation<br />

(estimation) of the power <strong>and</strong> energy consumption.<br />

.


file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 101 of 213<br />

optional settings<br />

[8.A2] OP16 Request for body weight before manual 0 = OFF. Request for body weight before starting a program is not<br />

or automatic start<br />

required. Calculation of energy consumption <strong>and</strong> power is based on<br />

the body weight entered in option no 15.<br />

1 = ON. Input of body weight before starting a program is required.<br />

Calculation of energy consumption <strong>and</strong> power is based on the entered<br />

body weight.<br />

OP17 Unit for energy consumption JOUL = kJoule is the unit for energy consumption<br />

OP18 Maximum speed in Cardio mode (default<br />

value)<br />

(this option is only available for treadmillergometers,<br />

not for ladder-ergometers)<br />

OP19 Setting of sender for polar W:I:N:D:<br />

system<br />

CALO = kcal is the unit for energy consumption<br />

= reports: 0.0 ... max<br />

for the default value for maximum speed in the Cardio mode.<br />

� adjusted unit is blinking<br />

� max. is blinking<br />

The value of max. speed in Cardio mode can be changed online by<br />

pressing d a<br />

0000 0000 = all senders are accepted (can also be set with UP<br />

<strong>and</strong> DOWN)<br />

xxxx xxxx = only specific sender with special ID is accepted<br />

(must be set with d <strong>and</strong> a)<br />

9999 9999 = next sender will be accepted, saved <strong>and</strong> filtered<br />

(can also be set with UP <strong>and</strong> DOWN)<br />

(further settings in user OP23 <strong>and</strong> administrator OP16 are necessary)


[8.A2] OP20 RS232 interface protocol: COM 1<br />

Various ECG <strong>and</strong> Ergospirometry<br />

systems support the h/p/cosmos coscom<br />

protocol. Individual adaptation of<br />

protocol on request. For further<br />

information about the h/p/cosmos<br />

coscom protocol see www.coscom.org<br />

We recommend to use the h/p/cosmos<br />

coscom protocol, as it provides a high<br />

safety <strong>and</strong> functionality st<strong>and</strong>ard.<br />

h/p/cosmos para control ® <strong>and</strong><br />

h/p/cosmos para graphics ® exclusively<br />

work with the h/p/cosmos coscom<br />

protocol.<br />

For advanced h/p/cosmos coscom v3<br />

connections with Baudrate 115200<br />

please select OP20=20 . Please note<br />

that the connected peripherie/software<br />

must be approved for h/p/cosmos<br />

coscom v3 with Baudrate 115200 (for<br />

example h/p/cosmos para control 4.0).<br />

OP21 RS232 interface protocol: COM 2<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 102 of 213<br />

optional settings<br />

§ indicates flashing: number of the RS232 interface protocol<br />

$ <strong>and</strong> % reports:<br />

OFF = RS232 not active / no protocol / interface deactivated<br />

1 = h/p/cosmos coscom v1, v2, v3 with Baudrate 9600 bps<br />

( st<strong>and</strong>ard setting COM 1 <strong>and</strong> COM 2)<br />

2 = JAEGER OXYCON (up to 1995); if available use: h/p/cosmos<br />

coscom (= 1)<br />

3 = Printer protocol (serial printer or converter required)<br />

4 = Free<br />

5 = Pacer treadmill emulation (max. 12.7 km/h)<br />

6 = MAX 1 (GE Marquette) EKG; if available use: h/p/cosmos<br />

coscom (= 1)<br />

7 = Trackmaster treadmill emulation in km/h<br />

8 = Trackmaster treadmill emulation in miles per hour<br />

9 = custo card ECG (MS DOS-Version); if available use: h/p/cosmos<br />

coscom (= 1)<br />

10 = Loop Back Test (special test plug required, available at<br />

h/p/cosmos)<br />

11 = SunTech Tango blood pressure monitor<br />

12 = Remote Control Hardware Terminal MCU 4<br />

(special hardware required)<br />

13 = Burdick T 500 treadmill emulation<br />

14 = Burdick T 600 treadmill emulation<br />

17 = Chipcard reader Ergofit (special hardware required) – only<br />

available at MCU4 (all firmware versions)<br />

or MCU5 firmware < v1.03.x<br />

18 = Chipcard reader Proxomed (special hardware required) – only<br />

available at MCU4 (all firmware versions) or MCU5 firmware <<br />

v1.03.x<br />

20 = h/p/cosmos coscom v3 with Baudrate 115200 bps<br />

See descriptions above<br />

OP23 RS232 interface protocol COM 4 § indicates flashing: number of the RS232 interface protocol<br />

$ <strong>and</strong> % reports:<br />

OFF = RS232 not active / no protocol / interface deactivated<br />

18 = Chip-card reader PROXOMED (special hardware required)<br />

20 = h/p/cosmos coscom v3 / Baudrate 115200 bps<br />

22 = Polar W:I:N:D: - System (weitere Einstellungen in Benutzer<br />

Option OP19 und Administrator Option OP16)


[8.A2] OP27 Minimum acceleration <strong>and</strong> deceleration<br />

level<br />

The selected minimum level is valid for<br />

all acceleration <strong>and</strong> deceleration<br />

processes in all modes <strong>and</strong> profiles.<br />

OP28 Maximum acceleration <strong>and</strong> deceleration<br />

level<br />

The selected maximum level is valid for<br />

all acceleration <strong>and</strong> deceleration<br />

processes in all modes <strong>and</strong> profiles<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 103 of 213<br />

optional settings<br />

§ blinking, reports the min. acceleration / deceleration for all<br />

modes <strong>and</strong> profiles (st<strong>and</strong>ard: level 1)<br />

Level of settings: 1 ... 5 but not higher than the value in option 28. Due<br />

to safety reasons, the acceleration / deceleration levels 5, 6 <strong>and</strong> 7 can<br />

not be selected. Note: The selected acceleration <strong>and</strong> deceleration level<br />

is NOT valid for control <strong>and</strong> operation via RS232-interface. In this case<br />

the acceleration <strong>and</strong> deceleration level is set in option no. 29 resp. set<br />

in the corresponding comm<strong>and</strong> of the h/p/cosmos coscom protocol.<br />

§ blinking, reports the max. acceleration / deceleration for all<br />

modes <strong>and</strong> profiles (st<strong>and</strong>ard: level 4)<br />

Note: Due to safety reasons, the acceleration / deceleration levels 5, 6<br />

<strong>and</strong> 7 can only be selected, if the safety of the subject is provided for<br />

by the use of the respective fall prevention (e. g. by the safety arch).<br />

The maximum acceleration <strong>and</strong> deceleration level is NOT valid for<br />

control <strong>and</strong> operation via V24/RS232 interface. In this case the<br />

acceleration <strong>and</strong> deceleration level is set in option no. 29 resp. set in<br />

the corresponding comm<strong>and</strong> of the h/p/cosmos coscom protocol.<br />

OP29 St<strong>and</strong>ard- Acceleration <strong>and</strong> deceleration The selected acceleration <strong>and</strong> deceleration level is valid for control <strong>and</strong><br />

level for RS 232 interface<br />

operation via RS232 interface. This option is very helpful, if the<br />

peripheral equipment (e.g. ECG, Ergospirometry, PC) does not offer a<br />

menu for acceleration <strong>and</strong> deceleration levels.<br />

§ blinking, reports: 1 ... 5, (St<strong>and</strong>ard: 1) for the acceleration <strong>and</strong><br />

deceleration level for all speed comm<strong>and</strong>s via RS 232. The maximum<br />

value adjustable depends on the setting of option 28.<br />

Note: If the peripheral equipment sends an acceleration <strong>and</strong><br />

deceleration comm<strong>and</strong> via the h/p/cosmos coscom protocol, the<br />

selected level in option number 29 is not valid for these speed<br />

comm<strong>and</strong>s.<br />

OP40 Locking <strong>and</strong> unlocking of the treadmill OFF = After switching on, the treadmill is completely locked / not<br />

accessible. To unlock the treadmill, press the buttons +, - <strong>and</strong> START<br />

together at the same time. While locked the display shows "no<br />

ACCESS"<br />

ON = treadmill is unlocked / accessible (st<strong>and</strong>ard)<br />

OP41 Locking <strong>and</strong> unlocking the manual mode OFF = manual mode is locked / not accessible<br />

ON = manual mode is unlocked / accessible (st<strong>and</strong>ard)<br />

OP42 Locking <strong>and</strong> unlocking the profile mode OFF = profile mode is locked / not accessible<br />

1 ... 6 = profile mode is unlocked / accessible up to the selected profile<br />

number St<strong>and</strong>ard: 6<br />

Example: selected profile number = 3: The profiles 1-3 can be<br />

selected, the profiles 4 – 6 can not be selected<br />

OP43 Locking <strong>and</strong> unlocking the cardio mode OFF = cardio mode is locked / not accessible<br />

ON = cardio mode is unlocked (st<strong>and</strong>ard)


[8.A2] OP44 Locking <strong>and</strong> unlocking the test mode<br />

OP45 Report mode display „index“<br />

"<br />

OP46<br />

Report mode display „elevation“ $<br />

in „Profile“ <strong>and</strong> „Test“ mode<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 104 of 213<br />

optional settings<br />

OFF = test mode is locked / not accessible<br />

1 ... 94 = test mode is unlocked / accessible up to the selected test<br />

profile number St<strong>and</strong>ard: 24<br />

selected test number = 5: The test profiles 1-5 can be selected, the<br />

test profiles 6 – 94 can not be selected<br />

0 = Display automatically switches through (Default)<br />

1 = Display permanently in MET<br />

2 = Display permanently in KJ<br />

3 = Display permanently in Watt<br />

After total switch off the default value 0 will be valid again.<br />

0 = Display automatically switches through (Default)<br />

1 = Display permanently in % or degree (°), depending on OP14<br />

2 = Display permanently in „step“<br />

After total switch off the default value 0 will be valid again.<br />

OP47 Sustain values in display resp. automatic OFF = Display values are deleted after pressing START again or<br />

„Reset“<br />

automatically after 2 minutes after having pressed STOP (Default)<br />

ON = Display values will be continued (added) after pressing START<br />

again <strong>and</strong> will not be automatically deleted by pressing STOP.<br />

Display values can only be deleted by pressing the STOP key twice<br />

(time, distance, energy).<br />

OP48 Count-Down of program-step OFF = The time-display counts up each program-step<br />

ON = The time-display counts down each program-step<br />

OP52 Output interval for printer protocol By entering a value between 0 <strong>and</strong> 100 the output interval is set in<br />

seconds for a printer directly connected to the treadmill. St<strong>and</strong>ard: 60<br />

(= Printout of all values once a minute). The value 0 disables the<br />

printing out of single values, but not the printing of headers <strong>and</strong> end<br />

results (UKK).<br />

OP53 Language settings for printer Select the language for printouts on a printer directly connected<br />

protocol<br />

to the treadmill. You can choose between six languages. Both<br />

the protocol printout as well as the test result <strong>and</strong> training<br />

recommendation of the UKK 2 km walk test are printed in the<br />

selected language.<br />

EnGL = Englisch (st<strong>and</strong>ard) SPAn = Spanish<br />

GErM = German POrt = Portuguese<br />

FrEn = French HUnG = Hungarian<br />

For a correct printout the connected printer must be compatible with<br />

PCL printer language. For special characters the ISO 8859-1 (Latin-1)<br />

font is used.


[8.A3] St<strong>and</strong>ard settings for the user options<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 105 of 213<br />

optional settings<br />

[8.A3] Option Character / Function of option Default setting Adjustable range<br />

OP 01 Set back of error messages.<br />

OP 02 Total distance (km) Report only 0 ... 4.294.976 km<br />

OP 03 Total hours of operation st<strong>and</strong>-by incl. runtime (h) Report only 0 ... 1.193.046 h<br />

OP 04 Total hours of operation runtime only (h) Report only 0 ... 1.193.046 h<br />

OP 05 Firmware Version Report only x.xx.x<br />

OP 06 Adjustment of the real time clock current date/time currently ... 31.12.2092<br />

OP 07 Acoustic Heart-Rate-signal OFF OFF or ON<br />

OP 08 Stopping / Deceleration Time 5 seconds 2 ... 30 seconds<br />

OP 09 Starting Speed (Modus Manual <strong>and</strong> Cardio) 0.5 km/h 0.0 km/h ... 5.0 km/h<br />

OP 11 Scaling of Profile mode 0 0. 1 or 2<br />

OP 12 Unit for display of speed – one decimal place rsp. none<br />

for 3 = m/min<br />

Unit for display of speed – two decimal places rsp.<br />

none for 3 = m/min<br />

0 = km/h 0 = km/h 1 = m/s<br />

2 = mph 3 = m/min<br />

20 = km/h 21 = m/s<br />

22 = mph 23 = m/min<br />

OP 13 Unit for display of distance 2: m 0: km;1: miles; 2: m<br />

OP 14 Unit for angle of elevation 0: % (Prozent) 0 = % / 1 = ° (Grad)<br />

OP 15 Subject's Body Weight 65 kg 10 ... 250 kg<br />

OP 16 Request for body weight before manual or automatic<br />

start<br />

OFF (no request) OFF / ON<br />

OP 17 Unit for energy consumption JOUL = kJoule JOUL = kJoule<br />

CALO = kcal<br />

OP 18 Maximum speed (default) in Cardio mode 6.0 km/h 0.0 ... max. speed<br />

OP19 Setting of Polar W:I:N:D System 0000 0000 xxxx xxxx<br />

OP 20 RS 232 interface protocol: COM 1 1 = h/p/cosmos coscom 1 ... 20<br />

OP 21 RS 232 interface protocol: COM 2 1 = h/p/cosmos coscom 1 ... 18<br />

OP 23 RS 232 interface protocol: COM 4 20 = h/p/cosmos coscom v3 OFF,. 18,. 20, 22<br />

OP 27 Minimum acceleration <strong>and</strong> deceleration level 1 1 ... 5<br />

OP 28 Maximum acceleration <strong>and</strong> deceleration level 4 1 ... 7<br />

OP 29 Acceleration <strong>and</strong> deceleration level for remote control<br />

via RS232 interface<br />

1 1 ... 5<br />

OP 40 Locking <strong>and</strong> unlocking the treadmill ON (unlocked) OFF = locked<br />

ON = unlocked<br />

OP 41 Locking <strong>and</strong> unlocking the Manual mode ON (unlocked) OFF = locked<br />

ON = unlocked<br />

OP 42 Locking <strong>and</strong> unlocking the Profile mode 6 (unlocked up to profile 6) 0 ... 6<br />

OP 43 Locking <strong>and</strong> unlocking the Cardio mode ON (unlocked) OFF = locked<br />

ON = unlocked<br />

OP 44 Locking <strong>and</strong> unlocking the Test mode 28 (unlocked up to test 28) 0 ... 94<br />

OP 45 Report mode Display „index“ 0 0 … 3<br />

OP 46 Report mode Display „elevation“ 0 0 … 2<br />

OP47 Sustain values in display resp. automatic „Reset“ OFF OFF = RESET automat.<br />

ON= RESET 2xSTOP<br />

OP48 Count-Down of program-step OFF OFF = count-up<br />

ON = count-down<br />

OP 52 Output interval for printer protocol 60 (seconds) 0 = no single values,<br />

1 ... 100<br />

OP 53 Language settings for printer protocol English English, German,<br />

French, Spanish<br />

Portuguese, Hungarian


[8.B] Administrator options<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 106 of 213<br />

optional settings<br />

The administrator options are only accessible by h/p/cosmos staff <strong>and</strong> authorised service engineers. The access is only<br />

possible with a special code which is not depicted here.<br />

Changing an administrator option might cause serious problems if you are not familiar with all the details. Therefore the<br />

h/p/cosmos service has to be contacted before doing any changes in the administrator options.<br />

For optional settings on devices without UserTerminal an external UserTerminal or a connected PC with the software<br />

h/p/cosmos para control is required.<br />

[8.B1] List of the administrator options<br />

[8.B1] Option Description Comment / Display<br />

OP01 Set back of error messages E 20, E 21,<br />

E 30, E 31, E 32, E 41, E 50 <strong>and</strong> E 51<br />

The required maintenance work has to be performed before deleting<br />

error messages. Confirmation in the display § with "donE"<br />

OP02 Selection of the OEM-Code<br />

This option only sets back the error message. If the error still exists,<br />

the error message will appear again<br />

By selecting the OEM code the functions <strong>and</strong> settings of the<br />

UserTerminal will be adapted to the requirements of the listed OEM<br />

partner.<br />

Following OEM codes can be selected:<br />

0 = h/p/cosmos (St<strong>and</strong>ard)<br />

1 = JAEGER / VIASYS / CARDINAL HEALTH<br />

2 = Ergo-Fit (1997-98)<br />

3 = Proxomed / Kardiomed<br />

4 = KISTLER Gaitway<br />

6 = COSMED<br />

9 = SCHILLER<br />

OP03 Selection of device type<br />

Not all companies listed above are or were OEM partners of<br />

h/p/cosmos.<br />

Selection of the correct treadmill device type according to the „list of<br />

device types” in the appendix. On the ladder-ergometer only the device<br />

type 0.1 <strong>and</strong> 1.1 can be confirmed..<br />

Note: Selection <strong>and</strong> confirmation of a device type sets back all user<strong>and</strong><br />

administrator options to default.


[8.B1] OP04 Maximum speed for reverse belt rotation<br />

(This option is only available for<br />

treadmills, not for a ladder-ergometer).<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 107 of 213<br />

optional settings<br />

Due to safety reasons, all treadmill models with a motor hood or a<br />

crossbar h<strong>and</strong>rail the speed must be limited to maximum speed of 5<br />

km/h while using belt rotation since they have an "obstacle" (motor<br />

hood or cross-bar h<strong>and</strong>rail) at the back of the subject during this mode.<br />

Changing this speed limitation for these models is only allowed if the<br />

safety of the runner is guaranteed (e.g. with a safety harness).<br />

= indicates: Maximum speed for reverse belt rotation.<br />

Adjustable from 1 km/h to maximum treadmill speed.<br />

� selected unit is blinking<br />

� max. is blinking<br />

OP05 Maximum Speed According to the selected device type (in OPTION 03) the st<strong>and</strong>ard<br />

maximum speed will be indicated in the SPEED Display.<br />

Example: h/p/cosmos mercury 4.0 = 22.0 km/h; h-p-cosmos discovery<br />

4.0 = 40 m/min<br />

For safety purposes (e.g. for beginners) the max. speed of every<br />

treadmill model can be limited.<br />

OP06 Maximum acceleration time in seconds<br />

(This option is only available for<br />

treadmills, not for a ladder-ergometer).<br />

OP07 Maximum Elevation<br />

(This option is only available for<br />

treadmills, not for a ladder-ergometer).<br />

= indicates: Maximum Speed<br />

Adjustable from 1 km/h to maximum treadmill speed<br />

� selected unit is blinking<br />

� max. is blinking<br />

§ reports e.g. 131 (max. acceleration increment)<br />

$ reports e.g. 5.0 (max. acceleration time in seconds.)<br />

% indicates: SEC.<br />

According to the selected device type (in OPTION 03) the st<strong>and</strong>ard<br />

maximum elevation will be indicated.<br />

Example: h/p/cosmos mercury = 25 %<br />

For safety purposes (e.g. for beginners, or in case the height of the<br />

room's ceiling is not enough when using an arch construction with<br />

safety harness) the max. elevation of every treadmill model can be<br />

limited.<br />

$ reports: maximum elevation<br />

Adjustable from 0 to maximum treadmill elevation


file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 108 of 213<br />

optional settings<br />

[8.B1] OP08 Heart Rate Display Interval § indicates: Heart Rate Display Interval in seconds<br />

Einstellbereich: 0 (beat to beat) .... 1 bis 9 Sekunden<br />

Adjustable from 0 (beat to beat) .... to 9 seconds<br />

Compare: An original POLAR watch has a display interval of 5<br />

seconds<br />

$ reports: SEC<br />

OP09 Test of all displays <strong>and</strong> LEDs<br />

OP10 Enter the serial number of the device Change the values with a or d<br />

according to the nameplate<br />

Confirm the values with c<br />

OP11 Activating <strong>and</strong> de-activating speed<br />

Example: Treadmill with serial number cos30000va08-0128 on the<br />

nameplate<br />

§ reports: 300 (item number of device)<br />

$ indicates: 0008 (item number of device)<br />

% indicates: 0128 (consecutive serial numbers of this device<br />

family)<br />

Normally the speed measurement is activated / de-activated by<br />

measurement<br />

selecting the correct device type (administrator option number 3). Only<br />

in very special cases a special adjustment in option number 11 is<br />

necessary. Speed measurement can only be activated if the device<br />

has a speed sensor<br />

§ reports:<br />

0 = without speed measurement / -sensor<br />

1 = with speed measurement / -sensor<br />

(permanent control)<br />

2 = with speed measurement / -sensor<br />

(control as of speed > 2 km/h, no control < 2km/h)<br />

OP12 Setting of allowance for speed-signal- The MCU5 control unit is comparing continuously the data of the<br />

comparison (Measurement / Control) speed measurement with a calculated reference value. In order to<br />

detect problems with the speed measurement or speed control or e.g.<br />

These values do not reflect the real performance limitations due to a bad voltage supply, an allowance<br />

speed accuracy of the <strong>running</strong> belt. range can be set in this option. In case the allowance is exceeded the<br />

Accuracy of <strong>running</strong>-belt speed is higher. error message E.31 (measured speed is too high) or E. 32 (measured<br />

speed is too low) is reported.<br />

§ reports: OFF = no comparison<br />

6 = 6 % deviation is allowed<br />

8 = 8 % deviation is allowed<br />

10 = 10 % deviation is allowed


file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 109 of 213<br />

optional settings<br />

[8.B1] OP13 Activating <strong>and</strong> de-activating motor break The motor break prevents the <strong>running</strong> belt from manual moving (for<br />

example because of the subjects bodyweight) when the speed value is<br />

set to “0”. If the treadmill recognizes a <strong>running</strong> belt movement it<br />

activates the motor break for 10 seconds.<br />

.<br />

OP14 Maximum acceleration time for RS232<br />

(option only available for treadmills, not<br />

for ladder ergometers)<br />

OP16 Activating <strong>and</strong> de-activating heart rate<br />

measurement<br />

Remark:<br />

The motor break is only working at <strong>running</strong> <strong>machine</strong>s with integrated<br />

speed measurement. The speed sensor must be switched on<br />

(OP11=1). If the <strong>running</strong> <strong>machine</strong> is not equipped with speed<br />

measuring system this could be retrofitted easily.<br />

To activate the motor break it is necessary to additionally change<br />

some parameters at the frequency inverter. Ask the service<br />

department of h/p/cosmos sports & medical gmbh for detailed<br />

instruction.<br />

Attention!<br />

If the device is switched off or disconnected from<br />

the power supply the motor break is not working!<br />

After 10 seconds of operation of the motor break,<br />

the break is released automatically <strong>and</strong> is only<br />

activated again after a belt movement. A short jerk<br />

may occure during reactivation.<br />

§ shows OFF = Motor break not active<br />

§ shows ON = Motor break active<br />

§ shows: e.g. 5 (max. acceleration-increment)<br />

$ shows: e.g. 131.1 (max. acceleration-time in sec.)<br />

% shows: SEC.<br />

= reports ...<br />

0 = heart rate measurement is de-activated / OFF<br />

1 = heart rate measurement is activated, circuit board HFU1 or<br />

RE06REC (default)<br />

2 = heart rate measurement is activated, circuit board HFUi<br />

auto = automatic detection <strong>and</strong> activation of circuit board<br />

4 = Polar W:I:N:D (further settings in user OP19 <strong>and</strong> OP23 necessary)<br />

OP33 Distance between rungs of the ladder For correct calculation of the number of rungs / steps, the distance<br />

ergometer<br />

between two rungs can be changed <strong>and</strong> confirmed in this option.<br />

(option only available for ladder St<strong>and</strong>ard: 254 for 25,4 cm<br />

ergometers, not for treadmills)


file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 110 of 213<br />

optional settings<br />

[8.B1] OP34 Manual speed calibration<br />

Normally the speed calibration values are set by selecting the correct<br />

(counted increments from the speed device type (administrator option number 3). Only for special speed<br />

sensor for 10 m belt movement) <strong>and</strong> special models the speed calibration values have to be changed. If<br />

you change the speed calibration value, please also pay attention to<br />

the correct settings in option number 5 (firmware speed limit), number<br />

11 (speed measurement; must be switched on) <strong>and</strong> option number 48<br />

(Frequency inverter speed limit) of the administrator mode. To change<br />

the speed calibration value please use the + <strong>and</strong> – buttons. For quick<br />

movements between st<strong>and</strong>ard values, please use the UP- <strong>and</strong> DOWNbuttons.<br />

If the speed measurement is switched off in OP11, then OP34 will only<br />

report internal program absolutes. In the other case following values<br />

are stored:<br />

N Running <strong>machine</strong> series h/p/cosmos mercury with<br />

special speed 10 km/h:<br />

Speed Calibration Value: 8.490<br />

Inverter Frequency: 80.2 Hz<br />

N Running <strong>machine</strong> series h/p/cosmos mercury with<br />

speed 22 km/h:<br />

Speed Calibration Value: 5.988<br />

Inverter Frequency: 123.3 Hz<br />

N Running <strong>machine</strong> series h/p/cosmos mercury with<br />

special speed 30 km/h:<br />

Speed Calibration Value: 4.165<br />

Inverter Frequency: 119.0 Hz<br />

N Running <strong>machine</strong> series h/p/cosmos quasar with<br />

special speed 10 km/h:<br />

Speed Calibration Value: 7.800<br />

Inverter Frequency: 80.0 Hz<br />

N Running <strong>machine</strong> series h/p/cosmos quasar with<br />

special speed 22 km/h:<br />

Speed Calibration Value: 4.495<br />

Inverter Frequency: 94.5 Hz<br />

N Running <strong>machine</strong> series h/p/cosmos quasar with<br />

speed 25 km/h:<br />

Speed Calibration Value: 4.495<br />

Inverter Frequency: 110.0 Hz<br />

N Running <strong>machine</strong> series h/p/cosmos quasar with<br />

special speed 30 km/h:<br />

Speed Calibration Value: 4.010<br />

Inverter Frequency: 115.0 Hz<br />

N Running <strong>machine</strong> series h/p/cosmos quasar with<br />

special speed 40 km/h:<br />

Speed Calibration Value: 2.874<br />

Inverter Frequency:<br />

Continuation next page…<br />

113 Hz


file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 111 of 213<br />

optional settings<br />

[8.B1] Continuation… N Running <strong>machine</strong> series h/p/cosmos quasar with<br />

special speed 45 km/h:<br />

Speed Calibration Value: 2.874<br />

Inverter Frequency: 125 Hz<br />

N Running <strong>machine</strong> series h/p/cosmos pulsar 3p with 40 km/h:<br />

Speed Calibration Value: 2.874<br />

Inverter Frequency: 110 Hz<br />

N Running <strong>machine</strong> series h/p/cosmos pulsar 3p with<br />

special speed 45 km/h:<br />

Speed Calibration Value: 2.874<br />

Inverter Frequency: 125 Hz<br />

OP35 Oil Interval<br />

The distance interval for indication of "OIL" service required can be<br />

adjusted by means of option 35.<br />

St<strong>and</strong>ard = 1000 km for treadmills, 100 km for ladder-ergometers. OIL<br />

intervals are indicated with error code "E01" "OIL" every 1000 km.<br />

§ reports: oil interval in km (St<strong>and</strong>ard = 1.000)<br />

� km is flashing<br />

Adjustment range: OFF, 100 ... 5.000 km<br />

Do not change the interval without prior communication with<br />

h/p/cosmos. For the models h/p/cosmos venus, saturn und orbiter,<br />

JAEGER LE 580 <strong>and</strong> bigger as well as COSMED T200 <strong>and</strong> bigger<br />

option number 35 is not adjustable, as the <strong>machine</strong>s are equipped with<br />

an automatic oil pump <strong>and</strong> a tank.


[8.B1] OP37 Interval for service message TSC<br />

(time-dependent)<br />

for technical safety control E.02 “HELP"<br />

OP38 Service Interval<br />

(distance-dependent)<br />

"E 02" "HELP"<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 112 of 213<br />

optional settings<br />

On sports devices <strong>and</strong>/or active medical devices with power supply<br />

connection used for either commercial, industrial, institutional <strong>and</strong><br />

medical applications regular technical safety controls have to be<br />

performed by a trained <strong>and</strong> authorised service technician. Appropriate<br />

protocols <strong>and</strong> inspection instructions are available at h/p/cosmos<br />

service. To provide for this an error message "E 02" "HELP" for the<br />

TSC-service interval will appear in the display after expiration of a time<br />

period of 12 month.<br />

By calling up OP37 the remaining time period (in month) till the next<br />

TSC/maintenance/inspection is reported.<br />

Reset of the interval <strong>and</strong> the error report „E02“ is possible via option<br />

OP47.<br />

On dem<strong>and</strong> the TSC report interval can be changed via option OP37:<br />

Select OP37 <strong>and</strong> change with a or d. Confirm with c<br />

Option OP37 can also be used for the adjustment of the maintenance<br />

intervals in month, if the mileage is small.<br />

§ reports: TSC-Interval in month (St<strong>and</strong>ard = 12 month)<br />

Adjustment range 0 to 48 month<br />

h/p/cosmos recommends regular maintenance (interior cleansing,<br />

checking drive belts, etc.) carried out by an authorized technician at<br />

least every 5.000 km (for ladder ergometers: 500 km). For this reason<br />

every 5.000 km (500 km) the error message "E 02" "HELP" is<br />

displayed. For detailed information, descriptions <strong>and</strong> specifications<br />

about maintenance, please contact the h/p/cosmos service<br />

department. To provide for this an error message "E 02" "HELP" for<br />

the maintenance interval will appear. TSC controls have to be<br />

performed every 12 month the latest.<br />

Reset of the interval <strong>and</strong> the error report „E02“ is possible via option<br />

OP47.<br />

On dem<strong>and</strong> the service maintenance interval can be changed via<br />

option OP38.<br />

Call up OP38 <strong>and</strong> change with a or d. Confirm with c<br />

§ reports: service interval in km (St<strong>and</strong>ard = 5.000 / 500)<br />

Adjustment range 1.000 to 9.900 km or 100 to 5.000 km for the ladder<br />

ergometer.


