5.3 Resetting the Accumulated<br />

Mass Totals<br />

Selecting and confirming the Reset option<br />

will reset the accumulated mass totals<br />

displayed on the main signal display screen.<br />

5.4 Taking a Gas Sample for<br />

Analysis<br />

For Sentrum to accurately track the<br />

fluctuating methane concentration in the<br />

gas transport pipe, it needs to acquire<br />

knowledge about the “Gas Signature” of the<br />

installation. This information is derived from<br />

gas chromatograph analysis of a gas sample<br />

collected from the installation.<br />

To collect a sample of gas for analysis the<br />

following procedure must be followed:<br />

1. Physically extract the sample of gas from<br />

the main line extraction point.<br />

2. Then enter the MDM Setup menu and<br />

confirm that the sample has been taken.<br />

3. Selecting and confirming the Sample<br />

Taken option will save the uncorrected<br />

methane sensor output with the<br />

corresponding time and date. This<br />

information is used by the controller to<br />

calibrate the correction factors once the<br />

results of the analysis are available.<br />

4. Information about the last sample taken<br />

can be viewed by selecting the View<br />

Sample option.<br />