REFERENCE MEASURING INSTRUMENT
REFERENCE MEASURING INSTRUMENT
REFERENCE MEASURING INSTRUMENT
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
140<br />
°C<br />
%RH<br />
td<br />
g/kg<br />
hPa<br />
m/s<br />
m³/h<br />
ppm<br />
CO<br />
ppm<br />
CO2<br />
rpm<br />
mA<br />
V<br />
The right probe for every application<br />
<strong>REFERENCE</strong> <strong>MEASURING</strong> <strong>INSTRUMENT</strong><br />
Highly accurate immersion/penetration probes<br />
with a system accuracy of 0.05 °C in the measuring range from<br />
0 to 100 °C and a resolution of up to 0.001 °C<br />
Quick-action surface probes<br />
to measure surface temperature<br />
Precision air probes<br />
to measure air temperature<br />
Globe thermometer<br />
to measure radiant heat<br />
Current/voltage cable (± 1 V, ± 10 V, 20 mA)<br />
for example, to check stationary transmitters<br />
CO2 probes<br />
to determine Indoor Air Quality and monitor the workplace<br />
Air humidity probes<br />
Measurements in room/ducts with (precision) humidity probe<br />
Pressure dew point probes<br />
to measure pressure dew point to - 60 °C tpd in compressed air systems<br />
Robust humidity probes<br />
For equilibrium moisture or duct measurements up to 180 °C<br />
Equilibrium probes<br />
To determine equilibrium moisture<br />
Differential (100 Pa / 10 hPa / 100 hPa) and absolute pressure probes<br />
For pressure and velocity measurement<br />
Refrigerant-proof high pressure probes<br />
For maintenance on refrigeration systems/water measurement<br />
Straight and Prandtl Pitot tubes to measure velocity in dirty air and<br />
temperatures up to 600 °C<br />
12/16/25 mm – Vane probes<br />
For measurements in ducts, with temperature measurement<br />
Vane probes 60/100 mm<br />
For integrating measurements on outlets<br />
Quick-action hot wire probes<br />
For low velocities in ducts<br />
Robust hot bulb probes<br />
For direction-independent velocity measurement<br />
Comfort level probes<br />
For turbulence level measurements to DIN 1946, Part 2<br />
The reference for HVAC units, testo 400<br />
Temperature measurement<br />
The PTB accredited DKD laboratory for temperature guarantees<br />
reliable readings<br />
First PTB accredited DKD laboratory for surface temperature,<br />
developed in cooperation with PTB and the University of Ilmenau<br />
Patented crossband probe for fast surface measurements<br />
Custom-designed temperature probes for your application<br />
System accuracy of testo 400 up to 0.05 °C with precision probe<br />
0614 0240<br />
Current and voltage measurement<br />
Optional connection of external transmitters, such as particle<br />
counters and pressure transmitters and scaling of input in<br />
instrument<br />
CO and CO 2 measurement<br />
Long-term stable 2 beam procedure to measure reference and<br />
and measurement duct for C02 Humidity measurement<br />
The first PTB accredited DKD laboratory for air moisture and<br />
dew point temperature guarantees reliable readings<br />
Worldwide patented (capacitive) Testo humidity sensor<br />
Inter-laboratory tests in national and international institutes<br />
confirm a sensor accuracy of ±1 %RH<br />
2 year guaranteed long-term stability of the Testo humidity<br />
sensor under normal conditions<br />
Easy calibration or adjustment of humidity probe (on site) with<br />
defined salt solutions (11.3 %RH, 33 %RH and 75.3 %RH)<br />
Pressure measurement<br />
Very high accuracy in lower measuring range (100 Pa)<br />
from +/- (0.3 Pa + 0.5 % of reading)<br />
Temperature-compensated pressure measurement<br />
Velocity measurement<br />
Efficient readings by one of the first PTB accredited<br />
laboratories for velocity<br />
Reference laser Doppler anemometer guarantees<br />
calibration accuracy from 0.5 % of reading<br />
Thermal probes for high accuracy up to<br />
+/- (0.03 m/s + 5% of reading) in measuring range up to 20 m/s<br />
Density independent measurement from 500 hPa absolute<br />
pressure<br />
or up to 350 °C ambient temperature with vanes in measuring<br />
range from 0.4 m/s to 60 m/s<br />
Straight Pitot tube with greatly improved accuracy compared to<br />
Prandtl Pitot tubes on account of Pitot tube factor of 0.67<br />
Comfort level measurement<br />
High accuracy to determine degree of turbulence<br />
from ± (0.03 m/s + 4% of reading)
<strong>REFERENCE</strong> <strong>MEASURING</strong> <strong>INSTRUMENT</strong><br />
The reference for HVAC units, testo 400<br />
Saves or prints at<br />
the touch of a<br />
button<br />
Clear graphics<br />
display<br />
3 user defined<br />
function buttons<br />
testo 400<br />
Power connection/<br />
quick battery<br />
recharge<br />
Attachable printer<br />
Readings can be printed in the<br />
matter of seconds on location<br />
Max/Min<br />
London 1<br />
Vol. Mean<br />
Mean<br />
10.4 m/s<br />
m/s<br />
°C<br />
2 user defined<br />
probe sockets<br />
London 1<br />
Easy<br />
operation<br />
with cursor<br />
Part no.<br />
Data<br />
communication<br />
with PC,<br />
barcode pen<br />
testo 400<br />
Multi-function measuring instrument,<br />
incl. battery, Li cell and cal. protocol<br />
Used for:<br />
Velocity, volume flow<br />
Humidity, pressure<br />
Temperature<br />
CO , rpm and<br />
2<br />
current/velocity<br />
0563 4001<br />
305,-<br />
VAC module<br />
Turbulence degree<br />
Data management<br />
Probes<br />
500,000 readings<br />
Reference measurement<br />
Precision reference class measuring<br />
instruments have everything the professional<br />
user needs to complete complicated<br />
measurement tasks efficiently, accurately and<br />
conveniently.<br />
testo 400 includes the parameters<br />
temperature, CO2 , rpm, current, voltage,<br />
relative humidity, pressure, velocity and<br />
volume flow.<br />
Intelligent electronics ensure the latest<br />
technology is used thanks to software<br />
updates. The measuring instrument can<br />
always keep up with the measurement tasks at<br />
hand thanks to upgrades.<br />
Upgradable and teachable, highly reliable<br />
and of the highest quality - they are the<br />
properties which guarantee that the customer<br />
is equipped for the future.<br />
Useful instrument functions:<br />
System accuracy up to 0.05 °C and up to a<br />
resolution of 0.001 °C<br />
All functions of testo 650 and testo 950<br />
Input of cross-sections to volume flow<br />
calculation<br />
Absolute pressure compensation in thermal<br />
probes<br />
Density calculation for velocity<br />
measurement with reference to<br />
temperature, humidity and absolute<br />
pressure<br />
Turbulence degree meas. to DIN EN 27726,<br />
DIN 1946 Part 2, ISO<br />
Assessment of volume flow measurements<br />
with calculation of total inaccuracy of<br />
measurement in accordance with EN 12599<br />
with VAC module (optional)<br />
141<br />
°C<br />
%RH<br />
td<br />
g/kg<br />
hPa<br />
m/s<br />
m³/h<br />
ppm<br />
CO<br />
ppm<br />
CO2<br />
rpm<br />
mA<br />
V
142<br />
°C<br />
%RH<br />
td<br />
g/kg<br />
hPa<br />
m/s<br />
m³/h<br />
ppm<br />
CO<br />
ppm<br />
CO2<br />
rpm<br />
mA<br />
V<br />
testo 400<br />
Location<br />
m 3 /h<br />
m/s<br />
Vol. Max/Min Mean<br />
°C<br />
<strong>REFERENCE</strong> <strong>MEASURING</strong> <strong>INSTRUMENT</strong><br />
The reference testo 400...<br />
. . . guarantees quality<br />
Proof of temperature<br />
Guaranteeing quality usually involves keeping to a certain temperature.<br />
This ranges in the food industry from frozen food temperatures to<br />
component temperatures in the electronics branch. Easy documentation<br />
and filing of this measurement data is possible with testo 400.<br />
Remaining moisture/high-level moisture<br />
Highly accurate and efficient measurement of humidity and<br />
temperature is required in many industrial processes to guarantee<br />
continuous quality. testo 400 calculates all the physical parameters<br />
in the Mollier diagram.<br />
Production and storage conditions<br />
Sensitive products require the right air conditions during production<br />
and storage. The testo 400 multi-function measuring instrument<br />
includes all the functions needed to analyse air with reference to<br />
temperature, humidity and air flow.
