D3Insert remaining screwsRepeat steps D1 and D2 for the remaining screws.D4Tighten screwsInstruments03.617.900 Screwdriver Stardrive, T8, self-holding,angled, with Sleeve03.110.002 Torque Limiter, 1.2 Nm, with AO/ASIFQuick Coupling03.617.902 Screwdriver Shaft Stardrive, T8,self-holding03.617.903 Handle with Quick CouplingTighten each screw to at least 1.2 Nm torque, to fully seatand lock the screw head in the plate.Note: It may be necessary (e. g. screw placed off angle) toapply a higher force than 1.2 Nm to fully seat the screw intothe plate. In such cases the screwdriver should be used withoutthe torque limiter.Synthes 25