<strong>APPROVAL</strong> <strong>REPORT</strong>Project No: 3040602Class: 3615Product Name:Product Type:Name of Report Holder:Address of Report Holder:Name of Manufacturer:Address of Manufacturer:640i and 780i Insertion and In-line Thermal MassFlowmetersInsertion and In-line Thermal Mass Flowmeters<strong>Sierra</strong> <strong>Instruments</strong> Inc.5 Harris Court, Building LMonterey, CA 93940USA<strong>Sierra</strong> <strong>Instruments</strong> Inc.5 Harris Court, Building LMonterey, CA 93940USACustomer ID: 1000002381Customer website:www.sierrainstruments.comPrepared by:Reviewed by:Marlon MitchellAdvanced EngineerAndrew LozinskiTechnical Team ManagerJames E. MarquedantGroup Manager28 December 2012Date of ApprovalFM Approvals1151 Boston-Providence TurnpikePO Box 9102Norwood, MA 02062Page 1 of 5


FM <strong>APPROVAL</strong>S PROJECT NO: 30406021 INTRODUCTION1.1 <strong>Sierra</strong> <strong>Instruments</strong> Inc. (manufacturer) requested FM Approval of their 640i and 780iInsertion and In-line Thermal Mass Flowmeters listed in Section 1.4 to be in compliancewith the applicable requirements of the standards listed in Section 1.3 as Explosionprooffor Class I, Division 1, Groups B, C and D hazardous (classified) locations, indoor andoutdoors (Type 4X).1.2 This report may be freely reproduced only in its entirety and without modification.1.3 Standards1.4 ListingTitle Number IssueElectrical Equipment for Use in Hazardous (Classified) FM Class 3600 2011Locations General Requirements.Explosionproof Electrical Equipment. FM Class 3615 2006Electrical Equipment for Measurement, Control and FM Class 3810 2005Laboratory UseEnclosures for Electrical Equipment (1000 VoltsMaximum)ANSI/NEMA 250 1991The product will appear in the Approval Guide, an on-line resource of FM Approvals, asfollows:Hazardous Location Equipment - United States Flowmeters Mass640i-a-FM-b-c-d-e-f-g-h-0-i-i-i. Insertion Thermal Mass Flowmeter.XP/I/1/BCD/T3C Ta = -40°C to +60°C; Type 4Xa = Multivariable VT or VTP.b = Probe L06, L09, L13, L18, L24, L36, L48, L(*) or L(*) M5. *Specify probe length (max48 inches).c = Mounting M0, M1, M1-M2(*), M3, M4(*), M8(*), M9 or M15(*). *Specify pipe O.D.d = Enclosure E2 or E4(*). *Specify cable length.e = Power P2 or P3.f = Output V4 or V6*. *VTP only.g = Display DD or NR.h = Pressure MP1*, MP2*, MP3*, MP4* or N/A † . *VTP only. † For VT or E4.i = Gas 0, 1, 1A, 2, 2A, 3, 4, 4A, 6, 6A, 7, 7A, 8, 8A, 10, 10A, 11, 12, 12A or 99.780i-a-FM-b-c-d-e-f-g-0-h-h-h. In-line Thermal Mass Flowmeter.XP/I/1/BCD/T3C Ta = -40°C to +60°C; Type 4Xa = Multivariable VT or VTP.b = Flow Body N2, N3, N4, N5, N6, N7, N8, N9, N10, F2, F3, F4, F5, F6, F7, F8, F9,F10, FD4, FD5, FD6, FD7, FD8, FD9 or FD10.c = Enclosure E2 or E4(*). *Specify cable length.d = Power P2 or P3.e = Output V4 or V6*. *VTP only.f = Display DD or NR.g = Pressure MP1*, MP2*, MP3*, MP4* or N/A † . *VTP only. † For VT or E4.h = Gas 0, 1, 1A, 2, 2A, 3, 4, 4A, 6, 6A, 7, 7A, 8, 8A, 10, 10A, 11, 12, 12A or 99.Page 2 of 5


FM <strong>APPROVAL</strong>S PROJECT NO: 30406022 DESCRIPTIONThe 640i and 780i Insertion and In-line Thermal Mass Flowmeters are used for gas flowmeasurement applications. The flowmeters integrate configuration, validation anddiagnostics in a probe-mounted or remotely-mounted housing. When remotelymounted,the electronics enclosure can be mounted up to 200 feet away from theremote probe enclosure. All variables can be monitored on an optional, viewabledisplay. The flowmeters can also provide 4-20 mA flow and temperature output, alarmoutput and an optional pressure output. The 640i is suitable for insertion into pipes orducts from three inches up to 72 inches while the 780i can go in-line to 8-inch flowbodies. For probes longer than 13 inches, a seal is required within 13 inches of theprobe and enclosure.The 640i and 780i Insertion and In-line Thermal Mass Flowmeters operate on 24 Vdc or100-240 Vac. The flowmeters are intended for use in an ambient temperature of -40°Cto +60°C and with process temperatures of -40°C to +200°C at a pressure of 300 psi (inline)or 500 psi (insertion).3 EXAMINATIONS AND TESTSSamples of the 640i and 780i Insertion and In-line Mass Thermal Flowmeters weresubmitted for examination and testing. The samples were considered to berepresentative of the product line and were examined, tested, and compared to themanufacturer's drawings. All data is on file at FM Approvals along with other documentsand correspondence applicable to this program.All testing and analysis considered appropriate was conducted and verified to be incompliance with the Standards defined in Section 1.3.4 MARKINGThe manufacturer’s metallic marking plates, as depicted on drawings 47-0444 and 47-0481, are shown below.Page 3 of 5


