SurgiVet ® <strong>V9400</strong><br />
<strong>Capnograph</strong> <strong>and</strong> <strong>Agent</strong> <strong>Monitor</strong><br />
The SurgiVet ® <strong>V9400</strong> <strong>Capnograph</strong><br />
<strong>and</strong> <strong>Agent</strong> monitor quickly <strong>and</strong><br />
reliably captures ETCO 2 numeric<br />
value, ETCO 2 waveform, inspired<br />
CO 2 , respiration rate, inspired<br />
<strong>and</strong> expired agent. Its fluorescent<br />
display assures easy viewing.<br />
Breath-by-breath CO 2 detection<br />
allows rapid response by clinicians<br />
to changes in patients. The SurgiVet ®<br />
<strong>V9400</strong> has the ability to monitor<br />
halogenated agents like isoflurane,<br />
sevoflurane or halothane.<br />
For more information, visit<br />
www.surgivet.com.<br />
<strong>VETERINARY</strong>
SurgiVet ® <strong>V9400</strong><br />
<strong>Capnograph</strong> <strong>and</strong> <strong>Agent</strong> <strong>Monitor</strong><br />
• Quiet System minimizes monitor noise<br />
• Visual <strong>and</strong> audible alarm signals<br />
• Easy-to-read vacuum fluorescent display<br />
• Sidestream technology accommodates<br />
intubated <strong>and</strong> non-intubated patients<br />
• Ability to transfer real time data via an<br />
optional serial cable<br />
• User selectable agent option including<br />
isoflurane, sevoflurane or halothane<br />
Technical Specifications<br />
DISPLAY & INDICATORS<br />
• Display<br />
Vacuum Fluorescent<br />
• Numeric<br />
ETCO 2 , inCO 2 , respiration,<br />
inspired/expired agents<br />
• LED's<br />
Alarms silenced (yellow)<br />
External power (green)<br />
High priority alarms (red)<br />
Medium/low priority alerts (yellow)<br />
• Waveform<br />
CO 2 (1, 2, 3, 4, 8 or 16 mm/sec.)<br />
• Keys<br />
On/Off<br />
Wave/Trend<br />
Alarm Silence<br />
Menu/Enter<br />
Up Arrow/Down Arrow<br />
SIDESTREAM CAPNOGRAPHY<br />
• Measurement<br />
Non-dispersive IR absorption<br />
• Measurement Range<br />
0 to 10% CO 2 , STPD (st<strong>and</strong>ard<br />
temperature <strong>and</strong> pressure dry)<br />
• Units<br />
mmHg, volume %, kPa (user selectable)<br />
• Display Range<br />
0 to 100 mmHg, 0 to 13.3 kPa, 0 to 10.0%<br />
• Accuracy<br />
±2 mmHg, ±0.3% kPa or 4% of reading,<br />
whichever is greater (0 to 36 BrPM)<br />
• Averaging<br />
4 breath average<br />
• N 2 O Compensation<br />
50% or Off (default: Off)<br />
• System Response Time<br />
360 ms to 90% of value (st<strong>and</strong>ard)<br />
• Flow Rate<br />
150 ± 20 ml/min<br />
• Calibration<br />
Manual 2 point<br />
RESPIRATION<br />
• Range<br />
0 to 150 breaths/min<br />
• Accuracy<br />
±1 BrPM (0 to 120 breaths/min)<br />
• Averaging<br />
4 breath average<br />
ANESTHETIC AGENTS<br />
• Technique<br />
Piezoelectric Crystal Resonance<br />
• Range<br />
0 to 8% Halothane, Isoflurane,<br />
or Sevoflurane<br />
• Accuracy<br />
Greater of ± 10% of reading<br />
or ± 0.2% from 0-6%<br />
Accuracy outside of range<br />
is unspecified<br />
• Calibration<br />
High/low calibration required<br />
every 100 hours of operation<br />
SERIAL OUTPUT<br />
• Data log (real-time) at selectable intervals:<br />
10-60 sec<br />
POWER REQUIREMENTS<br />
• AC Power<br />
105-125 V, 50/60 Hz Input, 24 VDC,<br />
1.3 A Output, with 4 kV isolation<br />
• Battery<br />
Not available<br />
DIMENSIONS<br />
3.5" H x 10.0" W x 5.5" D<br />
(8.9 cm H x 25.4 cm W x 14 cm D)<br />
WEIGHT<br />
2.65 lbs. (1.20 kg)<br />
ENVIRONMENTAL<br />
• Operating Temperature<br />
32°F to 122°F (0°C to 50°C)<br />
• Anesthetic <strong>Agent</strong>s<br />
Operating Temperature<br />
59°F to 86°F (15°C to 30°C)<br />
• Storage Temperature<br />
-40°F to 158°F (-40°C to 70°C)<br />
• Relative Humidity (non-condensing)<br />
15-95% (operating)<br />
10-95% (storage)<br />
CATALOG NUMBER<br />
• <strong>V9400</strong><br />
SPECIFICATIONS SUBJECT TO CHANGE WITHOUT NOTICE.<br />
Please read the Instructions for Use supplied with the product<br />
for detailed instructions, warnings <strong>and</strong> cautions.<br />
Smiths Medical PM, Inc.<br />
Veterinary<br />
N7W22025 Johnson Drive, Waukesha, Wisconsin 53186 USA<br />
Phone: 262-513-8500 Fax: 262-542-0718 Toll-Free USA: 888-745-6562<br />
www.surgivet.com<br />
Smiths Medical, part of the global technology business Smiths Group<br />
Product(s) described may not be licensed or available for sale outside of the United States.<br />
SurgiVet <strong>and</strong> the Smiths Medical design mark are trademarks of the Smiths Medical family of companies.<br />
The symbol ® indicates the trademark is registered in the U.S. Patent <strong>and</strong> Trademark Office <strong>and</strong> certain other countries.<br />
All other names <strong>and</strong> marks mentioned are the trade names, trademarks or service marks of their respective owners.<br />
© 2009 Smiths Medical family of companies. All rights reserved. <strong>V9400</strong> Rev.00 08/09