<strong>REFERENCE</strong> <strong>MEASURING</strong> <strong>INSTRUMENT</strong><br />
The reference testo 400...<br />
. . . sets standards . . . documents processes<br />
Precision temperature Production unit<br />
Testo 400, when used together with the 0614 0240 precision probe,<br />
has a system accuracy of 0.05 °C in the measurement range from 0<br />
to 100 °C and a resolution of up to 0.001 °C. This high accuracy<br />
level makes testo 400 ideal as a working standard.<br />
Reference humidity Compressed air unit<br />
testo 400 is setting new standards in accuracy and long-term stability<br />
with its worldwide inter-laboratory tests carried out in leading<br />
international institutes enabling accurate monitoring of air humidity<br />
fluctuations of ± 1 %RH.<br />
Comfort level Air conditioning systems<br />
Top performance requires good ambient conditions. testo 400 offers<br />
standard assessment of temperature, humidity, velocity and CO 2 for<br />
employees, employers and companies such as air conditioning and<br />
heating fitters or planning offices.<br />
Products and systems are subjected to different temperatures during<br />
a production process. testo 400 has extensive monitoring options,<br />
for example, a measurement program is started when a temperature<br />
value is exceeded.<br />
Extensive economic damage, depending on the application, could<br />
be caused by uncontrolled moisture ingress in the compressed air<br />
system. It is possible to measure and document readings over a<br />
longer period of time using testo 400.<br />
If the volume flow is too low in VAC systems, there will be<br />
insufficient fresh air. If the volume flow is too high, this will lead to<br />
non-energy optimised operation. testo 400 offers efficient and<br />
standard measurement of volume flow.<br />
143<br />
°C<br />
%RH<br />
td<br />
g/kg<br />
hPa<br />
m/s<br />
m³/h<br />
ppm<br />
CO<br />
ppm<br />
CO2<br />
rpm<br />
mA<br />
V
144<br />
°C<br />
%RH<br />
td<br />
g/kg<br />
hPa<br />
m/s<br />
m³/h<br />
ppm<br />
CO<br />
ppm<br />
CO2<br />
rpm<br />
mA<br />
V<br />
Structuring measurement data:<br />
Readings can be saved at individual locations<br />
- with guarantee of refinding.<br />
The “tree structure” - folders, sub-folders and measurement protocols - guarantees<br />
an uncomplicated view.<br />
Practical additional information such as measurement information or required value input<br />
can be saved with the location.<br />
The locations can be selected via barcode labels using the pen.<br />
It is easy to draw an effective tour plan using the locations list.<br />
Long-term control made easy:<br />
User-friendly data logging, not only for spot checks<br />
The beginning of the measurement can be...<br />
- determined manually each time.<br />
- activated if a user defined limit value is exceeded.<br />
- set according to date/time.<br />
The measurement is completed when...<br />
- the predefined number of readings is reached.<br />
- date/time is reached.<br />
- the memory is full.<br />
- ended manually.<br />
Non-stop measurement via wrap-around memory...<br />
- deletes the oldest respective value.<br />
- is deactivated manually.<br />
Documentation on-site:<br />
The individual measurement protocol can be either saved or deleted following analysis.<br />
The printer immediately supplies the documentation required.<br />
The attachable comfort printer also offers graphical analysis options.<br />
Thermal paper for long-term legible measurement data documentation<br />
of up to 10 years.<br />
<strong>REFERENCE</strong> <strong>MEASURING</strong> <strong>INSTRUMENT</strong><br />
Measurement data with the measuring instrument:<br />
Structure - Measure - Printout on-site<br />
1 testo 400<br />
/<br />
.Week 35/02<br />
+ HALL A1<br />
+ HALL A2<br />
+ HALL C3<br />
Page- Change Page+<br />
Program Start<br />
Speicher<br />
Fühler<br />
Read out<br />
Spezial<br />
Progr. Start<br />
Gerät<br />
Keylock Meas.rate<br />
Drucken<br />
Mode End<br />
Language<br />
Save<br />
Delete<br />
Save Delete<br />
0008 Mean<br />
0.22 m/s<br />
965.00 m3/h<br />
27.70 °C<br />
Continue New End<br />
2 testo 400<br />
/<br />
..HALL A1<br />
WINDOW L1<br />
WINDOW L2<br />
WINDOW L3<br />
WINDOW R1<br />
WINDOW R2<br />
Door 1<br />
Door 2<br />
Page- Change Page+<br />
Speicher<br />
Fühler<br />
Read out<br />
Spezial<br />
Progr. Start<br />
Gerät<br />
Keylock Messrate<br />
Drucken<br />
Modeand Falls Endeshort<br />
Exceeds<br />
Languae<br />
Speichern<br />
Manual Löschen<br />
Date/Time<br />
16.10.02 15:29:32<br />
Mean<br />
Multi-point<br />
WINDOW 1<br />
-- m3/h<br />
870 1070<br />
02.01.02 23:26:38<br />
16.10.02 15:29:32<br />
Time axis:<br />
0 sec / Div.<br />
Mean:<br />
- - - - - - - - - - -<br />
1 0.22 m/s<br />
2 965.00 m3/h<br />
3 27.70 °C<br />
- - - - - - - - - - -<br />
Info:<br />
Req. 1000 m3/h<br />
16.10.02<br />
15:29:32<br />
16.10.02 15:29:32<br />
Mean<br />
Multi-point<br />
WINDOW 1<br />
1 m/s2 m3/h 3 °C<br />
- - - - - - - - - - -<br />
01 0.23 1010 27.7<br />
02 0.22 960 27.7<br />
03 0.20 880 27.7<br />
04 0.21 920 27.7<br />
05 0.20 870 27.7<br />
06 0.24 1040 27.7<br />
07 0.25 1070 27.7<br />
08 0.23 970 27.7<br />
Mean:<br />
- - - - - - - - - - -<br />
1 0.22 m/s<br />
2 965.00 m3/h<br />
3 27.70 °C<br />
- - - - - - - - - - -<br />
Info:<br />
Req. value 1000 m3/h<br />
16.10.02<br />
15:29:32
<strong>REFERENCE</strong> <strong>MEASURING</strong> <strong>INSTRUMENT</strong><br />
Measurements with ComSoft 3 software:<br />
Preparation - Analysis - Filing - Documentation<br />
Easy reading management:<br />
Preparation of the measurement:<br />
- The measurement program is determined and loaded into instrument<br />
- Tour plan is drawn up based on locations and<br />
is loaded into instrument.<br />
The measuring instrument is downloaded once measuring is complete:<br />
- The saved protocols are conveniently filed via the software using<br />
"Drag & Drop" or are analysed in Data.<br />
The readings are determined using the measuring instrument and can also be displayed<br />
online using the software.<br />
Comprehensive analysis, easy filing:<br />
Analysis:<br />
- with calculation functions<br />
- with crosshairs<br />
- with mean calculation<br />
- with calculation of standard deviation<br />
- taking all conventional refrigerants into consideration (refrigeration module, optional)<br />
Display:<br />
- as table or as graphic<br />
- as digit field or as histogram<br />
- with analog display<br />
- Measurement channels can be activated or deactivated at the touch of a button<br />
Documenting:<br />
- Data is transferred to Excel table using “Copy and Paste”.<br />
Individual configuration options:<br />
Your company logo can be included on the printouts.<br />
Functions can be selected from the function list and the finished profile can be<br />
saved.<br />
The online interface is available for LabVIEW software.<br />
Menu can be individually tailored to your needs.<br />
145<br />
°C<br />
%RH<br />
td<br />
g/kg<br />
hPa<br />
m/s<br />
m³/h<br />
ppm<br />
CO<br />
ppm<br />
CO2<br />
rpm<br />
mA<br />
V
146<br />
°C<br />
%RH<br />
td<br />
g/kg<br />
hPa<br />
m/s<br />
m³/h<br />
ppm<br />
CO<br />
ppm<br />
CO2<br />
rpm<br />
mA<br />
V<br />
testo 400, multi-function measuring instrument,<br />
incl. battery, Li cell and calibration protocol<br />
Part no.<br />
0563 4001<br />
Attachable printer (securely attached) with 1 roll of<br />
thermal paper and batteries<br />
Quickly prints readings on location<br />
Part no. 0554 0570<br />
Measuring instrument<br />
testo 400, multi-function measuring instrument, incl. battery, Li cell and<br />
calibration protocol<br />
2 channel multi-function measuring instrument<br />
Accessories for measuring instrument<br />
Memory upgrade to 500,000 readings<br />
Upgrades memory capacity (via service)<br />
Rechargeable battery set for instrument (4 rechargeables 2.4V/700mAh)<br />
Selected for quick recharging in instrument<br />
Mains unit 230 V/ 8 V/ 1 A, for instrument (European plug)<br />
For mains operation and battery recharging<br />
Car charging adapter, ready to measure following recharging in car<br />
Battery is recharged while travelling in car<br />
Spare Li cell to save RAM data<br />
When changing battery or rechargeable battery<br />
Printer and accessories<br />
Attachable printer (securely attached) with 1 roll of thermal paper and batteries<br />
Testo printer with 1 roll of thermal paper and 4 AA size batteries<br />
Prints readings on location<br />
testo 575 fast printer, incl. 1 roll of thermal paper and batteries<br />
Infrared thermal line printer with graph function<br />
Recharger for printer (with 4 standard rech. batt.)<br />
Rechargeable batteries are recharged externally<br />
Spare thermal paper for printer (6 rolls)<br />
Spare thermal paper for printer (6 rolls)<br />
Measurement data documentation legible for up to 10 years<br />
Barcode pen to read in measurement locations<br />
Quick and accurate allocation of reading to<br />
location<br />
Part no. 0554 0460<br />
Label thermal paper (Testo patent) for testo 575 printer (6 rolls), can be applied<br />
directly<br />
Testo printer with 1 roll of thermal paper and 4 AA<br />
size batteries<br />
Part no. 0554 0545<br />
testo 575 fast printer, incl. 1 roll of thermal<br />
paper and batteries<br />
Part no. 0554 1775<br />
Part no.<br />
0563 4001<br />
Part no.<br />
0554 9481<br />
0554 0196<br />
0554 1084<br />
0554 0424<br />
0515 0028<br />
Part no.<br />
0554 0570<br />
0554 0545<br />
0554 1775<br />
0554 0110<br />
0554 0569<br />
0554 0568<br />
0554 0561<br />
<strong>REFERENCE</strong> <strong>MEASURING</strong> <strong>INSTRUMENT</strong><br />
ComSoft 3 - Professional with data management<br />
Incl. database, analysis and graphics function, data<br />
analysis, trend curve<br />
Part no. 0554 0830<br />
testovent 410, volume flow funnel, Ø 340mm/330<br />
x 330mm, incl. case<br />
Part no. 0554 0410<br />
testovent 415, volume flow funnel, Ø<br />
210mm/190x190mm, incl. case<br />
Part no. 0554 0415<br />
Softcase for instrument and printer<br />
SoftCase (protects instrument from impact) with carrier strap, magnetic holder<br />
and probe holder<br />
SoftCase for attachable printer (protects printer from dirt/impact)<br />
Barcode and accessories<br />
Barcode pen to read in measurement locations<br />
Quick and accurate allocation of reading to location<br />
Barcode labels, self-adhesive (1200 off)<br />
