Medium Priority AlertsThe following medium priority alert messages are accompanied by a triple beep,repeated every 25 seconds, and a flashing red light.Message Display & TextNo Breath DetectCheck CannulaOxygen ErrorService NeededO2 Delivery ErrorSystem ErrorSee ManualBattery HOTWarningCondition/Action/ExplanationConcentrator has not detected a breath for 60 seconds.Check that cannula is connected to concentrator, there areno kinks in tubing and cannula is positioned properly inyour nose.Oxygen output concentration has been below 50% for10 minutes. If condition persists, switch to your backupoxygen source and contact your equipment provider toarrange for service.A breath has been recognized, but proper oxygen deliveryhas not been detected.Concentrator is experiencing an error but is capable ofcontinued operation. Remove and re-insert battery and/or check external power supply connections. If conditionpersists, contact your equipment provider.Battery has exceeded temperature limit while concentratoris running on battery power. If possible, move concentratorto a cooler location or power unit with an external powersupply and remove battery. If condition persists, contactyour equipment provider.24