DrägerSensor ® <strong>CO</strong> – <strong>68</strong> <strong>09</strong> <strong>605</strong><br />
Data Sheet<br />
This data sheet is a sensor specific supplement to the "Instructions for Use" of the respective transmitter. Please consult in any event the<br />
"Instructions for Use" of the transmitter.<br />
Intended Use<br />
The DrägerSensor ® <strong>CO</strong> is an electrochemical three electrode sensor for the continuous real-time monitoring of carbon monoxide gas (<strong>CO</strong>) in<br />
ambient air. The sensor is designed to be used only in conjunction with a Dräger transmitter (e. g. Polytron 2).<br />
Technical Data<br />
Electrode reactions Measuring electrode: <strong>CO</strong> + H 2 O ––> <strong>CO</strong> 2 + 2 H + + 2 e -<br />
Counter electrode:<br />
O 2 + 4 H + + 4 e - ––> 2 H 2 O<br />
Measuring gas<br />
Measuring range:<br />
minimum<br />
default<br />
maximum<br />
Calibration interval: *<br />
minimum<br />
default<br />
maximum<br />
Ambient conditions:<br />
Operation<br />
Storage (in original container)<br />
Expected sensor life<br />
Lower detection limit **<br />
Linearity tolerance<br />
Repeatability:<br />
zero<br />
sensitivity<br />
Effect of temperature:<br />
zero –40 °C to 40 °C (–40 °F to 105 °F)<br />
40 °C to 65 °C (105 °F to 150 °F)<br />
sensitivity –40 °C to 0 °C (–40 °F to 30 °F)<br />
0 °C to 65 °C (30 °F to 150 °F)<br />
Effect of pressure:<br />
zero<br />
sensitivity<br />
Effect of humidity:<br />
zero<br />
sensitivity<br />
Effect of flow between 0 and 6 m/s (0 and 1180 ft/min):<br />
zero<br />
sensitivity<br />
Long-term drift:<br />
zero<br />
loss of sensitivity<br />
Response time t 0...90 with dust filter **<br />
Warm-up time:<br />
Sensor ready for operation after ***<br />
Sensor ready for calibration after ****<br />
<strong>CO</strong><br />
0 to 50 ppm<br />
0 to 300 ppm<br />
0 to 1000 ppm<br />
1 day<br />
6 months<br />
12 months<br />
–40 °C to 65 °C (–40 °F to 150 °F)<br />
700 hPa to 1300 hPa (20.7" Hg to 38.4" Hg at 32 °F)<br />
5 % r.h. to 95 % r.h., non condensating<br />
0 °C to 40 °C (30 °F to 105 °F)<br />
> 36 months<br />
5 ppm<br />
≤ ± 3 % of measured value<br />
≤ ± 2 ppm<br />
≤ ± 1 % of measured value<br />
≤ ± 0.1 ppm / K<br />
≤ ± 0.5 ppm / K<br />
≤ ± 1 % of measured value / K<br />
≤ ± 0.4 % of measured value / K<br />
no effect<br />
≤ ± 0.05 % of measured value / hPa<br />
no effect<br />
≤ ± 0.02 % of measured value / % r.h.<br />
no effect<br />
≤ ± 3 % of measured value<br />
≤ ± 1 ppm / month<br />
≤ 3 % / year<br />
≤ 30 seconds<br />
≤ 30 minutes<br />
≤ 10 hours<br />
Cross Sensitivities<br />
The table on the next page shows the response of the sensor to other gases than target gas (cross sensitivities). The values are typical and apply<br />
to new sensors. The table does not claim to be complete. Gas mixtures can be displayed as the sum of all components. Gases with negative cross<br />
sensitivities may displace a positive reading of the sensor.<br />
The selective filter A2F is an accessory (see Order List). The sensor has to be recalibrated subsequently when a selective filter is used.<br />
® DrägerSensor is a trademark of Dräger, registered in Germany.<br />
* Only applicable when used in conjunction with Polytron 2 or Polytron 2 XP.<br />
** At 25 °C (77 °F), 1013 hPa (29.2" Hg at 32 °F), 50 % r.h.<br />
*** At this time, the actual zero point error of the sensor is less than 4 times the value stated in this data sheet.<br />
