DrägerSensor ® <strong>CO</strong> – <strong>68</strong> <strong>09</strong> <strong>605</strong><br />

Data Sheet<br />

This data sheet is a sensor specific supplement to the "Instructions for Use" of the respective transmitter. Please consult in any event the<br />

"Instructions for Use" of the transmitter.<br />

Intended Use<br />

The DrägerSensor ® <strong>CO</strong> is an electrochemical three electrode sensor for the continuous real-time monitoring of carbon monoxide gas (<strong>CO</strong>) in<br />

ambient air. The sensor is designed to be used only in conjunction with a Dräger transmitter (e. g. Polytron 2).<br />

Technical Data<br />

Electrode reactions Measuring electrode: <strong>CO</strong> + H 2 O ––> <strong>CO</strong> 2 + 2 H + + 2 e -<br />

Counter electrode:<br />

O 2 + 4 H + + 4 e - ––> 2 H 2 O<br />

Measuring gas<br />

Measuring range:<br />

minimum<br />

default<br />

maximum<br />

Calibration interval: *<br />

minimum<br />

default<br />

maximum<br />

Ambient conditions:<br />

Operation<br />

Storage (in original container)<br />

Expected sensor life<br />

Lower detection limit **<br />

Linearity tolerance<br />

Repeatability:<br />

zero<br />

sensitivity<br />

Effect of temperature:<br />

zero –40 °C to 40 °C (–40 °F to 105 °F)<br />

40 °C to 65 °C (105 °F to 150 °F)<br />

sensitivity –40 °C to 0 °C (–40 °F to 30 °F)<br />

0 °C to 65 °C (30 °F to 150 °F)<br />

Effect of pressure:<br />

zero<br />

sensitivity<br />

Effect of humidity:<br />

zero<br />

sensitivity<br />

Effect of flow between 0 and 6 m/s (0 and 1180 ft/min):<br />

zero<br />

sensitivity<br />

Long-term drift:<br />

zero<br />

loss of sensitivity<br />

Response time t 0...90 with dust filter **<br />

Warm-up time:<br />

Sensor ready for operation after ***<br />

Sensor ready for calibration after ****<br />

<strong>CO</strong><br />

0 to 50 ppm<br />

0 to 300 ppm<br />

0 to 1000 ppm<br />

1 day<br />

6 months<br />

12 months<br />

–40 °C to 65 °C (–40 °F to 150 °F)<br />

700 hPa to 1300 hPa (20.7" Hg to 38.4" Hg at 32 °F)<br />

5 % r.h. to 95 % r.h., non condensating<br />

0 °C to 40 °C (30 °F to 105 °F)<br />

> 36 months<br />

5 ppm<br />

≤ ± 3 % of measured value<br />

≤ ± 2 ppm<br />

≤ ± 1 % of measured value<br />

≤ ± 0.1 ppm / K<br />

≤ ± 0.5 ppm / K<br />

≤ ± 1 % of measured value / K<br />

≤ ± 0.4 % of measured value / K<br />

no effect<br />

≤ ± 0.05 % of measured value / hPa<br />

no effect<br />

≤ ± 0.02 % of measured value / % r.h.<br />

no effect<br />

≤ ± 3 % of measured value<br />

≤ ± 1 ppm / month<br />

≤ 3 % / year<br />

≤ 30 seconds<br />

≤ 30 minutes<br />

≤ 10 hours<br />

Cross Sensitivities<br />

The table on the next page shows the response of the sensor to other gases than target gas (cross sensitivities). The values are typical and apply<br />

to new sensors. The table does not claim to be complete. Gas mixtures can be displayed as the sum of all components. Gases with negative cross<br />

sensitivities may displace a positive reading of the sensor.<br />

The selective filter A2F is an accessory (see Order List). The sensor has to be recalibrated subsequently when a selective filter is used.<br />

® DrägerSensor is a trademark of Dräger, registered in Germany.<br />

* Only applicable when used in conjunction with Polytron 2 or Polytron 2 XP.<br />

** At 25 °C (77 °F), 1013 hPa (29.2" Hg at 32 °F), 50 % r.h.<br />

*** At this time, the actual zero point error of the sensor is less than 4 times the value stated in this data sheet.<br />

