28.05.2015 Views

Planning Documents for OR

This Planning Document was developed to enhance your knowledge of the Dräger portfolio and to provide guidance when designing a new OR. This document contains an overview of Dräger products and their corresponding technical data. The last pages of this document show several exemplary Dräger solutions for different disciplines likes heart surgery or neurosurgery, including detailed technical drawings of the individual workplaces.

This Planning Document was developed to enhance your knowledge of the
Dräger portfolio and to provide guidance when designing a new OR. This
document contains an overview of Dräger products and their corresponding
technical data.
The last pages of this document show several exemplary Dräger solutions for
different disciplines likes heart surgery or neurosurgery, including detailed
technical drawings of the individual workplaces.

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

D-76332-2013<br />

ST-8934-2008<br />

D-85565-2013<br />

<strong>Planning</strong> <strong>Documents</strong><br />

<strong>for</strong> <strong>OR</strong>


PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> | PREAMBLE | 3<br />

Preamble<br />

This <strong>Planning</strong> Document was developed to enhance your knowledge of the<br />

Dräger portfolio and to provide guidance when designing a new <strong>OR</strong>. This<br />

document contains an overview of Dräger products and their corresponding<br />

technical data.<br />

The last pages of this document show several exemplary Dräger solutions <strong>for</strong><br />

different disciplines likes heart surgery or neurosurgery, including detailed<br />

technical drawings of the individual workplaces.


4 |<br />

PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> |<br />

INDEX OF CONTENTS<br />

INDEX OF CONTENTS<br />

OVERVIEW<br />

<strong>OR</strong> – The concept of <strong>OR</strong> planning 6<br />

Dräger – Workplace infrastructure 7<br />

DRÄGER P<strong>OR</strong>TFOLIO 8<br />

Overview Supply units 10<br />

<strong>OR</strong> lights and video systems/Gas Management Systems 11<br />

Ceiling supply units 12<br />

<strong>OR</strong> and examination lights 14<br />

Central axis system dedicated <strong>for</strong> Hybrid <strong>OR</strong> 16<br />

Display holder 18<br />

Ceiling fixtures 20<br />

Opera 23<br />

Light options 26<br />

Workstation components 27<br />

Gas Management Systems 30<br />

Overview anesthesia portfolio 32<br />

Accessories 34<br />

W<strong>OR</strong>KPLACE EXAMPLES 36<br />

Supply units Anesthesia/Docking 38<br />

Surgical workplace 39<br />

Display holder/<strong>OR</strong> lights 40<br />

Anesthesia devices 41<br />

GENERAL PLANNING 42<br />

General <strong>Planning</strong>: Ceiling supply units, <strong>OR</strong> lights/Cameras/Displays, 44<br />

Ceiling mount<br />

<strong>Planning</strong> an anesthesia workplace 45<br />

Special <strong>OR</strong> types 46


PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> | INDEX OF CONTENTS | 5<br />

INDEX OF CONTENTS<br />

GENERAL PLANNING<br />

Example layout <strong>for</strong> general surgery 48<br />

Example layout <strong>for</strong> vascular surgery 50<br />

Example layout <strong>for</strong> heart surgery 52<br />

Example layout <strong>for</strong> laparoscopic surgery 54<br />

Example <strong>for</strong> neurosurgery 56<br />

Example <strong>for</strong> orthopaedic surgery 58<br />

CAD Drawing – Configuration 1, Movita AK (L750-LiftExp.) 60<br />

CAD Drawing – Configuration 2, Agila 05 (L1000-LiftExp.) 61<br />

CAD Drawing – Configuration 3, Polaris® 700 - Polaris® 500 62<br />

CAD Drawing – Configuration 4, Agila 05 (L1000-Lift) - 63<br />

AC3000 2 Displays<br />

CAD Drawing – Configuration 5, Movita AK (L750-Lift) 64<br />

CAD Drawing – Configuration 6, Polaris® 700 - Polaris® 500 65<br />

CAD Drawing – Configuration 7, Movita 13MD (M500-Lift) - 66<br />

AC2000 1 Display<br />

CAD Drawing – Configuration 8, Movita 13 (M750-750) - 67<br />

AC2000 1 Display<br />

CAD Drawing – Configuration 9, Polaris® 700 - Polaris® 500 - Polaris® 500 68<br />

CAD Drawing – Configuration 10, Movita 13MD (M1000-Lift) - 69<br />

AC2000 1 Display<br />

CAD Drawing – Configuration 11, Movita 15 (M750-750) - 70<br />

AC2000 1 Display<br />

CAD Drawing – Configuration 12, Polaris® 500 - Polaris® 500 71<br />

CAD Drawing – Configuration 13, Movita 13 (M1250 - Lift) 72<br />

CAD Drawing – Configuration 14, Movita 13 (S1250 - 1000) 73<br />

CAD Drawing – Configuration 15, AC3000 Display 74<br />

CAD Drawing – Configuration 16, Polaris® 500 - Polaris® 700 75<br />

CAD Drawing – Configuration 17, Opera 4S 76<br />

CAD Drawing – Configuration 18, Polaris® 700 - Polaris® 500 77<br />

CONTACT 79


6 |<br />

PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> |<br />

OVERVIEW<br />

<strong>OR</strong><br />

The concept of <strong>OR</strong> planning:<br />

“Design from the inside out”<br />

Equipment<br />

Staff<br />

Intervention<br />

Patient<br />

NEW TECHNOLOGIES, NEW CONCEPTS<br />

Operating room design is changing in order to<br />

meet the requirements dictated by new procedures,<br />

technologies and methods. This trend<br />

will very likely continue <strong>for</strong> the <strong>for</strong>eseeable<br />

future. When designing a new operating room,<br />

it’s imperative to understand both the present<br />

and possible future needs of the client.<br />

FUTURE-PROOF DESIGN CONCEPTS<br />

Dräger offers a wide range of products<br />

designed with maximum flexibility in mind.<br />

This is true both in terms of daily clinical<br />

routine as well as long range planning. They<br />

ensure smooth workflow, superior ergonomics<br />

and offer custom solutions <strong>for</strong> just about<br />

any workplace design challenge.


PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> | OVERVIEW | 7<br />

Dräger<br />

Workplace infrastructure<br />

Dräger offers solutions <strong>for</strong> the entire hospital<br />

D-30018-2011


8 |<br />

PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> |<br />

DRÄGER P<strong>OR</strong>TFOLIO<br />

Dräger<br />

portfolio<br />

D-85458-2013


PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> | DRÄGER P<strong>OR</strong>TFOLIO | 9<br />

Dräger portfolio<br />

Dräger portfolio<br />

An overview of the Dräger portfolio <strong>for</strong> the <strong>OR</strong><br />

<strong>Planning</strong> an <strong>OR</strong> requires a thorough understanding<br />

of the Dräger product portfolio. This chapter will<br />

help you familiarize yourself with Dräger products<br />

and devices which are relevant <strong>for</strong> the operating<br />

room, such as supply units, <strong>OR</strong> lights, ceiling<br />

fixtures, gas management systems and anesthesia<br />

devices. We encourage you to take a close look<br />

at the Dräger products discussed in this chapter<br />

in order to maximize your potential planning<br />

competency.