[8.B1] OP39 Elevation: display as set point or true<br />

value<br />

(This option is only available for<br />

treadmills, not for a ladder-ergometer)<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 113 of 213<br />

optional settings<br />

0 = Display of elevation as set point<br />

1 = Display of elevation as true value rsp. flashing set point report, if<br />

the set point was changed with the keys <strong>and</strong> has not yet been<br />

reached. (Report: current elevation angle; recommended defaultsetting)<br />

OP40 Speed: display as set point or true value 0 = Display of speed as set point (Report: target speed)<br />

1 = Display of speed as true value rsp. flashing set point report, if the<br />

set point was changed with the keys <strong>and</strong> has not yet been reached.<br />

(Report: current speed; recommended default-setting)<br />

OP41 Elevation control after STOP<br />

0 = Elevation maintains at the present level after pressing the STOP<br />

(This option is only available for button (recommended; default-setting)<br />

treadmills, not for a ladder-ergometer). 1 = Elevation gets down to 0 % after pressing the STOP button.<br />

The setting of option 41 to value 1 effects the<br />

safety of the subject <strong>and</strong> of third parties. A downmovement<br />

of the elevation after pressing STOP<br />

can result in a jumpy reaction <strong>and</strong> in dangerous<br />

injuries <strong>and</strong> contusions.<br />

OP46 Display <strong>and</strong> setting of elevation for<br />

h/p/cosmos strongly recommends to keep the<br />

st<strong>and</strong>ard setting of this option. Stop means Stop<br />

for all movements <strong>and</strong> must not result in an active<br />

action of the <strong>running</strong> <strong>machine</strong>. It is only allowed<br />

to change option 41 after a written documented<br />

instruction about the danger warnings <strong>and</strong> after a<br />

written confirmation of the customer/user.<br />

$ indicates: Elevation of ladder ergometer, st<strong>and</strong>ard: 75 degree,<br />

ladder ergometer<br />

(only available for ladder ergometers,<br />

not for treadmills)<br />

adjustable from 0 to 180 degree<br />

OP47 Set back of oil- <strong>and</strong> service interval After a service with lubricating the treadmill <strong>and</strong> performing all<br />

maintenance work the oil <strong>and</strong> service interval can be set back with<br />

option number OP47.


[8.B1] OP48 Maximum speed during maximum<br />

activation of the frequency inverter.<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 114 of 213<br />

optional settings<br />

The maximum frequency inverter speed is achieved, if the MCU (within<br />

UserTerminal) sets the control voltage (analogue voltage for frequency<br />

inverter) at 10 volt, that is at maximum frequency inverter frequency.<br />

The maximum speed during maximum frequency inverter frequency is<br />

going to be set basically with the input of the correct device type at<br />

option number 3 at the administrator level. Option 48 is necessary at<br />

manual speed calibration or special speed transformation or drive<br />

motors (special speeds) only.<br />

When changing option 48, always pay attention to the correct setting of<br />

the options number 5 (maximum speed), number 11 (speed<br />

measurement; has to be switched on) <strong>and</strong> number 34 (speed<br />

calibration value).<br />

Examples for the different special speeds <strong>and</strong> the appropriate<br />

maximum inverter frequencies are to be found in list B3: List of device<br />

types.<br />

Procedure<br />

N Make sure that the correct device type is set at administrator<br />

option 3.<br />

N Go to administrator option 5 <strong>and</strong> set maximum possible speed.<br />

N Control at the manual mode the device with the key "+" to the<br />

maximum possible speed.<br />

N Check the analogue voltage at the MCU-FU-control with a<br />

voltmeter: it has to be 10 volt. Note down the voltage at the<br />

service report! (not possible with digital control via RS485)<br />

N Read off the speed <strong>and</strong> write down the value at the h/p/cosmos<br />

service report: OP 48: _____ km/h<br />

N Stop the <strong>running</strong> belt with the STOP key.<br />

N Go to administrator OPTION 48 <strong>and</strong> input the before measured<br />

value in km/h (maximum possible speed at full FU-activation).<br />

N Go again to administrator OPTION 05 <strong>and</strong> allow the desired /<br />

appropriate maximum speed.<br />

Info 1: If no speed scan (speed sensor) has to be set under option 11,<br />

the speed calibration value will be calculated automatically with option<br />

48.<br />

Info 2: Even if no speed scan is set, the speed calibration value has an<br />

internal significance. In that case the scanning impulses (actual<br />

impulse) are deduced or simulated directly from the analogue voltage<br />

<strong>and</strong> not measured via the speed sensor input.<br />

Range of adjustment: 1.00 to 99.99 km/h<br />

Fixed values selectable with Up- /Down-key: 20; 22; 22.5; 26; 30; 30.3;<br />

40; 40.9; 41.2; 45.3; 46; 50; 60; 80


[8.B1] OP49 Type of frequency inverter (FI) control Setting of FI control. Two types are available for selection:<br />

§ $ %<br />

Analogue 0 to 10 V AnA LOG<br />

Digital via RS-485 rS- 485<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 115 of 213<br />

optional settings<br />

OP51 Error statistics of the digital FI control The lower display row reports the occurred errors of three different<br />

error types.<br />

§ Number of NAK errors: the FI has sent an NAK (not<br />

acknowledged), i.e. the FI has not accepted the last comm<strong>and</strong><br />

(transmission error, …).<br />

$ Number of proof total errors: the proof total of the received FI<br />

messages is wrong (transmission error, …).<br />

% Number of Timeout errors: the FI has not answered within the<br />

Timeout period (60 ms).<br />

Note: At each emergency-stop the number of timeout errors is<br />

increased compulsorily by at least one numerator, because the<br />

transmission line to the FI is being interrupted by hardware.


[8.B1] OP89 System for elevation measurement<br />

(option not relevant for treadmills without<br />

elevation or ladder ergometers)<br />

Note:<br />

Normally the correct system for elevation<br />

measurement is set by selecting the<br />

correct device type (administrator option<br />

number 3). Only in very few cases <strong>and</strong><br />

for special models this option is needed.<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 116 of 213<br />

optional settings<br />

The following kinds of elevation measurement can be selected:<br />

N 0 = no elevation available.<br />

N 1 = elevation control: increment disc with light barrier 60 incr. per<br />

motor rotation or hall effect sensor with 60 incr. per motor<br />

rotation: elevation element with a 4 mm spindle for: 0 - 25%<br />

elevation, ( models QUASAR/PULSAR from 1991 to 2001)<br />

N 2 = elevation control: increment disc with light barrier 60 incr. per<br />

motor rotation or hall effect sensor with 60 incr. per motor<br />

rotation: elevation element with a 5 mm spindle for: 0 - 25%<br />

elevation, ( model h/p/cosmos saturn since 1994)<br />

N 3 = elevation control: increment disc with light barrier 60 incr. per<br />

motor rotation or hall effect sensor with 60 incr. per motor<br />

rotation: elevation element with a 5 mm spindle for: 0 - 35%<br />

elevation, ( models h/p/cosmos venus since 1997)<br />

N 4 = elevation control: linear motor for 0 - 24% elevation,<br />

increment sensor, lift range 240 mm (h/p/cosmos mercury<br />

models since 1997)<br />

N 5 = elevation control: linear motor for 0 - 28% elevation,<br />

increment sensor, lift range 300 mm (h/p/cosmos quasar 4.0<br />

since 2002)<br />

N 6 = elevation control: linear motor for 0 - 25% elevation,<br />

increment sensor, lift range 300 mm (h/p/cosmos pulsar 4.0<br />

since 2002)<br />

N 7 = Hydraulic-cylinder for 0 - 11 % elevation angle.<br />

lift range 430 mm (series PLM Mustang 2200)<br />

N 8 = Hydraulic-cylinder for 0 - 11 % elevation angle.<br />

lift range 302 mm (series PLM Trainer)<br />

N 9 = linear motor „HN“ + lifting arm. 0 - 25% elevation angle.<br />

lift range 252 mm (h/p/cosmos mercury since 2007)<br />

N 10 = linear motor „HN“ + lifting arm. -10…+18% elevation angle.<br />

lift range 300 mm (h/p/cosmos quasar since 2007)<br />

N 11 = linear motor „HN“ + lifting arm. -10…+15% elevation angle.<br />

lift range 300 mm (h/p/cosmos pulsar since 2007)<br />

N 12 = increment disc with light barrier or inductive sensor<br />

60 incr. per motor rotation: elevation element with 8mm spindle.<br />

-4…25 % elevation<br />

(series h/p/cosmos saturn 450/300rs as of 2008)


file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 117 of 213<br />

optional settings<br />

[8.B1] OP90 Load or save default settings With this option all option values (exceptions see below) can be saved<br />

or reloaded.<br />

With a or d the loading or saving is selected. After confirmation with<br />

c the safety request has to be confirmed again with c .<br />

§ $ %<br />

SAVE dEF AULt<br />

ArE YOU SUrE<br />

SAV Ed<br />

LOAD dEF AULt<br />

ArE YOU SUrE<br />

LOAd Ed<br />

If default values are supposed to be loaded without having been saved<br />

before following message will appear <strong>and</strong> no change can be<br />

performed.<br />

nO d EF S AVEd<br />

The key SW1 on the MCU 5 assembly board performs the same action<br />

as the loading of default values with the following exception: if there<br />

are no default values saved, the default values of the firmware will be<br />

taken over.<br />

Following option values are not h<strong>and</strong>led as default values <strong>and</strong> remain<br />

unchanged:<br />

User - OP 02: Total distance<br />

User - OP 03: Total time of operation<br />

User - OP 04: Total time of use<br />

Administrator-OP 10: Treadmill serial number<br />

Administrator-OP 91: MCU total time of operation<br />

Administrator-OP 92: Total distance<br />

Administrator-OP 93: Total time of operation<br />

Administrator-OP 94: Total time of use<br />

Administrator-OP 95: MCU serial number<br />

OP91 Total time of operation MCU (h) § <strong>and</strong> $ report: hours of operation<br />

= total time of switch on MCU % reports: h<br />

This value can not be changed!<br />

OP92* Total distance (km) § <strong>and</strong> $ report: total distance in km<br />

% reports km<br />

Difference to user option 02: the displayed value can also be changed.


file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 118 of 213<br />

optional settings<br />

[8.B1] OP93* Total time of operation (h)<br />

§ <strong>and</strong> $ report: hours of operation<br />

= St<strong>and</strong>-By-time including time of use of % reports: h<br />

the motor/ <strong>running</strong> belt<br />

Difference to user option 03: the displayed value can also be changed.<br />

OP94* Total time of use (h)<br />

§ <strong>and</strong> $ report: hours of operation<br />

= time of use of motor / <strong>running</strong> belt only % reports: h<br />

Difference to user option 04: the displayed value can also be changed.<br />

OP95 MCU Serial number Report <strong>and</strong> entering of 8-digit serial number with the following format:<br />

YYMM NNNN<br />

YY: 2-digit number of year of production lot<br />

MM: month of production lot<br />

NNNN: 4-digit serial number<br />

OP98 Setting of limitation of usage time Secret option<br />

OP99 Test-Profile 99 for endurance testing in With this option a special test profile (profile number 99) for endurance<br />

the factory / service<br />

testing in the factory or in service can be activated. This test profile is<br />

not for normal use, so it has to be de-activated again after using. If the<br />

device is switched off an on again, profile number 99 is deactivated<br />

automatically.<br />

Details of test profile 99:<br />

Elevation from 0 to maximum elevation every 20 min.<br />

Speed 3.0 km/h for 2 hours (for ladder ergometer: 3,0 m/min for 2<br />

hours)<br />

Speed 5.0 km/h for 3 hours (for ladder ergometer: 5,0 m/min for 3<br />

hours)<br />

Speed 8.0 km/h for 8 hours (for ladder ergometer: 8,0 m/min for 8<br />

hours)<br />

STOP after 13 hours.<br />

Adjustable settings:<br />

0 = Test-Profile 99 locked/ de-activated<br />

1 = Test-Profile 99 activated<br />

*Note: It is not allowed to change the total time of operation or the total distance of the <strong>running</strong>-<strong>machine</strong> to a value which<br />

does not match to the real value of the device.


[8.B2] St<strong>and</strong>ard settings of administrator options<br />

[8.B2] Option Description St<strong>and</strong>ard setting Adjustable Range<br />

OP 01 Set back of error messages E 20, E 21,<br />

E 30, E 31, E 32, E 41, E 50 <strong>and</strong> E 51<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 119 of 213<br />

optional settings<br />

OP 02 OEM-Code depending on the model 0 = h/p/cosmos (St<strong>and</strong>ard)<br />

1 = JAEGER / VIASYS /<br />

Cardinal Health<br />

2 = Ergo-Fit (1997-98)<br />

3 = Proxomed / Kardiomed<br />

4 = KISTLER Gaitway<br />

6 = COSMED<br />

9 = SCHILLER<br />

OP 03 Selection of device type<br />

- see list [6.B3] -<br />

depending on the model (depending on the model)<br />

OP 04 Maximum speed for reverse belt rotation depending on the device type in<br />

OP03<br />

OP 05 Maximum Speed<br />

depending on the device type in<br />

OP03<br />

1.0 km/h .... max. speed<br />

1.0 km/h .... max. speed<br />

OP06 Max. acceleration time in seconds Increment 131 = 5 sec. Increments: 255 ... 1<br />

Seconds: 2.6 ... 655.4<br />

OP 07 Maximum Elevation<br />

depending on the device type in<br />

OP03<br />

0 .... max. elevation<br />

OP 08 Heart Rate Display Interval 0 = beat to beat 0 or 1 ... 9 seconds<br />

OP 09 Display Test for all displays<br />

OP 10 Enter the serial number of the device 300… 12-digit serial number<br />

OP 11 Activating <strong>and</strong> de-activating speed depending on the device type in 0 = without speed measurement<br />

measurement<br />

OP03<br />

1 = with speed measurement<br />

OP 12 Setting of allowance for speed-signal- 6 %<br />

0 = no comparison<br />

comparison<br />

This value does not reflect the 6 = 6 % deviation allowed<br />

(Measurement / Control)<br />

speed accuracy of the <strong>running</strong>- 8 = 8 % deviation allowed<br />

belt.<br />

10 = 10 % deviation allowed<br />

OP13 Motor brake OFF OFF = not active<br />

ON = active<br />

OP 14 Maximum acceleration time for RS232 Increment 131 = 5 sec Increments: 255 ... 1<br />

Seconds: 2.6 ... 655.4<br />

OP 16 Activating <strong>and</strong> de-activating heart rate (depending on the model) 0 = HF measurement de-activated<br />

measurement<br />

1 = HF measurement activated,<br />

HFU1 + RE06REC<br />

2 = HF measurement activated,<br />

HFUi;<br />

auto = autom. detection<br />

OP 33 Distance between rungs (only for ladder<br />

ergometer)<br />

254 = 25.4 cm Distance between rungs in 0.1 cm<br />

OP 34 Manual Speed Calibration<br />

(depending on the device type;<br />

OP 03)<br />

Range from: 0 to 65535


[8.B2] OP 35 Oil Interval<br />

OP 37 Interval for service message TSC<br />

(time-dependent)<br />

for technical safety control E.02 “HELP"<br />

1.000 km (treadmills)<br />

100 km (ladder ergometer)<br />

0 = no supervision<br />

100 ... 5.000 km<br />

12 month 0 … 48 month<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 120 of 213<br />

optional settings<br />

OP 38 Service Interval (distance-dependent) 5.000 km (treadmill)<br />

1.000 ... 9.999 km (treadmill)<br />

"E 02" "HELP"<br />

500 km (ladder ergometer) 100... 5.000 km (ladderergometer)<br />

OP 39 Elevation: display as set point or true 1 (true value) 0 = set point<br />

value<br />

1 = true value<br />

OP 40 Speed: display as set point or true value 1 (true value) 0 = set point<br />

1 = true value<br />

OP 41 Elevation control after STOP 0 (Elevation maintains) 0 = Elevation maintains<br />

1 = Elevation gets down to 0 %<br />

OP 46 Display <strong>and</strong> setting of elevation for<br />

ladder ergometer<br />

75 (degree) 0 … 180 degree<br />

OP 47 Set back of oil- <strong>and</strong> service interval<br />

OP 48 Running-Belt-Speed at maximum (depending on the device type; (depending on the device type;<br />

frequency inverter speed<br />

OP 03)<br />

OP 03)<br />

1.00 … 99.99 km/h<br />

OP 49 Type of FI Control RS-485 Analogue, RS-485<br />

OP 51 Error statistics of the digital FI-control n.a. 0 to 9999 per type of error<br />

OP 89 System for elevation measurement (depending on the device type; (depending on the device type;<br />

OP 03)<br />

OP 03)<br />

OP 90 Load or save default setting n.a. n.a.<br />

OP 91 Total time of operation MCU<br />

(incl. st<strong>and</strong> by)<br />

n.a. 0 ... 1.193.046 h<br />

OP 92 Total Distance n.a. 0 ... 4.294.976 km<br />

OP 93 Total time of operation (incl. st<strong>and</strong>-by) n.a. 0 ... 1.193.046 h<br />

OP 94 Total time of use (motor + belt) n.a. 0 ... 1.193.046 h<br />

OP 95 MCU serial number n.a. 0000 0000 to 9999 9999<br />

OP 98 Setting for limitation of time of use n.a. n.a.<br />

OP 99 Test-Profile 99 for endurance testing in<br />

the factory / service<br />

0 = Test 99 de-activated 0 = Testprofile 99 de-activated<br />

1 = Testprofile 99 activated


[8.B3] List of <strong>running</strong> <strong>machine</strong>s types<br />

Model view<br />

Model series<br />

track in cm<br />

Device type<br />

Code<br />

Reverse speed<br />

Speed<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 121 of 213<br />

Elevation<br />

Speed sensor<br />

0 = no, 1 = yes<br />

Increment<br />

for 10 m<br />

Maximum inverter<br />

speed<br />

optional settings<br />

OP03 OP04 OP05 OP07 OP11 OP34 OP48 OP89 Ø<br />

User defined 0.1<br />

1.1 -5.0 km/h 10.0 km/h 24.0 % 1 8490 10.0 km/h 4<br />

1.2 -5.0 km/h 20.0 km/h 24.0 % 0 5988 20.0 km/h 4<br />

1.3 -5.0 km/h 22.0 km/h 24.0 % 0 5988 22.0 km/h 4<br />

150 / 50<br />

1.4<br />

1.5<br />

1.6<br />

1.7<br />

-5.0 km/h<br />

-5.0 km/h<br />

-5.0 km/h<br />

-5.0 km/h<br />

30.0 km/h<br />

10.0 km/h<br />

20.0 km/h<br />

22.0 km/h<br />

24.0 %<br />

25.0 %<br />

25.0 %<br />

25.0 %<br />

0<br />

1<br />

0<br />

0<br />

4165<br />

8490<br />

5988<br />

5988<br />

30.0 km/h<br />

10.0 km/h<br />

20.0 km/h<br />

22.0 km/h<br />

4<br />

9<br />

9<br />

9<br />

80 mm<br />

1.8 -5.0 km/h 30.0 km/h 25.0 % 0 4165 30.0 km/h 9<br />

2.1 -5.0 km/h 10.0 km/h 0.0 % 1 8490 10.0 km/h 0<br />

150 / 50<br />

2.2<br />

2.3<br />

2.4<br />

-5.0 km/h<br />

-5.0 km/h<br />

-5.0 km/h<br />

20.0 km/h<br />

22.0 km/h<br />

30.0 km/h<br />

0.0 %<br />

0.0 %<br />

0.0 %<br />

0<br />

0<br />

0<br />

5988<br />

5988<br />

4165<br />

20.0 km/h<br />

22.0 km/h<br />

30.0 km/h<br />

0<br />

0<br />

0<br />

3.1 no limit 40.0 km/h 35.0 % 1 2617 40.0 km/h 3<br />

200 / xx<br />

3.2<br />

3.3<br />

no limit<br />

no limit<br />

50.0 km/h<br />

60.0 km/h<br />

35.0 %<br />

35.0 %<br />

1<br />

1<br />

2617<br />

2617<br />

50.0 km/h<br />

60.0 km/h<br />

3<br />

3<br />

3.4 no limit 80.0 km/h 35.0 % 1 2617 80.0 km/h 3<br />

4.1 no limit 40.0 km/h 25.0 % 1 2617 40.0 km/h 2<br />

4.2<br />

4.3<br />

no limit<br />

no limit<br />

50.0 km/h<br />

60.0 km/h<br />

25.0 %<br />

25.0 %<br />

1<br />

1<br />

2617<br />

2617<br />

50.0 km/h<br />

60.0 km/h<br />

2<br />

2<br />

180 mm<br />

250 / xx 4.4 no limit 80.0 km/h 25.0 % 1 2617 80.0 km/h 2<br />

300 / xx 4.5 no limit 40.0 km/h 27.0 % 1 2617 40.0 km/h 2<br />

450 / xx 4.6 no limit 50.0 km/h 27.0 % 1 2617 50.0 km/h 2<br />

4.7 no limit 60.0 km/h 27.0 % 1 2617 60.0 km/h 2<br />

4.8 no limit 80.0 km/h 27.0 % 1 2617 80.0 km/h 2<br />

n.a. n.a. 40.0 km/h 25.0 % 1 1040 40.0 km/h 480 mm<br />

5.1 -5.0 km/h 10.0 km/h 28.0 % 1 7800 10.0 km/h 5<br />

5.2 -5.0 km/h 22.0 km/h 28.0 % 1 4495 22.5 km/h 5<br />

5.3 -5.0 km/h 25.0 km/h 28.0 % 1 4495 26.0 km/h 5<br />

5.4 -5.0 km/h 30.0 km/h 28.0 % 1 4010 30.3 km/h 5<br />

5.5 -5.0 km/h 40.0 km/h 28.0 % 1 2874 41.2 km/h 5<br />

170 / 65<br />

5.6<br />

5.7<br />

5.8<br />

-5.0 km/h<br />

-5.0 km/h<br />

-5.0 km/h<br />

45.0 km/h<br />

10.0 km/h<br />

22.0 km/h<br />

28.0 %<br />

25.0 %<br />

25.0 %<br />

1<br />

1<br />

1<br />

2874<br />

7800<br />

4495<br />

45.3 km/h<br />

10.0 km/h<br />

22.5 km/h<br />

5<br />

5<br />

5<br />

5.9 -5.0 km/h 25.0 km/h 25.0 % 1 4495 26.0 km/h 5<br />

5.10<br />

5.11<br />

-5.0 km/h<br />

-5.0 km/h<br />

30.0 km/h<br />

40.0 km/h<br />

25.0 %<br />

25.0 %<br />

1<br />

1<br />

4010<br />

2874<br />

30.3 km/h<br />

41.2 km/h<br />

5<br />

5<br />

100 mm<br />

5.12 -5.0 km/h 45.0 km/h 25.0 % 1 2874 45.3 km/h 5<br />

6.1 -5.0 km/h 10.0 km/h 0.0 % 1 7800 10.0 km/h 0<br />

6.2 -5.0 km/h 22.0 km/h 0.0 % 1 4495 22.5 km/h 0<br />

170 / 65<br />

6.3<br />

6.4<br />

6.5<br />

-5.0 km/h<br />

-5.0 km/h<br />

-5.0 km/h<br />

25.0 km/h<br />

30.0 km/h<br />

40.0 km/h<br />

0.0 %<br />

0.0 %<br />

0.0 %<br />

1<br />

1<br />

1<br />

4495<br />

4010<br />

2874<br />

26.0 km/h<br />

30.3 km/h<br />

41.2 km/h<br />

0<br />

0<br />

0<br />

6.6 -5.0 km/h 45.0 km/h 0.0 % 1 2874 45.3 km/h 0<br />

190 / 65<br />

7.1<br />

7.2<br />

-5.0 km/h<br />

-5.0 km/h<br />

40.0 km/h<br />

45.0 km/h<br />

25.0 %<br />

25.0 %<br />

1<br />

1<br />

2874<br />

2874<br />

40.9 km/h<br />

46.0 km/h<br />

6<br />

6<br />

For any questions about the correct device type, contact the h/p/cosmos service before doing any changes.<br />

Elevation system<br />

Code<br />

Drive roller (shaft)<br />

diameter


[9.]Maintenance <strong>and</strong> safety inspections<br />

maintenance & safety inspections<br />

N The maintenance <strong>and</strong> repair of the devices (also opening of the device) has to be performed only<br />

by the service engineers authorized <strong>and</strong> certified by h/p/cosmos, preferrably within the scope of a<br />

maintenance contract.<br />

N In case of any detected <strong>and</strong>/or assumed malfunctions <strong>and</strong>/or defects <strong>and</strong>/or unreadable safety<br />

warning labels the device has to be taken out of operation immediately, the device has to be<br />

marked <strong>and</strong> secured against operation <strong>and</strong> the supplier <strong>and</strong> authorized service personnel has to<br />

be informed in writing immediately.<br />

N Disregard of warnings, disregard of intended <strong>and</strong> forbidden use, safety precautions <strong>and</strong> also<br />

unauthorized or lack of maintenance <strong>and</strong>/or regular safety checks may lead to injuries or death<br />

<strong>and</strong>/or can damage the device <strong>and</strong> will result in loss of any liability <strong>and</strong> warranty.<br />

N Before intervening in the device for safety reasons switch the <strong>running</strong> <strong>machine</strong> off <strong>and</strong> pull the<br />

mains plug out.<br />

N During all maintenance works <strong>and</strong> safety tests make sure that no third parties are directly or<br />

indirectly in contact with the device under test <strong>and</strong>/or the technician performing the test. Keep a<br />

safety area of 2 m radius clear.<br />

Some <strong>running</strong> <strong>machine</strong> models are not equipped with an UserTerminal, so there is no keyboard or display available on these<br />

<strong>running</strong> <strong>machine</strong>s. It is only possible to control these <strong>running</strong> <strong>machine</strong>s via interface. With this interface you can use different<br />

external equipment, e.g. ECG, Ergospirometry, PC with h/p/cosmos software h/p/cosmos para graphics ® or h/p/cosmos para<br />

control ® . You can find important safety precautions <strong>and</strong> a complete list of all compatible external equipment in the chapter<br />

“compatibility RS232 interface” <strong>and</strong> in the internet www.coscom.org.<br />

For these models an additional keyboard or an external UserTerminal for remote control via RS232 is available as an option.<br />

If using an additional keyboard you might need to install additional wiring <strong>and</strong> a connector depending on the model.<br />

UserTerminal Touchpanel retro fit kit for h<strong>and</strong>rail, diameter: 60 mm [cos14569-02]<br />

Additional keyboard with 6 keys with 2 meter cable [cos10106]<br />

See website: http://www.h-p-cosmos.com/en/options/additional_user_terminal.htm<br />

For maintenance <strong>and</strong> diagnosis we recommend the use of PC Software h/p/cosmos para control ® (Freeware) which is<br />

available for download at: www.h-p-cosmos.com/en/software/para_control.htm<br />

[9.A] Preventive maintenance<br />

The h/p/cosmos authorised service engineers are happy to help you in the case of occurring problems.<br />

A preventive maintenance can avoid problems in the future <strong>and</strong> is indispensable for the safety of such technical devices.<br />

Therefore ask for an annual preventive maintenance contract, which is highly recommended by the manufacturer, from<br />

our service department. Some basic regular maintenance <strong>and</strong> regular safety checks are obligatory!<br />

Before switching the device on always check the circuit cable, plug, outlet socket <strong>and</strong> circuit entry of the device.<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 122 of 213


[9.B] Immediate maintenance<br />

maintenance & safety inspections<br />

Immediate maintenance is necessary if:<br />

N the device has been under high mechanical stress (push, power supply cable <strong>and</strong>/or interface cable defect through<br />

driving over it or pulling it)<br />

N fluid has got into the device<br />

N cable <strong>and</strong>/or connector plug have been damaged<br />

N coverings <strong>and</strong>/or safety warnings have fallen off<br />

N connections made of rubber show cracks (mainly the <strong>running</strong> belt <strong>and</strong> the drive belt)<br />

N the <strong>running</strong> belt does not run concentric<br />

N the <strong>running</strong> belt is not lubricated well<br />

N a defekt or malfunction of the device has been detected or is suspected<br />

Only a properly <strong>and</strong> regularly serviced device is safe. The maintenance of the devices has to be performed only by the<br />

service engineers authorized <strong>and</strong> certified by h/p/cosmos, preferrably within the scope of a maintenance contract.<br />

[9.C] Regular inspections/examinations<br />

For use in sports <strong>and</strong> medical areas as well as for private, public <strong>and</strong> military applications refer<br />

to the date at the inspection sticker at your device.<br />

To keep the condition of the device in due order, examinations have to be performed repeatedly<br />

according to the local laws <strong>and</strong> requirements in your country (e.g. in Germany based on BGV<br />

A3, regulations for prevention of accidents as well as safety requirement inspections in<br />

accordance with MDD <strong>Medical</strong> Device Directive, etc.).<br />

For h/p/cosmos sports <strong>and</strong> medical <strong>running</strong> <strong>machine</strong>s <strong>and</strong> ladder ergometers a maintenance interval respectively<br />

technical safety checks of one year has been set. These examinations are only to be performed by trained <strong>and</strong><br />

authorized electric technicians.<br />

The main inspection sticker on the device (e.g. <strong>running</strong> <strong>machine</strong>) certifies also the inspection of the optional equipment<br />

<strong>and</strong> the accessories. However, inspection intervals for optional equipment <strong>and</strong> accessories (e.g. rope of h/p/cosmos<br />

airwalk unweighting system, chest belt for safety harness, compressors, etc.) can deviate significantly from inspection<br />

intervals of the main device. Read respective manuals for details.<br />

For all required measurements <strong>and</strong> control steps consider also local requirements in your country <strong>and</strong> see<br />

detailed instruction <strong>and</strong> protocol form order no. [cos11690xx].<br />

Additionally to local requirements following examinations are to be performed <strong>and</strong> are to be documented in the detailed<br />

instruction <strong>and</strong> protocol form order no. [cos11690xx]:<br />

[9.C1] Visual inspections<br />

N Visual inspection for damages at device <strong>and</strong> complete accessories. Internal space, engine compartment,<br />

connecting lead <strong>and</strong> the right position of the tensile relief <strong>and</strong> plug, ground wire connections, etc.<br />

N Visual inspection of mechanics <strong>and</strong> wearing parts: driving belt, tension roller, <strong>running</strong> belt, lifting element with<br />

fixing bolts, welding seams at the frame, tight fitting of all screws <strong>and</strong> nuts. Pay attention to the appropriate<br />

maintenance list.<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 123 of 213


maintenance & safety inspections<br />

N Clean device <strong>and</strong> engine compartment/internal space. Remove dust <strong>and</strong> dirt from the cooler openings of the<br />

fan housing <strong>and</strong> the cooling air ducts of the drive motor, as well as from the ventilation slots <strong>and</strong> the cover of<br />

perforated plate of the frequency inverter.<br />

N Attach hazard notes completely <strong>and</strong> well legible according to the instructions. Check the warning signs,<br />

protective covers, step platform, engine hood <strong>and</strong> plastic-cover installation channel in the engine compartment<br />

for existence <strong>and</strong> for damage <strong>and</strong> replace as necessary.<br />

N All protective-resistance connectors, which can be reached from the outside, must examined for their correct<br />

value <strong>and</strong> correct labelling.<br />

N Check capture slot at the rear end <strong>and</strong> at reverse belt rotation in the front as well <strong>and</strong> adjust as necessary.<br />

Capture slot < 8 mm. According to regulation draft 60601-2-xx © IEC:200X 62D/479/NWIP 2003-05-18 <strong>and</strong><br />

according to EN 957-1. See chapter “test-finger“, which has a diameter of 9.5mm (+0.1) at the front.<br />

[9.C2] Protective Earth Resistance RPE measurement<br />

[9.C3] Isolation Resistance RISO measurement<br />

N Resistance between housing <strong>and</strong> ProtectiveEarth-<br />

Connection.<br />

N The low resistance pass is to be controlled according to<br />

VDE 0701/0702 (sports <strong>and</strong> fitness <strong>machine</strong>s) or VDE<br />

