16.06.2015 Views

Sample Preparation Into Ultra-thin Sections.

Sample Preparation Into Ultra-thin Sections.

Sample Preparation Into Ultra-thin Sections.

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

<strong>Sample</strong> <strong>Preparation</strong> <strong>Into</strong> <strong>Ultra</strong>-<strong>thin</strong> <strong>Sections</strong>.<br />

Contents<br />

Introduction<br />

Tissue selection<br />

Fixation<br />

Glutaraldehyde<br />

Osmium tetroxide OsO 4<br />

Potassium permanganate - KMnO 4<br />

Acroline<br />

Embedding<br />

Dehydration<br />

Resin infiltration<br />

Polymerization<br />

<strong>Ultra</strong>microtomy<br />

Trimming the capsule<br />

Making glass knives<br />

Thin sectioning<br />

Staining sections<br />

Miscellaneous procedures<br />

Thick sections - light microscopy<br />

Whole mounts on coated grids<br />

Arts and Formulae<br />

Photography<br />

Developing film<br />

Developing prints<br />

1


Introduction.<br />

The Electron Microscope was developed from the coalescence of<br />

several scattered ideas and hypotheses. The electron itself was discovered<br />

at about 1897 and was shown to have wave properties in 1924. Much of<br />

the early work with electrons centered on how to generate electrons and<br />

how to deflect or aim them. Various "lens" were contrived and by 1931,<br />

two German scientists were demonstrating the "first" EM which was largely<br />

a modified oscillograph. No specimens could be viewed but it was<br />

predicted that an electron microscope would have much better resolution<br />

than the light microscope. By the late 1930's and early 1940, commercial<br />

EMs were available with moderate resolution. These were used mainly to<br />

study electron optics and to make improvements. Procedures for specimen<br />

preparation came much later.<br />

For the biologist, the EM is one of the most powerful tools available<br />

for cell and tissue studies. It should be realized that electron microscopy is<br />

not a science on its own, but merely a technique or tool. A thorough<br />

understanding of the EM and of cells and their ultrastructures are an asset<br />

to understanding most other areas of biology.<br />

There are now many types of electron microscopes, but the two most<br />

common types are the transmission electron microscope (TEM or just EM)<br />

and the scanning electron microscope (SEM). Simply put, the SEM scans<br />

the surface of coated specimens with an electron beam and by detecting<br />

electrons scattered (reflected) by the object, forms an image on a TV like<br />

monitor. This image is usually aesthetically pleasing and has a resolution<br />

of 50 µ and up. The TEM transmits a beam of electrons through a<br />

specimen and forms an image based on the removal of electrons from the<br />

beam by the specimen (basically a high resolution shadow). Resolution can<br />

attain


of the tungsten filament (what happens if you crack the globe of an<br />

incandescent light bulb?). This vacuum is usually in the range of 10 -6 to<br />

10 -7 µ of mercury and is achieved using a "diffusion" pump. For all<br />

practical purposes, living tissues or cells cannot be viewed under the TEM<br />

since the specimens are subjected to such high vacuum, heat and intense<br />

radiation from the electron beam. This would suffice to kill the cells either<br />

by volatilization of water (and other low melting point substances),<br />

denaturation from heat, or ionizing radiation. In addition, the radiation<br />

undoubtedly causes many chemical changes to occur. Polymers often<br />

become insoluble and sublimation may occur.<br />

The most popular and useful procedure used with EM is the <strong>thin</strong><br />

sectioning technique. This technique can be broken down into the<br />

following sections and each will be dealt with separately.<br />

1. Tissue isolation<br />

2. Fixation with glutaraldehyde, OsO 4 and occasionally KMnO 4<br />

3. Embedding in a plastic resin<br />

4. <strong>Ultra</strong>microtomy<br />

5. Post-staining of <strong>thin</strong> sections<br />

TISSUE SAMPLES:<br />

Tissues must be killed and fixed in a way to stabilize their structures<br />

in the EM environment and to accurately reflect the true structure of the<br />

tissue. The time between isolation of the tissue (e.g., dissection) and<br />

addition of fixative should be minimized so as to avoid post-mortem<br />

changes. The tissue size should be kept small. Blocks should be cut to less<br />

than 1mm 3 in order to ensure thorough and quick penetration of fixatives<br />

and embedding solutions. Such size concerns are not pertinent to cell<br />

suspensions. It is usual practice to suspend the cells in the preparative<br />

solutions and to pellet the cells by centrifugation between steps. However,<br />

if a tissue block is too large, fixatives and embedding resins will usually<br />

not penetrate to the middle. Some tissues of low density (e.g., lung, root<br />

tips) are the exception. In the dense tissues, one can often note a "halo"<br />

effect.<br />

3


Cut in half<br />

HALO EFFECT<br />

Dark with<br />

Osmium<br />

TISSUE BLOCK<br />

Pale or White<br />

No Osmium<br />

Animal tissues pose a few unique problems for optimum preparation.<br />

Lacking rigid cell walls, the tissue is often limp and difficult to cut into<br />

small blocks (e.g., liver). This can be overcome by allowing larger pieces to<br />

be fixed for a short period of time in the primary fixative (glutaraldehyde,<br />