10 |<br />

PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> |<br />

DRÄGER P<strong>OR</strong>TFOLIO<br />

Overview<br />

Supply units<br />

D-3289-2011<br />

Ponta<br />

D-148-2010<br />

MT-5960-2008<br />

MT-6342-2008<br />

Opera®<br />

Agila<br />

Movita<br />

D-18477-2010<br />

D-36069-2011<br />

D-27741-2009<br />

Forta<br />

Linea<br />

Gemina® Duo


PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> | DRÄGER P<strong>OR</strong>TFOLIO | 11<br />

<strong>OR</strong> lights and video systems<br />

Dräger portfolio<br />

MT-6323-2008<br />

D-58476-2012<br />

D-31307-2011<br />

D-21704-2010<br />

Video systems<br />

Polaris® 50<br />

Polaris® 100/200<br />

Polaris® 5xx/7xx<br />

Gas Management Systems<br />

D-9489-2009<br />

D-9270-2009<br />

Terminal Units<br />

EcoPharm Tower<br />

D-9368-2009<br />

MT-0001-2006<br />

D-58913-2012<br />

Medical Air Guard<br />

Gas Control Station<br />

Area Control Unit with VoluCount®


12 |<br />

PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> |<br />

DRÄGER P<strong>OR</strong>TFOLIO<br />

Ceiling supply units<br />

D-52660-2012<br />

MT-6342-2008<br />

D-18460-2010<br />

Agila<br />

Movita<br />

Forta<br />

WHAT IS WHAT?<br />

Ceiling bearing<br />

Arm system<br />

Distance tube<br />

Lifting<br />

system<br />

Media head<br />

Media column<br />

D-18369-2010<br />

D-12307-2011<br />

Workstation<br />

components<br />

D-18551-2010


PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> | DRÄGER P<strong>OR</strong>TFOLIO | 13<br />

ARM SYSTEMS<br />

D-52660-2012<br />

D-52660-2012<br />

Single arm system<br />

Arm length: 500 mm, 750 mm, 1000 mm, 1250 mm<br />

Load capacity*: 120 kg - 270 kg<br />

Double arm system<br />

Arm length: min. 1000 mm - max. 2500 mm<br />

Load capacity*: 120 kg - 270 kg<br />

Dräger portfolio<br />

D-18369-2010<br />

Lift arm system<br />

Arm length: min. 1000 mm (lift arm) - max. 2500 mm<br />

Load capacity*: 80 kg - 270 kg<br />

*Load capacity = net load at media column/head<br />

strong<br />

medium<br />

light<br />

Arm types<br />

For single and double arm systems, you can choose between heavy-duty,<br />

medium and lightweight models. Load capacities are defined by the type<br />

of and length of arm.<br />

MT-5963-2008<br />

MT-6342-2008<br />

D-18468-2010<br />

Agila<br />

Movita<br />

Forta<br />

TECHNICAL DATA<br />

Product Agila Movita Forta<br />

Load capacity<br />

Media head 120 kg 120 kg –<br />

Column 120 kg 270 kg 270 kg<br />

Lift express/Lift 80 kg/120 kg 80 kg/180 kg –<br />

Lift speed 10/20 sec 10/20 sec 30 sec<br />

Height adjustment 600 mm 600 mm 600 mm<br />

Media outlets up to 51 up to 102 up to 32


14 |<br />

PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> |<br />

DRÄGER P<strong>OR</strong>TFOLIO<br />

<strong>OR</strong> and examination lights<br />

LIGHTS AND VIDEO OPTIONS<br />

D-21696-2010<br />

Video systems<br />

D-31029-2011<br />

D-58477-2012<br />

D-448-2014<br />

Polaris® 5xx/7xx<br />

Polaris® 100/200<br />

Polaris® 50<br />

TECHNICAL DATA<br />

Product Polaris® 5xx/7xx Polaris® 100/200 Polaris® 50<br />

Technology LED LED LED<br />

Light intensity 110.000 or 160.000 Lux 120.000 or 160.000 Lux 60.000 Lux<br />

Endo-light mode 3.000 Lux 300 Lux no<br />

Light field diameter 220 to 290 mm 200 mm 160 mm<br />

Color temperature 4600 K 4400 K or 5000 K or 5600 K (option) 4500 K<br />

Video options integrated SD or HD camera, external SD camera external SD camera no


PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> | DRÄGER P<strong>OR</strong>TFOLIO | 15<br />

D-21698-2010<br />

Dräger portfolio<br />

CENTRAL AXIS MODELS<br />

Central axis system<br />

– For Polaris, camera and displays,<br />

central axes are available<br />

Swivel arms<br />

– You can choose from 1 up to 4-fold axis.<br />

Standard length: 700, 850, 1000, 1150<br />

and 1450 mm<br />

Arm system<br />

Swivel arm length (mm)<br />

1-fold central axis (ø 110 mm) 850<br />

1150<br />

1450<br />

AC 2000/3000 1-fold 750<br />

2-fold central axis 700/850<br />

700/1150<br />

3-fold central axis 700/850/1000<br />

700/850/1150<br />

4-fold central axis 700/850/1000/1150<br />

All central axes are freely configurable. Display holders, external cameras, X-Ray protection shields or <strong>OR</strong> lights can be mounted on each position.<br />

A display holder must be to be mounted at either the lowest or highest position.


16 |<br />

PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> |<br />

DRÄGER P<strong>OR</strong>TFOLIO<br />

D-90850-2013<br />

D-442-2014<br />

Central axis system dedicated <strong>for</strong> Hybrid <strong>OR</strong><br />

Several axes especially designed <strong>for</strong> use in Hybrid <strong>OR</strong> are available.<br />

You can choose between a 1 and 2-fold axis. Available lengths: 1300 to 2200 mm<br />

These axes are also useful if two mounting points are outside the LAF-field (left and right)


PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> | DRÄGER P<strong>OR</strong>TFOLIO | 17<br />

D-22405-2010<br />

Dräger portfolio<br />

Arm system<br />

Swivel arm length (mm)<br />

1-fold central axis 1300<br />

1400<br />

–<br />

2200<br />

2-fold central axis 1150/1300<br />

1300/1450<br />

–<br />

2050/2200<br />

SPRING ARMS<br />

Light heads and display holders are<br />

mounted on swivel arms using spring arms.<br />

Spring arm length: approx. 910 mm<br />

(<strong>OR</strong> lights)


18 |<br />

PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> |<br />

DRÄGER P<strong>OR</strong>TFOLIO<br />

Display holder<br />

D-57969-2012<br />

D-90843-2013<br />

Pendula<br />

Display holder at CSU<br />

D-90844-2013<br />

D-069-2014<br />

Display holder at central axis (up to 24")<br />

Display holder at central axis (up to 32")<br />

PENDULA – CUSTOM SOLUTIONS F<strong>OR</strong> IMAGING SYSTEMS<br />

FEATURES<br />

– Stationary ceiling mount<br />

– Spring arm adapted to Dräger arm<br />

– Freely suspended system that swings away<br />

(horizontally and vertically) in case of<br />

collisions<br />

– The swinging mount underneath the lift<br />

arm supports the automatic upright positioning<br />

of the system and protect the displays<br />

and other equipment in case of collisions<br />

– Integrated cable management<br />

– Flexible arrangement of the displays in<br />

number and size (2-8 displays (19"/21") or<br />

1x 60" + backup screens)


PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> | DRÄGER P<strong>OR</strong>TFOLIO | 19<br />

D-35818-2011<br />

D-57963-2012<br />

D-57972-2012<br />

Dräger portfolio<br />

Front view 3-fold<br />

Front view 60 inch<br />

Front view 8-fold<br />

D-35825-2011<br />

D-57962-2012<br />

D-57973-2012<br />

Back view 3-fold<br />

Back view 60 inch<br />

Back view 8-fold<br />

VARIANTS<br />

– 4 arm lengths: 1750 mm-2500 mm<br />

– 7 display frames<br />

– 2-8 displays optimized <strong>for</strong> 19";<br />

21" displays can be also mounted<br />

– 60" large display<br />

TECHNICAL DATA<br />

Product Pendula Display holder at CSU Display arm at central axis<br />

Description Multi display holder Swivel and lift arms mounted on CSU Display arm at central axis<br />

Display sizes 19" up to 60" approx. 26" (VESA 100 adaptor), approx. 24" (larger displays<br />

max. 15 kg<br />

on request)<br />

Display quantity 1 (60") up to 8 (21") depending on column size 1 (19 to 32”) or 2 (19 to 24”)<br />