FM <strong>APPROVAL</strong>S PROJECT NO: 30406025 REMARKSInstallation shall be in accordance with the manufacturer's instructions and the NationalElectrical Code ® , ANSI/NFPA 70, as applicable.6 SURVEILLANCE AUDITThe design and manufacturing facilities at the following location is visited on a routinebasis. The facilities and quality control procedures in place have been found to besatisfactory to manufacture product identical to that tested and Approved. A Form 797shall be submitted to FM Approvals for requesting any additional manufacturing facilitieswhich are not listed below.Page 4 of 5


FM <strong>APPROVAL</strong>S PROJECT NO: 3040602Design & Manufacturing<strong>Sierra</strong> <strong>Instruments</strong> Inc.5 Harris Court, Building LMonterey, CA 93940USA7 MANUFACTURER’S RESPONSIBILITIES7.1 Documentation that is applicable to this approval is on file at FM Approvals and listed inthe Documentation File, Section 8, of this report. No changes of any nature shall bemade unless notice of the proposed change has been given and written authorizationobtained from FM Approvals. The Approved Product - Revision Report, FM ApprovalsForm 797, shall be forwarded to FM Approvals as notice of proposed changes.7.2 On 100% of production, the manufacturer shall inspect for the presence of the protectivegrounding terminals.7.3 On 100% of production, the AC version of the 640i and 780i Insertion and In-line MassThermal Flowmeters shall be dielectric tested. The input connections shall withstand,with no insulation breakdown, the application of 2000 Vdc, raised within 2 seconds andmaintained for 2 seconds, with respect to the protective ground conductor. Alternatively,test potentials 20% higher may be applied for at least one second. Routine dielectrictesting of the DC versions of the 640i and 780i Insertion and In-line Mass ThermalFlowmeters is waived as the voltage at which the equipment operates (24 Vdc) is belowthe minimum voltage considered hazardous live for wet locations (35 Vdc).WARNING: The dielectric test required may present a hazard of injury to personneland/or property and should only be performed under controlled conditions and bypersons knowledgeable of the potential hazards of such testing to minimize thelikelihood of shock and/or fire.8 DOCUMENTATIONSee attached blueprint report.9 CONCLUSIONThe equipment described in section 1.4 meets FM Approvals requirements. Since aduly signed Master Agreement is on file for this manufacturer, Approval is effective thedate of this report.PROJECT DATA RECORD: 3040602ATTACHMENTS: Blueprint Report - 3040602Page 5 of 5


Blueprint Report<strong>Sierra</strong> <strong>Instruments</strong> Inc (1000002381)Class No 3615Original Project I.D. 3040602Drawing No. Revision Level Drawing Title Last Report Electronic Drawing41-2203-XX-XX 8 SHAFT, INSERTION METERS 3040602 Yes (pdf)41-2205 8 PRESSURE SENSOR FITTING 3040602 Yes (pdf)41-2222-XX 4 SHAFT, IN-LINE METERS 3040602 Yes (pdf)41-4457 1 BUSHING, ENCLOSURE, 640i 3040602 Yes (pdf)41-4458 2 RETAINING RING, FLAME ARRESTOR 3040602 Yes (pdf)41-4463-X 9 ENCLOSURE, 640i 3040602 Yes (pdf)41-4487 2 DISK, FLAME ARRESTOR, 640i 3040602 Yes (pdf)41-4530 4 ADAPTER, BUSHING, 640I/780I REMOTE 3040602 Yes (pdf)41-4638 4 ENCLOSURE, REMOTE, 640i 3040602 Yes (pdf)43-0244-XX 12 WELDMENT, .5" AND .75" NPT, INLINE FLOWBODY 3040602 Yes (pdf)43-0247-XX-XX 6 PROBE WELDMENT AND ASSY, .75 O.D. INSERTION METERS 3040602 Yes (pdf)43-0249-XX 6 WELDMENT, SENSOR HOUSING, .75 O.D. IN-LINE METERS 3040602 Yes (pdf)43-0250-XX 5 WELDMENT, SENSOR HOUSING, .75 O.D. INSERTION METERS, 640i VTP 3040602 Yes (pdf)43-0256-XX 3 WELDMENT, .25 IN-LINE FLOW BODY 3040602 Yes (pdf)43-0268 1 ASSEMBLY, FLAME ARRESTOR, 640i 3040602 Yes (pdf)43-0274-X 3 SENSOR ASSEMBLY, 640i FR 3040602 Yes (pdf)43-0296-XX-XX 2 PROBE WELDMENT & ASSY. .75OD, W/FLANGE INSERTION METERS, 640i 3040602 Yes (pdf)47-0444 B LABEL, AGENCY <strong>APPROVAL</strong>, FM, ENCL BACK, 640i/780i 3040602 Yes (pdf)47-0481 A LABEL, RMT ENCL, 640i FM 3040602 Yes (pdf)52-0198 5 640I Terminal PCA 3040602 Yes (pdf)52-0199 4 640I Power PCA 3040602 Yes (pdf)52-0201 6 640I Display RS232 PCA 3040602 Yes (pdf)52-0202 8 640 I Main PCA 3040602 Yes (pdf)52-0209 1 640I Remote J-Box PCA 3040602 Yes (pdf)54-0372-XX 14 WIRING, SENSOR HOUSING, 640I/780I 3040602 Yes (pdf)99-1636 2 640i/780i ENCLOSURE USAGE 3040602 Yes (pdf)99-1685 3 640i/780i <strong>APPROVAL</strong> AGENCY OVER-VIEW 3040602 Yes (pdf)IM-640i/780i v3 QuadraTherm 640i/780i Series Insertion and In-Line Mass Flow Meters Instruction Manual 3040602 Yes (pdf)12/31/2012 Page 1 of 1