Location marked with barcode, printed using software<br />
Adhesive pockets (50 off) for printout, paper barcode labels...<br />
Software and accessories<br />
ComSoft 3 - Professional with data management<br />
Incl. database, analysis and graphics function, data analysis, trend curve<br />
RS232 cable<br />
Connects instrument to PC (1.8 m) for data transfer<br />
Electrical isolation for RS232 (connects measuring instrument to PC)<br />
VAC module<br />
Memory upgrade to 500,000 readings<br />
Upgrades memory capacity (via service)<br />
VAC module upgrade<br />
Volume flow calculation in ducts with error calculation function in instrument<br />
VAC module upgrade, PC software, (for ComSoft 3 software)<br />
Printout of standard measurement protocols<br />
Refrigeration module<br />
System case<br />
Transport case (plastic) for measuring instrument, probes<br />
System case (plastic) for measuring instrument, probes and accessories<br />
Probes in lid make it easy to find parts in case<br />
System case (aluminium) for measuring instrument, probes and accessories<br />
Probes in lid make it easy to find parts in case<br />
Ordering data<br />
SoftCase (protects instrument from impact) with<br />
carrier strap, magnetic holder and probe holder<br />
Part no. 0516 0401<br />
SoftCase for attachable printer (protects printer<br />
from dirt/impact)<br />
Part no. 0516 0411<br />
System case (plastic)<br />
Part no. 0516 0400<br />
System case (aluminium)<br />
Part no. 0516 0410<br />
"Refrigeration technology" update with saved curves of all usual refrigerants<br />
Part no.<br />
0516 0401<br />
0516 0411<br />
Part no.<br />
0554 0460<br />
0554 0411<br />
0554 0116<br />
Part no.<br />
0554 0830<br />
0409 0178<br />
0554 0006<br />
Part no.<br />
0554 9481<br />
0450 4010<br />
0554 4030<br />
Part no.<br />
0554 4035<br />
Part no.<br />
0516 0300<br />
0516 0400<br />
0516 0410
<strong>REFERENCE</strong> <strong>MEASURING</strong> <strong>INSTRUMENT</strong><br />
Calibration certificates<br />
Calibration certificates/Temperature<br />
Part no.<br />
ISO calibration certificate/Temperature<br />
0520 0001<br />
For air/immersion probes, calibration points -18°C; 0°C; +60°C<br />
ISO calibration certificate/Temperature<br />
0520 0021<br />
Measuring instruments with air/immersion probe; calibration points 0°C; +150°C; +300°C<br />
ISO calibration certificate/Temperature<br />
0520 0071<br />
Thermometers with surface probe; calibration points +60°C; +120°C; +180°C<br />
DKD calibration certificate/Temperature<br />
0520 0211<br />
Meas. instr. with air/immersion probe; calibration points -20°C; 0°C; +60°C<br />
DKD calibration certificate/Temperature<br />
0520 0271<br />
Contact surface temperature probes; calibration points +100°C; +200°C; +300°C<br />
Calibration certificates/Humidity<br />
Part no.<br />
ISO calibration certificate/Humidity<br />
0520 0106<br />
Calibration points freely selectable from 5 to 95 %RH<br />
+15 to +35 °C (max DP +70 °C/min DP -30 °C)<br />
-18 to +80 °C<br />
ISO calibration certificate/Humidity<br />
0520 0006<br />
Electronic hygrometers; calibration points 11.3%RH and 75.3%RH at +25°C<br />
ISO calibration certificate/Humidity<br />
0520 0116<br />
Pressure dew point measuring instruments; calibration points freely selectable<br />
-40 to 0 °C at 6 bar<br />
ISO calibration certificate/Humidity<br />
0520 0013<br />
Saturated saline solutions: calibration point 11.3%RH<br />
ISO calibration certificate/Humidity<br />
0520 0083<br />
Saturated saline solutions, calibration point 75.3%RH<br />
DKD calibration certificate/Humidity<br />
0520 0206<br />
Electronic hygrometers; calibration points 11.3%RH and 75.3%RH at +25°C<br />
DKD calibration certificate/Humidity<br />
0520 0216<br />
Calibration points freely selectable from 5 to 95%RH<br />
+25°C<br />
-18 to +70°C<br />
DKD calibration certificate/Humidity<br />
0520 0213<br />
Saturated saline solutions; calibration point 11.3%RH<br />
DKD calibration certificate/Humidity<br />
0520 0283<br />
Saturated saline solutions; calibration point 75.3%RH<br />
Calibration certificates/Pressure<br />
Part no.<br />
ISO calibration certificate/Pressure<br />
0520 0115<br />
Absolute pressure; 5 pt. distributed over the whole measurement range<br />
ISO calibration certificate/Pressure<br />
0520 0005<br />
Differential pressure; 5 points distributed over meas. range (-1 to 250 bar)<br />
DKD calibration certificate/Pressure<br />
0520 0225<br />
Differential and positive pressure; 6 measuring points distributed over meas. range (> 0.6% of fsv)<br />
DKD calibration certificate/Pressure<br />
0520 0212<br />
Absolute pressure; 11 measuring points distributed over meas. range (0.1 to 0.6% of fsv)<br />
DKD calibration certificate/Pressure<br />
0520 0215<br />
Differential and positive pressure; 11 measuring points distributed over the instrument measuring range<br />
Calibration certificates/Velocity<br />
ISO calibration certificate/Velocity<br />
All velocity probes, calibration points selectable from 0.3 to 50 m/s at +25°C<br />
ISO calibration certificate/Velocity<br />
Hot wire, vane anemometer, Pitot tube; calibration points 1; 2; 5; 10 m/s<br />
ISO calibration certificate/Velocity<br />
Hot wire, vane anemometer, Pitot tube; calibration points 5; 10; 15; 20 m/s<br />
ISO calibration certificate/Velocity<br />
Hot wire, vane anemometer; calibration points 0.5; 0.8; 1; 1.5 m/s<br />
DKD calibration certificate/Velocity<br />
Hot wire, vane anemometer; calibration points 0.5; 1; 2; 5; 10 m/s<br />
DKD calibration certificate/Velocity<br />
Hot wire, vane anemometer, Pitot tube; calibration points 2; 5; 10; 15; 20 m/s<br />
DKD calibration certificate/Velocity<br />
Hot wire anemometer; calibration points 0.1; 0.2; 0.5; 0.8; 1 m/s<br />
Part no.<br />
0520 0104<br />
0520 0004<br />
0520 0034<br />
0520 0024<br />
0520 0244<br />
0520 0204<br />
0520 0224<br />
147<br />
°C<br />
%RH<br />
td<br />
g/kg<br />
hPa<br />
m/s<br />
m³/h<br />
ppm<br />
CO<br />
ppm<br />
CO2<br />
rpm<br />
mA<br />
V
148<br />
°C<br />
%RH<br />
td<br />
g/kg<br />
hPa<br />
m/s<br />
m³/h<br />
ppm<br />
CO<br />
ppm<br />
CO2<br />
rpm<br />
mA<br />
V<br />
<strong>REFERENCE</strong> <strong>MEASURING</strong> <strong>INSTRUMENT</strong><br />
Suitable probes at a glance<br />
Air probes Illustration Meas. range Accuracy t99 Conn. Part no.<br />
NTC probes<br />
Highly accurate air probe for air and gas<br />
temperature measurements with bare,<br />
mechanically protected sensor<br />
Pt100 probes<br />
Standard air probe<br />
Precision air probe<br />
NiCr-Ni probes<br />
Super quick-action immersion/penetration probe<br />
for measurements in gases and liquids with a<br />
low-mass tip<br />
150 mm<br />
Ø 9 mm<br />
150 mm<br />
Ø 3 mm<br />
150 mm<br />
Ø 3 mm<br />
Thermocouple, made of fibre-glass insulated<br />
2000 mm<br />
thermal pipes, pack of 5<br />
Insulation: twin conductor, flat, oval, opposed and covered with fibre-glass, both conductors are wrapped together with fibre-glass and<br />
soaked with lacquer, please order adapter 0600 1693<br />
Adapter to connect NiCr-Ni thermocouples and<br />
probes with open wire ends<br />
Ø 9 mm<br />
Ø 9 mm<br />
-40... +125 °C To UNI curve 60 s Fixed cable 0610 9714<br />
-200... +600 °C Class A 75 s<br />
-100... +400 °C<br />
1/10 Class B (0 to<br />
100°C) 1/5 Class B<br />
(rem. range) to EN<br />
60751<br />
75 s<br />
Plug-in head,<br />
connection cable 0430<br />
0143 or 0430 0145<br />
required<br />
Plug-in head,<br />
connection cable 0430<br />
0143 or 0430 0145<br />
required<br />
0604 9773<br />
0628 0017<br />
150 mm<br />
20 mm -200... +600 °C Class 1 1 s Plug-in head,<br />
connection cable 0430<br />
0604 9794<br />
Ø 1.4 mm Ø 0.5 mm<br />
0143 or 0430 0145<br />
required<br />
0614 9794 *<br />
Surface probes Illustration Meas. range Accuracy t99 Conn. Part no.<br />
NiCr-Ni probes<br />
Quick-action surface probe with sprung<br />
thermocouple strip, measuring range short-term<br />
to +500°C<br />
Super quick-action surface probe, probe tip at<br />
90° angle, with sprung thermocouple strip<br />
Robust surface probe<br />
Robust surface probe, at 90° angle, suitable for<br />
inaccessible places<br />
Robust surface probe with sprung thermocouple<br />
strip for high temperature range up to +700°C<br />
Roller surface probe for measurements on rollers<br />
and rotating drums, max. circumferential velocity<br />
18 to 400m/min<br />
Magnetic probe, adhesive power approx. 20 N,<br />
with magnets, for measurements on metal<br />
surfaces<br />
Magnetic probe, adhesive power approx. 10 N,<br />
with magnets, for higher temperatures, measures<br />
on metal surfaces<br />
Miniature surface probe for measurements on<br />
electronic components, small motors...<br />
Adhesive thermocouple, pack of 2, carrier material:<br />
aluminium foil<br />
Is fixed at the measuring point using conventional adhesives or silicone heat paste 0554 0004<br />
Adapter to connect NiCr-Ni thermocouples and<br />
probes with open wire ends<br />
Pt100 probes<br />
Robust surface probe<br />
Infrared probes<br />
Infrared surface probe for fast non-contact temperature<br />
measurement on live, inaccessible and rotating parts<br />
Accessories<br />
Silicone heat paste (14g), Tmax = +260°C<br />
Improves heat transfer in surface probes<br />
35 mm<br />
75 mm<br />
150 mm<br />
50 mm<br />
Ø 10 mm<br />
100 mm Ø 10 mm<br />
150 mm<br />
Ø 4 mm<br />
130 mm<br />
Ø 4 mm<br />
200 mm<br />
274 mm<br />
270 mm<br />
Ø 5 mm<br />
150 mm<br />
Ø 4 mm<br />
Ø 33 mm<br />
Ø 20 mm<br />
Ø 21 mm<br />
Ø 0.8 mm<br />
Ø 4 mm<br />
Ø 4 mm<br />
Ø 15 mm<br />
Diameter extension 2 x 0.2<br />
mm, 0.1 mm thick<br />
Ø 9 mm<br />
-200... +400 °C Class 1 5 s Please order adapter<br />
0600 1693<br />
-200... +300 °C Class 2 3 s Plug-in head,<br />
connection cable 0430<br />
0143 or 0430 0145<br />
required<br />
-200... +300 °C Class 2 3 s Plug-in head,<br />
connection cable 0430<br />
0143 or 0430 0145<br />
required<br />
-200... +600 °C Class 1 25 s<br />
Plug-in head,<br />
connection cable 0430<br />
0143 or 0430 0145<br />
required<br />
-200... +600 °C Class 1 25 s Plug-in head,<br />
connection cable 0430<br />
0143 or 0430 0145<br />
required<br />
-200... +700 °C Class 2 3 s Fixed cable,<br />
coiled<br />
-50... +240 °C Class 2 Fixed cable,<br />
coiled<br />
0644 1109<br />
Fixed cable 0600 1693<br />
0604 0194<br />
0604 0994<br />
0604 9993<br />
0604 9893<br />
*<br />
0614 0194 *<br />
0614 9993<br />
0614 9893<br />