0751/IEC 60601-1 (medical devices) by the protectiveresistance<br />

measurement with a measuring device for the<br />

protective-earth-resistance measurement.<br />

N The connecting lead is to be moved while measuring for<br />

at least 5 sec. If the resistance changes hereby it is highly<br />

probable that the cable or the connectors have been<br />

damaged. In this case the cable has to be replaced <strong>and</strong><br />

the device has to be repaired immediately.<br />

N Resitance between „briged“ L+N <strong>and</strong> ProtectiveEarth-<br />

Connection.<br />

N Make sure that all isolations, that are under stress of the<br />

mains voltage, are being seized. All switches <strong>and</strong><br />

contactors should be connected.<br />

N The measuring is to be performed with measuring<br />

devices for the isolation-resistance measurement<br />

according to VDE 0701/0702 (sports <strong>and</strong> fitness<br />

<strong>machine</strong>s) or VDE 0751/IEC 60601-1 (medical devices).<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 124 of 213


[9.C4] Equivalent (alternative) leakage current IEDL measurement<br />

[9.C5] Leakage current measurement<br />

direct method: measurement in PE - line<br />

difference current method: supply/return line<br />

maintenance & safety inspections<br />

N Impedance measurement, indicating the current in the<br />

Protective¬Earth¬Cable<br />

N The measuring is to be performed by a measuring device<br />

for leakage current measuring according to VDE<br />

0701/0702 (sports <strong>and</strong> fitness <strong>machine</strong>s) or VDE<br />

0751/IEC 601-1 (medical devices).<br />

N The measurement is equivalent to the single fault earth<br />

leakage current based on IEC 60601.<br />

N The measuring is to be performed by a measuring device<br />

for leakage current measuring according to VDE<br />

0701/0702 (sports <strong>and</strong> fitness <strong>machine</strong>s) or VDE<br />

0751/IEC 60601-1 (medical devices).<br />

N The leakage current of the equipment is to be measured<br />

under operating conditions, if -despite closed switchingnot<br />

all isolations are seized.<br />

N Measurements must be made on both poles.<br />

N The unit to be tested must be isolated against earth<br />

potential.<br />

N The measurement is equivalent to the earth leakage<br />

current with grounded application parts based on<br />

IEC 60601.<br />

N During the direct method the device under test<br />

is temporarily without PE-connection. Especially devices<br />

without potential-isolation-transformer (sports devices)<br />

have temporarily a voltage of approx. 120 volts on the<br />

metal housing <strong>and</strong> frame parts. Do not touch the device<br />

under test during this measurement!<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 125 of 213


[9.C6] Electric Safety Tester <strong>and</strong> measurements<br />

maintenance & safety inspections<br />

Picture: example for electric safety tester based on IEC 60601-1.<br />

For all measurements read operation manuals of the measurement<br />

devices carefully <strong>and</strong> verify the values <strong>and</strong> intervals with the local<br />

guidelines <strong>and</strong> laws. An appropriate inspection record for all<br />

measurements <strong>and</strong> instructions are available at the manufacturers. See<br />

also detailed instruction <strong>and</strong> protocol form order no. [cos11690xx]. At<br />

first installation at the customer’s site <strong>and</strong> before first commissioning of<br />

the <strong>running</strong> <strong>machine</strong> device, the „first measured values“ have to be<br />

determined <strong>and</strong> recorded on a special protocol [cos11690xx] <strong>and</strong><br />

marked with „first measured values“.<br />

During each preventive maintenance <strong>and</strong>/or regular maintenance <strong>and</strong>/or also after each repair work on the device<br />

(even if it was only mechanical repair work!) all above mentioned electric safety measurements <strong>and</strong> checks (see<br />

protocol [cos11690xx]) have to be performed again. All values have to be compared with the „first measured<br />

values“ <strong>and</strong> allowed range of tolerance specified according to VDE 0701/0702 (sports <strong>and</strong> fitness <strong>machine</strong>s) or<br />

VDE 0751/IEC 60601-1 (medical devices).<br />

In case any or more of the measured values is/are not within the allowed range of tolerance specified according to<br />

VDE 0701/0702 (sports <strong>and</strong> fitness <strong>machine</strong>s) or VDE 0751/IEC 60601-1 (medical devices), the device has to be<br />

repaired until the values are within the specified allowed range of tolerance. In case the device can not be repaired<br />

accordingly <strong>and</strong> immediately, the device has to be marked <strong>and</strong> secured against operation <strong>and</strong> the supplier <strong>and</strong><br />

authorized service personnel has to be informed in writing.<br />

[9.C7] Building installation: Electric Checks, Protective Earth Function, RCD, Leakage Current<br />

The customer/operator has the operator duty to check the building installation (electrical installations <strong>and</strong><br />

stationary installations) in regular intervals according to regulation BGV A3 of the professional association<br />

respectively information of legal accident insurance every 4 years, on their own response <strong>and</strong> expenses, for<br />

functions <strong>and</strong> safety of the complete electrical building installation. At operating sites or rooms of special function -<br />

such as e.g. climate chamber, pressure chamber, installation with special endangering, etc. - the interval is<br />

prescribed one year or shorter if applicable.<br />

In any case use ground wire plugs with tested ground wires only. Existent earth leakage current protection<br />

switches (so called RCD or "residual-current-operated protective device") at the building installation must be tested<br />

by the customer/operator monthly (button at RCD switch in fuse box of the building installation) for correct <strong>and</strong><br />

proper function. As a rule the RCD switches at the fuse box are labelled with "press monthly".<br />

Normally a button at the RCD switch has to be pressed for testing (not the "main h<strong>and</strong>le"). With that an earth<br />

leakage current is simulated. The RCD switch must then switch-off the complete electricity supply. These tests<br />

must be carried out by the customer/operator at a time, when the normal operation is not disturbed, all electrical<br />

devices <strong>and</strong> computers are switched off <strong>and</strong> nobody is endangered.<br />

Due to the test of the RCD switch it possibly comes, depending on the circuit of the building electricity, to a switchoff<br />

of the complete or a part of the power supply of the building.<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 126 of 213


maintenance & safety inspections<br />

In this connection it is important to stress, that these are not regulations, st<strong>and</strong>ards <strong>and</strong> safety checks specifically<br />

for h/p/cosmos <strong>running</strong> <strong>machine</strong>s, but current regulations for all electrical devices <strong>and</strong> building installations<br />

worldwide <strong>and</strong> applicable for all electrical devices with metal housing.<br />

[9.D] Lubrication of the <strong>running</strong> belt/<strong>running</strong> surface<br />

Take care not to touch the moving <strong>running</strong> belt. This maintenance should be supervised by a<br />

second person, who could press the emergency button in the case of danger.<br />

The lubrication level must be checked periodically by the user. Procedure: Unplug <strong>running</strong> <strong>machine</strong> from power supply,<br />

place h<strong>and</strong> between <strong>running</strong> belt <strong>and</strong> <strong>running</strong> surface, ensure that surfaces are oiled <strong>and</strong> not dry. Also a special "tissue<br />

test" (order no. [cos14379]) is available at h/p/cosmos as an alternative way to check lubrication level of the <strong>running</strong> belt<br />

<strong>and</strong> deck. Use of bikes, wheelchairs or skis on the <strong>running</strong> <strong>machine</strong> as well as environmental conditions will affect the<br />

lubrication interval! Running <strong>machine</strong>s with automatic oiling systems (i.e. venus r <strong>and</strong> saturn r) must also be controlled<br />

periodically by the user to ensure correct lubrication levels (this can be compared to the need to check a car’s engine oil<br />

periodically as oil usage depends on driving style an can vary dramatically) as the type <strong>and</strong> frequency of use can raise<br />

or reduce the oil usage. In the event that either dry chafing noises are heard during use or that oil is noticed on the<br />

<strong>running</strong> deck h/p/cosmos must be contacted immediately. An h/p/cosmos authorized technical service engineer will then<br />

advise what corrective action needs to be taken. Only use original h/p/cosmos special oil. The use of other oils or the<br />

ignoring of the regular lubrication checks can lead to damage of the <strong>running</strong> belt, <strong>running</strong> deck or the lubrication system<br />

<strong>and</strong> a reduction of warranty liability. h/p/cosmos recommends a maintenance contract with an authorised h/p/cosmos<br />

service partner. Routine maintenance does not reduce the necessity to regularly control the lubrication levels. Regular<br />

maintenance <strong>and</strong> maintenance of the correct lubrication levels prolong the life of the <strong>running</strong> belt <strong>and</strong> <strong>running</strong> deck<br />

dramatically!<br />

[9.D1] Devices with UserTerminal <strong>and</strong> without automatic oil-pump<br />

The devices with UserTerminal are equipped with an automatic oil indication in the UserTerminal display. On<br />

st<strong>and</strong>ard setting every 1000 km (special device might have different intervals) the word OIL is flashing in the<br />

display. After lubrication the oil message has to be terminated with "OPTION 01". The device has no sensor <strong>and</strong><br />

the oil message will not terminate automatically after lubrication! The lubrication of the <strong>running</strong> belt <strong>and</strong> <strong>running</strong><br />

surface is done with the enclosed, to the extent of delivery, accessories (bottle of lubricant <strong>and</strong> 10ml syringe).<br />

Use the provided lubricant only. Other available oils <strong>and</strong><br />

lubricants can damage the <strong>running</strong> belt <strong>and</strong> <strong>running</strong> surface<br />

<strong>and</strong> therefore lead to a breakdown of the <strong>running</strong> <strong>machine</strong>.<br />

Lubricant available on request at h/p/cosmos. The <strong>running</strong><br />

belt should be lubricated after 1000 km (special device<br />

might have different intervals) at the latest or/<strong>and</strong> after<br />

hearing dry grinding noises during operation.<br />

The oil-filling hole [1] is placed at the right h<strong>and</strong> side in front<br />

of the motor hood [2] between the footboard [3] <strong>and</strong> the<br />

<strong>running</strong> belt [4]. The lower surface (gliding on the deck<br />

surface) of the <strong>running</strong> belt has a fabric construction <strong>and</strong><br />

accumulates the oil until the next interval is due.<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 127 of 213


maintenance & safety inspections<br />

Oiling amount<br />

3 syringes each filled with 10 ml silicone oil. (special devices might have different oiling amounts - see delivery<br />

papers) <strong>and</strong> 1 syringe filled with air (to empty the oil line in the <strong>running</strong> <strong>machine</strong>).<br />

Procedure<br />

N Start the <strong>running</strong> <strong>machine</strong> at a max. speed of 5 km/h.<br />

N Fill the syringe with 10ml of the lubricant.<br />

N Launch the syringe into the oil-filling hole <strong>and</strong> press the<br />

contents slowly in.<br />

N Fill the syringe again with the lubricant <strong>and</strong> slowly press<br />

the contents into the oil-filling hole.<br />

N Fill the syringe again with the lubricant <strong>and</strong> slowly press<br />

the contents into the oil-filling hole.<br />

N Fill the syringe again with air <strong>and</strong> slowly press the air<br />

into the oil-filling hole. This is to empty the small oiltube<br />

of the <strong>running</strong> <strong>machine</strong>.<br />

N Clean the area around the oil-filling-hole in order to make it free of overrunned oil.<br />

N Let the <strong>running</strong> <strong>machine</strong> operate for a few minutes at the speed of 5 km/h, so that the oil can spread itself on<br />

the inside of the <strong>running</strong> belt. Walk for 2 minutes at very low speed (ca. 2 km/h) <strong>and</strong> vary your positions on the<br />

<strong>running</strong> belt, so that the oil under the belt is spread evenly.<br />

N Check the position of the <strong>running</strong> belt after the lubrication. The belt should be in a centre position on the rear<br />

roller (tail shaft). In case the belt needs to be adjusted please follow the separate instruction.<br />

N After lubrication the oil message has to be terminated with "OPTION 01" (see optional functions). The device<br />

has no sensor <strong>and</strong> the oil message will not terminate automatically after lubrication. Refer to the procedure<br />

described at „Set back of oil-message“ in this chapter.<br />

This oil type must not be used for <strong>running</strong>-decks for cycle<strong>and</strong><br />

wheelchair - application (e.g. saturn 300/100r), <strong>and</strong><br />

not for oil-tanks with automatic oil-pump.<br />

Special oil 0.25 l - bottle<br />

for sliding plate with manual lubrication<br />

[cos000200008001]<br />

Syringe 10 ml for lubrication of sliding plate<br />

[cos12181]<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 128 of 213


[9.D2] Lubricant test<br />

maintenance & safety inspections<br />

The condition of the oil film between <strong>running</strong> surface <strong>and</strong> <strong>running</strong> belt can be established with the lubricant-test<br />

[cos14379] which is available at h/p/cosmos. The lubricant-test consists of a pre-wheighed commercial paper<br />

tissue. The tissue needs to be pulled through the gap between <strong>running</strong> belt <strong>and</strong> <strong>running</strong> surface in a defined lenght<br />

in order to absorb a certain quantity of lubricant. Based on previous experience h/p/cosmos is able to evaluate the<br />

quantity of the oil film by measuring the weight difference. However, it needs to be considered that this test can<br />

only find out approximate values <strong>and</strong> trends <strong>and</strong> can never replace the direct survey by the authorized h/p/cosmos<br />

technician in the prescribet time intervals.<br />

For security reasons, the <strong>running</strong> <strong>machine</strong> needs to be switched off <strong>and</strong> the power plug needs<br />

to be pulled out for full separation from the mains.<br />

[9.D2] Step Action Reaction/display<br />

[01]<br />

[02]<br />

Trittfläche / footboard<br />

1 meter<br />

Laufgurt / <strong>running</strong> belt<br />

Trittfläche / footboard<br />

L<br />

Taschentuch<br />

Tissue<br />

M<br />

R<br />

10 cm<br />

10 cm<br />

There are 3 pre-weighed paper tissues<br />

[cos14379] required for the test.<br />

Every single tissue is located in a separate<br />

plastic bag. The information (blue letters) in<br />

the left picture need to be entered to the<br />

sticker which is on the plastic bag.<br />

The tissue-test needs to be performed at 3<br />

places under the <strong>running</strong> belt:<br />

- Left edge<br />

- center<br />

- right edge<br />

For each test a separate tissue needs to be<br />

used!<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 129 of 213


[9.D2] [03]<br />

maintenance & safety inspections<br />

1. Take the tissue out of the plastic bag <strong>and</strong><br />

place it directly to the previously marked<br />

position under the <strong>running</strong> belt.<br />

2. Press with your h<strong>and</strong> on the <strong>running</strong> belt<br />

above the tissue <strong>and</strong> move the <strong>running</strong><br />

belt together with the tissue for exact 1<br />

meter.<br />

3. Take out the tissue <strong>and</strong> put it into the<br />

corresponding plastic bag. Take care that<br />

the plastic bag is closed sufficiently.<br />

Repeat steps 1 to 3 for the remaining<br />

positions.<br />

[04] After all required steps have been performed the tissues need to be returned to h/p/cosmos for further<br />

evaluation. The tissues will be weighed again <strong>and</strong> according to the weight difference h/p/cosmos can<br />

calculate if a modification of the lubrication interval is necessary.<br />

[9.D3] Devices without UserTerminal <strong>and</strong> without automatic oil-pump<br />

The devices without UserTerminal are equipped with an automatic oil message via acoustic signal (bleep)<br />

indicating the device needs to be lubricated. On st<strong>and</strong>ard setting every 1000 km (special device might have<br />

different intervals) after switching on the device (main power switch) the acoustic signal "5 times LONG (code for<br />

"0")" <strong>and</strong> 1 time SHORT + 4 times LONG (code for "1") will be repeated 3 times. The device has no sensor <strong>and</strong> the<br />

oil message will not terminate automatically after lubrication.<br />

The lubrication of the <strong>running</strong> belt <strong>and</strong> <strong>running</strong> surface is done with the enclosed, to the extent of delivery,<br />

accessories (bottle of lubricant <strong>and</strong> 10ml syringe). Use the provided lubricant only. Other available oils <strong>and</strong><br />

lubricants can damage the <strong>running</strong> belt <strong>and</strong> <strong>running</strong> surface <strong>and</strong> therefore lead to a breakdown of the <strong>running</strong><br />

<strong>machine</strong>.<br />

Please contact h/p/cosmos to receive the special oil. The <strong>running</strong> belt should be lubricated after 1000 km (special<br />

devices might have different intervals) at the latest or/<strong>and</strong> after hearing dry grinding noises during operation. After<br />

lubrication the oil message has to be terminated with "OPTION 01". An external UserTerminal or a PC-software<br />

h/p/cosmos para control ® is required for this.<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 130 of 213


[9.D4] Devices with UserTerminal & automatic oil-pump (models "r" for cycle & wheelchair)<br />

Running <strong>machine</strong>s which are designed for cycle <strong>and</strong><br />

wheelchair use are equipped with an automatic<br />

controlled oil-pump <strong>and</strong> an oil-tank.<br />

The oil-tank is located in the external control unit of the<br />

device below at the left side (At device up to the<br />

construction year 2002 <strong>and</strong> without external control unit<br />

the oil-tank is located at the rear position of the <strong>machine</strong>,<br />

close to the drive engine).<br />

The picture shows the oil-pump with the valve-cock (red<br />

arrow). With the valve-cock the valve for <strong>running</strong>-deck<br />

lubrication has to be positioned horizontally or vertically,<br />

depending on the application on the <strong>running</strong> surface<br />

(see illustrations next page). The valve-cock always<br />

must be rotated up to the mechanical stop. Intermediate<br />

positions ar not allowed <strong>and</strong> can result in serious<br />

damages of the <strong>machine</strong>.<br />

maintenance & safety inspections<br />

In particular at <strong>running</strong> <strong>machine</strong>s designed for cycle <strong>and</strong> wheelchair application, the correct quantity amd quality of<br />

oil is extremely important. Thus at those models a lubrication test (tissue-test) according the instruction (available<br />

from h/p/cosmos) must be performed in the course of the obligatory yearly maintenance. The test must be sent<br />

back to h/p/cosmos afterwards for evaluation purposes.<br />

Cooperation <strong>and</strong> inspection obligation of the user:<br />

Especially at treadmills designed for cycle <strong>and</strong> wheelchair application (all types of point loads on the <strong>running</strong><br />

surface or the <strong>running</strong> belt) a cooperation <strong>and</strong> inspection obligation of the user is due. The lubrication quantity <strong>and</strong><br />

quality between <strong>running</strong> belt <strong>and</strong> <strong>running</strong> surface needs to be observed <strong>and</strong> if necessary corrected accordingly.<br />

See separate instruction “lubricant test”.<br />

N Warning! Only special-treadmills with special labelling are adapted for cycling <strong>and</strong><br />

wheelchair applications! At treadmills which are not designed for cycling <strong>and</strong> wheelchair<br />

use or for use with spikes or ski sticks, those applications will result in serious damages<br />

of the treadmill.<br />

N Due to environmental influences, interior climate <strong>and</strong> type of application (for example<br />

frequent short-run operation or infrequent long-run operation, etc.) the required<br />

automatic oil-setting of the treadmill can deviate extensively from the factory setting.<br />

N In case the required lubrication quantity <strong>and</strong> quality is not assured due to missing<br />

inspections of the user <strong>and</strong>/or due to missing maintenance through authorized<br />

h/p/cosmos service-technicians, early damages of <strong>running</strong> belt <strong>and</strong> <strong>running</strong> surface can<br />

occur <strong>and</strong> warranty will expire.<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 131 of 213


unning, cycling, cross-country, skating [inline, roller ski]<br />

Valve-cock position [1] horizontally: The distributor [2] is<br />

adjusted for lubrication [3] only in the center of the <strong>running</strong>deck<br />

[4].<br />

The valve-cock position only changes the distribution of the<br />

oil amount to the various oil-outlets (tubes).<br />

The total oil-supply amount of the oil-pump always remains<br />

constant, independent from the valve-cock position.<br />

A nearly empty oil-tank will be indicated by the control lamp<br />

“Empty Oil-tank” at the external control unit. Refill the oiltank<br />

when control lamp indicates<br />

wheelchair<br />

Valve-cock position [1] vertically: The distributor [2] is<br />

adjusted for lubrication [3] over the entire width of the<br />

<strong>running</strong>-deck [4].<br />

The valve-cock position only changes the distribution of the<br />

oil amount to the various oil-outlets (tubes). The total oilsupply<br />

amount of the oil-pump always remains constant,<br />

independent from the valve-cock position.<br />

A nearly empty oil-tank will be indicated by the control lamp<br />

“empty oil-tank” at the external control unit. Refill the oiltank<br />

when control lamp indicates. Additionally at st<strong>and</strong>ard<br />

settings the word “OIL” will flash every 1000 km at the<br />

display. If this message occurs, check the oil-level in the oiltank<br />

<strong>and</strong> refill preterm if necessary.<br />

maintenance & safety inspections<br />

The device does not have a sensor, the oil-message does not extinguish automatically. After checking the oil-level,<br />

the message has to be quit with „OPTION 01“.<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 132 of 213


maintenance & safety inspections<br />

Refilling oil-tank<br />

N Follow this procedure to refill the oil-tank:<br />

N Unplug the device <strong>and</strong> wait for one minute.<br />

N The oil-tank is located in the external control unit of the device below at the left side (At device up to the<br />

construction year 2002 <strong>and</strong> without external control unit the oil-tank is located at the rear position of the<br />

<strong>machine</strong> close to the drive engine.).<br />

N Open the lid of the tank <strong>and</strong> refill the oil-tank with approximately 0.5 litre (at devices up to construction year<br />

2002 <strong>and</strong> without external control unit approximately 1 to 1.5 litre) with the provided original special oil, to<br />

receive at h/p/cosmos.<br />

N Close the lid of the oil-tank.<br />

N Plug in the power plug of the device <strong>and</strong> switch device on.<br />

N Check control lamp at the external control unit. If it is not indicating anymore, the oil-tank has been correctly<br />

refilled.<br />

N If necessary quit oil-message with „OPTION 01“.<br />

Special oil for sliding plate with automatic lubrication/0.25l-bottle [cos14007]<br />

Special oil for sliding plate with automatic lubrication/0.50l-bottle [cos13438]<br />

[9.D5] Set back of oil message MCU 4 without UserTerminal<br />

An external UserTerminal or the PC-software h/p/cosmos para control ® is required for this. Further instructions see<br />

"Devices with UserTerminal”.<br />

[9.D6] Set back of oil message MCU with UserTerminal<br />

Start: Running belt is not moving. One of the mode LED � of operation is flashing: (manual, profile, cardio, test)<br />

[9.D6] Step Activity Keys Response/Display<br />

[01] Select of the mode for "user<br />

options" (OPxx)<br />

= indicates: OP01 flashing (for option no. 01)<br />

a d b<br />

! <strong>and</strong> " indicate: E.rE SEt (for error reset)<br />

all 3 keys<br />

simultaneously for<br />

at least 3 seconds<br />

[02] Confirm option no. 1<br />

[03] Delete message<br />

[04] Quit option mode<br />

c<br />

c<br />

b<br />

§ indicates: donE to inform you, that you have to<br />

delete the oil message<br />

= indicates: OP01 flashing (for option no. 01)<br />

! <strong>and</strong> " indicate: E.rE SEt (or error reset)<br />

St<strong>and</strong>-by mode<br />

� manual, profil, cardio or test is flashing<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 133 of 213


[9.E] Control <strong>and</strong> tightening of the <strong>running</strong> belt<br />

maintenance & safety inspections<br />

The belt can loosen after some time of use or when it has been adjusted in a wrong way, so that a backlash occurs with<br />

every step between the driving shaft <strong>and</strong> the belt (slowing down of the belt at weight application through the step).<br />

In this case check the belt tension as follows:<br />

N Visual inspection of the surface for cracks. In case of a crack<br />

replace <strong>running</strong> belt immediately.<br />

N Open the motor hood in the front. Take care that nobody gets<br />

their h<strong>and</strong>s into the motor.<br />

N Now press the "START/Enter" key to get into manual mode<br />

<strong>and</strong> select with the help of the "+"/"-" keys a speed of 1 up to<br />

1.5 km/h.<br />

N St<strong>and</strong> on the <strong>running</strong> belt (if necessary with 2 people). Hold<br />

on to the side h<strong>and</strong>rails with both h<strong>and</strong>s <strong>and</strong> try to block the<br />

<strong>running</strong> belt by stemming yourself against the <strong>running</strong> belt<br />

rotation with your feet.<br />

N Try to block the <strong>running</strong> belt for max. 10 seconds. The driving<br />

shaft <strong>and</strong> the motor shaft should not be turning during that<br />

time. Otherwise the <strong>running</strong> belt has to be tightened (or the<br />

driving belt).<br />

If the <strong>running</strong> belt is being blocked for too long, then the motor regulation can switch off because<br />

of over current, which will be indicated by an error display. In this case, switch off the device for 5<br />

minutes <strong>and</strong> then switch it on again. The belt tension should not have more than 0.5 %, for<br />

otherwise it could lead to damages on the belt, on the shafting or the bearings!<br />

For models with a <strong>running</strong> surface up to 190 x 65 cm the belt is tightened with a hex socket (8 or 10 mm) at the rear<br />

drive shaft [2]. All bigger models are tightened at the front drive shaft [5]. Mark of 1000 mm [1] applied when belt [4]<br />

absolutely loose. Tighten belt, until mark exp<strong>and</strong>ed up to between 1004 <strong>and</strong> 1005 mm (= 0.4 to 0.5 %).<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 134 of 213


maintenance & safety inspections<br />

Procedure<br />

The right belt tension is between 0.4 - 0.5 %. That means that a length mark of 1000mm (applied on the belt with a pen)<br />

on a loose <strong>running</strong> belt will exp<strong>and</strong> to between 1004 <strong>and</strong> 1005mm at the right belt tension.<br />

Through turning of the left <strong>and</strong> right adjustment <strong>and</strong> tightening screws (hex socket 8mm) to the right the correct belt<br />

tension will be obtained. The belt can be tightened until the shaft does not turn anymore when <strong>running</strong> belt is being<br />

blocked (see control of <strong>running</strong> belt tension).<br />

The hex socket must be pulled out of the screw immediately after usage due to risk of injury!<br />

[9.F] Adjustment (centering) of the <strong>running</strong> belt<br />

Attention: Dangerous capture area at the tail shaft at the back of the devices (belt re-entry zone).<br />

Take care not to be caught with long hair or loose clothes in the capture area of the tail shaft. For<br />

safety reasons the procedure of adjustment has to be supervised by a second person that can<br />

press the emergency-stop in an emergency.<br />

Before adjustment of the <strong>running</strong> belt the belt has to be lubricated, if necessary. Lubrication of the belt may have impact<br />

on the centring of the belt.<br />

Adjust the <strong>running</strong> belt while the device is in operation with the help of the left trimming screw of the <strong>running</strong> <strong>machine</strong>.<br />

Use the enclosed hex socket ( 8 mm ). Operate <strong>running</strong> <strong>machine</strong> at 12 km/h without elevation. Watch the <strong>running</strong> belt<br />

for at least 2 minutes. The <strong>running</strong> belt should be in a centre position on the rear roller (tail shaft).<br />

If this is not the case please pay attention to the following points.<br />

N Move the <strong>running</strong> belt to the right by turning the LEFT trimming screw to the right.<br />

N Move the <strong>running</strong> belt to the left by turning the LEFT trimming screw to the left.<br />

N If the discrepancy is only minor the screw should only be turned max. ¼, in the case of major discrepancies with<br />

max. ½ a turn.<br />

N The <strong>running</strong> belt has to be observed after each alteration for at least 2 minutes. For the purpose of control operate<br />

the Running-Machine at 5 km/h <strong>and</strong> 20 km/h.<br />

N The procedure of adjustment is finished if the <strong>running</strong> belt stays in the centre of the rear roller (tail shaft) after<br />

having been operated at a speed of 12 km/h for some time (min. 4 min.).<br />

N Uphill run <strong>and</strong> different other <strong>running</strong> styles can lead to locomotion of the <strong>running</strong> belt. +/- 2cm are to be<br />

considered as an area of tolerance. As long as the <strong>running</strong> belt returns to the centre at a speed of 12 km/h it is not<br />

necessary to readjust. The <strong>running</strong> belt remains in the adjusted position for a period of months if it has been done<br />

correctly.<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 135 of 213


N During "reverse of belt rotation" (downhill run) a readjustment of the belt<br />

position (centring) is required.<br />

N Through evenly turning the left <strong>and</strong> the right trimming screw to the right you<br />

can tighten the <strong>running</strong> belt if necessary. See point <strong>running</strong> belt tension.<br />

maintenance & safety inspections<br />

Note: For <strong>running</strong> <strong>machine</strong>s with a <strong>running</strong> surface of 200 x 75 cm <strong>and</strong> bigger the trimming screw for the <strong>running</strong> belt is<br />

located at the front drive shaft. In this case the <strong>running</strong> belt movement is inverse after turning the screws.<br />

The hex socket must be pulled out of the screw immediately after usage due to risk of injury!<br />

[9.G] Control <strong>and</strong> tightening of the driving belt<br />

The correct tension of the driving belt is very important for a proper operation of the <strong>running</strong> <strong>machine</strong>. A too high tension<br />

can cause defects of the bearings from the motor <strong>and</strong> the roller. A too low tension results in backlash of the drive belt<br />

<strong>and</strong> therefore in higher abrasion <strong>and</strong> power losses.<br />

N This test can only be done with a dismounted motor hood. Don’t touch any parts in the motor<br />

compartment <strong>and</strong> take care not to be caught with long hair or loose clothes in the capture<br />

area of the danger zones.<br />

N For safety reasons the procedure of adjustment has to be supervised by a second person<br />

that can press the emergency-stop in an emergency.<br />

N Before intervening in the device for safety reasons switch the <strong>running</strong> <strong>machine</strong> off <strong>and</strong> pull<br />

the mains plug out.<br />

[9.G1] Driving belt with poly-V-belt system<br />

The poly-v-belt is provided with a belt adjuster <strong>and</strong> requires tightening usually very seldom. The belt adjuster offers<br />

the possibility of readjustment.<br />

N Checking the tightening of the drive belt with the „slipping-test“<br />

With the „slipping-test“ only a too low tension of the drive belt can be recognised. To be able to exclude a too high<br />

tension of the drive belt, the „frequency-test“ (see next point) must be used.<br />

The driving belt tension can be controlled according to point control of <strong>running</strong> belt tension.<br />

N Block the <strong>running</strong> belt at speed of 1 to 1.5 km/h.<br />

N No backlash must be between the motor shaft <strong>and</strong> the driving belt.<br />

N The tension at the bottom trumm (see picture) must be between 98 Hz <strong>and</strong> 103 Hz (frequency at oscillation).<br />

N The noise (at higher speeds) <strong>and</strong> the straight position of the driving belt need to be considered.<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 136 of 213


maintenance & safety inspections<br />

N Checking the tightening of the drive belt with the „frequency-test“<br />

Checking with measuring device for belt tension [cos14863]:<br />

N Switch off the <strong>running</strong> <strong>machine</strong> at the main switch.<br />

upper trumm<br />

N Visual inspection of the drive belt for abraison <strong>and</strong><br />

correct fitting.<br />

N Switch on the tuning device.<br />

N The tuning device must be in mode „BASS“,<br />

change with button GUITAR/BASS.<br />

N Hold the tuning device with the microphone (MIC)<br />

close to the bottom trumm of the drive belt <strong>and</strong><br />

pick the trumm (see picture).<br />

N The tension of the drive belt at the bottom trumm<br />

(see picture) needs to be between 98 Hz <strong>and</strong><br />

103 Hz (frequency during swinging).<br />

bottom trumm / place to pick<br />

N Check the tone in the display: Target = 1G – adjust the tension of the drive belt if necessary.<br />