**** At this time, the sensor has reached its specific accuracy.
Gas / Vapor Chemical Symbol Gas concentration Displayed value in ppm <strong>CO</strong><br />
with dust filter with selective filter A2F<br />
Acetaldehyde CH 3 CHO 50 ppm ≤ 45 no effect<br />
Acetone CH 3 <strong>CO</strong>CH 3 1 000 ppm ≤ 20 no effect<br />
Ammonia NH 3 60 ppm ≤ 1 no effect<br />
Arsine AsH 3 5 ppm ≤ 20 no effect<br />
Boron trifluoride BF 3 15 ppm ≤ 1 no effect<br />
Butadiene CH 2 CHCHCH 2 50 ppm ≤ 130 no effect<br />
Carbon dioxide <strong>CO</strong> 2 1.5 % by vol no effect no effect<br />
Chlorine Cl 2 5 ppm ≤ 1 (–) * no effect<br />
Diborane B 2 H 6 5 ppm ≤ 30 ≤ 30<br />
Dichlormethane CH 2 Cl 2 1 000 ppm ≤ 1 no effect<br />
Disilane Si 2 H 6 10 ppm ≤ 50 no effect<br />
Ethane C 2 H 6 2 000 ppm ≤ 1 no effect<br />
Ethanol C 2 H 5 OH 200 ppm ≤ 400 no effect<br />
Ethene C 2 H 4 1 000 ppm ≤ 1300 ≤ 15<br />
Ethine C 2 H 2 200 ppm ≤ 550 ≤ 40<br />
Ethyl acetate CH 3 <strong>CO</strong>OC 2 H 5 1 000 ppm ≤ 150 no effect<br />
Ethylene oxide C 2 H 4 O 20 ppm ≤ 20 ≤ 2<br />
Formaldehyde HCHO 20 ppm ≤ 30 no effect<br />
Hydrogen H 2 1 000 ppm ≤ 90 ≤ 90<br />
Hydrogen chloride HCl 40 ppm ≤ 5 no effect<br />
Hydrogen cyanide HCN 25 ppm ≤ 30 no effect<br />
Hydrogen fluoride HF 15 ppm ≤ 1 no effect<br />
Hydrogen selenide SeH 2 5 ppm ≤ 12 no effect<br />
Hydrogen sulfide H 2 S 20 ppm ≤ 90 no effect<br />
Methane CH 4 2 % by vol ≤ 1 no effect<br />
Methanol CH 3 OH 200 ppm ≤ 300 no effect<br />
Methylmethacrylate CH 2 C(CH 3 )<strong>CO</strong>OCH 3 60 ppm ≤ 60 no effect<br />
Nitrogen dioxide NO 2 20 ppm ≤ 3 no effect<br />
Nitrogen monoxide NO 20 ppm ≤ 30 no effect<br />
Phosgene <strong>CO</strong>Cl 2 5 ppm ≤ 1 no effect<br />
Phosphine PH 3 10 ppm ≤ 50 no effect<br />
Propane C 3 H 8 1 % by vol ≤ 1 no effect<br />
i-Propanol (CH 3 ) 2 CHOH 500 ppm ≤ 300 no effect<br />
Silane SiH 4 5 ppm ≤ 20 no effect<br />
Styrene C 6 H 5 CHCH 2 30 ppm ≤ 50 no effect<br />
Sulfur dioxide SO 2 20 ppm ≤ 16 no effect<br />
Tetrahydrofurane C 4 H 8 O 60 ppm ≤ 90 no effect<br />
Tetrahydrothiophene C 4 H 8 S 10 ppm ≤ 10 no effect<br />
Toluene C 6 H 5 CH 3 1 000 ppm ≤ 1 no effect<br />
Vinyl chloride C 2 H 3 Cl 50 ppm ≤ 50 no effect<br />
(–) * negative display.<br />
Order List<br />
DrägerSensor ® <strong>CO</strong> ............................................................. <strong>68</strong> <strong>09</strong> <strong>605</strong><br />
Dustfilter ............................................................................. <strong>68</strong> <strong>09</strong> 595<br />
Selective filter A2F .............................................................. <strong>68</strong> <strong>09</strong> <strong>68</strong>4<br />
For calibration:<br />
Calibration adapter .............................................................. <strong>68</strong> 06 978<br />
Remote calibration adapter .................................................. <strong>68</strong> 07 955<br />
Pressure reducer ................................................................ on request<br />
Test gas cylinder 99.9 % N 2 (zero gas), 4 L, 200 bar .......... <strong>68</strong> 05 113<br />
Calibration bottle for calibration with ampoules .................... <strong>68</strong> 03 407<br />
Test gas ampoule 100 ppm <strong>CO</strong> .......................................... <strong>68</strong> 07 920<br />
Test gas ampoule 300 ppm <strong>CO</strong> .......................................... <strong>68</strong> 07 921<br />
67 mm (2.64")<br />
Ø 27.6 mm (1.08")<br />
Ø 50 mm (1.97")<br />
00123456<br />
Dräger Safety AG & Co. KGaA – < Revalstraße 1, D-23560 Lübeck, Germany – y +49 451 8 82 - 27 94 – FAX +49 451 8 82 - 49 91<br />
! http://www.draeger.com/gds<br />
5th edition - January 2002<br />
Subject to alteration<br />
90 23 454 - TD 4<strong>68</strong>1.310 en/de<br />