10-15 minutes). The tissue will then be somewhat more firm and easily<br />

cut into smaller pieces.<br />

Some tissues are hard and mineralized (e.g., bone) and must be<br />

demineralized with chelating or acidic solutions. This is fairly rare<br />

however, and tissues such as hair and nails can be prepared without<br />

special treatment. Some animal tissues are fairly dense and longer<br />

treatment times may be necessary. This determination is made by<br />

experience and/or trial and error.<br />

Plant tissues also present some unique problems during preparation.<br />

There is usually a higher water content with most mature plant cells<br />

having a large central vacuole. This necessitates attention to the<br />

dehydration step of embedding. Another problem often encountered is the<br />

presence of wax on certain plant surfaces which may retard the<br />

penetration of fixatives and embedding solution. It may also cause the<br />

separation of the tissue from the surrounding plastic upon trimming and<br />

sectioning. The waxes (cutin and suberin) may be partially removed with<br />

organic solvents (acetone, ether) prior to processing. It is not necessary to<br />

remove all wax in that the solvents tend to make the waxy surface less of<br />

a barrier and more tenacious to the plastic (similar to cleaning and<br />

4


abrading a surface before painting or gluing).<br />

problems since they are hydrophilic gels.<br />

Cell walls pose no special<br />

FIXATION<br />

As mentioned earlier, living tissues cannot be viewed with the<br />

Electron Microscope. The goal of specimen preparation is to preserve the<br />

tissue in a form which hopefully represents its' natural, in vivo form.<br />

Fixation serves to "kill" the cells, and to stabilize and preserve them and<br />

their structures during subsequent preparation steps. Throughout the last<br />

three decades, several chemical fixatives have been studies for use in<br />

specimen preparation. Out of this, two fixatives have emerged as virtually<br />

"universal" and will be described here.<br />

Glutaraldehyde - a five carbon structure<br />

with an active aldehyde on each end.<br />

GLUTARALDEHYDE<br />

Glutaraldehyde is referred to as a bifunctional<br />

fixative due to the two terminal<br />

aldehydes and is usually used as the first of two fixatives. It is fairly<br />

stable in concentrated form and at cold temperatures (-20°C). At room<br />

temperature and especially when diluted to working strength (1-3%), it is<br />

unstable and impurities and polymers accumulate. Oxidation to glutaric<br />

acid is the usual consequence but this reaction is inhibited by low pH<br />

which is produced by the oxidation. Thus, it is somewhat of a self-limiting<br />

process, but also explains why buffered (pH 7) solutions are so unstable.<br />

The stock solution of glutaraldehyde (even when purchased new) should<br />

be checked often for impurities. This is easily done spectrophotometrically<br />

with a 0.5-1% solution. The aldehyde has an absorption maximum of about<br />

280 nm while impurities absorb maximally at 235 nm. The impurity peak<br />

should be half the height of the aldehyde peak. The pH of the stock<br />

solution should be above 3.5 as well. If impurities have accumulated to<br />

unacceptable levels, the glutaraldehyde can be easily redistilled in a<br />

vented fume hood with the distillate collected at 100 O C in small fractions<br />

until the pH of the distillate is less than 4.0. It is then convenient to freeze<br />

several aliquots at -20°C at which it is chemically stable for many months.<br />

Glutaraldehyde is relatively safe to use. Avoid skin and eye contact<br />

and never pipette by mouth. Use in a well ventilated area since strong<br />

fumes can irritate and mildly fix the epithelial lining of the<br />

nasopharyngeal tissues.<br />

5


Fixation and stabilization are due to the cross linking of structures<br />

that are reactive with aldehydes. Structures that are composed of proteins<br />

(enzymes, etc.), glycoproteins, nucleoproteins, lipoproteins, glycogen, and<br />

starch (occasionally) will react with the aldehydes. Most structures wi<strong>thin</strong><br />

the cell have these components. Membranes loose their fluidity and<br />

usually become very permeable. Structures that do not react with<br />

glutaraldehyde will then tend to diffuse out of the cell. It is important to<br />

follow this fixation step (often called a pre-fixation) with a second step<br />

(often called post-fixation) using another fixative (see next).<br />

Glutaraldehyde fixation does not cause significant shrinkage and can be<br />

carried out at room temperature. Cold temperatures cause much of the<br />

cytoskeleton of the cell to "dissolve" or disassemble, thus altering the<br />

ultrastructural representation. A 1-3% buffered solution (pH 7) is<br />

recommended and fixation should not exceed 60 minutes unless the tissue<br />

is naturally dense or impermeable (e.g., some insects and plant tissues).<br />

The choice of buffer is important. Veronal buffers (containing barbitals)<br />

should not be used with aldehyde fixatives and phosphate buffering may<br />

form a precipitate in the presence of calcium and uranyl ions. If the<br />

specimen is known to contain these ions, use a different buffer (e.g., Tris,<br />

Hepes, cacodylate). Also, the use of phosphate buffers with the<br />

glutaraldehyde fixative occasionally causes a precipitin to form during the<br />

second fixation step with osmium tetroxide. To prevent this, wash the<br />

specimen with saline or water after the first fixation so as to remove all<br />

traces of the phosphate. This problem rarely arises however and its cause<br />

is not understood.<br />

Osmium tetroxide - OsO 4<br />

Note: Osmium is extremely dangerous, the crystals, liquid<br />

and vapors are all hazardous. The vapors can fix the cornea and lens<br />

of the eye and both vapors and liquid are absorbed rapidly and act as a<br />

nerve gas (it was in fact used as this in the World Wars) and attacks the<br />

CNS. Use only with proper ventilation (e.g., fume hood). It<br />

usually is purchased in crystalline form in preweighed sealed glass<br />

ampoules. Always prepare the fixative solution in a fume hood. Report<br />

any and all accidents immediately.<br />

Usually a 1-2% buffered solution (see appendix) is used as the second<br />

fixative and the fixation should be complete in 60-90 minutes since the<br />

osmium molecule is small and penetrates rapidly. Most workers agree that<br />

osmium works by saturating double (or triple) bonds since it is such a<br />

strong oxidizing agent. For this reason, it is deposited at lipoidal sites quite<br />