N Check the amplitude of the needle: Target 440 ... 460 – adjust the tension of the drive belt if necessary.<br />

N When the tuning device is indicating the tone 1G <strong>and</strong> an amplitude of the needle of 440 ... 460 the tension of<br />

the drive belt is ok.<br />

Chart with tones to the according frequency:<br />

target value frequency drive belt tone<br />

LB<br />

scale<br />

70 2D 420<br />

76 2D 455<br />

79 4E 425<br />

81 4E 440<br />

82 4E 440<br />

85 4E 447<br />

87 4E 457<br />

90 1G 422<br />

92 1G 425<br />

95 1G 432<br />

X 98 1G 440<br />

X 100 1G 450<br />

X 103 1G 457<br />

104 3A 425<br />

106 3A 427<br />

109 3A<br />

HC<br />

437<br />

The target value of the frequency is between 98Hz <strong>and</strong> 102Hz. This is equal to the tone 1G. If the result of the test<br />

is out of this area, the tension of the drive belt must be adjusted.<br />

frequency < value 98: belt tension need correction.<br />

frequency > value 102: belt tension need correction.<br />

After changing the tension of the drive belt new check is necessary.<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 137 of 213


[9.G2] Driving belt with timing-belt system<br />

maintenance & safety inspections<br />

Timing belt drives have no automatic belt adjuster. The tension of the timing belt (depending on the model) needs<br />

to be adjusted by the manual belt adjuster or by changing the position of the drive motor.<br />

The driving belt tension can be controlled according to point control of <strong>running</strong> belt tension:<br />

N Block the <strong>running</strong> belt at speed of 1 to 1.5 km/h.<br />

N No backlash must be between the motor shaft <strong>and</strong> the driving belt.<br />

N Check if it is possible to move the drive belt with thumb <strong>and</strong> forefinger 90° sidewards by using moderate<br />

force. If the angle is more or less than 90°, the tension is too high or too low <strong>and</strong> the drive belt must be<br />

adjusted accordingly.<br />

N The noise (at higher speeds) <strong>and</strong> the straight position of the driving belt need to be considered.<br />

[9.H] Side step platforms: Check for non-slipping<br />

Each <strong>running</strong> <strong>machine</strong> has an anti-skid area on both sides next to the <strong>running</strong> surface. These step platforms offer safe<br />

halt when leaving the <strong>running</strong> surface in emergency cases. Check the anti-skid areas in regular intervals <strong>and</strong> replace<br />

them at signs of wearing off.<br />

[9.I] Hygiene, cleansing, cleansing of the interior<br />

For safety reasons switch off the device <strong>and</strong> pull the plug before cleansing, disinfecting <strong>and</strong><br />

before opening it. Pay attention to the instruction manual of the device.<br />

Pay attention to the notes of the operating instructions of the concerning devices <strong>and</strong> the accessories.<br />

Chemicals needed for application or cleansing have to be stored in appropriate reservoirs because of the danger of<br />

mixing them up. Interior Cleansing is required in intervals of 6 months. Depending on your model lift up the motor hood<br />

by lifting it up or remove bellow from upper frame.<br />

Clean the interior of the <strong>running</strong> <strong>machine</strong> by removing the dirt with the help of a vacuum cleaner. Pay special attention to<br />

the ventilation net of the drive motor <strong>and</strong> the cover of the frequency inverter. At devices with external control unit the<br />

internal space of the control unit must be cleaned as well.<br />

The cleaning of the surfaces of the treadmill can be done with a light moistened cloth. For disinfection h/p/cosmos<br />

recommends Bacillol plus [cos12179]. Before using cleaning <strong>and</strong> disinfection substances always test the compatibility<br />

of the material at bad viewable places.<br />

In case patients with communicable diseases are treated on the treadmill (or there is only the suspicion for<br />

communicable diseases), the treadmill has to be disinfected at the disrcretion of the doctor or medical stuff before <strong>and</strong><br />

after ther treatment.<br />

This covers all parts the patient might have contact to, such as front- <strong>and</strong> side-h<strong>and</strong>rails, arm support, user terminal<br />

(keyboard <strong>and</strong> display), emergency stop, chestbelt or other heart rate measurement systems, <strong>running</strong> belt, emergency<br />

stop with magnet switch, safety harness for fall stop prevention, vest for unweighting system, motor cover, frameparts,<br />

treads, wheelchair ramp, drinking bottle, etc.<br />

Cleaning procedure:<br />

Spray contact surfaces, impact, rub off.<br />

Spray disinfectant on a tissue <strong>and</strong> rub clean the possibly contaminated area.<br />

Always follow the disinfectants manufacturer’s manual.<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 138 of 213


Medium to achieve hygienic safety:<br />

Bacillol plus (h/p/cosmos order number cos12179)<br />

> See manufacturers safety warnings <strong>and</strong> data sheet<br />

Depollution of used chleaning medium:<br />

Bacillol plus (h/p/cosmos order number cos12179)<br />

> Manufacturers depollution regulations<br />

[9.J] Cleansing <strong>and</strong> adjustment of the speed-sensor/Light-barrier<br />

maintenance & safety inspections<br />

Depending on the model, the firmware version <strong>and</strong> the manufacturing date, the <strong>machine</strong>s are equipped with …<br />

N no speed sensor (so no cleansing required)<br />

N Light barrier <strong>and</strong> incremental waver mounted without housing at the front (drive shaft) of the drive engine.<br />

N Device from the construction year app. 2003 onwards have instead of the light barriers an induction sensor with<br />

metal toothed lock washer.<br />

N Only at oversize <strong>running</strong> <strong>machine</strong>s with a <strong>running</strong> surface of 200/75cm up to 300/125cm <strong>and</strong> up to the<br />

construction year 2002: Speed sensor mounted in its own housing at the rear part (drive shaft) of the drive engine.<br />

Due to the aspect, that the sensors have their own housing, they must not be cleaned.<br />

The light barrier of the drive motor is a pulse generator for the control unit. The light barrier <strong>and</strong> the incremental glass<br />

should be cleansed carefully with a cloth moistened with alcohol in intervals of 6 - 12 months. In "built in condition" of the<br />

glass you can also clean it with the help of a bristle brush, which has been soaked in alcohol. Clean the optical elements<br />

in between the glass <strong>and</strong> the light barrier carefully.<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 139 of 213


maintenance & safety inspections<br />

Alternatively you can also use a spray (with a thin spraying canal) of fat-dissolving fluid (e.g. braking detergent). The<br />

pressure on the detergent when coming through the spraying canal will cleanse the optics of the light barrier.<br />

When adjusting the light barrier take care to keep the<br />

following guidelines for the spaces between the glass [3]<br />

<strong>and</strong> the casing [4] of the light barrier: Cleft [1] axial<br />

approx. 1 mm/radial approx. 1 mm left <strong>and</strong> right<br />

according to illustration above. The imaginary extension<br />

of the light barrier has to pass through the centre [2] of<br />

motor shaft.<br />

Right h<strong>and</strong> side: The distance [1] between induction<br />

sensors [2] to the cogwheel [3] must not be bigger than<br />

0.8 mm.<br />

[9.K] Cleansing <strong>and</strong> adjustment of the light-barrier at belt-re-entry-zones<br />

Some models (<strong>running</strong> surface 200/75cm <strong>and</strong> bigger) are equipped with light-barriers at the <strong>running</strong> belt re-entry zone<br />

(rear position <strong>and</strong> front position for the reverse of belt rotation). As a safety measurement the light barriers effect a<br />

quick-stop function when the light stream is interrupted by a h<strong>and</strong>, a falling towel, a coat <strong>and</strong> of course, also by dust at<br />

the glass <strong>and</strong>/or reflector of the light barrier.<br />

To avoid unwanted switch off (quick stop), the light barrier <strong>and</strong> the reflector glass should be cleansed carefully with a<br />

cloth moistened with alcohol in intervals of 1 week, depending on the air <strong>and</strong> dust conditions in the room.<br />

Adjustment (direction <strong>and</strong> focus) of the light barrier can be checked when it is dark in the room.<br />

In case the cleansing of the light barrier does not result in success, you need to check <strong>and</strong> correct adjustment (direction<br />

<strong>and</strong> focus) of the light barrier. A special small screwdriver is required for adjustment of the focus. Talk about the<br />

procedure to your h/p/cosmos service partner.<br />

Light barrier must not be exposed to dazzling flash light of cameras. They might turn off <strong>and</strong> stop<br />

the <strong>running</strong> belt abruptly.<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 140 of 213


[9.L] Spare parts & consumables<br />

maintenance & safety inspections<br />

order number description picture<br />

cos000200008001<br />

cos12181<br />

cos13438<br />

Special oil (bottle 0.25 l)<br />

For h/p/cosmos <strong>running</strong> deck surface only for <strong>running</strong> applications<br />

(not for bicycle <strong>and</strong> wheel-chair use!) <strong>and</strong> manual lubrication with<br />

"h<strong>and</strong>-pump" syringe)<br />

Syringe 10ml<br />

For lubrication of <strong>running</strong> surface<br />

Special oil (bottle 0.50 l) for automatic lubrication<br />

This oil type (special lubricant) is only for <strong>running</strong> deck surfaces for<br />

applications with wheels (cycles, wheelchair, ski, skating, etc.) where<br />

the <strong>running</strong> <strong>machine</strong> has an oil tank <strong>and</strong> automatic oil pump.<br />

cos10905 POLAR chest belt XS<br />

cos10906 POLAR chest belt S<br />

cos10165 POLAR chest belt M<br />

cos10907 POLAR chest belt L<br />

cos10902<br />

cos15178<br />

POLAR Transmitter Set T31<br />

uncoded, with chest-belt rubber-b<strong>and</strong> size M<br />

POLAR Transmitter Set T34<br />

uncoded, extended range for <strong>running</strong> <strong>machine</strong>s, with chest-belt<br />

rubber-b<strong>and</strong> size M<br />

cos100420c POLAR W.I.N.D. WearLink chest belt<br />

cos100420b POLAR W.I.N.D. Transmitter TRX24<br />

cos14903-02-XXS<br />

cos14903-02-S<br />

cos14903-02-M<br />

cos14903-02-L<br />

cos14903-02-XL<br />

cos13752<br />

cos60000110029<br />

Chest belt system XXS<br />

size XXS (children) for h/p/cosmos safety arch<br />

Chest belt system S<br />

size S for h/p/cosmos safety arch<br />

Chest belt system M<br />

size M for h/p/cosmos safety arch<br />

Chest belt system L<br />

size L for h/p/cosmos safety arch<br />

Chest belt system XL<br />

size XL for h/p/cosmos safety arch<br />

airwalk, cable assembly<br />

for h/p/cosmos airwalk 35/70/160<br />

1 time maintenance + safety check<br />

treadmill series mercury / mercury med<br />

cos60000110030 1 time maintenance + safety check treadmill series quasar/med … pulsar 3p<br />

cos60000110032 1 time maintenance + safety check treadmill series venus/saturn up to 300/125<br />

cos11964-V4.05.1xxx<br />

MCU4 EPROM Firmware V4.05.1 treadmill<br />

For treadmill with MCU4 control board<br />

chest-belt rubber-b<strong>and</strong><br />

(only rubber b<strong>and</strong>, without transmitter)<br />

To be replaced every 6 months<br />

or earlier at the first sign of wear.<br />

To be replaced every 6 months<br />

or earlier at the first sign of wear.<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 141 of 213


cos16600m5tm- MCU5 Firmware V1.06.4 treadmill<br />

v1.06.4 For treadmill with MCU5 control board<br />

cos00097010034 Interface connection cable RS 232 5m<br />

cos10264-02 Runningbelt for treadmill 150-50 grooved<br />

cos10275-02 Running-belt for treadmill 170-65 LOGO<br />

cos10276-02 Running-belt for treadmill 190-65<br />

cos12399 Copper grease 100g tube<br />

cos13708 Servicebox treadmill 150/50<br />

cos13713 Servicebox treadmill 170&190-65<br />

cos100008-va11 Keyboard film & display for treadmills<br />

cos100456 emergency stop 16mm<br />

cos00098050043 safety stop: harness + waist-belt<br />

cos12343 Non slip step stripe 1350 x 25mm<br />

cos12344 Non slip step stripe 1350 x 76mm<br />

cos12345 Non slip step stripe 1570x25mm<br />

cos12346 Non slip step stripe 1570x76mm<br />

cos12347 Non slip step stripe 1770x25mm<br />

cos12348 Non slip step stripe 1770x76mm<br />

cos100037va01 mains cable 5m<br />

cos15505-01 Main lead 5m 400V/16A CEE<br />

cos15555-01 Main lead 5m 400V/32A CEE-plug<br />

maintenance & safety inspections<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 142 of 213


[10.]Trouble shooting<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 143 of 213<br />

trouble shooting<br />

In case of any detected <strong>and</strong>/or assumed malfunctions <strong>and</strong>/or defects or unreadable safety warning<br />

labels the device has to be taken out of operation, the device has to be marked <strong>and</strong> secured against<br />

operation <strong>and</strong> the supplier <strong>and</strong> authorised service personnel has to be informed in writing. See „option<br />

40“ in chapter „option settings“. With option 41 ... 44 you can lock also separate modes (manual,<br />

profile, cardio, test).<br />

[10.A] Mechanical / Noise problems<br />

In case you hear noises like knocking or rattling during the training please check if the <strong>machine</strong> has a firm st<strong>and</strong> <strong>and</strong><br />

follow exactly the advises in the chapter "Mechanical Installation" in this manual. Most time a false adjustment of the<br />

levelling sockets (adjustable "feet") at the rear of the Running-Machine are responsible for the knocking sound.<br />

[10.B] Running belt adjustment problems<br />

If the tension of the <strong>running</strong> belt is not correct it is difficult to keep the position of the belt centred.<br />

In case of reverse direction of the <strong>running</strong> belt: It is normal that the belt needs to be re-adjusted every time when using<br />

this option. Check <strong>and</strong> follow exactly the advises in the chapter "Control <strong>and</strong> Tightening of the Running Belt" in this<br />

manual.<br />

[10.C] Fuses<br />

Expulsion fuse / ON-OFF button 2-pole C16A Expulsion fuse, device fuse 3-pole C16A<br />

Secondary fuse isolation transformer 1-pole B16A Secondary fuse isolation transformer 3-pole C16A


Expulsion fuse at devices with external switching cabinet Fuses in cabinet for power supply unit<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 144 of 213<br />

trouble shooting<br />

The <strong>running</strong> <strong>machine</strong>s are equipped with an expulsion fuse at the front of the device / frontal section below the hood or<br />

in the interior section (depending on your model).<br />

Under the motor cover you can find the power supply for the electronic control unit, inside the housing of the power<br />

supply are one (sport <strong>and</strong> fitness <strong>machine</strong>s) / two (medical devices) secondary fuses. In the cabinet for the power<br />

supply is a primary fuse, <strong>and</strong> at devices for the medical application up to construction size 190/65 additionally two<br />

secondary fuses for 12V DC <strong>and</strong> 24V AC. Furthermore you will find the secondary fuse of the isolation transformer<br />

underneath the engine hood at devices for medical application.<br />

[10.D] Interference factor<br />

[10.D1] Electrostatic Discharge<br />

If the user moves around the devices they can be electro statically charged with up to several thous<strong>and</strong> volts. If<br />

then the user touches a metal piece, keys or display, it can lead to an electrostatic discharge between the user <strong>and</strong><br />

the device. Electrostatic discharges can in certain cases result in an interference of the device. Generally those<br />

electrostatic discharges are without harm for the user as well as for the device, but can be quite unpleasant. The<br />

main causes for electrostatic discharges are the choice of clothes, the sole of a shoe <strong>and</strong> the movement. Very dry<br />

air <strong>and</strong> many light fittings can also lead to the same results.<br />

Solution: Try different clothes or shoes, humidify the air in the room, <strong>and</strong> switch part of the light fittings off. Please<br />

inform the manufacturer if you detect such interference.<br />

[10.D2] Source of Interference<br />

The devices should not be installed near to e.g. an x-ray device, motors or transformer with high connection<br />

power, as the electric <strong>and</strong> magnetic interference can falsify measurements. Very strong sources of interference<br />

(e.g. above the limit according to EMT) can influence the functions of the device.<br />

High-tension power lines nearby <strong>and</strong> electrical devices without c - sign <strong>and</strong> without a certificate of compliance for<br />

electro-magnetic-tolerance should be avoided as well.<br />

Pay special attention to the EMC (electro magnetic compatibility) data <strong>and</strong> manufacturer’s declaration in the<br />

chapter „technical data / EMC“ of this manual.


[10.E] Voltage at the device housing / electric shock<br />

[10.E1] Open (interrupted) earth wire / ground<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 145 of 213<br />

trouble shooting<br />

In case the earth wire (ground) is not connected (for example due to defect in the building installation), on the<br />

metal frame of the device a voltage will be life via the „Y-noise-suppression-capacity“. This characteristic is very<br />

common with almost all electrical equipment with metal housing <strong>and</strong> EMC-fault clearance filters. The earth wire<br />

<strong>and</strong>/or contactor at the wall socket [1] in the building or at any other part of the AC-power line/feeder is open [2]. In<br />

this case a voltage of approximately 110 volt runs via the capacitor, of the EMC-fault clearance filter (to be found in<br />

the device) between housing <strong>and</strong> earth. Via the capacitors of the incorporated EMC-filters a voltage [5] of approx.<br />

110 volts AC is live between frame [3] <strong>and</strong> grounding [4], e.g. via the floor or heating radiator system or other<br />

grounded parts or <strong>machine</strong>s.<br />

a) If the device is flawless (at device category sports = without Potential-Isolation-Transformer) the current flow<br />

is significant noticeable in case of touching any bare metal parts (for example screws).<br />

b) If the device is flawless (at device category medical = with Potential-Isolation-Transformer) the current flow is<br />

almost not noticeable in case of touching any bare metal parts (for example screws).<br />

c) If the device is defective (at device category sports <strong>and</strong> also medical), e.g. because of a defective isolation<br />

inside the device, the current flow may be life-threatening high in case of touching any bare metal parts (for<br />

example screws).<br />

Take the device out of operation immediately, unplug it <strong>and</strong> secure it against new start-up.<br />

Order an authorised electrician to repair the contactor circuit in the building or/<strong>and</strong> at the<br />

device.<br />

Read <strong>and</strong> follow the directives <strong>and</strong> information concerning check of ground-wire terminals <strong>and</strong> functions in the<br />

chapters installation, maintenance <strong>and</strong> safety checks of this operation <strong>and</strong> service manual.<br />

[10.F] Malfunctions of the heart rate measurement system POLAR<br />

Possible sources of interference:<br />

N Screens, computer, printer, mobile phones <strong>and</strong> any radio engineering system<br />

N Electric devices, electric motors, transformer<br />

N High-voltage transmission lines, also from trains<br />

N Strong fluorescent tubes near by<br />

N Central heating radiators<br />

N Other electric devices


file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 146 of 213<br />

trouble shooting<br />

In order to prevent the <strong>running</strong> <strong>machine</strong> being interfered by such factors, place the device at some distance away from<br />

such interference sources. Do not rely on the indicated values if you suspect interference.<br />

Please also pay attention to the instruction of the manufacturer POLAR concerning interference at www.polar.fi or at<br />

http://www.polar-electro.de/3_Service_2003/FAQ/serv_FAQ_all.html<br />

[10.F1] Troubleshooting heart rate measurement system POLAR<br />

[10.F1] option Description Explanation / Display<br />

OP07 Activate: Acoustic heart rate signal % indicates: OFF or ON<br />

OFF: No acoustic heart-rate-signal<br />

ON: Acoustic heart-rate-signal for every beat. This function is normally<br />

used to control the regularity of the heart rate or to find reasons for<br />

transmitting problems (like mobile phones or computer monitors).<br />

[10.G] Malfunctions of the RS232 interface<br />

The most common causes for problem with the RS232 interface are<br />

N Wrong connection cable between the <strong>running</strong> <strong>machine</strong> <strong>and</strong> the periphery<br />

N Technical defects of the connection cable or male / female connector<br />

N False setting of protocols / driver at Running-Machine or periphery (ECG, PC, Ergospirometry)<br />

N False setting of COM port at periphery (ECG, PC, Ergospirometry)<br />

[10.G1] Troubleshooting <strong>and</strong> testing of the RS232 interface<br />

N Loop-Back Test: For testing the RS232 of the <strong>running</strong> <strong>machine</strong> a special RS232 test plug including testing<br />

instruction is available from the manufacturer. Connect the plug to the RS232 port <strong>and</strong> adjust option 20: 10<br />

N A blinking code <strong>and</strong> the instruction will tell you if input <strong>and</strong> output of the RS232 are working well.<br />

N PC software h/p/cosmos para control ® (Freeware): If you install the software to an external PC you can control<br />

the <strong>running</strong>-<strong>machine</strong>. If it works you know the <strong>running</strong>-<strong>machine</strong>, the RS232 interface connection cable <strong>and</strong> the<br />

RS232 interface card of the PC are working well.<br />

[10.H] Malfunctions of the control unit<br />

In rare cases, after several years of use, malfunctions of the control unit are possible because of worn battery power. In<br />

this case contact the service department to change the battery of the control unit.


[10.I] Error messages / Error codes<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 147 of 213<br />

trouble shooting<br />

Interferences <strong>and</strong> error codes might be caused by problems with the power supply or because of not enough<br />

maintenance (lubrication of the <strong>running</strong> surface, etc.).<br />

N Check voltage supply. Do not use extension cords or multi-way connectors. Connect the <strong>machine</strong> direct to the<br />

socket in the wall. Each <strong>machine</strong> should have an individual circuit.<br />

N Check mechanical parts for any malfunction. Check also for towels or other things, which might interfere the drive<br />

system.<br />

N Check if any light barrier are dusty or not correctly adjusted.<br />

N Check lubrication of the <strong>running</strong> belt <strong>and</strong> lubricate if required.<br />

N Also contact problems at connectors (loose connections) caused by vibrations can result in malfunctions. So<br />

please check cables <strong>and</strong> connectors for loose connections.<br />

The device has a self-diagnosis which recognizes some errors <strong>and</strong> shows their error messages on the display of the<br />

MCU UserTerminal (operating terminal) or the frequency changer / inverter (inside the device).<br />

[10.I1] Devices with UserTerminal<br />

Start: Running belt is not moving. One of the mode LED � of operation is flashing: (manual, profile, cardio, test)<br />

[10.I1] Message Description Explanation / Display / Activity<br />

E01 Oil interval due = „E 01“ ! „OIL“ flashing<br />

§ $ % „HELP“ flashing<br />

Running <strong>machine</strong> is still functioning<br />

N Every 1000 km (interval is adjustable) lubrication service is<br />

due. See chapter "lubrication of the <strong>running</strong> belt". Lubricate the<br />

belt.<br />

N Attention: Running <strong>machine</strong> has no sensor for oil. Lubrication<br />

only according to operation manual. Error message does not<br />

disappear automatically.<br />

N Set back error message (option 01)<br />

E02 Service interval due<br />

= „E 02“ ! „Service“ flashing<br />

or<br />

§ $ % „HELP“ flashing<br />

Safety inspection due<br />

Running <strong>machine</strong> is still functioning<br />

N Every 5000 km (interval is adjustable) a general service<br />

(interior cleansing, checking the driving belt, <strong>running</strong> belt, etc.)<br />

is due.<br />

N Every 12 months a electric saftey inspection is due. See<br />

chapter maintenance in this manual.<br />

N Contact authorized service engineer<br />

N Set back error message (option 01)


[10.I1] E20 Elevation system<br />

exceeded the max. value<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 148 of 213<br />

trouble shooting<br />

= „E 20“ ! „ELEV“ flashing<br />

§ $ % „HELP“ flashing<br />

Running <strong>machine</strong> is still functioning (except for the lift device)<br />

N Lifting element exceeded the maximum value. The combined<br />

0% / maximum % -contact had been activated<br />

N Incorrect measurement of angle of elevation<br />

N Maybe wrong adjusted, dirty or defective elevation sensor<br />

N Contact authorized service engineer<br />

E21 Error of elevation system = „E 21“! „ELEV“ flashing<br />

§ $ % „HELP“ flashing<br />

Running <strong>machine</strong> is still functioning (except for the lift device)<br />

N Maybe technical blocking of elevation (e.g. overload)<br />

E30 Error of<br />

speed / distance measurement<br />

N Maybe power supply too weak because of an extension cable<br />

etc.<br />

N Check power supply <strong>and</strong> power cord<br />

N Maybe elevation sensor adjustment is wrong or the sensor is<br />

dirty or defective<br />

N Contact authorized service engineer<br />

= „E 30“ ! „INCR“ flashing<br />

§ $ % „HELP“ flashing<br />

Running <strong>machine</strong> can still operate for a few seconds at speed of<br />

max. approx. 1 km/h<br />

N Maybe technical blocking of <strong>running</strong> belt (e.g. alien element,<br />

towel trapped in device)<br />

N Maybe power supply too weak because of an extension cable,<br />

etc.<br />

N Check power supply <strong>and</strong> power cord<br />

N Maybe speed sensor adjustment is wrong, the sensor may be<br />

dirty or defective or the speed calibration is wrong. OPTION 34<br />

N Safety delay time (inverter or MCU) is longer than the high<br />

acceleration comm<strong>and</strong> (SPRINT function or high acceleration<br />

level)<br />

N Trouble shooting see Administrator OPTION 48<br />

N Contact authorized service engineer<br />

N Check voltage supply. Do not use extension cords. Connect<br />

the <strong>machine</strong> direct to the socket in the wall<br />

N Check mechanical parts for any malfunction. Maybe several<br />

persons were st<strong>and</strong>ing on the <strong>machine</strong> for a moment?<br />

N Remove pleated bellow from upper frame (where applicable,<br />

depending on frame).<br />

N Attention: Dangerous voltage <strong>and</strong> danger of being hurt when<br />

the device is open. Do not touch the parts inside the device


file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 149 of 213<br />

trouble shooting<br />

[10.I1] N Clean the speed sensor (light barrier <strong>and</strong> disk, where<br />

applicable) <strong>and</strong> check the adjustment of the sensor according<br />

to the instruction.<br />

N Check if the safety-, acceleration- <strong>and</strong> deceleration time (of the<br />

frequency inverter <strong>and</strong>/or MCU) is higher than the highest<br />

acceleration level (e.g. sprints with very high acceleration<br />

levels, respectively at using the SPRINT-function if available).<br />

N Erase the message with "OPTION 01". In case the message<br />

appears again, contact authorized service engineer.<br />

N Check also Administrator OPTION 48<br />

N Check plug no. S6 on MCU <strong>and</strong> analogue voltage 0 ... 10 volts<br />

(for 0 ... max. speed) from MCU board to inverter drive.<br />

N Check plug no. S6 on MCU <strong>and</strong> “inverter on – signal / RFR”<br />

from MCU board to inverter drive<br />

E31 + tolerance aberration speed sensor<br />

The measured speed-value<br />

(instantaneous value) is at least the in<br />

Check mains power supply to frequency inverter<br />

= „E 31“ ! „INCR“ flashing<br />

§ $ % „HELP“ flashing<br />

administrator-option 12 set percentage<br />

BIGGER than the calculated value<br />

(debit-value)<br />

Error search see E30.<br />

E32 - tolerance aberration speed sensor<br />

The measured speed-value<br />

(instantaneous value) is at least the in<br />

= „E 32“ ! „INCR“ flashing<br />

§ $ % „HELP“ flashing<br />

administrator-option 12 set percentage<br />

LOWER than the calculated value (debitvalue)<br />

Error search see E30.