0600 0394<br />
0600 5093<br />
-50... +170 °C Class 2 Fixed cable 0600 4793<br />
-50... +400 °C Class 2 Fixed cable 0600 4893<br />
-200... +400 °C Class 2 3 s Fixed cable 0600 1494<br />
-200... +200 °C Class 1 Please order adapter<br />
0600 1693<br />
-50... +400 °C Class B 40 s<br />
-18... +260 °C ±2% of mv (+100.1... +260 °C)<br />
±2 °C (-18... +100 °C)<br />
Fixed cable 0600 1693<br />
Plug-in head,<br />
connection cable 0430<br />
0143 or 0430 0145<br />
required<br />
2 s Fixed cable,<br />
coiled<br />
0644 1607<br />
0604 9973<br />
0628 0018<br />
0600 0750<br />
0554 0004<br />
*with EEPROM: Precision adjustment for each probe at a measuring point; measuring range limits are saved in probe; t95 extrapolation; surface allowance in surface probe can be adapted to measuring task<br />
*<br />
*<br />
*
<strong>REFERENCE</strong> <strong>MEASURING</strong> <strong>INSTRUMENT</strong><br />
Suitable probes at a glance<br />
Pipe wrap probes Illustration Meas. range Accuracy t99 Conn. Part no.<br />
NTC probes<br />
Pipe wrap probe for pipes with diameter of up to<br />
2", for flow/return temperature measurement in<br />
hydronic systems<br />
Spare meas. head for pipe wrap probe<br />
Pt100 probes<br />
Velcro probe for pipes with diameter of max. 100<br />
mm<br />
Immers./penetr. probes Illustration Meas. range Accuracy t99 Conn. Part no.<br />
NiCr-Ni probes<br />
Fast response immersion/penetration probe<br />
Super quick-action immersion/penetration probe<br />
for measurements in liquids<br />
Super quick-action immersion/penetration probe<br />
for high temperatures<br />
Super quick-action immersion/penetration probe<br />
for measurements in gases and liquids with a<br />
low-mass tip<br />
Robust immersion/penetration probe made of<br />
V4A stainless steel, waterproof and oven-proof,<br />
e.g. for the food sector<br />
Smelting probe for measurements in non-ferrous<br />
melting baths, with exchangeable measuring tips<br />
Spare measuring tip for smelting probe<br />
Pt100 probes<br />
Standard immersion/penetration probe<br />
Standard immersion/penetration probe<br />
Highly accurate immersion/penetration probe<br />
incl. certificate<br />
Highly accurate immersion/penetration probe<br />
Flexible precision immersion probe, cable heatproof<br />
up to +300°C<br />
Robust immersion/penetration probe with<br />
sharpened measuring tip, waterproof and ovenproof<br />
1100 mm<br />
Ø 6.5 mm<br />
150 mm<br />
-200... +400 °C Class 1 3 s Plug-in head,<br />
connection cable 0430<br />
0604 0293<br />
Ø 3 mm<br />
0143 or 0430 0145<br />
required<br />
0614 0293 *<br />
150 mm<br />
-200... +600 °C Class 1 1 s Plug-in head,<br />
connection cable 0430<br />
0604 0493<br />
Ø 1.5 mm<br />
0143 or 0430 0145<br />
required<br />
0614 0493 *<br />
470 mm<br />
-200... +1100 °C Class 1 1 s Plug-in head,<br />
connection cable 0430<br />
0604 0593<br />
Ø 1.5 mm<br />
0143 or 0430 0145<br />
required<br />
0614 0593 *<br />
150 mm<br />
20 mm -200... +600 °C Class 1 1 s Plug-in head,<br />
connection cable 0430<br />
0604 9794<br />
Ø 1.4 mm Ø 0.5 mm<br />
0143 or 0430 0145<br />
required<br />
0614 9794 *<br />
150 mm<br />
Ø 3.5 mm<br />
Ø 6.5 mm<br />
200 mm Stainless steel<br />
Ø 3 mm<br />
200 mm<br />
Ø 3 mm<br />
295 mm<br />
Ø 4 mm<br />
200 mm<br />
Ø 3 mm<br />
1000 mm<br />
Ø 3.5 mm<br />
150 mm<br />
Ø 3.5 mm<br />
Nickel<br />
Stainless steel<br />
50 mm<br />
Ø 6 mm<br />
Ø 3 mm<br />
Ø 3 mm<br />
-200... +400 °C Class 1 3 s Fixed cable 0600 2593<br />
-200... +1250 °C Class 1 60 s Fixed cable 0600 5993<br />
-200... +1250 °C Class 1 60 s 0363 1712<br />
-200... +400 °C Class A 20 s<br />
-200... +600 °C Class A 20 s<br />
-40... +300 °C<br />
±0.05 °C (+0.01... +100 °C)<br />
±(0.05 °C ±0.05% of mv)<br />
(-40... 0 °C)<br />
±(0.05 °C ±0.05% of mv)<br />
(+100.01... +300 °C)<br />
-100... +400 °C 1/10 Class B (0 to<br />
100°C) 1/5 Class B<br />
(rem. range) to EN<br />
60751<br />
-100... +265 °C 1/10 Class B (0 to<br />
100°C) 1/5 Class B<br />
(rem. range) to EN<br />
60751<br />
60 s<br />
30 s<br />
80 s<br />
Plug-in head,<br />
connection cable 0430<br />
0143 or 0430 0145<br />
required<br />
Plug-in head,<br />
connection cable 0430<br />
0143 or 0430 0145<br />
required<br />
Plug-in head,<br />
connection cable 0430<br />
0143 or 0430 0145<br />
required<br />
Plug-in head,<br />
connection cable 0430<br />
0143 or 0430 0145<br />
required<br />
Plug-in head,<br />
connection cable 0430<br />
0143 or 0430 0145<br />
required<br />
0604 0273<br />
0604 0274<br />
0614 0240<br />
0628 0015<br />
0628 0016<br />
-200... +400 °C Class A 30 s Fixed cable 0604 2573<br />
Plug-in measuring tips Illustration Meas. range Accuracy t99 Conn. Part no.<br />
Plug-in measuring tip, 750mm long, flexible, for<br />
high temperatures, outer casing: stainless steel<br />
1.4541<br />
Plug-in measuring tip, 1200 mm long, flexible,<br />
for high temperatures, outer casing: stainless<br />
steel 1.4541<br />
Plug-in measuring tip, 550mm long, flexible, for<br />
high temperatures, outer casing: Inconel 2.4816<br />
Plug-in measuring tip, 1030mm long, flexible, for<br />
high temperatures, outer casing: Inconel 2.4816<br />
Handle for plug-in measuring tip<br />
35 mm<br />
15 mm<br />
750 mm<br />
Ø 3 mm<br />
1200 mm<br />
Ø 3 mm<br />
550 mm<br />
Ø 3 mm<br />
1030 mm<br />
Ø 3 mm<br />
-60... +130 °C Class 2 5 s Fixed cable 0600 4593<br />
-60... +130 °C Class 2 5 s 0602 0092<br />
-50... +150 °C Class B 40 s Fixed cable 0628 0019<br />
-200... +900 °C Class 1 4 s Please order handle<br />
with Part no. 0600 5593<br />
-330... +900 °C Class 1 4 s Please order handle<br />
with Part no. 0600 5593<br />
-200... +1100 °C Class 1 4 s Please order handle<br />
with Part no. 0600 5593<br />
-200... +1100 °C Class 1 4 s Please order handle<br />
with Part no. 0600 5593<br />
0600 5393<br />
0600 5493<br />
0600 5793<br />
0600 5893<br />
0600 5593<br />
*with EEPROM: Precision adjustment for each probe at a measuring point; measuring range limits are saved in probe; t95 extrapolation; surface allowance in surface probe can be adapted to measuring task<br />
*<br />
*<br />
149<br />
°C<br />
%RH<br />
td<br />
g/kg<br />
hPa<br />
m/s<br />
m³/h<br />
ppm<br />
CO<br />
ppm<br />
CO2<br />
rpm<br />
mA<br />
V
150<br />
°C<br />
%RH<br />
td<br />
g/kg<br />
hPa<br />
m/s<br />
m³/h<br />
ppm<br />
CO<br />
ppm<br />
CO2<br />
rpm<br />
mA<br />
V<br />
Current/voltage cable (±1 V, ±10 V, 20 mA)<br />
Cable, 1.5 m long, connects probe with plug-in head to meas. instrument<br />
Cable, 5 m long, connects probe with plug-in head to measuring instrument<br />
Extension cable, 5 m long, between plug-in head cable and instrument<br />
0430 0143<br />
0430 0145<br />
0409 0063<br />
<strong>REFERENCE</strong> <strong>MEASURING</strong> <strong>INSTRUMENT</strong><br />
Suitable probes at a glance<br />
Other temperature probes Illustration Meas. range Accuracy Conn. Part no.<br />
Globe thermometer to measure radiant heat Ø 150 mm<br />
More probes Illustration<br />
Meas. range Accuracy Conn. Part no.<br />
CO probe to measure CO level in ambient air<br />
CO2 probe measures indoor air quality and monitors the<br />
workplace. With plug-in head, connection cable 0430<br />
0143 or 0430 0145 required<br />
Mechanical rpm probe with plug-in head<br />
Included<br />
2 probe tips Ø 8 and Ø 12 mm<br />
1 hollow cone Ø 8 mm<br />
1 surface speed disc Ø 19 mm to measure rotational speed: rpm =<br />
rotational speed in mm/s<br />
190 mm<br />
0... +500 ppm CO ±5% of mv (+100.1... +500 ppm CO)<br />
±5 ppm CO (0... +100 ppm CO)<br />
0... +1 Vol. % CO2<br />
0... +10000 ppm<br />
CO2<br />
±(50 ppm CO2 ±2% of mv) (0... +5000 ppm<br />
CO2)<br />
±(100 ppm CO2 ±3% of mv) (+5001...<br />
+10000 ppm CO2)<br />
+20... +20000 rpm ± 1 digit Plug-in head,<br />
connection cable 0430<br />
0143 or 0430 0145<br />
required<br />
0640 0340<br />
0... +1000 mV ±1 mV (0... +1000 mV) 0554 0007<br />
0... +10 V<br />
±0.01 V (0... +10 V)<br />
0... +20 mA ±0.04 mA (0... +20 mA)<br />
Accessories for temperature probes Part no. Accessories for temperature probes<br />
PUR coating material<br />
PUR coating material<br />
PUR coating material<br />
Accuracy corresponds to<br />
ISO 7243, ISO 7726, DIN<br />
EN 27726, DIN 33403<br />
requirements<br />
Ø 25 mm<br />
Selecting the right temperature sensor<br />
Measuring range<br />
°C<br />
+1200 ––<br />
+1100 ––<br />
+1000 ––<br />
+900 ––<br />
+800 ––<br />
+700 ––<br />
+600 ––<br />
+500 ––<br />
+400 ––<br />
+300 ––<br />
+200 ––<br />
+100 ––<br />
0 ––<br />
-100 ––<br />
-200 ––<br />
NTC<br />
Pt100<br />
Type K<br />
Type J<br />
Type T<br />
Thermocouples<br />
0... +120 °C<br />
±0.5 °C (0... +49.9 °C)<br />
±1 °C (+50... +120 °C)<br />
Telescopic handle, max. 1 m, for probe with plug-in head<br />
Cable: 2.5 m long, PUR coating material<br />
Fixed cable 0632 1247<br />
Plug-in head,<br />
connection cable 0430<br />
0143 or 0430 0145<br />
required<br />
Glass pipe for immersion/penetration probe to protect from corrosive agents<br />
For probes with Part nos. 0604 0273 and 0628 0015<br />
Accuracy<br />
Accuracy data<br />
Sensor Temp.<br />
range<br />
Thermocouple<br />
Type K (NiCr-Ni)<br />
-40...+1200 °C<br />
-40...+1000 °C<br />
-40...+350 °C<br />
-40...+750 °C<br />
-100...+200 °C<br />
-200...+600 °C<br />
-50...-25,1 °C<br />
-25...+74,9 °C<br />
+75...+150 °C<br />
-30...-20,1 °C<br />
-20...0 °C<br />
+0,1...+75 °C<br />
+75,1...+275 °C<br />
Fixed cable 0554 0670<br />
0632 1240<br />
Part no.<br />
0430 0144<br />
0554 7072<br />
Select the sensor with the accuracy required for your application from the diagram or table.<br />
Type T<br />
Type J<br />
Pt100<br />
NTC<br />
(Standard)<br />
NTC<br />
(High temp.)<br />
Data for thermocouples to EN<br />
60584-1 (formerly IEC 584-1). Two<br />
values are given. One fixed value in<br />
°C and a formula.<br />
Class Maximum tolerances<br />
Fixed value Referred to temperature<br />
2<br />
1<br />
1<br />
1<br />
B<br />
A<br />
–<br />
–<br />
– °C<br />
±2,5 °C<br />
±1,5 °C<br />
±0,5 °C<br />
±1,5 °C<br />
± (0,3 + 0,005 ItI)<br />
± (0,15 + 0,002 ItI)<br />
±0.4 °C<br />
±0.2 °C<br />
±0.5 % of reading<br />
The larger value always applies.<br />
Data for Pt100 to EN 60751<br />
(formerly IEC 751).<br />
±0,0075 x ltl<br />
±0,004 x ltl<br />
±0,001 x ltl<br />
±0,004 x ltl<br />
±1 °C<br />
±0.6 °C<br />
±0.5 °C<br />
±0.5 °C ±0.5 % of reading<br />
t=Measuring temperature<br />
There is no standardization for NTC<br />
sensors.