6


heavily thus causing them to look dark. Thus, membranes and lipid<br />

droplets usually "stain" darkly with the osmium. It should be noted also<br />

that since osmium is used as an aqueous solution, it has a tendency to not<br />

reach the middle hydrophobic region of some membranes. This gives a<br />

tri-layered appearance (dark-light-dark) to these membranes and is easily<br />

seen at higher magnifications. The tissue specimen will begin to turn black<br />

almost immediately upon the addition of the osmium. The more dense the<br />

tissue, the darker it will appear. (That's how you will know if you get it on<br />

you, your skin, clothes, etc. will turn black. If you do spill some on<br />

yourself, don't panic, wash it off with lots of water and immediately notify<br />

your instructor or health office). The tissue often becomes brittle when<br />

over fixed in osmium. It is advisable to use this fixation at refrigeration<br />

temperatures so as to decrease the volatility of the solution. The unused<br />

solution should be stored at 4°C in a scrupulously clean, foil wrapped 50ml<br />

volumetric flask that is tightly corked (do not use ground glass or rubber<br />

stoppers - they leak). The long narrow neck of the flask retards<br />

evaporation of the fixative and offers a "handle" to the user. Use a long<br />

tipped Pasteur pipette, taking care not to draw the osmium up into the<br />

pipetting bulb (bippy). The flask should be stored in a refrigerator, and be<br />

sure it is well stoppered. If it leaks, the interior of the refrigerator will<br />

gradually turn black and there is a possibility that the osmium vapors can<br />

accumulate in the confined space to dangerous levels. The osmium<br />

solution will appear purple or violet in color when degraded or<br />

"exhausted". It should be carefully pipetted into a flask containing 95%<br />

ethanol. This will degrade the osmium for later disposal. Keep this waste<br />

flask in the fume hood at all times.<br />

Potassium permanganate - KMnO 4<br />

Occasionally, an investigation may center on the study of<br />

membranous structures and the cytoplasmic matrix is not of interest. In<br />

this circumstance, a 1-2% buffered solution of KMnO 4 can be used as the<br />

sole fixative or in tandem with OsO 4 (wash in between the two - they react<br />

together). This is a rapid process and KMnO 4 may also be used as a post<br />

stain to enhance contrast from the Glutaraldehyde/OsO 4 preparation. This<br />

is usually not necessary however.<br />

Most non-membranous structures are washed away in subsequent<br />

embedding steps. The time required for the fixation is usually 30 minutes<br />

and the KMnO 4 kills the cells quickly, being a strong oxidizing agent. It<br />

will permanently stain skin and clo<strong>thin</strong>g but it is not as hazardous as the<br />

7


other fixatives. In fact, it is often used to treat fungal infections of the<br />

skin. That doesn't mean that you can drink it however, so it should not be<br />

pipetted by mouth. It is chemically stable when kept at 4°C. It binds to<br />

the same structures as does OsO 4 but is not as electron dense.<br />

Acroline<br />

A very toxic, flammable and volatile substance, it penetrates very<br />

rapidly and is thus good for fixing large and dense tissues. It is not<br />

thought of as a "common" fixative since it is potentially hazardous to use.<br />

In addition, it does not fix lipids (in fact it dissolves them), denatures most<br />

enzymes to inactive form and does not preserve the cytoskeletal network<br />

very well. If it is deemed absolutely necessary to use this aldehyde, best<br />

results are obtained by combining it with other aldehydes and following<br />

with osmium post-fixation. Partially degraded and polymerized solutions<br />

appear to be as effective as fresh or redistilled acroline. Use extreme care<br />

when handling acroline. It is included here as a precaution to those who<br />

choose to use it and to those who read reference to it.<br />

EMBEDDING<br />

The fixatives as well as most cellular components are aqueous. The<br />

plastic resins that are used for embedding tissues are not miscible with<br />

water. Thus, an intermediate solvent that is miscible with both water<br />

and plastic resin is needed. Although there are many to choose from,<br />

acetone is probably the best. Some microscopists prefer ethanol, often out<br />

of habit from light microscopic procedures. Ethanol reacts with unbound<br />

OsO 4 to form a fine dense precipitate thus extensive washing after osmium<br />

fixation is needed. Acetone, which does not react, requires only minimal<br />

washing to remove the osmium and buffer salts. Methanol is less reactive<br />

than ethanol but has no advantage over acetone. In essence, water is<br />

replaced by solvent and solvent will be replaced with plastic.<br />

DEHYDRATION<br />

Use glass vials (or centrifuge tubes) since acetone can dissolve many<br />

plastics. Although a graded series of acetone solutions is commonly used -<br />

it is unnecessary. A three step dehydration process is adequate, -50%,<br />

95%, and two changes of 100% acetone (re-distilled - stored with molecular<br />

drying sieves, see appendix). It is essential that all water diffuses out of<br />

the tissue, otherwise holes will be created in the sections when viewed<br />

with the EM since residual water will vaporize under the extreme vacuum.<br />

8


Two changes of 100% acetone are a precautionary measure. As water<br />

diffuses into the first 100% acetone, it is no longer absolute. A second<br />

change dilutes out the water molecules even more. Use a Pasteur pipette<br />

to add and remove the acetone solutions to and from your specimen vial<br />

(as opposed to transferring the tissue block to a new vial with the next<br />

solution).<br />

Dehydration may cause some changes in the secondary and tertiary<br />

structures of macromolecules and usually causes some shrinkage of the<br />

tissue. The shrinkage is usually proportional to the water content of the<br />

specimen. Fixation lessens this effect. Do not let the tissue dry in air.<br />

Make transfers rapidly but neatly. During dehydration as during fixation<br />

and embedding steps, keep the vials capped. 100% acetone is hydroscopic<br />

and will absorb water from the air. Acetone will often dissolve unfixed or<br />

poorly fixed components of cells such as saturated lipids (which do not<br />

react with osmium) and chlorophyll and other lipoidal membrane<br />

components. Starch is difficult to fix but is often so highly polymerized<br />

and cross-linked in vivo that it is often "naturally fixed". However, it will<br />

occasionally be leached out during dehydration.<br />

RESIN INFILTRATION<br />

There are many types of plastic resins available for embedding<br />

tissue, each having attributes. The three principle types of resins used are<br />

the epoxy resins, polyester resins and methacrylate resins. The most<br />

commonly used resins are the epoxides Epon (Epon is no longer made but<br />

other similar resins are available with similar names e.g., Epox) and<br />

Araldite. They have adequate viscosity, are fairly stable under the intense<br />

electron beam, and are of very fine grain. The purpose of the embedding<br />

medium is to provide a stable, hard matrix throughout a tissue or cell in<br />

order that very <strong>thin</strong> sections may be cut, usually on the order of 400-800<br />