[10.I1] E41 Error during memory initialization<br />

SETUP ERROR“ MCU<br />

or<br />

low or defective battery<br />

E50 Error of<br />

frequency inverter / motor regulation<br />

At ANALOG control of frequency inverter<br />

= „E 41“ ! „type“ " „OP03“ blinking<br />

§ „SETU“ $ „P E“ % „rrOr“ blinking<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 150 of 213<br />

trouble shooting<br />

<strong>running</strong> <strong>machine</strong> is still working<br />

N At MCU5 control electronic (as of manufacturing date<br />

12/2007) this error exclusively shows a defect of the option<br />

memory (FRAM). In this case, please contact h/p/cosmos <strong>and</strong><br />

request a service technician.<br />

N Optional setup restores to default factory setting <strong>and</strong> device<br />

type<br />

N By pressing the key c the device can still be used, but it is<br />

likely that individual settings <strong>and</strong> program profiles may not be<br />

available any more. Also Display readings <strong>and</strong> accuracy of<br />

speed <strong>and</strong> elevation need to be checked.<br />

N Press Start <strong>and</strong> STOP, power off <strong>and</strong> power on the device.<br />

N In case the error message E41 is still indicated, contact an<br />

authorized service technician.<br />

Either the MCU electronic may be defective or the battery may be<br />

low or defective.<br />

= „E 50“ ! „FU“ flashing<br />

§ $ % „HELP“ flashing<br />

Running <strong>machine</strong> is not functioning<br />

N Check plug no. S6 at the MCU board<br />

N Maybe defective frequency inverter<br />

N Maybe power supply too weak because of an extension cable<br />

etc.<br />

N Check power supply <strong>and</strong> power cord<br />

N Contact authorized service engineer


[10.I1] E51 Error of<br />

frequency inverter / motor regulation<br />

Signal error between MCU5 <strong>and</strong><br />

frequency inverter.<br />

Only applicable at digital control (RS485)<br />

of MCU5/frequency inverter<br />

= „E 51“ ! „FU“ flashing<br />

§ $ % „HELP“ flashing<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 151 of 213<br />

trouble shooting<br />

Display " shows the hexadecimal number of the original error<br />

code of the frequency inverter (for example “b1” of error message<br />

“E.PUE” = PU disconnection).<br />

N Check RS485 plug at the MCU5 control board<br />

N Check RS485 plug at the frequency inverter<br />

N Check RS485 plug at the optocoupler board<br />

N Maybe defective frequency inverter<br />

N Maybe power supply too weak because of an extension cable<br />

etc.<br />

N Check power supply <strong>and</strong> power cord<br />

Contact authorized service engineer<br />

abbreviations: UT = UserTerminal / FI = Frequency Inverter<br />

indication<br />

display UT<br />

indication<br />

display FI<br />

00 - no alarm<br />

Meaning of the error code related to the frequency inverter (motor drive control) function<br />

/ status<br />

10 E.OC1 Overcurrent shut-off during acceleration<br />

11 E.OC2 Overcurrent shut-off during constant speed<br />

12 E.OC3 Overcurrent shut-off during deceleration or stop<br />

20 E.OV1 Regenerative overvoltage shut-off during acceleration<br />

21 E.OV2 Regenerative overvoltage shut-off during constant speed<br />

22 E.OV3 Regenerative overvoltage shut-off during deceleration or stop<br />

30 E.THT Inverter overload shut-off (electronic thermal relay function)<br />

31 T.THM Motor overload shut-off (electronic thermal relay function)<br />

40 E.FIN Fin overheat<br />

52 E.ILF Input phase loss<br />

60 E.OLT Stall prevention<br />

70 E.BE Brake transistor alarm detection<br />

80 E.GF Output side earth (ground) fault overcurrent<br />

81 E.LF Output phase loss<br />

90 E.OHT External thermal relay operation<br />

A1 E.OP1 Communication option fault<br />

B0 E.PE Parameter storage device fault<br />

B1 E.PUE PU disconnection<br />

B2 E.RET Retry count excess<br />

C0 E.CPU CPU fault<br />

F1 E.1 Option fault<br />

F6 E.6 CPU fault<br />

F7 E.7 CPU fault<br />

FD E.13 Internal circuit fault


[10.I1] E52 Error of<br />

Card Err<br />

01<br />

Card Err<br />

02<br />

Card Err<br />

03<br />

Card Err<br />

04<br />

RS485-cable motor control<br />

cable-defect or bad contact at the<br />

connection MCU/frequency-inverter.<br />

Only applicable at digital control (RS485)<br />

of MCU5/frequency inverter<br />

Error chip-card-drive<br />

(not all models)<br />

Error chip-card-drive<br />

(not all models)<br />

Error chip-card-drive<br />

(not all models)<br />

Error chip-card-drive<br />

(not all models)<br />

= „E 52“ ! „FU“ flashing<br />

§ $ % „HELP“ flashing<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 152 of 213<br />

trouble shooting<br />

The MCU5-control-boards always sends a reset-message at every<br />

switch-on of the device. After 10 time-out-failures in a row, the MCU<br />

generates the E52 error-message. This is equivalent to a missing<br />

connection (e.g. cable-defect) in the cable from MCU5 to frequencyinverter.<br />

N Check RS485 plug at the MCU5 control board<br />

N Check RS485 plug at the frequency inverter<br />

N Check RS485 plug at the optocoupler board<br />

N Contact authorized service engineer<br />

= „CARD“ ! „ERR“ " „1“<br />

Running <strong>machine</strong> is still functioning (except chip card function)<br />

N Communication malfunction with Chip-Card<br />

N Insert chip-card to the chip-card-drive<br />

N Replace chip-card<br />

N Replace chip-card-drive<br />

N Contact authorized service engineer<br />

= „CARD“ ! „ERR“ " „2“<br />

Running <strong>machine</strong> is still functioning (except chip card function)<br />

N Too many training weeks<br />

N Check period of training<br />

N Check system date <strong>and</strong> time settings of the <strong>running</strong> <strong>machine</strong><br />

N Re-program chip-card<br />

= „CARD“ ! „ERR“ " „3“<br />

Running <strong>machine</strong> is still functioning (except chip card function)<br />

N Test ident on the chip-card not accepted<br />

N False chip-card in the chip-card-drive<br />

N Insert the correct chip-card<br />

= „CARD“ ! „ERR“ " „4“<br />

Running <strong>machine</strong> is still functioning (except chip card function)<br />

N unknown chip-card-malfunction<br />

N no card identified<br />

N check chip-card<br />

N insert correct chip-card<br />

N reprogram chip-card


[10.I2] Devices without UserTerminal<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 153 of 213<br />

trouble shooting<br />

The devices without UserTerminal are equipped with an automatic error message codes via acoustic signal (bleep)<br />

indicating the device has a malfunction or needs to be lubricated or regular service. In case of error <strong>and</strong> service<br />

messages a beep-tone is given as acoustical signal. So the user of the <strong>running</strong> <strong>machine</strong> gets a feedback about<br />

error <strong>and</strong> service messages even without UserTerminal. The appropriate beep-code resounds the moment the<br />

error occurs, <strong>and</strong> afterwards when switching on the <strong>running</strong> <strong>machine</strong> if the error still exists three times in a row. To<br />

interpret the error without a display a beep-code is audible according to the error message <strong>and</strong> according to the list<br />

below:<br />

o = long beep / x = short beep<br />

0= ooooo / 1= xoooo / 2= xxooo / 3= xxxoo / 4= xxxxo / 5= xxxxx / 6 = oxxxx / 7= ooxxx / 8= oooxx / 9= oooox<br />

Example OIL message: On st<strong>and</strong>ard setting every 1000 km after switching on the device (main power switch) the<br />

acoustic signal "5 times LONG (code for "0")" <strong>and</strong> 1 time SHORT + 4 times LONG (code for "1") will be repeated 3<br />

times. Error Code list see: Devices with UserTerminal. The error messages have to be terminated with<br />

"OPTION 01". An external UserTerminal or a PC software h/p/cosmos para control ® is required for this.<br />

[10.J] Set back of error message<br />

[10.J1] Devices with UserTerminal<br />

Initiation: Running-Belt is not moving. One of the modes LED � is flashing: (manual, profile, cardio, test)<br />

[10.J1] Step Activity Keys Response / Display<br />

[01] Select of the mode for user<br />

options (OP xx)<br />

= indicates: OP01 flashing (for option no. 01)<br />

a d b<br />

! <strong>and</strong> " indicate: E.rE SEt (für Error reset)<br />

Press all 3 keys<br />

simultaneously for<br />

at least 3 seconds<br />

[02] Confirm option no. 1<br />

c<br />

§ indicates: donE to inform you that you have to<br />

delete the oil message<br />

[03] Delete message<br />

= indicates: OP 01 flashing (for option no. 01)<br />

[04] Quit option mode<br />

c<br />

b<br />

! <strong>and</strong> " indicate: E.rE SEt (für Error reset)<br />

St<strong>and</strong>-by mode<br />

� manual, profil, cardio or test is flashing<br />

With this procedure only the message E01-OIL, as well as the message E02-Service can be reset. All other error<br />

messages are caused by technical defects / malfunctions. These error messages can only be reset in the<br />

administrator-level. Please contact in this case the customer service in written form.<br />

[10.J2] Devices without UserTerminal<br />

The OIL-message as well as the SERVICE-message have to be quit by „OPTION 01”. Error messages have to be<br />

quit by „OPTION 01“ at the administrator-level. In both cases you need an „external UserTerminal“ or the PC-<br />

Software h/p/cosmos para control ® to do that . For further procedures see devices with UserTerminal.


[11.]Technical data<br />

[11.A] Running <strong>machine</strong>s, 150/50<br />

UserTerminal<br />

(Display & Keyboard)<br />

Power supply<br />

(st<strong>and</strong>ard, read nameplate)<br />

Dedicated line required.<br />

h/p/cosmos stratos<br />

h/p/cosmos stratos lt<br />

h/p/cosmos mercury<br />

h/p/cosmos mercury lt<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 154 of 213<br />

h/p/cosmos stratos med<br />

6 LCD displays / 6 keys / display of mode <strong>and</strong> unit with LED.<br />

„lt-models“ <strong>and</strong> „h/p/cosmos locomotion 150/50 E med“ have no UserTerminal (no display, no<br />

keyboard). Control only via interface.<br />

h/p/cosmos stratos med lt<br />

h/p/cosmos mercury med<br />

h/p/cosmos mercury med lt<br />

230 Volt AC 1~/N/PE f: 50/60 Hz 3680 VA<br />

Fuse (st<strong>and</strong>ard, read nameplate) 16 A<br />

Power consumption<br />

(St<strong>and</strong>ard, read nameplate)<br />

14.5 A 15 A 14.5 A 15 A 15 A<br />

[a] 14.5 A<br />

Power input 3335 VA 3450 VA 3335 VA 3450 VA 3450 VA<br />

[a] 3335 VA<br />

Drive motor capacity 3300 W<br />

Elevation motor capacity not applicable 390 W not applicable 390 W 390 W<br />

[a] not applicable<br />

Safety st<strong>and</strong>ards c<br />

IEC EN 60335-1(VDE 0700)<br />

IEC EN 60601-1-2,<br />

IEC EN 60601-1-4, VDE 0701,<br />

EN 957-1, EN 957-6<br />

h/p/cosmos Kistler Gaitway II F [a]<br />

h/p/cosmos Kistler Gaitway II S<br />

n<br />

MDD, directive 93/42/EEC + 2007/47/EC,<br />

IEC EN 60601-1 (VDE 0750), VDE 0751,<br />

IEC EN 60601-1-2, IEC EN 60601-1-4,<br />

EN 957-1, EN 957-6<br />

Leakage current 0.6 mA 0.2 mA<br />

Isolation transformer --- 2000 VA<br />

degree of protection: class I M / --- / IP 20 class I M / type B R / IP 20<br />

Overvoltage category Category III: Life-Earth < 300 Volt = transient over-voltage limit 4000 Volt<br />

Mode of operation acc. IEC 601-1 continuous operation<br />

Field of application / Accuracy sports <strong>and</strong> fitness: SA sports <strong>and</strong> medical: SIA<br />

Classification according to MDD --- --- II b<br />

Speed<br />

Definition<br />

0.0 ... 22.0 km/h<br />

0.1 km/h<br />

Allowance* +/- 5 %<br />

Option special speed<br />

Angle of inclination**<br />

Definition<br />

not applicable 0.0 ... 25.0 %**<br />

0.1 %<br />

0.0 ... 10 km/h or 0.0... 30 km/h<br />

not applicable 0.0 ... 25.0 %**<br />

0.1 %<br />

0.0 ... 25.0 %**<br />

0.1 %<br />

[a] not applicable<br />

technical data<br />

h/p/cosmos locomotion 150/50 E med<br />

h/p/cosmos locomotion 150/50 DE med<br />

E: no Terminal<br />

DE: Touchscreen<br />

15 A<br />

3450 VA<br />

390 W<br />

0.0 … 10.0 km/h<br />

0.1 km/h<br />

0.0 ... 22.0 km/h<br />

0.0 ... 30.0 km/h<br />

0.0 ... 15.0 %**<br />

0.1 %


Dimension of <strong>running</strong> surface<br />

length x width<br />

Dimension frame<br />

length x width x height<br />

(UserTerminal + 150 mm width)<br />

h/p/cosmos stratos<br />

h/p/cosmos stratos lt<br />

h/p/cosmos mercury<br />

h/p/cosmos mercury lt<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 155 of 213<br />

h/p/cosmos stratos med<br />

h/p/cosmos stratos med lt<br />

h/p/cosmos mercury med<br />

1500 x 500 mm<br />

2100 x 800 x 1370 mm<br />

Track access height from floor 180 mm<br />

h/p/cosmos mercury med lt<br />

h/p/cosmos Kistler Gaitway II F [a]<br />

h/p/cosmos Kistler Gaitway II S<br />

technical data<br />

Weight of device 180 kg 190 kg 190 kg 200 kg 350 kg 380 kg<br />

Weight load limits to floors<br />

(traffic load to DIN 1055 part 3)<br />

h/p/cosmos locomotion 150/50 E med<br />

h/p/cosmos locomotion 150/50 DE med<br />

2200 x<br />

1000 x<br />

1440 mm<br />

8.1 kN/m² 8.2 kN/m² 9.1 kN/m² 9.2 kN/m²<br />

In case of any further questions related to load limits onto floors, please ask our service department.<br />

Max. permissible user weight 200 kg (deviation on request specially made)<br />

Environmental conditions:<br />

Transport <strong>and</strong> storage<br />

Environmental conditions:<br />

Operation<br />

temperature: –30...+50 °C<br />

humidity: 0...95% - without condensation<br />

barometric pressure: 700...1060hPa<br />

temperature: +10...+40 °C (deviation on request)<br />

humidity: 30...70% - without condensation (up to 95% on request)<br />

barometric pressure: 700...1060hPa<br />

Maximum operating altitude: approx. 10,000 feet (3000m), without pressurization<br />

Pollution degree Degree 2: Normally only nonconductive pollution. Temporary conductivity caused by condensation is to be expected.<br />

Central lubrication system manual<br />

Store program space 6 training profiles / 10 test profiles / 8 user definable profiles<br />

RS232 Digital interface / serial port<br />

(USB-converter optional)<br />

COM1: st<strong>and</strong>ard;<br />

COM2: option;<br />

COM3: optional (st<strong>and</strong>ard for service);<br />

COM4: option (not for locomotion models)<br />

COM1, COM2:<br />

st<strong>and</strong>ard<br />

COM3: st<strong>and</strong>ard<br />

(for service)<br />

Heart rate measurement POLAR, 1 channel receiver, ECG precisely POLAR W.I.N.D.<br />

PC software<br />

(not for medical applications)<br />

h/p/cosmos para control ® included<br />

options: h/p/cosmos para graphics ®, h/p/cosmos para analysis ® <strong>and</strong> h/p/cosmos para motion ®<br />

Reverse belt rotation / downhill 1 not applicable option not applicable option option st<strong>and</strong>ard<br />

Safety Arch with harness*** option<br />

H<strong>and</strong>rails options: short / long / adjustable h<strong>and</strong>rails, crossbar no options<br />

Arm support option not applicable<br />

additional keyboard option st<strong>and</strong>ard<br />

* Overload <strong>and</strong> weak power supply (undervoltage) could lead to higher discrepancies concerning the speed-display or dripping of the fuse. ** The time of<br />

operation of the elevation motor (TO = 15 %) converts to approx. 5 full elevation cycles. Overload can result in temporary tripping of the fuse, but can be<br />

used again after short cool down period. Treadmills with a <strong>running</strong> surface 150/50 built approx. before June 2007 have a maximum elevation of 24%.<br />

*** For maximum stress tests <strong>and</strong> with oversize <strong>running</strong> <strong>machine</strong>s (track sizes) a safety facility (e.g. a safety harness with chest belt) is necessary <strong>and</strong> highly<br />

recommended. When using a front-bar h<strong>and</strong>rail (crossbar), the speed is limited to 5 km/h for safety reasons. Data of weight <strong>and</strong> measurements can deviate.<br />

Ask for further details <strong>and</strong> optional equipment or visit www.h-p-cosmos.com. Subject to technical alterations without prior notice. E & EO.


[11.B] Running <strong>machine</strong>s, 170-190/65<br />

UserTerminal<br />

(Display & Keyboard)<br />

Power supply<br />

(st<strong>and</strong>ard, read nameplate)<br />

Dedicated line required.<br />

h/p/cosmos stellar<br />

h/p/cosmos stellar lt<br />

h/p/cosmos quasar<br />

h/p/cosmos quasar lt<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 156 of 213<br />

h/p/cosmos stellar med<br />

h/p/cosmos stellar med lt<br />

6 LCD displays / 6 keys / display of mode <strong>and</strong> unit with LED.<br />

„lt-models“ have no UserTerminal (no display, no keyboard). Control only via interface.<br />

h/p/cosmos quasar med<br />

230 Volt AC 1~/N/PE f: 50/60 Hz 3680 VA<br />

Fuse (st<strong>and</strong>ard, read nameplate) 16 A<br />

Power consumption (St<strong>and</strong>ard,<br />

read nameplate)<br />

15.0 A 15.5 A 15.0 A 15.5 A 16 A 8.0 A<br />

h/p/cosmos quasar med lt<br />

h/p/cosmos pulsar<br />

h/p/cosmos pulsar lt<br />

technical data<br />

h/p/cosmos pulsar 3p<br />

h/p/cosmos pulsar 3p lt<br />

400 Volt AC3~<br />

/N/PE f:50/60Hz<br />

11000 VA<br />

Power input 3450 VA 3565 VA 3450 VA 3565 VA 3680 VA 5536 VA<br />

Drive motor capacity 3300 W 4300 W<br />

Elevation motor capacity not applicable 500 W not applicable 500 W 500 W<br />

Safety st<strong>and</strong>ards c<br />

IEC EN 60335-1(VDE 0700)<br />

IEC EN 60601-1-2,<br />

IEC EN 60601-1-4, VDE 0701,<br />

EN 957-1, EN 957-6<br />

n<br />

MDD, directive 93/42/EEC + 2007/47/EC,<br />

IEC EN 60601-1 (VDE 0750), VDE 0751,<br />

IEC EN 60601-1-2, IEC EN 60601-1-4,<br />

EN 957-1, EN 957-6<br />

Leakage current 0.6 mA 0.2 mA 0.2 mA 0.25 mA<br />

Isolation transformer --- 2000 VA 2000 VA 3 x 1500 VA<br />

Degree of protection class I M / --- / IP 20 class I M / type B R / IP 20<br />

Overvoltage category Category III: Life-Earth < 300 Volt = transient over-voltage limit 4000 Volt<br />

Mode of operation acc. IEC 601-1 continuous operation<br />

Field of application / Accuracy sports <strong>and</strong> fitness: SA sports <strong>and</strong> medical: SIA<br />

Classification according to MDD --- --- II b<br />

Speed<br />

Definition<br />

0.0 ... 25.0 km/h<br />

0.1 km/h<br />

0.0 ... 25.0 km/h<br />

0.1 km/h<br />

Allowance* +/- 5 %<br />

0.0 ... 40.0 km/h<br />

0.1 km/h<br />

Option special speed 0.0 ... 10.0 km/h , 0.0 ... 40.0 km/h , 0.0... 45.0 km/h 0.0 ... 10.0 km/h or 0.0... 45.0 km/h<br />

Angle of inclination**<br />

Definition<br />

not applicable 0.0 ... 28.0 %<br />

0.1 %<br />

not applicable 0.0 ... 28.0 %<br />

0.1 %<br />

- 25.0 % ... + 25.0 %<br />

0.1 %


Dimension of <strong>running</strong> surface<br />

length x width<br />

Dimension frame<br />

length x width x height<br />

h/p/cosmos stellar<br />

h/p/cosmos stellar lt<br />

h/p/cosmos quasar<br />

h/p/cosmos quasar lt<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 157 of 213<br />

h/p/cosmos stellar med<br />

h/p/cosmos stellar med lt<br />

h/p/cosmos quasar med<br />

h/p/cosmos quasar med lt<br />

h/p/cosmos pulsar<br />

h/p/cosmos pulsar lt<br />

1700 x 650 mm 1900 x 650 mm<br />

2300 x 1050 x 1450 mm<br />

(UserTerminal + 150 mm width)<br />

Track access height from floor 230 mm<br />

technical data<br />

h/p/cosmos pulsar 3p<br />

h/p/cosmos pulsar 3p lt<br />

2500 x1050 x 1450 mm<br />

(UserTerminal + 150 mm width)<br />

Weight of device 360 kg 370 kg 370 kg 380 kg 400 kg 440 kg<br />

Weight load limits to floors<br />

(traffic load to DIN 1055 part 3)<br />

6.4 kN/m² 6.1 kN/m²<br />

In case of any further questions related to load limits onto floors, please ask our service department.<br />

Max. permissible user weight 200 kg (deviation on request specially made)<br />

Environmental conditions:<br />

Transport <strong>and</strong> storage<br />

Environmental conditions:<br />

Operation<br />

temperature: –30...+50 °C<br />

humidity: 0...95% - without condensation<br />

barometric pressure: 700...1060hPa<br />

temperature: +10...+40 °C (deviation on request)<br />

humidity: 30...70% - without condensation (up to 95% on request)<br />

barometric pressure: 700...1060hPa<br />

Maximum operating altitude: approx. 10,000 feet (3000m), without pressurization<br />

Pollution degree Degree 2: Normally only nonconductive pollution. Temporary conductivity caused by condensation is to be expected.<br />

Central lubrication system manual<br />

Store program space 6 training profiles / 10 test profiles / 8 user definable profiles<br />

RS232 Digital interface / serial port<br />

(USB-converter optional)<br />

Heart rate measurement <strong>and</strong> heart<br />

rate dependent endurance control<br />

COM1: st<strong>and</strong>ard;<br />

COM2: optional;<br />

COM3: optional (st<strong>and</strong>ard for service);<br />

COM4: optional<br />

POLAR, 1 channel receiver, ECG precisely<br />

PC software h/p/cosmos para control ® included<br />

options: h/p/cosmos para graphics ® <strong>and</strong> h/p/cosmos para analysis ®<br />

COM1 <strong>and</strong> COM2: st<strong>and</strong>ard<br />

COM3: optional (st<strong>and</strong>ard for<br />

service);<br />

COM4: optional<br />

POLAR W.I.N.D. with WearLink<br />

transmission range approx. 10 m<br />

h/p/cosmos para control ® <strong>and</strong><br />

h/p/cosmos para graphics ® included<br />

Reverse belt rotation / downhill 1 not applicable option not applicable option st<strong>and</strong>ard<br />

Safety Arch with harness*** option<br />

H<strong>and</strong>rails options: short, long, crossbar (st<strong>and</strong>ard)<br />

Arm support / additional keyboard option<br />

* Overload <strong>and</strong> weak power supply (undervoltage) could lead to higher discrepancies concerning the speed-display or dripping of the fuse. ** The time of<br />

operation of the elevation motor (TO = 15 %) converts to approx. 5 full elevation cycles. Overload can result in temporary tripping of the fuse, but can be<br />

used again after short cool down period. Treadmills of the h/p/cosmos saturn series built approx. before January 2007 have a maximum elevation of 25%.<br />

*** For maximum stress tests <strong>and</strong> with oversize <strong>running</strong> <strong>machine</strong>s (track sizes) a safety facility (e.g. a safety harness with chest belt) is necessary <strong>and</strong> highly<br />

recommended. When using a front-bar h<strong>and</strong>rail (crossbar), the speed is limited to 5 km/h for safety reasons. Data of weight <strong>and</strong> measurements can deviate.<br />

Ask for further details <strong>and</strong> optional equipment or visit www.h-p-cosmos.com. Subject to technical alterations without prior notice. E & OE.


[11.C] Running <strong>machine</strong>s, oversize<br />

External control unit /<br />

UserTerminal (Display &<br />

Keyboard)<br />

Power supply<br />

(st<strong>and</strong>ard, read nameplate)<br />

Dedicated line required.<br />

h/p/cosmos venus 200/75<br />

h/p/cosmos venus 200/75 r<br />

h/p/cosmos venus 200/100<br />

h/p/cosmos venus 200/100 r<br />

h/p/cosmos saturn 250/75<br />

h/p/cosmos saturn 250/75 r<br />

h/p/cosmos saturn 250/100<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 158 of 213<br />

h/p/cosmos saturn 250/100 r<br />

h/p/cosmos saturn 250/125 r<br />

h/p/cosmos saturn 300/75<br />

h/p/cosmos saturn 300/75 r<br />

h/p/cosmos saturn 300/100<br />

6 LCD displays / 6 keys / display of mode <strong>and</strong> unit with LED<br />

400 Volt AC 3~/N/PE f: 50/60 Hz<br />

12800 VA (min)<br />

h/p/cosmos saturn 300/100 r<br />

technical data<br />

Fuse (st<strong>and</strong>ard, read nameplate) 32 A 64 A<br />

Power consumption (St<strong>and</strong>ard,<br />

read nameplate)<br />

17.0 A 18.0 A 20.0 A 21.0 A 22.0 A 23.0 A 31.3 A 62.0 A<br />

Power input 6760 VA 7160 VA 7800 VA 8360 VA 8750 VA 9200 VA 12500 VA 45000VA<br />

Drive motor capacity 7500 W 11000 W 30000W<br />

Elevation motor capacity 370 W 18000W<br />

Safety st<strong>and</strong>ards n<br />

MDD, directive 93/42/EEC + 2007/47/EC,<br />

IEC EN 60601-1 (VDE 0750), IEC EN 60601-1-2, IEC EN 60601-1-4<br />

VDE 0751, EN 957-1, EN 957-6<br />

Leakage current 0.25 mA on request<br />

Isolation transformer<br />

3 x 2000 VA 3 x 3000 VA<br />

Degree of protection class I M / type B R / IP 20<br />

Overvoltage category Category III: Life-Earth < 300 Volt = transient over-voltage limit 4000 Volt<br />

Mode of operation acc. IEC 601-1 Continuous operation<br />

Field of application / Accuracy<br />

according to EN 957<br />

h/p/cosmos saturn 300/125 r<br />

3 x 4200<br />

VA<br />

sports <strong>and</strong> medical S I A<br />

[r] = suitable for bicycle <strong>and</strong> wheel chair application (from <strong>running</strong> surface 3000 x 1250 mm st<strong>and</strong>ard)<br />

Classification according to MDD II b<br />

Speed / Definition 0.0 ... 40.0 km/h / 0.1 km/h<br />

0.0 ... 11.11 m/s / 0.1 m/s<br />

Allowance* +/- 5 %<br />

Option special speed 0.0 ... 30.0 or ... 50.0 or ... 60.0 or ... 80.0 km/h<br />

Angle of inclination**<br />

Definition<br />

Dimension of <strong>running</strong> surface<br />

length x width in mm<br />

[r] = suitable for bicycle <strong>and</strong> wheel<br />

chair application<br />

-35 % ... +35 %<br />

0.1 %<br />

-27 % ... +27%**<br />

0.1 %<br />

h/p/cosmos saturn 450/300 rs<br />

3 x 21000<br />

VA<br />

-5%(-25%)<br />

… +25%<br />

2000x750 2000x1000 2500x750 2500x1000 2500x1250 3000x750 3000x1000 3000x1250 4500x3000


Dimension frame<br />

length x width x height<br />

Track access height from floor<br />

(optional installation in pit)<br />

Weight of device<br />

Weight load limits to floors<br />

(traffic load to DIN 1055 part 3)<br />

Max. permissible user weight on<br />

<strong>running</strong> surface<br />

Environmental conditions:<br />

Transport <strong>and</strong> storage<br />

Environmental conditions:<br />

Operation<br />

h/p/cosmos venus 200/75<br />

h/p/cosmos venus 200/75 r<br />

2400<br />

x1150 x<br />

1380<br />

h/p/cosmos venus 200/100<br />

h/p/cosmos venus 200/100 r<br />

2400 x<br />

1400 x<br />

1380<br />

h/p/cosmos saturn 250/75<br />

h/p/cosmos saturn 250/75 r<br />

2900 x<br />

1150 x<br />

1380<br />

h/p/cosmos saturn 250/100<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 159 of 213<br />

h/p/cosmos saturn 250/100 r<br />

2900 x<br />

1400 x<br />

1380<br />

h/p/cosmos saturn 250/125 r<br />

2900 x<br />

1650 x<br />

1380<br />

h/p/cosmos saturn 300/75<br />

h/p/cosmos saturn 300/75 r<br />

3400<br />

x1150 x<br />

1380<br />

h/p/cosmos saturn 300/100<br />

h/p/cosmos saturn 300/100 r<br />

3400<br />

x1400 x<br />

1380<br />

technical data<br />

h/p/cosmos saturn 300/125 r<br />

3400 x<br />

1650 x<br />

1380<br />

h/p/cosmos saturn 450/300 rs<br />

on request<br />

dep. on<br />

installation<br />

480 mm on request<br />

850 kg 900 kg 950 kg 1000 kg 1050 kg 1100 kg 1150 kg 1250 kg<br />

on request<br />

dep. on<br />

installation<br />

9.0 kN/m² 7.5 kN/m² 7.7 kN/m² 6.5 kN/m² 5.6 kN/m² 7.0 kN/m² 5.8 kN/m² 5.1 kN/m² on request<br />

In case of any further questions related to load limits onto floors, please ask our service department.<br />

temperature: –30...+50 °C<br />

humidity: 0...95% - without condensation<br />

barometric pressure: 700...1060hPa<br />

280 kg<br />

(deviation on request specially made)<br />

temperature: +10...+40 °C (deviation on request)<br />

humidity: 30...70% - without condensation (up to 95% on request)<br />

barometric pressure: 700...1060hPa<br />

Maximum operating altitude: approx. 10,000 feet (3000m), without pressurization<br />

Pollution degree Degree 2: Normally only nonconductive pollution. Temporary conductivity caused by condensation is to be expected.<br />

Central lubrication system manual / <strong>running</strong> <strong>machine</strong>s with [r] have automatic central lubrication system<br />

Store program space 6 training profiles / 10 test profiles / 8 user defined profiles<br />

RS232 Digital interface / serial port COM1: st<strong>and</strong>ard; COM2: st<strong>and</strong>ard; COM3: optional (st<strong>and</strong>ard for service); COM4: optional<br />

Heart rate measurement <strong>and</strong> heart<br />

rate dependent endurance control<br />

(target pulse programming)<br />

PC software<br />

(not for medical application)<br />

POLAR W.I.N.D. with WearLink<br />

coded transmission<br />

transmission range approx. 10 m<br />

h/p/cosmos para control ® <strong>and</strong> h/p/cosmos para graphics ® included<br />

option: h/p/cosmos para analysis ®, h/p/cosmos para motion ®<br />

Reverse belt rotation / downhill st<strong>and</strong>ard<br />

Safety arch with harness*** option<br />

Wheel chair stabilizer option<br />

Monitor arm <strong>and</strong> laptop computer option<br />

* Overload <strong>and</strong> weak power supply (undervoltage) could lead to higher discrepancies concerning the speed-display or dripping of the fuse. ** The time of<br />

operation of the elevation motor (TO = 15 %) converts to approx. 5 full elevation cycles. Overload can result in temporary tripping of the fuse, but can be<br />

used again after short cool down period. Treadmills of the h/p/cosmos saturn series built approx. before January 2007 have a maximum elevation of 25%.<br />

*** For maximum stress tests <strong>and</strong> with oversize <strong>running</strong> <strong>machine</strong>s (track sizes) a safety facility (e.g. a safety harness with chest belt) is necessary <strong>and</strong> highly<br />

recommended. When using a front-bar h<strong>and</strong>rail (crossbar), the speed is limited to 5 km/h for safety reasons. Data of weight <strong>and</strong> measurements can deviate.<br />

Ask for further details <strong>and</strong> optional equipment or visit www.h-p-cosmos.com. Subject to technical alterations without prior notice. E & OE.


