DrägerSensor ® <strong>COCl</strong> 2 – <strong>68</strong> <strong>09</strong> <strong>930</strong><br />

Data Sheet<br />

This data sheet is a sensor specific supplement to the "Instructions for Use" of the respective transmitter. Please consult in any event the<br />

"Instructions for Use" of the transmitter.<br />

Intended Use<br />

The DrägerSensor ® <strong>COCl</strong> 2 is an electrochemical three electrode sensor for the continuous real-time monitoring of phosgene gas (<strong>COCl</strong> 2 ) in<br />

ambient air. The sensor is designed to be used only in conjunction with a Dräger transmitter (e. g. Polytron 2).<br />

Technical Data<br />

Electrode reactions Measuring electrode: <strong>COCl</strong> 2 + H 2 O + 2 Me ––> CO 2 + 2 H + + 2 MeCl + 2 e -<br />

Counter electrode:<br />

O 2 + 4 H + + 4 e - ––> 2 H 2 O<br />

Measuring gas <strong>COCl</strong> 2<br />

Measuring range:<br />

minimum<br />

default<br />

maximum<br />

Calibration interval: *<br />

minimum<br />

default<br />

maximum<br />

Ambient conditions:<br />

Operation<br />

Storage<br />

Expected sensor life<br />

Lower detection limit **<br />

Linearity tolerance<br />

Repeatability:<br />

zero<br />

sensitivity<br />

Effect of temperature:<br />

zero<br />

sensitivity<br />

Effect of pressure:<br />

zero<br />

sensitivity<br />

Effect of humidity<br />

zero<br />

sensitivity<br />

Effect of flow between 0 and 6 m/s (0 and 1180 ft/min):<br />

zero<br />

sensitivity<br />

Long-term drift:<br />

zero<br />

sensitivity<br />

Response time t 0...50 **<br />

Warm-up time:<br />

Sensor ready for operation after ***<br />

Sensor ready for calibration after ****<br />

0.5 ppm<br />

1 ppm<br />

1 ppm<br />

1 day<br />

3 months<br />

6 months<br />

–40 °C to 65 °C (–40 °F to 150 °F)<br />

700 hPa to 1300 hPa (20.7" Hg to 38.4" Hg at 32 °F)<br />

10 % r.h. to 95 % r.h., non condensating<br />

0 °C to 25 °C (30 °F to 80°F)<br />

700 hPa to 1300 hPa (20.7" Hg to 38.4" Hg at 32 °F)<br />

30 % r.h. to 70 % r.h., non condensating<br />

> 9 months, max. gas exposure 150 ppm x hours<br />

0.05 ppm<br />

≤ ± 20 % of full scale value<br />

≤ ± 0.02 ppm<br />

≤ ± 10 % of measured value<br />

≤ ± 0.002 ppm / K<br />

≤ ± 0.5 % of measured value / K<br />

no effect<br />

≤ ± 0.1 % of measured value / hPa<br />

≤ ± 0.0005 ppm/% r.h.<br />

≤ ± 0.3 % of measured value/% r.h.<br />

no effect<br />

≤ ± 20 % of measured value<br />

≤ ± 0.01 ppm / month<br />

≤ ± 5 % of measured value / month<br />

≤ 30 seconds<br />

≤ 15 minutes<br />

≤ 1 hour<br />

® DrägerSensor is a trademark of Dräger, registered in Germany.<br />

* Only applicable when used in conjunction with Polytron 2 or Polytron 2 XP.<br />

** At 25 °C (77 °F), 1013 hPa (29.2" Hg at 32 °F), 50 % r.h.<br />

*** At this time, the actual zero point error of the sensor is less than 4 times the value stated in this data sheet.<br />

**** At this time, the sensor has reached its specific accuracy.


Cross Sensitivities<br />

The table shows the response of the sensor to other gases than target gas (cross sensitivities). The values are typical and apply to new sensors.<br />

The table does not claim to be complete. Gas mixtures can be displayed as the sum of all components. Gases with negative cross sensitivity may<br />

displace a positive reading of the sensor.<br />

Gas / Vapor Chemical Symbol Gas concentration Displayed value in ppm <strong>COCl</strong> 2<br />

without selective filter K2F with selective filter K2F<br />

Carbon dioxide CO 2 5 000 ppm ≤ 0.01 no effect<br />

Carbon monoxide CO 1 000 ppm ≤ 0.01 no effect<br />

Chlorine Cl 2 0.1 ppm ≤ 0.05 ≤ 0.05<br />

Chloro-formic acid methl ester CH 3 O<strong>COCl</strong> 1 ppm ≤ 0.01 no effect<br />

Ethanol C 2 H 5 OH 5.8 % by vol ≤ 0.01 no effect<br />

Hydrogen H 2 1 000 ppm ≤ 0.01 no effect<br />

Hydrogen chloride HCl 1 ppm ≤ 0.5 no effect<br />

Hydrogen cyanide HCN 1 ppm ≤ 0.2 ≤ 0.2<br />

Hydrogen peroxide H 2 O 2 1 ppm ≤ 0.01 no effect<br />

Hydrogen sulfide H 2 S 1 ppm ≤ 1.0 no effect<br />

Hydrocarbons – – – % by vol. range ≤ 0.01 no effect<br />

Nitrogen dioxide NO 2 1 ppm ≤ 0.02 (–) * ≤ 0.02 (–) *<br />

Ozone O 3 1 ppm ≤ 0.01 (–) * no effect<br />

Phosphine PH 3 0.5 ppm ≤ 0.1 ≤ 0.1<br />

i-Propanol (CH 3 ) 2 CHOH 4.5 % by vol. ≤ 0.01 no effect<br />

Sulfur dioxide SO 2 1 ppm ≤ 0.01 no effect<br />

(–) * negative display.<br />

Order List<br />

DrägerSensor ® <strong>COCl</strong> 2 ....................................................... <strong>68</strong> <strong>09</strong> <strong>930</strong><br />

Selective filter K2F .............................................................. <strong>68</strong> <strong>09</strong> 933<br />

For calibration:<br />

Calibration adapter .............................................................. <strong>68</strong> 06 978<br />

Pressure reducer ................................................................ on request<br />

Test gas cylinder 99.9 % N 2 (zero gas), 4 L, 200 bar .......... <strong>68</strong> 05 113<br />

67 mm (2.64")<br />

00123456<br />

Ø 27.6 mm (1.08")<br />

Ø 50 mm (1.97")<br />

Dräger Safety AG & Co. KGaA – < Revalstraße 1, D-23560 Lübeck, Germany – y +49 451 8 82 - 27 94 – FAX +49 451 8 82 - 49 91<br />

! http://www.draeger.com/gds<br />

9th edition - September 2001<br />

Subject to alteration<br />

90 23 515 - TD 4<strong>68</strong>1.460 en/de<br />

© Dräger Safety AG & Co. KGaA