A. Wax such as the light microscopist paraffin is not firm enough for such<br />

<strong>thin</strong>ness and it will melt under the electron beam. Epon and Araldite are<br />

both epoxide resins and when polymerized are virtually indestructible and<br />

insoluble (as are tissues embedded in them). Remember that water is<br />

replaced with acetone and acetone is replaced by plastic.<br />

Note - Epoxys can be irritating to skin and eyes; use with caution. Many<br />

epoxides are known to be carcinogenic. The polymerized capsule however<br />

is not carcinogenic. Use acetone on a cloth or wipe to remove any resins<br />

from your skin. For eye contact - flush with warm (not hot) water.<br />

9


Do not pipette resins. They are viscous enough that pipetting is<br />

inaccurate. Simply pour the components into a 50ml disposable beaker<br />

according to the formula given. The plastic mixture to be used contains<br />

both epon and araldite along with DDSA (Dodecenyl succinic anhydride)<br />

and NMA (nadic methyl anhydride) which are curing agents, and increased<br />

amount of DDSA will result in softer plastic when polymerized. A<br />

polymerizing catalyst or accelerator, DMP-30 (dimethyl aminomethyl<br />

phenol) is used to speed the polymerization process.<br />

A four (4) step series of plastic concentrations is used for the<br />

embedding process:<br />

1. 3 part acetone - 1 part plastic mixture (w/o DMP-30)<br />

2. 1 part acetone - 1 part plastic mixture (w/o DMP-30)<br />

3. 1 part acetone - 3 parts plastic mixture (add 4 drops DMP-30<br />

for every 10ml used<br />

4. Pure plastic mixture with DMP-30 (two changes)<br />

The plastic mixture is:<br />

16 ml of a mixture of 5 parts Epon and 3 parts Araldite (506)<br />

9 ml NMA<br />

10 ml DDSA<br />

Total = 35 ml - stir exhaustively<br />

For pure plastic mixture steps of embedding (step 4) add 15-17<br />

drops of DMP-30 to the above mixture using a disposable Pasteur pipette.<br />

DMP-30 is kept refrigerated. Allow it to warm to room temperature prior<br />

to use to avoid condensation of water into the bottle. Epoxy resins,<br />

especially araldite are also somewhat hydroscopic. Be sure to stir the<br />

mixture well. It will quickly turn amber in color but will lighten to nearly<br />

colorless during polymerization if done slowly. Two changes of the pure<br />

plastic mixture with DMP-30 is recommended in order to ensure that all of<br />

the acetone is removed (i.e., replaced) in the tissue. Since the pure plastic<br />

mixture is viscous it is easier to remove the tissue block with a hooked<br />

needle (dissecting) and place it in a fresh vial containing the pure plastic.<br />

This will minimize the carryover of any acetone that had diffused into the<br />

first pure plastic step. Keep the vials capped during embedding.<br />

The times for each step will vary depending upon the density and<br />

permeability of the tissue. For most tissues a 30 minute period for each<br />

step is adequate. It is better to have each step longer than necessary than<br />

shorter than necessary - residual acetone can result in poor sections.<br />

10


For the final step, use the 00 size polyethylene capsule molds<br />

available. Fill them to about 2mm from the top and place your tissue block<br />

to the bottom of the mold. Place a small paper label (written in pencil)<br />

around the top perimeter of the mold. For cell suspensions, pellet the cells<br />

with centrifugation in the pure plastic mixture after the allotted time.<br />

Using a Pasteur pipette, draw off the pellet as a cell slurry and place 2-3<br />

drops of the slurry in the capsule molds. Layer pure plastic mixture over<br />

this to wi<strong>thin</strong> 2mm of the top and centrifuge the capsule mold in a tabletop<br />

clinical centrifuge on a setting of 5 or 6 until the cells are concentrated<br />

at the tip of the mold. Then place a label around the top rim of the<br />

capsule.<br />

POLYMERIZATION:<br />

Allow the tray(s) of capsules to stand (wrapped in aluminum foil)<br />

overnight at room temperature and then place the capsules in an oven at<br />

60°C for 2-3 days. If the correct amount of DMP-30 has been used, the<br />

capsules should be adequately polymerized wi<strong>thin</strong> 3 days and the plastic<br />

will have lost most of the amber color. A good test for correct<br />

polymerization is to try and dent one of the side ridges of the tip of the<br />

capsule with a fingernail (after removing the capsule from the mold of<br />

course). If there is an indentation from the fingernail the polymerization<br />

at 60°C should continue until the capsule is hard enough to show no<br />

indentations. To remove the capsule from the mold, carefully cut the mold<br />

lengthwise with a razor blade and peel the cut edges from the top (not the<br />

tip) of the capsule. The capsule can then be easily removed. If the side<br />

facets near the tip show cracks and/or bulging, it usually indicates too<br />

rapid polymerization.<br />

OTHER RESINS:<br />

Relatively recently, a new monomeric resin called LR-White has been<br />

introduced from England. It is a single solution that is stable at 4°C and is<br />

used with 4 to 6 changes after dehydration which must be carried out with<br />

absolute ethanol. Acetone can not be used since it generates free radicals<br />

which interfere with the polymerization reaction. Using ethanol<br />

necessitates that excess osmium be thoroughly removed by washing. The<br />

capsules may be polymerized by heating to 50°C overnight but<br />

polyethylene capsule molds should not be used since they are permeable<br />

to oxygen which also interferes with polymerization. The result will be<br />

"tacky" capsules and this may be avoided by using gelatin capsules as a<br />

11


mold. Their only drawback is that the ends are rounded and are more<br />

difficult to trim for sectioning.<br />

The LR-White resin is a general purpose embedding medium that can<br />

be used for light microscopy preparations. Under the electron microscope,<br />

the tissues have a tendency to look washed out or leached and they don't<br />

take up the post-strains (e.g., lead citrate) as well as the epoxy resins<br />

described above. Except for the convenience of not having to mix together<br />

the plastic resin components, there is no overriding advantage apparent<br />

for choosing LR-White. The same number of steps are needed for<br />

adequate infiltration. It is not as irritating or toxic as the epoxides,<br />

however, and this concern may merit its use for general studies or<br />

teaching.<br />

ULTRAMICROTOMY<br />

Now that the messy part is over with, it is time to master the skills of<br />

electron microscopy that require precision and perfection. Although it is<br />

most convenient to hire technicians to do the microtomy and microscopy,<br />

you will not have an adequate appreciation for the results unless it is<br />

learned first hand and it is truly one of the few procedures that are most<br />

easily learned correctly by doing them and making mistakes. This is due<br />

to the large number of variables that affect the quality of the result, i.e.,<br />