[11.D] EMC electromagnetic compatibility. Guidance <strong>and</strong> manufacturer's declaration<br />

Table 201: Guidance <strong>and</strong> manufacturer's declaration - electromagnetic emissions (for all equipment <strong>and</strong> systems)<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 160 of 213<br />

technical data<br />

The <strong>running</strong> <strong>machine</strong> is intended for use in the electromagnetic environment specified below. The customer or the user of the <strong>running</strong> <strong>machine</strong> should<br />

assure that it is used in such an environment.<br />

Emissions test Compliance Electromagnetic environment — guidance<br />

RF emissions CISPR11 Group 1 Class B<br />

RF emissions CISPR11 Group 1 Class B<br />

Harmonic emissions IEC 61000-3-2 Class A<br />

Voltage fluctuations/ flicker emissions<br />

IEC 61000-3-3<br />

Complies<br />

The <strong>running</strong> <strong>machine</strong> uses RF energy only for its internal function. Therefore, its RF<br />

emissions are low <strong>and</strong> are not very likely to cause any interference in nearby<br />

electronic equipment.<br />

The <strong>running</strong> <strong>machine</strong> is suitable for use in all establishments including domestic<br />

<strong>and</strong> those directly connected to the public low-voltage power supply network that<br />

supplies buildings used for domestic purposes.<br />

Table 202: Guidance <strong>and</strong> manufacturer's declaration - electromagnetic immunity (for all equipment <strong>and</strong> systems)<br />

The <strong>running</strong> <strong>machine</strong> is intended for use in the electromagnetic environment specified below. The customer or the user of the <strong>running</strong> <strong>machine</strong> should<br />

assure that it is used in such an environment.<br />

Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance<br />

Electrostatic discharge (ESD)<br />

IEC 61000-4-2<br />

Electrical fast transient/burst<br />

IEC 61000-4-4<br />

Surge<br />

IEC 61000-4-5<br />

Voltage dips,<br />

short interruptions <strong>and</strong><br />

voltage variations on<br />

power supply input lines<br />

IEC 61000-4-11<br />

Power frequency (50/60 Hz)<br />

magnetic field<br />

IEC 61000-4-8<br />

±6 kV contact<br />

±8 kV air<br />

±2 kV for power supply lines<br />

±1 kV for input/output lines<br />

±1 kV differential mode<br />

±2 kV common mode<br />

95 % dip in UT)<br />

for 0.5 cycle<br />

40 % UT<br />

(60 % dip in UT)<br />

for 5 cycles<br />

70 % UT<br />

(30 % dip in UT)<br />

for 25 cycles<br />

95 % dip in UT)<br />

for 5 sec<br />

±6 kV contact<br />

±8 kV air<br />

±2 kV for power supply lines<br />

±1 kV for input/output lines<br />

±1 kV differential mode<br />

±2 kV common mode<br />

95 % dip in UT)<br />

for 0.5 cycle<br />

40 % UT<br />

(60 % dip in UT)<br />

for 5 cycles<br />

70 % UT<br />

(30 % dip in UT)<br />

for 25 cycles<br />

95 % dip in UT)<br />

for 5 sec<br />

3 A/m Not applicable<br />

NOTE UT is the A.C. mains voltage prior to application of the test level.<br />

Please pay special attention also to the other EMC related chapters of this manual:<br />

a) Chapter: Safety, Installation, Operation, POLAR heart rate monitor: EMC<br />

b) Chapter: Error, Malfunction, possible interference<br />

Floors should be wood, concrete or ceramic tile. If<br />

floors are covered with synthetic material, the<br />

relative humidity should be at least 30 %.<br />

Mains power quality should be that of a typical<br />

commercial or hospital environment.<br />

Mains power quality should be that of a typical<br />

commercial or hospital environment.<br />

Mains power quality should be that of a typical<br />

commercial or hospital environment. If the user of<br />

the <strong>running</strong> <strong>machine</strong> requires continued<br />

Operation during power mains interruptions, it is<br />

recommended that the <strong>running</strong> <strong>machine</strong> be<br />

powered from an uninterruptible power supply or<br />

a battery. Caution! <strong>running</strong> <strong>machine</strong>s require high<br />

capacity UPS because of high capacity motor<br />

drive.<br />

Power frequency magnetic fields should be at<br />

levels characteristic of a typical location in a<br />

typical commercial or hospital environment.


file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 161 of 213<br />

technical data<br />

Table 204: Guidance <strong>and</strong> manufacturer's declaration -electromagnetic immunity (for all equipment <strong>and</strong> systems that are not life supporting)<br />

The <strong>running</strong> <strong>machine</strong> is intended for use in the electromagnetic environment specified below. The customer or the user of the <strong>running</strong> <strong>machine</strong> should<br />

assure that it is used in such an environment.<br />

Immunity test IEC 60601 test level Compliance level Electromagnetic environment — guidance<br />

Conducted RF<br />

IEC 61000-4-6<br />

Radiated RF<br />

IEC 61000-4-3<br />

3 Vrms<br />

150 kHz to 80 MHz<br />

3 V/m<br />

80 MHz to 2.5 GHz<br />

3 Vrms<br />

3 V/m<br />

Portable <strong>and</strong> mobile RF communications equipment should be<br />

used no closer to any part of the <strong>running</strong> <strong>machine</strong>, including<br />

cables, than the recommended separation distance calculated<br />

from the equation applicable to the frequency of the transmitter.<br />

Recommended separation distance<br />

d = 1.17 1/V * √P<br />

d = 1.17 m/V * √P for 80 MHz to 800 MHz<br />

d = 2.33 m/V * √P for 800 MHz to 2.5 GHz<br />

where P is the maximum output power rating of the transmitter in<br />

watts (W) according to the transmitter manufacturer <strong>and</strong> d is the<br />

recommended separation distance in meters (m).<br />

Field strengths from fixed RF transmitters, as determined by an<br />

electromagnetic site survey, a should be less than the compliance<br />

level in each frequency range. b<br />

Interference may occur in the vicinity of equipment marked with<br />

the following symbol:<br />

NOTE 1: At 80 MHz <strong>and</strong> 800 MHz, the higher frequency range applies. NOTE 2: These guidelines may not apply in all situations.<br />

Electromagnetic propagation is affected by absorption <strong>and</strong> reflection from structures, objects <strong>and</strong> people.<br />

a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones <strong>and</strong> l<strong>and</strong> mobile radios, amateur radio, AM <strong>and</strong> FM<br />

radio broadcast <strong>and</strong> TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF<br />

transmitters, an electromagnetic site survey should be considered, If the measured field strength in the location in which the <strong>running</strong> <strong>machine</strong> is used<br />

exceeds the applicable RF compliance level above, the <strong>running</strong> <strong>machine</strong> should be observed to verify normal operation. If abnormal performance is<br />

observed, additional measures may be necessary, such as reorienting or relocating the <strong>running</strong> <strong>machine</strong>.<br />

b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.<br />

Table 206: Recommended separation distances between portable <strong>and</strong> mobile RF communications equipment <strong>and</strong> the <strong>running</strong> <strong>machine</strong><br />

(for all equipment <strong>and</strong> systems that are not life supporting)<br />

The <strong>running</strong> <strong>machine</strong> is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of<br />

the <strong>running</strong> <strong>machine</strong> can help prevent electromagnetic interference by maintaining a minimum distance between portable <strong>and</strong> mobile RF communications<br />

equipment (transmitters) <strong>and</strong> the <strong>running</strong> <strong>machine</strong> as recommended below, according to the maximum output power of the communications equipment.<br />

Rated maximum output power<br />

of transmitter (in Watt)<br />

W<br />

Separation distance (in meters) according to frequency of transmitter<br />

150 kHz to 80 MHz<br />

d = 1.17 1/V * √P<br />

80 MHz to 800 MHz<br />

d = 1.17 m/V * √P<br />

800 MHz to 2.5 GHz<br />

d = 2.33 m/V * √P<br />

0.01 W 0.12 m 0.12 m 0.23 m<br />

0.1 W 0.37 m 0.37 m 0.74 m<br />

1 W 1.17 m 1.17 m 2.33 m<br />

10 W 3.70 m 3.70 m 7.37 m<br />

100 W 11.7 m 11.7 m 23.3 m<br />

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the<br />

equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the<br />

transmitter manufacturer.<br />

NOTE 1: At 80 MHz <strong>and</strong> 800 MHz, the separation distance for the higher frequency range applies. NOTE 2: These guidelines may not apply in all<br />

situations. Electromagnetic propagation is affected by absorption <strong>and</strong> reflection from structures, objects <strong>and</strong> people.


[11.E] Compatibility interface<br />

technical data<br />

Products for application in combination with h/p/cosmos treadmill:<br />

h/p/cosmos offers the treadmill in combination with components the system can be used with (components as well from<br />

h/p/cosmos as from other manufacturers).<br />

The clinical evaluation refers only to the h/p/cosmos treadmill <strong>and</strong> the „intersection risks“ (e.g. risk at stresstest until fatigue in<br />

treadmill ergometry with cardiorespiratory diagnostic) <strong>and</strong> does not extend to the optional components themselves.<br />

The products mentioned in article 12 of the MDD 93/42/EEC are listed in the following table.<br />

<strong>Medical</strong> Product<br />

Ergospirometry<br />

System<br />

CareFusion (VIASYS<br />

/ JAEGER) Oxycon<br />

mobile CPX<br />

Ergospirometry-<br />

System<br />

COSMED K4 b² /<br />

Quark b²<br />

Ergospirometry-<br />

System<br />

CORTEX<br />

SCHILLER CS 200<br />

Excellence<br />

POLAR Heart Rate<br />

Monitor Type T31,<br />

T34 <strong>and</strong> W.I.N.D with<br />

WearLink<br />

Pneumex<br />

Unweighting System /<br />

BWS Body Weight<br />

Support System<br />

PneuWeight (airwalk)<br />

70 <strong>and</strong> PneuWeight<br />

(airwalk) 160.<br />

Producer<br />

manufacturer:<br />

CareFusion Germany<br />

234 GmbH / Höchberg<br />

manufacturer:<br />

COSMED s.r.l<br />

Rome / Italy<br />

manufacturer:<br />

CORTEX Biophysik<br />

GmbH<br />

manufacturer:<br />

SCHILLER AG<br />

Baar / Switzerl<strong>and</strong><br />

manufacturer:<br />

POLAR Electro OY<br />

Oulu / Finl<strong>and</strong><br />

sales through:<br />

h/p/cosmos sports &<br />

medical gmbh /<br />

Germany<br />

manufacturer:<br />

Pneumex, Inc. /<br />

S<strong>and</strong>point / USA.<br />

placing on EC-market:<br />

EUROLINK (Europe)<br />

Ltd. / Martin Green /<br />

Watchfield Swinden / UK<br />

sales through:<br />

h/p/cosmos sports &<br />

medical gmbh /<br />

Germany<br />

Intended Use<br />

applicable in<br />

ergospirometry<br />

applicable in<br />

ergospirometry<br />

applicable in<br />

ergospirometry<br />

applicable in ECG stress<br />

testing<br />

applicable in stress testing<br />

<strong>and</strong> cardiac rehabilitation<br />

applicable in orthopedic<br />

<strong>and</strong> neurologic<br />

rehabilitation<br />

CE-class<br />

IIa<br />

CE0123<br />

IIa<br />

CE0476<br />

IIa<br />

CE0124<br />

IIa<br />

CE0123<br />

IIa<br />

CE 0537<br />

Configuration<br />

via coscom v3 interface:<br />

release for complete<br />

medical treadmill series.<br />

via coscom v3 interface:<br />

release for complete<br />

medical treadmill series.<br />

via coscom v3 interface:<br />

release for complete<br />

medical treadmill series.<br />

via coscom v3 interface:<br />

release for complete<br />

medical treadmill series.<br />

wireless transmission.<br />

Type T31, T34 released for<br />

all h/p/cosmos treadmills of<br />

150/50 series (except<br />

locomotion) <strong>and</strong> 170/65<br />

series. W.I.N.D. system<br />

released for all locomotion,<br />

pulsar, venus <strong>and</strong> saturn<br />

models.<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 162 of 213 LN<br />

Im<br />

CE<br />

no physical contact to<br />

h/p/cosmos treadmill, just<br />

used “over” the treadmill.<br />

released for all h/p/cosmos<br />

treadmills with <strong>running</strong> deck<br />

size 150/50cm <strong>and</strong><br />

170…190/65cm


[11.E1] Interface additional information:<br />

technical data<br />

All other systems <strong>and</strong> devices, which are not in the above stated list, must not be connected to h/p/cosmos<br />

treadmills.<br />

For getting the latest updated list of compatible devices <strong>and</strong> systems in writing please contact<br />

service@h-p-cosmos.com<br />

Only connect <strong>and</strong>/or use accessories <strong>and</strong> software <strong>and</strong> host equipment if it is approved as compatible by all<br />

manufacturers. Non compliance to this clause can lead to serious injury or death <strong>and</strong> furthermore manufacturer<br />

warranty <strong>and</strong> liability expires.<br />

Pay attention to further safety regulations <strong>and</strong> operation instructions in the appendix of this instruction manual. Pay<br />

special attention to all instructions for operation, maintenance <strong>and</strong> service also in manuals of all accessories<br />

<strong>and</strong>/or software.<br />

Accessory equipment connected to the analog <strong>and</strong> digital interfaces must be certified according to the respective<br />

IEC st<strong>and</strong>ards, e.g. IEC 950 for data processing equipment <strong>and</strong> IEC 60601-1 for medical equipment. Furthermore<br />

all configurations shall comply with the valid version of the system st<strong>and</strong>ard IEC 60601-1-1. Always use IEC<br />

60601-1 approved potential isolation components when linking medical devices via interface. It is not allowed to<br />

connect sports treadmills to medical devices, for instance ECG-systems.<br />

Everybody who connects additional equipment to the signal input port or signal output port or via any other linkage<br />

possibility, configures a medical system, <strong>and</strong> is therefore responsible that the system complies with the<br />

requirements of the valid version of the system st<strong>and</strong>ard IEC 60601-1-1.<br />

More recent computers will have only an USB interface instead of the RS232 interface. For that case an „USB to<br />

RS232 interface-adapter-cable“ is available at h/p/cosmos under item no. cos12769. To control via USB interface<br />

you need a processor Pentium 1.8 GHz or higher.<br />

h/p/cosmos only approves the interface protocol „coscom v3“ for medical applications <strong>and</strong> herewith confirms the -<br />

since March 21, 2010 (MDD / Directive 2007/47/EC) m<strong>and</strong>atory- conformity of coscom v3 according to the norms<br />

EN 62304 (Life Cycle Requirements for <strong>Medical</strong> Device Software; <strong>Medical</strong> Device Software Life Cycle Processes)<br />

<strong>and</strong> EN 14971 (risk management for medical devices) as well as the compliance to a very high safety st<strong>and</strong>ard.<br />

All other interface protocols (older versions coscom v2, coscom v1, coscomekg.dll, Trackmaster, Pacer, etc.) are<br />

not validated by h/p/cosmos according to the norms EN 62304 <strong>and</strong> EN 14971 <strong>and</strong> must not be used for medical<br />

applications (e.g. ergometry, stress-tests, cardiorespiratory diagnostic, performance diagnostic, etc.) since March<br />

21, 2010 in connection with h/p/cosmos treadmills or h/p/cosmos OEM treadmills, except where explicitly approved<br />

<strong>and</strong> confirmed in writing by h/p/cosmos <strong>and</strong> the respective manufacturer!<br />

The support for all older coscom ® files (including coscomekg.dll version 1.00 <strong>and</strong> coscom.dll version 1.2.9 <strong>and</strong> also<br />

including coscom v2) ended on 31.12.2010. Please upgrade to the new coscom v3 .NET Objects <strong>and</strong> coscom v3<br />

.NET Controls.<br />

Please note also, that the older versions of coscom v1 <strong>and</strong> coscom v2 are not designed <strong>and</strong> not documented<br />

based on IEC EN 62304:2006 <strong>Medical</strong> device software - Software life cycle processes. Therefore since March 21,<br />

2010 it is obligatory to upgrade to h/p/cosmos coscom v3 in order to stay compatible with the latest norms, latest<br />

safety st<strong>and</strong>ards <strong>and</strong> Regulatory Affair Issues when connecting to h/p/cosmos medical treadmills.<br />

We strongly recommend to update to the latest h/p/cosmos coscom v3 interface protocol. The latest coscom.dll<br />

<strong>and</strong> implementation notes can be found on www.coscom.org. In case you have any questions please contact<br />

service@h-p-cosmos.com<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 163 of 213 LN


technical data<br />

<strong>Medical</strong> <strong>running</strong> <strong>machine</strong>s (treadmills) are active therapeutic devices <strong>and</strong> active diagnostic devices of the risk class<br />

IIb.<br />

Modern PC do not have a RS232 interface, but an USB interface. In this case h/p/cosmos offers a “interface<br />

converter USB / RS232”, article number cos12769. For treadmill control via USB interface the processor must be<br />

at least Pentium 1.8 GHz or higher.<br />

If the equipment is used in the field of medicine, all connected devices <strong>and</strong> host equipment must be linked with<br />

potential equalization cables. First connect the potential equalization cable with the corresponding potential<br />

equalization plug pin (next to the main switch at the front of the h/p/cosmos device), then connect the entire<br />

medical system to the main potential equalization terminal of the medical room <strong>and</strong> then connect the mains voltage<br />

power plug of the medical system.<br />

When connecting medical devices to a medical system <strong>and</strong>/or when putting a company sign/logo/mark on a<br />

medical device, you become the manufacturer of this new device/system <strong>and</strong> you are responsible for CE-marking,<br />

technical documentation, risk management <strong>and</strong> all obligations based on MDD <strong>and</strong> the respective local laws <strong>and</strong><br />

norms.<br />

----<br />

Article 12 of MDD Directive 93/42/EEC:<br />

Particular procedure for systems <strong>and</strong> procedure packs<br />

1. By way of derogation from Article 11 this Article shall apply to systems <strong>and</strong> procedure packs.<br />

2. Any natural or legal person who puts devices bearing the CE marking together within their intended purpose <strong>and</strong><br />

within the limits of use specified by their manufacturers, in order to place them on the market as a system or<br />

procedure pack, shall draw up a declaration by which he states that:<br />

(a) he has verified the mutual compatibility of the devices in accordance with the manufacturers' instructions <strong>and</strong><br />

has carried out his operations in accordance with these instructions; <strong>and</strong><br />

(b) he has packaged the system or procedure pack <strong>and</strong> supplied relevant information to users incorporating<br />

relevant instructions from the manufacturers; <strong>and</strong><br />

(c) the whole activity is subjected to appropriate methods of internal control <strong>and</strong> inspection.<br />

Where the conditions above are not met, as in cases where the system or procedure pack incorporate devices<br />

which do not bear a CE marking or where the chosen combination of devices is not compatible in view of their<br />

original intended use, the system or procedure pack shall be treated as a device in its own right <strong>and</strong> as such be<br />

subjected to the relevant procedure pursuant to Article 11.<br />

3. Any natural or legal person who sterilized, for the purpose of placing on the market, systems or procedure packs<br />

referred to in paragraph 2 or other CE-marked medical devices designed by their manufacturers to be sterilized<br />

before use, shall, at his choice, follow one of the procedures referred to in Annex IV, V or VI. The application of the<br />

abovementioned Annexes <strong>and</strong> the intervention of the notified body are limited to the aspects of the procedure<br />

relating to the obtaining of sterility. The person shall draw up a declaration stating that sterilization has been<br />

carried out in accordance with the manufacturer's instructions.<br />

4. The products referred to in paragraphs 2 <strong>and</strong> 3 themselves shall not bear an additional CE marking. They shall<br />

be accompanied by the information referred to in point 13 of Annex I which includes, where appropriate, the<br />

information supplied by the manufacturers of the devices which have been put together. The declaration referred<br />

to in paragraphs 2 <strong>and</strong> 3 above shall be kept at the disposal of competent authorities for a period of five years.<br />

---- end of Article 12 of MDD Directive 93/42/EEC ---<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 164 of 213 LN


[12.]Accessories <strong>and</strong> options<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 165 of 213<br />

accessories & options<br />

Only connect <strong>and</strong>/or use accessories <strong>and</strong> software <strong>and</strong> host equipment if it is approved as<br />

compatible by all manufacturers. Non compliance to this clause can lead to serious injury or death<br />

<strong>and</strong> furthermore manufacturer warranty <strong>and</strong> liability expires.<br />

Pay attention to further safety regulations <strong>and</strong> operation instructions in the appendix of this<br />

instruction manual. Pay special attention to all instructions for operation, maintenance <strong>and</strong> service<br />

also in manuals of all accessories <strong>and</strong>/or software.<br />

[12.A] Arm support for body weight support <strong>and</strong> safety<br />

The h/p/cosmos arm supports with additional keyboard <strong>and</strong><br />

additional stop-button offers a wide range of applications. It<br />

is possible for example for patients with over-weight or<br />

people with orthopaedic problems or cardiovascular<br />

diseases to exercise in gentle way. Therapies can start<br />

earlier <strong>and</strong> the therapist is supported efficiently in his work.<br />

The h/p/cosmos arm supports are available for the model<br />

series h/p/cosmos mercury, h/p/cosmos quasar <strong>and</strong><br />

h/p/cosmos pulsar with long <strong>and</strong> st<strong>and</strong>ard h<strong>and</strong>rails.<br />

The arm supports are adjustable exactly to respective needs <strong>and</strong> physical dem<strong>and</strong>s. This is possible due to a patented<br />

easily adjustable system. Three joints with scale on each side offer individual adjustments. The arm supports are<br />

applicable especially in the field of rehabilitation <strong>and</strong> walking therapy, as well as for the safety of seniors <strong>and</strong> insecure<br />

persons. The h/p/cosmos arm supports (patent no.: DE 199 16 508 A1) are available for the h/p/cosmos <strong>running</strong><br />

<strong>machine</strong>s with <strong>running</strong> deck size 150/50, 170/65 <strong>and</strong> 190/65 cm but not for <strong>running</strong> <strong>machine</strong>s with adjustbale h<strong>and</strong>rails.<br />

h/p/cosmos arm support [cos00098060044] for h<strong>and</strong>rails with 10° angle<br />

h/p/cosmos arm support [cos12013] for h<strong>and</strong>rails with 0° angle (for example long h<strong>and</strong>rails with 2 pillars)<br />

Chest-harnesses, patient-vests, waist-belts with leash <strong>and</strong> cuffs are not designed for direct skin or<br />

mucous membrane contact. Wear appropriate clothing.<br />

[12.A1] Danger precaution <strong>and</strong> operation of the adjustable arm supports<br />

N The maximum load is limited to persons not more than 140 kg body weight.<br />

N Do not adjust the arm supports under strain (the upholstered arm support must not be weighted by a person)!<br />

N Do only use the arm supports under supervision of a trained staff member. Only trained staff is allowed to<br />

adjust the arm supports.<br />

N Do only loosen the gear tooth locking elements to adjust the angle, if the you hold the arm-supports with one<br />

h<strong>and</strong> at the h<strong>and</strong>le or at the upholstered underarm support. Turn the respective lever with the h<strong>and</strong>, that holds<br />

the h<strong>and</strong>le, in the desired position.<br />

N Before weighting make sure that after the adjustment the gear tooth locking elements are correctly engaged.<br />

N While adjusting make sure that you do not pen in yourself or any other person at the danger spots (squeezing<br />

<strong>and</strong> jamming spots).<br />

N Take care that the underarm overlay at the left <strong>and</strong> right arm supports are approximately hoizontal <strong>and</strong> evely<br />

adjusted.


file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 166 of 213<br />

accessories & options<br />

N Do not use the arm supports at reverse belt rotation.<br />

N For safety reasons the use of the arm supports is only admitted for walking <strong>and</strong> not for <strong>running</strong>.<br />

N While <strong>running</strong> the arm supports have to be folded vertically downwards on the outside [1]. At any case the arm<br />

supports must never loom into the area of the <strong>running</strong> surface [2].<br />

[12.A2] Reproducibility of adjustments<br />

To obtain reproducibility you have to determine the individually optimal adjustments (see scaling at the locking<br />

elements) for each subject <strong>and</strong> note it down appropriatly (e.g. in a patient file). We have deliberately refrained from<br />

giving you st<strong>and</strong>ard settings for particular heights, because the adjustment can widly differ according to the<br />

posture <strong>and</strong> the individual situation of the testing person.<br />

The torsional elements have a scaling at any joint, which is afferent when turned in direction to the <strong>running</strong><br />

surface. The adjustments of the height <strong>and</strong> width occurs gradually, because the interlocking of the joints works in<br />

6-degree steps. In the illustration below, which you can copy, e.g. for patient files, you can fill in the scaling datas<br />

for the individual adjustments of a patient.


[12.B] Additional keyboard <strong>and</strong> add. stop-button for arm support <strong>and</strong> remote control<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 167 of 213<br />

accessories & options<br />

To complement the arm supports <strong>and</strong> to increase the safety an additional keyboard has been developed. With this<br />

keyboard it is possible to control the speed <strong>and</strong> elevation from a supported <strong>and</strong> safe position at the arm supports. The<br />

additional stop-button for the case of an emergency is accessible at any time with the thumb.<br />

picture type of stop function interface<br />

communication<br />

Additional stop<br />

button for arm<br />

support,<br />

incorporated on<br />

top of the<br />

h<strong>and</strong>grip<br />

N Stop button is pushed<br />

N <strong>running</strong> belt & inclination<br />

system stop fast<br />

N <strong>machine</strong> is NOT cut off from<br />

main power of drive motor<br />

<strong>and</strong> inclination system<br />

N display active <strong>and</strong> shows<br />

“PULL STOP”<br />

N interface active, but no start,<br />

speed or elevation signal<br />

executed<br />

N keyboard inactive<br />

communication<br />

NOT interrupted<br />

during<br />

“pull stop” stage<br />

how to restart<br />

N release push force<br />

from the button<br />

N press START key,<br />

N or new comm<strong>and</strong><br />

through interface<br />

The additional keyboard can also be taken off from its mounting on the arm-support <strong>and</strong> can be fixed on the h<strong>and</strong>rail or<br />

next to the motor cover left or right with the help of additional mountings. With the long connection cable the therapist<br />

has the possibility to take over the control of the <strong>running</strong> <strong>machine</strong> e.g. from a kneeling position during locomotion<br />

therapy.


file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 168 of 213<br />

accessories & options<br />

article number<br />

h/p/cosmos additional keyboard with 6 keys incl. 2m spiral cord cos10106<br />

mount for additional keyboard on arm support cos10111<br />

h/p/cosmos additional stop button (red button on top of h<strong>and</strong>grip) for arm support right cos10107<br />

h/p/cosmos additional stop button (red button on top of h<strong>and</strong>grip) for arm support left cos10108<br />

mount for additional stop button on h<strong>and</strong>rail with clamp for circular tube 60mm ¬ cos14135<br />

mount for additional keyboard on motor cover left cos14327<br />

mount for additional keyboard on motor cover right cos11750<br />

extension spiral cord 2m for additional keyboard cos12922<br />

[12.C] H<strong>and</strong>rails adjustable with gear tooth locking elements<br />

The h<strong>and</strong>rails adjustable in height <strong>and</strong> width, are only<br />

available for the <strong>running</strong> <strong>machine</strong>s with surface 150/50cm (e.g.<br />

h/p/cosmos mercury med). The h<strong>and</strong>rails offer best possible<br />

safety at best attainable freedom of body motion.<br />

The adjustable h<strong>and</strong>rails are useful for insecure or elderly<br />

people <strong>and</strong> especially for all therapies <strong>and</strong> exercise with<br />

children.<br />

The h<strong>and</strong>rails are to be adjusted individually at the two joints<br />

with scales at the back end. The two stabilising elements at<br />

the front are responsible for the fixation of the adjusted<br />

position.<br />

If using the adjustable h<strong>and</strong>rails, the UserTerminal is placed at the crossbar in front of the testing person. Because of<br />

the tremendous variability <strong>and</strong> broad adjustments the h<strong>and</strong>rails are applicable for most body heights. Especially for<br />

children they offer additional safety in therapy <strong>and</strong> exercise.<br />

Adjustable h<strong>and</strong>rails at the h/p/cosmos mercury med Adjustment with scales / back Stabilising elements / front<br />

h/p/cosmos adjustable h<strong>and</strong>rails [cos10030]


[12.C1] Danger precaution <strong>and</strong> operation<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 169 of 213<br />

accessories & options<br />

N The maximum load is limited to persons not more than 140 kg body weight.<br />

N Do not adjuts the h<strong>and</strong>rails under strain. The h<strong>and</strong>rail must not be weighted by a person.<br />

N Do only use the adjustable h<strong>and</strong>rails under supervision of a trained staff member. Only trained staff is allowed<br />

to adjust the h<strong>and</strong>rails.<br />

N Do only loosen the gear tooth locking elements to adjust the angle, if the you hold the h<strong>and</strong>rail with one h<strong>and</strong>.<br />

Turn the respective lever with the h<strong>and</strong>, that holds the h<strong>and</strong>le, in the desired position.<br />

N Before weighting make sure that after the adjustment the gear tooth locking elements at the rear end are<br />

correctly engaged <strong>and</strong> that the two stabilization elements at the front (“clamping screws”) are tightened <strong>and</strong><br />

secured again.<br />

N While adjusting make sure that you do not pen in yourself or any other person at the danger spots (squeezing<br />

<strong>and</strong> jamming spots).<br />

N Using the h/p/cosmos arm support with the adjustable h<strong>and</strong>rails is not possible.<br />

[12.C2] Reproducibility of adjustments<br />

To obtain reproducibility you have to determine the individually optimal<br />

adjustments (see scaling at the locking elements at the back) for each<br />

subject <strong>and</strong> note it down appropriatly (e.g. in a patient file). We have<br />

deliberately refrained from giving you st<strong>and</strong>ard settings for particular heights,<br />

because the adjustment can widly differ according to the posture <strong>and</strong> the<br />

individual situation of the testing person.<br />

The torsional elements have a scaling at any joint, which is afferent when turned in direction to the <strong>running</strong><br />

surface. The adjustments of the height <strong>and</strong> width occurs gradually, because the interlocking of the joints works in<br />

6-degree steps.<br />

In the illustration below, which you can copy, e.g. for patient files, you can fill in the 4 datas of the scaling for the<br />

individual adjustments of a patient.


[12.D] H<strong>and</strong>rails adjustable. Parallel bars with gas struts <strong>and</strong> locking lever<br />

[12.D1] Safety precautions <strong>and</strong> operation<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 170 of 213<br />

accessories & options<br />

The bar h<strong>and</strong>rails adjustable in height <strong>and</strong> width are available for<br />

the h/p/cosmos <strong>running</strong> <strong>machine</strong> of the “locomotion series”<br />

exclusively. The individually adjustable h<strong>and</strong>rails offer utmost safety<br />

at best possible freedom of movement. Well suited for therapy with<br />

insecure <strong>and</strong> elderly patients as well as for therapy with children.<br />

Picture: <strong>running</strong> <strong>machine</strong> h/p/cosmos locomotion 150/50 DE med<br />

in locomotion application <strong>and</strong> in combination with<br />

“h/p/cosmos airwalk se 135” unweighting (body weight support)<br />

system, seats for the therapist <strong>and</strong> a wheelchair ramp.<br />

N The max. load is limited to 140 kg body weight.<br />

N The positioning of the adjustable bar h<strong>and</strong>rails must not be executed under load or pressure on the h<strong>and</strong>rail<br />

N The adjustment <strong>and</strong> use of the adjustable bar h<strong>and</strong>rails is reserved for trained <strong>and</strong> experienced personnel,<br />

resp. should always be done under supervision.<br />

N Pay attention to always look the locking levers before using the bar h<strong>and</strong>rails<br />

N At least once per year you have to check the function of the bar h<strong>and</strong>rails by exposing the bar h<strong>and</strong>rail to a<br />

load of approx. 200 kg (e.g. 3 persons) with locked levers.<br />

N While adjusting the h<strong>and</strong>rails take care that neither you nor others get caught or trapped at the danger zones<br />

(jamming or squeezing zones).<br />

N If turing the UserTerminal to outside position, then at least 1 emergency-stop-button has to remain in reach of<br />

the subject at any time.<br />

N The use of the h/p/cosmos arm supports is not possible with the adjustable bar h<strong>and</strong>rails.<br />

[12.D2] Adjustment of the bar h<strong>and</strong>rails<br />

Point Illustration Action<br />

[01]<br />

N Release the locking lever at the h<strong>and</strong>rail by lifting up the lever<br />

Locking lever


[02]<br />

[03]<br />

[04]<br />

[05]<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 171 of 213<br />

accessories & options<br />

N Adjust the bar h<strong>and</strong>rail to the desired height by pressing the<br />

push button.<br />

Push button<br />

N Move the bars horizontally to adjust the requested width.<br />

N Lock the locking lever after positioning the bar h<strong>and</strong>rails by<br />

pushing the lever down.<br />

N At the rear end of the parallel bars two telescope tubes are<br />

mounted inside the tubes of the h<strong>and</strong>rails. They can serve the<br />

subject when mounting the <strong>running</strong> surface with a wheelchair<br />

for example.<br />

N After mounting the <strong>running</strong> surface the telecope h<strong>and</strong>rails have<br />

to be inserted back to the h<strong>and</strong>rail tubes to avoid any obstacles<br />

<strong>and</strong> risks of injury for the therapist walking <strong>and</strong> moiving around<br />

the <strong>running</strong> <strong>machine</strong>.


