27.11.2012 Views

REFERENCE MEASURING INSTRUMENT

REFERENCE MEASURING INSTRUMENT

REFERENCE MEASURING INSTRUMENT

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

<strong>REFERENCE</strong> <strong>MEASURING</strong> <strong>INSTRUMENT</strong><br />

Technical data<br />

Probe type<br />

Meas. range<br />

Accuracy<br />

± 1 digit<br />

Resolution<br />

Probe type NTC<br />

Pt100<br />

Meas. range -40... +150 °C<br />

-200... +800 °C<br />

Accuracy ±0.2 °C (-10... +50 °C)<br />

±0.1 °C (-49.9... +99.9 °C)<br />

± 1 digit<br />

±0.4 °C (-40... -10.1 °C)<br />

±0.4 °C (-99.9... -50 °C)<br />

±0.4 °C (+50.1... +150 °C) ±0.4 °C (+100... +199.9 °C)<br />

±1 °C (-200... -100 °C)<br />

±1 °C (+200... +800 °C)<br />

Resolution 0.1 °C (-40... +150 °C)<br />

0.01 °C (-99.9... +300 °C)<br />

0.1 °C (-200... -100 °C)<br />

0.1 °C (+300.1... +800 °C)<br />

Probe type<br />

Meas. range<br />

Accuracy<br />

± 1 digit<br />

Resolution<br />

PC RS232 interface<br />

Oper. temp. 0... +50 °C<br />

Storage temp. -25... +60 °C<br />

Battery type 1.5 V AA<br />

Battery life 18 h<br />

Vane<br />

0... +60 m/s<br />

CO2 probe<br />

Weight 500 g<br />

Warranty 3 years<br />

See Probes data for system<br />

accuracy<br />

0.01 m/s (for Ø 60/100 mm),<br />

0.1 m/s (for remaining<br />

probes)<br />

Probe type Testo humid. sensor, cap.<br />

Meas. range 0... +100 %RH<br />

Accuracy See probe data<br />

± 1 digit<br />

Resolution 0.1 %RH (0... +100 %RH)<br />

0... +1 Vol. % CO2<br />

0... +10000 ppm CO2<br />

See probe data<br />

Thermal<br />

0... +20 m/s<br />

±0.01 m/s (0... +1.99 m/s)<br />

±0.02 m/s (+2... +4.9 m/s)<br />

±0.04 m/s (+5... +20 m/s)<br />

0.01 m/s (0... +20 m/s)<br />

Pressure<br />

0... +2000 hPa<br />

Probe 0638 1345<br />

Probe 0638 1445<br />

Probe 0638 1545<br />

Probe 0638 1645<br />

±0.1% of mv<br />

Probe 0638 1740<br />

Probe 0638 1840<br />

Probe 0638 1940<br />

±0.2% of mv<br />

0.001 hPa (0638 1345 probe)<br />

0.001 hPa (0638 1445 probe)<br />

0.01 hPa (0638 1545 probe)<br />

1 hPa (0638 1645 probe)<br />

0.01 bar (0638 1740 probe)<br />

0.01 bar (0638 1840 probe)<br />

0.01 bar (0638 1940 probe)<br />

CO probe<br />

0... +500 ppm CO<br />

±5% of mv (0... +500 ppm CO)<br />

Memory space in basic version: 128 kB<br />

corresponds to approx. 45,000 readings<br />

Memory space, extended: 1 MB, corresponds<br />

to approx. 500,000 readings<br />

Other features: automatic recognition of all<br />

connected probes<br />

Power supply: Battery/rech. batt., alternatively<br />

8 V mains unit<br />

Battery life in continuous operation with 2<br />

thermocouple probes<br />

aw value<br />

0... +1 aW<br />

See probe data<br />

Type K (NiCr-Ni)<br />

-200... +1370 °C<br />

±0.4 °C (-100... +200 °C)<br />

±1 °C (-200... -100.1 °C)<br />

±1 °C (+200.1... +1370 °C)<br />

0.1 °C (-200... +1370 °C)<br />

Mechanical<br />

+20... +20000 rpm<br />

(+20... +20000 rpm)<br />

1 rpm (+20... +20000 rpm)<br />

Type S (Pt10Rh-Pt)<br />

0... +1760 °C<br />

±1 °C (0... +1760 °C)<br />

1 °C (0... +1760 °C)<br />

Current/voltage measurement<br />

0... +20 mA<br />

±0.04 mA (0... +20 mA)<br />

0.01 mA (0... +20 mA)<br />

Type J (Fe-CuNi)<br />

-200... +1000 °C<br />

±0.4 °C (-150... +150 °C)<br />

±1 °C (-200... -150.1 °C)<br />

±1 °C (+150.1... +1000 °C)<br />

0.1 °C (-200... +1000 °C)<br />

Current/voltage measurement<br />

0... +10 V<br />

±0.01 V (0... +10 V)<br />

0.01 V (0... +10 V)<br />

159<br />

°C<br />

%RH<br />

td<br />

g/kg<br />

hPa<br />

m/s<br />

m³/h<br />

ppm<br />

CO<br />

ppm<br />

CO2<br />

rpm<br />

mA<br />

V