Fixation analogue to CSU fixation at CSU – Media column at central axis of <strong>OR</strong> light<br />

or separate


20 |<br />

PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> |<br />

DRÄGER P<strong>OR</strong>TFOLIO<br />

Ceiling fixtures<br />

Ceiling fixtures <strong>for</strong> supply units<br />

PRODUCT DESCRIPTION<br />

The universal Dräger ceiling mount supports<br />

the installation of Agila, Motiva and Forta<br />

supply units as well as the Pendula. They are<br />

designed <strong>for</strong> either anchor (fig. B + D) or boltthrough<br />

(fig. A + C) mounting. The ceiling<br />

supply can be directly attached to a concrete<br />

ceiling (direct mounting, see A1 and B1 in<br />

figure below).<br />

A variable fixture kit enables mounting up to<br />

1200 mm under the concrete ceiling in order<br />

to bridge the space between it and a false<br />

ceiling. A media flange, located between the<br />

raw and false ceilings, offers enough space <strong>for</strong><br />

connecting supply voltage, medical gas, vacuum<br />

and data cables. The installation manuals<br />

are included in the delivery.<br />

PLANNING GUIDELINES<br />

Dräger ceiling supply units are validated and<br />

certified <strong>for</strong> use with these ceiling fixtures.<br />

This certification is invalidated if non-Dräger<br />

ceiling mounting systems are used.<br />

The ceiling mount must meet the following<br />

criteria:<br />

– 12.500 Nm of torque at the lower end<br />

of the fixture<br />

– 9.000 N tensile <strong>for</strong>ce at the ceiling<br />

If the building design is not compatible with<br />

a Dräger ceiling mount, a custom design can<br />

be substituted. Certification of the two conditions<br />

stated above is mandatory. The per<strong>for</strong>mance<br />

specifications of a Dräger ceiling<br />

supply unit cannot be guaranteed if a mounting<br />

design is used which does not meet the<br />

required criteria listed above. In such cases,<br />

any given approvals are automatically invalidated.<br />

Available <strong>for</strong>:<br />

– Anchor mounting: Anchor mounting is<br />

possible if the ceiling is constructed<br />

with at least 250 mm of B 25 concrete<br />

– Bolt-through mounting: Bolt-through<br />

mounting is required if the ceiling<br />

thickness is less than 250 mm or the<br />

concrete quality is less than B 25.<br />

Without ceiling fixture set<br />

With ceiling fixture set<br />

A<br />

C<br />

Nm<br />

max. 1200 mm<br />

Bolt-through mounting<br />

N<br />

B<br />

D<br />

Anchor mounting<br />

TECHNICAL DATA<br />

Product<br />

Span intermediate ceiling distance<br />

Max. torque<br />

Max. tensile <strong>for</strong>ce at the ceiling<br />

Hole pattern<br />

Ceiling fixtures<br />

up to 1200 mm<br />

12.500 Nm<br />

9.000 N<br />

Ø 470 mm


PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> | DRÄGER P<strong>OR</strong>TFOLIO | 21<br />

Ceiling hood/Flat cover<br />

Dräger portfolio<br />

TYPES OF CEILING COVERS<br />

D-90841-2013<br />

D-90845-2013<br />

D-90848-2013<br />

D-90849-2013<br />

D-90846-2013<br />

Ceiling hood (height: 170 mm)<br />

Flat cover (height: 2 mm)<br />

Single/double/triple CSU<br />

Dräger ceiling fixtures are designed <strong>for</strong><br />

universal compatability. You can choose<br />

between single, double or triple type.<br />

Single CSU<br />

Double CSU<br />

Double CSU with third arm<br />

OPTIONS<br />

Single Double Triple<br />

CSU<br />

CSU/CSU<br />

Light CSU /Light CSU /CSU /Light<br />

Display CSU /Display CSU /CSU /Display<br />

X-Ray protection CSU /X-Ray protection CSU /CSU /X-Ray protection


22 |<br />

PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> |<br />

DRÄGER P<strong>OR</strong>TFOLIO<br />

Ceiling fixtures <strong>for</strong> <strong>OR</strong> lights/display arms<br />

PRODUCT DESCRIPTION<br />

The Dräger ceiling mount supports the installation<br />

of lights and display holders (on the<br />

central axis). The <strong>OR</strong> light/display arm can be<br />

fixed directly under the concrete ceiling<br />

(direct mounting). The distance tube allows<br />

mounting at heights of up to 1200 mm under<br />

a concrete ceiling in order to bridge the space<br />

above a false ceiling. The installation manuals<br />

are included with delivery.<br />

MOUNTING OPTIONS<br />

Direct mount<br />

This system is recommended as a direct<br />

mount or <strong>for</strong> use with a false ceiling up to a<br />

distance of 600 mm between the concrete<br />

and false ceiling or in combination with a laminar<br />

flow system. The suspension tube is available<br />

in five standard lengths (220/460/630/<br />

800/1000 mm). For service purposes, it<br />

should be considered that the electrical components<br />

must be installed in a switch box near<br />

the <strong>OR</strong> (max. 40 m).<br />

D-90851-2013<br />

With ceiling cover<br />

D-90851-2013<br />

With flat cover<br />

With laminar flow ceiling<br />

Distance tube<br />

This system is recommended <strong>for</strong> use with distances<br />

> 600 mm between the concrete and<br />

false ceiling or in combination with a laminar<br />

flow system. The distance tube is available in<br />

two standard lengths (600/1200 mm). For<br />

service purposes, it should be considered that<br />

the electrical components must be installed in<br />

a switch box near the <strong>OR</strong> (max. 40 m).<br />

D-90851-2013<br />

D-90851-2013<br />

With ceiling cover<br />

With flat cover<br />

With laminar flow ceiling<br />

Axis system Arm lengths Spring load arm lengths Vertical <strong>for</strong>ce (F) Torque (M)<br />

1-fold axis<br />

1f axis xxx 850 mm 930 mm 960 N 910 Nm<br />

1f axis AC2000 display 700 mm 930 mm 870 N 460 Nm<br />

1f axis AC3000 display 750 mm 930 mm 940 N 700 Nm<br />

1f axis xxx 1150 1150 mm 930 mm 1000 N 760 Nm<br />

1f axis xxx 1450 1450 mm 930 mm 1030 N 640 Nm<br />

2-fold axis<br />

2f axis xxx/xxx 700/850 mm 930 mm 1530 N 1280 Nm<br />

2f axis xxx/xxx 1150 700/1150 mm 930 mm 1400 N 1260 Nm<br />

3-fold axis<br />

3f axis xxx/xxx/xxx 700/850/1000 mm 930 mm 1920 N 1850 Nm<br />

4-fold axis<br />

4f axis xxx/xxx/xxx/xxx 700/850/1000/1150 mm 930 mm 2490 N 2680 Nm


PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> | DRÄGER P<strong>OR</strong>TFOLIO | 23<br />

Opera<br />

Additional to conventional ceiling supply systems Dräger has a beam system dedicated <strong>for</strong> the<br />

<strong>OR</strong> environment.<br />

D-148-2010<br />

Dräger portfolio<br />

The Opera system is the perfect solution <strong>for</strong><br />

operating rooms with laminar air flow ceilings<br />

and enhanced hygiene requirements.<br />

D-144-2010


24 |<br />

PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> |<br />

DRÄGER P<strong>OR</strong>TFOLIO<br />

The system is available in various configurations.<br />

A beam system with four beams is called<br />

Opera 4S.<br />

MT-1047-2008<br />

Opera 4S<br />

A system with three beams is the Opera 3S.<br />

It is also possible to have a two beam system<br />

called Opera 2S. For this variant the beams<br />

will be connected in the shape of an “L”.<br />

D-19626-2010<br />

Opera 3S<br />

TECHNICAL DATA<br />

Product<br />

Max. load suspension tube<br />

Opera<br />

500 kg<br />

Beam<br />

Max. load capacity<br />

300 kg<br />

Length<br />

2400 mm to 3600 mm<br />

Materials<br />

Aluminum, copper, silicon, plastic (ASA)<br />

Color RAL 9002<br />

Terminal units<br />

Beam profile designed <strong>for</strong> installations on the exterior of the beam.<br />

Max. number of terminal units per meter: 10 pcs.<br />

It must be considered that there are restrictions <strong>for</strong> installations at the ends of the beams.<br />

Max. number of terminal units (electric + gases + IT) based on components acc. to DIN standard.


PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> | DRÄGER P<strong>OR</strong>TFOLIO | 25<br />

The Opera system is a beam system with<br />

optional equipment carriers. These carriers<br />

are mounted to the beam with shuttles which<br />

can be moved along rails at the bottom of the<br />

beam or on the inner rails of the system.<br />

D-18767-2010<br />

Dräger portfolio<br />

D-18769-2010<br />

D-158-2010<br />

The shuttles on the underside of the beam can be equipped with a single pole or a rack system which contains two parallel tubes.<br />

These shuttles can be moved along the entire beam by using rotary corners.<br />

D-18761-2010<br />

D-18772-2010<br />

At the inner side of the beam there are other shuttle systems possible.<br />

You are able to mount a purisole lift system, a monitor arm or just an infusion pole.<br />

TECHNICAL DATA<br />

Shuttle pole /rack<br />

Shuttle pole rack MPL (Monitor, Purisole Lift)<br />

Max. load capacity 150 kg 60 kg 15 kg (Monitor)/50 kg (Purisole Lift)<br />

Swivel range 330° 0° 0°<br />

Brakes Linear movement/Rotation Linear movement Linear movement<br />

automatic+mech./mech. automatic+mech. mechanic


26 |<br />

PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> |<br />

DRÄGER P<strong>OR</strong>TFOLIO<br />

Light options<br />

The Opera supply system offers numerous integrated lighting options. You can choose<br />

between indirect light in different variations. In addition to an optional night light in the<br />

beam, you can also select an RGB lighting package to create a number of com<strong>for</strong>table<br />

and pleasing lighting scenes. Another option is to integrate a green LED stripe which<br />

can be used as an Endo-light.<br />

NIGHT LIGHT<br />

INDIRECT LIGHT<br />

LED-STRIPE (ENDO)<br />

D-145-2010<br />

D-147-2010<br />

D-146-2010<br />

RGB – LIGHT<br />

D-151-2010<br />

D-150-2010<br />

D-149-2010<br />

Colored lighting can generate a pleasant<br />

working atmosphere in the <strong>OR</strong> during long<br />

operations and is also useful when procedures<br />

are per<strong>for</strong>med on patients who are not fully<br />

sedated.<br />

TECHNICAL DATA<br />

Light module:<br />

Control:<br />

Electr. power:<br />

Operating panel:<br />

Voltage:<br />

3 colored fluorescent lamps (red - green - blue)<br />

Switchable (key button)<br />

3x24 W per beam<br />

8 fix colors and 3 color scenarios<br />

220-240 V, ± 10 % - 50/60 Hz


PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> | DRÄGER P<strong>OR</strong>TFOLIO | 27<br />

Workstation components<br />

Dräger offers a wide range of workstation components. These components are freely<br />

configurable and help to ensure an ergonomic workplace design.<br />

Dräger portfolio<br />

SHELVES<br />

D-27735-2009<br />

D-27733-2009<br />

D-27732-2009<br />

Our shelf portfolio includes a wide range of options. You can choose from different shelf sizes, the number of standard rails or<br />

add drawers according to your needs.<br />

SHELF SIZES/UASBLE SURFACE (W X D)<br />

430 x 340 mm<br />

630 x 340 mm<br />

430 x 480 mm<br />

530 x 480 mm<br />

630 x 480 mm<br />

860 x 480 mm<br />

STANDARD RAILS<br />

no rails<br />

side rails<br />

front rail<br />

control handles at front<br />

DRAWER OPTIONS<br />

1 drawer<br />

1 drawer with lock (incl. 2 keys)<br />

2 drawers<br />

2 drawers and 1 lock<br />

2 drawers and 2 locks<br />

drawers and 2 locks on a shelf<br />

(860 x 480 mm)<br />

1 drawer and 1 documentation module<br />

1 documentation module<br />

D-26256-2009


28 |<br />

PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> |<br />

DRÄGER P<strong>OR</strong>TFOLIO<br />

ARM TYPES<br />

INFUSION ARMS<br />

For all infusion arms, the options are almost<br />

limitless. These arms are available in both<br />

single- and double-arm configurations. Additionally,<br />

you can select the desired swivel arm<br />

length: 200/300/400 mm. All arms feature<br />

adjustable friction brakes (ball bearings) and<br />

integrated cable management.<br />

D-32268-2011<br />

Double swivel arm <strong>for</strong> 25 mm pole (infusion), mounted on a 38 mm pole<br />

MONIT<strong>OR</strong> ARMS<br />

For all monitor arms, a variety of options are<br />

available. These arms are available as lifting<br />

arms or lifting arms with additional swivel arm.<br />

You can also choose an optional keyboard holder.<br />

Additionally, you can select the desired<br />

swivel arm length: 200/300/400 mm. All<br />

arms feature adjustable friction brakes (ball<br />

bearings) and integrated cable management.<br />

D-111-2010


PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> | DRÄGER P<strong>OR</strong>TFOLIO | 29<br />

CABLE MANAGEMENT<br />

D-26583-2009<br />

D-26600-2009<br />

Dräger portfolio<br />

Cable channel<br />

Cable boxes<br />

COL<strong>OR</strong> CODE CONCEPT<br />

D-18601-2010<br />

Handles<br />

2-1298-94, 2-1201-94, 2-1065-96<br />

D-18609-2010<br />

Control at column<br />

ACCESS<strong>OR</strong>IES<br />

In addition to the components mentioned<br />

above, further components are available<br />

(e.g. infusion poles, application rods, racks,<br />

clamping rails, equipment holders etc.).<br />

Compact rail, cable holder, rail holder


30 |<br />

PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> |<br />

DRÄGER P<strong>OR</strong>TFOLIO<br />

Gas Management Systems<br />

Dräger Gas Management Systems (GMS) are designed to manage<br />

the hospital’s gas supply needs. Dräger offers complete gas supply<br />

systems which include all necessary components and distribution<br />

piplines.<br />

MT-0018-2006<br />

Gas Terminal Units<br />

<strong>Planning</strong> and designing Gas Management<br />

Systems <strong>for</strong> either complete hospitals or<br />

individual areas like a new IC or <strong>OR</strong> area is<br />

a challenging task. Different international<br />

standards and guidelines provide different<br />

in<strong>for</strong>mation, require different system configurations<br />

and different approaches to all the<br />

challenges occurring.<br />

Dräger is at your side when it comes to solving<br />

these challenges and optimizing the system<br />

design <strong>for</strong> the benefit of the hospital.<br />

We have more than 50 years experience in<br />

designing, building and testing Gas<br />

Management Systems.<br />

As every project is different and the systems<br />

tailor-made <strong>for</strong> the hospital, it is difficult to<br />

provide generic planning details. There<strong>for</strong>e<br />

we kindly ask you to contact your local<br />

Dräger representative <strong>for</strong> individual planning<br />

support.<br />

GAS TERMINAL UNITS<br />

The locations and quantities of gas terminal<br />

units are essential parts of the planning of<br />

every Gas Management System. Depending<br />

on the clinical procedures and applications,<br />

different gases are required in different hospital<br />

areas and the amount of units provided<br />

needs to match processes and devices used<br />

there. Dräger builds world-renowned gas terminal<br />

units since over 40 years and is offering<br />

outlets according to DIN, British Standard and<br />

other international types.<br />

D-1217-2009<br />

Alarm Management<br />

ALARM MANAGEMENT SYSTEM<br />

A complex system like a medical Gas<br />

Management System requires good operational<br />

management and provides a host of in<strong>for</strong>mation.<br />

Dräger Alarm Management Systems<br />

consolidate all this in<strong>for</strong>mation like operational<br />

and emergency alarms. Analogue operational<br />

data (e.g. pressure and flow) can also easily<br />

be integrated. All in<strong>for</strong>mation can be displayed<br />

where ever required with LED-, text- or touch<br />

display monitors and connected to integrated<br />

building management systems.<br />

D-16746-2010<br />

Distribution systems<br />

DISTRIBUTION SYSTEM<br />

Medical grade copper tubes according to ISO<br />

13348 with defined cleanliness and thorough<br />

testing regime in combination with high quality<br />

installation (e.g. inert gas shielding during<br />

welding) <strong>for</strong>m the backbone of the infrastructure<br />

which ensures that all gases are<br />

available at the required points of use. Proper<br />

system calculation according to Dräger<br />

Design Guidelines and international standards<br />

and comprehensive testing and installation<br />

ensure quality and per<strong>for</strong>mance of the overall<br />

system.


PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> | DRÄGER P<strong>OR</strong>TFOLIO | 31<br />

D-507-2014<br />

MT-7526-2004<br />

Dräger portfolio<br />

Reliable monitoring throughout the entire system<br />

AREA CONTROL UNITS<br />

All clinical areas need to be controlled individually<br />

by Area Control Units. This includes<br />

pressure gauges <strong>for</strong> each gas and alarms <strong>for</strong><br />

pressure deviations as well as lock-off valves<br />

and emergency inlet points.<br />

MT-5164-2004<br />

MT-5206-2004<br />

D-1216-2009<br />

Vacuum pumps<br />

Compressed air system<br />

Cylinder Manifold <strong>for</strong> oxygen<br />

VACUUM SYSTEM<br />

Vacuum is indispensable <strong>for</strong> surgical application<br />

(wound field aspiration) as well as bronchus<br />

suctioning in anaesthesia and intensive<br />

care treatment. Dräger vacuum systems with<br />

triplex or quadruplex pumps fulfill the ISO<br />

7396-1 and match all requirements in terms of<br />

safety and efficiency.<br />

COMPRESSED AIR SYSTEM<br />

Medical air systems <strong>for</strong> breathing purpuoe<br />

(4/5 bar) as well as <strong>for</strong> surgical tool air (7/8<br />

bar) must fulfill the ISO 7396-1, breathing air<br />

must also comply with the European Pharmacopoiea.<br />

Dräger medical air systems combine<br />

proven design, safe operations, and efficient<br />

reliability.<br />

CYLINDER MANIFOLDS<br />

Oxygen, Nitrous Oxyde and Carbon Dioxide<br />

can be supplied by 3-fold cylinder manifolds<br />

or by a liquid oxygen tank (V.I.E.) and a 2-fold<br />

cylinder manifold with a Dräger Gas Control<br />

Station as the core of the system.


32 |<br />

PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> |<br />

DRÄGER P<strong>OR</strong>TFOLIO<br />

Overview anesthesia portfolio<br />

ANESTHESIA DEVICES<br />

D-11215-2010<br />

D-47654-2012<br />

MT-0818-2004<br />

Fabius® plus<br />

Fabius® plus XL<br />

Fabius® Tiro<br />

D-12281-2011<br />

MT-6421-2008<br />

D-7656-2010<br />

Fabius® GS premium<br />

Primus®<br />

Apollo®<br />

D-64715-2012<br />

D-7116-2011<br />

D-64716-2012<br />

Primus® Infinity® Empowered<br />

Perseus® A500<br />

Zeus® Infinity® Empowered


PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> | DRÄGER P<strong>OR</strong>TFOLIO | 33<br />

MONIT<strong>OR</strong>ING & IACS<br />

D-90837-2013<br />

D-19739-2009<br />

Dräger portfolio<br />

Infinity® Delta<br />

Provide continuous monitoring and the same<br />

bedside level of care during transport<br />

Infinity® Acute Care System monitoring solution<br />

Created to meet the needs of all care areas – both<br />

at bedside and during transport<br />

VAP<strong>OR</strong>IZER AND SMARTPILOT® VIEW<br />

MT-2295-2008, D-879-2010<br />

D-7613-2011<br />

Vaporizers<br />

Designed with flexibility in mind – like Dräger<br />

anesthesia workstations<br />

SmartPilot® View<br />

Helps facilitate anesthesia induction, maintenance<br />

and recovery<br />

SPECIALTIES<br />

D-12287-2011<br />

D-9375-2009<br />

MT-1100-2007<br />

Fabius® MRI<br />

MRI-compatible anesthetic management requires a<br />

special room layout plan which con<strong>for</strong>ms to the<br />

patient‘s treatment path<br />

Fabius® Tiro M<br />

Offers the full spectrum of anesthesia ventilation to<br />

military anesthesiologists/cRNAs in the field<br />

The Surgical Display Controller enables you to<br />

display patient vital sign in<strong>for</strong>mation on a largescreen<br />

display in the Operating Room


34 |<br />

PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> |<br />

DRÄGER P<strong>OR</strong>TFOLIO<br />

D-60193-2012<br />

Accessories<br />

Day in, day out in the hospital you are faced with enormous time and cost pressures at the same time<br />

you are taking care of the welfare of your patients. You need technical medical accessories which you<br />

can use to unlock the full potential of your equipment, which work smoothly, guarantee the best possible<br />

care of your patients and help you to improve your processes. In short: accessories you can rely on,<br />

Dräger can supply it <strong>for</strong> you.<br />

SODA LIME<br />

SENS<strong>OR</strong>S<br />

Whether <strong>for</strong> patient monitoring, care and<br />

treatment of neonates, anesthesia, intensive<br />

care, the emergency department or on the<br />

ward: each original accessory from Dräger<br />

doesn't just fulfil all the official quality standards.<br />

We also measure our accessories<br />

against our own strict Dräger quality standards.<br />

MT-2004-2008<br />

MT-2086-2007<br />

E.g. CLIC Absorber<br />

E.g. flow sensor Spirolog®


PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> | DRÄGER P<strong>OR</strong>TFOLIO | 35<br />

VENTILATION<br />

D-19281-2009<br />

MT-11164-2006<br />

D-7059-2009<br />

Breathing circuits<br />

Breathing system filters and HMEs<br />

CO2 cuvettes<br />

D-12465-2014<br />

D-957-2010<br />

D-402-2010<br />

Catheter mount ErgoStar®<br />

Breathing masks<br />

Intubation kits<br />

MONIT<strong>OR</strong>ING<br />

WARD EQUIPMENT<br />

D-32549-2011<br />

D-10282-2011<br />

D-50637-2012<br />

Dräger portfolio<br />

RAIL ACCESS<strong>OR</strong>IES AND<br />

EXAMINATION LIGHTS<br />

E.g. Masimo® Sensors<br />

E.g. medical aspiration<br />

E.g. medical examination light VarioLux®


36 |<br />

PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> |<br />

W<strong>OR</strong>KPLACE EXAMPLES<br />

Workplace<br />

examples<br />

<strong>for</strong> <strong>OR</strong><br />

D-94674-2013


PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> | W<strong>OR</strong>KPLACE EXAMPLES | 37<br />

Workplace examples <strong>for</strong> the <strong>OR</strong><br />

The following pages will provide you with some in<strong>for</strong>mation and ideas <strong>for</strong><br />

several standard workplace configurations.<br />

Workplace examples<br />

The following pages show typical solutions <strong>for</strong><br />

various types of workplaces. They can be used as<br />

a starting point <strong>for</strong> new design concepts based on<br />

customer needs.<br />

Combine elements of two or more examples to<br />

develop the best possible solution <strong>for</strong> your client.


38 |<br />

PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> |<br />

W<strong>OR</strong>KPLACE EXAMPLES<br />

Supply units<br />

ANESTHESIA<br />

D-90855-2013<br />

D-90854-2013<br />

D-90853-2013<br />

AN-CSU with lifting arm with docked anesthesia<br />

device, monitoring and infusion (docking via<br />

equipment holder or MediaDocking)<br />

AN-CSU with media head and tubes <strong>for</strong> shelves,<br />

monitoring and infusion pumps<br />

AN-CSU with long column <strong>for</strong> shelves and<br />

infusion pumps. The anesthesia device is<br />

placed beside the supply unit<br />

DOCKING<br />

D-85706-2013<br />

Equipment holder <strong>for</strong> docking carts or anesthesia devices with<br />

magnetic docking control<br />

D-18338-2010<br />

Equipment holder <strong>for</strong> docking carts and anesthesia devices with<br />

automatic connection of all media via special MediaDocking interface


PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> | W<strong>OR</strong>KPLACE EXAMPLES | 39<br />

SURGICAL W<strong>OR</strong>KPLACE<br />

Workplace examples<br />

D-92502-2013<br />

D-90852-2013<br />

CSU with lifting arm, short column and shelf<br />

CSU with long column and three shelves<br />

MT-221-2001<br />

D-12307-2011<br />

CSU with lifing arm, long column, MediaDocking<br />

and docked MovaCart<br />

CSU with media head and tubes <strong>for</strong> shelves and<br />

infusion pumps


40 |<br />

PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> |<br />

W<strong>OR</strong>KPLACE EXAMPLES<br />

DISPLAY HOLDER<br />

D-57969-2012<br />

Multi display holder <strong>for</strong> up to 8 displays or one<br />

large display (60" screen size)<br />

D-16304-2009<br />

D-90843-2013<br />

D-90844-2013<br />

Short column with four separate display arms<br />

Display arm on central axis<br />

<strong>OR</strong> LIGHTS<br />

D-449-2014<br />

Single light head on central axis<br />

Double light combination with integrated HD-camera<br />

D-103620-2013<br />

Double light combination with display arm, without camera


PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> | W<strong>OR</strong>KPLACE EXAMPLES | 41<br />