[12.D3] Reproducibility of adjustments<br />

[01]<br />

[12.E] Therapist’s seat <strong>and</strong> foot support adjustable<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 172 of 213<br />

accessories & options<br />

N To obtain reproducibility you have to determine the individually<br />

optimal adjustments (see scaling at the gas pressure zylinders)<br />

for each testing person <strong>and</strong> note it down appropriatly (e.g. in a<br />

patient file). We have deliberately refrained from giving you<br />

st<strong>and</strong>ard settings for particular heights, because the adjustment<br />

can widly differ according to the posture <strong>and</strong> the individual<br />

situation of the testing person.<br />

The bilaterally positioned <strong>and</strong> ergonomically formed therapist’s seats in combination with the adjustable foot supports provide<br />

the therapist with a back friendly seating possibility while executing a treadmill therapy at a patient. The adjustable therapist’s<br />

seats <strong>and</strong> foot supports are fixed to the treadmill via a guide rail.<br />

[12.E1] Safety precautions <strong>and</strong> operation<br />

N The max. load is limited to 200 kg body weight.<br />

N The adjustment <strong>and</strong> use of the adjustable foot support/ therapist’s seat is reserved for trained <strong>and</strong> experienced<br />

personnel, resp. should always be done under supervision<br />

N If the clamp h<strong>and</strong>les are not locked sufficiently, there is danger of squeezing due to unintentional movement of<br />

the therapist’s seat or the foot support. Open <strong>and</strong> overhanging clamp h<strong>and</strong>les are a source of danger as well.<br />

Therefore pay attention to always lock the fixation screws before using the seat <strong>and</strong> foot support.<br />

N Especially during adjusting the seat <strong>and</strong> foot support take care that neither you nor others get caught or<br />

trapped at the danger zones (jamming or squeezing zones). Special attention is required at following danger<br />

zones: push button at extreme tangential deviation of the h<strong>and</strong>rail; moveable holder of the UserTerminal to the<br />

motor hood <strong>and</strong> to the pillars of the h<strong>and</strong>rails; joint heads of the h<strong>and</strong>rails, moveable foot support <strong>and</strong><br />

therapist’s seat to the main frame <strong>and</strong> to the pillars of the h<strong>and</strong>rails.<br />

[12.E2] Operation <strong>and</strong> adjustment<br />

Point Illustration Action<br />

[01]<br />

N Release the fixation screw underneath the foot support to<br />

adjust the optimal position.


[02]<br />

[03]<br />

Total view <strong>running</strong> <strong>machine</strong> h/p/cosmos locomotion 150/50 DE med<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 173 of 213<br />

accessories & options<br />

N Release the fixation screw underneath the seat to adjust the<br />

optimal seat position.<br />

N Fix the adjusted seat <strong>and</strong> foot support position by tightening the<br />

corresponding fixation screws


[12.F] Special h<strong>and</strong>rails<br />

h/p/cosmos offers a number of optional h<strong>and</strong>rails <strong>and</strong> custom made devices on request.<br />

h<strong>and</strong>rail-cross-bar adjustable h<strong>and</strong>rail-cross-bar bended with clamp<br />

fixation<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 174 of 213<br />

accessories & options<br />

h<strong>and</strong>rail on left side shorted,<br />

„speed-grip” at the left<br />

h<strong>and</strong>rail short h<strong>and</strong>rail long rolled, 2 pillars h<strong>and</strong>rail detachable / separable<br />

Details <strong>and</strong> order numbers are available at the manufacturer h/p/cosmos or at your local authorized distributor.<br />

N The crossbar always needs to be positioned in that way that at least 2/3 of the <strong>running</strong> surface<br />

behind the crossbar is free for the user.<br />

N It is not allowed to run with the back to the crossbar.<br />

N The h<strong>and</strong>rail-cross-bar should be dismounted if a safety-arch with chest-belt <strong>and</strong> harness is<br />

installed <strong>and</strong> used.<br />

N There must be a gap of at least 4cm between the crossbar <strong>and</strong> the UserTerminal due to danger of<br />

bruising.


[12.G] h/p/cosmos airwalk se 135 (available as of 10/2008)<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 175 of 213<br />

accessories & options<br />

The „h/p/cosmos airwalk se 135“ is a user friendly <strong>and</strong> novel dynamic unweighting system with central point suspension<br />

for manual locomotion therapy. It improves the manual training due to constant un-weighting of the patient in the vertical<br />

movement area.<br />

N The dynamic unweighting system with low inertia allows precise unweighting of the patient <strong>and</strong> enables a natural<br />

vertical movement for a more physiological motion.<br />

N The stageless adjustable unweighting also allows children <strong>and</strong> lightweighted subjects an effective motion training.<br />

N Easy operation: The subject can be lifted as well as lowered by a simple key stroke. This eases the training<br />

preparation <strong>and</strong> supports the training itself.<br />

N The unweighting can be levelled precisely to the needs of the patient <strong>and</strong> ensures an optimum training<br />

environment.<br />

N High patient comfort for more motivation.<br />

N The high-class, breathable <strong>and</strong> very comfortable unweighting vest has a physiological form <strong>and</strong> is easily<br />

adjustable. This avoids dermal irritations <strong>and</strong> pressure marks to a large extent.<br />

N The proper fitting of the unweighting vest enhances the mobility of the patient <strong>and</strong> allows a more physiological<br />

movement.<br />

N Better ergonomic for better training results.<br />

N The parameters of the <strong>running</strong> <strong>machine</strong> <strong>and</strong> the unweighting system can be changed during the training (speed,<br />

elevation, unweighting).<br />

N The training phases can be extended which results in a more effective therapy for the patient.<br />

N The „h/p/cosmos airwalk se 135“ can be used for a very wide field of applications. Therapeutic applications like<br />

functional motion therapy, rehabilitation or neurology as well as competitive sports (e.g. coordination training for<br />

athlets, etc.).<br />

h/p/cosmos airwalk se [cos30017va01]


[12.H] Unweighting system h/p/cosmos airwalk<br />

Please refer to the additionally provided manual for the unweighting system<br />

for all hazard notes <strong>and</strong> instructions for use for the unweighting system<br />

including the accessories. Change cable assemblies every 6 months. In<br />

case of any sign of wear change cable assemblies even earlier than within 6<br />

months.<br />

The h/p/cosmos airwalk is a pneumatic unweighting system with dynamic<br />

central-point suspension. With the help of the system locomotion therapy<br />

can start at an early stage, all exercises can be done in an upright functional<br />

position <strong>and</strong> a wider range of exercises is possible.<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 176 of 213<br />

accessories & options<br />

It is possible to use the h/p/cosmos airwalk with the <strong>running</strong> <strong>machine</strong> series with track size 150/50cm or 170/65cm or<br />

190/65cm, <strong>and</strong> as well with most <strong>running</strong> <strong>machine</strong>s produced by other manufacturers.<br />

[12.H1] Vests for subjects for h/p/cosmos airwalk<br />

The vests offer high compatibility at best attainable freedom of body motion. They<br />

do not have undercutting straps during gait <strong>and</strong> have an optimum fitting due to the<br />

individually adjustable strap system.<br />

4 different vests provide adaptation to all body sizes <strong>and</strong> forms. A special vest for<br />

children is also available. Additional straps for gait correction vest are included in<br />

the shipment. The unweighting is adjustable by the testing person itself.<br />

Models<br />

For the models h/p/cosmos airwalk 70 <strong>and</strong> h/p/cosmos airwalk 160 a compressed air supply or a separate compressor<br />

is necessary.<br />

h/p/cosmos airwalk 70 [cos10092]<br />

h/p/cosmos airwalk 160 [cos10093]<br />

Compressor [cos10094]<br />

h/p/cosmos airwalk vest red, waist 55-88 cm [cos10095]<br />

h/p/cosmos airwalk vest blue, waist 81-112 cm [cos10096]<br />

h/p/cosmos airwalk vest yellow, waist 112-145 cm [cos10097]<br />

h/p/cosmos airwalk vest children [cos10112]<br />

Chest-harnesses, patient-vests, waist-belts with leash <strong>and</strong> cuffs are not designed for direct skin or<br />

mucous membrane contact. Wear appropriate clothing.


[12.I] h/p/cosmos robowalk ® exp<strong>and</strong>er<br />

[12.J] Wheelchair ramp<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 177 of 213<br />

accessories & options<br />

how does the h/p/cosmos robowalk ® exp<strong>and</strong>er work?<br />

First, elastic cables are attached to patient´s legs with comfortable leg<br />

cuffs. As the patient walks, the cables at the front assist the movement of<br />

the legs with support. The cables at the rear can be used also as<br />

resistance <strong>and</strong> for gait correction training. Both the front <strong>and</strong> back<br />

system can be utilized together for even greater training effects. By<br />

adjusting the angle of the support/resistance cables either vertically or<br />

horizontally, movement correction is possible. The patented tension<br />

adjustment module involves readable scales on each cable for tension<br />

monitoring. Use only with safety arch with chest belt <strong>and</strong> fall stop!<br />

Detailed instructions see separate operation manual for robowalk.<br />

When using the wheelchair access ramp pay particular attention to the gap<br />

between ramp <strong>and</strong> <strong>running</strong> belt that h<strong>and</strong>s, hair, clothing or other items do not get<br />

caught. The ramp should not touch the <strong>running</strong> belt as this will lead to excessive<br />

wear. Ensure that the ramp is stable <strong>and</strong> cannot slip. Be aware of danger of<br />

slipping or tripping on ramp. Do not attempt to mount ramp from the side. Only<br />

original h/p/cosmos wheelchair access ramps should be used.<br />

wheelchair ramp 150/50 [cos16186]<br />

[12.K] h/p/cosmos wheel chair stabilizer for oversize <strong>running</strong> <strong>machine</strong>s<br />

With the help of the h/p/cosmos wheel chair stabilizer it is easy<br />

to keep wheel chairs with 4 or 3 wheels safely in lane. The<br />

guide rail is fixed between the two pillars of the right-h<strong>and</strong> side<br />

h<strong>and</strong>rail. The metal spring, which has to be mounted on the<br />

guide rail, prevents the wheel chair from rolling too far back.<br />

The wheel chair stabilizer is available for the h/p/cosmos<br />

<strong>running</strong> <strong>machine</strong> series with track size 200/75cm (<strong>and</strong> bigger)<br />

with the mark “r”. The wheelchair application is only allowed in<br />

combination with the safety harness (chest belt <strong>and</strong> fall-stop).<br />

Take care that the brakes at the wheelchair are deactivated<br />

(e.g. demount brake suspension).<br />

Before using the <strong>running</strong> <strong>machine</strong> for any application without a wheelchair (<strong>running</strong>, cycling,<br />

skating etc.), the wheelchair stabilizer has to be dismantled for safety reasons. No remaining<br />

dangerous barrier, crossbar or other part of the wheelchair stabilizer must obstruct or injure the subject.


[12.L] Safety arch with chest belt <strong>and</strong> harness<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 178 of 213<br />

accessories & options<br />

N Every time before using the safety arch with harness a visual examination of the system has<br />

to done (examination for damages <strong>and</strong> wear). Especially the rope, the harness <strong>and</strong> all links<br />

like the snap hook <strong>and</strong> the rope brake have to be checked.<br />

N All wearing parts of the system (rope, harness <strong>and</strong> all links like the snap hook <strong>and</strong> the rope<br />

brake) have to be changed immediately in case of any damages or wear. At least every two<br />

years all wearing parts have to be changed.<br />

Indispensable accessory for stress testing, accident<br />

prevention <strong>and</strong> for the general safety of athletes: The<br />

safety arch with chest belt <strong>and</strong> harness. Automatic stop of<br />

the <strong>running</strong> belt in the moment a fall occurs. The chest<br />

belt secures the subject <strong>and</strong> prevents falling over to the<br />

front. The h/p/cosmos safety arch is available for all<br />

h/p/cosmos <strong>running</strong> <strong>machine</strong>s.<br />

The chest belt is connected via harness with the<br />

emergency stop at the crossbar of the safety arch. If<br />

there is more than 8 kg / 17.6 lb traction pulling on the<br />

harness, the switch will stop the <strong>running</strong> belt immediately.<br />

The length of the harness is individually adjustable.<br />

The chest belt straps do not pinch or restrict the athlete. The chestbelt has to be<br />

put on with the h/p/cosmos logo at the front. The vertical belt (1), to attach the<br />

carabiner, has to be put on below the shoulder<br />

belts (2), facing the body, to unload the seem at<br />

the joint (3). If the chestbelt is put on back to front<br />

or inside out or twisted, a shearing strain can occur<br />

at the joints <strong>and</strong> then lead to a loosening of the<br />

joints.<br />

Despite the fast stop of the <strong>running</strong> belt skin-abrasion at the legs might occur due<br />

to the <strong>running</strong> out of the belt, because a fallen test person still touches <strong>running</strong> out<br />

<strong>running</strong> belt for a few seconds.<br />

Nevertheless the risk of injuries is minimized drastically compared to a fall at systems without the arch <strong>and</strong> compared<br />

with falls outdoors. The maximum body weight is 140 kg / 309 lbs.<br />

The ceiling height should be at least 2.46m / 97.6 inches (0 % elevation) rsp. 2.57 / 101.6 inches (24 %) if used with<br />

h/p/cosmos mercury <strong>and</strong> 2.51 / 96.5 inches (0 %) rsp. 2.68 m / 105.5 inches (max. elevation) if used with <strong>running</strong><br />

<strong>machine</strong>s h/p/cosmos quasar or h/p/cosmos pulsar.


picture type of stop function interface<br />

communication<br />

Quick Stop<br />

on Safety Arch<br />

with chest belt,<br />

harness/rope <strong>and</strong><br />

fall stop<br />

prevention.<br />

= pull-cord<br />

operated switch<br />

N subject wears chest belt.<br />

safety rope is pulled with<br />

more than approx. 8 kg<br />

N <strong>running</strong> belt & inclination<br />

system stop very fast<br />

N drive motor is used to stop<br />

the <strong>running</strong> belt very fast on<br />

the oversize <strong>machine</strong>s<br />

venus/saturn<br />

N <strong>machine</strong> is NOT cut off from<br />

main power of drive motor<br />

<strong>and</strong> inclination system<br />

N display active <strong>and</strong> shows<br />

“PULL STOP”<br />

N interface active, but no start,<br />

speed or elevation signal<br />

executed<br />

N keyboard inactive<br />

h/p/cosmos safety arch with chest belt <strong>and</strong> harness [cos10079]<br />

(indicate model of <strong>running</strong> <strong>machine</strong>)<br />

communication<br />

NOT interrupted<br />

during Quick-<br />

Stop <strong>and</strong>/or<br />

“pull stop” stage<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 179 of 213<br />

accessories & options<br />

how to restart<br />

N release weight<br />

from safety rope<br />

N press START key,<br />

N or new comm<strong>and</strong><br />

through interface<br />

Chest-harnesses, patient-vests, waist-belts with leash <strong>and</strong> cuffs are not designed for direct skin or<br />

mucous membrane contact. Wear appropriate clothing.<br />

[12.M] Running belt rotation reversible for downhill walk<br />

N As the subject is <strong>running</strong> against the usual <strong>running</strong> belt rotation <strong>and</strong> therefore has no control<br />

of the UserTerminal it is necessary to have a supervisor who controls the <strong>running</strong> <strong>machine</strong><br />

during the training. For safety reasons the max. speed has been limited with the models<br />

possessing a cross-bar h<strong>and</strong>rail or a motor case cover. Use a safety arch with harness for<br />

additional safety.<br />

N Turn the key-operated switch for the reverse belt rotation only when the <strong>running</strong> belt is not<br />

moving.<br />

The function <strong>running</strong> belt rotation reverse (optional) permits a simulation of a downhill run. To prevent any abuse, the<br />

function can only be activated by a key-operated switch (on the external UserTerminal of the oversize <strong>running</strong> <strong>machine</strong><br />

by rotary switch).<br />

N If the key-operated switch is being activated at st<strong>and</strong> still, the <strong>running</strong> belt rotation reverse will be in function at the<br />

next start.<br />

N If the key-operated switch is being activated while the Running-Machine is in operation, the Running-Machine stops<br />

<strong>and</strong> the <strong>running</strong> belt rotation reverse will be in function at the next start.


file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 180 of 213<br />

accessories & options<br />

The <strong>running</strong> belt rotation reverse will be announced by an acoustic signal <strong>and</strong> the display = "r." Speed will be<br />

indicated as a negative value while in operation = „-0.0 ... -5.0“.<br />

[12.M1] Pull Cord Actuator (lanyard + magnet)<br />

This stop function is a safety feature that is<br />

used with a waist belt. Especially with<br />

reverse belt rotation (downhill) it is also<br />

important to connect. The athlete or patient<br />

is connected to the magnetic switch at the<br />

UserTerminal via a ripcord at the waist belt.<br />

In case of an emergency the magnet gets<br />

ripped of from the UserTerminal <strong>and</strong> the<br />

<strong>running</strong> <strong>machine</strong> stops immediately. The<br />

ripcord is not a prevention of falling. At<br />

removed magnet „PULL STOP“ will occur in<br />

the display.<br />

Safety harnesses such as safety stop cord type (safety lanyard with pull-cord actuator) <strong>and</strong>/or safety arch with<br />

harness, chest belt <strong>and</strong> stop function <strong>and</strong>/or unweighting systems with patient vest <strong>and</strong> harness shall be adjusted<br />

in a way, that the subject can not use the rear 30% of the <strong>running</strong> deck length.<br />

Pull Cord Actuator (lanyard + magnet) with STOP function: [cos00098050043]<br />

Chest-harnesses, patient-vests, waist-belts with leash <strong>and</strong> cuffs are not designed for direct<br />

skin or mucous membrane contact. Wear appropriate clothing.<br />

picture type of stop function interface<br />

communication<br />

Pull-cord<br />

actuator (lanyard<br />

with<br />

magnetic switch<br />

on the<br />

UserTerminal <strong>and</strong><br />

waist belt on the<br />

subject)<br />

N magnet is pulled off through<br />

the cord<br />

N <strong>running</strong> belt & inclination<br />

system stop fast<br />

N <strong>machine</strong> is NOT cut off from<br />

main power of drive motor<br />

<strong>and</strong> inclination system<br />

N display active <strong>and</strong> shows<br />

“PULL STOP”<br />

N interface active, but no start,<br />

speed or elevation signal<br />

executed<br />

N keyboard inactive<br />

communication<br />

NOT interrupted<br />

during<br />

“pull stop” stage<br />

how to restart<br />

N reconnect magnet<br />

to its position on<br />

UserTerminal<br />

N press START key,<br />

N or new comm<strong>and</strong><br />

through interface


[12.M2] Change from forward to reverse rotation<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 181 of 213<br />

accessories & options<br />

N Turn the key-operated switch <strong>and</strong> start the <strong>machine</strong> without a subject on the <strong>running</strong> belt only when the belt is<br />

not moving.<br />

N Observe the belt position for 2 minutes <strong>and</strong> correct the belt position with the belt adjustment screw (see<br />

chapter maintenance <strong>and</strong> belt adjustment). It is normal, that after changing the belt rotation, the belt will move<br />

to one side <strong>and</strong> therefore the <strong>running</strong> belt needs to be readjusted to the center of the <strong>running</strong> track.<br />

N Once the belt is readjusted, STOP the belt <strong>and</strong> operate the <strong>machine</strong> with the subject on the belt.<br />

[12.M3] Change from reverse to forward rotation<br />

N Turn the key-operated switch <strong>and</strong> start the <strong>machine</strong> without a subject on the <strong>running</strong> belt only when the belt is<br />

not moving.<br />

N Observe the belt position for 2 minutes <strong>and</strong> correct the belt position with the belt adjustment screw (see<br />

chapter maintenance <strong>and</strong> belt adjustment). It is normal, that after changing the belt rotation, the belt will move<br />

to one side <strong>and</strong> therefore the <strong>running</strong> belt needs to be readjusted.<br />

N Once the belt is readjusted, STOP the belt <strong>and</strong> operate the <strong>machine</strong> with the subject on the belt.<br />

[12.N] Bottle holder<br />

In most cases the readjustment of the <strong>running</strong> belt requires the same revolutions of the belt adjustment screw.<br />

Therefore we recommend to note the required revolutions of the belt adjustment screw. This will help to avoid the<br />

waiting time <strong>and</strong> can reduce the belt adjustment procedure to some seconds effort.<br />

The bottle holder is available for all h/p/cosmos <strong>running</strong> <strong>machine</strong>s <strong>and</strong><br />

can be easily fixed with a clamp on the h<strong>and</strong>rail. The position of the<br />

bottle holder must not be at the rear end of the h<strong>and</strong>rail due to this end<br />

must be reachable for the subject.<br />

bottle holder for h<strong>and</strong>rail 60mm [cos11020]<br />

[12.O] Retrofitting UserTerminal<br />

All lt-models can be retro-fitted with a UserTerminal. h/p/cosmos<br />

oversized <strong>running</strong> <strong>machine</strong>s, which are equipped with an external control<br />

unit, can be complemented optionally with an additional UserTerminal at<br />

the h<strong>and</strong>rail.<br />

In this case the external UserTerminal is connected via interface to the<br />

<strong>running</strong> <strong>machine</strong>. To connect other devices, e.g. PC or ECG, an<br />

additional interface (com 2 or com3) at the <strong>running</strong> <strong>machine</strong> is<br />

necessary.


file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 182 of 213<br />

accessories & options<br />

picture description order number<br />

blind cover cos100565 for "lt-<br />

UserTerminal":<br />

display circuit board cos14292:<br />

plus keyboard film cos10198-01:<br />

UserTerminal Touchpanel retro fit kit for h<strong>and</strong>rail,<br />

diameter: 60 mm<br />

with PC software h/p/cosmos para control. connected via<br />

RS232 interface to the <strong>running</strong>-<strong>machine</strong><br />

including external power supply [cos100820] 90...260<br />

VAC to 12 VDC / 5 A<br />

Mounting at h<strong>and</strong>rail, diameter 60 mm<br />

for h/p/cosmos retro-fitting additional UserTerminal<br />

(MCU4), or for h/p/cosmos retro-fitting additional<br />

UserTerminal TouchPanel<br />

UserTerminal Touchpanel remote control<br />

complete retrofitting kit for table<br />

(without holder for h<strong>and</strong>rail of the treadmill) with st<strong>and</strong> for<br />

desktop with PC software h/p/cosmos para control.<br />

connected via 5 meter RS232 interface to the <strong>running</strong><strong>machine</strong><br />

including external power supply [cos100820]<br />

90...260 VAC to 12 VDC / 5 A<br />

UserTerminal Display Keyboard MCU5 retrofit kit.<br />

Retrofit-kit UserTerminal Display Keyboard, extra charge<br />

for travel costs <strong>and</strong> installation;<br />

For h/p/comos <strong>running</strong> <strong>machine</strong>s with MCU5 <strong>and</strong> preinstalled<br />

UserTerminal housing <strong>and</strong> blind cover<br />

cos100565. h/p/cosmos "lt-models" (example h/p/cosmos<br />

mercury lt med) manufacturing date from approx. middle<br />

of January 2009.<br />

Instead of the existing blind-cover on the pre-installed<br />

UserTerminal a circuit board with displays cos14292<br />

including 6 pieces of LCD displays (4 digits, 7-segments),<br />

24 LEDs <strong>and</strong> a keyboard film cos10198-01 with 6 keys<br />

will be installed on the UserTerminal. Travel expenses<br />

<strong>and</strong> labor costs are not yet included in this article <strong>and</strong><br />

have to be charged extra. When placing the order for this<br />

article please include the serial number of the <strong>running</strong><br />

<strong>machine</strong>.<br />

cos14569-02<br />

cos13514<br />

cos10002-02<br />

cos15312


[12.P] Retrofitting / additional safety stop / emergency stop<br />

order number product description picture<br />

cos100548 emergency stop-button magnet holder 10m (incl. 10m conecting cable)<br />

external emergency stop button with magnetic holder for easy fixation on<br />

h/p/cosmos <strong>running</strong> <strong>machine</strong>s, for example on h<strong>and</strong>rail with diameter 60mm<br />

Cable 10 m (no spiral cord)<br />

Plug: 3 pin XLR male<br />

cos15933 emergency stop-button magnet holder 5m (incl. XLR-plug 5m spiral cord)<br />

external emergency stop button with magnetic holder for easy fixiation on<br />

h/p/cosmos <strong>running</strong> <strong>machine</strong>s<br />

Cable 5 m spiral cord<br />

Plug: 3 pin XLR male<br />

cos15021 Emergency stop extern 2m mount 60mm (incl. clamp fixation)<br />

External emergency stop for h<strong>and</strong>rail fixation. Only for h<strong>and</strong>rails with rail<br />

diameter d=60mm<br />

To connect the external emergency stop with the treadmill you need to make<br />

sure that your treadmill has a rear panel terminal for external emergency<br />

stops. If there is no socket for external emergency stops you need to order the<br />

special rear panel terminal for add. keyboard cos10109.<br />

cos15294L10m<br />

(10 m cable)<br />

cos15294L15m<br />

(15 m cable)<br />

Emergency stop ext. 10 or 15m without fixation (incl. XLR-plug)<br />

External emergency stop without fixation on the treadmill.<br />

Including cable 10 or 15 meters <strong>and</strong> 3-pin XLR plug male.<br />

To connect the external emergency stop with the treadmill you need to make<br />

sure that your treadmill has a rear panel terminal for external emergency<br />

stops. If there is no socket for external emergency stops you need to order the<br />

special rear panel terminal for add. keyboard cos10109. In case of retrofitting<br />

please stipulate the serial number of the device.<br />

Cable 10 or 15 m, no spiral cord; Plug: 3 pin XLR male<br />

picture type of stop function interface<br />

communication<br />

Safety Stop =<br />

Emergency Stop<br />

External <strong>and</strong>/or<br />

additional<br />

emergency stop<br />

mushroom head<br />

type.<br />

Quick Stop<br />

External <strong>and</strong>/or<br />

additional<br />

emergency stop<br />

mushroom head<br />

type, connected<br />

to the “quick stop<br />

port” of the<br />

N mushroom switch is hit<br />

N <strong>running</strong> belt & inclination<br />

system stop fast<br />

N <strong>machine</strong> is cut off from main<br />

power of drive motor <strong>and</strong><br />

inclination system<br />

N display, keyboard <strong>and</strong><br />

interface inactive<br />

N mushroom switch is hit<br />

N <strong>running</strong> belt & inclination<br />

system stop extremely fast<br />

N drive motor is used to stop<br />

the <strong>running</strong> belt very fast<br />

N <strong>machine</strong> is NOT cut off from<br />

main power of drive motor<br />

<strong>and</strong> inclination system<br />

communication<br />

interrupted<br />

during cut off<br />

from main<br />

power<br />

communication<br />

NOT interrupted<br />

during Quick-<br />

Stop stage<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 183 of 213<br />

accessories & options<br />

how to restart<br />

N wait 1 minute<br />

N pull mushroom<br />

switch<br />

N push “power on”<br />

button / main<br />

power<br />

N press START key,<br />

N or new comm<strong>and</strong><br />

through interface<br />

N pull mushroom<br />

switch<br />

N press START key,<br />

N or new comm<strong>and</strong><br />

through interface


[12.Q] Monitor / laptop arm<br />

UserTerminal of<br />

oversize model<br />

families<br />

200-450/75-300<br />

(venus/saturn).<br />

N display active <strong>and</strong> shows<br />

“PULL STOP”<br />

N interface active, but no start,<br />

speed or elevation signal<br />

executed<br />

N keyboard inactive<br />

A movable monitor <strong>and</strong> laptop arm for the external control unit<br />

of the oversize <strong>running</strong> <strong>machine</strong>s h/p/cosmos saturn <strong>and</strong><br />

h/p/cosmos venus is optionally available. Using this arm, the<br />

monitor also can be turned in the field of vision of the athlete or<br />

patient.<br />

Monitor arm (movable) for PC monitor or Laptop<br />

[cos13320-01]<br />

Laptop computer [cos13476]<br />

[12.R] Safety multiple socket<br />

The stable safety multiple socket [cos100157] can be used for<br />

the safe connection of a medical h/p/cosmos treadmill with<br />

peripheral devices such as PCs, printer or monitors.<br />

It meets all requirements for multiple sockets for creating<br />

connections of a medical electrical system within the patients<br />

environment.<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 184 of 213<br />

accessories & options<br />

safety multiple socket [cos100157]<br />

As claimed in the norm EN60601-1 for medical electrical devices, the multiple socked is accessible only with the help of<br />

a tool. With this method an unintentionally made connection of further devices to the system is avoided, which lowers<br />

the risk of an exceedance of the maximum allowed patient-leakage-current of 0.5mA at the complete system.<br />

N Due to the very high load of treadmills it is never allowed to connect more than 1 treadmill or<br />

other consumers with high load do the safety multiple socket. Use a dedicated line only.<br />

N The power consumption of the consumers which are connected in addition to the treadmill-<br />

ergometer to the safety multiple socket must be in total lower than 300W.<br />

N When connecting treadmills to the safety multiple socket, there are no performance-limitations<br />

to expect at low-performance applications (walking, slow <strong>running</strong>). For high-performance<br />

applications (fast <strong>running</strong>, sprints, etc.) the use of the safety multiple socket can result in<br />

performance limitations.<br />

N For high performance applications a 3-phase powered treadmill is recommended.