<strong>REFERENCE</strong> <strong>MEASURING</strong> <strong>INSTRUMENT</strong><br />
Suitable probes at a glance<br />
Probes Illustration Meas. range Accuracy t90 Conn. Part no.<br />
Air probes<br />
Standard indoor air quality probe up to +70°C<br />
Duct humidity/temperature probe, can be<br />
connected to telescopic handle<br />
Telescopic handle 0430 9715, see Ordering data for Accessories<br />
Thin humidity probe incl. 4 attachable protection caps for<br />
ambient air measurements, measurements in exhaust air<br />
ducts and equilibrium moisture measurements<br />
Highly accurate reference humidity/temp. probe<br />
incl. calibration certificate<br />
Humidity/temperature probe<br />
Process humidity<br />
Standard pressure dew point probe for<br />
measurements in compressed air systems<br />
Precision pressure dew point probe for<br />
measurements in compressed air systems incl.<br />
cert. with test point -40°C tpd<br />
High humidity level probe with heated sensor<br />
element, no condensation on sensor<br />
Robust high temperature/humidity probe up to<br />
+180°C<br />
Flexible humidity probe (does not retain shape)<br />
for measurements in inaccessible places<br />
Material and equilibrium moisture<br />
Flexible humidity probe with mini module for<br />
meas. e.g. on material testing rigs, module cable<br />
length 1500mm, probe tip 50x19x7mm<br />
Sword probe for measuring humidity and<br />
temperature in stacked material<br />
Robust humidity probe e.g. for measuring<br />
equilibrium moisture or for measurements in<br />
exhaust ducts to +120°C<br />
Material moisture probe<br />
Material/building moisture cable<br />
aw value<br />
aw value set: pressure-tight precision humidity<br />
probe with certificate, measurement chamber and<br />
5 sample bowls (plastic)<br />
Pressure probe, 100 hPa, measures differential<br />
pressure<br />
Pressure probe, 2000 hPa, measures absolute<br />
pressure<br />
Low pressure probe, refrigerant-proof stainless<br />
steel, without cable<br />
300 mm<br />
300 mm<br />
Ø 21 mm<br />
Ø 21 mm<br />
180 mm<br />
Ø 12 mm<br />
300 mm<br />
Ø 12 mm<br />
300 mm<br />
Ø 12 mm<br />
Ø 12 mm<br />
1500 mm 100 mm<br />
Ø 12 mm<br />
320 mm<br />
Ø 18 mm<br />
300 mm<br />
Ø 12 mm<br />
0... +100 %RH ±1 %RH (+10... +90 %RH) * ±0.4 °C (-10... +50 °C)<br />
-20... +70 °C<br />
±2 %RH (0... +9.9 %RH) ±0.5 °C (-20... -10.1 °C)<br />
±2 %RH (+90.1... +100 %RH) ±0.5 °C (+50.1... +70 °C)<br />
-20... +70 °C<br />
0... +100 %RH<br />
-30... +50 °C tpd<br />
0... +100 %RH<br />
-60... +50 °C tpd<br />
0... +100 %RH<br />
-20... +85 °C<br />
0... +100 %RH<br />
-20... +180 °C<br />
0... +100 %RH<br />
-20... +180 °C<br />
0... +100 %RH<br />
-20... +125 °C<br />
0... +100 %RH<br />
-20... +70 °C<br />
0... +100 %RH<br />
-20... +120 °C<br />
±2.5 %RH (0... +100<br />
%RH)<br />
±2 %RH (+2... +98<br />
%RH)<br />
±2 %RH (+2... +98<br />
%RH)<br />
±2 %RH (+2... +98<br />
%RH)<br />
±2 %RH (+2... +98<br />
%RH)<br />
±2 %RH (+2... +98<br />
%RH)<br />
±0.9 °C tpd (+0.1... +50 °C tpd)<br />
±1 °C tpd (-4.9... 0 °C tpd)<br />
±2 °C tpd (-9.9... -5 °C tpd)<br />
±3 °C tpd (-19.9... -10 °C tpd)<br />
±4 °C tpd (-30... -20 °C tpd)<br />
±0.8 °C tpd (-4.9... +50 °C tpd)<br />
±1 °C tpd (-9.9... -5 °C tpd)<br />
±2 °C tpd (-19.9... -10 °C tpd)<br />
±3 °C tpd (-29.9... -20 °C tpd)<br />
±4 °C tpd (-40... -30 °C tpd)<br />
±0.4 °C (-10... +50 °C)<br />
±0.5 °C (-20... -10.1 °C)<br />
±0.5 °C (+50.1... +85 °C)<br />
±0.4 °C (+0.1... +50 °C)<br />
±0.5 °C (-20... 0 °C)<br />
±0.5 °C (+50.1... +180 °C)<br />
±0.4 °C (+0.1... +50 °C)<br />
±0.5 °C (-20... 0 °C)<br />
±0.5 °C (+50.1... +180 °C)<br />
±0.4 °C (-10... +50 °C)<br />
±0.5 °C (-20... -10.1 °C)<br />
±0.5 °C (+50.1... +125 °C)<br />
12 s Plug-in head, 0636 9741<br />
connection cable 0430<br />
0143 or 0430 0145<br />
required<br />
300 s Plug-in head, 0636 9840<br />
connection cable 0430<br />
0143 or 0430 0145<br />
required<br />
300 s Plug-in head, 0636 9841<br />
connection cable 0430<br />
0143 or 0430 0145<br />
required<br />
30 s Plug-in head, 0636 2142<br />
connection cable 0430<br />
0143 or 0430 0145<br />
required<br />
30 s Plug-in head, 0628 0021<br />
connection cable 0430<br />
0143 or 0430 0145<br />
required<br />
30 s Plug-in head, 0628 0022<br />
connection cable 0430<br />
0143 or 0430 0145<br />
required<br />
20 s Plug-in head, 0628 0013<br />
connection cable 0430<br />
0143 or 0430 0145<br />
required<br />
±0.4 °C (-10... +50 °C)<br />
±0.5 °C (-20... -10.1 °C)<br />
±0.5 °C (+50.1... +70 °C)<br />
12 s Plug-in head, 0636 0340<br />
connection cable 0430<br />
0143 or 0430 0145<br />
required<br />
±0.4 °C (-10... +50 °C) 30 s<br />
±0.5 °C (-20... -10.1 °C)<br />
±0.5 °C (+50.1... +120 °C)<br />
Plug-in head, 0636 2140<br />
connection cable 0430<br />
0143 or 0430 0145<br />
required<br />
1500 mm Free scaling, reference<br />
measurement, no water<br />
level<br />
Screw-in<br />
thread 7/16"<br />
UNF<br />
0 to 100 k Ohm =<br />
100 to 0 %<br />
±0.4 °C (+0.1... +50 °C)<br />
±0.5 °C (-20... 0 °C)<br />
±0.5 °C (+50.1... +70 °C)<br />
Plug-in head, 0636 9742<br />
connection cable 0430<br />
0143 or 0430 0145<br />
required<br />
Displayed values in<br />
instrument mean: 100 to 66<br />
wet; 0 to 1 very dry<br />
0636 0365<br />
0636 0565<br />
0... +1 aW ±0.01 aW (+0.1... +0.9 ±0.4 °C (-10... +50 °C) Reproducibility of 0628 0024<br />
0... +100 %RH<br />
-20... +70 °C<br />
aW)<br />
±0.02 aW (+0.9... +1 aW)<br />
±0.5 °C (-20... -10.1 °C)<br />
±0.5 °C (+50.1... +70 °C)<br />
aw value ±0.003<br />
* in the temperature range from +10°C to +30°C<br />
Description Illustration Probe type Meas. range Accuracy Conn. Part no.<br />
Precision pressure probe, 100 Pa, measures<br />
differential pressure<br />
Pressure probe, 10 hPa, measures differential<br />
pressure<br />
High pressure probe, refrigerant-proof stainless<br />
steel, up to 30 bar, without cable<br />
High press. probe, refrigerant-proof st. steel, up<br />
to 40 bar, w/o cable<br />
250 mm<br />
Ø 4 mm<br />
Screw-in<br />
thread 7/16"<br />
UNF<br />
Screw-in<br />
thread 7/16"<br />
UNF<br />
0... +100 %RH<br />
-20... +70 °C<br />
±2 %RH (+2... +98<br />
%RH)<br />
±0.4 °C (-10... +50 °C)<br />
±0.5 °C (-20... -10.1 °C)<br />
±0.5 °C (+50.1... +70 °C)<br />
12 s Plug-in head, 0636 9740<br />
connection cable 0430<br />
0143 or 0430 0145<br />
required<br />
0... +100 %RH ±2 %RH (+2... +98 ±0.4 °C (-10... +50 °C) 12 s Fixed cable 0636 9715<br />
-20... +70 °C<br />
%RH)<br />
±0.5 °C (-20... -10.1 °C)<br />
±0.5 °C (+50.1... +70 °C)<br />
0... +100 %RH<br />
-20... +70 °C<br />
±2 %RH (+2... +98<br />
%RH)<br />
±0.4 °C (-10... +50 °C)<br />
±0.5 °C (-20... -10.1 °C)<br />
±0.5 °C (+50.1... +70 °C)<br />
15 s Plug-in head, 0636 2130<br />
connection cable 0430<br />
0143 or 0430 0145<br />
required<br />
Differential 0... +100 Pa ±(0.3 Pa ±0.5% of mv)<br />
pressure probe<br />
(0... +100 Pa)<br />
Differential<br />
pressure probe<br />
Absolute pressure<br />
probe<br />
Low pressure<br />
probe<br />
High pressure<br />
probe<br />
0... +100 hPa ±0.5% of mv (+20...<br />
+100 hPa)<br />
±0.1 hPa (0... +20 hPa)<br />
0638 1345<br />
Differential<br />
pressure probe<br />
0... +10 hPa ±0.03 hPa (0... +10 hPa) 0638 1445<br />
High pressure<br />
probe<br />
0638 1545<br />
0... +2000 hPa ±5 hPa (0... +2000 hPa) 0638 1645<br />
-1... +10 bar ±1% of f.v. (-1... +10 bar) Plug-in head,<br />
0638 1740<br />
Overload<br />
connection cable 0409<br />
±32 bar (-1... +10 bar) 1745 required<br />
0... +30 bar ±1% of f.v. (0... +30 bar) Plug-in head,<br />
0638 1840<br />
Overload<br />
connection cable 0409<br />
±70 bar (0... +30 bar) 1745 required<br />
0... +40 bar ±1% of f.v. (0... +40 bar) Plug-in head,<br />
0638 1940<br />
Overload<br />
connection cable 0409<br />
±70 bar (0... +40 bar) 1745 required<br />
151<br />
°C<br />
%RH<br />
td<br />
g/kg<br />
hPa<br />
m/s<br />
m³/h<br />
ppm<br />
CO<br />
ppm<br />
CO2<br />
rpm<br />
mA<br />
V
152<br />
°C<br />
%RH<br />
td<br />
g/kg<br />
hPa<br />
m/s<br />
m³/h<br />
ppm<br />
CO<br />
ppm<br />
CO2<br />
rpm<br />
mA<br />
V<br />
<strong>REFERENCE</strong> <strong>MEASURING</strong> <strong>INSTRUMENT</strong><br />
Suitable probes at a glance<br />
Caps for humidity probes Ø 12 and 21 mm Illustration For humidity probes:<br />
Part no.<br />
Metal protection cage, Ø 21 mm for humidity probes, material: stainless steel V4A. Quick adjustment<br />
time, robust and temperature-proof. Used when measuring velocities of less than 10 m/s<br />
Metal protection cage, Ø 12 mm for humidity probes, material: stainless steel V4A. Quick adjustment<br />
time, robust and temperature-proof. Used when measuring velocities of less than 10 m/s.<br />
Wire mesh filter, Ø 21 mm, insertable filter for metal protection cage and plastic cap. Material:<br />
stainless steel V4A, quick adjustment time, protects from dirt and damage. Applications: meteorology,<br />
splashwater, condensation.<br />
Cap with wire mesh filter, Ø 12 mm<br />
Teflon sintered filter, Ø 21 mm, PTFE. Not affected by condensation, water-repellent, resistant to<br />
corrosive substances. Applications: compressed air measurements, high humidity range (continuous<br />
measurements), high velocities<br />
Teflon sintered filter, Ø 12 mm, PTFE. Not affected by condensation, water-repellent, resistant to<br />
corrosive substances. Applications: compressed air measurements, high humidity range (continuous<br />
measurements), high velocities<br />
Stainless steel sintered cap, Ø 21 mm, made of stainless steel V2A. Highly robust, suitable for<br />
penetration, clean with compressed air, mechanical protection of sensor. Applications: high<br />
mechanical loads, high velocity speeds.<br />
Stainless steel sintered cap, Ø 12mm, made of stainless steel V2A. Highly robust, suitable for<br />
penetration, should be cleaned with compressed air, mechanical protection of sensor. Applications:<br />
high mechanical loads, high velocity speeds.<br />
Teflon cap, Ø 5 mm, attachable, PTFE material, (5 off). Applications: dust protection, high humidity<br />
level measurements, high velocities<br />
Cable, 1.5 m long, connects probe with plug-in head to meas. instrument<br />
PUR coating material<br />
Cable, 5 m long, connects probe with plug-in head to measuring instrument<br />
PUR coating material<br />
Extension cable, 5 m long, between plug-in head cable and instrument<br />
PUR coating material<br />
Telescopic handle, max. 1 m, for probe with plug-in head<br />
Cable: 2.5 m long, PUR coating material<br />
Telescopic handle, 340 - 800mm long<br />
Adapter for surface humidity measuring, for humidity probes Ø 12mm<br />
