Inogen G3
Inogen G3
Inogen G3
- No tags were found...
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
Notifications (continued)The <strong>Inogen</strong> One® <strong>G3</strong> monitors various parameters during operation and utilizes anintelligent alarm system to indicate a malfunction of the concentrator. Mathematicalalgorithms and time delays are used to reduce the probability of false alarms while stillensuring proper notification of an alarm condition.If multiple alarm conditions are detected, the highest priority alarm will be displayed.The following notification messages are accompanied by a single, short beep.Message Display & Text<strong>Inogen</strong> One®Shutting DownHH:MMVx.x:Serial NumberCondition/Action/ExplanationOn/Off button has been pressed for two seconds.Concentrator is performing system shut down.Audible Alarm button has been pressed for five seconds.Low Priority AlertsThe following low priority alert messages are accompanied by a double beep and asolid red light.Message Display & TextBattery LowAttach PlugChange O2Column SoonCondition/Action/ExplanationBattery power is low, with less than 10 minutes remaining.Attach external power supply or power down and inserta fully charged battery.Column maintenance is required within 30 days. Press andhold the light button for 3 seconds to clear message fromdisplay. The message will be silenced for 72 hours. Contactyour equipment provider to arrange for service.22