ANESTHESIA DEVICES<br />

MT-0555-2008<br />

D-85719-2013<br />

Primus® on CSU. Connection via standard docking or MediaDocking<br />

(also as mobile version available)<br />

Perseus® A500 ceiling mount. Ergonomic anesthesia workstation<br />

with integrated remote control <strong>for</strong> ceiling supply unit<br />

D-7118-2011<br />

Workplace examples<br />

D-64716-2012<br />

Perseus® A500. Mobile flexible positioning of monitors and infusion pumps<br />

Zeus®


42 |<br />

PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> |<br />

GENERAL PLANNING<br />

General<br />

planning<br />

D-62809-2012


PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> | GENERAL PLANNING | 43<br />

General planning<br />

Learn more about planning an <strong>OR</strong>.<br />

<strong>Planning</strong> an <strong>OR</strong> is a complex and technically<br />

challenging process. This chapter will address the<br />

many questions encountered when designing a<br />

new <strong>OR</strong>.<br />

An intimate knowledge of the Dräger product portfolio<br />

and a thorough understanding of customer<br />

needs are vital prerequisites <strong>for</strong> success.<br />

General planning


44 |<br />

PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> |<br />

GENERAL PLANNING<br />

General planning<br />

It is important to keep the following in mind:<br />

CEILING SUPPLY UNITS<br />

– Which anesthesia device (lifting or mobile version)?<br />

– Which surgical devices have to be placed<br />

(size, weight, media connections)?<br />

– Is a cart (MovaCart) in use which could be docked<br />

at the supply unit?<br />

– Are there infusion pumps which have to be mounted?<br />

– What is the most ergonomic placement of devices?<br />

– Select a long or short column/head/equipment rack<br />

(e.g. AN-device under or in front column)<br />

– Select an arm system (where must the supply unit to<br />

be placed, where are parking the positions?)<br />

– Select height adjustable or non-adjustable system<br />

(e.g. different staff and heights)<br />

– Select workplace components (shelves, infusion poles)<br />

– Select handles/grips <strong>for</strong> brakes/lift (from where will<br />

the units be positioned? back, front)<br />

– Select cable management (are many cables and hoses to be<br />

expected? How can they be managed? Cable boxes,<br />

cable holder, channels?)<br />

CEILING MOUNT<br />

– Design the ceiling fixture <strong>for</strong> max. load capacity (upgradeability)<br />

– Select a uni-flange <strong>for</strong> possible double CSU (upgradeability)<br />

– Design of double CSU: clarify dead swivel range<br />

of arm system > long distance tubes on rear or front side<br />

– Consider a second anesthesia CSU <strong>for</strong> TIVA<br />

(Total intravenous Anesthesia) or infusion management<br />

– Possible parking positions <strong>for</strong> supply units <strong>for</strong> <strong>OR</strong> change<br />

> stops in arm systems have to be set accordingly<br />

– For MediaDocking: The column has to be equipped with the same<br />

media outlets. If a replacement anesthesia device is necessary<br />

hoses and cables must be available to ensure media supply<br />

– Consider possible collisions of pendants with air flow ceiling<br />

curtains or other interior fixtures<br />

– Consider space (clearance height) under arm systems<br />

(arm/motor hood) and grips of <strong>OR</strong> lights, space under columns<br />

<strong>for</strong> cleaning or further devices such as patient warmer or<br />

suction systems<br />

<strong>OR</strong> LIGHTS/CAMERAS/DISPLAYS<br />

– How many light heads are required?<br />

– Which size is necessary?<br />

– Do you have a central axis?<br />

(is it possible? movement of C-Bow)<br />

or a separate mounting on two sides?<br />

– Do you need a third light at foot end?<br />

(e.g. <strong>for</strong> vein harvesting)<br />

– Do you need a small examination light?<br />

(e.g. preparation of catheter)<br />

– Do you need an integrated camera or external camera?<br />

– How will the lights be controlled?<br />

(at cardanic, wall control panel, remote control, touch screen<br />

(integrated <strong>OR</strong>))<br />

– Where will displays be placed?<br />

(size and number is important)<br />

– Select a display holder<br />

(e.g. Pendula, display holder at central axis or holder<br />

at supply unit)<br />

D-7118-2011


PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> | GENERAL PLANNING | 45<br />

General planning<br />

D-8900-2011<br />

<strong>Planning</strong> an anesthesia workplace<br />

ELEMENTS OF AN ANESTHESIA<br />

W<strong>OR</strong>KPLACE<br />

– Anesthesia device (mobile or at CSU)<br />

– Infusion pumps (at CSU or AN-device, separate)<br />

– Anesthesia supply unit<br />

– Anesthesia trolley<br />

– Blood gas analyser<br />

– Cellsaver<br />

– Sonography device<br />

– Space, about 2 x 4 m<br />

– Must be easily accessible and an ergonomic operation<br />

of equipment must be ensured. Specify any meaningful parking<br />

position <strong>for</strong> angiography system<br />

– Standard hose lengths up to 3 m, special lengths on request<br />

MONIT<strong>OR</strong>ING/ANESTHESIA/IT<br />

– Different workstation configurations are possible<br />

(especially with Dräger Perseus A500)<br />

– Position of control cockpit has to be clarified<br />

– Position of patient monitoring has to be clarified<br />

– Position of IT portal has to be clarified<br />

– Position of infusion pumps has to be clarified<br />

– Optimize workplace <strong>for</strong> connection to ceiling supply unit,<br />

e.g. select shelf, drawer, small appliances rod


46 |<br />

PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> |<br />

GENERAL PLANNING<br />

Special <strong>OR</strong> types<br />

Most common are Hybrid, Integrated and Modular <strong>OR</strong>s.<br />

HYBRID <strong>OR</strong><br />

D-62775-2012<br />

This type of <strong>OR</strong> is the combination of a<br />

conventional operating room (e.g. heart<br />

surgery) and a imaging system (e.g. catheter<br />

lab). In this working environment, many devices<br />

are require red and their placement must<br />

be carefully considered. Imaging devices will<br />

require mobile and ceiling-based x-ray shielding<br />

and walls and windows must also contain<br />

lead shielding. For the interventionist, a<br />

sufficient number of monitors is essential.<br />

The main monitor is often quite large (up tp<br />

60"). The ceiling-mounted rails which support<br />

the imaging equipment necessitate long swivel<br />

arms in order <strong>for</strong> them to be able to reach<br />

the operating field.<br />

INTEGRATED <strong>OR</strong><br />

An Intergrated <strong>OR</strong> is a complete system.<br />

Nearly all functions in the room are controlled<br />

centrally by one system. This system controls<br />

the doors, blinds, light, table, <strong>OR</strong> lights and<br />

many other functions. In addition, video<br />

management of the room is done by these<br />

integration systems. These rooms will typically<br />

contain a large number of devices (e.g. control<br />

units) and interfaces. Providing sufficient<br />

shelf space <strong>for</strong> device placement can be challenging<br />

in rooms such as this. The devices<br />

involved often require special media outlets<br />

and cables (e.g. fibre optic).<br />

D-452-2014


PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> | GENERAL PLANNING | 47<br />

MODULAR <strong>OR</strong><br />

Modular <strong>OR</strong>s are comprised out of single<br />

modules. These modules are prefabricated.<br />

The prefabricated walls are already equipped<br />

with electrical sockets and gas outlets. This<br />

type of wall can allow the integration of<br />

modern materials such as glass, artwork or<br />

even illumination.<br />

The advantages of such Ors include short<br />

planning and construction times and flexible<br />

room design.<br />

General planning<br />

D-109188-2013


48 |<br />

PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> |<br />

GENERAL PLANNING<br />

General surgery<br />

Example layout <strong>for</strong> general surgery<br />

1. Anesthesia<br />

2. Surgery<br />

3. <strong>OR</strong> light combination


PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> | GENERAL PLANNING | 49<br />

