REFERENCE MEASURING INSTRUMENT
REFERENCE MEASURING INSTRUMENT
REFERENCE MEASURING INSTRUMENT
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
<strong>REFERENCE</strong> <strong>MEASURING</strong> <strong>INSTRUMENT</strong><br />
Technical data<br />
Probe type<br />
Meas. range<br />
Accuracy<br />
± 1 digit<br />
Resolution<br />
Probe type NTC<br />
Pt100<br />
Meas. range -40... +150 °C<br />
-200... +800 °C<br />
Accuracy ±0.2 °C (-10... +50 °C)<br />
±0.1 °C (-49.9... +99.9 °C)<br />
± 1 digit<br />
±0.4 °C (-40... -10.1 °C)<br />
±0.4 °C (-99.9... -50 °C)<br />
±0.4 °C (+50.1... +150 °C) ±0.4 °C (+100... +199.9 °C)<br />
±1 °C (-200... -100 °C)<br />
±1 °C (+200... +800 °C)<br />
Resolution 0.1 °C (-40... +150 °C)<br />
0.01 °C (-99.9... +300 °C)<br />
0.1 °C (-200... -100 °C)<br />
0.1 °C (+300.1... +800 °C)<br />
Probe type<br />
Meas. range<br />
Accuracy<br />
± 1 digit<br />
Resolution<br />
PC RS232 interface<br />
Oper. temp. 0... +50 °C<br />
Storage temp. -25... +60 °C<br />
Battery type 1.5 V AA<br />
Battery life 18 h<br />
Vane<br />
0... +60 m/s<br />
CO2 probe<br />
Weight 500 g<br />
Warranty 3 years<br />
See Probes data for system<br />
accuracy<br />
0.01 m/s (for Ø 60/100 mm),<br />
0.1 m/s (for remaining<br />
probes)<br />
Probe type Testo humid. sensor, cap.<br />
Meas. range 0... +100 %RH<br />
Accuracy See probe data<br />
± 1 digit<br />
Resolution 0.1 %RH (0... +100 %RH)<br />
0... +1 Vol. % CO2<br />
0... +10000 ppm CO2<br />
See probe data<br />
Thermal<br />
0... +20 m/s<br />
±0.01 m/s (0... +1.99 m/s)<br />
±0.02 m/s (+2... +4.9 m/s)<br />
±0.04 m/s (+5... +20 m/s)<br />
0.01 m/s (0... +20 m/s)<br />
Pressure<br />
0... +2000 hPa<br />
Probe 0638 1345<br />
Probe 0638 1445<br />
Probe 0638 1545<br />
Probe 0638 1645<br />
±0.1% of mv<br />
Probe 0638 1740<br />
Probe 0638 1840<br />
Probe 0638 1940<br />
±0.2% of mv<br />
0.001 hPa (0638 1345 probe)<br />
0.001 hPa (0638 1445 probe)<br />
0.01 hPa (0638 1545 probe)<br />
1 hPa (0638 1645 probe)<br />
0.01 bar (0638 1740 probe)<br />
0.01 bar (0638 1840 probe)<br />
0.01 bar (0638 1940 probe)<br />
CO probe<br />
0... +500 ppm CO<br />
±5% of mv (0... +500 ppm CO)<br />
Memory space in basic version: 128 kB<br />
corresponds to approx. 45,000 readings<br />
Memory space, extended: 1 MB, corresponds<br />
to approx. 500,000 readings<br />
Other features: automatic recognition of all<br />
connected probes<br />
Power supply: Battery/rech. batt., alternatively<br />
8 V mains unit<br />
Battery life in continuous operation with 2<br />
thermocouple probes<br />
aw value<br />
0... +1 aW<br />
See probe data<br />
Type K (NiCr-Ni)<br />
-200... +1370 °C<br />
±0.4 °C (-100... +200 °C)<br />
±1 °C (-200... -100.1 °C)<br />
±1 °C (+200.1... +1370 °C)<br />
0.1 °C (-200... +1370 °C)<br />
Mechanical<br />
+20... +20000 rpm<br />
(+20... +20000 rpm)<br />
1 rpm (+20... +20000 rpm)<br />
Type S (Pt10Rh-Pt)<br />
0... +1760 °C<br />
±1 °C (0... +1760 °C)<br />
1 °C (0... +1760 °C)<br />
Current/voltage measurement<br />
0... +20 mA<br />
±0.04 mA (0... +20 mA)<br />
0.01 mA (0... +20 mA)<br />
Type J (Fe-CuNi)<br />
-200... +1000 °C<br />
±0.4 °C (-150... +150 °C)<br />
±1 °C (-200... -150.1 °C)<br />
±1 °C (+150.1... +1000 °C)<br />
0.1 °C (-200... +1000 °C)<br />
Current/voltage measurement<br />
0... +10 V<br />
±0.01 V (0... +10 V)<br />
0.01 V (0... +10 V)<br />
159<br />
°C<br />
%RH<br />
td<br />
g/kg<br />
hPa<br />
m/s<br />
m³/h<br />
ppm<br />
CO<br />
ppm<br />
CO2<br />
rpm<br />
mA<br />
V