**** At this time, the sensor has reached its specific accuracy.


Gas / Vapor Chemical Symbol Gas concentration Displayed value in ppm <strong>CO</strong><br />

with dust filter with selective filter A2F<br />

Acetaldehyde CH 3 CHO 50 ppm ≤ 45 no effect<br />

Acetone CH 3 <strong>CO</strong>CH 3 1 000 ppm ≤ 20 no effect<br />

Ammonia NH 3 60 ppm ≤ 1 no effect<br />

Arsine AsH 3 5 ppm ≤ 20 no effect<br />

Boron trifluoride BF 3 15 ppm ≤ 1 no effect<br />

Butadiene CH 2 CHCHCH 2 50 ppm ≤ 130 no effect<br />

Carbon dioxide <strong>CO</strong> 2 1.5 % by vol no effect no effect<br />

Chlorine Cl 2 5 ppm ≤ 1 (–) * no effect<br />

Diborane B 2 H 6 5 ppm ≤ 30 ≤ 30<br />

Dichlormethane CH 2 Cl 2 1 000 ppm ≤ 1 no effect<br />

Disilane Si 2 H 6 10 ppm ≤ 50 no effect<br />

Ethane C 2 H 6 2 000 ppm ≤ 1 no effect<br />

Ethanol C 2 H 5 OH 200 ppm ≤ 400 no effect<br />

Ethene C 2 H 4 1 000 ppm ≤ 1300 ≤ 15<br />

Ethine C 2 H 2 200 ppm ≤ 550 ≤ 40<br />

Ethyl acetate CH 3 <strong>CO</strong>OC 2 H 5 1 000 ppm ≤ 150 no effect<br />

Ethylene oxide C 2 H 4 O 20 ppm ≤ 20 ≤ 2<br />

Formaldehyde HCHO 20 ppm ≤ 30 no effect<br />

Hydrogen H 2 1 000 ppm ≤ 90 ≤ 90<br />

Hydrogen chloride HCl 40 ppm ≤ 5 no effect<br />

Hydrogen cyanide HCN 25 ppm ≤ 30 no effect<br />

Hydrogen fluoride HF 15 ppm ≤ 1 no effect<br />

Hydrogen selenide SeH 2 5 ppm ≤ 12 no effect<br />

Hydrogen sulfide H 2 S 20 ppm ≤ 90 no effect<br />

Methane CH 4 2 % by vol ≤ 1 no effect<br />

Methanol CH 3 OH 200 ppm ≤ 300 no effect<br />

Methylmethacrylate CH 2 C(CH 3 )<strong>CO</strong>OCH 3 60 ppm ≤ 60 no effect<br />

Nitrogen dioxide NO 2 20 ppm ≤ 3 no effect<br />

Nitrogen monoxide NO 20 ppm ≤ 30 no effect<br />

Phosgene <strong>CO</strong>Cl 2 5 ppm ≤ 1 no effect<br />

Phosphine PH 3 10 ppm ≤ 50 no effect<br />

Propane C 3 H 8 1 % by vol ≤ 1 no effect<br />

i-Propanol (CH 3 ) 2 CHOH 500 ppm ≤ 300 no effect<br />

Silane SiH 4 5 ppm ≤ 20 no effect<br />

Styrene C 6 H 5 CHCH 2 30 ppm ≤ 50 no effect<br />

Sulfur dioxide SO 2 20 ppm ≤ 16 no effect<br />

Tetrahydrofurane C 4 H 8 O 60 ppm ≤ 90 no effect<br />

Tetrahydrothiophene C 4 H 8 S 10 ppm ≤ 10 no effect<br />

Toluene C 6 H 5 CH 3 1 000 ppm ≤ 1 no effect<br />

Vinyl chloride C 2 H 3 Cl 50 ppm ≤ 50 no effect<br />

(–) * negative display.<br />

Order List<br />

DrägerSensor ® <strong>CO</strong> ............................................................. <strong>68</strong> <strong>09</strong> <strong>605</strong><br />

Dustfilter ............................................................................. <strong>68</strong> <strong>09</strong> 595<br />

Selective filter A2F .............................................................. <strong>68</strong> <strong>09</strong> <strong>68</strong>4<br />

For calibration:<br />

Calibration adapter .............................................................. <strong>68</strong> 06 978<br />

Remote calibration adapter .................................................. <strong>68</strong> 07 955<br />

Pressure reducer ................................................................ on request<br />

Test gas cylinder 99.9 % N 2 (zero gas), 4 L, 200 bar .......... <strong>68</strong> 05 113<br />

Calibration bottle for calibration with ampoules .................... <strong>68</strong> 03 407<br />

Test gas ampoule 100 ppm <strong>CO</strong> .......................................... <strong>68</strong> 07 920<br />

Test gas ampoule 300 ppm <strong>CO</strong> .......................................... <strong>68</strong> 07 921<br />

67 mm (2.64")<br />

Ø 27.6 mm (1.08")<br />

Ø 50 mm (1.97")<br />

00123456<br />

Dräger Safety AG & Co. KGaA – < Revalstraße 1, D-23560 Lübeck, Germany – y +49 451 8 82 - 27 94 – FAX +49 451 8 82 - 49 91<br />

! http://www.draeger.com/gds<br />

5th edition - January 2002<br />

Subject to alteration<br />

90 23 454 - TD 4<strong>68</strong>1.310 en/de<br />

© Dräger Safety AG & Co. KGaA