This setup is an example <strong>for</strong> general surgery rooms. The configuration is not heavily equipped but flexible <strong>for</strong> different kinds<br />

of interventions.<br />

1. ANESTHESIA<br />

A single ceiling mount is selected to attach<br />

the swiveling and liftable anesthesia supply<br />

unit. The anesthesia device stands in below<br />

the supply unit. The supply unit has an horizontal<br />

media head – Movita head. This unit is<br />

equipped with all necessary media.<br />

Devices at supply unit<br />

Additional devices supplied by unit<br />

Anesthesia device<br />

Infusions<br />

Monitoring (often at AN-device)<br />

IT/Documentation<br />

Patient warmer<br />

Endotracheal aspiration (often at AN-device)<br />

2. SURGERY<br />

There is a single ceiling mount to attach a<br />

supply unit <strong>for</strong> surgery. The surgery supply<br />

unit is a short, liftable column with one shelf.<br />

The column itself is a slim Agila column.<br />

Devices at supply unit<br />

HF surgery device<br />

Additional devices supplied by unit<br />

Surgical suction device<br />

3. <strong>OR</strong> LIGHTS<br />

A double light solution is optimal. There is a<br />

main light (160.000 lux) and a satellite<br />

(110.000 lux).<br />

General planning


50 |<br />

PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> |<br />

GENERAL PLANNING<br />

Vascular surgery<br />

Example layout <strong>for</strong> vascular surgery<br />

4. Surgery & displays<br />

5. Anesthesia<br />

6. <strong>OR</strong> light combination


PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> | GENERAL PLANNING | 51<br />

This setup is an example <strong>for</strong> general surgery rooms. The configuration is not heavily equipped but flexible <strong>for</strong> different kinds<br />

of interventions. In addition to the standard general surgery room there is a mobile C-Bow in the room <strong>for</strong> x-ray imaging.<br />

X-ray protection is there<strong>for</strong>e mandatory (lead vests, skirts, leaded glass windows and walls. Images must be displayed on<br />

monitors.<br />

4. SURGERY AND DISPLAYS<br />

There is a double ceiling mount to attach a<br />

supply unit <strong>for</strong> surgery and a display holder.<br />

The surgery supply unit is a short, liftable<br />

column with one shelf. The column itself is a<br />

slim Agila column.<br />

SURGERY<br />

Devices at supply unit<br />

HF surgery device<br />

DISPLAYS<br />

Devices at supply unit<br />

Displays (2 x)<br />

Additional devices supplied by unit<br />

Surgical suction device<br />

Mobile C-Bow (often with wall socket)<br />

5. ANESTHESIA<br />

A single ceiling mount is selected to attach<br />

the swiveling and liftable anesthesia supply<br />

unit. The anesthesia device stands in below<br />

the supply unit. The supply unit has an horizontal<br />

media head – Movita head. This unit is<br />

equipped with all necessary media.<br />

Devices at supply unit<br />

Additional devices supplied by unit<br />

Anesthesia device<br />

Monitoring<br />

IT/Documentation<br />

Endotracheal aspiration (often at AN-device)<br />

Further infusions<br />

Patient warmer<br />

6. <strong>OR</strong> LIGHTS<br />

A double light solution is optimal. There is a<br />

main light (160.000 lux) and a satellite<br />

(110.000 lux).<br />

General planning


52 |<br />

PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> |<br />

GENERAL PLANNING<br />

Heart surgery<br />

Example layout <strong>for</strong> heart surgery<br />

7. Anesthesia/<br />

Cardio & display<br />

8. Surgery & display<br />

9. <strong>OR</strong> light combination


PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> | GENERAL PLANNING | 53<br />

The room is set up to allow <strong>for</strong> a large amount of equipment positioned at the patient’s head which must be operated by the<br />

anesthetist. As in vascular surgery rooms there is a mobile C-Bow in the room <strong>for</strong> x-ray imaging. There<strong>for</strong>e cardiac surgery<br />

protection is mandatory (Protection clothes, lead glass walls). Images, hemodynamic in<strong>for</strong>mation and other vital parameters<br />

must be displayed and be visible from several angles.<br />

7. ANESTHESIA/CARDIO<br />

A double ceiling mount is selected to attach<br />

the liftable anesthesia supply unit and a display<br />

arm. The anesthesia device stands in<br />

front of the supply unit. The supply unit has a<br />

long column. At this column an infusion pole is<br />

mounted. This column supplies the anesthesia<br />

and cardio workplace with all necessary<br />

media.<br />

Devices at supply unit<br />

Monitoring (or at AN-device)<br />

IT/Documentation (or at AN-device)<br />

Pacemaker<br />

Infusions (or at AN-device)<br />

DISPLAY HOLDER<br />

Display (1 x)<br />

Additional devices supplied by unit<br />

Anesthesia device<br />

Blood gas analysis<br />

Endotracheal aspiration (often at AN-device)<br />

Further infusions<br />

Patient warmer<br />

Heart lung machine HLM<br />

Hypothermia device<br />

8. SURGERY AND DISPLAY<br />

There is a double ceiling mount to attach a<br />

supply unit <strong>for</strong> surgery and a display holder.<br />

The surgery supply unit is a long, liftable<br />

column with a minimum of 3 shelves. The<br />

display is used <strong>for</strong> surgeon <strong>for</strong> vital parameters<br />

or x-ray imaging.<br />

9. <strong>OR</strong> LIGHTS<br />

A minimum of two light heads are necessary.<br />

The main light has a light intensity of 160.000<br />

lux and the satellite 110.000 lux. Preferably, a<br />

third light head with 110.000 lux <strong>for</strong> a second<br />

operating field (vein harvesting) should also<br />

be available.<br />

SURGERY<br />

Devices at supply unit<br />

HF surgery device<br />

ECG<br />

Defibrillator<br />

DISPLAY HOLDER<br />

Display (1 x)<br />

Additional devices supplied by unit<br />

Surgical suction device<br />

Sonographer<br />

Further infusions<br />

Cellsaver<br />

General planning


54 |<br />

PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> |<br />

GENERAL PLANNING<br />

Laparoscopy<br />

Example layout <strong>for</strong> laparoscopic surgery<br />

10. Anesthesia & display<br />

11. Surgery & display<br />

12. <strong>OR</strong> light combination


PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> | GENERAL PLANNING | 55<br />

This setup is organized <strong>for</strong> laparoscopic intervention. At both sides of the table is a display holder to mount displays.<br />

In this configuration, the endoscopic images are well within the surgeons’ field of view.<br />

10. ANESTHESIA AND DISPLAY<br />

A double ceiling mount is selected to attach<br />

the liftable anesthesia supply unit and swivel<br />

arm <strong>for</strong> a display. The anesthesia device is<br />

lifted and docked to the supply unit (via<br />

MediaDocking). The supply unit has a long<br />

column. At this column, a shelf is mounted.<br />

Devices at supply unit<br />

Anesthesia device<br />

Monitoring (or at AN-device)<br />

IT/Documentation (or at AN-device)<br />

Infusions (or at AN-device)<br />

DISPLAY HOLDER<br />

Display (1 x)<br />

Additional devices supplied by unit<br />

Endotracheal aspiration (often at AN-device)<br />

Patient warmer<br />

11. SURGERY AND DISPLAY<br />

There is a double ceiling mount to attach the<br />

surgery supply unit and a display holder. This<br />

unit is a long, swiveling column with 5 shelves<br />

a large amount of endoscopic equipment.<br />

SURGERY<br />

Devices at supply unit<br />

HF surgery device<br />

Endoscopic devices<br />

Control units<br />

Insufflator<br />

Light source<br />

Additional devices supplied by unit<br />

Surgical suction device<br />

Ultrasound surgical device<br />

12. <strong>OR</strong> LIGHTS<br />

A double light solution is optimal. The light<br />

heads have a light intensity of 110.00 lux and<br />

160.000 lux.<br />

DISPLAY HOLDER<br />

Display (1 x)<br />

General planning


56 |<br />

PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> |<br />

GENERAL PLANNING<br />

Neurosurgery<br />

Example <strong>for</strong> neurosurgery<br />

13. Anesthesia<br />

14. Surgery<br />

15. Display holder<br />

16. <strong>OR</strong> lights


PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> | GENERAL PLANNING | 57<br />

