3. At the recovery, the process is repeated. The sonde, when recovered, would be externally wipedclean (not the sensors) and then wrapped again in the wet towel and placed back in the white cooler.The DO data now collected by the sonde will be used as part of the post calibration record. Byhaving this information recorded on the actual data set the QA will always be available for futurereference.4. The other sensors will be post-calibrated when the sonde is returned to the lab. These tests willdetermine what effect (if any) fouling might have had on the data. When the post-calibration is<strong>com</strong>pleted the sonde can be returned to the saturated water bath for a final verification.©<strong>YSI</strong> Inc.28