the micrograph. The steps needed to master this section include: trimming<br />

the capsule so as to expose the tissue for proper sectioning, making glass<br />

knife edges fitted with a water boat (for sections to float on when cut from<br />

the tissue (capsule) "face", the actual sectioning using the ultramicrotome<br />

and placing the sections on "grids".<br />

TRIMMING THE CAPSULE<br />

This is easier to demonstrate than to explain in written form. Excess<br />

plastic surrounding the tissue must be trimmed away in a fashion that will<br />

yield a square or rectangular section. The capsule mold produced a 1mm 2<br />

face on the tip (see figure). This must be trimmed to a pyramid where the<br />

pyramid tip and sides are exposed tissue. The angle of the pyramid sides<br />

(called facets) should be about 45°. Too steep of an angle will not allow<br />

enough lateral support when sectioning while too flat (or low) of an angle<br />

will cause the "face" being sectioned to enlarge too quickly during<br />

sectioning. The tip of the pyramid may be a point (giving square sections)<br />

or a ridge (giving rectangular sections).<br />

12


TRIMMING<br />

Top view<br />

Side view<br />

Gives rectangular gives square<br />

<strong>Sections</strong><br />

sections<br />

Trim the capsule while viewing under the dissecting microscope using old<br />

glass knives or knives not suitable for sectioning. Use smooth slicing (not<br />

chiseling) strokes that cut through the plastic in one stroke. Take very<br />

<strong>thin</strong> slices so as to leave a smooth side surface (important for good<br />

sectioning). Your instructor will demonstrate.<br />

MAKING GLASS KNIVES<br />

Although most electron microscope laboratories have automatic knife<br />

makers, it is good practice to learn the art of making knives by hand. Not<br />

all types of glass are suitable for knives and despite occasional claims,<br />

hardware store plate glass is rarely adequate. In theory, a semi-liquid<br />

13


knife edge is made by bringing two natural fractures to a 45° apex. The<br />

quality (smoothness, sharpness and durability) of the edge depends upon<br />

the density, temper and composition of the glass. For these reasons, most<br />

labs purchase good quality glass from vendors of EM supplies. The glass<br />

usually comes as one inch wide strips varying in thickness; usually 1/4,<br />

5/16, or 3/8 inches. The strips must be scrupulously cleaned with acetone<br />

or alcohol. The glass can then be scored using a diamond glass scribe<br />

either free hand or by using a simple Plexiglas scoring guide. Two scores<br />

are made, one across (perpendicular to the length) the glass strip to yield a<br />

1" square piece and one diagonal score towards the first score. Both scores<br />

are made at the same time and should be made with enough pressure so as<br />

to just see and "hear" the score. Be sure to align the edge of the diamond<br />

scribe flush with the guide edges to ensure a straight and precisely placed<br />

score.<br />

14


Use the glaziers pliers to first make the cross (perpendicular) fracture and<br />

then the diagonal fracture. Do not touch the knife edge or sides with your<br />

fingers. The contaminants of the fingerprint will prevent adhesion of the<br />

water boat to be mounted on the knife. Examine the knife edge for its<br />

shape and horizontal angle. Note the size of the spur, fracture ridge (burr<br />

line) and curve of the edge (see the diagram).<br />

The boat can be made with short sections of black vinyl electrical tape cut<br />

in half lengthwise. Wrap the tape around the knife (diagonal side) so that<br />

the top edge of the boat is perpendicular with the vertical side of the knife.<br />

Do not leave a gap at the back of the boat and use your fingernail to seal<br />

the adhesive against the glass sides (air bubbles disappear). Seal the back<br />

and sides of the boat with nail polish and allow to dry. The knife may then<br />

be used or stored under cover. It is not a good practice to store knives for<br />

long periods of time (no more than 2-3 days) since they have a tendency<br />

to clutter up the microtome area and to get dull.<br />

THIN SECTIONING<br />

(using the Sorvall Porter-Bloom MT-1 ultramicrotome)<br />

<strong>Ultra</strong>microtomy is one of the most difficult techniques to master<br />

since there are many variables contributing to the cutting process. Some<br />

of these are:<br />

plastic hardness<br />

knife quality<br />

knife angle<br />

boat water level<br />

trimmed edge smoothness<br />

vibration<br />

temperature<br />

humidity<br />

cutting speed<br />

tongue in wrong position<br />

15


A good deal of patience is necessary along with steady hands. First,<br />

read the instruction manual for the ultramicrotome and memorize each<br />

control and component. It is convenient to begin a microtomy session by<br />

resetting the specimen holder arm to the rearmost position. Remove the<br />

knife holder and secure the specimen capsule in the collet holder. Be sure<br />

that the knurled ring securing the ball and socket pivot is tight. This<br />

should be done with the specimen arm hook clamp in place so as to not<br />

damage the lead alloy threaded advancing rod inside the microtome. A<br />

new knife may be secured in the holder and placed in its locking<br />

mechanism. You should not have to move or adjust the cool light source<br />

which may be turned on at the start of the session. Unhook the specimen<br />

arm and rotate the sectioning knob to bring the specimen to knife edge<br />

height. Then manually advance the knife stage to wi<strong>thin</strong> 1-2 mm of the<br />