[12.S] Digital (serial) RS232 interface / USB converter<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 185 of 213<br />

accessories & options<br />

Connecting a <strong>running</strong> <strong>machine</strong> to a medical device results in a medical system. Always use IEC<br />

601-1 approved potential isolation components when linking medical devices via interface. Only<br />

authorised trained staff is entitled to perform this connection. This medical system, when<br />

connecting a <strong>running</strong> <strong>machine</strong> to a medical device, has to be connected via a potential-balance-<br />

cable with the provided connector bolt <strong>and</strong> with the provided bearing within the medically used<br />

room.<br />

The interface serves as a data transmitter from the <strong>running</strong> <strong>machine</strong> to a PC or other controlling devices. You can also<br />

control the Running-Machine from another controlling device (ECG. o.a.) or with the appropriate Software from a PC, via<br />

interface. St<strong>and</strong>ard for all devices is ONE serial interface COM 1. Optionally there are further serial interfaces COM 2,<br />

COM3 <strong>and</strong> COM4 available. For further information please contact your local dealer or h/p/cosmos. The interface<br />

protocol h/p/cosmos coscom © is available on request or can be found at www.coscom.org. The incorporated protocols<br />

(e.g. for ECG-devices) are listed in the list of options under OPTION 20 / OPTION 21. See the compatibility list of the<br />

RS232 interface in the chapter „Technical data“. More recent computers will have only an USB interface instead of the<br />

RS232 interface. For that case an „USB to RS232 interface-adapter-cable“ is available at h/p/cosmos under order<br />

number cos12769. To control via USB interface the processor must be a Pentium 1.8 GHz or higher.<br />

[12.S1] Interface cable RS232 for connection to PC<br />

To be used for h/p/cosmos para graphics ® , h/p/cosmos para control ® , PC, ECGs, Ergospirometry devices. For<br />

connecting your PC to the <strong>running</strong> <strong>machine</strong> you need an interface cable (optional equipment) 9-pole Sub- D (male<br />

- female) with crossed transmission- receive- cable (PIN 2 <strong>and</strong> PIN 3).<br />

6<br />

7<br />

8<br />

9<br />

1<br />

2<br />

3<br />

4<br />

5<br />

PC COM1 oder COM2<br />

CONNECTOR DB9<br />

Buchse / female<br />

[12.S2] Interface cable RS232 for connection to ECG Schiller<br />

R x D<br />

T x D<br />

GND<br />

[cos00097010034] for 5 m / [cos00097010035] for 10 m<br />

1<br />

2<br />

3<br />

4<br />

5<br />

6<br />

7<br />

8<br />

9<br />

Laufb<strong>and</strong> / treadmill<br />

CONNECTOR DB9<br />

Stecker / male<br />

Connection cable between <strong>running</strong> <strong>machine</strong> <strong>and</strong> ECG Schiller AT 10 / AT 60 or CS 100 / CS 200<br />

6<br />

7<br />

8<br />

9<br />

1<br />

2<br />

3<br />

4<br />

5<br />

EKG / ECG Schiller<br />

CONNECTOR DB9<br />

Stecker / male<br />

R x D<br />

T x D<br />

GND<br />

[cos00097010036] for 5 m<br />

1<br />

2<br />

3<br />

4<br />

5<br />

6<br />

7<br />

8<br />

9<br />

Laufb<strong>and</strong> / treadmill<br />

CONNECTOR DB9<br />

Stecker / male


file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 186 of 213<br />

accessories & options<br />

[12.S3] Interface cable for connection Ergospirometry JAEGER/VIASYS/CardinalHealth/CareFusion OXYCON<br />

OXYCON Alpha / Delta / Champion / Pro: Connect the <strong>running</strong> <strong>machine</strong> interface with an available serial interface<br />

(COM1 ... COM4) at the PC of the OXYCON. Do not use the female connector at the OXYCON with the letters<br />

"treadmill / <strong>running</strong> <strong>machine</strong>". Important: From January 2000 use h/p/cosmos coscom © RS232 interface protocol<br />

6<br />

7<br />

8<br />

9<br />

1<br />

2<br />

3<br />

4<br />

5<br />

OXYCON<br />

CONNECTOR DB9<br />

Buchse / female<br />

R x D<br />

T x D<br />

GND<br />

[cos00097010034] for 5 m / [cos00097010035] for 10 m<br />

1<br />

2<br />

3<br />

4<br />

5<br />

6<br />

7<br />

8<br />

9<br />

Laufb<strong>and</strong> / treadmill<br />

CONNECTOR DB9<br />

Stecker / male


[12.T] Printer protocol<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 187 of 213<br />

accessories & options<br />

Connect the printer (with PCL printer language <strong>and</strong> with a serial interface) via interface cable with the serial interface<br />

RS232 of the <strong>running</strong> <strong>machine</strong>. Attention: For a printer with a parallel interface you will need a commercial interfaceconverter,<br />

available in every computer-store, at your local dealer or from the manufacturer (h/p/cosmos order no.<br />

[cos10056]). If Option 20: 3 is chosen (printer protocol see list of options), following data will be printed out:<br />

Interface converter for printer [cos10056]


[12.U] POLAR Heart rate measurement system<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 188 of 213<br />

accessories & options<br />

Pay attention to the safety instructions in the instruction manual. Stop training immediately if<br />

feeling sick or dizzy <strong>and</strong> see your doctor. In the case of interference do not rely on the indicated<br />

values of the wireless heart rate transmission. Precision of heart rate measuring: + / - 1 %, resp.<br />

+ / - 1 beat per minute. Pay attention to the advices about safety <strong>and</strong> warnings <strong>and</strong> possible<br />

interferences in the chapter errors, technical data, EMC, <strong>and</strong> installation of this manual.<br />

WARNING: Heart rate monitoring systems can be inaccurate. Incorrect <strong>and</strong>/or forbidden use<br />

<strong>and</strong>/or overtraining can lead to serious injury or even death. If you feel close to fainting, the<br />

training has to be stopped immediately.<br />

[12.U1] Note from POLAR for patients with pacemaker<br />

The expert´s report from the Pacemaker-Institute in Rothenberg / Germany stated, that a negative influence or<br />

danger for patients with implanted pacemakers is impossible. (Pacemaker-Institute, Rothenberg Süd 18, 82431<br />

Kochel am See / Germany). Nevertheless we recommend all patients with pacemakers to consult their doctor<br />

before using the POLAR heart rate measurement system.<br />

[12.U2] Chest belt <strong>and</strong> transmitter<br />

You can use all POLAR transmitters (coded <strong>and</strong> un-coded) with your <strong>running</strong> <strong>machine</strong>. For <strong>running</strong> <strong>machine</strong>s of<br />

generation 4.0 with firmware-version lower than V 3.01.1 an EPROM update is available for the use of coded<br />

transmitters. Note: Even with using the coded senders the transmission of the heart rate to the treadmill is still uncoded.<br />

The models h/p/cosmos locomotion <strong>and</strong> h/p/cosmos pulsar <strong>and</strong> h/p/cosmos venus <strong>and</strong> h/p/cosmos saturn have<br />

the POLAR W.I.N.D. transmitter with WearLink technology <strong>and</strong> coded transmission.<br />

The transmitter is only activated if placed correctly on the body. The chest belt is washable. In this case loosen the<br />

belt from the transmitter. Please take care not to crush the two electrodes of the belt. After having loosened the<br />

transmitter off the belt, wash the belt <strong>and</strong> especially the electrodes with warm water <strong>and</strong> mild soap. Do not clean<br />

the electrodes mechanically. Do not use alcohol.<br />

[12.U3] Radius of transmission<br />

The transmitter radius amounts to approx. 80 to max. 120 cm. If you run several Running-Machines or other<br />

training devices on the same heart rate measurement system, you should keep a gap of at least 100 cm between<br />

the devices to exclude interference.<br />

The POLAR W.I.N.D. transmitter has a transmission range of approx. 10 meters.


[12.U4] Battery of the transmitter<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 189 of 213<br />

accessories & options<br />

At an average application of 2 hours per day the life-span of the battery amounts to approx. 1 year. The receiver is<br />

incorporated in the Running-Machine <strong>and</strong> requires no battery. It is supplied by the <strong>running</strong> <strong>machine</strong>.<br />

If an irregular pulse display occurs in spite of an unobjectionable technical condition, please<br />

check your pulse manually or in case of doubts see your doctor for a check.<br />

[12.U5] Correct placement of transmitter belt<br />

Adjust the belt length so that the belt fits<br />

tightly but does not confine you. The belt<br />

should not loosen while exercising. Close<br />

the belt with transmitter placed outwards<br />

(POLAR-Logo in right position).<br />

In order to allow an optimal skin contact the<br />

skin should be moist. Contact gel, as used<br />

for ECG, is an excellent solution. Moisten<br />

the two electrodes <strong>and</strong> the skin with water<br />

or contact gel, which is available in chemistries.<br />

Place the transmitter so that it is right below the pectoral muscle (chest), as is shown in the illustration above. For<br />

ca. 85 – 88 % of the subjects the "normal" placement of the transmitter belt is recommended.<br />

[12.U6] Transmission problems<br />

N For ca. 10 – 15 % of the subjects the placement "POLAR-logo upside down" is recommended.<br />

N For ca. 1 – 2 % of the subjects the placement "transmitter at the back of the subject" is recommended.<br />

However, try the "normal" position always first.<br />

[12.V] PC Software h/p/cosmos para control ® 4.1<br />

The free PC software h/p/cosmos para control ® is the result of a very large enhancement <strong>and</strong> function expansion of the<br />

already well-known version 3.0. In addition to the pure remote control function of h/p/cosmos treadmill-ergometers over<br />

the pc the software has integrated now a lot of additional functions such as an option setting configuration tool with<br />

integrated storage function, firmware update over flash eprom, quick-stop-key, cool-down-key, pause-key, target speed<br />

settings, target elevation settings, stop-count-down, display reset, etc.<br />

The software is not designed for medical use.<br />

Most of the additional functions of h/p/cosmos para control ® require an interface connection over the sophisticated<br />

coscom v3 protocol, which is integrated in all h/p/cosmos treadmill-ergometers with MCU5 control electronic (as of<br />

manufacturing date 12/2007). Treadmills which are equipped with MCU4 control electronic (until manufacturing date<br />

12/2007) have integrated the forerunner protocol coscom v2 but they can be refitted with coscom v3 by a simple<br />

firmware-update.


file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 190 of 213<br />

accessories & options<br />

h/p/cosmos para control ® can be found on the h/p/cosmos demo & info DVD [cos10067] which is delivered together<br />

with every h/p/cosmos treadmill-ergometer. Alternatively the software can be downloaded at<br />

www.h-p-cosmos.com/en/software. At aforementioned link also a manual can be found in which all functions of the new<br />

para control are described in detail.<br />

For warnings, system requirements <strong>and</strong> application regulations of the software please read the separate software<br />

manual.<br />

For the data transmission a connection cable from the <strong>running</strong> <strong>machine</strong> to the personal computer is required. This<br />

connection cable is available as optional equipment (see also optional equipment serial interface).<br />

Interface connection cable RS 232 5m [cos00097010034]<br />

Interface connection cable RS 232 10m [cos00097010035]<br />

USB to RS232 interface adapter-cable [cos12769-01]<br />

Applications<br />

N Remote control of all <strong>running</strong> <strong>machine</strong> functions during tests or exercise via interface from the PC.<br />

N Demonstrations for big groups (students, audience, etc.) with the help of a big screen or beamer.<br />

N Control of multiple h/p/cosmos treadmill-ergometers over 1 PC.<br />

N Firmware-update over Flash-Eprom.<br />

N Service & maintenance work, setup, programming <strong>and</strong> display of error codes (obligatory for <strong>running</strong>-<strong>machine</strong>s<br />

without hardware UserTerminal).<br />

N Testing the RS232 interface of the PC, testing the RS232 interface cable in case of communication problems<br />

between ECG or Ergospirometry.<br />

h/p/cosmos para control ® 4.1 [cos10071 -V4.1.0]


[12.W] PC Software h/p/cosmos para graphics ®<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 191 of 213<br />

accessories & options<br />

Especially in performance diagnostics the software h/p/cosmos para graphics ® supports the user with processes<br />

automation. Therefore the user is able to fully concentrate on the test person. But also in sports science applications or<br />

jus to control devices <strong>and</strong> for documentation of exercise programs, the software is perfectly suitable. h/p/cosmos para<br />

graphics ® permits an on-line registration <strong>and</strong> administration of Heart-Rate-values <strong>and</strong> endurance-parameters. The<br />

treadmill can be controlled automatically. The software is not designed for medical use.<br />

The data can be converted by the software <strong>and</strong> be taken over <strong>and</strong> evaluated e.g. by EXCEL, or h/p/cosmos para<br />

analysis software. A take over <strong>and</strong> processing of the data within a spreadsheet (EXCEL, etc.) or an editor is also<br />

possible with the help of the h/p/cosmos para graphics ® .<br />

Further information is available in the separate instruction manual of the software h/p/cosmos para graphics ® . Also a<br />

demo version is available for further tests.<br />

For the data transmission a connection cable from the <strong>running</strong> <strong>machine</strong> to the personal computer is needed. This<br />

connection cable is available as optional equipment (see also optional equipment serial interface).<br />

For danger precautions, regulations for application <strong>and</strong> system requirements of the software read the separate software<br />

manual.<br />

h/p/cosmos para graphics ® [cos10156]


[12.X] PC Software h/p/cosmos para analysis ®<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 192 of 213<br />

accessories & options<br />

The software h/p/cosmos para analysis ® is an evaluation software for performance diagnostics <strong>and</strong> training control that<br />

allows the creation of individual training plans.The evaluation of heart rate <strong>and</strong> lactate curves after graded test is<br />

possible on the basis of individual selected threshold models in graphics <strong>and</strong> tables. It is possible to enter the test data<br />

in h/h/p/cosmos para analysis ® usually or by importing it from h/p/cosmos para graphics ® clear design of the program<br />

makes the evaluation easy. The comprehensive printout simplifies the interpretation <strong>and</strong> is the basis for best possible<br />

coaching.<br />

The software is not designed for medical use.<br />

The software enables to adapt the integrated test methods <strong>and</strong> to generate individual ones. Test devices - even field<br />

test as well as time <strong>and</strong> performance parameters are free definable. It is possible to enter the test data manually or by<br />

importing it from h/p/cosmos para graphics ® . The entry <strong>and</strong> evaluation of desired parameters, e.g. lactat or heart rate as<br />

well as blood pressure or the Borg scale can be defined or added. Data of the software PA7000 <strong>and</strong> of the devices of<br />

the POLAR S Series can be imported as well.<br />

The models for the evaluation of the individual anaerobic threshold - also several simultaneously for the comparison<br />

option - are free definable. Besides the different models to evaluate the individual anaerobic threshold at the lactate<br />

performance diagnostics it is possible to create fixed thresholds. It is possible to make an evaluation according to<br />

Conconi. The evaluation occurs graphical <strong>and</strong> tabulated. The scaling of the diagram can be adjusted individually. The<br />

clear graphics of training categories <strong>and</strong> intensities are the basis for successful training plans. For danger precautions,<br />

regulations for application <strong>and</strong> system requirements of the software read the separate software manual.<br />

h/p/cosmos para analysis ® [cos10155]


[12.Y] PC Software h/p/cosmos para motion<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 193 of 213<br />

accessories & options<br />

The h/p/cosmos para motion is a software for recording <strong>and</strong> evaluation of video analysis with individual report for<br />

printout, incl. control functions of h/p/cosmos <strong>running</strong> <strong>machine</strong>s. The software is not designed for medical use.<br />

Basic functions <strong>and</strong> features<br />

N Digital, software-based video recordings with treadmill control for all h/p/cosmos treadmills parallel to recording<br />

with feedback.<br />

N Creation of individual analysis patterns for different applications.<br />

N Intelligent video management system in flexible data base - the video clips are saved in the data base in<br />

accordance with the chosen analysis pattern, separate saving of data can be omitted.<br />

N <strong>Medical</strong> history tool with information database: accumulation with the most common gait <strong>and</strong> motion specific<br />

disorders with transfer possibility to the individual report.<br />

N Integrated player for the analysis of several video clips.<br />

N Video overlay of up to 8 videos.<br />

N Full screen mode.<br />

N Various 2D analysis tools such as gridlock <strong>and</strong> grid.<br />

N Flexible 2D painting tools such as rectangles, tracking lines, arrows or text fields.<br />

N Tools <strong>and</strong> measurement tools for the fast detection of angles <strong>and</strong> distances.<br />

N Definition of gait specific situations (e.g. first peak, st<strong>and</strong>ing phase) for faster synchronisation <strong>and</strong> easier analysis.<br />

N Report function for creation of individual reports.<br />

For control <strong>and</strong> data communication an interface connection cable from the <strong>running</strong> <strong>machine</strong> to the personal computer<br />

is required. The interface cable <strong>and</strong> USB-RS232 Converter <strong>and</strong> second port COM2 are available as optional accessory.<br />

h/p/cosmos para motion "basic"<br />

Item numbers<br />

cos15508<br />

h/p/cosmos para motion "st<strong>and</strong>ard" cos15509<br />

h/p/cosmos para motion "professional" cos15510<br />

RS232 interface connection cable 5 m cos00097010034<br />

USB to RS232 interface adapter-cable, 0,5 m cos12769-01<br />

Detailed specifications <strong>and</strong> operation instructions can be found in the separate software operation manual <strong>and</strong> on the<br />

website www.h-p-cosmos.com/en/software/para_motion.htm


[13.]Certificates<br />

[13.A] Certificate of the TSA (TÜV) according to ISO 9001<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 194 of 213<br />

certificates


[13.B] Certificate of the TSA (TÜV) according to EN ISO 13485<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 195 of 213<br />

certificates


[13.C] EC Certificate of the TSA (TÜV)<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 196 of 213<br />

certificates


[13.D] Certificate of compliance for sports <strong>and</strong> fitness devices<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 197 of 213<br />

certificates


[13.E] Certificate for medical devices<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 198 of 213<br />

certificates


file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 199 of 213<br />

certificates


file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 200 of 213<br />

certificates


[14.]Disposal<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 201 of 213<br />

disposal<br />

By request <strong>and</strong> at the expense of the client h/p/cosmos might take over the disposal of old devices <strong>and</strong> devices not longer<br />

functioning. Please contact service@h-p-cosmos.com for a detailed offer. Note the information for possible disposal of the<br />

<strong>running</strong> <strong>machine</strong> parts or components through the client or a subcontractor.<br />

The h/p/cosmos devices are marked with following sign / symbol on the name plate:<br />

Symbol for collection, treatment, recycling <strong>and</strong> disposal of waste electrical <strong>and</strong> electronic equipment (WEEE)<br />

as set out in Directive 2002/96/EC of January 27, 2003 of the European Parliament <strong>and</strong> of the Council on<br />

waste electrical <strong>and</strong> electronic equipment are necessary to reduce the waste management problems linked to<br />

the heavy metals concerned <strong>and</strong> the flame retardants concerned.<br />

h/p/cosmos EAR WEEE-Reg.-No. DE 42594388<br />

[14.A] Disassemble <strong>and</strong> cut up<br />

Use personal protective equipment, when cutting up material of any kind with the appropriate tools (eye-protection, dustmask,<br />

etc.). Contact service@h-p-cosmos to receive the safety-data-sheet according to European Commission Directive<br />

91/155/EEC for a material.<br />

[14.B] h/p/cosmos <strong>running</strong> <strong>machine</strong><br />

h/p/cosmos <strong>running</strong> <strong>machine</strong>s consist among other things of powder coated <strong>and</strong> galvanized metals from different<br />

producers <strong>and</strong> qualities, stainless steel parts, aluminium parts, plastics, rubber, electronics with cables, boards <strong>and</strong><br />

condensers as well as batteries. These materials can be recycled by h<strong>and</strong>ing them to the official municipal valuable<br />

substance collection or to authorized disposal partners of valuable substance disposal. Pay attention to the regulations<br />

of the disposal company. The material of the <strong>running</strong> surface (sliding plate) has the disposal key number 57109 <strong>and</strong><br />

may be recycled as well.<br />

[14.C] h/p/cosmos service oil / grease<br />

Service-oils for the belt lubrication have the disposal key numbers 130202 <strong>and</strong> 130203 <strong>and</strong> have been assigned to<br />

class I, which means, that they can be recycled according to the local provisions. Both products have not to be marked<br />

for identification according to the EU guidelines / ordinance on Hazardous Substances. For oversize <strong>running</strong> <strong>machine</strong>s<br />

with <strong>running</strong> deck size 200/75cm <strong>and</strong> bigger a bit of copper paste grease is used for the elevation system.<br />

[14.D] h/p/cosmos <strong>running</strong> belt<br />

If necessary please request the materials of the different <strong>running</strong> belts directly at service@h-p-cosmos.com, because<br />

depending on the application they have a different consistence. Correct information about the consistence you will get<br />

according to your request under naming the serial number of the <strong>running</strong> <strong>machine</strong>.


[15.]Contact<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh author: fh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 202 of 213<br />

contact<br />

For additional orders <strong>and</strong> technical enquiries please have the model type, the serial number <strong>and</strong> the date of installation of<br />

your <strong>running</strong> <strong>machine</strong> ready. In case of any further questions about delivery dates, service or maintenance, orders for<br />

consumables etc., please contact the corresponding phone, fax or email for qualified help.<br />

[15.A] Service department<br />

phone +49 86 69 86 42 0<br />

phone direct +49 86 69 86 42 21<br />

fax +49 86 69 86 42 49<br />

email service@h-p-cosmos.com<br />

[15.B] Sales department<br />

phone +49 86 69 86 42 0<br />

fax +49 86 69 86 42 49<br />

email sales@h-p-cosmos.com<br />

[15.C] Headquaters<br />

h/p/cosmos sports & medical gmbh<br />

Am Sportplatz 8<br />

DE 83365 Nussdorf-Traunstein<br />

Germany<br />

phone +49 18 05 16 76 67<br />

fax +49 18 05 16 76 69<br />

email@h-p-cosmos.com<br />

www.h-p-cosmos.com<br />

Building 1 (top picture)<br />

h/p/cosmos development & production<br />

Am Sportplatz 8<br />

DE 83365 Nussdorf-Traunstein<br />

Building 2 (picture below)<br />

h/p/cosmos sales & service<br />

Feldschneiderweg 5<br />

DE 83365 Nussdorf-Traunstein


[16.]Appendix – Instruction <strong>and</strong> commissioning<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh author: fh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 203 of 213<br />

appendix<br />

Once the installation of the <strong>running</strong> <strong>machine</strong> has been completed, the h/p/cosmos employee / h/p/cosmos partner starts with<br />

the instruction <strong>and</strong> commissioning of the device. It is important to include all people in the instruction <strong>and</strong> commissioning,<br />

who are going to work with the <strong>running</strong> <strong>machine</strong>. After the instruction is completed the instruction protocol is to be signed by<br />

the h/p/cosmos technician <strong>and</strong> all trained persons <strong>and</strong> sent back, together with the signed delivery note <strong>and</strong> the registration<br />

form, to h/p/cosmos.<br />

[16.A] Instruction into general operation<br />

[01]<br />

[02]<br />

[03]<br />

H<strong>and</strong> over instruction manual.<br />

Inform about instruction <strong>and</strong> service manual (always keep one within<br />

reach of all users).<br />

accomplished �<br />

Refer to general warnings <strong>and</strong> safety precautions<br />

according to the instruction manual. Place the print-out of<br />

the safety precautions (DIN A4 form in the delivery folder)<br />

close to the treadmill. Children are not allowed near or on the<br />

treadmill. For higher-risk applications a safety arch with chest<br />

belt <strong>and</strong> fall-stop-prevention has to be used.<br />

accomplished �<br />

Especially point out the necessary safety-zone of min. L: 200 cm x<br />

W: 100 cm (or wider if <strong>running</strong> deck is wider) in case of fall.<br />

accomplished �


[16.B] Instruction into functions<br />

[01]<br />

[02]<br />

[03]<br />

[04]<br />

[05]<br />

How to switch the <strong>running</strong> <strong>machine</strong> ON <strong>and</strong> OFF properly.<br />

1 - 2 minutes warming-up (delay) time for medical <strong>machine</strong>s.<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh author: fh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 204 of 213<br />

appendix<br />

Different ON/OFF button <strong>and</strong>/or expulsion fuse for sports <strong>and</strong> medical<br />

devices as well as ON/OFF button at the UserTerminal.<br />

Explain safety features that the actual model is equipped with for<br />

example: emergency-stop, quick-stop, rehab-stop, safety-harness, etc.<br />

Do not use the Emergency-Stop for common switching OFF of the<br />

device.<br />

1 - 2 minutes warming-up time (delay before next power on) for<br />

medical <strong>machine</strong>s.<br />

Explain the display-panels. What does which display st<strong>and</strong> for?<br />

Explain manual operation of <strong>running</strong> <strong>machine</strong>.<br />

Information about, limit of max. speed for beginners, possibility of<br />

locking of each single mode or all modes on request, <strong>and</strong> the<br />

possibility or the necessity to lock the device against unauthorised<br />

users.<br />

Operation of <strong>running</strong> <strong>machine</strong> via program mode.<br />

Refer to the list of programs in the instruction manual.<br />

accomplished �<br />

accomplished �<br />

accomplished �<br />

accomplished �<br />

accomplished �


[06]<br />

[07]<br />

[08]<br />

[09]<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh author: fh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 205 of 213<br />

appendix<br />

Operation of <strong>running</strong> <strong>machine</strong> via cardio-mode (heart rate controlled<br />

load).<br />

Inform about correct heart rate measurement <strong>and</strong> limitations: correct<br />

adjustment of chest belt, how to h<strong>and</strong>le problems <strong>and</strong> disorders,<br />

possible causes for interferences <strong>and</strong> failures (PCs, quartz watches,<br />

monitors, cables, etc.).<br />

Explanation of test-mode <strong>and</strong> the integrated test-programs, inform<br />

about interface <strong>and</strong> compatibility list. In medical field potential isolation<br />

for any interface linkage is obligatory! Only connect to coscom v3<br />

compatible <strong>and</strong> authorized host equipment. Do not connect to<br />

equipment or software, which is not explicitly declared as compatible<br />

by h/p/cosmos <strong>and</strong> by the other manufacturer. Inform about the<br />

h/p/cosmos software programs h/p/cosmos para control, h/p/cosmos<br />

para graphics, h/p/cosmos para analysis <strong>and</strong> h/p/cosmos para motion<br />

for documentation <strong>and</strong> visualization of data.<br />

If device includes reverse-belt rotation: adjustment of <strong>running</strong> belt.<br />

Customer must try to adjust at least once. Note down necessary<br />

settings <strong>and</strong> adjustments.<br />

rotations: � ¼ � ½ � ¾ � 1 �_______<br />

direction: � clockwise � anticlockwise<br />

accomplished �<br />

accomplished �<br />

accomplished �<br />

accomplished �


[16.C] Instruction into maintenance works<br />

[01]<br />

[02]<br />

[03]<br />

[04]<br />

[05]<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh author: fh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 206 of 213<br />

appendix<br />

H<strong>and</strong> over <strong>and</strong> explain service-box. The manufacture recommends a<br />

service contract through an authorised h/p/cosmos service technician.<br />

Lack of maintenance, poor maintenance or unauthorised<br />

servicing/repairs together with improper or lack of safety checks can<br />

lead to a risk of injury to users or damage to the <strong>running</strong> <strong>machine</strong>, this<br />

could also lead to loss of manufactures warranty <strong>and</strong> /or third party<br />

liability.<br />

Correct adjustment of leveling sockets.<br />

Check weight load on the leveling sockets by trying to lift up the<br />

treadmill frame.<br />

Proper adjustment of <strong>running</strong> belt with regulation screw <strong>and</strong> Allen-key.<br />

For each change of <strong>running</strong> belt direction there are two adjustment<br />

steps:<br />

a) Turn key at the back of the UserTerminal<br />

b) Adjust regulation screw according to previously defined value<br />

c) The allen-key must be pulled out of the screw immediately after<br />

usage due to risk of injury!<br />

Lubrication of <strong>running</strong> belt when "OIL"-message (quantity, intervals<br />

according to instruction manual, spread the oil by "Duck-walking").<br />

Use only original h/p/cosmos lubricant authorized for this model of<br />

<strong>running</strong> <strong>machine</strong>. Observe dem<strong>and</strong> for additional lubrication based on<br />

manual.<br />

After <strong>running</strong> belt lubrication the position of the <strong>running</strong> belt must be<br />

checked once more <strong>and</strong> be adjusted if necessary.<br />

Delete „oil“-message with Option 01.<br />

accomplished �<br />

accomplished �<br />

accomplished �<br />

accomplished �<br />

accomplished �


[06]<br />

[07]<br />

[16.D] Advice <strong>and</strong> support documents<br />

[01]<br />

Inform about cleaning of <strong>running</strong>-<strong>machine</strong>, area of the motor <strong>and</strong><br />

especially the light-barriers <strong>and</strong> ventilation grid.<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh author: fh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 207 of 213<br />

appendix<br />

Before opening the motor hood always pull the plug from the mains.<br />

Maintenance on medical products <strong>and</strong> electrical devices may only be<br />

performed by authorised <strong>and</strong> trained personnel who are certified by<br />

h/p/cosmos.<br />

accomplished �<br />

Inform about periodically recurring maintenance intervals <strong>and</strong><br />

obligatory safety controls <strong>and</strong> the risks <strong>and</strong> dangers in case of<br />

disregarding these points.<br />

Offer <strong>and</strong> recommend maintenance contract through authorized <strong>and</strong><br />

trained personnel.<br />

accomplished �<br />

H<strong>and</strong> over <strong>and</strong> explain delivery papers (folder).<br />

Fill out registration form <strong>and</strong> send it to h/p/cosmos sales.<br />

Let the customer sign the delivery note <strong>and</strong> instruction protocol <strong>and</strong><br />

send it back to h/p/cosmos sales immediately.<br />

Ask for brochure of customer <strong>and</strong> send it to the sales department at<br />

h/p/cosmos.<br />

accomplished �


[16.E] Confirmation of commissioning <strong>and</strong> introduction / instruction<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh author: fh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 208 of 213<br />

appendix<br />

By signing this protocol the authorised h/p/cosmos technician <strong>and</strong> the h/p/cosmos customer confirm the receipt <strong>and</strong> underst<strong>and</strong>ing of all<br />

warnings, safety precaution information, the performed instruction <strong>and</strong> commissioning according to form cos15228-01. The customer <strong>and</strong><br />

user confirm the receipt of the listed devices including all accessories <strong>and</strong> options according to the h/p/cosmos delivery note. Disregard of<br />

warnings, disregard of intended <strong>and</strong> forbidden use, safety precautions <strong>and</strong> also unauthorized or lack of maintenance <strong>and</strong>/or regular safety<br />

checks may lead to injuries or even death <strong>and</strong>/or can damage the device <strong>and</strong> will result in loss of any liability <strong>and</strong> warranty. Please fill out<br />

the instruction protocol <strong>and</strong> send back to h/p/cosmos via Fax to +49 86 69 86 42 49, via email to sales@h-p-cosmos.com or via post<br />

h/p/cosmos sports & medical gmbh Am Sportplatz 8 DE 83365 Nussdorf-Traunstein / Germany<br />

h/p/cosmos sports & medical gmbh<br />

Am Sportplatz 8<br />

DE 83365 Nussdorf-Traunstein<br />

Germany<br />

h/p/cosmos device, model name device serial number<br />

instructor<br />

instructed persons / customer / user / operator<br />

customer’s (end-user’s) stamp / customer address:<br />

name in clear block letters h/p/cosmos dealer / technician date <strong>and</strong> signature<br />

name in clear block letters position / function / department date <strong>and</strong> user’s signature


[16.F] Service report – cos15531<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh author: fh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 209 of 213<br />

appendix


[16.G] Control protocol – cos11690en-02<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh author: fh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 210 of 213<br />

appendix


[16.H] Wiring diagram load circuit quasar med 5.0 / pulsar 5.0<br />

Color Code not US / not Canadian st<strong>and</strong>ard!<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh author: fh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 211 of 213<br />

appendix


[16.I] Wiring diagram control circuit quasar med 5.0 / pulsar 5.0 – page 1<br />

Color Code not US / not Canadian st<strong>and</strong>ard!<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh author: fh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 212 of 213<br />

appendix


[16.J] Wiring diagram control circuit quasar med 5.0 / pulsar 5.0 – page 2<br />

Color Code not US / not Canadian st<strong>and</strong>ard!<br />

file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_<strong>running</strong>_<strong>machine</strong>.doc<br />

© 2012 h/p/cosmos sports & medical gmbh author: fh email@h-p-cosmos.com created 15.02.2012 printed 15.02.2012 page: 213 of 213<br />

appendix