Workload Survey• Found total W to be far different from the 1000mA·min/wk blindly assumed– Radiographic Rooms: 277 mA·min/wk– Chest Rooms: 45 mA·min/wk– Cardiac Angio Rooms: 3050 mA·min/wk• Found kVp distribution of workloads to be atpotentials significantly below the single kVpoperating value usually assumed31