specimen. Both should now be in view through the dissecting microscope.<br />

Adjust the microscope to the highest magnification and focus. Add fresh,<br />

clean distilled water to the boat so as to have a silver reflection from the<br />

surface. This will occur with the water surface slightly concave and the<br />

water should be adjacent to the knife edge.<br />

Once the knife and specimen are roughly aligned manually the upper<br />

half of the knife stage may be advanced manually using the course and<br />

fine advance controls until cutting the first sections from the tip of the<br />

specimen. With the first piece of plastic section (it will probably be fairly<br />

thick) the knife may then be advanced using the fine advance adjustment<br />

which is calibrated in microns. Advance the fine control one half micron at<br />

a time to create a "face". If done correctly, each time the face comes into<br />

view under the microscope it should appear mirror-like. At this point it is<br />

advisable to back the knife edge away from the specimen using the fine or<br />

coarse adjustment control and to move the knife edge laterally to a new<br />

area. Secure the knife using the locking lever and slowly advance the<br />

knife to the specimen face once more. If the light and water level are<br />

adjusted correctly, as you advance the knife you should eventually see the<br />

reflection of the knife edge in the mirror-like face. The knife edge and its<br />

reflection can be thought of as two lines which should be parallel. If they<br />

are parallel, it will ensure that the first section that is taken by the new<br />

knife edge will not section only part of the face. Although it is difficult to<br />

have the lines exactly parallel one can usually come close by adjusting the<br />

specimen block using the ball socket pivot. It is extremely important that<br />

the first section using the new area of the knife be as <strong>thin</strong> as possible.<br />

Otherwise the knife edge will dull. Most problems arise with beginning<br />

microtomy due to improper water level, too fast a cutting stroke, wrong<br />

knife angle (this should be about 5°) and a dull knife edge. Unfortunately,<br />

16


this technique is one which is learned most rapidly by doing it and making<br />

mistakes. Your instructor will help you understand the problems as they<br />

arise.<br />

If you are successful at getting good sections they will appear gold or<br />

silver. The thickness of the sections is determined by their refractive color<br />

as calibrated on the thickness chart provided with the microtome. Gold<br />

sections are 800-900 A thick, yield better contrast but slightly less<br />

resolution. If maximum resolution is needed, gray sections on the order of<br />

300-400 A can be attempted. This, however, is extremely difficult and<br />

requires a good deal of experience in attaining.<br />

The sections should float on the water and should adhere to one<br />

another in the form of a ribbon. As the knife edge cuts through the plastic,<br />

it causes the sections to become compressed. This may be alleviated by<br />

exposing the sections to vapors of organic solvents such as chloroform or<br />

ether. This is done by merely dipping a long handled cotton swab in the<br />

solvent and holding it close to but not in contact with the sections near the<br />

knife edge while viewing through the microscope. The sections will appear<br />

to become smooth and large and this will be very noticeable to the eye.<br />

The sections are then ready to be maneuvered to the center of the boat to<br />

be picked up on the grids.<br />

PLACING SECTIONS ON GRIDS<br />

If a noticeable amount of water has evaporated from the boat<br />

causing a more concave surface, it is advisable to add a small amount of<br />

water to the boat so the surface is nearly level before moving the sections.<br />

The sections may be maneuvered using a fine needle (a 0000 stainless<br />

steel insect mounting pin slightly bent at the tip pressed onto a wooden<br />

handle is convenient.) After centering the sections they may be adhered to<br />

the proper grid by grasping the grid in fine tweezers (Pick up the grid only<br />

by the very edge) and bending the grid against the bottom of the petri<br />

dish so that it may be placed dull side down on the surface of the water<br />

and sections. Be careful as you touch down on the surface of the sections<br />

that they are oriented in the middle of the grid. Do not push hard enough<br />

that you break the surface tension of the water. You need only to barely<br />

touch the surface and the sections along with a small drop of water will<br />

adhere to the grid. If the water droplet does not spread out evenly on the<br />

grid it signifies that the grids are dirty. If this is the case you will usually<br />

find that the sections have become wrinkled and do not span over the<br />

holes. If this is the case the grids must be cleaned (see appendix). The<br />

sections on the grids are now ready to be stained.<br />

17


STAINING<br />

Additional contrast of the EM image can be gained by staining the<br />

tissue sections with heavy metals. The strains most commonly used are<br />

uranyl acetate and lead citrate. Both metals apparently bind at sites of<br />

osmium deposition and lead also binds with (i.e. stains) nucleic acids and<br />

glycogen.<br />

Lead Citrate<br />

Lead citrate (Reynold's) is perhaps the best stain available since it<br />

can be used at a high pH and stains a wide variety of cellular components<br />

including nuclear components, ribosomes, membranes, microfilaments and<br />

glycogen. The precise chemical nature of the binding is not well<br />

understood. Care must be taken since lead citrate will react with<br />

atmospheric CO 2 to form a fine precipitate of lead carbonate.<br />

To Stain <strong>Sections</strong>:<br />

Pour a generous amount of either sodium or potassium hydroxide<br />

pellets around the perimeter of a plastic disposable petri dish. Place the<br />

cover on the dish. Carefully place (don't drop) a 2-3 mm drop of lead<br />

citrate (one for each grid to be stained) on the center surface using a<br />

Pasteur pipet and lifting the plate cover just enough to give the pipet<br />

clearance. The drops will not spread but will remain as droplets. Place the<br />

grids, with section side down, on top of the droplets and cover the plate.<br />

The grids will float. Do not breath onto the petri dish while placing grids<br />

on the drops of stain. Stain for 15-20 minutes. Remove the grids and<br />

immediately but gently dip and stir them in a weak NaOH solution (or KOH<br />

- one or two pellets in 30-40 ml dHOH in a small beaker). Only a few<br />

seconds are needed. Rinse the grids in distilled water by gently dipping<br />

and stirring; blot dry by pressing the surface of the grid not having<br />

sections on a piece of filter paper. This will "wick" most of the water away.<br />