This setup is optimized <strong>for</strong> a neurosurgery intervention at the patient’s head. To have free access to the head end the<br />

anesthesia workplace is located at the feet end. The images from navigation system or microscope can be seen at both<br />

sides on screens which are mounted to display holders.<br />

13. ANESTHESIA<br />

A single ceiling mount is selected to attach<br />

the swiveling and liftable anesthesia supply<br />

unit. The anesthesia device stands in front<br />

of the supply unit. The supply unit has long<br />

Movita column. This unit is equipped with all<br />

necessary media.<br />

Devices at supply unit<br />

Monitoring (or at AN-device)<br />

IT/Documentation (or at AN-device)<br />

Additional devices supplied by unit<br />

Anesthesia device<br />

Infusions<br />

Patient warmer<br />

Endotracheal aspiration<br />

14. SURGERY<br />

A single ceiling mount is selected to attach a<br />

swiveling surgery supply unit. The supply<br />

unit has a long column. On this column, we<br />

have three shelves <strong>for</strong> devices.<br />

Devices at supply unit<br />

HF surgery device<br />

Neuro monitoring<br />

Additional devices supplied by unit<br />

Surgical suction device<br />

15. DISPLAY HOLDER<br />

There are to single display holders at each<br />

side of the operating table. The staff will<br />

have access to all required images in the<br />

room.<br />

DISPLAY HOLDER<br />

2 x Display (1 x)<br />

16. <strong>OR</strong> LIGHTS<br />

A double light solution is optimal. The light<br />

heads have a light intensity of 110.00 lux and<br />

160.000 lux.<br />

General planning


58 |<br />

PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> |<br />

GENERAL PLANNING<br />

Orthopaedic surgery<br />

Example <strong>for</strong> orthopaedic surgery<br />

17. Beam system<br />

18. <strong>OR</strong> lights


PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> | GENERAL PLANNING | 59<br />

This setup is optimized <strong>for</strong> an <strong>OR</strong> with highest hygienic requirements. A laminar air flow ceiling can be used without disturbing<br />

arms from supply units. With an <strong>OR</strong> table with a column a flexible setup of the room is possible. This flexibility is supported by a<br />

beam system with media outlets all over the beam. In addition, there are two shuttles with equipment carriers to place devices<br />

which can be moved along the entire beam. On the interior side, there are two more shuttles with display holders.<br />

17. BEAM SYSTEM<br />

A four-sided beam system with two equipment<br />

carriers which can be moved along the<br />

entire system. In addition, are two more shuttles<br />

<strong>for</strong> display holders. For a pleasent workplace<br />

atmosphere a number of lighting options<br />

have been integrated (e.g. indirect light,<br />

RGB light).<br />

Devices at supply unit<br />

Monitoring (or at AN-device)<br />

IT/Documentation (or at AN-device)<br />

Tourniquets (at shuttle)<br />

Endotracheal aspiration (at shuttle or AN-device)<br />

HF surgery device (at shuttle)<br />

Additional devices supplied by unit<br />

Anesthesia device<br />

C-Bow<br />

Surgical suction device<br />

Further infusions<br />

Patient warmer<br />

18. <strong>OR</strong> LIGHTS<br />

A double light solution is optimal. The light<br />

heads have a light intensity of 110.00 lux and<br />

160.000 lux.<br />

General planning


60 |<br />

PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> |<br />

GENERAL PLANNING<br />

CAD Drawings – Configuration 1<br />

In this chapter, all configurations shown in “Workplace Solutions” are listed.<br />

The numbers correspond to the numbers in the ground plans.<br />

Movita AK (L750-LiftExp.)


PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> | GENERAL PLANNING | 61<br />

Configuration 2<br />

General planning<br />

Agila 05 (L1000-LiftExp.)


62 |<br />

PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> |<br />

GENERAL PLANNING<br />

Configuration 3<br />

Polaris® 700 - Polaris® 500


PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> | GENERAL PLANNING | 63<br />

Configuration 4<br />

General planning<br />

Agila 05 (L1000-Lift) - AC3000 2 Displays


64 |<br />

PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> |<br />

GENERAL PLANNING<br />

Configuration 5<br />

Movita AK (L750-Lift)


PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> | GENERAL PLANNING | 65<br />

Configuration 6<br />

General planning<br />

Polaris® 700 - Polaris® 500


66 |<br />

PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> |<br />

GENERAL PLANNING<br />

Configuration 7<br />

Movita 13MD (M500-Lift) - AC2000 1 Display


PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> | GENERAL PLANNING | 67<br />

Configuration 8<br />

General planning<br />

Movita 13 (M750-750) - AC2000 1 Display


68 |<br />

PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> |<br />

GENERAL PLANNING<br />

Configuration 9<br />

Polaris® 700 - Polaris® 500 - Polaris® 500


PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> | GENERAL PLANNING | 69<br />

Configuration 10<br />

General planning<br />

Movita 13MD (M1000-Lift) - AC2000 1 Display


70 |<br />

PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> |<br />

GENERAL PLANNING<br />

Configuration 11<br />

Movita 15 (M750-750) - AC2000 1 Display


PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> | GENERAL PLANNING | 71<br />

Configuration 12<br />

General planning<br />

Polaris® 500 - Polaris® 500


72 |<br />

PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> |<br />

GENERAL PLANNING<br />

Configuration 13<br />

Movita 13 (M1250 - Lift)


PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> | GENERAL PLANNING | 73<br />

Configuration 14<br />

General planning<br />

Movita 13 (S1250 - 1000)


74 |<br />

PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> |<br />

GENERAL PLANNING<br />

Configuration 15<br />

AC3000 Display


PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> | GENERAL PLANNING | 75<br />

Configuration 16<br />

General planning<br />

Polaris® 500 - Polaris® 700


76 |<br />

PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> |<br />

GENERAL PLANNING<br />

Configuration 17<br />

Opera 4S


PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> | GENERAL PLANNING | 77<br />

Configuration 18<br />

General planning<br />

Polaris® 700 - Polaris® 500


78 |<br />

PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong>


PLANNING DOCUMENTS F<strong>OR</strong> <strong>OR</strong> | CONTACT | 79<br />

Contact<br />

For planners and architects:<br />

CONTACT F<strong>OR</strong> PROJECTS AND TECHNICAL PLANNING:<br />

Marco Patzer<br />

Project Engineer<br />

+49 451 882 2419<br />

marco.patzer@draeger.com<br />

Sebastian Rahn<br />

Project Engineer<br />

+49 451 882 2446<br />

sebastian.rahn@draeger.com<br />

WEB:<br />

www.draeger.com/planning<br />

Subject to change.<br />

Not all products are available in all parts of the world.<br />

More in<strong>for</strong>mation: www.draeger.com


C<strong>OR</strong>P<strong>OR</strong>ATE HEADQUARTERS<br />

Drägerwerk AG & Co. KGaA<br />

Moislinger Allee 53–55<br />

23558 Lübeck, Germany<br />

www.draeger.com<br />

Manufacturer:<br />

Dräger Medical GmbH<br />

Moislinger Allee 53–55<br />

23558 Lübeck, Germany<br />

As of August 2015:<br />

Dräger Medical GmbH changes<br />

to Drägerwerk AG & Co. KGaA.<br />

REGION EUROPE CENTRAL<br />

AND EUROPE N<strong>OR</strong>TH<br />

Dräger Medical GmbH<br />

Moislinger Allee 53–55<br />

23558 Lübeck, Germany<br />

Tel +49 451 882 0<br />

Fax +49 451 882 2080<br />

info@draeger.com<br />

REGION EUROPE SOUTH<br />

Dräger Médical S.A.S.<br />

Parc de Haute Technologie d’Antony 2<br />

25, rue Georges Besse<br />

92182 Antony Cedex, France<br />

Tel +33 1 46 11 56 00<br />

Fax +33 1 40 96 97 20<br />

dlmfr-contact@draeger.com<br />

REGION MIDDLE EAST, AFRICA<br />

Dräger Medical GmbH<br />

Branch Office<br />

P.O. Box 505108<br />

Dubai, United Arab Emirates<br />

Tel +971 4 4294 600<br />

Fax +971 4 4294 699<br />

contactuae@draeger.com<br />

REGION ASIA / PACIFIC<br />

Draeger Medical<br />

South East Asia Pte Ltd.<br />

25 International Business Park<br />

#04-27/29 German Centre<br />

Singapore 609916, Singapore<br />

Tel +65 6572 4388<br />

Fax +65 6572 4399<br />

asia.pacific@draeger.com<br />

REGION CENTRAL<br />

AND SOUTH AMERICA<br />

Dräger Panama Comercial S. de R.L.<br />

Complejo Business Park,<br />

V tower, 10th floor<br />

Panama City<br />

Tel +507 377 9100<br />

Fax +507 377 9130<br />

contactcsa@draeger.com<br />

90 68 999 | 1411-1 | Communications & Sales Marketing | AW | LE | Printed in Germany | Chlorine-free – environmentally compatible | Subject to modifications | © 2014 Drägerwerk AG & Co. KGaA

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!