P6701B, P6703B & P6723 O/E Converters Service ... - Tequipment.net
P6701B, P6703B & P6723 O/E Converters Service ... - Tequipment.net
P6701B, P6703B & P6723 O/E Converters Service ... - Tequipment.net
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
<strong>P6723</strong> Performance VerificationUse the following procedure to verify the warranted specification (outputamplitude) of the <strong>P6723</strong> O/E converter. Before beginning these procedures, referto page 15 and photocopy the test record and use it to record the performance testresults. The recommended calibration interval is one year.Equipment RequiredTable 8 lists the equipment required to perform the performance verificationprocedure. The types and quantities of connectors may vary depending on thespecific equipment you use.The instrument should be tested at an ambient temperature between 20 and 30degrees C.Table 8: Test equipmentDescription Minimum requirements Example productOptical power meter withhead and adapters1310 nm digital opticalsignal sourceSampling oscilloscopeAdjustable opticalattenuatorAccuracy > 2.5%, resolution >5 pW, Max power > 1 mW, calibratedfrom 700 nm - 1600 nmHP 8153A with powersensor HP 81530A andHP 81531A52 MHz - 622 MHz Tektronix CTS710/CTS750oscilloscope with Tekprobe IIinterface4 decades, 50 μm core fiber,FC-style connectorsOptical cable FC-FC multimode, 62.5 μm, 2metersTektronix TDS784xJDS 5000L with 62 μmfiber with FC connectors174-2322-00<strong>P6701B</strong>, <strong>P6703B</strong>, and <strong>P6723</strong> <strong>Service</strong> Manual 13