Also blot (wick) away the water between the tweezer tips so the grid can<br />

be placed in a holder without being wetted by a fountain pen-like action.<br />

The grid will air dry quickly and is ready for viewing under the EM.<br />

Throw the petri dish away after taping it shut.<br />

18


Uranyl Acetate<br />

Uranyl acetate may be used during the dehydration process by<br />

making the 50% acetone up to 2% with the stain. Uranyl acetate is not<br />

soluble in pure acetone. An aqueous solution of 2% concentration can be<br />

used to float or dip sections mounted on grids. Epon and Araldite do not<br />

take up aqueous stains well unless they are alkaline. Organic solvent<br />

solutions will usually leave a fine precipitate on the sections and is not<br />

often used. Rinse the solutions well with distilled water. Caution - uranyl<br />

acetate is radioactive. Do not pipette by mouth or spill.<br />

Phosphotungstic acid<br />

May be used as a "negative" stain in that it does not bind particularly<br />

well to any<strong>thin</strong>g but instead caused areas other than cellular (organic)<br />

material to appear dark. It is especially useful for viewing molecules such<br />

as proteins (e.g., antibody, DNA) and suspensions of subcellular structures<br />

such as membranes (e.g., mitochondria - elementary particles, etc.) Usually<br />

a 1-2% aqueous solution is used to stain a tissue block during dehydration<br />

for about 30 minutes.<br />

OTHER PROCEDURES<br />

Thick <strong>Sections</strong>:<br />

It is generally not advisable to begin a structure oriented study at<br />

the electron microscope level. Some form of light microscopy is usually<br />

performed to become familiar with the general structure, orientation and<br />

on occasion, to locate a specific site wi<strong>thin</strong> the specimen. Specimens<br />

prepared for the electron microscope can be thick sectioned and<br />

appropriately stained for light microscopic examination with ease.<br />

The trimmed specimen capsule is mounted in the ultramicrotome the<br />

same as for <strong>thin</strong> sectioning. A knife is also secured and may or may not<br />

have a boat. The "face" is formed in the usual manner and thick sections<br />

(2-3 µ) are made using the fine advance adjustment control. The sections<br />

often have a tendency to roll up like a scroll. This can be prevented by<br />

using a new knife edge and a slow smooth cutting stroke. The sections are<br />

then placed on a clean microscope slide with enough water present to<br />

permit positioning of the sections in an orderly arrangement. The slide is<br />

dried, lightly heat fixed and stained for 5 minutes with Toluidine blue.<br />

Excess dye is rinsed off with distilled water and the sections are then<br />

destained (excess stain in the sections) with ethanol (90-95%) for 2<br />

19


minutes. The slide is dried and a cover slip is secured with a small drop of<br />

the pure plastic mixture used for the embedding procedure. A vial of this<br />

mixture (with DMP-30 added) may be kept for several months in the<br />

freezer (-20°C). The section may be viewed to see if they are of<br />

appropriate quality and can be labeled and stored after curing the plastic<br />

mixture (used for mounting the cover slip) on a warming plate set at 60°C.<br />

Place a weight on the cover slip to insure that it will press the sections flat<br />

against the glass slide.<br />

Whole mounts:<br />

It is often necessary to view a whole or solid object such as a<br />

bacteria, virus, or molecules. For this, it is routine to place the specimen on<br />

a film-coated grid and to coat the object with a <strong>thin</strong> film of metal. The<br />

grids are usually coated with a Formvar film of varying thickness. A 0.25%<br />

solution in anhydrous ethylene dichloride is kept in a tightly sealed<br />

volumetric. Water will cause holes in the film. A small drop of the<br />

solution is gently laid onto the convex (mounded) surface of distilled water<br />

in a petri dish. The Formvar will spread into a very <strong>thin</strong> film and the<br />

solvent will evaporate almost immediately. Carefully place clean grids<br />

onto the film surface dull side down using tweezers. The film is thicker at<br />

the center of the dish. Do not allow vibrations while making the film. The<br />

coated grids can be recovered from the plate surface by carefully dropping<br />

a piece of filter paper onto (over) the surface allowing it to wet completely.<br />

Then remove the paper, invert it (grids face up) and place it on absorbent<br />

towels to blot excess water away. After the paper has dried, the grids may<br />

be recovered by raising them straight up off of the paper using sharp<br />

tweezers. The grids may then be used or stored.<br />

Specimens are usually applied to the coated grids by aspiration or by<br />

placing a small drop of water suspension on the surface and allowing it to<br />

dry.<br />

20


Arts and Formulae<br />

PBS -phosphate buffered saline, use either potassium or sodium phosphate.<br />

(9.0 g NaC1, 2.7 g KH 2 PO 4 , d HOH to 1 L, adjust pH)<br />

Formvar resin solution for coating EM grids<br />

0.25% dissolved in ethylene dichloride (0.157 g in 50ml = 0.25%)<br />

Glutaraldehyde<br />

23 ml PBS<br />

2 ml 25% Glut.<br />

OsO 425<br />

ml PBS, 0.5 g OsO4 crystals<br />

CAUTION<br />

Lead Citrate Post Stain:<br />

1.33g Lead nitrate (Pb(NO3)2)<br />

1.76g Sodium citrate (Na 3 (C 6 H 5 O 7 )2H 2 O<br />

30 ml dHOH in 50ml volumetric<br />

Shake vigorously for 2 min. then at intervals for 1/2 hr.<br />

8 ml of 1 N NaOH_ (0.4g/10ml dHOH-fresh)<br />

Dilute to 50 ml w dHOH clear, pH=12<br />

KMnO 4<br />

- 1 g in 50 ml PBS<br />

Plastic Mixture:<br />

16 ml of a 5:3 Epon:Araldite mixture<br />

9 ml NMA<br />

10 ml DDSA<br />

35 ml total - stir exhaustively<br />

15-17 drops DMP-30 - stir again, will turn dark amber.<br />

Toluidine blue for thick sections<br />

1 g borax (sodium carbonate)<br />

100 ml HOH-dissolve<br />

1 g Toluidine blue dye<br />

Stir - filter if necessary<br />

21


Cleaning copper grids<br />

Immerse grids in a 2-4 N HC1 solution for 5 minutes (a drop of detergent<br />

will help the grids sink)<br />

Rinse with distilled water and then acetone in a filter funnel attached to a<br />

vacuum flask. Use filter paper in the funnel and allow the grids to air dry.<br />

