procedure: several teeth missingSite Preparation of First Implant (cont.)• Place the (2) Drill Guide RP to ∅ 2 mm in the same templatesleeve.• Mount a Drill Stop ∅ 2 on the drill for a safe and accuratedrilling procedure.• Thereafter, use the (2) Guided Twist Drill 2 × (10+) 7–18mm and drill to the desired depth under profuseirrigation. Use the drill guide for guidance.The depth markings on the drills correspond to7, 10, 13,15 and 18 mm implants and shouldbe measured level with the drill guide.SEVERALNote! (10+) indicates that the drills extend anadditional 10 mm.• Shift to (3) Guided Drill Guide RP to ∅ 3.2 mm and drill tothe stop with the (3) Guided Twist Drill 3 × (10+) 7–18 mm.• Mount a Drill Stop ∅ 3.2 on the drill for a safe andaccurate drilling procedure.OPTIONALA (4) Guided Screw Tap RP ∅ 4.0 7–13 mm may be used insituations where the jaw bone is judged dense or locallydense bone. Tap to the desired depth.• Switch handpiece to reverse mode and back screw tap out.20