Instrument Range Concentration Calc<br />

• FP628 range:<br />

– 500 mg sample mass<br />

• <strong>Nitrogen</strong> (80ppm to 10%)<br />

– Low end = (0.04 mg N / 500 mg sample) * 100 = 0.08% or 80 ppm N<br />

– High end = (50 mg N / 500 mg sample) *100 = 10% N