Place grids in a covered glass (paper bottom) petri dish (plastic allows<br />

static charge to occur).<br />

Specimen <strong>Preparation</strong> with L. R. White<br />

I. Fixation<br />

A. Dissect tissue into 1mm 3 cubes shortly after procurement.<br />

B. Place cubes in Phosphate Buffered Saline (PBS) for a 10 minute<br />

wash.<br />

C. Transfer to a solution of 2% glutaraldehyde in Phosphate<br />

Buffered Saline. (60 min.)<br />

D. Wash in PBS for 10 minutes.<br />

E. Transfer blocks of tissue to an osmium tetroxide solution;<br />

consisting of the following: (60-90 minutes)<br />

1. 5 ml. of PBS<br />

2. 5 ml. of 0.1 N HC1<br />

3. 15 ml. of distilled HOH<br />

4. 0.5g OsO 4 crystals<br />

II.<br />

Dehydration<br />

A. Transfer blocks to 50% ethanol (15 min.)<br />

B. Transfer blocks to 75% ethanol (15 min.)<br />

C. Transfer blocks to 95% ethanol (15 min.)<br />

D. Transfer blocks to 100% ethanol (Twice) (15-20 min./switch)<br />

III. Infiltration<br />

A. Transfer blocks to L.R. White resin (45 min.) four (4) times in<br />

oven at 60° C.<br />

B. Place blocks in gelatin capsules (1/cap.) and fill to rim with L.<br />

R. White resin.<br />

IV.<br />

Polymerization<br />

A. Polymerize 20-24 hours at 60°C.<br />

B. Remove the capsule when fully polymerized.<br />

22


Photography<br />

Making negatives:<br />

The films should be exposed according to the instructions in the EM<br />

operating manual. Generally speaking, contrast is enhanced by longer<br />

exposures with low illumination. However, exposures longer than 1 second<br />

are more prone to blurring from vibration in the building, thus negating<br />

the fine resolution of the EM. It is desirable to have as much contrast in<br />

the recorded image as possible. It is often advantageous to slightly over<br />

expose the film so as to have a "dense" negative. In this case, you would<br />

adjust the illumination (with the condenser lens control) to indicate just<br />

less than a 1 second exposure on the exposure meter and expose (timer<br />

set) for 1 second.<br />

Film development:<br />

The 35mm film is developed as follows:<br />

Dektol (D-72 straight) = 2 minutes<br />

Water (tepid)= 2 changes<br />

Hypo (Fixer)= 5 minutes (save hypo)<br />

Water wash= 5 minutes<br />

Distilled water rinse - dry<br />

Plate film is developed in trays or tanks as follows:<br />

D-19 (2:1 water:D-19)= 4 minutes<br />

Water wash (stop bath)= 10 seconds<br />

Hypo (Fixer - save)= 8 minutes<br />

Water wash= 5 minutes<br />

Printing - Black and White<br />

Additional contrast can be gained during the printing process by<br />

using high contrast paper or polycontrast paper and high contrast filters.<br />

Also, a certain amount of contrast can be gained by again using a longer<br />

exposure with low illumination.<br />

The exposure and illumination settings can be determined by using<br />

"test strips" or an exposure guide on a test print. This consumes a sizeable<br />

quantity however, and it is best to acquire a "feel" for proper exposures by<br />

practice. Most print papers have a generously wide exposure latitude but<br />

the exposure should be such so as to allow the image to develop wi<strong>thin</strong> 1-<br />

23


1.5 minutes in developer. Among the most often used print papers is<br />

Kodak Kodabromide polycontrast RCII paper in E or F surface. RC stands<br />

for "resin coated" which eliminates the need for a print dryer and it<br />

shortens processing as well. The E and F surfaces differ in that the E<br />

surface is a "matte" surface and the F surface is more "glossy".<br />

Paper development is as follows:<br />

Dektol (D-72, 2:1, water:D-72)= 1-1.5 min.<br />

Stop bath (indicator)= 10-15 sec.<br />

Hypo (can be used)= 3-5 min.<br />

Water wash= 5 min.<br />

Squeegee and air dry<br />

The instructor will demonstrate the use of the photographic enlarger.<br />

It is fairly straight-forward. Take care to remove dust from the negative<br />

and lenses. Always leave the darkroom clean. Wipe up hypo spills since<br />

dry hypo is a fine white powder and can easily contaminate surfaces of<br />

negatives and lenses.<br />

Rinse hands well before handling paper or touching the enlarger or<br />

light switch. It is alright to transfer paper and hands from developer to<br />

stop bath to hypo but it is not good practice to go the other way. Rinse<br />

hands thoroughly of hypo before going back to the developer or stop bath.<br />

Prints may be conveniently labeled using transfer letters and<br />

numbers available in most office supply stores. These transfer markings<br />

are virtually permanent but can be removed by gently scraping with a<br />

scalpel or razor blade.<br />

Magnification (recorded when the film was exposed) can be<br />

multiplied by the magnification factor of the enlarger and represented by<br />

a "micron" or other unit measure on the print as a line or bar.<br />

Prints should be stored out of direct sunlight in a dry environment.<br />

There may be a tendency for the prints to curl or roll up due to absorption<br />

of moisture by the paper backing while the emulsion remains relatively<br />

moisture free (since it is fairly <strong>thin</strong>). The prints may be mounted on<br />

various types of poster board by glue adhesives or by heat sensitive "dry<br />

mounting" sheets (essentially a wax that melts and adheres to the print<br />

and mount). The print image is virtually permanent if stored or cared for<br />

properly.<br />

24

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!