Locates damp spots on walls, for example<br />
Cap for bore holes, for humidity probe with 12mm diameter<br />
Measures equilibrium moisture in bore holes<br />
Control and humidity adjustment set 11.3%RH/75.3%RH incl. adapter for<br />
humidity probes<br />
Control and storage humidity (33%RH) for humidity probes<br />
0430 0143<br />
0430 0145<br />
0409 0063<br />
0430 0144<br />
0430 9715<br />
0628 0012<br />
0554 2140<br />
0554 0660<br />
0554 0636<br />
Ø 21 mm<br />
Ø 12 mm<br />
Ø 21 mm<br />
Ø 21 mm<br />
Ø 12 mm<br />
Ø 21 mm<br />
Ø 12 mm<br />
Ø 5 mm<br />
All humidity probes with Ø 21 mm 0554 0665<br />
0636 9740, 0636 9715 0554 0755<br />
All humidity probes with Ø 21 mm 0554 0667<br />
All humidity probes with Ø 12 mm 0554 0757<br />
All humidity probes with Ø 21 mm 0554 0666<br />
0636 9769, 0636 9740, 0636 9715 0554 0756<br />
All humidity probes Ø 21 mm 0554 0640<br />
0636 9740, 0636 9715 0554 0647<br />
0636 2130 0554 1031<br />
Accessories for humidity probes/sensors Part no.<br />
Accessories for pressure probes<br />
Part no.<br />
Connection cable for pressure probes 0638 1740, 0638 1840, 0638 1940<br />
Magnetic holder for pressure probes<br />
For pressure probes 0638 1345/..1445/..1545/..1645<br />
Adapter for pressure probes, 1/2" outer thread, 1/4" inner thread<br />
Why you should choose humidity meas. instr. from Testo<br />
09/00<br />
12/00<br />
Inter-lab. tests<br />
KRISS<br />
NRCCRM<br />
08/01<br />
04/00<br />
PTB<br />
10/00<br />
05/00<br />
JQA<br />
03/01<br />
03/99<br />
08/96<br />
PTB<br />
04/96<br />
10/98<br />
INTA<br />
10/98<br />
Three precision probes were subjected to extensive inter-laboratory tests at the PTB in Berlin, NIST in<br />
the USA, the French national institute CETIAT, the Italian institute IMGC, the English national institute<br />
NPL, the Spanish national institute INTA, JQA in Japan, KRISS in Korea, NRCCRM in Peking and in<br />
Testo`s DKD calibration laboratory. The results confirm an accuracy of ±1 %RH for the probes, as<br />
indicated by Testo.<br />
10/96<br />
09/98<br />
CETIAT<br />
12/96<br />
07/97<br />
NPL<br />
10/96<br />
09/98<br />
NIST<br />
IMGC<br />
05/97<br />
10/97<br />
Deviation from resp. reference in %RH<br />
3<br />
2<br />
1<br />
0<br />
-1<br />
-2<br />
Long-term stability of Testo humidity sensor<br />
Reference humidity probes for highest precision<br />
- Accuracy ±1 %RH<br />
- 2 year guaranteed long-term stability under<br />
normal conditions<br />
0409 1745<br />
0554 0225<br />
0699 3127<br />
PTB; 23.08.1996<br />
JQA; 10.04.2000<br />
25°C<br />
-3<br />
0 10 20 30 40 50 60 70 80 90 100<br />
Results of the worldwide inter-laboratory test on 3<br />
precision humidity probes 1996-1997
158<br />
°C<br />
%RH<br />
td<br />
g/kg<br />
hPa<br />
m/s<br />
m³/h<br />
ppm<br />
CO<br />
ppm<br />
CO2<br />
rpm<br />
mA<br />
V<br />
Outgoing air<br />
Ingoing air/VAC->insulator<br />
Example of layout of a clean room<br />
Lab A2<br />
44.4 %<br />
22.5 °C<br />
Clean room Class D<br />
43.1 %<br />
Mean<br />
Isolator Class A<br />
Checking the flow speed using the hot wire probe Part no.: 0635 1041<br />
Ingoing air/VAC->clean room<br />
Outgoing air<br />
testo 400 display during the calibration of a<br />
stationary transmitter:<br />
Left display half: Reference humidity probe<br />
Right display half: 4-20 output measurement<br />
in a transmitter using current/voltage cable<br />
(scaling 0-100%RH)<br />
<strong>REFERENCE</strong> <strong>MEASURING</strong> <strong>INSTRUMENT</strong><br />
Defined process ambient conditions must be<br />
assured for the qualification and validation<br />
of the high quality standards of production<br />
units in clean rooms.<br />
Air exchange and the resulting air flow are<br />
linked directly to air temperature and air<br />
moisture. Specified air flows produce<br />
defined positive pressures which prevent the<br />
ingress of impurities from outside.<br />
Testo's measurement technology has proven<br />
to be ideal for testing process ambient<br />
conditions.<br />
With the testo 400 reference measuring<br />
instrument, you have the possibility of<br />
connecting 2 probes simultaneously. The<br />
measuring instrument can then be used to<br />
monitor measurements on-site or for longterm<br />
measurements thanks to the integrated<br />
readings memory with capacity for 48,000<br />
data.<br />
Typical measurement tasks:<br />
differential pressure monitoring<br />
using the 100 Pa probe<br />
The testo 100 Pa probe with an accuracy of<br />
±(0.3 Pa + 0.5 % of the reading) is the ideal<br />
solution.<br />
Position dependencies are completely<br />
eliminated thanks to the revolutionary<br />
double membrane technology and<br />
fluctuations in temperature no longer have<br />
any influence on the measured result thanks<br />
to temperature compensation.<br />
Recommended kit:<br />
The Pro Set for clean room systems<br />
testo 400, multi-function measuring instrument, incl. battery, Li cell and<br />
calibration protocol<br />
Precision pressure probe, 100 Pa, measures differential pressure and velocities<br />
(in connection with Pitot tube)<br />
Precision air probe<br />
Plug-in head, connection cable 0430 0143 or 0430 0145 required<br />
Highly accurate reference humidity/temp. probe incl. calibration certificate<br />
Plug-in head, connection cable 0430 0143 or 0430 0145 required<br />
Cable, 1.5 m long, connects probe with plug-in head to meas. instrument<br />
Cable, 1.5 m long, connects probe with plug-in head to meas. instrument<br />
Quick-action hot wire probe, Ø 10 mm, with telescopic handle, for<br />
measurements in the lower velocity range with direction recognition<br />
Bendable vane probe (can be bent by 90°), Ø 100 mm, attachable to handle or<br />
telescopic handle, for measurements on ventilation outlets<br />
Professional telescopic handle for plug-in vane probes, max. 1 m long,<br />
extension on request<br />
Current/voltage cable (±1 V, ±10 V, 20 mA)<br />
System case (aluminium) for measuring instrument, probes and accessories<br />
Probes in lid make it easy to find parts in case<br />
ComSoft 3 - Professional with data management<br />
Incl. database, analysis and graphics function, data analysis, trend curve<br />
Accurate air temp. measurement<br />
testo 400 achieves a system accuracy of<br />
0.1°C and a resolution of 0.01°C when used<br />
together with the precision air probe (Pt100<br />
Class B 1/10).<br />
Accurate air moisture measurement<br />
The task at hand is to monitor exactly the<br />
fluctuations in air moisture with an accuracy<br />
of up to ±1%RH. testo 400 sets new<br />
standards in terms of accuracy and longterm<br />
stability. The worldwide inter-laboratory<br />
test with the patented humidity sensor in<br />
leading, international institutes confirm the<br />
stated values.<br />
Measurement of ideal air supply<br />
testo 400, with its thermal, vane and pitot<br />
tube measurements, has all the technology<br />
available to measure air flow. A calibration<br />
accuracy from 0.5% of the reading is<br />
assured thanks to the first PTB accredited<br />
DKD laboratory for flow.<br />
Function test on lab extractors<br />
Regular testing of the air function can be<br />
carried out as a differential pressure or<br />
speed measurement in the ventilation<br />
connection above the extractor or at the<br />
sliding front opening. Based on DIN 12950<br />
and DIN 12924, we can recommend the hot<br />
wire probe 0635 1041, vane probe Ø 100<br />
mm 0635 9340 and the differential pressure<br />
probes 0638 1345 / 0638 1445.<br />
Stationary transmitters<br />
The check is carried out using the<br />
current/voltage cable (0 to 20 mA, 0 to 1 V, 0<br />
to 10 V) and there is a possibility of<br />
integrating additional parameters.<br />
Part no.<br />
0563 4001<br />
0638 1345<br />
0628 0017<br />
0636 9741<br />
0430 0143<br />
0430 0143<br />
0635 1041<br />
0635 9340<br />
0430 0941<br />
0554 0007<br />
0516 0410<br />
0554 0830<br />
RS232 cable<br />
0409 0178<br />
Connects instrument to PC (1.8 m) for data transfer<br />
We recommend: DKD calibration certificates for temperature, humidity, velocity, pressure<br />
(See Calibration)
<strong>REFERENCE</strong> <strong>MEASURING</strong> <strong>INSTRUMENT</strong><br />
The professional set for comfort level measurements and occupational safety and health<br />
The thermal well-being of humans depends<br />
to a great extent on ambient air flow.<br />
Humans react sensitively to draughts.<br />
Draught air is the most common reason for<br />
complaints about climate.<br />
Testo´s direction-independent comfort probe<br />
has been specially designed to analyse these<br />
draughts. When used together with the testo<br />
400 reference measuring instrument, it is<br />
possible to set up a reading sequence and to<br />
calculate the corresponding mean.<br />
However, the mean air flow alone is not<br />
sufficient to assess the effect on people. The<br />
fluctuations in ambient air flow is also<br />
important. The degree of turbulence required<br />
in the respective standards and guidelines is<br />
a measure of this. It is also calculated<br />
automatically by the testo 400 reference<br />
measuring instrument.<br />
Standards recommend that air flow is<br />
measured directly at the workplace at a<br />
height of 0.1 m, 0.6 m and 1.1 m (for seated<br />
persons) or 0.1 m, 1.1 m and 1.7 m (for<br />
standing persons).<br />
Th maximum mean air flow depends on the<br />
air temperature measured and the degree of<br />
turbulence calculated in each case (see<br />
graphics).<br />
Recommended kit:<br />
testo 400, multi-function measuring instrument, incl. battery, Li cell and<br />
calibration protocol<br />
2 channel multi-function measuring instrument<br />
Comfort level probe for measuring degree of turbulence levels, with telescope<br />
and stand. Fulfills DIN 1946 Part 2 or VDI 2080 requirements<br />
Attachable printer (securely attached) with 1 roll of thermal paper and batteries<br />
We recommend:<br />
CO2 probe measures indoor air quality and monitors the workplace. With plugin<br />
head, connection cable 0430 0143 or 0430 0145 required<br />
Cable, 1.5 m long, connects probe with plug-in head to meas. instrument<br />
PUR coating material<br />
Standard indoor air quality probe up to +70°C<br />
Measures all physical parameters in the Mollier diagram<br />
Quick-action surface probe with sprung thermocouple strip, measuring range<br />
short-term to +500°C<br />
Cable, 1.5 m long, connects probe with plug-in head to meas. instrument<br />
PUR coating material<br />
- Draught prevention in the workplace<br />
- Measure ambient air flow in airconditioned<br />
rooms in accordance with DIN<br />
1946 Part 2, ISO 7726<br />
- Automatic calculation of degree of<br />
turbulence (with testo 400)<br />
- Fulfills the accuracy specifications of DIN<br />
1946 Part 2, VDI 2080, ISO 7726<br />
Part no.<br />
0563 4001<br />
0628 0009<br />
0554 0570<br />
0632 1240<br />
0430 0143<br />
0636 9740<br />
0604 0194<br />
0430 0143<br />
Mean air flow in m/s<br />
The normally time-consuming degree of turbulence calculation is carried out automatically<br />
by the testo 400 measuring instrument.<br />
Example: air temp.: 24.4°C, turbulence: 10%<br />
Degree of turbulence calculation: maximum mean air flow 0.26 m/s<br />
12.07.00 01:51:02<br />
Turbulence<br />
OFFICE 1<br />
Mean:<br />
1<br />
2<br />
Turbulence:<br />
10%<br />
Info:<br />
0.26<br />
24.4<br />
m/s<br />
°C<br />
testo 400 set up with turbulence degree probe<br />
Air temperature in °F<br />
Degree of turbulence T<br />
Air temperature in °C<br />
Mean air flow in fpm<br />
10.0% turbulence<br />
0.26 m/s<br />
24.4°C<br />
Continue ➔ End<br />
testo 400 printout e.g. on attachable printer testo 400 display with calculated degree of<br />
turbulence, mean air flow and air<br />
temperature<br />
157<br />
°C<br />
%RH<br />
td<br />
g/kg<br />
hPa<br />
m/s<br />
m³/h<br />
ppm<br />
CO<br />
ppm<br />
CO2<br />
rpm<br />
mA<br />
V
156<br />
°C<br />
%RH<br />
td<br />
g/kg<br />
hPa<br />
m/s<br />
m³/h<br />
ppm<br />
CO<br />
ppm<br />
CO2<br />
rpm<br />
mA<br />
V<br />
Measure with testo 400 and WBGT probe<br />
30.09.02 13:42:30<br />
Hall 2<br />
40.0<br />
26.4<br />
Tg<br />
20.5<br />
Tw<br />
°C<br />
°C<br />
WBGT<br />
25.1<br />
Ta °C<br />
WBGTS<br />
Mean<br />
24.9<br />
testo 400 on-site printout e.g. with<br />
attachable printer<br />
°C<br />
°C<br />
The WBGT case for fast assessment of workplaces<br />
Hall 2<br />
40.0<br />
Tg<br />
20.5<br />
Tw<br />
°C<br />
°C<br />
WBGT<br />
25.1<br />
Ta °C<br />
WBGTS<br />
Mean<br />
<strong>REFERENCE</strong> <strong>MEASURING</strong> <strong>INSTRUMENT</strong><br />
26.4<br />
24.9<br />
°C<br />
°C<br />
Calculated WBGT and WBGTS indices are<br />
displayed directly<br />
Wet Bulb Globe Temperature<br />
The professional set for assessing workplaces subjected to heat<br />
The measurement task involves assessing<br />
workplaces, particularly those subjected to<br />
heat radiation:<br />
The WBGT probe is used to determine the<br />
WBGT (Wet Bulb Globe Temperature)<br />
climate index in accordance with DIN 33403<br />
and ISO 7243.<br />
The WBGT index is used to determine the<br />
maximum allowable exposure time at<br />
workplaces subjected to heat.<br />
Possible applications are e.g.:<br />
- Steel industry<br />
- Foundries<br />
- Glass industry<br />
- Furnaces<br />
- Ceramics industry.<br />
Heat radiation causes an increase in<br />
temperature based on:<br />
- the thermal influence of the surroundings<br />
- Work intensity<br />
- Thermal transfer of clothing<br />
- Duration of exposure.<br />
If this burden is too high, there is a risk of a<br />
circulatory collapse, heat cramps or heat<br />
stroke.<br />
Recommended kit:<br />
testo 400, multi-function measuring instrument, incl. battery, Li cell and<br />
calibration protocol<br />
Wet Bulb Globe temperature probe to assess workplaces subjected to heat, in<br />
accordance with ISO 7243 or DIN 33403, incl. WBGT case<br />
Attachable printer (securely attached) including 1 roll of thermal paper and<br />
batteries<br />
Quickly prints readings on location<br />
Three different temperatures are required in<br />
order to be able to calculate WBGT:<br />
- Temperature of a naturally ventilated,<br />
moisturised thermometer (Tw), natural wet<br />
bulb temperature<br />
- Globe temperature (Tg)<br />
- Air temperature (Ta).<br />
The calculation is carried out inside and<br />
outside buildings without solar radiation:<br />
WBGT = 0.7 Tw + 0.3 Tg<br />
Outside buildings with solar radiation:<br />
WBGTS = 0.7 Tw + 0.2 Tg + 0.1 Ta<br />
The testo 400 measuring instrument<br />
calculates indices and shows them in the<br />
display.<br />
The measurement should be the following<br />
for standing work: 1.7/1.1/0.1 m and<br />
1.1/0.6/0.1 m for sedentary work.<br />
Part no.<br />
0563 4001<br />
0635 8888<br />
ID No.<br />
0699 4239/1<br />
0554 0570
<strong>REFERENCE</strong> <strong>MEASURING</strong> <strong>INSTRUMENT</strong><br />
For fast measurements on VAC systems<br />
testo 400 with its VAC module is<br />
currently the only measuring system<br />
worldwide with which a quick and<br />
objective assessment of the<br />
functionality of a VAC system can be<br />
carried out without the need for any<br />
additional calculations. Measurement<br />
stipulations are based on<br />
internationally recognised standards;<br />
VDI 2080 in Germany, the 12599 Draft<br />
EN standard for Europe as well as<br />
Ashrae in the US. These are the first<br />
automated measurement procedures<br />
which fulfill the requirements of these<br />
standards.<br />
The testo 400 measuring instrument is<br />
simply positioned on the ventilation<br />
duct via a magnetic holder which<br />
means that the user always has one<br />
hand free. The location data which<br />
includes location name, measurement<br />
coordinates, duct area, correction<br />
factor etc. are automatically read into<br />
the measuring instrument.<br />
The measurement data saved in testo<br />
400 is transmitted to your PC at the<br />
touch of a button. Time-consuming<br />
writing is eliminated, the calculations<br />
are done automatically by testo 400.<br />
The measurement results are printed<br />
using a VDI or EN standard layout.<br />
Recommended kit:<br />
testo 400, multi-function measuring instrument, incl. battery, Li cell and<br />
calibration protocol<br />
Memory upgrade to 500,000 readings<br />
Upgrades memory capacity (via service)<br />
VAC module upgrade<br />
Volume flow calculation in ducts with error calculation function in instrument<br />
ComSoft 3 - Professional with data management<br />
Incl. database, analysis and graphics function, data analysis, trend curve<br />
VAC module upgrade, PC software, (for ComSoft 3 software)<br />
Printout of standard measurement protocols<br />
RS232 cable<br />
Connects instrument to PC (1.8 m) for data transfer<br />
Bendable vane probe (can be bent by 90°), Ø 100 mm, attachable to handle or<br />
telescopic handle, for measurements on ventilation outlets<br />
Vane/temperature probe, Ø 16 mm, attachable to handle or telescopic handle<br />
Professional telescopic handle for plug-in vane probes, max. 1 m long,<br />
extension on request<br />
Attachable printer (securely attached) with 1 roll of thermal paper and batteries<br />
SoftCase (protects instrument from impact) with carrier strap, magnetic holder<br />
and probe holder<br />
SoftCase for attachable printer (protects printer from dirt/impact)<br />
Protects from impact and falls<br />
System case (aluminium) for measuring instrument, probes and accessories<br />
Probes in lid make it easy to find parts in case<br />
Part no.<br />
0563 4001<br />
0554 9481<br />
0450 4010<br />
0554 0830<br />
0554 4030<br />
0409 0178<br />
0635 9340<br />
0635 9540<br />
0430 0941<br />
0554 0570<br />
0516 0401<br />
0516 0411<br />
0516 0410<br />
We recommend: DKD calibration certificate/Temperature<br />
0520 0201<br />
El. resistance thermometer, el. thermometer; cal. points selectable from -80 to +1000°C<br />
Mean<br />
2.0 +<br />
- 0.8 m/s<br />
14760 +<br />
- 5525 m3/h<br />
Assessment of measurement directly<br />
on location with integrated uncertainty<br />
calculation<br />
1/1 (H 250 / V 500 mm)<br />
x + + +<br />
+ + + +<br />
The coordinates required for the grid<br />
measurement are shown in the<br />
instrument display. The depth<br />
information on the vane telescopic<br />
handle makes the task that much<br />
easier in practice.<br />
The specialist case for fast assessment of ventilation/air conditioning systems<br />
155<br />
°C<br />
%RH<br />
td<br />
g/kg<br />
hPa<br />
m/s<br />
m³/h<br />
ppm<br />
CO<br />
ppm<br />
CO2<br />
rpm<br />
mA<br />
V
154<br />
°C<br />
%RH<br />
td<br />
g/kg<br />
hPa<br />
m/s<br />
m³/h<br />
ppm<br />
CO<br />
ppm<br />
CO2<br />
rpm<br />
mA<br />
V<br />
Pitot tube, stainless steel, 360 mm long, measures<br />
velocity with temperature, for pressure probes 0638<br />
1345/..1445/..1545<br />
Pitot tube, stainless steel, 500 mm long, measures<br />
velocity with temperature, for pressure probes 0638<br />
1345/..1445/..1545<br />
Pitot tube, stainless steel, 1000 mm long, measures<br />
velocity with temperature, for pressure probes 0638<br />
1345/..1445/..1545<br />
Comfort level measurement<br />
3-function probe for simultaneous measurement of<br />
temperature, humidity and velocity. With plug-in head,<br />
0430 0143 connection cable required<br />
Shell anemometer, 3 m cable, for meteorological<br />
wind measurement<br />
270 mm<br />
Ø 21 mm<br />
<strong>REFERENCE</strong> <strong>MEASURING</strong> <strong>INSTRUMENT</strong><br />
360 mm<br />
500 mm<br />
1000 mm<br />
Ø 8 mm<br />
Ø 8 mm<br />
Ø 8 mm<br />
Type K (NiCr-Ni)<br />
Type K (NiCr-Ni)<br />
Type K (NiCr-Ni)<br />
Hot bulb<br />
Testo humid. sensor, cap.<br />
NTC<br />
Hot wire<br />
NTC<br />
Suitable probes at a glance<br />
Pitot tube measurement Illustration Probe type Meas. range Accuracy Part no.<br />
Straight Pitot tubes<br />
Comfort level probe for measuring degree of<br />
turbulence levels, with telescope and stand.<br />
Fulfills DIN 1946 Part 2 or VDI 2080<br />
Wet Bulb Globe temperature probe to assess<br />
workplaces subjected to heat, in accordance with<br />
ISO 7243 or DIN 33403, incl. WBGT case<br />
890 mm<br />
Ø 90 mm<br />
-40... +600 °C 0635 2040<br />
-40... +600 °C 0635 2140<br />
-40... +600 °C 0635 2240<br />
Accessories for pressure probes Part no. Accessories for pressure probes Part no.<br />
Connection hose, silicone, 5m long<br />
Silicone hose connects Pitot tube and pressure probe, 5 m long<br />
Other<br />
0554 0440<br />
Magnetic holder for pressure probes<br />
For pressure probes 0638 1345/..1445/..1545/..1645<br />
0... +10 m/s<br />
0... +100 %RH<br />
-20... +70 °C<br />
0... +5 m/s<br />
0... +50 °C<br />
Accessories for 3-function probe Part no. Accessories for 3-function probe<br />
Adapter for humidity adjustment of 3-function probe 0635 1540<br />
Order with adjustment set<br />
0554 0661<br />
Measuring and application ranges of velocity probes<br />
100<br />
60<br />
40<br />
20<br />
15<br />
10<br />
0<br />
m/s<br />
Thermal<br />
probes<br />
up to +70°C Up to +140°C Up to +350°C<br />
The flow measuring range 0 to 100 m/s can be<br />
divided into three sections:<br />
- Low-speed velocity 0 to 5 m/s<br />
- Mid-speed velocity 5 to 40 m/s<br />
- High-speed velocity 40 to 100 m/s<br />
Ø 150 mm<br />
Vane probes<br />
up to +60°C<br />
Thermal probes are used for<br />
accurate measurements in the range<br />
0 to 5 m/s. Vane probes are ideal for<br />
velocities from 5 to 40 m/s. The<br />
measuring range of the Pitot tube<br />
depends on the differential probe<br />
used. The new 100 Pa probe can<br />
±(0.03 m/s ±5% of mv) (0... +10 m/s)<br />
±2 %RH (+2... +98 %RH)<br />
±0.4 °C (0... +50 °C)<br />
±0.5 °C (-20... 0 °C)<br />
±0.5 °C (+50.1... +70 °C)<br />
±(0.03 m/s ±4% of mv)<br />
(0... +5 m/s)<br />
±0.3 °C (0... +50 °C)<br />
0... +120 °C In accordance with<br />
ISO 7243 or DIN<br />
33403<br />
Cable, 1.5 m long, connects probe with plug-in head to meas. instrument<br />
PUR coating material<br />
Vane/temperature probes<br />
therefore be used for the exact<br />
measurement of the flow speed from<br />
approx. 1 m/s to 12 m/s. The Pitot<br />
tube yields optimum results in the<br />
higher velocity range. An additional<br />
criterion when selecting the correct<br />
velocity probe is the temperature.<br />
Vane +0.7... +30 m/s ±(0.3 m/s ±5% of mv)<br />
(+0.7... +30 m/s)<br />
Pitot tube<br />
to +600°C<br />
Thermal sensors can normally be<br />
used at up to approx. +70°C.<br />
Special design vane probes can be<br />
used to maximum +350°C. Pitot<br />
tubes are used for temperatures<br />
above +350°C.<br />
0554 0225<br />
0635 1540<br />
0628 0009<br />
0635 8888<br />
ID No.<br />
0699 4239/1<br />
Part no.<br />
0430 0143<br />
0635 9045<br />
NiCr-Ni Pitot<br />
tube<br />
-40 to +600°C
<strong>REFERENCE</strong> <strong>MEASURING</strong> <strong>INSTRUMENT</strong><br />
Suitable probes at a glance<br />
Vane probes Illustration Probe type Meas. range Accuracy Part no.<br />
Vane probe, Ø 12 mm, can be attached to handle<br />
or telescopic handle<br />
Vane/temperature probe, Ø 16 mm, attachable to<br />
handle or telescopic handle<br />
Vane/temperature probe, Ø 25 mm, can be<br />
attached to handle or telescopic handle<br />
Bendable vane probe (can be bent by 90°), Ø 60<br />
mm, attachable to handle or telescopic handle, for<br />
measurements on ventilation outlets<br />
Bendable vane probe (can be bent by 90°), Ø 100<br />
mm, attachable to handle or telescopic handle, for<br />
measurements on ventilation outlets<br />
Vane probe, Ø 16 mm, for stationary assembly, 3<br />
m cable (PVC)<br />
High temperature vane probe, Ø 25 mm, with<br />
handle for continuous measurements up to<br />
+350°C<br />
Ø 60 mm<br />
180 mm<br />
180 mm<br />
180 mm<br />
Ø 100 mm<br />
Accessories/Vane probes Part no. Accessories/Vane probes<br />
Professional telescopic handle for plug-in vane probes, max. 1 m long,<br />
extension on request<br />
Extension for telescopic handle, 2 m long<br />
Please also order the 0409 0063 extension cable<br />
Handle for plug-in vane probes<br />
250 mm<br />
Ø 16 mm<br />
560 mm<br />
0430 0941<br />
0430 0942<br />
0430 3545<br />
Ø 12 mm<br />
Ø 16 mm<br />
Ø 25 mm<br />
Vane<br />
+0.6... +20 m/s ±(0.2 m/s ±1% of mv)<br />
Oper. temp.<br />
-30... +140 °C<br />
(+0.6... +20 m/s)<br />
Vane +0.4... +60 m/s ±(0.2 m/s ±1% of mv)<br />
Type K (NiCr-<br />
Ni)<br />
-30... +140 °C<br />
(+0.4... +60 m/s)<br />
Vane +0.4... +40 m/s ±(0.2 m/s ±1% of mv)<br />
Type K (NiCr-<br />
Ni)<br />
-30... +140 °C<br />
(+0.4... +40 m/s)<br />
Vane +0.25... +20 m/s ±(0.1 m/s ±1.5% of mv)<br />
Oper. temp.<br />
0... +60 °C<br />
(+0.25... +20 m/s)<br />
Vane<br />
Vane<br />
Type K (NiCr-<br />
Ni)<br />
+0.1... +20 m/s<br />
Oper. temp.<br />
0... +60 °C<br />
±(0.1 m/s ±1.5% of mv)<br />
(+0.1... +20 m/s)<br />
0635 9443<br />
0635 9540<br />
0635 9640<br />
0635 9440<br />
0635 9340<br />
+0.4... +60 m/s ±(0.2 m/s ±1% of mv)<br />
(+0.4... +60 m/s)<br />
0628 0036<br />
Swan neck, flexible connection between probe and connection part<br />
Magnetic probe holder for vane probes<br />
+0.6... +20 m/s<br />
-40... +350 °C<br />
±(0.3 m/s ±1% of f.v.)<br />
(+0.6... +20 m/s)<br />
0635 6045<br />
0430 0001<br />
0554 0430<br />
Thermal probes Illustration Probe type Meas. range Accuracy Part no.<br />
Robust hot bulb probe, Ø 3 mm, for<br />
measurements in the lower velocity range, 2m<br />
cable (PVC)<br />
Affordable, robust hot bulb probe, Ø 3 mm, for<br />
measurements in the lower velocity range, with<br />
handle<br />
Robust hot bulb probe, Ø 3 mm, with handle and<br />
telescopic handle for measurements in the lower<br />
velocity range<br />
Quick-action hot wire probe, Ø 10 mm, with<br />
telescopic handle, for measurements in the lower<br />
velocity range with direction recognition<br />
Pitot tube measurement Illustration Probe type Meas. range Accuracy Part no.<br />
Differential pressure probes<br />
Precision pressure probe, 100 Pa, measures<br />
differential pressure and velocities (in connection<br />
with Pitot tube)<br />
Pressure probe, 10 hPa, measures differential<br />
pressure and velocities (in connection with Pitot<br />
tube)<br />
Pressure probe, 100 hPa, measures differential<br />
pressure and velocities (in connection with Pitot<br />
tube)<br />
Prandtl Pitot tubes<br />
Pitot tube, 300 mm long, stainless steel,<br />
measures velocity in connection with pressure<br />
probes 0638 1345/..1445/..1545<br />
Pitot tube, 350 mm long, stainless steel,<br />
measures velocity flow in connection with<br />
pressure probes 0638 1345/..1445/..1545<br />
Pitot tube, 500 mm long, stainless steel,<br />
measures velocity in connection with pressure<br />
probes 0638 1345/..1445/..1545<br />
Pitot tube, 1000 mm long, stainless steel,<br />
measures velocity in connection with pressure<br />
probes 0638 1345/..1445/..1545<br />
850 mm<br />
760 mm<br />
150 mm<br />
Ø 3 mm<br />
150 mm<br />
Ø 4 mm<br />
300 mm Ø 4 mm<br />
350 mm Ø 7 mm<br />
500 mm Ø 7 mm<br />
1000 mm Ø 7 mm<br />
Ø 3 mm<br />
Ø 25 mm<br />
Ø 3 mm<br />
Ø 10 mm<br />
Hot bulb<br />
NTC<br />
Hot bulb<br />
NTC<br />
Hot bulb<br />
NTC<br />
Hot wire<br />
NTC<br />
Differential 0... +100 Pa ±(0.3 Pa ±0.5% of mv)<br />
pressure probe<br />
(0... +100 Pa)<br />
Part no.<br />
0638 1345<br />
Differential<br />
pressure probe<br />
0... +10 hPa ±0.03 hPa (0... +10 hPa) 0638 1445<br />
Differential<br />
pressure probe<br />
0... +10 m/s<br />
-20... +70 °C<br />
0... +10 m/s<br />
-20... +70 °C<br />
0... +10 m/s<br />
-20... +70 °C<br />
0... +20 m/s<br />
-20... +70 °C<br />
0... +100 hPa ±0.5% of mv (+20...<br />
+100 hPa)<br />
±0.1 hPa (0... +20 hPa)<br />
0638 1545<br />
Oper. temp. 0635 2245<br />
0... +600 °C<br />
Oper. temp. 0635 2145<br />
0... +600 °C<br />
Oper. temp.<br />
0... +600 °C<br />
Oper. temp.<br />
0... +600 °C<br />
±(0.03 m/s ±5% of mv)<br />
(0... +10 m/s)<br />
±(0.03 m/s ±5% of mv)<br />
(0... +10 m/s)<br />
±(0.03 m/s ±5% of mv)<br />
(0... +10 m/s)<br />
±(0.03 m/s ±4% of mv)<br />
(0... +20 m/s)<br />
0628 0035<br />
0635 1549<br />
0635 1049<br />
0635 1041<br />
0635 2045<br />
0635 2345<br />
153<br />
°C<br />
%RH<br />
td<br />
g/kg<br />
hPa<br />
m/s<br />
m³/h<br />
ppm<br />
CO<br />
ppm<br />
CO2<br />
rpm<br />
mA<br />
V
<strong>REFERENCE</strong> <strong>MEASURING</strong> <strong>INSTRUMENT</strong><br />
Technical data<br />
Probe type<br />
Meas. range<br />
Accuracy<br />
± 1 digit<br />
Resolution<br />
Probe type NTC<br />
Pt100<br />
Meas. range -40... +150 °C<br />
-200... +800 °C<br />
Accuracy ±0.2 °C (-10... +50 °C)<br />
±0.1 °C (-49.9... +99.9 °C)<br />
± 1 digit<br />
±0.4 °C (-40... -10.1 °C)<br />
±0.4 °C (-99.9... -50 °C)<br />
±0.4 °C (+50.1... +150 °C) ±0.4 °C (+100... +199.9 °C)<br />
±1 °C (-200... -100 °C)<br />
±1 °C (+200... +800 °C)<br />
Resolution 0.1 °C (-40... +150 °C)<br />
0.01 °C (-99.9... +300 °C)<br />
0.1 °C (-200... -100 °C)<br />
0.1 °C (+300.1... +800 °C)<br />
Probe type<br />
Meas. range<br />
Accuracy<br />
± 1 digit<br />
Resolution<br />
PC RS232 interface<br />
Oper. temp. 0... +50 °C<br />
Storage temp. -25... +60 °C<br />
Battery type 1.5 V AA<br />
Battery life 18 h<br />
Vane<br />
0... +60 m/s<br />
CO2 probe<br />
Weight 500 g<br />
Warranty 3 years<br />
See Probes data for system<br />
accuracy<br />
0.01 m/s (for Ø 60/100 mm),<br />
0.1 m/s (for remaining<br />
probes)<br />
Probe type Testo humid. sensor, cap.<br />
Meas. range 0... +100 %RH<br />
Accuracy See probe data<br />
± 1 digit<br />
Resolution 0.1 %RH (0... +100 %RH)<br />
0... +1 Vol. % CO2<br />
0... +10000 ppm CO2<br />
See probe data<br />
Thermal<br />
0... +20 m/s<br />
±0.01 m/s (0... +1.99 m/s)<br />
±0.02 m/s (+2... +4.9 m/s)<br />
±0.04 m/s (+5... +20 m/s)<br />
0.01 m/s (0... +20 m/s)<br />
Pressure<br />
0... +2000 hPa<br />
Probe 0638 1345<br />
Probe 0638 1445<br />
Probe 0638 1545<br />
Probe 0638 1645<br />
±0.1% of mv<br />
Probe 0638 1740<br />
Probe 0638 1840<br />
Probe 0638 1940<br />
±0.2% of mv<br />
0.001 hPa (0638 1345 probe)<br />
0.001 hPa (0638 1445 probe)<br />
0.01 hPa (0638 1545 probe)<br />
1 hPa (0638 1645 probe)<br />
0.01 bar (0638 1740 probe)<br />
0.01 bar (0638 1840 probe)<br />
0.01 bar (0638 1940 probe)<br />
CO probe<br />
0... +500 ppm CO<br />
±5% of mv (0... +500 ppm CO)<br />
Memory space in basic version: 128 kB<br />
corresponds to approx. 45,000 readings<br />
Memory space, extended: 1 MB, corresponds<br />
to approx. 500,000 readings<br />
Other features: automatic recognition of all<br />
connected probes<br />
Power supply: Battery/rech. batt., alternatively<br />
8 V mains unit<br />
Battery life in continuous operation with 2<br />
thermocouple probes<br />
aw value<br />
0... +1 aW<br />
See probe data<br />
Type K (NiCr-Ni)<br />
-200... +1370 °C<br />
±0.4 °C (-100... +200 °C)<br />
±1 °C (-200... -100.1 °C)<br />
±1 °C (+200.1... +1370 °C)<br />
0.1 °C (-200... +1370 °C)<br />
Mechanical<br />
+20... +20000 rpm<br />
(+20... +20000 rpm)<br />
1 rpm (+20... +20000 rpm)<br />
Type S (Pt10Rh-Pt)<br />
0... +1760 °C<br />
±1 °C (0... +1760 °C)<br />
1 °C (0... +1760 °C)<br />
Current/voltage measurement<br />
0... +20 mA<br />
±0.04 mA (0... +20 mA)<br />
0.01 mA (0... +20 mA)<br />
Type J (Fe-CuNi)<br />
-200... +1000 °C<br />
±0.4 °C (-150... +150 °C)<br />
±1 °C (-200... -150.1 °C)<br />
±1 °C (+150.1... +1000 °C)<br />
0.1 °C (-200... +1000 °C)<br />
Current/voltage measurement<br />
0... +10 V<br />
±0.01 V (0... +10 V)<br />
0.01 V (0... +10 V)<br />
159<br />
°C<br />
%RH<br />
td<br />
g/kg<br />
hPa<br />
m/s<br />
m³/h<br />
ppm<br />
CO<br />
ppm<br />
CO2<br />
rpm<br />
mA<br />
V